PXIe-4080 4082
PXIe-4080 4082
PXIe-4080 4082
PXIe-4080/4082
PXIe, 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C
Measurement Support, PXI Digital Multimeter
This document contains the verification and adjustment procedures for the PXIe-4080/4082.
Refer to ni.com/calibration for more information about calibration solutions.
Contents
Required Software.....................................................................................................................2
Related Documentation.............................................................................................................2
Password................................................................................................................................... 2
Calibration Interval................................................................................................................... 2
Test Equipment..........................................................................................................................2
Calibration Procedures.............................................................................................................. 3
Initial Setup............................................................................................................................... 4
Test Conditions..........................................................................................................................4
As-Found and As-Left Limits................................................................................................... 5
Verification................................................................................................................................5
Verifying DC Voltage........................................................................................................5
Verifying AC Voltage........................................................................................................7
Verifying 4-Wire Resistance........................................................................................... 10
Verifying 2-Wire Resistance........................................................................................... 11
Verifying DC Current......................................................................................................14
Verifying AC Current......................................................................................................16
Verifying Capacitance and Inductance............................................................................17
Adjustment.............................................................................................................................. 18
Adjusting DC Voltage..................................................................................................... 18
Adjusting Resistance.......................................................................................................19
Adjusting AC Voltage..................................................................................................... 23
Adjusting Current............................................................................................................25
Adjusting Capacitance and Inductance........................................................................... 26
Completing the Adjustment Procedures......................................................................... 27
Reverification..........................................................................................................................27
Worldwide Support and Services............................................................................................ 27
Required Software
Calibrating the PXIe-4080/4082 requires you to install the following software on the
calibration system:
• NI-DMM version 15.2 or later
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
• NI Digital Multimeters Help
• NI PXIe-4080/4081/4082 Getting Started Guide
• PXIe-4080 Specifications
• PXIe-4082 Specifications
Password
The default password for password-protected operations is NI.
Calibration Interval
Recommended calibration interval 2 years
Test Equipment
The following table lists the equipment required for calibrating your device. If you do not have
the recommended instruments, use these specifications to select a substitute calibration
standard.
Two sets of low Fluke 5440 Voltage, Current Shielded twisted pair
thermal electromotive (1 set), copper cables with copper
force (EMF) copper Resistance (2 or gold-plated copper
cables sets) banana plugs
270 pF, 1 nF, 100 nF, IET SCA series Capacitance ≤ 0.1% uncertainty
10 µF, 1000 µF
capacitors
Calibration Procedures
The calibration process includes the following steps:
• Initial Setup - Set up the test equipment.
• Verification Procedures - Verify the existing operation of the device. This step confirms
whether the device is operating within its specified range prior to calibration.
Initial Setup
Note Ensure that both the calibrator and the PXIe-4080/4082 are warmed up for at
least 60 minutes before you begin this procedure.
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-4080/4082
meets published specifications:
• Ensure that the PXI Express chassis fan speed is set to HIGH and that the fan filters are
clean.
• Use PXI Express filler panels in all vacant slots to allow proper cooling.
• Plug the PXI Express chassis and the calibrator into the same power strip to avoid ground
loops.
• Power on and warm up both the calibrator and the PXIe-4080/4082 for at least 60
minutes before beginning this calibration procedure.
• Maintain an ambient temperature of 23 ±1 °C.
• Maintain an ambient relative humidity of less than 80%.
• Allow the calibrator to settle fully before taking any measurements. Consult the calibrator
user documentation for instructions.
• Allow the thermal EMF enough time to stabilize when you change connections to the
calibrator or the PXIe-4080/4082. The suggested time periods are stated where necessary
throughout this document.
• Keep a shorting bar connected between the V-GUARD and GROUND binding posts of
the calibrator at all times.
• Clean any oxidation from the banana plugs on the cables before plugging them into the
binding posts of the calibrator or the connectors of the PXIe-4080/4082. Oxidation
Verification
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.
You can use the verification procedures described in this section for both pre-adjustment and
post-adjustment verification. You can omit sections (for example, the Verifying AC Current
section), however, any function not verified during post-adjustment will no longer be traceable
and should not be used for measurements.
Note Self-calibrate the PXIe-4080/4082 before performing verification procedures.
