Service Manual: BA-88A Semi-Auto Chemistry Analyzer
Service Manual: BA-88A Semi-Auto Chemistry Analyzer
Service Manual: BA-88A Semi-Auto Chemistry Analyzer
Service Manual
, , , , , BeneView, WATO,
BeneHeart, are the registered trademarks or trademarks owned by Mindray in China
and other countries. All other trademarks that appear in this manual are used only for
editorial purposes without the intention of improperly using them. They are the property of
their respective owners.
WARNING:
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or personal injury.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
any product of any other manufacturer.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
1 Return authorization: Contact the Customer Service Department and obtain
a Customer Service Authorization number. This number must appear on the
outside of the shipping container. Returned shipments will not be accepted if
the number is not clearly visible. Please provide the model number, serial
number, and a brief description of the reason for return.
2 Freight policy: The customer is responsible for freight charges when this
product is shipped to Mindray for service (this includes customs charges).
3 Return address: Please send the part(s) or equipment to the address offered
by the Customer Service department
ii
EC Representative
Phone: 0049-40-2513175
Fax: 0049-40-255726
iii
Bold and Italic font indicates text displayed on the screen, such as Sample Request.
Safety Symbols
This chart explains the symbols used in this manual.
Serial Number
Foreward 1
Manufacturer
ON (Power)
Graphics
All graphics, including screens and printout, are for illustration purposes only and must
not be used for any other purpose.
EC Representative
Name: Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestraβe 80, 20537 Hamburg Germany
Phone: 0049-40-2513175
Fax: 0049-40-255726
2 Foreward
WARNING
If the instrument is used in a manner not specified by our company, the
protection provided by the system may be impaired.
WARNING
When the instrument is turned on, users must not open the cover.
Spillage of reagent or sample on the analyzer may cause equipment
failure and even electric shock. Do not place sample and reagent on
the analyzer. In case of spillage, switch off the power immediately,
remove the spillage and contact our company customer service
department or your local distributor.
This instrument is supplied with a slow-blow fuse (250V, 3.15A), which
must not be replaced by the user.
Power supply: 100-240V~, 50/60Hz.
The instrument is supplied with a three-wire power cord and should be
properly grounded during application.
WARNING
Do not put your finger or hand into any open part when the system is in
operation.
Foreward 3
WARNING
Light sent by the photometer lamp may hurt your eyes. Do not stare into
the lamp when the system is in operation.
If you want to replace the photometer lamp, first switch off the Main
Power and then wait at least 15 minutes for the lamp to cool down
before touching it. Do not touch the lamp before it cools down, or you
may get burned.
Preventing Infection
Please observe the following instructions to protect against the biohazardous infection.
BIOHAZARD
Inappropriately handling samples may lead to biohazardous infection. Do
not touch the sample, mixture or waste with your hands. Wear gloves and
lab coat and, if necessary, goggles.
In case your skin contacts the sample, follow standard laboratory safety
procedures and consult a doctor.
WARNING
Reagents and enhanced wash solution may hurt human skins. Exercise
caution when using the reagents and enhanced wash solution. In case
your skin or clothes contact them, wash them off with clean water. In
case the reagents or wash solution spill into your eyes, rinse them with
much water and consult an oculist.
4 Foreward
BIOHAZARD
Some substances in reagent, control, enhanced wash solution and
waste are subject to regulations of contamination and disposal. Dispose
of them in accordance with your local or national guidelines for
biohazard waste disposal and consult the manufacturer or distributor of
the reagents for details.
Wear gloves and lab coat and, if necessary, goggles.
WARNING
Materials of the analyzer are subject to contamination regulations. Dispose
of the waste analyzer in accordance with your local or national guidelines
for waste disposal.
WARNING
Ethanol is flammable substance. Please exercise caution while using the
ethanol.
Foreward 5
Intended Use
WARNING
The system is an analyzer designed for in vitro quantitative
determination of clinical chemistries in serum, plasma, urine and CSF
samples. Please consult Mindray first if you want to use the system for
other purposes.
