Professional Documents
Culture Documents
TE 135 Service Manual
TE 135 Service Manual
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Table of Contents
1. General
2. Specifications
Book for 3. Special-purpose Jigs and Tools and Measuring Instruments Needed
Beginner 4. Description of Various Modes
Engineers 5. Operating Principles
6. Maintenance and Inspection
7. Replacing Procedures
8. Troubleshooting
8.1 Troubleshooting table by symptoms
8.2 Alarms and error messages
Books for 9. Maintenance Parts List
Senior
Engineers 9.1 Maintenance parts list
10. Disassembly, Reassembly, and Adjustment Procedure
10.1 Disassembly procedure
10.2 Reassembly and adjustment procedure
Note) Books for Senior Engineers are distributed to those who have been trained and passed
exams. Consequently, only qualified engineers are allowed to carry out repairs,
troubleshooting, etc. of the TERUFUSION Infusion Pump.
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
1. General
This Service Manual is written to provide the engineer with the on-the-spot maintenance and
servicing of TERUFUSION Infusion Pump (TE-135).
In the event that any complaints including failure and others are received from customers, in
principle, take actions in compliance with the following procedures.
(1) Before setting out to customers for emergency maintenance, carefully re-read the operating
instructions as well as this Manual and be thoroughly familiar with their content.
(2) Make sure that the customer uses and handles the apparatus strictly in compliance with the
operating instructions of the present apparatus.
(3) In the event that abnormal conditions result from erroneous handling, remove the causes,
and then, re-teach the customer how to correctly handle the apparatus and allow the
customer to be thoroughly familiar with the usage.
(4) Before undertaking emergency maintenance, reconfirm the content of customer complaints
and gain a correct understanding of abnormal conditions to start troubleshooting.
(5) In the event that the apparatus exhibits abnormal behavior even though the apparatus is
properly handled, be sure to carry out fault diagnosis strictly in compliance with the present
Manual.
(6) In the event that the apparatus is unable to be repaired by the use of this Manual, the
apparatus must be repaired at plant. State the trouble content and what have been done
for cures, and send back the apparatus to the plant.
(7) Disinfect as much as possible the replaced defective parts and others which came in contact
with patients’ bodily fluids, etc. and are returned to the plant. Hermetically seal and attach
a note of caution when any article is returned to the plant without disinfection.
(General precautions)
(1) Do not remove or insert connectors, printed circuit boards, and others with power supply
turned ON.
(2) First of all, visually check any sign of disconnection, short-circuiting, loose connectors, parts
breakage, and others, and at the same time, check for any abnormal noise and abnormal
smell (burning smell of resistors, etc.).
(3) Upon completion of inspection and repair, make sure all the switches are set to the specified
condition. Clean the apparatus and the vicinity, and report the results to the person in
charge.
(4) In the event that semiconductors, particularly, MOS-IC, are handled, take utmost care to
electrostatic discharge damage, etc.
* Ground human body, soldering iron, etc. via 1 MΩ resistance (this does not apply to
insulation type soldering irons but be sure to confirm insulation conditions.)
* When IC and printed circuit boards are transported, keep them in electrically conductive
containers or wrap them with sheets (aluminum foil, etc.).
(Avoid using plastics such as vinyl, styro-pads, etc.)
(5) In the event that printed circuit boards are replaced and trouble is corrected, return the
original boards and make sure whether or not the same trouble occurs. If the same trouble
does not occur, contact failure or others may be suspected. Further investigation is
required.
* In this Service Manual, the following SI units [the international System of Units, SI (System
International d’Unites in French)] are adopted.
SI units
Quantity
Name Symbol Remarks
Force Newton N 1 Kgf = 9.80665 N
2
Pressure Pascal Pa 1 Kgf/cm = 98066.5 Pa
1-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. Specifications
Model TE-135
Occlusion detection Adjustable between 30 and 140 kPa (0.30 - 1.43 kgf/cm2)
pressure (Ambient temperature : approximately 25ºC)
Alarms Battery, Air-in-line, Occlusion, Door, Infusion Completion(Keep Open Rate function),
Repeat(activated if the cause of the initial alarm remains for about 2 minutes),
Start Reminder(activated if about 2 minutes is wasted after infusion becomes available),
Battery Error, and D.Rate / D.Limit Comparison alarms; Empty, Flow error (free flow).
(An alarm during infusion is recognized by the relevant alarm indicator or display, the
Operation indicator, a buzzer, etc. The pump is stopped by any alarm other than the Battery
(except Shutdown and Infusion Completion alarms.)
2-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Power consumption 24VA(AC 100 to 120V), 28VA(AC 100 to 240VA), 11W(DC power)
2-2
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Nurse call output Connectable with nurse call via the computer interface / nurse call connector.
Contact capacity: DC12V, 1A
Units available
Reorder number Voltage Plug type Socket Computer interface
type
TE*135NX1 100 to 120V American type -
TE*135NW2 100 to 240V UK type -
TE*135NW3 100 to 240V European type -
TE*135CX1 100 to 120V American type Yes
TE*135CW2 100 to 240V UK type Yes
TE*135CW3 100 to 240V European type Yes
Conformity standard
IEC 60601: 1988, Amd.1 : 1991, Amd.2 : 1995
IEC 60601-1-1: 2000
IEC 60601-1-2: 2001
IEC 60601-2-24: 1998
MEDICAL DEVICE DIRECTIVE 93/42/EEC: 1993
2-3
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Designation Specifications
1) Digital volt-meter Voltage range: 20 V; Resistance range: 1Ω
2) Synchroscope Frequency range 500 kHz
3) TERUFUSION infusion set for pump TS*PA200L, TS*PA300L
4) Small-size syringe (MYJECTOR)
5) Pressure gauge Pressure range: 0 – 294.2 kPa (0-3.0 Kgf/cm2)
6) Occlusion adjusting jig 1.8 mm, 2.8 mm
7) Standard tool set
3-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Rotary SW
DIPSW(DS2)
DIPSW(DS1)
“Operation panel”
T TERUMO
AIR EMPTY FLOW ERR.
