Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power
Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power
Agilent Technologies 1141A Differential Probe and 1142A Probe Control and Power
For Safety and Regulatory information, see the pages at the end
of this book.
© Copyright Agilent Technologies 2000-2004
All Rights Reserved
This manual contains information for use and service of the differential
probe system, the 1141A Differential Probe and 1142A Probe Control
and Power Module. In this document, the two models will be treated as
a system.
Each of the two instrument models that make up the differential probe
system has a serial number sticker. The sticker for the 1141A
Differential Probe is inside the probe, in the bottom cover. (See
chapter 3 for disassembly procedure.)
The 1141A/1142A probe system allows measurement of small
differential signals in the presence of much larger common-mode
signals. It has the following major features:
• 200 MHz bandwidth
• Variable offset
• dc reject
• ac coupling
2
Contents
Calibration Tests 27
dc Gain Accuracy 28
Bandwidth 34
CMRR Test 36
Calibration Test Record 39
Adjustments 40
Probe Adjustment 40
Adjustment Procedure 42
Attenuator Adapter Adjustment 49
3 Service
Introduction 54
General Characteristics 58
3
Contents
Theory of Operation 59
Differential Probe 60
Control and Power Module 60
Attenuator Adapters 61
Test Board 62
Service Policy 63
Troubleshooting 64
Probe Troubleshooting 64
Probe Control and Power Module Troubleshooting 64
Replaceable Parts 71
Parts List 71
Ordering Information 71
Direct Mail Order System 71
Manufacturers’ Codes 72
Exploded View 73
4
1
5
Operating the Probe
Accessories Supplied
Introduction
This chapter shows you how to connect and operate the 1141A
Differential Probe and 1142A Probe Control and Power Module as a
differential probe system.
Accessories Supplied
The following items are supplied as part of the 1141A/1142A probe
system. Item numbers refer to the numbers in Figure 1-1 on page -7
and Figure 1-2 on page -8. Those without item numbers are supplied
but not shown in figures. See the Replaceable Parts List for parts not
listed below.
1 Differential Probe 1
Carrying case
6
Operating the Probe
Accessories Supplied
Figure 1-1
7
Operating the Probe
Accessories Available
Figure 1-2
Figure 1-3
Accessories Available
The following accessories can be ordered.
• 5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5
8
Operating the Probe
To inspect the probe
Recommended
Agilent
Equipment Required Critical Specifications Model/Part Use
Signal Generator 1 to 200 MHz, 400 m Vrms 8648A P
Function Generator 2.5 to 500 kHz, 600 m Vp-p to 16 Vp-p 33120A A
Oscilloscope >400 MHz bandwidth, 1 mV/div sensitivity 54830B P,A
dc source ± 100 mV to 7 Vdc E3632A P
9
Operating the Probe
Recommended Test Equipment
Recommended
Agilent
Equipment Required Critical Specifications Model/Part Use
Test Board No substitute (supplied accessory) 01141-66504 A
Load BNC Feedthrough, 50 Ω Pasternack P
Enterprises PE6008-
50 or Huber+Suhner
22543742
Cables (2) BNC, 50 Ω 36-inch 10503A P,A
Cable BNC, 50 Ω 9-inch 10502A A
Cable Type N (m) 24-inch 11500B P
Adapter Type N (m) to BNC (f) 1250-0780 A
Adapter Type N (f) to BNC (m) 1250-0077 P
Adapters (2) BNC (f) to dual banana (m) 1251-2277 P
Alignment tool Small flat blade (supplied accessory) 8710-1961/ A
Sprague-Goodman
part number GTT-5G
P=Calibration Tests, A=Adjustments
Power Requirements
The 1141A/1142A probe system (specifically the 1142A) requires a power source of
either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.
CAUTION Before connecting power to this instrument, be sure the line voltage switch on the rear
panel of the instrument is set properly.
10
Operating the Probe
Recommended Test Equipment
WARNING Before connecting this instrument, the protective earth terminal of the instrument must
be connected to the protective conductor of the (Mains) power cord. The Mains plug
must be inserted in a socket outlet provided with a protective each contact. The
protective action must be negated by the use on an extension cord (power cable)
without a protective conductor (grounding). Grounding one conductor of a two-
conductor outlet does not provide an instrument ground.
Procedure
1 Use the power cord to connect the 1142A to the ac mains.
2 Connect the 1141A probe cable power connector to the PROBE connector on
the rear panel of the 1142A power module.
3 Connect the output of the probe to the input of the oscilloscope.
4 Set the input impedance of the oscilloscope to 50 Ω.
If the oscilloscope does not have a selectable 50 Ω input impedance, connect a 50 Ω BNC
feedthrough termination between the probe output and the input of the oscilloscope.
5 If making an initial equipment setup, continue with the initial adjustment in
the following section.
Initial Adjustment
For a given combination of 1141A Differential Probe and 1142A Probe Control and Power
Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need
to make these adjustments only once, before the probe is first used. You can make them
any time to optimize the system. These adjustments do not affect the specifications of
the probe system.
11
Operating the Probe
Recommended Test Equipment
• Offset null zeroes the dc level at the output of the probe. The range of adjustment is
about ±4 mV.
• DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc
component of an input signal. The range of adjustment is about ±0.5%.
Figure 1-5
Equipment Needed
The following equipment is necessary for initial adjustment.
• 5 V power supply
• DVM that can measure 25 µV
• 50 Ω BNC feedthrough terminator
• 01141-66504 test board
Equipment Setup
Use the following procedure to setup the differential probe system for initial adjustment.
CAUTION Do not exceed ±7 V when using the test board for this procedure. If the voltage is too
high, it will cause excessive power dissipation in the 50 Ω termination on the test board.
12
Operating the Probe
Recommended Test Equipment
Adjustment
Warm up the 1141A for 30 minutes before making adjustments.
1 With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for
a minimum reading on the DVM.
The voltage swing of the adjustment is approximately ±4 mV.
2 On the 1142A, under DC Reject, press 5.0 Hz.
3 Read and record the reading on the DVM, _________ mV.
4 Connect the probe to the test board in the position shown below
(signal to + input).
Figure 1-6
Signal to + input
5 After the DVM reading stabilizes, adjust DC Reject Gain to the reading recorded
in step 3.
With a 5 V supply, the voltage swing is approximately ±12.5 mV. With a lower supply, the
voltage swings proportionally less.
13
Operating the Probe
Recommended Test Equipment
Figure 1-7
Probe Tips
Probe tips fit into the receptacles in the probe and are held in place with probe tip caps.