To draw a clinical conclusion, please also refer to the patient’s clinical
symptoms and other test results.
Operator
WARNING
The system is to be operated only by clinical professionals, doctors or
laboratory experimenters trained by our company or our authorized
distributors.
Environment
CAUTION
Please install and operate the system in an environment specified by this
manual. Installing and operating the system in other environment may lead
to unreliable results and even equipment damage.
To relocate the system, please contact our customer service department or
your local distributor.
6 Foreward
CAUTION
Electromagnetic noise may interfere with operations of the system. Do
not install devices generating excessive electromagnetic noise around
the system. The electromagnetic environment should be evaluated
prior to operation of the device. Do not use such devices as mobile
phones or radio transmitters in the room housing the system. Do not
use other CRT displays around the system. The electromagnetic noise
might lead to system failures.
Do not use other medical instruments around the system that may
generate electromagnetic noise to interfere with their operations.
NOTE
It is the manufacturer's responsibility to provide equipment
electromagnetic compatibility information to the customer or user.
NOTE
It is the user's responsibility to ensure that a compatible
electromagnetic environment for the equipment can be maintained in
order that the device will perform as intended.
CAUTION
Operate the system strictly as instructed by this manual. Inappropriate use
of the system may lead to unreliable test results or even equipment
damage or personal injury.
Before using the system for the first time, run the calibration program and
QC program to make sure the system is in normal status.
Be sure to run the QC program every time you use the system, otherwise
the result may be unreliable.
Do not touch the screen with wet hands or hands contaminated by
chemicals.
Do not place the Power to ON again within 10 seconds since placing it to
OFF;
Foreward 7
CAUTION
Maintain the system strictly as instructed by this manual. Inappropriate
maintenance may lead to unreliable results, or even equipment damage
and personal injury.
To wipe off dust from the system surface, use a soft, clean and wet (not too
wet) cloth, soaked with mild soap solution if necessary, to clean the
surface. Do not use such organic solvents as ethanol for cleaning. After
cleaning, wipe the surface with dry cloth.
Switch off all the powers and unplug the power cord before cleaning. Take
necessary measures to prevent water ingression into the system,
otherwise it may lead to equipment damage or personal injury.
Replacement of such major parts as lamp assembly must be followed by a
calibration.
Check the pump tubing for leakage as needed and replace the tubing in
time. Otherwise, the normal aspiration of the system might be affected. It is
recommended that the inner system tubing should be replaced every 24
months to avoid possible blockage or invalidation brought about by aging.
CAUTION
To define such parameters as calculation method and wavelength, follow
the instructions in this manual and the package insert of the reagents.
8 Foreward
CAUTION
Use samples that are completely free of insoluble substances like fibrin, or
suspended matter; otherwise the probe may be blocked.
Medicines, anticoagulants or preservative in the samples may lead to
unreliable results.
Hemolysis, icterus or lipemia in the samples may lead to unreliable test
results, so a sample blank is recommended.
Store the samples properly. Improper storage may change the
compositions of the samples and lead to unreliable results.
Sample volatilization may lead to unreliable results. Do not leave the
sample open for a long period.
Some samples may not be analyzed on the system based on parameters
the reagents claim capable of testing. Consult the reagent manufacturer or
distributor for details.
Certain samples need to be processed before being analyzed by the
system. Consult the reagent manufacturer or distributor for details.
CAUTION
Use appropriate reagents, calibrators and controls on the system.
Select appropriate reagents according to performance characteristic of the
system. Consult the reagent suppliers, our company or our authorized
distributor for details, if you are not sure about your reagent choice.
Store and use reagents, calibrators and controls strictly as instructed by
the suppliers. Otherwise, you may not obtain reliable results or best
performance of the system.
Improper storage of reagents, calibrators and controls may lead to
unreliable results and bad performance of the system even in validity
period.
Perform a calibration after changing reagents. Otherwise, you may not
obtain reliable results.
Contamination caused by carryover among reagents may lead to
unreliable test results. Consult the reagent manufacturer or distributor for
details.
