CV-200 Ventilator Technical Specifications
CV-200 Ventilator Technical Specifications
CV-200 Ventilator Technical Specifications
200 Ventilator System
CV 2 0 0 V E N T I L A T O R S Y S T E M
CV 200 Ventilator System consists of CV 200 Ventilator Unit and PAVAN Air Medical Compressor
mounted on a trolley. CV 200 Ventilator Unit comes with a built‐in battery for 2 hours backup.
Half an hour UPS backup is provisioned for PAVAN Medical Air Compressor.
TECHNICAL SPECIFICATIONS
A CV 200 VENTILATOR UNIT
Sl No. Parameter Specification
Parameters
Assisted/Controlled Mandatory
Ventilation‐Volume Control (A/CMV‐VC)
Volume Control‐Synchronized
Intermittent Mandatory Ventilation (VC‐SIMV)
1 Modes Assisted/Controlled Mandatory
Ventilation‐Pressure Control (A/CMV‐PC)
Pressure Control‐Synchronized
Intermittent Mandatory Ventilation (PC‐SIMV)
Pressure Support Ventilation (PSV)
Input gas supply should be oxygen and air at 2.75 to
5 bar (40 to 70 psi). Input gas supply must be connected to
CV 200 ventilator unit from direct line maintained at
specified pressure and can be used from external panel
2 Input Gas Supply
outlets and/or cylinders with necessary attachments and
adapters for both oxygen and air. Pressure difference
between air and oxygen must be maintained within 20 psi for
correct FiO2 delivery.
4 to 100 b/min (measured value can be within
3 Respiratory Rate (RR)
±2 bpm of the set bpm)
4:1 and 1:1 to 1:8 (measured value can be within ±0.25 of
4 I:E ratio, I:E
set value)
50 to 1600 mL (measured value can be within ±10% of
5 Tidal Volume (VT)
set value)
10 to 65 cmH2O (measured value can be within ± 5 cmH2O
6 Peak Inspiratory Pressure, PIP
of set value) (5 cmH2O less than Plimit)
CV 200 Ventilator System
Sl No. Parameter Specification
Positive End Expiratory Pressure, 0 to 50 cmH2O (measured value can be within ± 2 cmH2O of
7
PEEP set value) (increments of 1 cmH2O)
8 Pressure Support, Psupport 5 to 40 cmH2O
0.3 to 5 sec ((measured value can be within ± 0.05 sec of
9 Inspiratory Time, INSP TIME
set value)
10 Peak Flow Rate 135 L/min
11 Pressure limit, Plimit 10 to 70 cmH2O (VC) (5 cmH2O above PIP in PC)
12 Pressure Trigger, PTrig OFF, 1 to 15 cmH2O
13 Alarm Volume 1 to 12 (minimum to maximum)
14 Standby/Start Start or Pause Ventilation
15 Inspiratory Pause Available
16 Volume Accuracy 2% to 3% of the full scale between 10 L/min to 80 L/min
17 Patient type Paediatric & Adult
18 Intended use ICU or Recovery room
19 Breathing circuit type Dual‐limb circuit, disposable
20 Mechanical pressure relief 80 ± 5 cmH2O
70 cmH2O (Breathe terminated when Paw reaches
21 Maximum working pressure
maximum working pressure)
22 Inspiratory valve response time < 10 ms
23 Minimum expiratory time 20% of cycle time, 0.2 to 0.8 s
24 Flow measurement dP sensing proximal to patient
25 Airway pressure measurement Proximal sensing
26 Time measurement Microprocessor (inside ventilator)
27 Oxygen monitoring Galvanic cell
28 Display 6.5 inch Colour TFT
Touch screen for parameter selection encoder for value
29 User interface
increment/decrement
30 Nebulizer Ultrasonic nebulisation (Jet not available)
31 Flow 2.5 to 65 L/min
Waveforms and Loops
Paw vs time, Flow vs time, Volume vs time, P‐V loop and
32 Graphics
F‐V loop
X : 0 to 10 s
33 Airway pressure vs time
Y : ‐5 to 80 cmH2O
CV 200 Ventilator System
Sl No. Parameter Specification
X : 0 to 10 s
34 Flow vs time
Y : ‐150 to 150 L/min
X : 0 to 10 s
35 Volume vs time
Y : 0 to 1.6 L
X : 0 to 1600 mL
36 Pressure‐Volume loop
Y : ‐10 to 80 cmH2O
X : 0 to 1600 mL
37 Flow‐Volume loop
Y : ‐150 to 150 L/min
Alarms
38 AC Mains OFF Power supply unplugged
39 Patient disconnected PIP < 4 cmH2O in 3 consecutive breathes
40 High Inspiratory Pressure Airway pressure exceeds the set Plimit
41 Negative Airway Pressure Pressure falls below ‐10 cmH2O
42 High Air/O2 Supply System pressure (internally regulated) exceeds 45 psi
43 Low Air/O2 Supply System pressure (internally regulated) drops to 20 psi or less
Measured PEEP is more than the Set High PEEP limit for
44 High PEEP
3 consecutive breathes
Measured PEEP is less than the Set Low PEEP limit for
45 Low PEEP
3 consecutive breathes
Measured VTe is more than the Set High VTe limit for
46 High VTe
3 consecutive breathes
Measured VTe is less than the Set Low VTe limit for
47 Low VTe
3 consecutive breathes