If necessary, you can solder the probe tips into a circuit or wires can be soldered to the
tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.
Note Because of the close tolerances between the probe tip caps and probe tips, it will be
difficult to separate the probe tips and caps once the probe tips have been soldered.
Ground Leads
The circular end of the ground lead fits over the screw on the top side of the probe.
Extension Leads
The extension leads provide a flexible connection between circuitry and the probe.
• To provide a male connection to other circuitry, connect the extension lead over the
probe tips.
• To provide a female connection, remove the probe tip caps and probe tips and connect
the extension leads to the probe.
14
Operating the Probe
Recommended Test Equipment
Note Use extension leads and similar connection accessories carefully. Extension leads
compromise the high-frequency specifications of the probe. CMRR is particularly
sensitive to unbalanced input parameters.
To prevent pickup of stray fields when you use extension lead, either the ones supplied
with the 1141A or others, dress them carefully as follows:
• Connect the leads at right angles to the circuitry under test.
• Keep the leads as parallel as possible before they connect to the probe.
Mini-Grabbers
Mini-grabbers can be attached to the probe or adapter through the extension leads.
1 Remove the probe tip caps and tips.
2 Attach the extension leads to the probe or adapter.
3 Attach the mini-grabbers to the extension leads.
15
Operating the Probe
Recommended Test Equipment
Adapters
There are three adapters for use with the differential probe. Two adapters are
attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe
input.
The adapters are installed on the probe after the probe tip caps and probe tips have been
removed. The adapter fastens to the probe using a thumb wheel located on the underside
of the adapter. the figure below shows a good way to hold the probe while attaching the
adapter.
1 Remove the probe tip caps and probe tips, from the probe.
2 Fit the adapter over the end of the probe and rotate the thumb wheel with your
finger until the adapter fits snugly.
A snug fit is important because the ground is maintained through the thumb wheel screw.
A loosely attached adapter compromises the mechanical and electrical integrity of the
combination.
Figure 1-8
16
Operating the Probe
Recommended Test Equipment
Adapter Combinations
The figure below shows the allowed adapter and probe connections. There are two
specific combinations that should not be used.
• Do not attach the ac adapter between an attenuator adapter and the probe.
An attenuator adapter must be terminated by the input resistance of the probe. The
ac adapter isolates the probe input resistance.
• Do not cascade two attenuator adapters.
The attenuator adapters are designed to be terminated by the 1 MΩ resistance of the
probe. The input resistance of the attenuator adapter is 9 MΩ for the 10x adapter and
10 MΩ for the 100x adapter.
Figure 1-9
17
Operating the Probe
Recommended Test Equipment
Connector Compatibility
The following are general connector characteristics for the probe, adapters, and
accessories.
• The female connectors on the probe, adapters, and other accessories are designed to
mate with 0.030-inch round or 0.0250-inch square pins.
• The probe, adapter, and extension lead pins are 0.030-inch round.
• The strip of circuit connection posts provided as an accessory has 0.025-inch square
pins.
• The mini-grabber has a 0.25-inch square pin.
• The ground connection at the end of the probe and adapters (where the adapters
fasten) accepts an M3 metric screw.
Test Board
The primary use of the test board is to apply test and calibration signals to the input of
the probe or adapters. Specific use of the test board is covered wherever it applies.
Grounding
Grounding is very important when probing circuitry. Improper grounding can increase
the common mode signal level. This reduces the effectiveness of the differential probe.
The mechanical connections at the input of the probe are ground for probe signals. The
screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the
attenuator and ac adapter fasten to this ground through the screw connection and the
ground is carried through each adapter to its front.
Figure 1-10
Probe Grounds
Coupling Functions
There are three methods for blocking or compensating for the dc component of a signal.
Each has specific advantages.
18
Operating the Probe
Recommended Test Equipment
dc Reject
dc Reject is the best method of eliminating the dc component of a signal when dc is not
a factor in the measurement.
The key characteristics are:
• The low-frequency component (from dc to the selected corner frequency) is
automatically nulled by the dc reject circuitry.
• Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is uses.
• There is a selectable low frequency corner with -3dB points at 0.05 Hz, 0.5 Hz,
or 5.0 Hz.
• The voltage reject range is ±20 V with the probe alone, ±200 V with the 10x attenuator,
and ±500 V with the 100x attenuator.
19
Operating the Probe
Recommended Test Equipment
To use dc reject:
1 Remove the ac adapter if it is installed.
2 On the front panel of the 1142A, press Local.
3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or
5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz.
Within the frequency and voltage characteristics noted elsewhere in this manual, low
frequencies are nulled from the input signal.
Offset
Offset is the best method to use when the low-frequency corners associated with dc
reject and the ac adapter interfere with the measurement.
The key characteristics are:
• The user manually null the dc component with the offset adjustment.
• Offset is dc coupled so there is no low frequency roll-off.
• Probe CMRR specifications are not compromised as happens when the ac coupling
adapter is used.
• The voltage offset range is ±20 V with the probe alone, ±200 V with the 10x attenuator,
and ±500 V with the 100x attenuator. (With the 100x attenuator, the offset range is
restricted by the maximum input voltage rating rather than the operating range of
the offset).
To use offset:
1 Remove the ac adapter if it is installed.
2 On the front panel of the 1142A, press Local and Variable offset.
3 Adjust the Coarse and Fine Variable Offset until the signal is displayed on the
screen of the oscilloscope.
ac Adapter
The ac adapter must be used when the dc component of the signal exceeds the operating
range of the dc reject or offset methods. The ac adapter block the dc and low frequency
component of the input by forming a high pass filter with the input impedance of the
probe or adapter.
The key characteristics are:
• The ac adapter safely blocks ±200 Vdc when attached directly to the probe or
±500 Vdc when attached to a 10x or 100x adapter.
• The probe and adapters have different input impedances, so they have different low
frequency corners with the ac adapter. When the ac adapter is directly on the probe
the -3dB corner is 15 Hz. When the ac adapter is on an attenuator the corner is 1.5 Hz.
• The low-frequency CMRR when using the ac adapter is not as good as when using the
probe alone or the probe with a 10x or 100x adapter.
CAUTION If you measure a node having a high dc potential, the blocking capacitors in the ac
adapter will charge to that potential. After making such measurements, discharge the
capacitors by grounding both inputs of the ac adapter. This will prevent damage by a
high voltage discharge into sensitive circuitry when the next measurement is made.
