Unique flexibility in the smallest of spaces –

Standard-compliant endoscope and instrument


reprocessing

WD 425: Washer-disinfector for reprocessing endoscopes, accessories, instruments


and other medical devices in accordance with EN ISO 15883-1, -2 and -4
Standard-compliant reprocessing
of endoscopes
Endoscopic examinations and interventions are today indispensable for diagnostics and therapy.
The quality of endoscope reprocessing is particularly important in this context. Effective prevention
of infection transmission from the patient to the endoscopy personnel or to other patients through
contaminated endoscopic instruments is essential.

Infection prevention in gastroenterology High capacity and cleaning performance


The best and safest way of reprocessing endoscopes is, As one of the leading suppliers of system solutions in the
according to the HYGEA survey (Hygiene in Gastroenterol- field of infection prevention, Belimed has developed the
ogy – Endoscopic Reprocessing), manual pre-cleaning cost-effective WD 425 washer-disinfector especially for
immediately after use on the patient. This is also medical practices and hospitals. With simultaneous
emphasized by acknowledged guidelines, bodies such as reprocessing of up to 3 endoscopes, it enables a high
the Robert Koch Institute and various gastroenterological turnover. The endoscope channels are each individually
societies. This should be followed by fully automatic, cleaned, disinfected and dried. With its high cleaning
chemo-thermal reprocessing of the endoscopes, including performance, WD 425 fully meets the requirements for
final rinsing with thermally disinfected water. standard-compliance.

The WD 425 is available as under-bench machine (left) or mounted on a base designed to allow for ergonomic loading height (right).

2
Great to be so flexible.
But in more ways than the choice of color.

High flexibility
One device for everything. Whether for reprocessing flexible or rigid endoscopes, surgical or
MIS instruments, anesthesia materials or equipment – WD 425 convinces by its broad range
of applications.

Flexible application Simple installation


WD 425 is suitable in any situation where both endo- The automated endoscope reprocessor WD 425 is easy
scopes and a wide variety of medical devices have to be and fast to install. Apart from water, RO water, drainage
reprocessed within extremely limited space. The broad and electrical power, no further connections are required.
range of application of WD 425 enables real cost savings. The exhaust-air condenser is integrated as standard.
Both chemo-thermal procedures for endoscopes as well as Optional connection to the building exhaust-air system is
thermal procedures for instruments can be applied. also possible.
WD 425 is suitable for reprocessing all types of commer-
cially available endoscopes from a wide range of
manufacturers.

Procedures using glutaraldehyde or peracetic acid


WD 425 is perfectly equipped for heavy-duty use. Thanks
to high-quality components and parts, various procedures
using peracetic acid or glutaraldehyde can be applied.
Confirmation of compatibility of materials must be
obtained from the endoscope manufacturers prior to
reprocessing.

3-level rack for 3 flexible MIS rack 8 DIN tray rack


endoscopes
Hygienic, safe and reliable
Our top priority is to protect patients and personnel. This we ensure through reliable process quality and
simple operation. Our consistent endeavor to ensure safe and reliable operation of the device is
revealed in every detail.

A new dimension of hygiene Simple, safe and reliable operation


Each channel is individually cleaned and disinfected by the Operation of the endoscope reprocessor WD 425 is simple,
Belimed channel rinsing system which thereby ensures a intuitive and well arranged. It fulfills all criteria for high
hygienically flawless cleaning result. The new chamber user comfort. Just pressing one button is enough to start
design with rounded corners was especially developed to the appropriate program. Important process data such as
prevent cross-contamination and carry-over of rinsing cycle time, program status, temperature and remaining
media. The standard washing chamber is made of cycle time are indicated on the display in clearly readable
high-quality 316L stainless steel. This design, which is form. Remaining cycle time, loading or removal readiness
based on pharmaceutical technology, ensures fast and and error messages are additionally displayed via our
residue-free drainage in every process step. patented status display that is well visible from afar.

Standard-compliant – requirements of current Automatic program selection and economic


standards fully met process cycle
Verified and certified by an independent hygiene The benefit of simple, safe and reliable operation is
institute – WD 425 meets the requirements of all systematically optimized: automatic instead of manual pro-
applicable guidelines – both at the international and at gram selection is optionally available. When automatic
the country-specific level. EN ISO 15883, parts 1, 2 and 4, program selection is activated, the control identifies the
as most important standard, has been fully implemented. respective rack using special sensors and automatically
begins the cleaning process, based on rack identification.

Easy and ergonomic operation: the operating panel is located at the optimal height from an ergonomic point of view.

