HC-300 Ta
HC-300 Ta
HC-300 Ta
Instructions
Please read the manual carefully before operation as incorrect operation may affect the accuracy and precision of
test results and even cause the damages of the analyzer or personal injury.
After reading, please reserve the manual properly for reference at any time.
Address of Registered Person/Manufacturer: 95 Yunhe Street, New & High Tech. Development Zone,
Changchun
Address: Fascinatio Boulevard 522, Unit 1.7, 2909VA Capelle aan den IJssel, The Netherlands
Production Address: 3333 Yiju Road, New & High Tech. Development Zone, Changchun Post Code: 130103
Website: http://www.dirui.com.cn
E-mail: dirui@dirui.com.cn
Fax: 0431-85173354
For effective protection against biological hazards, please observe the following precautions.
● Improper use of samples may lead to infection. Do not touch samples, controls and waste strips with bare hands.
Be sure to wear gloves and work clothes to prevent infection, and protective glasses if necessary.
● If the sample accidentally comes into contact with the skin, please handle it according to the user's work
standard immediately and consult a doctor.
For effective protection against chemical hazards, please observe the following precautions.
Certain cleansers and detergents can damage the skin. Be careful when handling cleansers and detergents to
prevent direct contact of hands and clothing with cleansers and detergents. In case of accidental contact with
hands or clothes, rinse them immediately with clean water. In case of accidental contact with eyes, rinse
immediately with plenty of water and consult an ophthalmologist.
(3) Environmental safety precautions — treatment of waste liquid and solid waste
In order to prevent waste liquid and solid waste from causing environmental pollution and personal injury, please
observe the following precautions when handling waste liquid and solid waste.
● Some substances in quality controls, cleansers, detergents and waste solutions are subject to pollution
regulations and discharge standards. Please comply with local discharge regulations and consult Dirui’s customer
service department.
● Used solid wastes such as waste strips, test tubes, etc. are subject to pollution regulations. Please follow local
waste disposal regulations to dispose of solid wastes.
● When handling waste liquid and solid waste, be sure to wear gloves and work clothes to prevent infection, and
protective glasses and masks if necessary.
To prevent environmental pollution and personal injury from scrapped analyzers, observe the following
precautions when disposing of the instrument.
Some substances in the scrapped analyzer are regulated by contamination regulations. Dispose of the used
analyzer in accordance with local waste disposal regulations.
Alcohol is flammable. Great care must be taken when using alcohol near the Analyzer to avoid fire and explosion.
Precautions for use
(1) User
● The Analyzer shall be used by professional medical examination personnel or trained doctors, nurses and testers.
● As the Analyzer has biological and chemical risks, the operator shall be trained and use personal protective
appliance to reduce the risk.
● Only trained operators are allowed to conduct dangerous operations, such as moving parts.
● If the transportation or storage humidity is greater than 70%, the Analyzer shall be turned on for testing only after
it is stored in a normal working environment for 24 hours (drying process at room temperature), because the
equipment may not meet the safety requirements soon after the drying process.
● The protection measures provided for the Analyzer may become invalid if the Analyzer is not used according to
the manual.
● The Analyzer shall be used under a well-grounded condition, independent power supply shall be used and the
input voltage shall meet requirements. Use specified fuse.
● The Analyzer shall be regularly maintained in strict accordance with the manual, or the Analyzer may have faults
or its test precision and accuracy of the Analyzer may be affected.
● Please use the Analyzer under conditions regulated in the manual. If not, the Analyzer may not operate normally,
the test results may not be reliable, the components of the Analyzer may be damaged and personal injuries may be
caused.
● Do not use turpentine, benzene or other organic solvent to clean the surface of the analyzer because they may
cause housing color or shape change. Wipe with soft cloth or moistened cloth. In case of severe dirt, diluted
alcohol can be used to clean it.
● As for the usage and storage of reagent strip and control, please refer to the relevant instructions. Confirm that
they are within shelf life.
● If any serious incident that has occurred in relation to the Analyzer, the user shall report it to the manufacturer and
the competent authority of the country.