20
Operating the Probe
Recommended Test Equipment
To use ac coupling:
1 Attach the ac coupling adapter to the input of the probe or the input of the
attenuator adapter.
2 On the 1142A, press Local and Zero offset.
Remote operation
For automatic test applications, the coupling and offset functions provided by the 1142A
Probe Control and Power Module can be remotely controlled through a connector on the
rear panel of the module. The connection is through a standard 9-pin female
D-subminiature connector. This style is the same as that used on some personal
computer monitor cables, which provides an economical way to connect the 1142A to
the controller interface on an automatic test system.
The following table gives the connections.
NOTE To minimize dc offset errors and potential noise coupling, electrically isolate all
connections between the Remote Input connector and the controlling system.
21
Operating the Probe
Recommended Test Equipment
Function Select
The easiest way to control the function select lines is contact closures between the lines
and Digital common (pin 2) of the remote input connector. (TTL compatible control
signals can be used; but to avoid problems with ground loops, they must be electrically
isolated.) The following truth table shows the functions provided by the function select
lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.
Variable Offset
The remote variable offset can be used when the dc couple function is remotely selected.
The offset voltage must be referenced to the External offset common (pin 4) of the
remote input connector. It must be electrically isolated from the controlling system. The
following table shows the offset range and remote offset requirements for probe and
adapter combinations.
Cleaning Requirements
If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean
the external surfaces of the instrument with a soft cloth dampened with a mixture of
mild detergent and water. (3) Make sure that the instrument is completely dry before
reconnecting it to a power source.
22
Operating the Probe
Recommended Test Equipment
23
Operating the Probe
Recommended Test Equipment
24
2
25
Calibration Tests and Adjustment
Equipment Required
Introduction
This chapter is divided into two sections. The first section gives
calibration tests and the second adjustment procedures for the 1141A
Differential Probe and 1142A Probe Control and Power Module.
Equipment Required
A complete list of equipment required for the calibration tests and adjustments is listed
in “Recommended Test Equipment” on page 9. Equipment required for individual
procedures is listed at the procedure. Any equipment satisfying the critical specifications
listed may be substituted for the recommended model.
CAUTION The power rating of the 50 Ω termination is 1.0 W. Keep the signal input below 7 Vdc
or rms to avoid degrading the termination.
Once the probe tip caps and probe tips have been removed, the probe can be connected
to the test board in one of three ways, as shown in the figure below. At each position of
the probe, probe inputs are connected to a different combination of signal and ground.
A separate terminal on the test board connects the ground of the probe to the signal
ground.
Figure 2-1
26
Calibration Tests and Adjustment
The Test Board
Calibration Tests
These procedures test the probe’s electrical performance using applicable specifications
given in “Performance Specifications and Characteristics” on page 55 as performance
standards. Specifications applicable to individual tests are noted at the test for reference.
Testing Interval
The calibration testing procedures may be performed for incoming inspection of the
instrument and should be performed periodically thereafter to ensure and maintain peak
performance. The recommended test interval is yearly or every 2,000 hours of operation.
Amount of use, environmental conditions, and the user’s experience concerning need
for testing will contribute to verification requirements.
NOTE Allow the instrument to warm up for at least 30 minutes prior to beginning calibration
tests.
27
Calibration Tests and Adjustment
dc Gain Accuracy
dc Gain Accuracy
This test checks the dc gain accuracy of the differential probe and the dc accuracy of
the differential probe with attenuator and adapters.
Specification: Probe alone, ±2%; with attenuator adapter, ±4%
Equipment Required
Recommended
Equipment Agilent
Required Critical Specifications Model/Part
ac/dc Calibrator 100 mV to 7 V
or E3632A
dc Power Supply 100 mV to 7 V
DVM 0.5% accuracy, 10 µV resolution 3458A or
E34401A
Load BNC Feedthrough, 50 Ω Pasternack
Enterprises
PE6008-50 or
Huber+Suhner
22543742
Cables (2) BNC 50 Ω 10503A
Adapters (2) BNC (f) to dual banana (m) 1251-2277
Test Board No substitute 01141-66514
28
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-2
29
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-3
Signal to - input
Record the result of this calculation in the “Calibration Test Record” on page 39.
To pass this test, the probe gain = 0.98 to 1.02
NOTE Failure of the gain accuracy test can be caused by mis-adjustment of the probe. Perform
the Probe Adjustment procedure in the Adjustments section later in this chapter and
retest.
NOTE If the gain test for the probe fails, the 10x Attenuator Accuracy Test will fail or the
results will be poor. Do not continue until the probe passes the dc gain test.
1 Disconnect the probe from the test board and connect the 10x attenuator
adapter to the probe.
2 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to + input).
30
Calibration Tests and Adjustment
dc Gain Accuracy
Figure 2-4
Signal to + input
Signal to - input
7 Record the Vout2 measurement from the bottom DVM in figure 2-2.
8 Calculate the 10x attenuator gain as
Record the result of this calculation in the “Calibration Test Record” on page 39.
NOTE Failure of the accuracy test for the 10x attenuator can be caused by mis-adjustment of
the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter
Adjustment procedure in the Adjustments section later in this chapter then retest the
attenuator adapter. If if continues to fail, repair is necessary.
31
Calibration Tests and Adjustment
dc Gain Accuracy
NOTE If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator
adapter. Do not continue until the probe passes the gain test.
1 Disconnect the probe/attenuator from the test board. Remove the 10x
attenuator adapter from the probe and connect the 100x attenuator adapter.
2 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to + input).
Figure 2-6
Signal to + input
CAUTION Avoid excessive power dissipation in the termination on the test board. Keep the voltage
input at or below ±7 Vdc.
32
Calibration Tests and Adjustment
dc Gain Accuracy
6 Carefully connect the input of the probe/attenuator to the test board in the
position shown in the figure below (signal to - input).
Figure 2-7
Signal to - input
7 Record the Vout2 measurement from the bottom DVM in figure 2-2.
8 Calculate the 100x attenuator gain as
Record the result of this calculation in the “Calibration Test Record” on page 39
NOTE Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment
of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter
Adjustment procedure in the Adjustments section later in this chapter then retest the
attenuator adapter. If if continues to fail, repair is necessary.
33
Calibration Tests and Adjustment
Bandwidth
Bandwidth
This test checks the high-frequency response of the 1141A Differential Probe. The
bandwidth of the oscilloscope is characterized first so it is not a factor in the
measurement.