4
There are many ways to protect life.
Ours are the most effective.

Validated final rinse – recontamination ruled out Consistent data monitoring


Disinfected RO water is used for the final rinse, as The essential program data are monitored by means of the
recommended by the Robert Koch Institute (RKI). This printer that is integrated as standard. For paperless
ensures consistent quality of the final rinse water and documentation and archiving, WD 425 can be connected
effectively prevents recontamination of the endoscopes. up to the Belimed ICS 8535 cycle documentation system.
The process, including indirect re-cooling, takes place in The cycle data can be presented graphically or in form of a
parallel to the ongoing cleaning process. Cycle times can detailed cycle report.
thus be shortened by about 15 minutes.
Fast data recording
A0 value-controlled thermal disinfection The use of barcode scanners enhances the efficiency and
For surgical utensils and instruments, an A0-value of 3000 reliability of data recording – the barcodes on the
is normally accepted. Once this value has been reached, endoscopes can be read in within a matter of seconds. The
the controller terminates the disinfection phase. Approx. data are automatically recorded and documented and can
5 minutes of processing time and 0.75 kWh of electrical thus be clearly retraced.
energy per cycle can thus be saved compared to the
previously customary control by temperature and time.

Flexible individual endoscope-channel connection concept: endoscopes High cleaning performance: The endoscope channels are each
from all manufacturers can be easily and quickly connected up. individually cleaned, disinfected and dried.

5
That’s true greatness.
Often concentrated in the smallest of spaces.

Maximum capacity in the smallest space


Compact design with high turnover: WD 425 is the most powerful and most compact automated washer-
disinfector in its class. Up to 3 endoscopes or 8 DIN instrument trays can be reprocessed simultaneously.

Professional reprocessing – even under conditions Greater efficiency due to pre-loading outside
of limited space the machine
The WD 425 endoscope reprocessor combines high Only a limited number of endoscopes is available in an
technology with an extremely compact design. Only 90 cm endoscopy department, and it is thus all the more essential
wide and measuring 70 cm in depth, the machine features that endoscopes are reprocessed quickly so that they can
an excellent relationship between payload space and room be used again. An important precondition for this is that
space required. the capacity of the washer-disinfector is optimally utilized.
The flexible loading system allows the WD 425 device to
be utilized continuously. There are no downtimes due to
loading. The endoscopes can already be loaded onto the
racks and connected up to the individual channels in
advance outside the machine. Thus, they are immediately
ready for the next reprocessing cycle.

Technical Data
Dimensions
Outer dimensions, under-bench model, without
860 × 900 × 700
cover, H × W × D (mm)
Outer dimensions, with base, free-standing, with
1450 × 900 × 700
cover, H × W × D (mm)
Chamber dimensions, H × W × D (mm) 635 × 580 × 550
Chamber volume (liters) 210
Items to be washed Intake capacity

Flexible endoscopes (pieces) 3


Rigid endoscopes (pieces) 36
MIS instruments (sets) 2–3
Surgical instruments (DIN trays) 4 / 6 / 8
Endoscope accessories on 2 tiers
Anesthesia material (sets) 3

For more technical data, see page 7 The endoscopes can already be loaded in advance outside the machine.
This saves time and costs.

6
Technical Data
Equipment WD 425
Type tested according to EN ISO 15883-1, 2, 4 ■
Medical product certificate CE 0044, VDE, EMV, DVGW, SVGW ■
Front-loading single-door model with door interlock ■
Manually operated full-glass hinged door made of safety glass ■
Wash chamber made of high quality stainless steel 1.4404 (AISI 316L); execution meets pharmaceutical industry design ■
Wash chamber illumination ■
Fast-filling valves for cold, warm and RO water with temperature control during filling (16L/min) ■
«Dynamic Filling» system to reduce media consumption ■
Procedures using glutaraldehyde or peracetic acid ■
Dosing pumps for cleaning agents and disinfectants (standard/additional as an option) 2/1
Flow control for dosing units (standard/additional as an option) 2/1
Empty level control of process chemical containers (standard/additional as an option) 2/1
Washing system designed for 3 flexible endoscopes with 7 channels each ■
Washing system designed for 8 DIN instrument trays on 4 levels ■
Leakage test for 3 endoscopes ■
Foam control system ■
Drainage system with valve and pump ■
Guaranteed drainage of residual liquid to prevent carry-over of rinsing water ■
Electrically heated tank (kW) 8.5
Thermal RO water disinfection with re-cooling function ■
High-performance drying ventilator with HEPA filter H13 (operational capacity in m3/h) approx. 200
Drying system, electrically heated (kW) 3.5
Air flow surveillance to avoid overheating ■
Vapor condenser for exhaust-air cooling ■
Patented process status display with LED technology ■
Foil keypad with 12 direct keys and tactile feedback ■
Selection of programs via coded racks ■
A0 value control function and controller for thermal programs ■
Two-line VFD monitor display ■
Temperature monitoring for control and surveillance ■
Thermal protection against overheating ■
Interfaces: RS 232 for printer and barcode reader, RS 485 for PC cycle documentation ■
Barcode reader for recording goods to be cleaned O
Built in cycle printer integrated into front side of device ■
PC cycle documentation for recording and archiving all process data O
Independent process data monitoring using redundant sensor system ■
Controller-independent process recording of conductivity O
Acoustic signal at program end ■
Validation nozzle for external sensors ■
Device for taking water samples ■
Modem for remote maintenance and diagnosis O
Automatic maintenance reminder ■
Front, side and rear paneling made of AISI 304 stainless steel, smoothed, 4N ■
Stainless steel base (H = 540 mm) with drawer for chemicals O
Machine top cover (W = 900 mm × H = 50 mm) made of stainless steel O
Machine top cover (W = 1200 mm × H = 50 mm) made of stainless steel O
Storage cabinet (W = 300 mm × H = 860 mm) made of stainless steel, for storing the RO water cartridge O
Storage cabinet (W = 300 mm × H = 860 mm) made of stainless steel, for storing process chemicals O
Floor pan with leakage sensor O
Connection data
Cold water 200 – 500 kPa, ¾" nipple
Warm water 200 – 500 kPa, ¾” nipple
RO water 200 – 500 kPa, ¾” nipple
Drain DI 19, with siphon on building side
Electrical connection 400 V 3N AC 50 Hz, 10.5 kW, 16A
Additional connection (not mandatory)
Connection to building exhaust-air system DN 50

