What Is Quality Control
What Is Quality Control
What Is Quality Control
The other responsibilities of Quality Control are sampling of Raw & packing
material, Testing of raw material, packing material, In process, Finished
product& Stability batches, Sampling & testing of water, Calibration of
Instruments, Preparation of Specification of Raw, Packing, In process &
Finished products, Preparation of Standard Test procedure of Raw, Packing,
In process & Finished products and reporting of result after analysis &
preparation of COA.
It is the time required for the Tablet / Capsule to break into particles, the
disintegration test is a measure of the time required under a given set of
conditions (Temperature) for a group of tablets/capsules to disintegrate into
particles.
Cycle of shaft holding the tube basket limit is 29-32 cycles per minutes and
distance covered by shaft basket is 50-60 mm and beaker temperature is 35 to
39 º C.
Where,
W1 = Weight of Tablets (Initial / Before Tumbling) &
W2 = Weight of Tablets (After Tumbling or friability)
Tablets with individual weight equal to or less than 650 mg then take the
sample of whole corresponding to as near as 6.5 gram equivalent and tablets
with individual weight more than 650 mg then take sample of 10 whole tablets
to perform friability test. Tablets must be de-dusted prior to and after use.
6. What is Incident :
A laboratory Incident is an event in the laboratory that occurs for two primary
reasons either due to analyst error or instrument error.
7. What is Calibration:
The demonstration that a particular instrument or device produces results
within specified limits by comparison with those produced by a traceable
standard over an appropriate range of measurements.
8. What is Qualification :
The action of proving that any equipment or process work correctly and
consistently and produces the expected result. Qualification is part of, but not
limited to a validation process, i.e. Installation Qualification (IQ), Operation
Qualification(OQ) and Performance Qualification (PQ).
The act of planning, carrying out and recording the results of tests on
equipment to confirm its capabilities and to demonstrate that it will perform
consistently as intended use and against predefined specification.
9. What is Deviation:
After Final approval of change control the changes can be made in documents
and change control can be closed after completion of required action plan
which is mentioned in the Change control form.
Preventive action:
Before start of analysis of the Chromatographic system like HPLC &GC system
suitability has to perform to know that the system is working properly or to
know the performance.
The amount of time it takes for the compound to pass through the column is
the retention time (RT). The relative retention time (RRT) is the comparison
of the RT of one compound to another.
The different Types of HPLC Columns are Normal phase, Reverse Phase, Ion
Exchange and Size Exclusion columns.
The drug is placed within the medium in the vessels after it has reached
sufficient temperature and then the dissolution apparatus is operated. Sample
solutions collected from dissolution testing are commonly analyzed by HPLC
or Ultraviolet–visible spectroscopy.
If the initial sample analysis, known as S1 or stage 1 testing fails to meet the
acceptable value for Q, then additional testing known as stage 2 and 3 testing
is required. S3 testing is performed only if S2 testing fails in Q parameter. If
there is a deviation from the acceptable Q values at S3, then an OOS (Out of
Specification) investigation is generally initiated.
In gas chromatography, the mobile phase is a carrier gas, usually an inert gas
such as helium or an un reactive gas such as nitrogen.
Karl Fischer titration is a classic titration method in chemical analysis that use
s coulometric or volumetric titration to determine trace amounts of water in a
sample. It was invented in 1935 by the German chemist Karl Fischer.
The endpoint of the titration is reached when all the water is consumed and
the process uses an organic base (B), sulphur dioxide, iodine and an alcohol.
Out of Specification (OOS) means the test result that falls outside the
specifications or acceptance criteria which has been specified in the official
monographs or the Blend, In process, Raw material, Packing material,
Stability and finished product specification.
During analysis if any OOS observed then it should be investigated to find out
the root cause and required Corrective & preventive actions shall be taken to
avoid the reoccurrence.
Out of Trend (OOT) means the test result that is within the specification limit
or acceptance criteria as mentioned in the Blend, In process, Raw material,
Packing material, Stability and finished product specification but outside the
trend of previously tested batches.
Suppose X Product has the Specification limit 95 to 105 % & we have tested
many batches of product X and the trend result shows is 98 to 102 %. Suppose
X Product current result is 97.5 % so in this case it is called OOT.
The period of time during which a drug product, if stored correctly, is expected
to comply with the specifications determined by stability studies on a number
of batches of the product. The shelf life is used to establish the expiry date of
each batch.
A 5% change in assay from its initial value, any degradation product exceeding
its acceptance criterion. Failure to meet the acceptance criteria for
appearance, physical attributes. Failure to meet the acceptance criteria for
dissolution.
13. Question 13. In Reverse Phase Hplc, Which Type Of Stationary Phase
Is Used And Give Example?
Answer :
Non polar stationary phase used
Ex: Silica gel C-18
14. Question 14. What Are The Detectors Used In Hplc?
Answer :
UV detector, IR detector, Fluorescence detector, Mass spectroscopy, LC
MS etc.
15. Question 15. How To Calculate Retention Factor In Paper
Chromatography?
Answer :
Rf = Distance travelled by solute/ Distance travelled by solvent.