Specification (-3dB, dc coupled): dc to 200 MHz
Equipment Required
Procedure
This test depends on the accuracy of the termination on the test board and the
termination in the oscilloscope. Both should be with 1%.
1 With the N cable and N-to-BNC adapter, connect the signal generator to the
oscilloscope channel 1 input.
2 Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms).
3 Set the 1142A front panel switches to Local and Zero offset.
4 On the oscilloscope press AUTOSCALE, then set the following parameters.
34
Calibration Tests and Adjustment
Bandwidth
Signal to + input
NOTE Failure of the bandwidth test can be caused by mis-adjustment of the probe. Perform
the Probe Adjustment procedure in the Adjustments section later in this chapter.
35
Calibration Tests and Adjustment
CMRR Test
CMRR Test
This test checks the CMRR at 1 MHz and 100 MHz.
Specification 3000:1 at 1 MHz, 10:1 at 100 MHz
Equipment Required
Recommended
Equipment Agilent
Required Critical Specifications Model/Part
Oscilloscope 400 MHz bandwidth at 1 mV/div 54830B
Signal Generator 1-100 MHz at = 400 mVrms 8648A
Test Board No substitute 01141-66504
Cable Type N (m) 24-inch 11500B
Adapter Type N (f) to BNC (m) 1250-0077
Procedure
1 Connect the probe power connector to the PROBE connection on the rear of
the 1142A Probe Control and Power Module.
2 Set the 1142A front panel switches to Local and Zero offset.
3 Connect the probe output to the oscilloscope channel 1 input.
4 Connect the input of the probe to the test board in the position shown in the
figure below (signal to + input).
Figure 2-9
Signal to + input
36
Calibration Tests and Adjustment
CMRR Test
37
Calibration Tests and Adjustment
CMRR Test
15 Connect the input of the probe to the test board in the position shown in the
figure below (signal to + input).
Figure 2-11
Signal to + input
16 Set the signal generator for 100 MHz at 0.0 dBm (about 224 mVrms, 632 mVp-p).
17 Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div.
18 After the measurement settles (averaging is complete), note the V P-P reading.
Vp-p (1) = _____________ mV
19 Connect the input of the probe to the test board in the position shown in the
figure below (signal to both inputs).
Figure 2-12
38
Calibration Tests and Adjustment
Calibration Test Record
39
Calibration Tests and Adjustment
Probe Adjustment
Adjustments
This section provides adjustment procedures for the 1141A Differential Probe and
attenuator adapters. There are no service adjustments for the 1142A Probe Control
Module.
Adjustment Interval
None of the adjustment procedures that follow should be considered for a routine
maintenance plan. The differential probe and attenuator adapters should be adjusted
under conditions specified at the beginning of the respective procedures.
NOTE Warm up the instrument for 30 minutes before starting adjustment procedures.
Probe Adjustment
This procedure adjust the high-frequency and low-frequency paths on the 1141A
Differential Probe.
NOTE Do not perform this procedure as a part of routine maintenance. Perform the procedure
only if the probe does not meet specifications or has been repaired.
CAUTION You are going to remove the covers of the probe, so the assembly inside will be exposed
while under power. The PC assembly will be electrically and mechanically vulnerable.
Do these adjustment procedures in an ESD-safe area.
Avoid inadvertent contact between the powered assembly and nearby tools and
equipment.
Avoid mechanical damage by carefully handling the exposed assembly and cables.
40
Calibration Tests and Adjustment
Probe Adjustment
The following equipment is required for this procedure. Procedures are based on the
model or part number recommended.
Equipment Required
Probe Preparation
The probe cover must be removed before adjustment. Drift due to temperature
differences with and without covers is negligible.
1 Remove the probe tip caps and probe tips.
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-
clockwise) and slide it down the cable.
3 Remove the bottom cover.
a At the cable end of the probe, separate the covers about centimeter (1/2 inch).
b Slide the bottom cover toward the cable end of the probe until the locator pins at
the probe input clear the holes. Then, remove the cover.
NOTE Note the position of the ground block at the input end of the probe. The ground block
is held, through the PC assembly, by the grounding screw on the top of the probe. The
ground block must be reinstalled on the PC assembly after the top cover is removed.
Handle the PC assembly by the edges of the PC board.
41
Calibration Tests and Adjustment
Adjustment Procedure
6 As shown in the figure below, use the grounding screw to reinstall the ground
block on the PC assembly.
Figure 2-13
The ground block provides a mechanical and electrical connection when the probe PC
assembly is connected to the test board.
7 Connect the probe power connector to the PROBE connection on the rear of
the 1142A Probe Control and Power Module.
8 Connect the mains power to the 1142A.
9 Set the 1142A front panel switches to Local and Zero offset.
Adjustment Procedure
Unless specified elsewhere, the procedures must be followed in the order given.
The only adjustment which may be done separately is HF COMP, the high-frequency
compensation.
42
Calibration Tests and Adjustment
Adjustment Procedure
3 On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set
up the following parameters.
Signal to + input
43
Calibration Tests and Adjustment
Adjustment Procedure
9 Adjust R9, HF GAIN so the Vp-p (1) measurement is the same as in step 4, within
±0.5%.
Make the adjustment slowly so the oscilloscope display has time to react to signal
averaging. Press CLEAR DISPLAY occasionally to restart averaging, which gives a
quicker indication of changes.
10 Carefully connect the probe to the test board in the position shown in the figure
below (signal to both inputs).
Figure 2-16
44
Calibration Tests and Adjustment
Adjustment Procedure
3 Carefully connect the input of the probe to the test board in the position shown
in the figure below (signal to + input).
Figure 2-17
Signal to + input
4 Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top
(see figure below). Again, adjust slowly and press CLEAR DISPLAY frequently
to restart averaging.
Figure 2-18
45
Calibration Tests and Adjustment
Adjustment Procedure
5 Carefully connect the probe to the test board in position shown in the figure
below (signal to both inputs).
Figure 2-19
46
Calibration Tests and Adjustment
Adjustment Procedure
Signal to + input
R13 Adjustment
The minimum allowable amplitude is 212 mVp-p. Typical values will be between 275 and
325 mVp-p. the probe needs repair if the minimum cannot be reached.
Probe Reassembly
1 Disconnect the probe power cable at the rear panel of the 1142A
2 Remove the probe PC assembly from the test board. Be sure the probe input
connectors remain attached to the probe.