Foreward 9
WARNING
External equipment connected to the analogue and digital interfaces
must be complied with the relevant Safety and EMC standards (e.g., IEC
60950 Safety of Information Technology Equipment Standard and CISPR
22 EMC of Information Technology Equipment Standard (CLASS B)).
Any person, who connects additional equipment to the signal input or
output ports and configures an IVD system, is responsible for ensuring
that the system work normally and complies with the safety and EMC
requirements. If you have any problem, consult the technical services
department of your local representative.
Communication interface
CAUTION
The system is equipped with two USB ports which can be used in
connecting the keyboard, mouse, printer and other external
equipments or in system upgrading. RS232 is used in connecting the
PC with the analyzer to transfer data.
These three ports should not be used to operate the system for usage
other than those mentioned above. Otherwise, system might be
damaged.
10 Foreward
Contents.................................................................................................................. I
1 System Specifications .............................................................................. 1-1
1.1 System specifications............................................................................................. 1-1
1.2 Parameters ............................................................................................................ 1-1
1.3 Instrument Configuration ........................................................................................ 1-2
1.4 Overview................................................................................................................ 1-2
2 System Installation.................................................................................... 2-1
2.1 Basic...................................................................................................................... 2-1
2.2 Installation Procedures ........................................................................................... 2-2
3 System Description ................................................................................... 3-1
3.1 Operating Procedures ............................................................................................ 3-1
3.2 Component Structure and Functions....................................................................... 3-2
3.2.1 Photometer Assembly ................................................................................ 3-2
3.2.2 Lamp assembly ......................................................................................... 3-3
3.2.3 Lamp Base Assembly ................................................................................ 3-5
3.2.4 Flow Cell Base Assembly........................................................................... 3-9
3.2.5 Base Plate Assembly ............................................................................... 3-12
3.2.6 Enclosure Assembly ................................................................................ 3-13
4 Hardware .................................................................................................... 4-1
4.1 Overview................................................................................................................ 4-1
4.2 Safety Precautions ................................................................................................. 4-1
4.3 Introduction of the Modules .................................................................................... 4-1
4.4 CPU Board............................................................................................................. 4-2
4.4.1 LED Indicator............................................................................................. 4-2
4.4.2 Memory Module ......................................................................................... 4-2
4.4.3 Clock Module............................................................................................. 4-3
4.4.4 Interface Module ........................................................................................ 4-3
4.5 CPU Extension Board ............................................................................................ 4-3
4.5.1 Overview ................................................................................................... 4-3
4.5.2 LED Indicator............................................................................................. 4-4
4.5.3 USB Functions........................................................................................... 4-4
4.5.4 LCD&Touchscreen Driver Control .............................................................. 4-5
4.5.5 Serial Port Communication ........................................................................ 4-5
4.5.6 Driver of the Step Motor ............................................................................. 4-6
4.5.7 Driver of the Peristaltic Pump ..................................................................... 4-6
4.5.8 Driver of the Refrigeration Plate ................................................................. 4-6
4.5.9 Detection of the Input Signal ...................................................................... 4-7
4.5.10 Analogue Board Control........................................................................... 4-7
Contents I
II Contents
Weight: ≤7Kg
1.2 Parameters
Reaction type: Endpoint, Kinetics, Fixed-time, Absorbance;
1.4 Overview
Colorimetric system: The colorimetric analysis is realized through the
combination of optical collimating, mono-color filter, mechanical parts and
software, hardware module.
2.1 Basic
Make sure the installation site of the hospital meet the system installation
requirements on space, power supply and environment. Please refer to the
operator’s manual for details.
Memory: at least 1G for Vista operating system; at least 512 M for other
operating system.
Other
tests
All the
Lamp tests
stabilization Reaction finished
Temperature
stabilization
Procedures:
1 Empty the waste. Connect the waste tubing to the waste container. Prepare
the samples, reagents and distilled water;
2 Connect power cable correctly. Power on the analyzer. The instrument will
under go the initialization processes, including; system startup, hardware
self-check, parameters download, dark current testing. After initialization,
wait until the temperature and the light source are stable. The whole
process lasts for about 15 minutes;
3 After the system is stable, set the parameters of the tests, such as
wavelength, temperature, methods and etc. Start the test.
