Phyto PAM
Phyto PAM
Phyto PAM
PHYTO-PAM
and
Phyto-Win
Software V 1.45
System Components
and
Principles of Operation
2.130 / 01.99
2. Edition: July 2003
phyto_4e.doc
Printed in Germany
NOTES
I
CONTENTS
I
CONTENTS
II
CONTENTS
III
CHAPTER 1 SAFETY INSTRUCTIONS
1 Safety instructions
1
CHAPTER 1 SAFETY INSTRUCTIONS
2
CHAPTER 2 INTRODUCTION
2 Introduction
The pulse-amplitude-modulation (PAM) measuring principle is
based on selective amplification of a fluorescence signal which is
measured with the help of intense, but very short pulses of measuring
light. In the PHYTO-PAM Phytoplankton Analyzer µsec measuring
light pulses are generated by an array of light-emitting diodes (LED)
featuring 4 different colors: blue (470 nm), green (520 nm), light red
(645 nm) and dark red (665 nm). The differently colored measuring
light pulses are applied alternatingly at a high frequency, such that
quasi-simultaneous information on chlorophyll (Chl) fluorescence
excited at the 4 different wavelengths is obtained. This feature is
very useful for distinguishing algae with different types of light
harvesting pigment antenna. For example, in green algae Chl
fluorescence is much more effectively excited by blue and red light
(470, 645 and 665 nm) than by green light (520 nm). In the case of
cyanobacteria, almost no Chl fluorescence is excited by blue light
(470 nm), while excitation at 645 nm is particularly strong due to
phycocyanin and allophycocyanin absorption. On the other hand, in
diatoms and dinoflagellates excitation by blue (470 nm) and green
(520 nm) is relatively high due to strong absorption by fucoxanthin,
Chl c and carotenoids. While this multi-excitation approach opens
new ways in basic research, it also has considerable potential for
practical applications. Phytoplankton in natural surface waters
displays dynamic heterogeneities, depending on time, location and a
number of natural and man-made environmental factors. The
fluorescence signals measured by the 4-wavelengths excitation
method carry the information to differentiate between the
contributions of the main types of phytoplankton with different
pigment systems. Furthermore, following proper calibration, also the
Chl content of the various types can be estimated. And, last but not
least, being a PAM-Fluorometer, the PHYTO-PAM also offers the
3
CHAPTER 2 INTRODUCTION
4
CHAPTER 2 INTRODUCTION
5
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
4 3
1
6
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
7
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
8
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
9
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
10
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
11
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
1 Additional filter RG 9
2 Special filter combination (consisting of BG 3, R 65 and RG 9)
3 Black anodized aluminum cover
The blue-glass filter BG 3 absorbs scattered measuring light,
while passing most of long-wavelength fluorescence. The dichroic
filter R65 serves the purpose of reflecting scattered excitation light,
thus preventing excitation of fluorescence in the RG9 filter.
Therefore, it is essential that the blue-glass filter and the dichroic
filter are facing towards the perspex light-guide (CUV. SIDE). For
optimal signals, the tube-port of the Optical Unit with the perspex
light guide should be gently pushed into the opening of the filter
housing of the Photomultiplier-Detector until it touches the filter,
pressing it carefully against the wall of the housing. The nylon
screws serve for fixing the position.
12
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
13
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
14
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
15
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
16
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
17
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
18
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
19
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
3.3.1 PHYTO-ED
The PHYTO-ED contains all essential components, which in the
standard System I correspond to the Optical Unit, the Measuring and
Actinic LED-Array-Cone and the Photomultiplier-Detector. It weighs
only 600 g as compared to almost 6 kg of the equivalent components
of the standard version and, hence, is particularly well suited for field
applications.
20
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
• PVC centering ring with o-ring sealing against the inner wall of
the Measuring Head, serving as a guide for the cuvette and as an
adapter for mounting the optional Miniature Stirring Motor
Water-S and the optional Spherical Micro Quantum Sensor US-
SQS/W.
21
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
22
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
23
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
facilitated by forcing gently a thin flat body into the slit (like finger
nail or thin screw driver).
It should be noted that at high photomultiplier gain the paddle of
the WATER-S will cause some increase of noise. This is due to the
fact that some measuring light is reflected from the paddle towards
the photodetector, such that the background signal is approximately
doubled and the electronic noise is correspondingly increased.