3 Remove the grounding screw and ground block from the PC assembly.
47
Calibration Tests and Adjustment
Adjustment Procedure
a Insert the input connectors first, and seat the cable end of the PC assembly over the
pins at the rear of the cover.
b Position the ground block at the center-front of the PC assembly.
c Insert the grounding screw through the top cover and PC assembly and screw it into
the ground block as shown.
5 Replace the bottom cover.
a Position the cable strain relief and with one hand, hold the cable and top together.
The flange on the strain relief has a notch that fits around a protrusion in the top cover.
CAUTION Note where the two pins at the rear of the top cover enter the holes in the PC assembly.
Position the cable wires away from these two areas. Otherwise, when the bottom cover
is closed, part of it will pinch wires that are laying over these areas.
b Insert the pins at the front of the bottom cover into the holes at the front of the top
cover.
c Close the two covers together and fasten with the probe clamp ring.
48
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
NOTE Attenuator adapters should not be adjusted as part of routine maintenance. The design
necessary to give the high CMRR and high impedance of the attenuators makes
adjustment delicate.
Additionally, because of broadband noise, a spectrum analyzer should be used to verify
adjustment quality.
Each attenuator contains two identical voltage dividers on a ceramic substrate. There
are three adjustments. The low-frequency CMRR adjust the low-frequency balance
between the two voltage dividers. The two high-frequency adjustments are electrically
identical. Each adjusts the high-frequency compensation of one of the voltage dividers.
To meet both pulse response and CMRR characteristics they are adjusted differently.
Briefly, the adjustment procedure is:
a Adjust the positive high-frequency response (+ HF RESP) for the best pulse response
using a 3.5 kHz square wave.
b Adjust the negative high-frequency response (-HF RESP) and Low-frequency CMRR
(LF CMRR) for best CMRR using a 3.5 kHz square wave.
The following equipment is required for this procedure. Procedures are based on the
model or part number recommended.
Equipment Required
49
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
Adjustment Procedure
NOTE The attenuator must be adjusted when installed on the 1141A probe with which it will
be used. The specifications and characteristics will not be met if the attenuator adapter
is adjusted with one differential probe then used with another.
1 Remove the probe pins from the attenuator adapter and differential probe, then
attach the adapter to the probe.
2 Set the 1142A front panel switches to Local and Zero offset.
3 Use the 9-inch BNC cable to connect the function generator to the test board.
The short cable minimized ground-loop voltages.
4 Set up the function generator.
• Square wave
• 3.5 kHz
• 3.0 Vp-p for 10x adapter and 16 Vp-p for a 100x adapter.
5 Use the long BNC cable to connect the Trig Out of the function generator to
the EXT TRIG of the oscilloscope.
6 Set up the oscilloscope, then set the following parameters.
50
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
7 Connect the adapter/probe combination to the test board in the position shown
in the figure below.
Figure 2-23
Signal to + input
8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top.
Use the figure below for adjust locations.
Figure 2-24
Adjustment Locations
51
Calibration Tests and Adjustment
Attenuator Adapter Adjustment
11 Connect the adapter/probe combination to the test board in the position shown
in the figure below (signal to both inputs).
Figure 2-25
12 Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal
on the oscilloscope. Repeat the adjustments until the signal is optimized to a
minimum. Each adjustment should be set to minimize the component of the
signal it affects most. Some high-frequency components of the signal are not
affected by either adjustment.
52
3
Service
53
Service
Introduction
Introduction
This section provides troubleshooting, service, and repair information for the
1141A Differential Probe and 1142A Probe Control and Power Module. The
troubleshooting information is provided to isolate a faulty assembly. When a
faulty assembly has been located, the disassembly/assembly procedures help
direct replacement of the assembly.
WARNING Maintenance should be performed by trained service personnel aware of the hazards
involved (for example, fire and electric shock). When maintenance can be performed
without power applied, the power cord must be removed from the instrument.
54
Service
Performance Specifications and Characteristics
55
Service
Performance Specifications and Characteristics
Figure 3-1
Legend
A. CMRR specification for probe with no input adapters.
B. Low-frequency CMRR specification for probe with the ac adapter.
C. Typical CMRR characteristic for differential probe with no input adapters
D. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
E. Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.
56
Service
Performance Specifications and Characteristics
Figure 3-2
Legend
A. Input voltage limits for probe alone.
B. Input voltage limits for 10x adapter.
C. Input voltage limits for 100x adapter.
57
Service
General Characteristics
General Characteristics
The following characteristics apply to the 1141A Differential Probe with the 1142A Probe
Control and Power Module.
Environmental Conditions
Operating Non-operating
Temperature 0°C to +55 C° (32°F to +131°F) -40°C to +70°C (-40°F to +158°F)
Humidity up to 95% relative humidity (non- up to 90% relative humidity at +65°C
condensing) at +40°C (+104°F) (+149°F)
Altitude up to 4,600 meters (15,000 ft) up to 15,300 meters (50,000 ft)
Vibration Random vibration 5 to 500 Hz, 10 Random vibration 5 to 500 Hz, 10 min.
minutes per axis, 0.3grms. per axis, 2.41 grams. Resonant search
5 to 500 Hz swept sine, 1Octave/min.
sweep rate, (0.75g), 5 min. resonant
dwell at 4 resonances per axis.
Power Voltage: 90 to 132/198 to 264 Vac, 47 to 440 Hz
Requirements Power: 25 VA maximum
Weight Net: approximately 1.8 kg (4.0 lb.)
Shipping: approximately 2.7 kg (6.0 lb.)
Dimensions Refer to the outline drawings below.
Figure 3-3
Mechanical Dimensions
58
Service
Theory of Operation
Theory of Operation
The following discussion covers block-level theory for the 1141A/1142A differential
probe system. Refer to the block diagram below.
The differential probe system consists of two units, the 1141A Differential Probe with
its accessories and the 1142A Probe Control and Power Module. For purposes of the
following discussion, these will be called the probe and the control module respectively.
Figure 3-4
59
Service
Theory of Operation
Differential Probe
The probe contains a two-path differential amplifier with unity gain. It is implemented
on a double-sided surface-mount PC board with the high-frequency path on one side and
the low-frequency path on the other. The two paths are split directly after the differential
input connections.
High-Frequency Path
The positive and negative inputs are ac-coupled at 33 Hz into identical impedance
converters. The HF CMRR adjustment balances the gain at the outputs of the impedance
converters. The impedance feed a semi-custom differential amplifier. An additional
negative input to the differential amplifier brings in the sum of the low-frequency and
feedback signals. The FREQ COMP adjustment provides variable high-frequency peaking
of the differential amp. The output amp provides two signals. A feedback signal is
summed with the low-frequency signals and the output signal is the final output of the
probe. Overall probe gain is set by the HIGH FREQ GAIN adjustment.