3.2.1.1 Introduction
Photometer assembly is the core of BA-88A, integrating optical system, temperature
control system and photoelectrical collection system. The photometer assembly is the
key for reliability of the system. Photometer assembly includes lamp assembly, lamp
base assembly, flow cell assembly and other components, which are used to position
and install the optical components, thermal components and analogue board.
1 When disassembling the instrument, do not drag the cable with excessive force.
If necessary, you can disconnect the connection of some cables in advance. Do
not make any cable pressed, especially the cable under the photometer base.
2 When disassembling the instrument, do not press the air duct or conduct any
operation that might distort the air duct, otherwise the refrigeration plate might be
damaged.
3.2.2.1 Introduction
The lamp assembly is the light source of the instrument, providing energy source for
the detection system. Among the components of the lamp assembly, the lamp should
be replaced regularly, so when we design the system, we take the manual
replacement into consideration, taking the assembly as a whole connected with the
lamp base, fixing it with a long hand screw to facilitate manual disassembly.
Components relevant to lamp replacement are shown in the following figure. For
more details please refer to 5.3.3.
WARNING
If you want to replace the lamp, first switch off the power and then wait at
least 15 minutes for the lamp to cool down before touching it. Do not touch
the lamp before it cools down, or you may get burned.
NOTE
When replacing the lamp assembly, do not touch the surface of the bulb to
avoid contamination.
When dismounting the restraining screw of the lamp base, please take extra
care to avoid element falling into the inside of the instrument. If it is difficult to
dismount, some tools like the clipper might be used. The photoelectric gain
should be adjusted after the lamp is replaced. For more details, please refer
to the operator’s manual.
3.2.3.1 Introduction
The lamp base assembly supports the installation and positioning of the lamp
assembly and in the meantime supports the installation and positioning for filter wheel
and driver motor.Integration structure design will increase the positioning accuracy of
the optical elements.
1 If fixture is not used when you assemble or disassemble the filter, extra care
should be taken lest the filter is scratched by the screw driver. Please follow the
correct order to install the filter. The order is shown in the following table.
No. 1 2 3 4 5 6 7 8
Wavelength
340 405 546 670 450 510 578 630
nm
450 nm and 670 nm are optional wavelengths. When these two wavelengths are
not configured, relevant position is shielded by stop plate; when relevant filters are
installed, the arrow on the circle should point outward of the hole.
2 When assembling or disassembling the home position sensor, place the surface
with text onward (refer to the base plate).
3 When assembling or disassembling the lens, make sure the spherical surface
face the outside. Press it until secure with clamping ring. Make sure that the
outer surface of the clamping ring is at the same level with or higher than the
5 During assembly and disassembly, ensure that optical elements like filter, lens
and bulb are properly protected. When installation is completed, use balloon to
blow away the dust or other impurities from the surface of the optical elements.
3 Click H pos..
4 When the system is in Standby (if Lamp stabilizing is displayed, tests can only
be started when lamp is stabilized), Put the aspiration tubing into a tube filled
with about 2 ml of distilled water. Click Asp.. When the aspiration is completed,
click Start to calculate the deviation.
6 It will take 3 minutes to complete the test. When the test is completed, the
deviation of the filter wheel will be displayed in Deviation. The deviated steps
should be within 0-63, otherwise, the position of the home position sensor should
be adjusted before testing.
Tips: when the value of the deviated steps is higher than 63, move the sensor
downward; when the value is lower than 0, move it upward. When the above
requirements are met, click Save.
7 When operations above are completed, click Return to go back to the main
screen. The analyzer will restart automatically. When analyzer is stable (after
about 15-20 minutes), click Maintenance-Maintenance on the analyzer to enter
the maintenance screen. Select Gain adjust on the screen to adjust the
photoelectric gain of the system. When adjustment is completed, wait at least 15
minute to start testing. Please refer to the operator’s manual for details on
photoelectric adjustment.
