OX100 Series: Zirconia-Sensor Environmental Oxygen Analyzer
OX100 Series: Zirconia-Sensor Environmental Oxygen Analyzer
OX100 Series: Zirconia-Sensor Environmental Oxygen Analyzer
OX100 Series
Zirconia-Sensor
Environmental Oxygen Analyzer
• One Instrument measures 100 ppm to 25% Oxygen
• Simple "One Touch" Calibration
• Self Diagnostics Assure High Reliability
• "One Touch" Sensor Replacement
• Compact Size
• Accessories to Assure Accuracy
• Dedicated Sampling Unit (Separate from Oxygen Analyzer)
(Removes Dust, Mist and Organic Solvents)
• Calibration Kit
Simple On-Site Calibration
(Using 1 liter Gas Cylinder) Bulletin 11M10A01-03E
www.yokogawa.com
Low range 0 to 100 ppm Spans 1000 ppm to 25%
Autoranging in four ranges from 0-100 ppm to 0-25% Autoranging in two ranges, 0-1000 ppm and 0-25%
High and Low Alarms High Alarm
Oxygen ppm
Environment : Alcohol / Rosin / Flux, atmospheric pressure
20mm/hr
1000
Oxygen 100 ppm Gas flow : 30 Nl/min. (N2)
Oxygen ppm
500
Easy maintenance for Sampling unit Fast response
A Difference furnace which causes oxygen to rise to 200 ppm, A
(easy replacement of filter, activated carbon) so can help with welding quality assurance.
0
Pressure meter
Sampling unit (K9424GA) (flow adjustment)
Power supply
1 Gas outlet cock
2
Measurement
gas inlet Sensor (K9424JC, JD)
connector
Filter
Pump
3 Activated charcoal, 4 Flow
membrane filter meter
1000 ppm O2/N2
Measurement Needle valve Quantity 6, one-liter
gas outlet Calibration gas
cylinder (K9424JP)
ALM FAIL 1 9
2 10
3 11
%
1 9
OXYGEN ANALYZER SNSR
2 10
4-20mA
ppm 3 11
4 12
MODEL OX100 HTR ALM
96 SUFFIX
No
-1151J
4
5
12
13
FAIL
5 13 91.6
6 14
PART No K9424AA 7 15
100VAC
SET AIR 1000ppm 4-20mA 8 16 6 14
17
Made in Japan
7 15
8 16
MODE ENT.
17
1~10
23 Panel thickness Weight : Approx 0.6kg
Other instrument
+1
89 0
100 *
Unit : mm
Sensor
OX102 EXTERNAL DIMENSIONS
Convertor Unit : mm
88 103
(212)
96 10 180 22
1
+ 9
% 4-20 5 + ALM 13
NC
2 mA 1-5V
ppm - 6
10
14
100-
3 + - 11 240
96 88 4
RNG 7
NC
FAIL
12
15
VAC 7
H.ALM L.ALM FAIL PASS 7 - 8
NC
16
G
17
SNSR
ALM 4-20mA % ppm
CAL ENT
+1
89 0
100 *
Unit : mm
Sensor
(1190)
81
(150)
Sensor parameter Table
Sensor NO. XXXX
1.015
1.043
1.002
0.990
1.429
-19
-17
-65
-25
⭋8⫾0.5 41
A
B
A
B
A
B
A
B
Gain1
Gain2
Gain3
Gain4
Gain5
50
Temp. 250⬚C Max. Temp. 150⬚C Max. Temp. 70⬚C Max.
Max temp. 70⬚C when plastic nipple
(option code "/J" or P/N L9811LA) is used.
OX102-13 29
Unit:mm
Sampling Unit (K9424GA)
Sample inlet (Rc 1/4) Power line connector
Sample outlet (Rc 1/4)
(240 )
(290 )
(220) Needle valve
Activated
carbon filter
(136 )
(146 )
Sensor
Power line switch
Flowmeter
Connector to joint sensor (ø8)
Unit : mm
Calibration Kit (K9424JA, JB) Standard Gas
( 6 cylinders )
(310)
OX100/OX102 Sensor
(90)
K9424JC (255)
NEEDLE VALVE ASSY
Gas vent tube With Pressure gage and Capillary
Pressure
Approx. 180
gage
Approx. 300 1/8 Rc Piping connection for K9242JD
ø75
Approx. 340
K9424JD
NEEDLE VALVE ASSY
OX102 Basic Specifications
Model OX102
Meas. principle: Zirconium sensor, current-limited operation
Sampling: Natural diffusion
Gas sample: Temperature 0 - 70°C, Humidity 20 - 80% RH (dew free), Pressure 1013 ± 40 hPa
Measured gas: O2 in N2 or mixed gases; can't contain silicone vapor, flammable gas or organic solvent
Mounting: Indoor, panel mounting
Meas. ranges: 0-100ppm 0-1000ppm 0-10000ppm 0-25%
Ranging: 0 - 25%
Range/ 20-100ppm 100-1000ppm 0-10000ppm 1-25%
Repeatability: ±15ppm ±25ppm ±3% of FS (for reading of 1,000-5,000) ±0.25% of FS
Calibration: 1. Atmospheric pressure calibration (between 15 to 25% O2)
2. 100 ppm calibration (between 80 and 120 ppm)
3. 1000 ppm calibration (between 900 and 1100 ppm)
4. 9000 ppm calibration (between 8500 and 9500 ppm)
5. Atmospheric pressure, 20, 100, 500, 1000 and 9000 ppm calibration
Fine tune mode : The measuring value can be adjusted to true analyzed value.
External output: 1. 4-20 mA isolated, fixed range
(0-100, 0-1000, 0-10,000 ppm, 0-25% selectable),or 4-20 mA autoranging.
2. 1-5 V, fixed range (0-100, 0-1000, 0-10,000 ppm, 0-25% selectable)
3., Zone signal output for 4-20 mA autoranging
Contact output: Two outputs, rated 110 V AC, 24 V DC, 1 A
Abnormal Hi or Lo alarm
Self-diagnostics: Sensor resistance monitored (open circuit check)
Hi, Lo alarm settings: Alarm contact output if sensor open circuit. Setting range 1 ppm to 24.9%
Storage environment: Sensor and converter temperature 0 to 70°C (dew free)
Power: 100-240V AC ±10%, 50/60Hz 15VA
Converter size: 96 x 96 x 190 mm
Accessories
Part Specifications
K9346WM Activated charcoal element, five per pack: for sampling unit
K9424JC Gas outlet cock (with flow adjuster): qty. one
K9424JD Gas outlet cock (without flow adjuster): qty. one
K9424JP Standard gas cylinder ( 1000 ppm O2/N2 ) : qty. six
K9424BA Sensor assy. cable 3 m for OX100 only
K9424CA Extension cable, 2 m: for OX100 only
B9877BA Bracket for mounting OX100, two per set.
K9424LA Sensor assy. One per set : for OX102 only
K9424LB Extension cable, 4 m: for OX102 only
K9424LC Extension cable, 9 m: for OX102 only
K9424LD Extension cable, 29 m: for OX102 only
K9424DB Sensor holder (metal) for OX100 or OX102
L9811LA Sensor holder (plastic) for OX100 or OX102
Restrictions
Be sure to read the manual and follow the safety precautions in it.
Can't be used in flammable or exhaust gas, corrosive gas, or organic solvent environment
To use in an environment with organic silicone, flow furnace, or N2 reflow furnace:
(1) Use a sampling unit, Can't use the OX100-1.
(2) The maintenance requirements (filter and activated charcoal) depend on the measured gas.
Don't hesitate to contact us if you have any questions.