OCCLUSION BATTERY DOOR
MICRO COMPLETION
F STANDARD
( CLEAR
2 sec. ) “Volume delivered/Delivery limit” display
D.LIMIT Σ mL
INPUT
Σ mL
D.LIMIT
mL
D.RATE
INPUT “Flow rate” display
mL/h
PURGE STOP/
START
SILENCE
POWER
F E
・
・
AC/DC BATTERY
4-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
4-2
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
4-3
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Start Green
Date/time
Date is displayed.
Time is displayed only while “Function” switch is
held depressed.
PURGE STOP/
START
SILENCE
History No.
POWER History No. is displayed only while “PURGE”
F E
・
・ switch is held depressed.
AC/DC BATTERY
4-4
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
D.RATE y Delivery-limit
INPUT
Displayed only while “Delivery-limit setting” switch is held
mL/h
depressed.
(“Delivery-limit” display goes on.)
y Volume-delivered
Displayed only while “Volume-delivered” switch is held
depressed.
(“Volume-delivered” display goes on.)
PURGE STOP/
START
SILENCE
y Flow rate
POWER y Others
F E
・
・ Error displays, etc.
AC/DC BATTERY
Start-reminder alarm ( )
TERUFUSION INFUSION PUMP TE-135
Shock detection ( )
Shutdown ( )
Start/Stop Power switch operation ( ), ( )
Applicable lamp blinks.
(Goes out during quick delivery.)
Verification of operating time
y Displayed while “Start” switch is held depressed.
Power supply (History is not displayed during this period (lamp goes out).)
Goes on while AC or DC is being used. (Ex. in the case of 123,456 hours)
(Goes out when battery is used.) “Delivery-limit” display: ( )
“Flow rate” display: ( )
4-5
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Switch Description
MICRO MICRO MODE
0.1 to 99.9mL/h (0.1mL/h step)
STANDARD STANDARD MODE
1 to 999mL/h (1mL/h step)
y With DIP-switch DS1-4 OFF, the infusion mode cannot be switched to the
ON MICRO mode.
OFF 1 2 3 4 5 6 7 8
1 8
Change-over Occlusion detection pressure can be subtly adjusted in 10 stages by rotary switch
of occlusion on the main-body bottom.
detection
pressure
4 56
Alarm becomes Alarm becomes
2 3
7 8
4-6
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Volume
Mode DIP-SW(DS2) Flow rate display DIP-SW(DS2-4): ON DIP-SW: OFF
delivered display
Self-diagnosis Self-diagnosis check Regular mode
check mode OFF mode (standard
y When power is setting)
1 2 3 4
ON turned on, software
1 4 checking processing
is carried out. In
such event, regular
product software is
not affected.
(Default)
OFF
y The mode is not shut
down even when the
ON 1 2 3 4 pump is operated on
battery.
1 4
y Check mode is
changed over by
“Flow Rate Setting”
switch.
With “Flow Rate Setting” switch, following steps are implemented.
Delivery-limit/
Check and
Step Volume Flow rate display Description and operation
change items
delivered display
Initial display Step 0 Initial value y Start/stop operation is effective at each
display step.
Battery voltage Step 1 Battery voltage y Software is displayed in decimal
A/D value A/D value notation at last three digits of
(decimal notation) Delivery-limit/Volume delivered display.
(Ver 2.1 as of 2006.6.30)
**.* y Battery voltage A/D value is displayed in
decimal notation.
y Normal value:
With AC connected: 165-185
With battery connected: 100-140
Occlusion Step 2 Occlusion sensor y Output voltage A/D value of occlusion
sensor A/D voltage A/D value sensor voltage is displayed in decimal
value (decimal notation) notation.
y Output value: 70-110 (water-filled tube)
Air-in-line Step 3 Occlusion sensor y Output voltage A/D value of air-in-line
sensor A/D voltage A/D value sensor voltage is displayed in decimal
value (decimal notation) notation.
y Output value: 100 or more
(water-filled tube)
Temperature Step 4 Temperature y DIP-SW setting is displayed in
sensor A/D sensor voltage hexadecimal notation at last three digits
value A/D value (***) of Delivery-limit/Volume delivered
*** (decimal notation) display.
y Output voltage A/D value of temperature
sensor is displayed in decimal notation.
y Normal value: 40-65
4-7
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Delivery-limit/
Check and
Step Volume Flow rate display Description and operation
change items
delivered display
Total rotating Step 5 Total rotating y Total rotating count of cam shaft is
count of drive count display displayed (last three digits).
(low-order) y Last three digits of 0-999999 rotations
(decimal notation) are displayed.
Step 6 Total rotating y Total rotating count of cam shaft is
count display displayed (middle three digits).
(mid-order) y Middle three digits of 0-999999 rotations
(decimal notation) are displayed.
Step 7 Total rotating y Total rotating count of cam shaft is
count display displayed (first three digits).
(high-order) y First three digits of 0-999999 rotations
(decimal notation) are displayed.
Operating time Step 8 Operating time y Total operating time of drive is displayed
(low-order) (last three digits).
(decimal notation)
Step 9 Operating time y Total operating time of drive is displayed
(high-order) (first three digits).
(decimal notation)
Number of Step 10 Number of times y Number of times data is written to
times data is data is written to EEPROM is displayed (last 3 digits).
written to EEPROM
EEPROM (low-order)
(decimal notation)
Step 11 Number of times y Number of times data is written to
data is written to EEPROM is displayed (middle 3 digits).
EEPROM
(middle-order)
(decimal notation)
Step 12 Number of times y Number of times data is written to
data is written to EEPROM is displayed (first 3 digits).
EEPROM
(high-order)
(decimal notation)
EEPROM data Step 13 EEPROM data y EEPROM address is displayed in
(hexadecimal Delivery-limit/Volume delivered display
notation) (hexadecimal notation).
y Last two digits (**) of designated
address data are displaced at flow rate
display (hexadecimal notation).
** y First two digits (**) of designated
address data are displaced at number of
drops display (hexadecimal notation).
<Change of address by operation of
setting switches>
y Confirmation of EEPROM data.
Key entry data Step 14 Switch entry code y Pressed switch other than “Flow rate
setting” switch is displayed (decimal
notation).
y State of infusion set display selector
switch and rotary switch is displayed at
number of drops display.