NOTE
The assembly or disassembly of the filter wheel, motor or home position
sensor should be followed with home position deviation adjustment.
Otherwise, the test result might not be reliable.
The home position deviation adjustment of the filter wheel should be followed
with the adjustment of photoelectric gain. Otherwise, the test result might not
be reliable.
3.2.4.1 Introduction
The flow cell, temperature control elements, analogue board and etc are installed and
positioned by the flow cell base assembly. Two modes—flow cell and cuvette are
available on the system. Manual replacement is supported. The temperature control
system is composed of semi-conductor refrigeration plate, air duct, temperature
sensor and insulation material, providing 37℃, 30℃ and 25℃ which are needed on
the system. Analogue board collects the data of absorbance and temperature.
2 When installing the refrigeration plate, please pay attention to the installing
direction. The cable should project at the side of the analogue board and the red
cable should be lower;
3 The semi-conduct refrigeration plate should be installed between the air duct
and the flow cell base. Before installing, apply heat conducting gel to both sides
of the refrigeration plate (the thickness is about 0.05-0.1mm, the gel should be
evenly applied and no granule or nodular matter is allowed; the gel should not be
too watery). Correctly stick the plate inside the groove of the flow cell assembly
(follow step 2).Stick the air duct to the other side of the plate and tighten it with
spring and spring lightener. The two longer spring lightener should be inside the
hole on the upper side of the air duct and insulation is needed. Use straight slot
head screw driver to tighten the screws.
6 Do not drag the cable with excessive force. If necessary, you can disconnect the
connection of some cables in advance.
7 When replacing the temperature sensor, please take care to dismount the
insulation cotton, otherwise damage might occur; when reinstalling, use double
glue to stick the insulation back.
3 Select Temp. to enter the following screen. Select Cell. Enter the A, B, C
parameters on the flow cell accessory into their respective positions, making sure
that they are in accordance with those on the accessory. Click Save.
5 After above operations are completed, click Return. Click View para. and make
sure the inquired A, B and C parameters are in accordance with those on the
accessory. Click Return.
6 Exit the Maitenance screen and the system will restart. Follow the instruction to
complete the configuration of the temperature.
3.2.5.1 Introduction
The base plate assembly supports the system, providing fixing and support for
photometer assembly, hardware board, interface and other components.
1 When assembling and disassembling the photometer assembly, do not drag the
cable with excessive force, especially the cable of the semiconductor
WARNING
When the enclosure is opened, do not touch the boards, sockets, switch
and etc, when the system is powered on, otherwise, electric shock might
occur.
3.2.6.1 Introduction
The enclosure assembly supports and fixes the touchscreen, recorder, aspirate
button, other input and output components and etc. Optical window cover and lamp
replacement window cover are designed to facilitate the replacement of flow cell,
cuvette and lamp.
Enclosure
Aspirate button fixing screw
hole
1 If the system is configured with flow cell, before opening the enclosure, move the
flow cell out of the installing location and place it in the hole for holding flow cell
temporarily. Otherwise, the tubing connection will be dragged off.
2 When opening the enclosure, do not drag the cable with excessive force.
When installing the LCD/touchscreen module, pay attention to the following as shown
in figure 3-14.
1 When installing LCD cable, the blue face (untouchable face) of the FPC should
face outward;
2 When installing the LCD data cable of LCD&touchscreen board, the metal face
of the FPC cable (touchable face) should face outward;
4 When installing the supporting plate of the touchscreen, tighten the four screws
gradually one by one to avoid uneven force on the screen and excessive
tightening force on the screw which will result in false reaction of the
touchscreen.
The blue face The metal face The metal face Touchscreen
should face should face should face FPC cable
outward outward outward
4.1 Overview
This chapter describes the function of circuit boards in the BA-88A.
4 Hardware 4-1
The CPU communicates with the CPU extension board via CPU external interface in
multiple-buses communication mode.
4-2 4 Hardware
USB functions;
Data collection;
4 Hardware 4-3
USB power control and ESD protector is added between the USB controller and the
USB interface.