Verifying DC Voltage
Complete the following steps to verify DC voltage:
1. Plug in the shorting bar across the HI and LO terminals on the PXIe-4080/4082.
2. Wait two minutes for thermal EMF to stabilize.
3. Reset the PXIe-4080/4082.
4. Initialize the PXIe-4080/4082 for DC Voltage with a range of 100 mV and resolution of
6.5 digits.
5. Verify the accuracy for each configuration in the following table:
a) Acquire the measurement at each specified range and input resistance.
b) Compare to the specified limits.
OUTPUT SENSE
A
NI DMM Calibrator
You have completed verifying the DC voltage mode of the PXIe-4080/4082. Select one of the
following options:
• If you want to continue verifying other modes, go to the Verifying AC Voltage section.
• If you do not want to verify any additional modes, and you are performing a pre-
adjustment verification, close the session.
• If you do not want to verify any additional modes, and you want to update the verification
time stamp, complete the following steps:
1. Open a new calibration session to the instrument (with correct calibration password).
2. Close the calibration session with the Action set to Save.
Verifying AC Voltage
Complete the following steps to verify AC voltage:
1. Set the calibrator to Standby mode.
1 For the ± 100 mV test points, lock the calibrator range to 2.2 V.
5V DC 5V 30 Hz 4.99450 5.00550
AC / DC 5V 50 Hz 4.99650 5.00350
5V 1 kHz 4.99650 5.00350
50 V 5V 1 kHz 4.9875 5.0125
5V 5V 20 kHz 4.99650 5.00350
5V 50 kHz 4.99450 5.00550
5V 100 kHz 4.97400 5.02600
5V 300 kHz 4.84750 5.15250
50 V DC 50 V 30 Hz 49.9450 50.0550
AC /DC 50 V 50 Hz 49.9650 50.0350
50 V 1 kHz 49.9650 50.0350
300 V 50 V 1 kHz 49.915 50.085
50 V 50 V 20 kHz 49.9650 50.0350
50 V 50 kHz 49.9450 50.0550
50 V 100 kHz 49.7400 50.2600
50 V 300 kHz 48.4750 51.5250
300 V DC 219 V 30 Hz 218.751 219.249
AC / DC 300 V 50 Hz 299.790 300.210
219 V 1 kHz 218.831 219.170
219 V 20 kHz 218.831 219.170
219 V 50 kHz 218.743 219.257
219 V 100 kHz 217.845 220.155
70 V 300 kHz 67.750 72.250
You have completed verifying the AC voltage mode of the PXIe-4080/4082. Select one of the
following options:
OUTPUT SENSE
A
NI DMM Calibrator
HI HI
LO LO
HI SENSE HI SENSE
LO SENSE LO SENSE
Not Connected V-GUARD
10 MΩ Off 0Ω On None - 10 - 2
1 MΩ Off 0Ω On None - 10 - 2
100 Off 0Ω On None - 8 - 4
kΩ
10 kΩ On 0Ω On None - 3 - 2.2
1 kΩ On 0Ω On None - 3 - 2.2
100 Ω On 0Ω On None - 13.4 - 11.2
You have completed verifying the 4-wire resistance mode of the PXIe-4080/4082. Select one
of the following options:
• If you want to continue verifying other modes, go to the Verifying 2-Wire Resistance
section.
• If you do not want to verify any additional modes and you are performing a pre-
adjustment verification, close the session.
• If you do not want to verify any additional modes, and you want to update the verification
time stamp, complete the following steps:
1. Open a new calibration session to the instrument (with correct calibration password).
2. Close the calibration session with the Action set to Save.
100 MΩ Off 10 10
10 MΩ Off 10 2
1 MΩ Off 10 2
100 kΩ Off 10 6
10 kΩ On 23 22
1 kΩ On 203 202
100 Ω On 2015 2010
5. Remove the shorting bar from the PXIe-4080/4082.
6. Set the calibrator to Standby.
7. Connect the PXIe-4080/4082 to the calibrator using low thermal EMF cables, as shown in
Figure 1. on page 6.
8. Wait two minutes for thermal EMF to stabilize.
9. Verify the accuracy for each configuration in the following table:
a) Acquire the measurement at each specified configuration.
b) Compare to the specified tolerances.