4-8
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Delivery-limit/
Check and
Step Volume Flow rate display Description and operation
change items
delivered display
Communication Step 15 Communication y Setting related to communication is
setting setting results made.
y Setting of blinking digit can be changed
by operating [UP] / [DOWN] switches
(low-order).
Baud rate: 2 ...19200
1 ... 9600 (default value)
0 ... 4800
Parity: 2 ... Even
1 ... Odd
0 ... None (default value)
Stop bit: 1 ... 1 bit
0 ... 2 bits (default value)
y Operating [UP]/[DOWN] switches
(high-order), set items (digits) can be
changed. (Blinking digits are
changed.)
y Pressing “PURGE” switch stores the
setting currently displayed (3 digits). In
such event, a buzzer comes on for 1
second.
y Pressing “Flow rate setting” switch
moves operation to the next step.
y In the event that operation is moved to
the next step without carrying out
storage operation, settings are not
changed.
Step 16 Last three digits of y Setting related to communication is
apparatus made.
address (decimal y Setting of blinking digit can be changed
notation) by operating [UP]/[DOWN] switches
y When moved (low-order).
from last y Operating [UP]/[DOWN] switches
three digit (high-order), set digits can be changed.
display to (Blinking digits are changed.)
middle digits: Low-order digit high-order digit:
[1] [2] [3]
High-order digit low-order digit:
y When moved [3] [2] [1]
from middle y Pressing “PURGE” switch stores 8-digit
three digit setting of first 2 digits, middle 3 digits,
display to and last 3 digits which have been set.
high-order In such event, a buzzer comes on for 1
digits. second.
y In the event that operation is moved to
the next step without carrying out
storage operation, settings are not
changed.
4-9
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Delivery-limit/
Check and
Step Volume Flow rate display Description and operation
change items
delivered display
System Step 17 y Pressing “FUNCTION” switch changes
reservation Delivery-limit/Volume delivered display
. to “b000 b001 b002 b003
b000 ...”
y Make sure all are [1.00].
Air-in-line Step 18 y To display the value set at “4.2.1
detection Air-in-line detection sensitivity
sensitivity change-over function.”
setting y Make sure either [5] or [10] is displayed.
Buzzer sound Step 19
y Selection of buzzer sound (setting can
selection
be changed).
1: Continuous sound
0: Intermittent chords (default)
4-10
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Delivery-limit/
Check and
Step Volume Flow rate display Description and operation
change items
delivered display
System Step 22
0: (default) Do not change.
reservation
4-11
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
5. Operating Principles
Motor rotation
detection circuit
Drip sensor CPU monitor circuit
Air-in-line detection
Reinforced insulation circuit Motor drive circuit Motor
Occlusion detection
circuit Buzzer sound circuit Buzzer
*1
Power
supply
Power supply circuit Main unit
# treat IV line attached to main unit as attached part. (Specified in JIS T 1653(1991)8.2(1))
5-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
5-2
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
+12 V to 15 V GND
GND
Main board
(2) Power supply to circuits
Power is supplied to circuits via FET by the power input section (switching power supply, external
DC, and battery). Power supply to control circuits is further stepped down to +5 V by a DC/DC
converter.
+5 V Switching
+ Switching
circuit
regulator
-
3-terminal regulator AC/battery
RTC (realtime clock) discrimination FU1
backup Noise
OUT IN filter +
Memory backup External
Battery monitor IC - DC
Main board
5-3
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
+5 V (Normally supplied)
Main CPU
R86
DC/DC Q30
3-terminal +5 V
regulator converter
+5 V R87
Q D
CK
Q
IC20 C64
R S
5-4
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
* Maintenance parts are shipped from the factory with CN11 turned OFF. When CN11 is kept in
the ON state on the main board alone, the lithium battery lasts for about 8 years.
* Backup by the lithium battery is carried out only when connector CN3 is connected or CN11 is
short-circuited on the ON side. To retain the clock or history data when the main board is
detached, make sure CN11 is connected to the ON side and then, detach CN3.
* When the main board is replaced and mounted, turn ON CN11 and connect CN3, and then,
turn ON the power supply to start up CPU once.
IC39
3-terminal Battery
OUT IN regulator
Lithium
battery 3.2 V (1) (1)
CN11 CN3 CN11 CN3
Main Board
5-5
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
AC/DC
R12 IC1
Load
R13
IC40
Charge Constant- Vcc Communication
circuit voltage circuit
Main
Display control CPU
R155, R156 GND
Battery
COM
Battery voltage
detection
R157
80%
Current 40%
detection 0%
I Main
BAT V
BAT
CPU
R154
Main and display board
IC40 has a function to learn the capability of battery capacity, and when the battery capacity drops
due to battery deterioration, all three battery residual power lamps may not go on even when the battery
is fully charged. When battery connectors are mounted or detached for battery replacement, etc.,
inspection set forth in Section 6, Maintenance and Inspection is required for accurate residual power
indication, because it is indicated on the basis of “nominal capacity” of the battery.
CPU carries out the following operations by communication with IC40.
[1] Battery trouble detection: When following conditions are detected, CPU allows three
“Battery” lamps to blink simultaneously.
y Abnormal charging current
y Short-circuited cell
y Battery life
Buzzer sound: none
[2] Low battery detection: When residual battery capacity is detected to be about 15% or
less, CPU activates the low battery alarm.
Indication: “BATTERY” blinks.
Buzzer sound: Changes from intermittent sound to continuous
sound in about 10 minutes.
[3] Shutdown detection: When voltage per one battery is detected to be not more than 1.05
V (1.00-1.10 V), CPU activates the shutdown alarm.
Indication: [S-d] goes on.
Buzzer sound: After the buzzer continues sounding for 3 minutes
at maximum, CPU shuts down the infusion pump.
5-6
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Constant current
Constant current
charging circuit
charging circuit
I1 I2
Timer
Battery monitor IC
Trickle-charging current: I 1
5-7
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Memory
Motor
drive IC
History
Time is up.
Main Board
Watchdog “time’s
up” signal
5-8
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Reset
Motor
Main CPU
IC30 IC29
5-9
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
* The relay contact capacity is 1 A at DC12 V. Take utmost care to prevent relays from
exceeding the capacity when they are connected to any external loads.