Statement
Dirui Company declares that it will be responsible for the safety, reliability and performances of Urine Analyzer
(model: HC-300) only if all the following requirements are met:
(1) The installation, commissioning and servicing of the Analyzer are undertaken by professional personnel of
Dirui Company.
No further notice will be provided in case of any changes to the software interface.
User Manual
Contents
I
User Manual
II
User Manual
1.1 Overview
1.1.1 Brief introduction
Urine Analyzer (Model: HC-300) is an in vitro diagnostic (IVD) device.
Urine Analyzer (Model: HC-300) should be used together with Dirui urinalysis strips.
Adopting the advanced “high luminosity cold light source reflection determination” technology, the analyzer has a
long lifespan. It can finish the testes of biochemical components in urine within 30 seconds, and it also can revise
the effects on the test result which is caused by pH, hematuria, and abnormally colored sample. It can also be
connected with FUS series urine sediment analyzer.
Contraindication: None.
1.1.2 Intended purpose
The product is a semi-automated instrument for the semi-quantitative analysis of chemical components in human
urine. It can screen normal people and patients with urine-related diseases.
The product has no independent measuring function and uses the supporting reagent strips for urinalysis
manufactured by DIRUI to finish testing. Detectable chemical indicators include (vitamin C (VC), microalbumin
(MALB), creatinine (Cr) and urinary calcium (Ca) are optional): urobilinogen, bilirubin, ketone body, creatinine,
blood, protein, microalbumin, nitrite, leukocyte, glucose, specific gravity (colorimetry), pH, vitamin C, and
urinary calcium.
1-1
User Manual
conversion, the reflectivity of a module can be calculated based on these conversion data. The analyzer will
determine the content of relevant components in the urine according to the reflectivity.
Dual-wavelength testing method is used in application, that is, final test data is obtained according to the ratio of
the reflectivity of the strip block when two kinds of light are used. The formula used for calculation is shown
below.
Tm C r
R
Tr C m
Where:
R Test data of strip block
Tr Reflected intensity of strip block to reference light
Cr Reflected intensity of white benchmark to reference light
Tm Reflected intensity of strip block to test light
Cm Reflected intensity of white benchmark to test light
1 housing 2 strip holder 3 LCD 4 SD card slot 5 thermal printer 6 power switch
7 12V DC power socket 8 RS-232 interface 9 barcode reader interface
Fig. 1-4-1
1.5 Symbols
Table 1-5-1
Symbol Meaning
Biohazard, reminding the user to pay attention; otherwise there is risk of potential bio-infectivity
1-2
User Manual
Symbol Meaning
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
PROTECTIVE GROUND
MANUFACTURER
The symbol of the crossed out wheeled bin indicates that the product (electrical and electronic
equipment) should not be placed in municipal waste. Please check local regulations for disposal of
electronic products.
CATALOGUE NUMBER
“ON”(POWER)
“OFF”(POWER)
1-3
User Manual
1-4
User Manual
2.2 Unpack
Please take out the analyzer and its accessories, and check them with the packing list. If there is damage to the
following parts, please contact the supplier immediately.
1 2 3 4 5
6 7 8 9
1 analyzer 2 power adapter 3 serial port cable 4 barcode reader (optional)
5 User Manual/certificate, etc. 6 sample of strip 7 printing paper
8 standard strip box (with strips inside) 9 CD
Fig. 2-2-1
2.3 Installation
2.3.1 Install printer paper
(1)Obtain a roll of thermal printer paper (Figure 2-3-1) with width of 57mm and diameter less than 45mm.
(2)Press to open the cover of the printer.
(3)Put the new roll of paper into the paper groove, with paper output upwards and near the printer.
(4)Put the printing paper in the printer. Pull the paper up and down to facilitate paper output.
(5)Pull the paper through the hole on the cover the printer and close the printer cover.
2-1
User Manual
Fig. 2-3-1
Fig. 2-3-2
Fig. 2-3-3
2-2
User Manual
Fig. 2-3-4
The barcode reader (optional) that the analyzer uses is a laser one. Do not look directly at the reading
window of the barcode reader when it is power on.