The entire signal portion of the high-frequency path is implemented on a hybrid IC.
Support circuitry includes bias for the impedance converters and a bias supply for
current sources on the hybrid.
Low-Frequency Path
The bandwidth of the low-frequency path is approximately 75 KHz. The positive and
negative inputs are dc coupled into identical inverting op-amps with gain of 0.5. They
provide a precision 1 M Ω input impedance for the probe. The LOW FREQ BANDWIDTH
adjustments match the gain and phase of the low-frequency path to that of the feedback
from the probe output. One of the two adjustments is set to match the properties of the
feedback and the other is to match the two low-frequency paths. These adjustments
affect the CMRR quality of the probe. The inputs of the inverting op-amps include
protection for ESD and over-voltage conditions.
The inverting op-amps feed a precision differential amplifier with unity gain and a single-
ended output. The output is fed to the summing amp and to the control module for use
in the dc reject circuit.
Summing Amp
The summing amp combines the feedback signal, the low-frequency signal, and the offset
signal. The LOW FREQ GAIN adjustment matches the gain of the low-frequency path to
the overall gain.
60
Service
Theory of Operation
Offset Functions
There are two offset functions developed in the control module: variable offset and dc
reject. A variable offset voltage with coarse and fine adjustments can be selected by the
front panel controls. The offset level is buffered by U8 and selected by multiplexer U3
as the input to offset amp U7. The output of the offset amp is summed with the low-
frequency signal and feedback which gives dc coupling in the probe. Front panel
screwdriver adjustment Offset Null zeros the dc output from the probe when the dc input
and offset are zero.
For dc Reject, an output from the low-frequency amplifier in the probe (LFSIG) is used
to develop a voltage used to null the dc component of the input signal. LFSIG is an input
to U6, an inverting amplifier and low-pass filter. Multiplexer U3 selects one of three
capacitors to set a roll-off frequency of 0.05, 0.5, or 5 Hz. The output of U6 is selected,
again by U3, as the input to the offset amp. When the output of the offset amp is summed
into the low frequency path, the result is cancellation of the dc component of the input
signal. Front panel screwdriver adjustment DC Reject Gain adjusts the gain of the dc
reject circuit.
Power Supply
The supply provides ±6 V and ±15 V for the probe and analog control circuitry as well
as +5 V for the digital control circuitry.
Attenuator Adapters
The 10x and 100x Attenuator Adapter are similar. A ceramic substrate carries two
attenuators, one for each input polarity. A variable resistor adjusts the low-frequency
balance (LF CMRR) between the two attenuators. The high-frequency adjustments are
the same for each attenuator. Each attenuator is adjusted differently. One attenuator is
adjusted for optimum pulse response and the other for best high-frequency CMRR.
Figure 3-5
61
Service
Theory of Operation
Test Board
The test board is a device for conveniently connecting test signals to the differential
probe. The probe can be connected to the board with the signal to the positive, negative,
or both inputs.
Figure 3-6
62
Service
Service Policy
Service Policy
For parts of the 1141A/1142A probe system that are complex, the service policy is for
assembly-level repair. For parts of the system with simple circuitry, the service policy is
component-level repair.
The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies
include the PC assembly and cable. The PC assembly is an “exchange assembly.” A
repaired and tested assembly is shipped upon receipt of the defective assembly.
The attenuator and ac coupling adapters are shipped as complete assemblies.
The service policy for the 1142A Probe Control and Power Module is component-level
repair.
63
Service
Troubleshooting
Troubleshooting
Use the following paragraphs to assist in troubleshooting problems with the 1141A/1142A
Differential Probe.
Probe Troubleshooting
To troubleshoot the probe:
1 Apply a known signal to the input of the probe.
2 Check for an identical output at the output coax to the cable. This connection
is the one soldered to the PC board. If the probe output cable is not terminated,
or the coax is open, the output signal will be about twice the amplitude of the
input signal.
3 If the signal is incorrect, check the power supply voltages from the 1142A Probe
Control and Power Module. Use the cable diagram on the next page.
4 Troubleshoot the cable with an ohmmeter. Use the cable diagram on the next
page.
64
Service
Troubleshooting
Figure 3-7
65
Service
Removing and Replacing Assemblies
CAUTION Never remove or install any assembly with the instrument power ON. Component
damage can occur.
Differential Probe
Use the following procedure to remove and replace the amplifier PC board in the
differential probe.
CAUTION Handle the differential probe carefully once it has been disassembled. If unsupported,
the weight of the cable can put strain on the PC board.
Disassemble Probe
1 Remove the probe tip caps and probe tips.
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-
clockwise) and slide it down the cable.
3 Remove the bottom cover.
a At the cable end of the probe, separate the covers about one centimeter (1/2 inch).
b Slide the bottom cover toward the cable end of the probe until the locator pins at
the probe input clear the holes. Then, remove the cover.
Note the way the cable strain relief is keyed and held at the rear of the top cover.
4 Remove the probe top cover.
The ground screw passes through the top cover and PC board and screws into the ground
block.
a Remove the ground screw on the top of the probe.
At the cable end, the PC board fits over the pins in the top cover.
b Lift the board off of the pins and slide it in the direction of the cable until the input
connectors clear the front of the probe.
5 Un-solder the two connections where the coaxial output cable connects to the
PC board.
6 Disconnect the cable connector from the probe PC board.
66
Service
Removing and Replacing Assemblies
Reassemble Probe
The ground screw passes through the top cover and PC board and screws into the ground
block.
1 If replacing the PC board, remove the input connectors from the old board and
put them on the new one.
2 If replacing the cable, note the orientation of the probe clamp ring on the old
cable, remove the ring and put it on the new cable.
3 Connect the cable connector to the PC board.
4 Solder the two connections of the coaxial cable to the PC board.
On a new probe cable, the conductors of the coaxial cable are connected by a heavy
single wire. Cut the heavy wire so it matches the wire on the cable that was removed.
5 Assemble the PC assembly into the top cover.
The large hybrid is exposed when the assembly is in the top cover. The figure on the
below shows the sequence of the ground screw, top cover, PC board, and ground block.