+5 V
1
4
RY1
Main CPU control Nurse call
6 output
signal under
normal condition 9
CN2
Main CPU
runaway stop
signal
Main CPU
Watchdog “time’s
up” signal
Main board Communication board
5-10
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
In addition, in the event that runaway stop/watchdog “time’s up” signal occur irrespective of main
CPU condition,
buzzer drive voltage is fixed to +5 V and
drive frequency to 3.9 KHz, and
the buzzer sounds.
Q14 BZ
Buzzer circuit
3.9 KHz
Main Board
5-11
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
y It is possible to confirm operating conditions, various alarm conditions, and others by the use of
external communication (RS-232C or RS-485).
y The pump is connected with external medical monitors by a 3-conductor system comprising
two data wires and one ground wire, and there is no control signal.
Data
y The pump does not transmit data unless a command is provided from the outside.
y For data reliability and safety, a “reverse nibble format” is adopted to carry out data check for
each byte (nibble means 4 bits).
y When multiple infusion pumps are used or TERUFUSION infusion pump is used with other
apparatus, the individual pieces can be identified by setting ID number (apparatus No.: 2 digits;
serial No.: 8 digits).
* Apparatus No. of 2 digits become a fixed value that represents TE-135.
y In the event that no suitable connection cables are commercially available, they must be
fabricated to order by cable manufacturers, medical monitor manufacturers, and others.
In the event that operation is confirmed on an experimental level, commercially available cables
may be modified to connect.
y In the event that multi-drops (RS-485) are carried out, for example, wire as follows:
To pump
Communication cable
KSP-3104
External
medical
monitor, etc.
5-12
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
By setting DIP-switch DS1 on the communication board, either RS-232C or RS-485 can be
chosen for communication specifications.
DS1
1 2 3 4 5 6 7 8
RS232C ON ON ON OFF OFF OFF OFF OFF
RS485
OFF OFF OFF ON ON ON OFF OFF
(with no terminating resistor)
RS485
↑ ↑ ↑ ↑ ↑ ↑ ↑ ON
(with terminating resistor)
*1. The 8th switch of DS1 is turned ON when terminating resistor is used at RS-485.
*2. To choose RS-232 or 485, change over DS1 by a small flat-head screwdriver or a hex key.
<RS-232C specification>
DS1 RS-232C driver CN2
1 Rin 1 Vcc
1
RXD 2
2
Tout Tin 2
TXD 3 3 RXD
3
4 Rout 4 TXD
4
GND 5 5 RE DE
5
6 6 GND
6
7
7
8
8
9
Communication connector
<RS-485 specification>
DS1 RS-485C driver CN2
1 1 Vcc
1
Transmission and
2 2
reception + 2 D1
Transmission and
3 3 RXD
reception - 3
4 A RE 4 TXD
4
GND 5 5 RE DE
5 B DE
6 6 GND
6
7
7
Ro
8
8
Terminating
9 resistor
Communication
connector Communication board
5-13
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
+5V
8 data lines L
7 scan lines L
Goes Goes Goes
ON OUT OUT
CPU
Sensor input
5-14
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
* The shock sensor is intended to function when the main body falls from altitude about 1-meter
high. In the event that the shock sensor detects any shock, no audible alarm is activated but
the history can be checked in the history mode (see Section 4.2.1 for history function).
+5 V
(normally entered)
+5 V
+5 V (normally entered)
(normally entered)
-
-
- +
+
+ IC18
IC19 IC18 (1)
(2)
S
Main CPU Q IC20
R
Shock
sensor reset
5-15
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
5V >1.8 V
0V 0V
0V 0V
AMP
One- - G=100 G=8
shot + Light receiving
element
Reference voltage
Comparator
Sample/hold circuit
Light
emitting
element
generation
Pulse
978 Hz
31.25 KHz
160 µs
5V
DROP PROBE01
Connection check signal
5-16
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
voltage circuit
Constant-
Tube
Differential
AMP
+
-
Buffer IC2
Main CPU
Occlusion
Hall effect plunger
element
Sensor output
characteristics Occlusion board
Output
voltage
Occlusion sensor
Plunger position
Temperature
sensor
Vcc
To Main CPU
OUT
Temperature
detection
GND
5-17
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Oscillation
circuit Waveshaping 1/2 division
3.2 MHz Door unit
Transmission
Tube
Reception
Smoothing
circuit Air-in-line sensor unit
AMP
Main CPU VR1
Monitoring speed of
rotation
To main CPU Photo interrupter
Detecting
high-speed rotation Photo interrupter
To main CPU
Reflector
Main Board Rotation sensor board
5-18
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Vcc
To Main CPU OUT
Door open
detection
GND
5-19
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
5-20
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
This section describes maintenance and inspection of TE-135. You are urged to periodically
carry out the matters to be set forth hereunder and to detect trouble early and take corrective actions.
6.1 Maintenance
6.1.1 Infusion set used for maintenance and inspection
Caution y When maintenance and inspection is undertaken, use TERUFUSION infusion sets for pump.
y To prevent danger of infections, etc., use sterilized infusion sets and others. In addition, do not
reuse infusion sets but use them only once because reused infusion sets would affect the
measurement accuracy, too.
6.1.2 When fall and impact are applied to the infusion pump:
Caution Even if no abnormality is observed in main body appearance and operation, immediately stop using
the infusion pump, inspect in accordance with Section 6.2 Periodical Inspection, and ensure the
safety.
Number of years
Parts name Reasons for replacement
elapsed
Battery 1.0 to 1.5 years y Persistent “Battery” alarms blink and audible alarms are
activated in a short time despite repeated full charges.
(Good-rule-of-thumb of next replacement time is affixed
to the main body.)
Motor unit 2 to 3 years y Flow rate trouble occurs and abnormal noise is
generated during operation.
Pole clamp 2 to 3 years y Discoloration and damaged portions by solution are
found.
y Infusion pump is unable to be securely fixed to the pole.
Door seal rubber 2 to 3 years y Breakage and deformation are present.
Drip sensor unit 2 to 3 years y Flow rate trouble occurs.
y The unit is unable to be securely mounted to the drip
chamber.
* The replacement time varies depending on use frequency and use environment of each unit.
