Wall Wash Sample Water Miscibility-Hydrocarbon Procedure
Wall Wash Sample Water Miscibility-Hydrocarbon Procedure
Wall Wash Sample Water Miscibility-Hydrocarbon Procedure
The mechanism of this procedure for wall wash analysis is as follows. The wall
wash solvent for the hydrocarbon test is usually methanol or acetone and both
are completely soluble in water. As organic solvents, they also have the ability to
dissolve / extract hydrocarbon residues from the surface of the cargo tank,
(always bearing in mind the solubility of the hydrocarbon in the wall wash
solvent directly impacts how much of the hydrocarbon will be extracted).
When the wall wash sample is mixed with water, any hydrocarbon contaminant
that is soluble in the wall wash solvent, but insoluble in water, is forced out of
the wall wash solvent and shows up as a hazy / turbid emulsion in the solvent-
DI water solution. If the wall wash sample shows any sign of haziness, it is said
to contain water immiscible (hydrocarbon) contaminants and will fail the test.
Equipment Required
This Document remains the Property of L&I Maritime Services Inc. and L&I Maritime (UK) Ltd. 13 of 27
All Rights Reserved. Version 4. 07/07/14
Tel: +971 (0)9 2220 262 Fax: +971 (0)9 2221 252 Email: asiaops@limaritime.com
Tel: +44 (0) 1909 532 003 Fax: +44 (0) 1909 500 945 Email: operations@limaritime.com
L&I
Procedure
1. For each cargo tank sampled, mix 10ml of wall wash / cargo sample with
90ml of lab grade DI Water in the 100ml measuring cylinder. Insert the
stopper and invert the cylinder two or three times, without shaking.
3. Leave the wall-wash / cargo and reference samples to stand for 20 minutes
before assessing the result.
Reporting
In order to pass the test, the wall wash / cargo samples must be free from any
haziness or turbidity after the 20 minutes standing time, compared to the
reference.
If any hydrocarbons are present, there will be a visual difference between the
test sample(s) and the reference sample. The presence of low levels of
hydrocarbons will typically be identified by the presence of a transparent blue
coloured tinge, compared to the presence of higher levels of hydrocarbons which
will be identified by the presence of an opaque white “milky” emulsion.
Some laboratories shine the beam of a pen-light / torch horizontally through the
measuring cylinder against a black background, which will show up trace levels
of hydrocarbons quite clearly. Strictly speaking, this is not in accordance with
ASTM D1722, but as the test can only be reported as “pass” or “fail” it is difficult
to argue against a rejection on the grounds of subjectivity.
Note. If there is any evidence of foam or bubbles on the top of the liquid
inside the measuring cylinder, this is generally an indication that tank
cleaning chemicals have not been completely rinsed away. This is far more
common in coated cargo tanks compared to stainless steel.
The results of ASTM D 1722 are given as either “Pass” or “Fail” only. There is no
allowance for the degree of failure.
This Document remains the Property of L&I Maritime Services Inc. and L&I Maritime (UK) Ltd. 14 of 27
All Rights Reserved. Version 4. 07/07/14
Tel: +971 (0)9 2220 262 Fax: +971 (0)9 2221 252 Email: asiaops@limaritime.com
Tel: +44 (0) 1909 532 003 Fax: +44 (0) 1909 500 945 Email: operations@limaritime.com
L&I
1 2 3 4 5
. . . .
Pic 1. (Above)
Pic 2 (Right)
Sample 2. 20 FTU on
spectrometer.
1 2
. .
This Document remains the Property of L&I Maritime Services Inc. and L&I Maritime (UK) Ltd. 15 of 27
All Rights Reserved. Version 4. 07/07/14
Tel: +971 (0)9 2220 262 Fax: +971 (0)9 2221 252 Email: asiaops@limaritime.com
Tel: +44 (0) 1909 532 003 Fax: +44 (0) 1909 500 945 Email: operations@limaritime.com