Figure 3-8
a Insert the input connectors first, and seat the cable end of the PC assembly over the
pins at the rear of the cover.
b Position the ground block at the front of the PC assembly.
c Insert the grounding screw through the top cover and screw it into the ground block
as shown in the figure above.
6 Replace the bottom cover.
a Position the cable strain relief and with one hand, hold the cable and top cover
together.
The flange on the strain relief has a notch that fits over a protrusion in the top cover.
67
Service
Removing and Replacing Assemblies
CAUTION Note where the two pins at the rear of the top cover enter the holes in the PC assembly.
Position the cable wires away from these two areas. When the bottom cover is closed,
part of it will pinch wires that are laying over these areas.
b Insert the pins at the front of the bottom cover into the holes at the front of the top
cover.
c Close the covers together and fasten with the probe clamp ring. If the covers do not
fit together tightly and easily, check for pinched wires (see caution above).
Probe Adapters
Use the following procedure to disassemble the probe adapters. The adapter housing
consists of two plastic parts, one of which slides into the other. The parts are held
together by the spring effect of two plastic tabs on the inner part.
Mechanically, the ac adapter is about the same as the 10x and 100x attenuator adapters.
The attenuator adapters have an extra ground connector which connects the substrate
ground to the thumbwheel screw and plating inside the housing.
Disassemble Adapter
1 Remove the probe tip caps and probe tips from the adapter input.
2 Hold the adapter in one hand taking care not to block the output end of the
adapter (the end which attaches to the probe).
3 Note the view of the input end of the adapter in the figure below. The arrows
indicate the holding tabs.
Figure 3-9
Disassembling Adapters
4 With the thumb and forefinger, squeeze the tabs together, as indicated by the
arrows. Simultaneously, push the tabs into the outer housing so the inner
housing begins to slide out.
5 While holding the outer housing, push back against the thumbwheel until the
inner housing can be grasped and removed.
Reassemble Adapter
Reassembling the adapter is slightly harder because you have to align the connector pins
and thumbwheel screw, while sliding the inner housing and outer housing together.
1 Be sure the input connectors and output pins are present and seated on the
substrate or PC board.
68
Service
Removing and Replacing Assemblies
2 Combine the thumbwheel screw and thumbwheel and insert them into the hole
in the outer housing.
3 Insert the substrate/board into the outer housing. Slip the attenuator ground
(attenuator adapters only) over the thumbwheel screw and seat the input
connectors in the proper holes in the housing.
Figure 3-10
Do not force reassembly of the adapter. The housing halves will slide together with
moderate friction.
4 Align the inner housing tabs with the grooves in the side of the outer housing
and slide the two partly together.
While seating the thumbwheel screw and pins, it will help to hold the assembly vertically,
with the input end of the adapter down. This will allow the assemblies to sit vertically
and more easily align with the holes in the inner housing.
5 Seat the thumbwheel screw first, then the output pins, into the appropriate
holes in the inner housing.
6 Once the two housings are nearly together, press them together firmly until
the tabs click into place.
69
Service
Removing and Replacing Assemblies
WARNING Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely
to the following procedures.
WARNING Be sure to reconnect the safety ground when reassembling the instrument.
3 Unplug the safety ground from the tab on the rear panel of the instrument.
4 Note the orientation of the knobs. Remove the two knobs.
5 On the bottom of the instrument, remove the 5mm screw that fastens the
transformer support.
6 Remove the following pan-head screws.
• Three directly on the PC board.
• Two on the ac input connector.
• Two on the heatsink.
7 Remove the PC board. Slide it slightly forward so parts will clear the rear panel,
then lift the rear of the board out while sliding it backwards.
8 Remove the two heatsink spacers from the standoffs that were directly under
the heatsink.
9 Reverse the procedure to reassemble the control and power supply.
70
Service
Replaceable Parts
Replaceable Parts
This section contains information for ordering parts. Service support for the 1141A
Differential Probe is to the assembly level. Service support for the adapters is as complete
assemblies, except for the probe tip caps and the probe tips. Service support for the
1142A Probe Control and Power Module is to the component level.
Parts List
The replaceable parts lists include all parts relevant to the applicable service levels. The
information given for each part consists of the following:
• Reference designator
• Part number
• Total quantity (Qty) in instrument or on assembly. The total quantity is given once
and at the first appearance of the part number in the list.
• Description of part
• Typical manufacturer of part in a five-digit code.
Ordering Information
To order a part in the material part list, quote the part number, indicate the quantity
desired, and address the order to the nearest Agilent Technologies Sales Office.
To order a part not listed in the material list, include the instrument part number,
instrument serial number, a description of the part (including its function), and the
number of parts required. Address the order to the nearest Agilent Technologies Sales
Office.
71
Service
Replaceable Parts
Mail order forms and specific ordering information are available through your local
Agilent Technologies Sales Office. Addresses and telephone numbers are located in a
separate document shipped with the manuals.
Manufacturers’ Codes
A list of manufacturers’ codes is given the table below. The codes are given for parts in
the parts lists. The table gives the manufacturer and address for each code.