2-3
User Manual
2-4
Clear QC records
Calibration
Strip type
Clear records
Chapter 3 Functions & Settings of Analyzer
LIS standard
QC record query
QC
Main interface
Fig. 3-1-1
More settings
3-1
LIS switch
Settings of standards
Strip registration
Clarity switch
Unit system stettings
Color switch
3.1 Block diagram of functions
3.2 Settings
Fig. 3-2-1
On the main screen (Fig. 3-2-1), press to enter the main menu, and the following is displayed on the
screen:
Fig. 3-2-2
To give up the setting, click to return to the main menu (Fig. 3-2-2). The original number does
not change.
3.2.2 Edit of ID, color and clarity
Press to enter the edit interface of ID, color and clarity, and the screen display is shown in the figure
below:
3-2
User Manual
Fig. 3-2-3
Press to finish editing the sample and to enter the page of editing the next sample.
At most 30 pieces of information can be entered each time; the sample beyond the range is invalid.
(1) Manual input
a) ID edit
Numbers, upper-case letters and lower-case letters can be entered in ID, 20 digits at most.
b) Color edit
When editing color, “On” should be selected for the color status switch; otherwise, no color valve will be
displayed.
c) Clarity edit
When editing clarity, “On” should be selected for the clarity status switch; otherwise, no color valve will be
displayed.
As the next test is under execution, the set ID number, color and clarity will be displayed on the next result.
Table 3-2-1
Color Clarity
YELLOW CLEAR
ORANGE SL CLOUDY
RED CLOUDY
GREEN TURBID
BLUE OTHER
3-3
User Manual
Color Clarity
BROWN /
OTHER /
On the main menu (Fig. 3-2-1) select , and the screen display is shown in the figure below:
Fig. 3-2-4
The user can query records by “Current Results”, “Query by sample NO.” and “Query by ID”.
3.2.3.1 Current Results
From Fig. 3-2-4 select , the screen displays the current record, press to query
records forwards and backwards, press to print the record shown on the screen, and click to
return to Fig. 3-2-4.
3.2.3.2 Query by sample NO.
From Fig. 3-2-4 select , and the screen display is shown in the figure below:
3-4
User Manual
Fig. 3-2-5
Enter the initial sequence number, press to search for sequence records of finished tests. If the
analyzer has the record, the screen displays the test record of the sequence number; if the analyzer does not have
the record, the screen displays “Blank record” (as shown in Fig. 3-2-6); press to return to Fig.
3-2-4.
Fig. 3-2-6
3.2.3.3 Query by ID
From Fig. 3-2-4 select , and the screen display is shown in the figure below:
Fig. 3-2-7
3-5
User Manual
After entering the ID number, press . If the analyzer has the record corresponding to the ID
number, the screen will display the record. If the analyzer does not have the record corresponding to the number,
the screen will display “Blank record” as shown in Fig. 3-2-6.
3.2.3.4 QC query
The analyzer can save 50 QC records. From Fig. 3-2-4 select , press to query
records forwards and backwards, press to print the record displayed on the screen, and press
to return; if the analyzer does not have the record corresponding to the number, the screen displays “Blank
record”.
3.2.3.5 Record clearing
From Fig. 3-2-4 select or and the screen display is shown in the figure
below:
Fig. 3-2-8
Select to keep the saved records in the analyzer; select to clear the saved
records.
3.2.4 Standard strip test and control liquid test
“Standard strip test” and “Control test” will be introduced in “Chapter 4 QC Monitoring”.
3.2.5 System settings
On the main menu (Fig. 3-2-1) select , and the screen display is shown in the figure below:
3-6
User Manual
Fig. 3-2-9
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
Fig. 3-2-10
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
Fig. 3-2-11
3-7
User Manual
● The matching strips of the analyzer are Dirui reagent strips for urinalysis. To guarantee accurate test
results, please don’t use strips of other models.
● The strip selecting function needs the supplier’s authorization.