Furthermore, there is an increase of sample noise caused by the
movement of cells or cell groups.
24
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
3.4.1 PHYTO-EDF
2
3
7
6
4
8 1
25
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
26
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
27
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
3
9
28
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
• Stand with Base Plate (ST-101) (5) and Dark Box (6), on
which the fiberoptics (2) are mounted. A special Mounting Ring
(7) is provided, which holds the Fiberoptics/perspex-rod-adapter
(3) (see Fig. 4). At its upper side, this Mounting Ring features a
flat adjustment screw on which the perspex-rod-adapter (3)
rests. By moving this screw up/down, the distance between
sample and exit plane of the perspex rod can be adjusted. This
distance determines signal amplitude and actinic light intensity.
Both parameters are also influenced by the interface between
perspex-rod and sample. A drop of water leads to a substantial
increase of both parameters.
The performance of the PHYTO-PAM with respect to sensitivity
and stability is affected by any modulated and non-modulated
background signals. The former is due to non-chlorophyll
29
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
30
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
31
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
32
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
33
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
When the program is for the first time started on a particular PC,
the user is asked which communication port (Com Port) is going to
be used:
One of the Com 1-Com 8 ports
can be selected. The same
query also appears when no
instrument is connected via the
RS 232 interface cable, as the
user may just start the program
for viewing stored data. In this
case, the View mode button has
to be pressed.
Once a Com Port was defined, this information is stored and used
for further program starts, as it is assumed that the same Com-port is
also used in the future. If for some reason the communication with
the selected Com-Port does not work or if the instrument (Power-
and-Control-Unit) is switched off, there is a warning:
34
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
35
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
36
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
37
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
38
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
39
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
40
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
41
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
42
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
43
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
44
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
not specific for Chl (Chl a, Chl b and Chl c), but rather provides
a measure of the concentration of all antenna pigments that
transfer absorbed energy via Chl a to the photosynthetic reaction
centers. This aspect is particularly relevant for assessment of
cyanobacteria, the major light-harvesting antenna of which (the
phycobilisomes) do not contain Chl.
(7) It also has to be considered, that the overall Chl is distributed
between PS I and PS II, and that most of the measured
fluorescence reflects PS II Chl and not PS I Chl. Therefore, any
change in the ratio of Chl (PS I) to Chl (PS II) will affect the
Chl determination via fluorescence, as the Chl/F is changed. For
example, it is known that Chl (PS I)/Chl (PS II) increases in
diatoms with the irradiance level during growth.
At the factory, only a coarse Chl calibration for green algae was
carried out and identical calibration factors for the three main algae
groups were assumed. This is alright for first orienting
measurements, when information on relative Chl concentrations is
essential. For quantitative work, more accurate and algae-specific
calibration is recommended (see 4.9).
The actual Chl determination is very simple, provided the proper
Chl calibration file is selected. It is important that the Measuring
Light Frequency (MF1, MF2, MF4, MF8, MF16, MF32, MF64 or
MF128) used for calibration and for the actual measurements are
matching. The Chl determination is started by pressing the Chl(MF)-
button. Then automatically the deconvoluted fluorescence
amplitudes of the three different groups of phytoplankton are
sampled and the corresponding Chl concentrations are calculated on
the basis of the stored calibration factors. The measured and
calculated values are displayed on the "Algae"-window and also
stored in the Report-file (line starting with cF).
45
CHAPTER 3 COMPONENTS OF THE PHYTO-PAM
46
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
47
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
48
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
49
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
3) Below and at the right hand side of the central output window a
number of elements for system operation and display of
instrument status are located, which are always accessible and do
not change when various output windows are selected.
These elements will be briefly described starting from the lower
left corner and ending at the upper right corner:
Light:
On/Off switches of Measuring Light
(ML) and Actinic Light (AL),
respectively.