DMM Settings Calibrator Settings Tolerances
As-Found As-Left
Range OCO Resistance Ext Comp ppm of ppm of ppm of ppm of
Sense Reading Range Reading Range
OUTPUT SENSE
A
NI DMM Calibrator
10 kΩ Off 10 kΩ On 2-Wire 80 23 12 22
1 kΩ Off 1 kΩ On 2-Wire 80 203 12 202
100 Ω Off 100 Ω On 2-Wire 80 2015 15 2010
You have completed verifying the 2-wire resistance mode of the PXIe-4080/4082. Select one
of the following options:
• If you want to continue verifying other modes, go to the Verifying DC Current section.
• If you are performing a pre-adjustment verification, close the session.
• If you do not want to verify any additional modes, and you want to update the verification
time stamp, complete the following steps:
1. Open a new calibration session to the instrument (with correct calibration password).
2. Close the calibration session with the Action set to Save.
Verifying DC Current
Complete the following steps to verify DC current:
1. Unplug all cables from the PXIe-4080/4082.
2. Reset the PXIe-4080/4082.
3. Initialize the PXIe-4080/4082 for DC current with a range of 20 mA and resolution of 6.5
digits.
4. Verify the accuracy for each configuration in the following table:
a) Acquire the measurement at each specified range.
b) Compare to the specified limits.
DMM Range As-Found Limits As-Left Limits
Lower Upper Lower Upper
OUTPUT SENSE
A
NI DMM Calibrator
DMM Range Calibrator Output As-Found Limits (A) As-Left Limits (A)
Lower Upper Lower Upper
You have completed verifying the DC current of the PXIe-4080/4082. Select one of the
following options:
Verifying AC Current
Complete the following steps to verify AC current:
1. Set the calibrator to Standby.
2. Connect the PXIe-4080/4082 to the calibrator using low leakage cables as shown in
Figure 4. on page 15.
3. Reset the PXIe-4080/4082.
4. Initialize the PXIe-4080/4082 for AC Current with a range of 10mA and a resolution of
6.5 digits.
5. Verify the accuracy for each configuration in the following table:
a) Set the PXIe-4080/4082 to the listed configuration and commit.
b) Set the Calibrator to the listed configuration and allow it to settle.
c) Acquire the specified measurement and compare to the limits.
d) Set the calibrator to Standby if you have completed all test points in a DMM
Range.
You have completed verifying AC current for the PXIe-4080/4082. Select one of the following
options:
The following verification procedure assumes the use of verification capacitors with the
following values: 270 pF, 1 nF, 100 nF, 10 µF, and 1000 µF.
The configuration of the cables and fixtures should be consistent throughout each
measurement. If you are using cables to connect the verification capacitors to the inputs of the
PXIe-4082, minimize noise by ensuring that the cables remain fixed and do not move during
the measurement.
Keep direct contact with the verification capacitors to a minimum so that they are constantly
kept at the ambient temperature. After connecting a capacitor to the PXIe-4082, NI
recommends waiting 30 seconds for the capacitor temperature to stabilize.
Note You should know the total capacitance up to the banana connectors that plug
into the PXIe-4082 before performing the verification procedure.
1. Disconnect any fixtures or cables from the device.
2. Reset the PXIe-4082.
3. Verify the accuracy for each configuration in the following table:
a) Connect the indicated standard to the device and wait the recommended settling
time.
b) Configure the device as indicated in DMM Range and take a measurement.
c) Compare the acquired value to the specified tolerances.
d) Repeat the previous steps for each standard listed in the table.
You have completed verifying the capacitance and inductance of the PXIe-4082. Select one of
the following options:
• If you are performing a pre-adjustment verification, close the session.
• To update the verification time stamp, complete the following steps:
1. Open a new calibration session to the instrument (with correct calibration password).
2. Close the calibration session with the Action set to Save.
Adjustment
This section explains how to adjust the PXIe-4080/4082. You can choose to perform these
adjustment procedures with or without performing the verification procedure first.
Note Repeat the verification procedures after you perform these adjustment
procedures. NI recommends that you perform a post-adjustment verification to
ensure that the device you have calibrated is operating within specifications after
adjustments.
Caution Do not skip any of the steps within a section of the adjustment
procedures.