72
Service
Replaceable Parts
Exploded View
Figure 3-11
73
Service
Replaceable Parts
Figure 3-12
74
Service
Replaceable Parts
Figure 3-13
Cabling Diagram
75
Service
Replaceable Parts
Figure 3-14
76
Service
Replaceable Parts
Figure 3-15
77
Service
Replaceable Parts
AC Adapter
01141-26102 2 PROBING PIN 28480 01141-26102
01141-24102 2 CAP-PROBING PIN 28480 01141-24101
78
Service
Replaceable Parts
79
Service
Replaceable Parts
80
Service
Replaceable Parts
81
Service
Replaceable Parts
82
Service
Replaceable Parts
83
Service
Replaceable Parts
84
Service
Replaceable Parts
85
Service
Replaceable Parts
86
Index
Numerics C E
100x attenuator test, 32 cabling diagram, 75 environmental characteristics, 58
10x attenuator test, 30 calibration exchange assemblies, 63
test record, 39 extension leads, 14
A tests, 27
ac adapter, 20 characteristics F
ac coupling adapters, 19 environmental, 58 features, 2
ac low frequency, 55 general, 58 frequency
accessories performance, 55 high compensation, 47
available to order, 8 probe connectors, 18 low response, 44
circuit connector posts, 15 circuit connector posts, 15 function select, 22
extension leads, 14 cleaning instrument, 22
ground leads, 14 CMRR
G
mini grabbers, 15 definition, 23
high frequency gain, 42 gain, 42
probe tips, 14 gain accuracy specification, 55
shielded signal leads, 15 low frequency response, 44
specification, 55 general characteristics, 58
supplied with probe, 6 ground leads, 14
test board, 26 test, 36
common mode rejection ratio, 23 grounding, 18
using, 13
accuracy test, attenuator, 30, 32 compensating dc, 18
component locator H
adapter combinations, 17
adapters, 16 power & control board, 74 high frequency
attenuator, 61 connector compensation, 47
adjustment, 11 characteristics, 18 gain &CMRR adjustment, 42
attenuator adapter, 49 compatibility, 18 high-frequency path, 60
initial, 13 control module, 60 humidity, 58
probe, 40, 42 coupling functions, 18
altitude, 58 I
amplifiers, 23 D
impedance, output, 55
assemblies, removing & replacing, dc gain accuracy, 28 initial adjustment, 11
66 dc mode, 55 input impedance, 55
attenuator dc offset, 19, 55 input range, differential, 55
adapter adjustment, 49 dc reject, 19, 55 input voltage, maximum, 55
adapters, 61 dc reject gain, 11 inspecting, 9
differential amplifiers, 23
B differential input range, 55 L
differential probe
bandwidth
disassembly, 66 line voltage, 10
specification, 55 local control, 61
test, 34 reassembly, 67
dimensions, 58 locator table, 85
blocking dc, 18 low frequency response, 44
block-level theory, 59 direct mail order parts, 71
displayed noises, displayed, 55 low-frequency path, 60
87
Index
88
Notices Warranty commercial license terms, and
non-DOD Departments and
The material contained in this
© Agilent Technologies, Inc. document is provided “as is,” and Agencies of the U.S. Government
2000-2004 is subject to being changed, will receive no greater than
without notice, in future editions. Restricted Rights as defined in
No part of this manual may be FAR 52.227-19(c)(1-2) (June
Further, to the maximum extent
reproduced in any form or by any 1987). U.S. Government users will
permitted by applicable law,
means (including electronic receive no greater than Limited
Agilent disclaims all warranties,
storage and retrieval or Rights as defined in FAR 52.227-
either express or implied, with
translation into a foreign 14 (June 1987) or DFAR 252.227-
regard to this manual and any
language) without prior 7015 (b)(2) (November 1995), as
information contained herein,
agreement and written consent applicable in any technical data.
including but not limited to the
from Agilent Technologies, Inc. as
implied warranties of
governed by United States and Manual Safety Notices
merchantability and fitness for a
international copyright laws.
particular purpose. Agilent shall CAUTION:
not be liable for errors or for
Manual Part Number incidental or consequential A CAUTION notice denotes a
01141-97002, July 2004 damages in connection with the hazard. It calls attention to an
furnishing, use, or performance operating procedure, practice, or
Print History the like that, if not correctly
of this document or of any
01141-97000, June 2000 information contained herein. performed or adhered to, could
01141-97001, September 2002 Should Agilent and the user have result in damage to the product or
01141-97002, July 2004 a separate written agreement loss of important data. Do not
with warranty terms covering the proceed beyond a CAUTION
Agilent Technologies, Inc. notice until the indicated
1900 Garden of the Gods Road material in this document that
conflict with these terms, the conditions are fully understood
Colorado Springs, CO 80907 USA and met.
warranty terms in the separate
agreement shall control. WARNING:
A WARNING notice denotes a
Technology Licenses hazard. It calls attention to an
The hardware and/or software operating procedure, practice, or
described in this document are the like that, if not correctly
furnished under a license and performed or adhered to, could
may be used or copied only in result in personal injury or death.
accordance with the terms of Do not proceed beyond a
such license. WARNING notice until the
indicated conditions are fully
Restricted Rights understood and met.
Legend
If software is for use in the
performance of a U.S.
Government prime contract or
subcontract, Software is delivered
and licensed as “Commercial
computer software” as defined in
DFAR 252.227-7014 (June 1995),
or as a “commercial item” as
defined in FAR 2.101(a) or as
“Restricted computer software”
as defined in FAR 52.227-19
(June 1987) or any equivalent
agency regulation or contract
clause. Use, duplication or
disclosure of Software is subject
to Agilent Technologies’ standard
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Product Safety any unintended operation. Safety Symbols
Service instructions are for
Notices trained service personnel. To !
avoid dangerous electric shock,
This apparatus has been designed do not perform any service unless Instruction manual symbol: the
and tested in accordance with qualified to do so. Do not attempt product is marked with this
IEC Publication 1010, Safety internal service or adjustment symbol when it is necessary for
Requirements for Measuring unless another person, capable of you to refer to the instruction
Apparatus, and has been supplied rendering first aid and manual in order to protect
in a safe condition. This is a resuscitation, is present. against damage to the product.
Safety Class I instrument
(provided with terminal for Do not install substitute parts or
protective earthing). Before perform any unauthorized
applying power, verify that the modification to the instrument.
correct safety precautions are Hazardous voltage symbol.
Capacitors inside the instrument
taken (see the following may retain a charge even if the
warnings). In addition, note the instrument is disconnected from
external markings on the its source of supply.
instrument that are described Earth terminal symbol: Used to
Do not operate the instrument in indicate a circuit common
under "Safety Symbols."
the presence of flammable gasses connected to grounded chassis.
or fumes. Operation of any
Warnings electrical instrument in such an
Before turning on the instrument, environment constitutes a definite
you must connect the protective safety hazard.
earth terminal of the instrument
Do not use the instrument in a
to the protective conductor of the
manner not specified by the
(mains) power cord. The mains
manufacturer.
plug shall only be inserted in a
socket outlet provided with a
protective earth contact. You To clean the instrument
must not negate the protective If the instrument requires
action by using an extension cord cleaning: (1) Remove power from
(power cable) without a the instrument. (2) Clean the
protective conductor external surfaces of the
(grounding). Grounding one instrument with a soft cloth
conductor of a two-conductor dampened with a mixture of mild
outlet is not sufficient protection. detergent and water. (3) Make
sure that the instrument is
Only fuses with the required rated
completely dry before
current, voltage, and specified
reconnecting it to a power source.
type (normal blow, time delay,
etc.) should be used. Do not use
repaired fuses or short-circuited
fuseholders. To do so could cause
a shock or fire hazard.
If you energize this instrument by
an auto transformer (for voltage
reduction or mains isolation), the
common terminal must be
connected to the earth terminal of
the power source.
Whenever it is likely that the
ground protection is impaired,
you must make the instrument
inoperative and secure it against