4-4 4 Hardware
4 Hardware 4-5
The control module of the CPU extension board (including: MCU and FPGA) outputs
motor control signal, which can control the step motor of the filter wheel via step
motor driver circuit.
The control module of the CPU extension board (including: MCU and FPGA) outputs
peristaltic pump control signal, which can control the motor of peristaltic pump via
power driver circuit.
The control module of the CPU extension board (including: MCU and FPGA) outputs
motor control signal, which can control the refrigeration plate via power driver circuit.
4-6 4 Hardware
The system should detect the signal from filter position sensor, aspirate button,
cuvette temp. controlling fan blocking. Before detected by control module, all input
signals should go through the level translation circuit.
Three signals are controlled by analogue board: digital potentiometer control signal
(DCPx), channel selection signal (CH), AD collection and control signal (AD_Ox and
AD_lx). All control signals are generated by control module. The signals between the
control module end and analogue board should go through level translation (TTL to
LVTTL).
The photodiode transforms the light signal which has passed through the flow
cell (or cuvette) into voltage signal. The voltage signal will be filtered,
amplified inversely, and finally output at the outputting end of the AD
converter through selectable switch.
4 Hardware 4-7
The AD converter receives control signal from the CPU extension board and
samples the processed photoelectric/temperature signal and sends the AD
value to CPU extension board for processing.
4-8 4 Hardware
The power board provides stable and reliable DC power to the modules and devices
of the system. The circuit is of Flyback converter structure, providing +24V, ±12V
4.8.2.2 Output
A5V output – for circuit board
Over-voltage protection
Over-heating protection
4 Hardware 4-9
1 Install all the boards back and ensure all the connections are well connected;
2 Set the digital universal meter to volt measurement. Connect the connectors of
the universal meter to LCD background cable connectors. (“+” is connected to
red cable; “-“ is connected to white cable). Turn on the power of the analyzer.
3 Observe the voltage result measured by digital universal meter, if the voltage is
within the range of 9.80V≤V≤10.10V, it meets the requirement; Otherwise, the
voltage should be adjusted by operating software. The procedure for adjusting is
shown as follows:
5 Click Optional to enter the advanced setting interface. Use “+” or “-“in Screen
brightness area to adjust the background voltage. Click “+” to increase the LCD
background voltage (LCD background is brightening); click “-“to decrease the
LCD background voltage (LCD background is darkening). The adjustment range
is: 9.80V≤V≤10.10V;
4-10 4 Hardware
4 Hardware 4-11
4-12 4 Hardware
4 Hardware 4-13
To ensure reliability, good performance and service life of the system, regular
maintenance is required. Be sure to follow the instructions given below to maintain
the system. Even you’re only an operator, it’s very important for you to learn this
chapter. Your thorough understanding will help you obtaining the best performance of
the system.
WARNING
Do not perform any maintenance procedures that are not described
in this chapter.
Do not touch the components other than the ones specified in this
chapter.
Performing unauthorized maintenance procedures may damage
your system, void any applicable warranty or service contract and
even cause personal injury.
After performing any maintenance actions or procedures, ensure that
the system runs normally.
Do not spill water or reagent on mechanical or electrical components
of the system.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles during
maintaining process.
2 Wash the tubing with distilled water or DI water before shutting down the
analyzer (for instrument installed with flow cell).
3 When the system is not in use, make sure that the tubing and the flow cell is
filled with clean distilled water (or DI water).
2、 If the instrument will not be in use for a long time. Detach the pump head from
the pump to elongate the service life of the pump tubing. Refer to the step 2 in
5.3.1 for details about detaching pump head. Do not disconnect the connection
of the tubing.
Contamination of the flow cell or bubble in the flow cell might cause low
background. You can use the absolute alcohol to clean the flow cell ( also
applies for bubble in the flow cell). The procedure is as follows:
A) Click Wash on the software to wash the tubing. After 10 seconds, click
Stop.