50
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
51
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
52
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
53
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
4.2 Channels-window
54
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
To determine Zoff, the cuvette is filled either with pure water or with
the filtrate of a natural water sample. By clicking the Zoff button the
4 background signals are measured and substracted from the original
Ft-signals, such that these are suppressed to zero. It is recommended
to determine Zoff at the same Gain as used for the actual
measurements (see also 3.6.1). There is also the possibility to recall
previously determined Zoff-values from the Report-file using the
Copy Zoff command (see 4.4). If it is foreseeable, that with the same
sample measurements at different Gain-settings will take place, a
number of Zoff measurements at various Gain-settings should be
carried out with the same filtrate. Then at any later time at a given
Gain-setting the corresponding Zoff values can be recalled via the
Copy Zoff command. Please note that the Zoff values also are
adjusted automatically to a changed Gain-setting on the basis of the
known Photomultiplier-Gain characteristic. However, the accurracy
55
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
56
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
57
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
58
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
4.3 Algae-window
The Algae-window shows the deconvoluted fluorescence
information for cyanobacteria, green algae and
diatoms/dinoflagellates. The deconvolution is based on Reference
Excitation Spectra (see 4.6) that were previously measured with the
same instrument. The principle of distinguishing between different
groups of phytoplankton is outlined in sections 3.6.2. and 4.7.
59
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
60
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
61
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
62
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
63
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
4.4 Report-window
In the Report-file all measured data are stored. It can be edited by
the user and exported into other programs. The data stored in the
Report-file are the basis for display and analysis of data in the
VIEW-mode of the PhytoWin-program (see 4.11).
64
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
65
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
66
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
67
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
68
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
69
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
70
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
71
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
72
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
73
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
74
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
75
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
76
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
PAR
ETR =
a ∗ PAR + b ∗ PAR + c
2
with the coefficients a, b and c being fitted for least square deviation.
The fitted curves are displayed as continuous lines superimposed on
the data points of the Yield- and ETR-Light Curves (see Fig. 13).
The original version of the model had to be modified in order to take
account of the fact that some types of phytoplankton (particularly
cyanobacteria) do not show maximal PS II quantum yield at PAR=0,
as assumed by the model, but at significant levels of PAR (ca. 20-40
µmol quanta m-2s-1), corresponding to measuring light frequencies
MF32-MF64) (see also 4.5.1). Hence, there is an initial rise and peak
of Yield, before the usual decline sets in at higher PAR values (see
typical example in the Light Curve of the cyanobacteria in Fig. 13).
This phenomenon is likely to reflect a state 2 - state 1 pigment
change. The fitting routine applied by the PhytoWin program ignores
the data points in the rising part of the Yield Light Curve, including
the peak value.
A speed-button is provided for opening a window listing the L
Curve Fit Parameters for Ch1-Ch4 and for the deconvoluted
types of phytoplankton (Blue, Green and Brown). The same window
can be opened via the Main menu. The parameter α (alpha) reflects
the maximal slope of the ETR Light Curve that, as outlined above, in
phytoplankton samples is not necessarily observed close to PAR=0.
The numerical value of α is equivalent to the maximal Yield
multiplied by a PS II absorptivity term. If, as usually the case,
relative ETR is determined, a value of 0.42 is assumed for this term
(see 4.3.2). For measurement of absolute ETR the optical cross-
section of PS II must be known, the value of which can be entered
under Options/ETR Parameters (see 4.3.2). ETRmax represents
maximal electron transport rate (relative units). Ik corresponds to the
particular PAR-value at the crossing point of the lines defined by the
77
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
Fig. 15 Light Curve Fit Parameters calculated for the Light Curve
recordings displayed in Figs. 12 and 13
The quality of Light Curves for Blue, Green and Brown, just like
the deconvoluted values for fluorescence yield of Blue, Green and
Brown (see section 4.3), strongly depends on proper choice of
Reference Spectra (see 4.7). Actually, in many cases the noise
introduced by the fitting may prevent a satisfactory assessment of
Light Curves for all three types of phytoplankton. This will be
particularly true for a component at low content, when another type
is dominating. The less Reference Spectra are used for fitting, the
lower will be the fitting noise. Hence, if e.g. it is known from
microscopic inspection that no or only few cyanobacteria are present,
the Blue reference spectrum (see 4.7) should be eliminated, to
improve the results for green algae and diatoms.
78
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
4.6 Settings-window
79
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
80
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
81
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
82
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
purpose are not equal. For the sake of higher and more uniform
fluorescence yields, the intensities of the 470 nm and 520 nm beams
generally are increased relative to the 645 nm and 665 nm beams.