6-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
1. Cut a part of an empty infusion bag with a retractable knife and pour 200 mL of water into the
bag.
2. Mount the water-containing bag to the infusion stand.
3. Close the roller clamp of the infusion set with a needle cannula connected; then, insert the
cap spike very far into the insertion port.
4. Gradually unclamp to allow the water to prime through the entire infusion set and close the
roller clamp. In such event, make sure there is no air bubble in the tube. If any air bubble
is present, unclamp and continue discharging water to remove air bubbles.
6-2
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Inspection procedure]
1. Make sure main body exterior, drip sensor (TE*975), pole clamp, and AC power cord are
free of any breakage and others.
2. Make sure the infusion pump can be operated on battery power supply.
3. Turning ON power supply with the door open causes the infusion pump to carry out the
following self-check operations.
y All the displays blink three times and fingers slightly move.
y [Door], [Air], and [Occlusion] indicator lamps blink.
4. When AC power cord is connected, the “AC/DC” lamp does not go on with the power switch
not turned ON.
5. When the infusion pump is operated on AC power supply, “AC/DC” lamp goes on.
6. When the tube clamp is released and the door is closed and opened, the tube clamp unit is
closed.
7. “Start” switch and “Stop/Buzzer Silence” switch can be operated.
6.2.4 Tube clamp function inspection method (Once in every two months)
Once every two months, inspect tube clamp functions in compliance with the following procedure.
[Inspection procedure]
1. Mount the prepared infusion set to the infusion pump, close the door, and open the roller
clamp.
2. Turn on power supply and carry out quick-delivery until water comes out from the tip end of
the infusion set (priming).
3. When fingers are released from the “quick-delivery” switch, make sure no water trickles at
the tip end of the infusion set and drip chamber.
4. When the door is opened, make sure no water trickles from the tip end of infusion set and
from the nozzle of the drip chamber.
6-3
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Inspection procedure]
1. Make sure the rotary SW on the main body bottom is set to “5.” If it is not “5,” after
recording the number once set, rotate the rotary SW to change the number to “5.”
2. Turn on power supply.
3. Mount the pressure gauge with a water containing tube connected to the main body in such
a manner that the pressure gauge comes about 100 cm below the main body, and close the
door.
4. Set the flow rate to 100 [mL/h] and the delivery limit to “...” (no setting), press the “start”
switch, and operate the infusion pump until the “occlusion” lamp blinks, the buzzer comes on,
and the pump operation is stopped.
5. Make sure the detected pressure value is included within the judgment value of the table
below when the “occlusion” lamp blinks, the buzzer comes on, and the pump operation is
stopped, and record the result. In the event that the detected pressure value is not
included in the judgment value at one try, change the tube clipping position or replace the
tube, and carry out the inspection several times, and make sure the reproducibility of
detected pressure value is obtained.
Occlusion detection
Flow rate setting Delivery-limit pressure value
2
kPa ([Kgf/cm ])
“----” 60 to 120
100 mL/h
No setting (0.61 to 1.22)
6-4
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
6.2.6 Inspection method of flow rate accuracy (Once in every two months)
Inspect the flow rate accuracy in compliance with the following procedure once every two months.
[Inspection procedure]
1. Turn on the power supply.
2. Mount a prepared infusion set to the infusion pump.
3. Attach a needle cannula to the infusion set tube end.
4. Close the door and loosen the roller clamp.
5. Deliver until water comes out from the head end for priming.
6. Press the “Volume delivered” switch to set the volume delivered to “0” and set the flow rate
to 120 [mL/h] and delivery-limit to 20 [mL].
7. Place the head end of the infusion set into a graduated cylinder, press the “START” switch to
start infusion, and at the same time, start the stopwach.
8. Upon completion of infusion, measure the volume of water in the graduated cylinder.
50 to 70 cm
50 to 70 cm
Measurement
Flow rate setting Delivery-limit Rated value
time
120 mL/h 20 mL 10 minutes 18 to 22 mL
6-5
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Inspection procedure]
1. Mix about 15 mm air bubbles in the prepared infusion set.
2. Loosen the roller clamp.
3. Turn on the power supply.
4. Set the flow rate to 300 mL/h, and delivery-limit to “---.”
5. Press the “START” switch to operate the pump.
6. Make sure the audible “Air bubble” alarm is activated. In addition, open the door and make
sure air bubbles are present in the vicinity of the air-in-line sensor.
6-6
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Inspection procedure]
1. With the power supply turned off, connect an AC power cord to the infusion pump (in such
event, make sure the “AC/DC” lamp goes on) and charge the battery for at least 15 hours.
2. Mount the prepared infusion set to the infusion pump, set the drip sensor, close the door,
and loosen the tubing clamp.
3. Turn on the power supply.
4. Set the flow rate to 25 mL/h, and delivery-limit to “----,” and press the “START” switch to start
infusion.
5. Make sure the infusion pump operates for not less than 120 minutes after the AC power cord
is disconnected. Thereafter, let it stand until the power supply is turned off, and when the
power supply is turned off, connect the AC power cord again, charge the battery for more
than 15 hours, and use the infusion pump. When the operating time before the power
supply is turned off is 120 minutes or less, replace the battery.
6-7
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Inspection procedure]
1. Make sure the drip sensor is not contaminated with solutions, etc., and if contaminated, wipe
away stains with gauze, etc. immersed with lukewarm water.
2. Mount a prepared infusion set to the infusion pump and close the door.
3. Set the drip sensor to the drip chamber of the infusion set and loosen the roller clamp.
4. Set the “[MICRO/STANDARD] switch to “STANDARD,” flow rate to 100 mL/h, and
delivery-limit to “---.”
5. Make sure the solution drips in the drip chamber and every time the solution drips, the green
lamp of drip sensor blinks.
6. Make sure [FLOW ERR.] alarm blinks, buzzer comes on, and the pump stops when the drip
sensor connector is connected or disconnected during operation.
Note) For safer and longer use of the infusion pump, carry out periodical inspection once a year as a
good-rule-of-thumb. When periodical inspection is carried out, check and replace periodically replaced
parts.
6-8
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
6.3 Cleaning
Caution Before cleaning, be sure to turn off the “Power” switch and unplug the AC power cord.
