Dräger Evita XL

The new generation of excellence in Dräger ventilation

MT-0156-2004
TECHNICAL DATA EVITA XL

Patient type – Adults, children, infants (body weight of at least 3 kg/6.6 lbs)
– Premature infants with NeoFlow option

Ventilation settings
Ventilation mode – IPPV, IPPVAssist /CMV, CMVAssist
– SIMV, SIMVPsupp
– MMV, MMVPsupp
– BIPAP1), BIPAP1)ASB, BIPAP1)Assist / PCV+, PCV+Psupp, PCV+Assist
– APRV
– CPAP, CPAPASB / CPAP/Psupp, CPAP/Psupp
– ILV
– PPS (optional)
Enhancements – AutoFlow™ – Automatic adaptation of inspiratory flow
in volume controlled modes
– ATC™ – Automatic Tube Compensation™
– NIV – Mask Ventilation (optional)
– SmartCare®/PS– Automated clinical protocol
in CPAP/ASB / CPAP/Psupp (optional)
– Lung Protection Package –
MT-0068-2008

Recruitment maneuvre and Low Flow maneuvre (optional)


Ventilation frequency (f) 0 to 100/min, 0 to 150/min (Neonatal)
Inspiration time (Tinsp) 0.1 to 10 s
Tidal volume (VT) (BTPS*) – 0.1 to 2.0 L (Adult) / 0.02 to 0.3 L (Pediatric)
– 0.003 to 0.1 L (Neonatal)
Inspiratory flow – 6 to 120 L/min (Adult)
– 6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure 0 to 95 mbar/cmH2O
PEEP / intermittent PEEP 0 to 50 mbar/cmH2O
Pressure assist ASB/Psupp 0 to 95 mbar/cmH2O
Rise time for inspiratory pressure 0 to 2 s
O2 concentration 21 to 100 Vol.% Dräger Evita XL
Multi-sense Trigger Criteria Internal automatic pressure trigger, Flow, Volume
(Flow adjustable 0.3 to 15 L/min)

Measured values displayed


Airway pressure Peak pressure, plateau pressure, mean pressure,
PEEP, min. pressure (-45 to 110 mbar/cmH2O)
Minute volume (MV), (BTPS*) MV, MVspont (0 to 120 L/min, MVleak (0 to 99 L/min)
Tidal volume (VT), (BTPS*) Vtasb 0 - 10, respectively 0 - 3999 ml
02 | DRÄGER EVITA XL

Breathing frequency (f) ftotal, fspon, fmand. (0 to 300 bpm)


O2 concentration (FiO2) Inspired O2 concentration (15 to 100 Vol.%)
Lung mechanics – Resistance (0 to 600 mbar/cmH2O L/s)
– Compliance (0 to 300 mL/mbar/cmH2O)

90 51 738 | 09.10-4 | Marketing Communications | LSL | LE | Printed in Germany | Chlorine-free – environmentally compatible | Subject to modifications | © 2010 Drägerwerk AG & Co. KGaA
Breathing gas temperature 18 °C to 51 °C
Capnography (etCO2) (optional) – 0 to 100 mmHg
CO2 production (VCO2) – 0 to 999 mL/min, STPD*
Serial dead space Vds – 0 to 999 mL, BTPS*
Dead space ventilation (Vds/VT) – 0 to 99%
Weaning parameters – RSB (0 to 9999 (min x L)) / NIF (-45 to 0 mbar/cmH2O)

Alarms / Monitoring
Airway pressure High / Low
Expired minute volume High / Low
Tidal volume High
Apnea alarm Time 5 to 60 s
Spontaneous breath frequency High
Inspired O2 concentration High / Low
Breathing gas temperature High
SpO2 pulse (optional) High / Low
etCO2 (optional) High / Low

Performance data
Valve response time T0...90 ≤ 5 ms
Control principle Time cycled, volume constant, pressure-controlled
Safety relief valve 100 mbar/cmH2O
Leakage and hose system compensation compliance automatic
Max. flow for pressure support and spontaneous breathing 180 L/min
Outlet for pneumatic nebulizer

Operating data
Mains power connection 100 to 240 V, 50/60 Hz, 10 to 30 V DC
Power consumption Approx. 125 W
Gas supply operating pressure O2, air: 2.7 to 6 bar / 39 to 87 PSI

Physical specifications
Dimensions ventilator (W x H x D) 530 x 315 x 450 mm / 20.9 x 12.4 x 17.7 inches (without trolley)
Diagonal screen size 15" TFT color touch screen
Weight basic unit Approx. 50 kg / 64 lbs
Machine outputs:
Digital output Output and reception via an RS 232 C interface
Digital output Output for independent lung ventilation (ILV)
Digital output (optional) For output and reception via two RS 232 C interfaces
Analog output (optional) For analog output of two measured values
1)
BIPAP, trademark used under license. ATC™, trademarked by Dräger. AutoFlow™, trademarked by Dräger. BTPS* (Body Temperature Pressure Saturated).
Measured values relating to the conditions of the patients lung, body temperature 37 °C, steam-saturated gas, ambient pressure. STPD* (Standard Temperature, Pressure, Dry).
Measured values based on normal physical conditions: 0 °C, 1013 hPa, dry.

HEADQUARTERS REGION EUROPE CENTRAL REGION MIDDLE EAST, AFRICA, REGION NORTH AMERICA
Drägerwerk AG & Co. KGaA AND EUROPE NORTH CENTRAL AND SOUTH AMERICA Draeger Medical, Inc.
Moislinger Allee 53–55 Dräger Medical GmbH Dräger Medical GmbH 3135 Quarry Road
23558 Lübeck, Germany Moislinger Allee 53–55 Branch Office Dubai Telford, PA 18969-1042, USA
23558 Lübeck, Germany Dubai Healthcare City, P.O. Box 505108 Tel +1 215 721 5400
www.draeger.com Tel +49 451 882 0 Dubai, United Arab Emirates Toll-free +1 800 437 2437
Fax +49 451 882 2080 Tel + 971 436 24 762 Fax +1 215 723 5935
info@draeger.com Fax + 971 436 24 761 info.usa@draeger.com
contactuae@draeger.com

REGION EUROPE SOUTH REGION ASIA / PACIFIC Manufacturer:


Dräger Médical S.A.S. Draeger Medical Dräger Medical GmbH
Parc de Haute South East Asia Pte Ltd 23542 Lübeck, Germany
Technologie d’Antony 2 25 International Business Park The quality management system at Drä-
25, rue Georges Besse #04-27/29 German Centre ger Medical GmbH is certified accor-
92182 Antony Cedex, France Singapore 609916, Singapore ding to ISO 13485, ISO 9001
Tel +33 1 46 11 56 00 Tel +65 6572 4388 and Annex II.3 of Directive 93/42/EEC
Fax +33 1 40 96 97 20 Fax +65 6572 4399 (Medical devices).
dlmfr-contact@draeger.com asia.pacific@draeger.com

You might also like