Spesifikasi Ventilator Draeger Evita V300 (CO2)

Items Description
Patient Type Neonate, Infant, Pediatric and Adult (Bodyweight ≈ 0,2 to 300 kg)

Ventilation Settings
Invasive Ventilation Modes All Ventilation Modes (Neonate to Adult)
All Ventilation Modes (Pediatric to Adult),
Non-Invasive Ventilation Modes
Pressure Control and Spontaneous/CPAP (Neonate/Infant)
Basic Ventilation Modes :
• Volume (A) Control VC-CMV, VC-AC
• Pressure (A) Control PC-CMV, PC-AC
• Volume SIMV VC-SIMV
• Pressure SIMV PC-SIMV
• Spontaneous & CPAP SPN-CPAP/PS
• Apnea Backup Ventilation Yes, Automatic return to previous mode
Advance Ventilation Modes :
• BIPAP/Assist PC-AC
• BIPAP/PS PC-BIPAP
• APRV PC-BIPAP with inverse Ratio
• Pressure Control with VG
PC-AC/VG, VC-AC/AutoFlow
(BIPAP/Assist - Volume Guarantee)
• Pressure SIMV with VG
PC-SIMV/VG, VC-SIMV/AutoFlow
(BIPAP/PS - Volume Guarantee)
• Volume Support SPN-CPAP/VS
• Minute Volume Guarantee VC-MMV/AutoFlow
• AutoWeaning with EtCO2 SmartCare/PS 2.0

AutoFlow & Volume Guarantee Automatic Tube Compensation Capnograph with mainstream EtCO2
Enhancement
Graphical Loops Overlaid with Reference for Comparison
Parameters Setting :
• Ventilation Frequency (RR)
0.5/minute to 150/minute
• Inspiration Time (Ti)
0.1 second to 10 second 2 ml to 3000 ml
• Tidal Volume
2 L/minute to 120 L/minute
• Inspiratory Flow
1 cmH2O to 95 cmH2O, absolute (include PEEP, not above PEEP) 2 cmH2O to 100 cmH2O
• Inspiratory Pressure
0 to 50 cmH2O
• Insp.Pressure Limit (Pmax)
0 to 95 cmH2O (above PEEP)
• PEEP (Intermittent PEEP)
0 second to 2 second
• Pressure Support
21% to 100%
• Pressure Rise Time (Slope)
0.2 L/minute to 15 L/minute
• O2 Concentration
5% to 70% of Peak Inspiratory Flow
• Inspiratory Trigger Sensitivity
• Termination Criteria (Expiratory Trigger)
Items Description
Ventilation Monitoring
Displayed Parameters :
PEEP,
Minimal Pressure (Pmin), Mean Pressure (Pmean), Plateau Pressure (Pplat),
• Pressure
End Inspiratory Pressure (EIP), Peak Inspiratory Pressure (PIP), Pressure Low (Plow),
Pressure High (Phigh)
Expiratory Minute Volume (MVe), Inspiratory Minute Volume (MVi), Minute Volume (MV),
Expiratory Mandatory MV (MVemand), Percentage Spontaneous MV (%MVspon), Leakage Minute
• Minute Volume & Flow
Volume (MVleak),
%Leak
Volume Tidal (VT), Mandatory VT (VTmand), Spontaneous VT (VTspon), Inspiratory VT (VTi),
Inspiratory Mandatory VT (VTimand), Inspiratory Spontaneous VT (VTispon), Expiratory VT (VTe),
• Tidal Volume
Expiratory Mandatory VT (VTemand), Expiratory Spontaneous VT (VTespon), VT per kg Bodyweight
(VT/kg BW)
Respiratory Rate (RR), Mandatory Rate (RRmand), Spontaneous Rate (RRspon),
Mandatory Triggered Rate (RRtrig), Spontaneous Insp.Time (Tispon), Time Low (Tlow),
• Respiratory Rate & Time
Time Constant, Tau (TC), Expiratory Time Constant (TCe), Ratio Insp. To Exp. Time (I:E),
Ratio Spontaneous IE Ratio (I:Espon)
Inspiratory O2 Concentration (FiO2), Rapid Shallow Breathing (RSB), Dynamic Compliance (Cdyn),
Index last 20%C to Cdyn (C20/Cdyn), Total Resistance (R),
• Others
Patient Airway Resistance (Rpat),
Least Mean Square R,C & CT (r2)
Items Description
Spontaneous Rate (SC-RRspon), Volume Tidal (SC-VT),
• Smarcare/PS Data
End Tidal CO2 (SC-etCO2), Pressure Support (SC-Psupp)
Deadspace Volume (Vds),
Ratio Deadspace to Exp.VT (Vds/VTe), End Tidal CO2 (etCO2),
• End Tidal CO2 Data
Expired CO2 per breath (VTCO2), Expired CO2 per Minute (V'CO2), Increased Measured CO2 (CO2
slope)
Numeric/Graphical Displayed :
• Simultaneous, up to 8 Numerical Data Pressure, Minute Volume & Flow, Tidal Volume, RR & Time, etc Selectable, Pressure, Flow, Volume
• Simultaneous, up to 3 Waveforms 2h, 4h, 8h, 12h, 1day or 7days of Selectable Trend data
• Simultaneous, up to 3 Graphical Trends Pressure, Flow & Volume Combination (Overlaid with Reference)
• Simultaneous, up to 2 Graphical Loops

Alarms / Monitoring
Expired minute volume High / Low
Airway pressure High / Low
Inspired O2 concentration High / Low
Tachypnoea Monitoring High
Volume Monitoring High / Low
Apnea alarm time 5 to 60 second

Performance data
Control principle Time cycled, volume constant, pressure-controlled
Intermittent PEEP duration 1 - 20 expiratory cycles
Medicament Nebulisation Selectable 5, 10, 15, 30 minute for Adult, Pediatric and Infant
Bronchial Suction Manuever Yes
Magnetic Field (Paramagnetic),
O2 Measurement
Lifetime (approx. 5 years and no yearly replacement needed)
Open if Medical Oxygen and Air supply is not sufficient, enables spontaneous
Safety Valve
breathing with ambient air.
Items Description
Operating data
Main Power Connection 100 to 240 Volt, 50/60 Hz
Current Consumption 1.6 Ampere max. at 230 Volt
Up to 240 minutes, integrated on Trolley.
Battery Backup
Capability to Supply Compressor (if ordered/included) for Up to 120 minutes
Gas supply :
2.7 to 6 Bar
• Oxygen Pressure
2.7 to 6 Bar
• Medical Air Pressure
Machine outputs
Digital output and input via an RS232 C interface Draeger MEDIBUS and MEDIBUS.X
Digital output
protocol
Accessories
Humidification System Humidifier F&P MR850
Reusable Breathing Circuit Complete with Chamber for Adult/Pediatric
Breathing Circuit
Reusable Breathing Circuit Complete with Chamber for Infant
Test Lung for Testing Reusable Test Lung for Adult/Pediatric
Ventilation Mask Reusable Size S and M
Nebulisation Reusable Nebulizer
Gas Hose Oxygen Supply Hose (White) Air Supply Hose (Black)
Gas Connector CIG / BritishStandard / DIN

You might also like