Caution y Do not wipe with organic solvents such as alcohol, thinner, etc.
y Before undertaking cleaning, be sure to turn off the power supply and detach AC power cord.
Caution Never attempt to rub with something hard or sharp, such as forceps made of metal, etc.
6-9
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
[Cleaning method]
In the event that the outer slide case drags or is excessively contaminated, disassemble and clean
the drip sensor. To detach the drip sensor outer case unit from the drip sensor unit, disengage fixing
pawls while pressing the slide case with fingers from right and left (for the detail, see Section 7.1
Replacement Procedure).
Wipe the drip sensor main unit with gauze, etc. soaked with lukewarm water. Wash the slide
case and drip sensor spring and thoroughly dry, and then, reassemble. For reassembly, set the drip
sensor spring in the slide case and push the slide case until the main body stopper fits into the slide case
groove.
* The slide case can be removed by allowing the stopper portion to suspend while pressing the slide
case from right and left without using a screwdriver.
Flat-head screwdriver
y After protecting the case with soft cloth, gauze, y Since this rib serves as a stopper, lift the slide
etc., lift it up with a screwdriver. case in such a manner as to detach this.
6-10
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
6-11
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Check items
1 Is there any breakage or other trouble in infusion pump main unit or pole clamp? OK NG
3 Does the infusion pump properly self-check when power switch is turned on? OK NG
4 When AC power cable is connected, does “AC/DC” lamp go on even with power OK NG
switch not turned ON?
6 When the door is closed and opened once, is the tube clamp closed? OK NG
12 Is infusion pump able to be operated for at least 120 minutes with battery (for nor OK NG
less than 120 minutes at flow rate of 25 mL/h)?
Remarks
Warning: Should any abnormality be found, immediately stop using the infusion pump.
6-12
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
7. Replacing Procedures
Caution When the following maintenance parts are replaced, adjustment is required. Be sure to carry out
adjustment in compliance with the specified procedure.
Caution The detailed maintenance parts list will be distributed when senior engineer training is completed.
Unless you are qualified for senior engineer, you are not allowed to procure maintenance parts and
distribute manuals of disassembly, reassembly, adjustment procedures, and others related to
maintenance parts.
7-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. After replacement, connect the drip sensor to the drip sensor connector on the back surface of the
infusion pump.
7-1-(1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution When the slide case of drip sensor is removed, take utmost care to prevent missing of the drip
sensor spring.
2. Set the drip sensor spring to the drip sensor spring fixing section located inside the new outer case,
insert the drip sensor main body with care to prevent the spring from collapsing, and fix with the
pawl of the drip sensor main body.
* Method to remove the drip sensor outer case unit without using a screwdriver
TE-135 drip sensor can be removed without using a screwdriver.
Press the outer case from right and left with fingers and pull out the slide case while slightly
twisting the outer case so that it suspends at the pawl portion of the main body.
7-1-(2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution When the fuse holder is removed, take utmost care to prevent damage to the fuse holder, infusion
pump, and others.
2. Take out the fuse from the fuse holder and check for any sign of disconnection.
3. If disconnection is suspected, replace with two new [32] fuses.
4. Mount the fuse holder to the AC inlet, press the “POWER” switch after connecting the AC power
cable to the infusion pump, and make sure the power is turned on.
Caution When the fuse holder is mounted, make sure the fuse holder is securely locked (prevent pawls from
being suspended).
7-1-(3)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Battery cover
unit [7]
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer Bolt M3-L8)
Screw tightening torque : 60 to 69 cN.m
2. Take out [24] battery set in the infusion pump battery housing, release the connector, and replace
with a new battery.
Caution y After replacing the battery, be sure to fully charge the battery (charge for more than 15 hours with
the power supply tuned OFF) after complete discharging.
y When the battery is replaced, be sure to replace the battery replacement recommendation label,
too. When the battery is replaced, peel off the battery replacement recommendation label
affixed before replacement and affix a new battery replacement recommendation label.
7-1-(4)
8. Troubleshooting
SY14-9B002
Lot No. 0607 –
Applicable Lot
matter
SY14-9B002
Lot No. 0607 –
Applicable Lot
is not outputted across
drain and GND of
6P of main board.
6. Breakage or dislocation of 6. Visual 6. Replace [26] timing belt.
timing belt
9 • Er5 is displayed. 1. Trouble of EEPROM 1. Er5 is displayed. 1. Replace [6] main board.
10 • Er7 is displayed. 1. Runaway of main CPU 1. Er7 is displayed. 1. Turn on power supply
again or replace [6] main
board.
11 • Er8 is displayed. • If [01] through [14] are not
displayed in response to
the switch when switch is
pressed in self-diagnosis
mode [9000], confirm the
SY14-9B002
Lot No. 0607 –
Applicable Lot
following:
SY14-9B002
Lot No. 0607 –
Applicable Lot
disconnected.
SY14-9B002
Lot No. 0607 –
Applicable Lot
2. In the case of other than 2. In the case of other than 2. In the case of other than
SY14-9B002
Lot No. 0607 –
Applicable Lot
1. Defective tact switch 1. Check continuity (display 1. Replace [5] display
boards CN1-9, 11P) board.
SY14-9B002
Lot No. 0607 –
Applicable Lot
Prepared in Feb., 2007
Operation/check Symptoms Possible causes Determining method Remedial method
25 Set flow rate to 120 mL/h and • Does not start. • If [03] is not displayed when
delivery-limit to 20 mL, and switch is pressed in
hold “START” switch self-diagnosis mode [9000],
depressed. confirm the following:
SY14-9B002
Lot No. 0607 –
Applicable Lot
3. If other than the above, see 3. If other than the above, see 3. If other than the above,
SY14-9B002
Lot No. 0607 –
Applicable Lot
tube and nor less than
about 0.2V for empty tube
SY14-9B002
Lot No. 0607 –
Applicable Lot
IC1-78, 79, 80, 91P
SY14-9B002
Lot No. 0607 –
Applicable Lot
4. Broken main board 4. Defective if no data IO is 4. Replace [6] main board.
2. Defective battery charging 2. Defective when battery 2 Replace [6] main board.
circuit charging current is not from
100 to 250 mA.