■ = Standard, O = Optional
Subject to modification

7
www.belimed.com

AUSTRIA Branch Office HUNGARY UNITED KINGDOM


Belimed GmbH Belimed SAS Belimed Infection Control Kft. Belimed Limited
Grüne Lagune 1 ZAC Saumaty Séon Gyergyó utca 5. Unit 4 Newbuildings Place
8350 Fehring 19 rue Gaston Castel 1026 Budapest Dragons Green Road
Tel. +43 3155 40 699 0 13016 Marseille Tel. +36 1 318 86 97 Shipley
Fax +43 3155 40 699 10 Tel. +33 4 96 15 22 10 Fax +36 1 318 87 75 West Sussex, RH13 8GQ
info@belimed.at Fax +33 4 96 15 22 19 info@belimed.hu Tel. +44 1403 738 811
info.sud@belimed.fr Fax +44 1403 730 830
BELGIUM NETHERLANDS info@belimed.co.uk
NV Belimed SA Branch Office Belimed B.V.
Rue de Clairvaux 8 Belimed SAS Vlambloem 65 USA
1348 Louvain-La-Neuve Parc Espale 3068 JG Rotterdam Belimed, Inc.
Tel. +32 10 42 02 40 1, av. Pierre Pflimlin Tel. +31 10 286 17 50 2284 Clements Ferry Road
Fax +32 10 42 02 49 68390 Sausheim Fax +31 10 456 56 97 Charleston, SC 29492
info@belimed.be Tel. +33 3 89 63 65 40 info@belimed.nl Tel. +1 843 216 7424
Fax +33 3 89 63 65 41 Fax +1 843 216 7707
CHINA info@belimed.fr SLOVENIA info@belimed.us
Belimed Medical Equipment Belimed d.o.o.
(Shanghai) Co., Ltd GERMANY Kosovelova cesta 2 OTHER COUNTRIES
CaiLun Road 780, Belimed Deutschland GmbH 1290 Grosuplje Medical/Medical Lab
5th floor, Room H Edisonstrasse 7a Tel. +386 1 7866 010 Belimed AG
ZhangJiang Hi-Tech Park 84453 Mühldorf am Inn Fax +386 1 7866 011 Dorfstrasse 4
201203 Pudong, Shanghai Tel. +49 8631 9896 0 info@belimed.si 6275 Ballwil
Tel. +86 21 513 709 98 Fax +49 8631 9896 300 SWITZERLAND
Fax +86 21 513 709 96 info@belimed.de SWITZERLAND Tel. +41 41 449 78 88
info@belimed.cn Belimed Sauter AG Fax +41 41 449 78 89
Branch Office West Zelgstrasse 8 info@belimed.ch
FRANCE Belimed Deutschland GmbH 8583 Sulgen
Belimed SAS Emil-Hoffmann-Strasse 27 Tel. 0848 55 88 11 OTHER COUNTRIES
Parc GVIO 50996 Köln Fax +41 71 644 86 06 Pharma/Pharma Lab
330 Allée des Hêtres, Hall E Tel. +49 2236 9642 0 contact@belimed-sauter.ch Belimed Sauter AG
69760 Limonest Fax +49 2236 9642 200 Zelgstrasse 8
Tel. +33 4 37 41 63 03 info.west@belimed.de 8583 Sulgen
Fax +33 4 37 41 63 04 SWITZERLAND
info.pharma@belimed.fr Tel. +41 71 644 85 00
MP 2002.05.1109E Subject to modification

Fax +41 71 644 86 00


info@belimed-sauter.ch

You might also like