BCC3000 Training Material
BCC3000 Training Material
BCC3000 Training Material
BCC3000
Automatic Hematology Analyzer
Contents
Composition
Maintenance
BCC3000 Working Principle
WBC/RBC / PLT Testing- Impedance Method
Electrical Impedance Method is based on
the measurement of changes in electrical
resistance produced by a particle passing
through an aperture.
As blood cells are poor conductive, when
they pass through the aperture, they
cause an increased impedance, the
impedance directly proportioned to the cell
size.
Working principle
Size of Cells
An important feature of the electrical pulses is the relationship of
the pulse height to the size of the cell passing through the
aperture. Larger particles or cells generate higher pulse
amplitudes. Fortunately, each cell type falls within a predictable
range of sizes making it possible to distinguish one type of cells
from another. During cells sensing period, approximate normal
size ranges for the different cells are:
WBC’ s: 30-400 femtoliter
RBC’ s: 30-250 femtoliter
PLT’ s: 2-25 femtoliter
The number of pulses is equivalent to the cell number
through the aperture.
Reagent :
Diluent : diluting, rinsing, offering conductive
environment, cells equal volume processing
Quantitative tube
Hemoglobin Concentration Testing-Colorimetry
The diluted sample is mixed with lyse reagent, RBC dissolve, releas
ing hemoglobin.
Hemoglobin combined with lyse reagent to form hemoglobin complex.
The hemoglobin complex is irradiated by monochromatic light with a w
avelength of 546nm .
Phototube is used to receive transmission light, and convert the light sig
nal into voltage signal.
By comparing with the voltage produced by background transmission li
ght intensity before sample adding (only diluent exists), the hemoglobin
concentration can be obtained(HGB), unit is g / L.
Dilution rations :
Whole Blood Capillary Blood
WBC/HGB 1:308 1:428
RBC/PLT 1:44872 1:43355
Lyse Volume Whole Blood Capillary Blood
0.5 ml 0.36 ml
Parameters
Hematocrit HCT
Platelet PLT
Mean Platelet Volume MPV
Platelet Distribution Width PDW
Plateletcrit PCT
Platelet Larger cell ratio P-LCR
WBC Histogram
RBC Histogram
PLT Histogram
Storage: 1000 testing results, 570 QC results
Language: English
Operation Environment:
Temperature : 15℃ ~ 30℃ ;
Humidity : 30 %~ 85 %;
Air Pressure : 76kPa ~ 106kPa
Input/Output: RS-232, Internal thermal printer, 1 bar code reader (optional),
1 parallel printer(optional)
Data transmitting rate: 19200bps
Power supply: 100~240VAC , 50/60Hz
Fuse: 250V 2A
Power: 200W
Weight: 25Kgs
Dimension: 460(L)×390(W)×410(H)
Analysis and display range
parameter Display range Linear range Background range
WBC ( 109/ 0.0~99.9 0.0~99.9 ≤0.3×109/L
L)
RBC ( 1012/ 0.00~9.99 0.00~7.00 ≤0.03×1012/L
L)
HGB ( g/ 0~300 0~300 ≤1g/L
L)
MCV ( fL ) 0.0~300.0
4
3
2
5
View of left door open 1-Pressure bottle
2-negative pressure Pump
3-positive pressure pump
2
3
1--- valve 8 2---volumetric unit
3---vacuum chamber 4---valve 14
5---valve 13 6---valve 9
7---valve 12 8---valve 11
9---valve 10 10--- valve 2
11---2.5ml and 50ul motor
12---10ml motor
13---2.5ml syringe 14---50ul syringe
15---10ml syringe 16---valve 4
17---valve 3 18---valve 1
19---valve 6 20---valve 5
21---valve 15 22---valve 16
23---valve 17 24---valve 7
25--valve 18
BCC-3000B electrical board
Diluent 3.5 mL
Confirm Exit
Description and use of menu of instrument system check and diagnosis
Total 19 tests, such as counter hardware, machinery parts and liquid line
components, are tested by system, and it is shown that whether the component is in
normal state or whether the time of the test is in the range of time.
System testing is fluid path for a total of 19 tests showed or to the time required for
this operation. Check whether 2.5ml and 50ul, 10ml, deflection and up-down motors
run normally in all directions.
System Testing ~
2 、 QC
Return
Troubleshooting
Malfunction and resolvent
If analyzer detects some unnormal situations, the relevant malfunction
reminding will show on the screen. The following is the guidelines for
resolving problem.
crystal .
Fluctuating motor/ rotatory motor malfunction