3.2.5.3 Status settings
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
Fig. 3-2-12
In status settings, the user can set the switches of ID, cutoff, plus system, internal printing, color, turbidity, and
LIS port.
(1) : When “ID” is set on “On”, if the barcode reader is connected, the barcode on a test tube
can be scanned; the output result shows the report ID number (Fig. 3-2-13). When “ID” is set on “Off”, the output
result does not show the ID number (Fig. 3-2-14). As to all the following reports, H11 is taken as an example for
explanation.
The quantity of barcodes scanned each time cannot exceed 30. After sample testing, you can continue
scanning a barcode, but the total number of barcodes for testing cannot exceed 30. Once exceeding 30, the
barcode scanned is invalid; the length of a barcode should be less than 20 characters. It indicates scanning
failure if no response sound is sent out by the analyzer when scanning a barcode.
(2)
3-8
User Manual
When “Abnormal” is set on “On”, if the test value is equal to or more than the cutoff set by the user, “*” is printed
before the test item (Fig. 3-2-13); When “Abnormal” is set on “Off”, “*” is not printed in the test result (Fig.
3-2-14).
(3)
When “Plus Used” is set on “On”, the analyzer display and print the test value and “+” corresponding to the test
value under the current unit system (Fig. 3-2-13); When “Plus Used” is set on “Off”, the analyzer only displays
and prints the test value under the current unit system (Fig. 3-2-14).
(4)
When “Printer” is “On”, the built-in printer of the analyzer prints test data; when “Printer” is “Off”, the built-in
printer of the analyzer does not print test data.
(5)
When “LIS Port” is “On”, the analyzer can transmit data to the computer; when “LIS Port” is “Off”, the analyzer
cannot transmit data to the computer.
(6) ,
When “Color” is set on “On”, the “Color” item is printed simultaneously in the test result; when “Clarity” is set
on “On”, the “Clarity” item is printed simultaneously in the test result. The inspector can write the test results by
hand according to the “Color” and “Clarity” of the sample tested. When “Color” or “Clarity” is set on “Off”,
“Color” or “Clarity” will not be printed in the test result.
3.2.5.4 Selection of language
From Fig. 3-2-9 select , the screen displays the following figure:
Fig. 3-2-15
The analyzer offers eight languages. The default is English. Other languages need upgrade of language packs.
In Chinese mode, the screen of the analyzer displays Chinese prompts and Chinese reports are output.
In mode of other languages, the screen of the analyzer displays English prompts and English reports are output.
3.2.5.5 Cutoff settings
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
3-9
User Manual
Fig. 3-2-16
Press “ ” to select a level, and press . Press and the setting is not saved.
3.2.5.6 Unit settings
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
Fig. 3-2-17
From Fig. 3-2-9 select , and the screen display is shown in the figure below:
3-10
User Manual
Fig. 3-2-18
(1) Test mode: The analyzer offers two kinds of test mode. You can switch between “Fast mode” and “Slow
mode”.
a) In fast mode, after testing one strip, the analyzer can continuously test next ones with no need of pressing any
button; the fast mode is suitable for continuous tests of a large amount of urine samples; the test time of each
strip is 30 seconds.
b) In slow mode, the analyzer stops after each test. You need to press for the next test; the slow
mode is applicable to inconsecutive tests and the test time of each strip is 60 seconds.
(2) Baud rate: used to set analyzer’s transmission speed; the baud rate is switched between “9600” and “1200”.
(3) LIS: Standard or Simple can be selected.
3.2.6 Registration of strips
Fig. 3-2-19
On the main screen (Fig. 3-2-19), press and the screen displays the following:
3-11
User Manual
Fig. 3-2-20
Recharge steps:
(1) Press , put the recharge card near the card reading area of the analyzer, and the card
information will be displayed on Fig. 3-2-20 after successive reading.
(3) Press , put the recharge card near the card reading area of the analyzer, and you can take away
the card after successive recharge.