B) Prepare about 5 ml absolute alcohol in the tube. Put the aspiration tubing
into the tube and click Wash. After 5 seconds, click Stop. Wait for 10
seconds.
C) Wash with absolute alcohol for 5 seconds again, and then wait for 10
seconds. Wash the tubing with DI water for 10-20 seconds. The cleaning is
completed.
Before switching to other tests, it is recommended the flow cell be washed with
distilled water (or DI water). This is very necessary for tests which carryover
might take place, or samples whose concentration differs a lot. The cleaning
should last for no less than 5 seconds. The amount of distilled water (or DI water)
should be around 1.5 ml. You can also use the wash solution specifically for
chemistry analyzer to wash first, then with the distilled water (or DI water).
NOTE
Only the copy the software that needs to be upgraded to the
relevant directory. The upgrading time is in direct proportion to
the size of the file.
7 The upgrading time is in direct proportion to the size of the file. It will take 10
minutes to upgrade the operating software. After upgrading is completed, the
following interface is displayed, click OK. And the system will be restarted.
Remove the U disk.
If not, execute the CPU board software upgrading again. But only FPGA software or
lower computer software is needed to be copied the U disk.
Click Advanced to pop up the dialog box shown in the following figure. Enter the
password—analyzer and click OK.
Click start and a dialog box notifying the user to shut down lower computer
communication will be displayed.
Click Start and a dialog box notifying the user to shut down lower computer
communication will be displayed.
After upgrading is completed, a dialog box will pop up to notify the user to restart.
5.3 Maintenance
5.3.1 Replacing the tubing
The tubing is a component that should be replaced regularly. When the instrument is
in use, if you notice that the instrument can not aspirate or aspiration volume
decreases obviously, check for leakage in the peristaltic pump. If yes, the pump
tubing might be broken and it should be replaced. A pump tubing is in the accessory
package (outer diameter 3.2mm, length 150-160mm, yellow).The procedure for
replacing is shown as follows:
1 Shutdown the instrument and disconnect the power cable between the
instrument and the power supply.
Pinch the
buckles to
detach the
pump head
3 After detaching the pump head, unplug the tubing (yellow) whose adapters
are closer to the pump head.
Unplug the
tubing
(yellow)
whose
adapters
are closer
to the pump
head.
Pinch the
buckles on
the pump
head and
detach the
pump shell.
5 After detaching the pump shell, press the roller inside the pump to take the
used pump tubing out. Remove the red ring on the used tubing and install
them on the new tubing. Install the new pump tubing on the pump and
carefully pull it to make the tubing fit well with the inside of the pump without
any twisting. Make sure the lengths exposed at both ends are roughly the
same. If necessary, pull both ends of the tubing slightly.
Pinch the
buckles to lock
the pump head
7 After installing the pump head, connect the tubing with the tubing that goes
through the backboard and the waste tubing. The pump tubing on the left
should be connected to the tubing that goes through the backboard, the
pump tubing on the right to the waste tubing.
The pump tubing on
the left should be
connected to the
tubing that goes
through the
backboard.
8 After connecting the tubing, slip the tubing that goes through the backboard
into the analyzer until the adapters does not expose to the outside.
9 Check the installation for any errors. Connect the analyzer to the power
supply and calibrate the pump. Please refer to the relevant part of the
operator’s manual for calibration method.
NOTE
The service life of the pump tubing is 18 month. Tubing
should be checked irregularly and replaced in time.
Do not use excessive force while pulling, slipping the
tubing or doing any operation that might pulling the
tubing. Otherwise, the tubing might be damaged or the
inner connection might be cut off.
While connecting the tubing, insert the tubing into the
adapter until the tubing reaches the bottom of the adapter
so as to ensure the reliability of the connection.
While installing the pump tubing into the pump, avoid
twisting. If necessary, pull the tubing back and forth
slightly.
Calibrate the pump after replacing the pump tubing.
Otherwise, the aspiration volume might not be correct
and the test may be affected.