3. Reset of battery monitor IC 3. When power supply is 3. Charge battery for 15
turned on, three battery hours in compliance with
LEDs blink. “Section 6.
Maintenance,
Inspection, and
Cleaning.”
SY14-9B002
Lot No. 0607 –
Applicable Lot
Prepared in Feb., 2007
SY14-9B002
Applicable Lot
Lot No. 0609 –
Prepared in Nov., 2006
8.2.1 Alarms
Alarms Description
End of infusion y Alarm is activated when volume delivered exceeds delivery-limit.
Alarm is automatically deactivated when delivery-limit exceeds volume delivered.
Air-in-line alarm y Alarm is activated when air bubble that passes the sensor is detected by air-in-line sensor.
Occlusion alarm y Alarm is activated when increase of inner tube pressure on the downstream side of pump
is detected when solution is being delivered by “START” or “Quick Delivery.”
Door alarm y Opening or closure of door is detected by the door sensor, and alarm is activated when the
door is held open.
Flow error alarm(*) y Alarm is activated when flow error (free-flow, empty container) is detected by the drip
Empty container alarm(*) sensor.
y The condition in which trickling is detected while the pump is stopped is judged as
“free-flow.”
y The condition in which no flow of solution drops is detected during start is judged as
“empty container.”
y Following conditions are exhibited when alarm is activated.
Low battery alarm y As the internal battery charge decreases, “BATTERY” display blinks and the audible alarm
is activated. The audible alarm sound changes from intermittent to continuous as time
passes. In addition, the operation indicator does not show any change.
(*)Alarm when using drip sensor (option).
8-13
9. Maintenance Parts List
This section shows a table of parts that compose TE-135/135C. When maintenance is carried out, see the following table and “Section 10.
Disassembly, Reassembly, and Adjustment Procedure,” and take appropriate measures.
* “No.” on the left side in the table coincides with “No.” in “Section 10. Disassembly, Reassembly, and Adjustment Procedure.”
SY14-9B002
Lot No. 0607 –
Applicable Lot
• Door sensor board [23]
3 Door unit (for TE-135) - 06PC503 0607 lots - • Door case unit [4]
• Display board [5]
• Flexible cable unit [9]
• Crescent hook unit [11]
1
• Air-in-line sensor unit [18]
• Door seal rubber [27]
• Door body unit [51]
• Buffer plate unit [53]
4 Door case unit (for TE-135) - 06PC504 0607 lots - • Switch spacer [56]
1
• Front panel [57]
5 Display board - 05PB05 0607 lots - 1
11 Crescent hook unit - 05PB11 0607 lots - • Crescent hook (B) [12] 1
SY14-9B002
Lot No. 0607 –
Applicable Lot
14 Infusion pump cover unit - 06PC514 0607 lots - • Door seal rubber [27] 1
29 Cam shaft retainer set - 05PB29 0607 lots - 1 Two in one set
SY14-9B002
Lot No. 0607 –
Applicable Lot
37 Drip sensor outer case unit - 06PC536 0607 lots - • Outer case 1 Dedicated for TE*975
51 Door body unit - 05PB51 0607 lots - • Door seal rubber [27] 1
58 Connector for main board - 05LD58 0607 lots - • Connector for flexible 1 TE-171/172 in common
flexible cable cable For flexible cable
(on main board side)
59 Connector for display board - 05LD59 0607 lots - • Connector for flexible 1 TE-171/172 in common
9-4
SY14-9B002
Lot No. 0607 –
Applicable Lot
-- Pressure gauge - 05LD371 0607 lots - 1 TE-171/172 in common
SY14-9B002
Lot No. 0607 –
Applicable Lot
Prepared in Feb., 2007
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
10-1
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. Then, detach [8] drip-proof cap from the fixing pawl inside the case unit, and remove the drip-proof
cap from the case unit.
Caution When the drip-proof cap is detached, take utmost care to prevent damage to the drip-proof cap.
10-1-1 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping screws, nominal
3-L10 × 2 pcs., inner clip washer × 2 pcs.)
2. Detach the connector (CN2) on the communication board side and remove from the infusion pump
chassis.
Communication
board [34]
10-1-2 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive
spring washer and plain washer M3-L5)
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L6)
Rotation sensor
board [25]
10-1-3 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Connector
(CN2, 7, 9) Connector
(CN1)
2. Detach screws (Cross-recessed head tapping screws, nominal 3-L6 × 2 pcs.) that fix [6] main
board.
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L6 × 2 pcs.)
10-1-4 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. Slide [6] main board to the upper side of infusion pump and lightly lift up, and detach harness
connectors (CN3, 4, 5) connected to remove the main board.
10-1-4 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Connector (CN1)
2. After detaching screws (Cross-recessed head tapping screws, nominal 3-L5, nominal 3-L6) that fix
the switching regulator, lift up and remove.
Cross-recessed head
tapping screws Connector
(nominal 3-L5) (CN1, 2)
10-1-5 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Connector
(CN2)
2. Remove screws (flat countersunk head screw M3-L6 × 2 pcs.) that fix [22] AC inlet unit.
Fixing screw
(flat countersunk head
screw M3-L6 2 pcs.)
10-1-6 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. After lightly lifting [22] AC inlet unit to the infusion pump chassis back surface side, pull in the
lateral direction (main board fixing side) to remove.
Removing direction
10-1-6 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping
screws, nominal 3-L8 × 2 pcs.)
2. Open the door and remove screws (Cross recessed head screws with captive spring washer and
plain washer M4-L16 × 2 pcs.) that fix [4] door unit.
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M4-L16 2 pcs.)
10-1-7 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive
spring washer and plain washer M3 × L10)
10-1-8 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping
screws, nominal 3-L8 × 2 pcs.)
2. Lightly lift [18] air-in-line sensor unit, disconnect air-in-line sensor unit connector (CN1), and
detach the unit from the door body.
* [18] air-in-line sensor unit can be removed even after the door case is separated from the door body.
The procedure after separating the door case and the door body is described in Section 7.2.10
Separation of door case and door body.”
Caution When the air-in-line sensor unit is removed, take thoroughgoing care to disconnect connected
harness.