: Strip table moved out; remove the strip table and clean it
3-12
User Manual
Chapter 4 QC Monitoring
Place a standard strip flat in the center of the strip holder, push it to the top of the strip holder, select on
the main menu (Fig. 3-2-1), and the analyzer carries out testing.
After testing, if the calibration passed, the result would be “Calibration OK”; if it failed, the result is “Calibration
Not OK”.
In the case of calibration failure, please execute calibration testing again. If it still failed, please contact the Dirui
customer service department or the distributor.
Put a strip with control liquid in the strip holder, select from the main menu (Fig. 3-2-1), and the result
will be output after testing. For QC results refer to the reference values in the instructions about the control liquid.
If QC failed, please use control liquid of another lot number to execute QC testing again. If it still failed, please
contact the Dirui customer service department or the distributor.
4-1
User Manual
4-2
User Manual
● Urine may be potentially infectious; please take protective measures during testing, cleaning or
maintenance.
● Please dispose of urine samples and used strips in accordance with local laboratory regulations.
This chapter introduces the routine urine test method of using the analyzer. Before using it for the first time,
please read “Chapter 3 Functions & Settings of Analyzer” in detail.
● When the self-test fails, the screen displays an error message code, please refer to “Appendix D Fault
Information Table” for processing.
● Do not place the analyzer in direct sunlight during the test, so as not to affect the test accuracy.
● Please check the urinalysis strip model before the test to avoid incorrect test results due to the incorrect
urinalysis strip model.
● Do not use urinalysis strips beyond the expiration date or that deteriorate.
Fig. 5-2-1
5-1
User Manual
Step 1: Press for about two seconds. After hearing the prompt tone, completely immerse the reagent
area of the strip in a fresh, well-mixed and uncentrifuged sample, immediately take it out, place it in the center of
the strip holder, and push forward until it touches the top of the strip holder (this is the first strip).
Step 2: After hearing the prompt tone for the second time, immerse the second strip in the next sample,
immediately take it out and put it on the absorbent paper to wait for the test (this is the second strip).
Step 3: When hearing the prompt tone for the third time, immerse another strip and put it on the absorbent paper;
at this time, the first test is completed and the test result is printed; the strip table is removed, remove the first strip
from the strip holder, and put the second strip on the strip holder for testing.
Step 4: Repeat the operation of Step 3 every time you hear the prompt tone.
The analyzer stops testing: Remove the strip from the strip holder, the screen displays “No strip” or “Strip
not in place”, and the analyzer stops testing.
5.3.2 Slow mode
The test time for each strip is 60 seconds. The operation method is as follows:
On the main menu, press for about 2 seconds; after hearing the prompt sound, completely immerse the
reagent area of the strip in a fresh, well-mixed, uncentrifuged sample, take it out immediately, place it in the
center of the strip holder, and push it forward until it touches the top of the strip holder (this is the first strip).
After waiting for 40 seconds, the analyzer starts the test. After the test is completed, the test result is printed out,
the strip table is moved out, and the analyzer returns to the main interface. To test the next sample,
needs to be pressed again.
● The strip to be tested must be placed before the strip table moves.
● If there is no strip on the strip holder, after the analyzer detects the strip holder, the screen displays “No
strip” or “Strip not in place”, and the strip holder moves out. Press to continue the test.
● During the test of the analyzer, the keyboard is in an invalid state, and the analyzer will not respond if
any key is pressed.
● Do not hit the strip table during the test. If the strip table is bumped, the screen will display “The strip
table is blocked”.
5-2
User Manual
● Do not spill water, urine samples or other solutions on the machine or electric parts of the analyzer in
case of any damage.
● During operation, operators shall take preventive measures like wearing protective gloves and working
suits in case of infection due to touching the polluted area or solution. The operator shall rinse with water
and take disinfection measures after contact with the polluted or corrosive liquid due to carelessness.
● If the analyzer is installed with accessories which are not provided or recommended by the manufacturer
or the analyzer is not used in the manner specified by the manufacturer, the relating protection may be
weaken.
● During cleaning and maintenance, please check whether there are hazards caused by inefficiency of parts
filled with liquid.