1 Open the optical window cover and press the bottom of the flow cell ( if flow
cell fastening screw is installed, you do not have to press), then carefully pull
the metal inlet tubing which is connected with the flow cell ( the slimmer one
which is closer to the front of the instrument)
Pull the
tubing
out
Aspiration
deflection
tube
BIOHAZARD
When replacing the pump tubing, wear gloves and lab coat and, if
necessary, goggles. In case your skin contacts the waste, follow
standard laboratory safety procedure and consult a doctor.
NOTE
After replacement is completed, please make sure that the flow cell
is pressed to the bottom of the installation hole (the top of the flow
cell should be on the same level with the installation surface).
Otherwise the test result might not be reliable.
NOTE
When connecting the tube, the tube should reach the root of the
metal tube to ensure reliability.
After the aspiration tubing is installed, the bending part should be
transitioned smoothly, no sharp angle, corrugation is allowed
Unscrew the
two screws
and remove
the
replacement
window cover
2 Unscrew the fastening screw of the lamp base (if it is difficult to dismount, some
tools like the clipper might be used). Put the screws in a proper place outside the
instrument. Pinch the lamp base directing line and then carefully take the lamp
assembly out).
Lamp
base
fastening
screw
Lamp
base
directing
line
Wire
joint
4 Connect the new lamp with the wire joint of the lamp assembly and wire joint in
the instrument. Install the lamp into the installing hole in the lamp base. Fasten
the retaining screw on the lamp base.
5 After installation, power on the instrument. Adjust the gain. Check the
background after the instrument is stabilized to check whether the replacement
is successful.
WARNING
If you want to replace the lamp, first switch off the power and then
wait at least 15 minutes for the lamp to cool down before touching it.
Do not touch the lamp before it cools down, or you may get burned.
NOTE
When replacing the lamp assembly, do not touch the surface of the
bulb.
When dismounting the restraining screw of the lamp base, please
take extra care to avoid element falling into the inside of the
instrument. If it is difficult to dismount, some tools like the clipper
might be used. The photoelectric gain should be adjusted after the
lamp is replaced. For more details, please refer to 5.3.2 Adjusting the
Photoelectric Gain.
1 Find the “sensor parameter A, B and C”, “filter wheel home position deviation”
and “340 nm peak wavelength” in parameter configuration based on the SN.
5 Adjust the photoelectric gain and refer to 5.1.3.2 for details; calibrate the pump
and refer to 5.1.3.3 for details.
NOTE
If the temperature sensor or analogue board has been replaced
before replacing the CPU board, enter new parameters after
replacement
If the home position sensor, motor, filter wheel and sensor stop
plate has been dismounted and repaired before replacing the
CPU board, enter new parameters after replacement.
This chapter presents all warning messages and recommended corrective actions,
which should be taken in time once any error occurs.
When an error or failure occurs, relevant alarm message will be displayed, and the
system will take corresponding actions.The logs will record the time, level, code and
detailed information of each warning to help user record and search errors.
In case of an error/warning message, check its error code on the Logs screen, and
then check the table below for recommended corrective measures.
WARNING
BIOHAZARD
6.2 Levels
Of all the system failures, the causes of many failures are similar. Classifying the
failure into different levels and allocating all failures to different levels will facilitate
troubleshooting and make failures of the same level have common corrective
measures.
All the failures recorded in the log can be classified into 3 levels.
6 Troubleshooting 6-1
The phenomenon and the error code will be displayed in the dialog box. Users can
use the error code to search the detailed information and corrective measures in the
logs.
The error information recorded in the logs includes: alarm date, alarm time, alarm
name, code, level, and details. The details include: error description, causes and
corrective measures.
6.3.2 Sound
The sound alarm is available on the system as well. You can enable or disable the
sound alarm in Setup.
6-2 6 Troubleshooting
6 Troubleshooting 6-3
6-4 6 Troubleshooting
6 Troubleshooting 6-5
The accidental
Filter wheel does not communication between the
Filter wheel does not
respond in maximum CPU board and MCU or
015 1 respond in maximum
delay and the failure FPGA is not normal. The
delay!
can be recovered system can be recovered
6-6 6 Troubleshooting