Measured MVe is more than the Set High MVe limit for
48 High MVe
3 consecutive breathes
Measured MVe is less than the Set Low MVe limit for
49 Low MVe
3 consecutive breathes
Measured RR is more than the Set High RR limit for
50 High Respiratory Rate
3 consecutive breathes
Measured RR is less than the Set Low RR limit for
51 Low Respiratory Rate
3 consecutive breathes
52 High (FiO2) Measured FiO2 is more than the Set High FiO2 limit
53 Low (FiO2) Measured (FiO2) is less than the Set Low FiO2 limit
54 Check Sensor (FiO2) Oxygen sensor is not connected
55 Low Battery Remaining battery level is below 14.5 V ± 0.5 V tolerance
Remaining battery level is below 8 V ± 0.5 V tolerance or
56 Check Battery
Battery is disconnected
CV 200 Ventilator System
Sl No. Parameter Specification
57 Check Flow Sensor connection Flow Sensor is not connected
58 Reverse Flow Sensor Flow Sensor is reversely connected
59 Alarm mute symbol Displays status of audible alarm
60 Volume symbol Displays status of audible alarm volume
61 Battery symbol Displays charge status of battery
62 Visual alarm Flashing RED message display
Power Supply Requirements*
63 Power supply 220 V to 240 V AC, 50 Hz
64 Internal battery 14.8 V/2.2 Ah for more than 2 hours backup
65 Charging time 10 hours (charges automatically during mains operation)
66 Power consumption 16 W (typical)
Mechanical Dimensions
67 Dimension 421 x 328 x 195 mm (W x D x H)
68 Weight 11 kg (approximately)
B PAVAN MEDICAL AIR COMPRESSOR
Sl No. Parameter Specification
Operating Parameters
1 Output Pressure 44 to 65 psi (3 to 4.5 bar) internally adjustable
2 Output Flow (continuous) >50 lpm at 50 psi (3.5 bar)
3 Internal Tank Size 2 litres
4 Oil and Particulate Filtration 5 micron
5 Air Drier System Electronic drier
6 Dew Point Depression < 5 °C below ambient
(a) Normal Mode
7 Modes of Operation
(b) Service Mode
8 Display 2 Line LCD display (16x2 character)
9 Stand‐by Mode Compressor stand‐by option available
10 Cut‐in Pressure 40 ± 3 psi (2.8 ± 0.2 bar)
11 Cut‐out Pressure 48 ± 3 psi (3.3 ± 0.2 bar)
CV 200 Ventilator System
Sl No. Parameter Specification
Equipment Classification
12 Noise Level < 60 dBA at 1 m distance
13 Protection against Electric Shock Class I
14 Mode of Operation Long Term and Continuous
15 Degree of Mobility Mobile (with Trolley)
Parameters Displayed
16 Outlet Pressure 0 to 999 psi (Compressor air outlet pressure)
Alarms Displayed
Alarms when the pressure is low for more than 1 sec in
17 Low Pressure normal running.
Low Pressure : < 30 ± 3 psi (2 ± 0.2 bar)
Alarms when the temperature of the pump is more than
18 High Pump Zone Temperature the specified temperature for more than 30 sec.
Pump Zone Temperature : > 90 °C ± 2 °C
Service Mode
Indicates the temperature of the pump chamber. Display
19 Pump Zone Temperature
range : 0 °C to 99 °C
Indicates the temperature of the ambient air.
20 Ambient Temperature
Display range : 0 °C to 99 °C
Indicates temperature of compressed air at the cooler
21 Dew Point Temperature outlet, i.e., < 5 °C below ambient.
Display range : 0 °C to 99 °C
Indicates the output pressure of the compressor. 50 psi
22 Compressor Outlet Pressure (Factory set)
Display range : 0 ‐ 99 psi
Displays compressor usage in hours
23 Hour Meter
Display range : 0 to 99999 hours
Mechanical Dimensions
24 Dimension 425 x 475 x 448 ± 2 mm (L x B x H)
25 Weight 42.2 ± 0.5 kg (approximately)
Power Supply Requirements
26 Voltage 220 V to 240 V AC, 50 Hz
27 Fuse 5 A, 250 V fuse in‐line and neutral
28 Maximum current 2.5 A at 220 V nominal
CV 200 Ventilator System
C UNINTERRUPTED POWER SUPPLY (UPS)
Sl No. Parameter Specification
1 Type of UPS Online
2 Input Voltage 230 V AC ± 10%, 1 ph, 50 Hz
3 Output Voltage 230 V AC, 1 ph, 50 Hz pure sine wave
4 Output Power Rating 1000 VA
5 Output Voltage Regulation ± 1% both mains and battery
6 Battery Integrated battery required as single UPS unit
7 Battery Type Sealed Maintenance Free
8 Battery Backup Time @750 W 30 minutes (minimum)
9 Display LCD/LED
10 Dimension (including battery) 200 x 420 x 420 mm (W x D x H) (maximum)
11 Weight < 40 kg (including battery)
12 Noise Level 52 dB maximum at 1 meter from UPS surface
13 Input connector & cable length Indian plug, minimum 3 m length
14 Output Connectors Minimum 4 Nos. (5 A or 6 A) Indian Type
15 Operating Temperature 10 °C to 40 °C
16 Certification CE/BIS