Furthermore, due to differences in individual LED intensities, each
instrument features somewhat different relative intensities of the four
excitation beams. Therefore, the References of the same sample
measured with different instruments are similar, but not identical.
They differ substantially between the different types of heads
(standard Optical Unit, PHYTO-ED and PHYTO-EDF). The
difference between a Reference Spectrum and the corresponding
Excitation Spectrum can be described by "transfer-factors" that are
specific for each individual Measuring Head and contained in the so-
called Trans-file of a particular Measuring Head (see below).
83
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
84
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
85
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
86
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
87
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
Ref2-file. The 2 stands for the two References (F and dF) contained
in this file. With PhytoWin software versions issued before July
2003, only the F-Reference could be measured in form of a Ref2-
file.
88
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
89
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
90
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
91
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
92
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
93
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
94
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
The calculation of Act. Chl. for Blue, Green and Brown relies on
information on the relative extent of variable fluorescence, dF,
typically displayed by the given types of phytoplankton. For
example, it is known that many cyanobacteria are characterized by
quite small dF/F values, while green algae usually show rather large
values. However, considerable differences in dF/F are possible
within one group of phytoplankton and even within the same species,
depending on the physiological health and the light adaptation state.
Therefore, the calculated Act. Chl. may differ substantially from the
Chl concentration determined on the basis of the fluorescence yield
reached after dark-adaptation or to adaptation to the Measuring
Light, Chl (MF) (see 4.9.1). The Act. Chl. parameter not only reflects
Chl concentration, but also the effective quantum yield of a
particular type of phytoplankton. If, e.g. a water sample is poisoned
by a herbicide, dF and the calculated Act. Chl. will tend towards
zero, irrespective of the actual Chl concentration. The same
95
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
Just like with the normal mode of operation, also in the Delta
F-mode proper choice of Reference Spectra is essential. It should be
noted, that choice of the Zoff-Reference does not make sense, as the
background signal does not display any variable component.
In practice, the Delta F-mode is most useful for assessment of
relative changes in phytoplankton composition and activity at very
low Chl contents. Ideally, the Reference Spectra for pure cultures of
the major types of phytoplankton (Blue, Green and Brown) should be
known (see 4.7). Also the Chl/F Factors (see 4.9) as well as the dF/F-
values should have been determined. However, in particular with
respect to Chl/F and dF/F, the absolute values may not be of primary
importance, as most of the time relative changes in content and
activity are investigated. If alternative methods for assessment of Chl
distribution and concentration (flow cytometry, HPLC) are available,
96
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
97
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
98
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
99
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
100
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
101
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
102
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
103
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
104
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
105
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
106
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
than when submersed in water (factor 1.5). This has to be taken into
account, if the sensor is recalibrated.
During Light Calibration and as well as during Actinic
Illumination the program recognizes whether the Actinic-LED-
Array-Cone PHYTO-AL is connected or not. Correspondingly, the
newly calibrated Internal PAR list applies either for the instrument
with and without PHYTO-AL. Please note that for recognition of the
PHYTO-AL it is important that the connector is completely pushed
into the AL Array socket and that the threaded ring has to be
fastened.
4.11 VIEW-mode
The PhytoWin software can be used in two different modes, the
MEASURE-mode and the VIEW-mode. While the MEASURE-
mode requires connection of the PC with the turned-on PHYTO-
PAM via the RS 232 interface cable, for the VIEW-mode only the PC
is required. In the MEASURE-mode, all data automatically are
written into the current Report (see 4.4) from where they can be
further saved in form of RPT-files (Save Report function in File-
submenu) and reloaded in the VIEW-mode.
In the VIEW-mode not only the data can be viewed in the form
as originally recorded and saved in the Report-file, but also in
modified form after selection of different Reference files (see 4.7) or
Chl calibration files (see 4.9.1). In this way, older data can be
analysed on the basis of new information on the composition and
properties of a particular sample.
A Report-file selected in the VIEW-mode can be exported in
form of a csv-file (comma separated values). In this form, the data
can be further analysed with a spread-sheet program like Excel.