Adjusting DC Voltage
Complete the following steps to adjust DC Voltage:
1. Disconnect all inputs from the PXIe-4080/4082.
2. Reset the calibrator.
10 V DC Volts 10 V >10 GΩ 10 V
-10 V -10 V
8. Disconnect the cables from the PXIe-4080/4082, leaving the other end of the cable
connected to the calibrator binding posts.
9. Plug in the shorting bar across the HI and LO terminals on the PXIe-4080/4082.
10. Wait two minutes for the thermal EMF to stabilize.
11. Call the niDMM Cal Adjust Misc VI or the niDMM_CalAdjustMisc function with
Type set to NIDMM_EXTCAL_MISCCAL_VREF_DC_VOLTAGE.
12. Self-calibrate the PXIe-4080/4082.
13. Call the niDMM Cal Adjust Offset VI or the niDMM_CalAdjustOffset function
with the following parameters:
• Function: DC Volts
• Range: 100 mV
• Input Resistance: >10 GΩ
You have completed adjusting the DC voltage mode of the PXIe-4080/4082. Proceed to one of
the following sections:
• Adjusting Resistance
• Adjusting AC Voltage
• Adjusting Current
If you are not performing additional adjustments, proceed to the Completing the Adjustment
Procedures section to commit the calibration constants.
Adjusting Resistance
Note You must adjust DC voltage before you adjust resistance.
DMM Settings
Function Input Resistance Range
You have completed adjusting the resistance of the PXIe-4080/4082. Proceed to one of the
following sections:
• Adjusting AC Voltage
• Adjusting Current
• Adjusting Capacitance and Inductance
If you are not performing additional adjustments, proceed to the Completing the Adjustment
Procedures section to commit your calibration constants.
Note If you do not use the AC voltage modes for any measurements, or the
accuracy of these modes is irrelevant, you can skip this section in the calibration
procedure and go directly to the Adjusting Current section.
AC Volts 50 mV 50 mV 1 kHz
5 kHz
20 kHz
50 kHz
100 kHz
200 kHz
300 kHz
500 kHz
If you are not performing additional adjustments, proceed to the Completing the Adjustment
Procedures section to commit the calibration constants.
Adjusting Current
Note You must adjust DC voltage before you adjust current.
Note If you do not use the current modes (DC and AC), or the accuracy is
insignificant for your application you can skip this section and go directly to the
Adjusting Capacitance and Inductance section.
Complete the following steps to adjust the current modes of the device:
1. Reset the calibrator.
2. Disconnect all inputs of the PXIe-4080/4082 .
3. Call the niDMM Cal Adjust Misc VI or niDMM_CalAdjustMisc function with Type
set to NIDMM_EXTCAL_MISCCAL_VREF_DC_CURRENT.
4. Repeat the previous step, setting Type to
NIDMM_EXTCAL_MISCCAL_VREF_AC_CURRENT.
5. Connect the PXIe-4080/4082 to the calibrator as shown in Figure 4. on page 15.
DC Current 20 mA N/A 20 mA 20 mA
-20 mA -20 mA
200 mA N/A 200 mA 200 mA
-200 mA -200 mA
1A N/A 1A 1A
-1 A -1 A
7. Self-calibrate the PXIe-4080/4082.
You have completed adjusting the current modes of the PXIe-4080/4082. To continue
adjusting other modes, proceed to the Adjusting Capacitance and Inductance section.
If you are finished adjusting modes, proceed to the Completing the Adjustment Procedures
section to commit the calibration constants.
25 Ω NIDMM_EXTCAL_LC_25OHM
125 Ω NIDMM_EXTCAL_LC_125OHM
5 kΩ NIDMM_EXTCAL_LC_5KOHM
100 kΩ NIDMM_EXTCAL_LC_100KOHM
You have completed adjusting the capacitance and inductance modes of the PXIe-4082.
Proceed to the Completing the Adjustment Procedures section to commit the calibration
constants.
Reverification
Repeat the Verification section to determine the as-left status of the PXIe-4080/4082.
Note If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-4080/4082 to NI. Refer to
the Worldwide Support and Services section for information about support resources
or service requests.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
can obtain the DoC for your product by visiting ni.com/certification. If your product supports
calibration, you can obtain the calibration certificate for your product at ni.com/calibration.
Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for
information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your
software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find
information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer
to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how
to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS
TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.