Unique Flexibility in The Smallest of Spaces - Standard-Compliant Endoscope and Instrument Reprocessing
Unique Flexibility in The Smallest of Spaces - Standard-Compliant Endoscope and Instrument Reprocessing
Unique Flexibility in The Smallest of Spaces - Standard-Compliant Endoscope and Instrument Reprocessing
The WD 425 is available as under-bench machine (left) or mounted on a base designed to allow for ergonomic loading height (right).
2
Great to be so flexible.
But in more ways than the choice of color.
High flexibility
One device for everything. Whether for reprocessing flexible or rigid endoscopes, surgical or
MIS instruments, anesthesia materials or equipment – WD 425 convinces by its broad range
of applications.
Easy and ergonomic operation: the operating panel is located at the optimal height from an ergonomic point of view.
4
There are many ways to protect life.
Ours are the most effective.
Flexible individual endoscope-channel connection concept: endoscopes High cleaning performance: The endoscope channels are each
from all manufacturers can be easily and quickly connected up. individually cleaned, disinfected and dried.
5
That’s true greatness.
Often concentrated in the smallest of spaces.
Professional reprocessing – even under conditions Greater efficiency due to pre-loading outside
of limited space the machine
The WD 425 endoscope reprocessor combines high Only a limited number of endoscopes is available in an
technology with an extremely compact design. Only 90 cm endoscopy department, and it is thus all the more essential
wide and measuring 70 cm in depth, the machine features that endoscopes are reprocessed quickly so that they can
an excellent relationship between payload space and room be used again. An important precondition for this is that
space required. the capacity of the washer-disinfector is optimally utilized.
The flexible loading system allows the WD 425 device to
be utilized continuously. There are no downtimes due to
loading. The endoscopes can already be loaded onto the
racks and connected up to the individual channels in
advance outside the machine. Thus, they are immediately
ready for the next reprocessing cycle.
Technical Data
Dimensions
Outer dimensions, under-bench model, without
860 × 900 × 700
cover, H × W × D (mm)
Outer dimensions, with base, free-standing, with
1450 × 900 × 700
cover, H × W × D (mm)
Chamber dimensions, H × W × D (mm) 635 × 580 × 550
Chamber volume (liters) 210
Items to be washed Intake capacity
For more technical data, see page 7 The endoscopes can already be loaded in advance outside the machine.
This saves time and costs.
6
Technical Data
Equipment WD 425
Type tested according to EN ISO 15883-1, 2, 4 ■
Medical product certificate CE 0044, VDE, EMV, DVGW, SVGW ■
Front-loading single-door model with door interlock ■
Manually operated full-glass hinged door made of safety glass ■
Wash chamber made of high quality stainless steel 1.4404 (AISI 316L); execution meets pharmaceutical industry design ■
Wash chamber illumination ■
Fast-filling valves for cold, warm and RO water with temperature control during filling (16L/min) ■
«Dynamic Filling» system to reduce media consumption ■
Procedures using glutaraldehyde or peracetic acid ■
Dosing pumps for cleaning agents and disinfectants (standard/additional as an option) 2/1
Flow control for dosing units (standard/additional as an option) 2/1
Empty level control of process chemical containers (standard/additional as an option) 2/1
Washing system designed for 3 flexible endoscopes with 7 channels each ■
Washing system designed for 8 DIN instrument trays on 4 levels ■
Leakage test for 3 endoscopes ■
Foam control system ■
Drainage system with valve and pump ■
Guaranteed drainage of residual liquid to prevent carry-over of rinsing water ■
Electrically heated tank (kW) 8.5
Thermal RO water disinfection with re-cooling function ■
High-performance drying ventilator with HEPA filter H13 (operational capacity in m3/h) approx. 200
Drying system, electrically heated (kW) 3.5
Air flow surveillance to avoid overheating ■
Vapor condenser for exhaust-air cooling ■
Patented process status display with LED technology ■
Foil keypad with 12 direct keys and tactile feedback ■
Selection of programs via coded racks ■
A0 value control function and controller for thermal programs ■
Two-line VFD monitor display ■
Temperature monitoring for control and surveillance ■
Thermal protection against overheating ■
Interfaces: RS 232 for printer and barcode reader, RS 485 for PC cycle documentation ■
Barcode reader for recording goods to be cleaned O
Built in cycle printer integrated into front side of device ■
PC cycle documentation for recording and archiving all process data O
Independent process data monitoring using redundant sensor system ■
Controller-independent process recording of conductivity O
Acoustic signal at program end ■
Validation nozzle for external sensors ■
Device for taking water samples ■
Modem for remote maintenance and diagnosis O
Automatic maintenance reminder ■
Front, side and rear paneling made of AISI 304 stainless steel, smoothed, 4N ■
Stainless steel base (H = 540 mm) with drawer for chemicals O
Machine top cover (W = 900 mm × H = 50 mm) made of stainless steel O
Machine top cover (W = 1200 mm × H = 50 mm) made of stainless steel O
Storage cabinet (W = 300 mm × H = 860 mm) made of stainless steel, for storing the RO water cartridge O
Storage cabinet (W = 300 mm × H = 860 mm) made of stainless steel, for storing process chemicals O
Floor pan with leakage sensor O
Connection data
Cold water 200 – 500 kPa, ¾" nipple
Warm water 200 – 500 kPa, ¾” nipple
RO water 200 – 500 kPa, ¾” nipple
Drain DI 19, with siphon on building side
Electrical connection 400 V 3N AC 50 Hz, 10.5 kW, 16A
Additional connection (not mandatory)
Connection to building exhaust-air system DN 50
■ = Standard, O = Optional