6-1
User Manual
6.3 Sterilization
Because the strip holder is in frequent contact with urine samples, it should be sterilized.
One of the following solutions can be used for disinfection:
(1) 2% glutaraldehyde solution.
(2) 5% sodium hypochlorite solution.
The disinfection steps are as follows:
(1) Inject a 10 cm high disinfectant solution into a suitable container.
(2) Immerse the strip holder in the disinfectant and ensure that the white benchmark does not touch the
disinfectant.
(3) Soak for 10 minutes.
(4) After taking out, rinse the disinfectant on the strip holder with clean water.
(5) Dry the strip holder and white benchmark with a soft cloth.
(6) Install the strip holder.
(7) Restart the instrument for self-testing.
The waste generated by the operation of the analyzer has the risk of biological and chemical pollution,
please handle it as required.
6-2
User Manual
7.1 Transportation
The transportation of the analyzer is specified in the contract, during which fierce vibration and extrusion shall be
prevented, and also the analyzer shall be handled gently during loading/unloading.
7.2 Storage
Store the analyzer in a clean and ventilated room, free of chemicals and corrosive gas, and with ambient
temperature of -40℃~55℃ and relative humidity not more than 93%.
7-1
User Manual
7-2
User Manual
A-1
User Manual
A-2
User Manual
B-1
User Manual
(2) SI, Conventional and Plus communication format (barcode scanning on):
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
STX CR LF
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP × × × × CR LF
SP I D : × × × × × × × × × × × × × × × × × × CR LF
SP U B G SP × × × × × × × × × × × × × × × × CR LF
SP B I L SP × × × × × × × × × × × × × × × × CR LF
SP K E T SP × × × × × × × × × × × × × × × × CR LF
(SP C R E SP) CR LF
SP B L D SP × × × × × × × × × × × × × × × × CR LF
SP P R O SP × × × × × × × × × × × × × × × × CR LF
SP N I T SP × × × × × × × × × × × × × × × × CR LF
SP L E U SP × × × × × × × × × × × × × × × × CR LF
SP G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP P H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
(SP A L B SP) × × × × × × × × × × × × × × × × CR LF
(SP C a SP SP) CR LF
ETX
B-2
User Manual
Notes:
The above two tables list the communication format of the 11 tests of the analyzer. If there are less than 11
test items, the line corresponding to the name of the item not tested is not output.
Note:
When the strip is 11MA(N) or 11MA, the line of VC only shows the test item, no output content.
STX=0X02 CR=0X0D LF=0X0A SP=0X20
SP1=0XAB ETX=0X03 -=0X2D X=any ASCII character
B-3
User Manual
B-4
User Manual
Conventional SI
Item Abbr. Plus
Plus on Plus off Plus on Plus off
Normal 0.2 mg/dL 0.2 mg/dL Normal 3.4μmol/L 3.4μmol/L Normal
Normal 1 mg/dL 1 mg/dL Normal 17 μmol/L 17 μmol/L Normal
Urobilinogen UBG 1+ 2 mg/dL 2 mg/dL 1+ 34 μmol/L 34 μmol/L 1+
2+ 4 mg/dL 4 mg/dL 2+ 68 μmol/L 68 μmol/L 2+
3+ >=8 mg/dL >=8 mg/dL 3+ >=135 μmol/L >=135 μmol/L 3+
Neg Neg Neg Neg Neg
1+ 1 mg/dL 1 mg/dL 1+ 17 μmol/L 17 μmol/L 1+
Bilirubin BIL
2+ 3 mg/dL 3 mg/dL 2+ 51 μmol/L 51 μmol/L 2+
3+ >=6 mg/dL >=6 mg/dL 3+ >=103 μmol/L >=103 μmol/L 3+
C-1
User Manual
C-2
User Manual
The power supply is not (1)Check whether the power supply is normal;
1 The screen has no display. connected. (2)If it cannot be solved, please contact the
Mainboard CPU fault. supplier.
(1)Check the front end of the strip table. If there
are obstacles, remove them.