Actually, when such a program is installed on the PC, a csv-file is
107
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
108
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
109
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
110
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
111
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
112
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
113
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
114
CHAPTER 4 FEATURES OF THE SOFTWARE PHYTOWIN
115
CHAPTER 5 TECHNICAL SPECIFICATIONS
5 Technical Specifications
Environmental
humidity: up to 31 °C ≤ 80%,
linearly decreasing to 50 % at 40 °C
Maximal altitude
During operation: 4000 m
In resting state: 15000 m
Mains voltage
fluctuations: max. ±10 %
Overvoltage category: II
Contamination level: 1
116
CHAPTER 5 TECHNICAL SPECIFICATIONS
Windows-Software PhytoWin
PC Requirement: Pentium 600 MHz processor (minimum);
128 MB RAM (minimum); Windows 98,
Me, 2000 or XP
Features: Seven main windows for data display and
analysis
• Channels: Original, unbiased fluorescence information at 4
different excitation wavelengths
• Algae: Deconvoluted fluorescence information for green algae,
diatoms and cyanobacteria based on previously recorded
reference excitation spectra
117
CHAPTER 5 TECHNICAL SPECIFICATIONS
118
CHAPTER 5 TECHNICAL SPECIFICATIONS
Photomultiplier-Detector PM-101P
Design: Mounted in aluminum housing containing
pulse-signal preamplifier; featuring on/off
push-buttons and special circuitry for
automatic overload switch-off; with light-
tight Filter Box and adapter for mounting
on Optical Unit
Signal detection: Miniature photomultiplier with high red
sensitivity (type H6779-01, Hamamatsu)
Detector filter: Combination of three filters passing wave-
lengths above 710 nm, optimized for low
background signal
Dimensions: 100 mm x 66 mm x 108 mm (L x W x H)
Weight: 490 g (incl. cable, 1.5 m long)
119
CHAPTER 5 TECHNICAL SPECIFICATIONS
5.2.2 Accessories
Actinic LED-Array-Cone PHYTO-AL (strongly recommended)
Design: Array consisting of 37 actinic LEDs
peaking at 655 nm, (max. intensity 2000
µmol quanta m-2s-1 PAR), with light-
guiding perspex cone narrowing beam
down to 13 mm Ø; with short-pass filter
(λ < 695 nm) at cone-exit; mounted in
black-anodized aluminum housing
Dimensions: Ø 59 mm, length 190 mm
Weight: 600 g (incl. cable, 1.5 m long)
120
CHAPTER 5 TECHNICAL SPECIFICATIONS
Windows-Software PhytoWin
see 5.2.1
121
CHAPTER 5 TECHNICAL SPECIFICATIONS
5.3.2 Accessories
Spherical Micro Quantum Sensor US-SQS
see 5.2.2
122
CHAPTER 5 TECHNICAL SPECIFICATIONS
Windows-Software PhytoWin
see 5.2.1
123
CHAPTER 5 TECHNICAL SPECIFICATIONS
124
CHAPTER 6 RECHARGEABLE BATTERY
6 Rechargeable battery
The Phytoplankton Analyzer PHYTO-PAM is equipped with a
rechargeable sealed-lead acid battery.
125
CHAPTER 7 WARRANTY CONDITIONS
7 Warranty conditions
All products supplied by the Heinz Walz GmbH, Germany, are
warranted by Heinz Walz GmbH, Germany to be free from defects in
material and workmanship for one (1) year from the shipping date
(date on invoice).
The warranty is subject to the following conditions:
1. This warranty applies if the defects are called to the attention of
Heinz Walz GmbH, Germany, in writing within one year (1) of
the shipping date of the product.
2. This warranty shall not apply to any defects or damage directly
or indirectly caused by or resulting from the use of unauthorized
replacement parts and/or service performed by unauthorized
personnel.
3. This warranty shall not apply to any product supplied by the
Heinz Walz GmbH, Germany which has been subjected to
misuse, abuse, abnormal use, negligence, alteration or accident.
4. This warranty does not apply to damage caused from improper
packaging during shipment or any natural acts of God.
5. This warranty does not apply to underwater cables, batteries,
fiberoptic cables, lamps, gas filters, thermocouples, fuses or
calibrations.
To obtain warranty service, please follow the instructions below:
1. The Warranty Registration form must be completed and returned
to Heinz Walz GmbH, Germany.
2. The product must be returned to Heinz Walz GmbH, Germany,
within 30 days after Heinz Walz GmbH, Germany has received
written notice of the defect. Postage, insurance, custom duties,
126
CHAPTER 7 WARRANTY CONDITIONS
127