The strip table is not in The motor movement is
2 (2)Check the mechanical operation part. If there
place. blocked.
are obstacles, remove them.
(3)Click the “Start” key, and test again.
The screen displays “Strip (1)Place a strip.
3 No strip on strip table
Missed”. (2)Press the “Start” key to continue testing.
(1)Error in position for strip (1)Take a new strip and place correctly.
The screen displays “Strip
4 (2)Strip not completely (2)Immerse the strip completely in the sample.
error”.
immersed in sample (3)Press the “Start” key to continue testing.
Check white benchmark, make sure that there is
5 2-33 Low white benchmark
no foreign matter, and reboot the instrument.
The standard strip test White benchmark block dirty (1)Wipe clean the white benchmark block.
9 value is beyond the Luminotron aging (2)Replace the strip holder.
reference range. Standard strip dirty (3)Replace the standard strip.
11 2-35 Oblique strip Check the strip placing position. Test again.
D-1
User Manual
D-2
User Manual
E-1
User Manual
E-2
User Manual
Cycle of Way of
Name of Part Position Remarks
Replacement Replacement
Strip holder Instrument’s front end outlet 5 years
Strip Strip holder Timely
Thermal printer paper Built-in printer Timely 2.3.1
Standard strip Standard strip box Timely
F-1
User Manual
F-2
User Manual
Radiated
IEC 3V/m, 80MHz~2.0GHz,
Shell electromagnetic A Applicable
61000-4-3 80%AM
fields
Rated
IEC
power-frequency 3A/m, 50Hz A Applicable
61000-4-8
magnetic field
0% for 1 cycle; B
IEC
Voltage sag 40% for 5 cycles; B Applicable
61000-4-11
70% for 25 cycles C
Voltage IEC
5%, duration: 250 cycles C Applicable
interruption 61000-4-11
AC
power IEC
Pulse train ±1kV(5/50ns,5kHz) B Applicable
supply 61000-4-4
Radio-frequency IEC
3V, 150kHz~80MHz, 80%AM A Applicable
conduction 61000-4-6
IEC
Pulse train ±1kV(5/50ns, 5kHz) B ——
61000-4-4
DC
IEC Line to ground: ±2 kV
power Surge B ——
61000-4-5 Line to line: ±1kV
supply
Radio-frequency IEC
3V, 150kHz~80MHz, 80%AM A ——
conduction 61000-4-6
IEC B
Pulse train ±0.5kV(5/50ns, 5kHz)
61000-4-4
I/O ——
signal
Radio-frequency IEC A
3V, 150kHz~80MHz, 80%AM
conduction 61000-4-6
IEC
I/O Pulse train ±1kV(5/50ns,5kHz) B ——
61000-4-4
signal
to main
power Radio-frequency IEC
3V, 150kHz~80MHz, 80%AM A ——
conduction 61000-4-6
Remark: “——” means the item is not applicable according to the standard stipulations.
G-1
User Manual
Applicable or
Test items EMC standard Test value
not
Voltage fluctuation
IEC 61000-3-3:2005 Meets the requirements in IEC 61000-3-3:2005. ——
and flicker
Remark: “——” means the item is not applicable according to the standard stipulations.
G-2
User Manual
Harmful materials
Name of part Hexavalent Polybrominated Polybrominated
Lead Mercury Cadmium
chromium biphenyls diphenyl ethers
(Pb) (Hg) (Cd)
(Cr (VI) ) (PBB) (PBDE)
LCD screen ○ ○ × ○ ○ ○
Main control
× ○ × ○ ○ ○
board
Sampling board × ○ ○ ○ ○ ○
Electrical
○ ○ × ○ ○ ○
interface board
○: Indicates that the content of the hazardous substance in all homogeneous materials of the part is below the limit requirement
specified in IEC 32321-2.
×: Indicates that the content of the hazardous substance in at least one homogeneous material of the part exceeds the limit
requirement specified in IEC 32321-2, but is within the allowable use range.
H-1
User Manual
H-2
User Manual
I-1
User Manual
I-2