10-1-9 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L8, Cross recessed head screws
with spring and plain washer M3-L16)
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L6)
2. Remove the screws that fix the door case unit to the door body unit (Cross recessed head screws
with captive spring washer and plain washer M3-L8 × 3 pcs.).
Fixing screw
(Cross recessed head screws with spring and
plain washer M3-L8 × 3 pcs.)
10-1-10 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. Remove [9] flexible cable unit that is connected to the display board from the connector (CN1).
Connector
(CN1)
10-1-10 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
10-1-11 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution To detach, take utmost care to prevent missing of washer because the washer is mounted to the
upper-side door hinge only.
Fixing pin
10-1-12 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(Cross-recessed head tapping screws, nominal
3-L6 × 2 pcs.)
10-1-13 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L6 × 2 pcs.)
10-1-14 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L8 × 6 pcs.)
2. Lift up [5] display board to detach from the door case unit and remove [56] switch spacers (2
places) mounted to [4] door case unit.
10-1-15 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
1. Remove [15] motor unit fixing screws (Cross recessed head screws with captive spring washer
and plain washer M3-L10 × 2 pcs.), and after detaching [26] timing belt mounted, remove the
motor unit.
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L10 × 2 pcs.)
2. Remove [60] stepping motor mount mounted to [15] motor unit by removing fixing screws (Cross
recessed head screws with captive spring washer and plain washer M3-L6 × 2 pcs.).
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L6 × 2 pcs.)
10-1-16 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(flat countersunk head screw
M3-L6 × 2 pcs.)
10-1-17 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L12 × 4 pcs.)
10-1-18 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution To detach, take utmost care to prevent missing of spring mounted to occlusion
plunger.
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Occlusion sensor
cover [10]
10-1-19 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Infusion Infusion
pump bottom pump top
surface side surface side
2. After removing pawls (a total of 3 places) that fix [14] infusion pump cover unit to the infusion pump
chassis and detaching the infusion pump cover unit, remove [23] door sensor board mounted to
the infusion pump chassis and remove the connected connector (CN1).
Caution When door sensor board is removed, take utmost care to missing of O-ring for fixing boards, and
mount the O-ring to the door/temperature sensor board mounted section of the infusion pump
chassis as required after the board is removed.
10-1-20 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with
captive spring washer and plain washer
M3-L8 × 2 pcs.)
10-1-21 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross-recessed head tapping screws, nominal
3-L14 × 2 pcs.)
Air-in-line sensor
board [19]
10-1-22 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive
spring washer and plain washer M3-L10 × 4 [55] cam unit
pcs.)
2. Remove the timing pulley mounted to [55] cam unit by removing the fixing screws (Cross recessed
head screws with captive spring washer and plain washer M3-L5).
10-1-23 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Tube clamp
unit [13]
10-1-24 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
* See “Section 7.1 Replacement procedure” for battery installation and drip sensor installation methods.
10-2
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-1 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. After fixing [55] cam unit, mount the cam unit with [29] cam shaft retainer set and fixing screws
(Cross recessed head screws with captive spring washer and plain washer M3-L10 × 4 pcs.).
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L10 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
Fixing screws
(Cross recessed head screws with captive spring washer
and plain washer M3-L10 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
Caution After fixing cam unit, never fail to apply grease to the cam portion.
10-2-2 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Air-in-line sensor
board [19]
Fixing screws
(Cross-recessed head tapping screws, nominal
3-L14 × 2 pcs.)
Screw tightening torque : 20 to 29 cN・m
10-2-3 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution If occlusion board is mounted at lower side, volume adjustment cannot be carried out on the occasion of
“10.2.26 Adjustment of occlusion sensor”. Make sure to mount it in correct direction.
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Screw tightening torque : 50 to 59 cN・m
10-2-4 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution The lead wire assembly may have been mounted to the occlusion unit on the occasion of “10.2.5
Installation of occlusion unit.” In such event, detach the lead wire assembly from the occlusion
unit and connect the assembly to the door sensor board in advance. After the infusion pump
cover unit is mounted, the assembly is unable to be mounted to the door sensor board.
2. Mount [14] infusion pump cover unit to the infusion pump chassis. Allow the occlusion unit board
mounted to the infusion pump cover unit to fit into the groove of the infusion pump chassis. In
addition, allow the lead wire assembly from the door sensor board to come out from the groove
where the occlusion unit is fitted into the infusion pump chassis side surface (main board mounting
side), and make sure the lead wire assembly has come out, and fix the infusion pump cover unit
securely to the infusion pump chassis with the infusion pump cover unit fixing pawls (a total of 3
places).
Fixing pawls
10-2-5 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. After making sure the infusion pump cover unit is securely fixed with fixing pawls, fix the infusion
pump top side and infusion pump bottom side tube clamp unit with the infusion pump cover unit
fixing screws (Cross recessed head screws with captive spring washer and plain washer M3-L8 ×
2 pcs.). After fixing with screws, apply a small amount of screw lock (blue) to prevent screws
from loosening (2 places).
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-5 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution On the occasion of mounting, take utmost care to prevent missing of the spring attached to the
occlusion plunger.
Occlusion sensor
cover [10]
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 2 pcs.)
Screw tightening torque : 20 to 29 cN・m
10-2-6 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L12 × 2 pcs.)
Screw tightening torque : 50 to 59 cN・m
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L12 × 2 pcs.)
Screw tightening torque : 50 to 59 cN・m
Caution y Take utmost care to prevent overtightening so that [54] pump unit would not deform.
10-2-7 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(flat countersunk head M3-L6 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-8 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
1. Mount [60] stepping motor mount to a motor with fixing screws (Cross recessed head screws with
captive spring washer and plain washer M3-L6 × 2 pcs.).
Stepping motor
mount [60]
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L6 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
2. Upon completion of mounting [60] stepping motor mount, mount [26] timing belt between [15]
motor unit and [55] cam unit. After positioning the timing belt by the condition not to provide
over-tension, mount the motor unit with fixing screws (Cross recessed head screws with captive
spring washer and plain washer M3-L10 × 2 pcs.). After fixing screws, fix with screw lock (blue).
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L10 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-9 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution Make sure the seal packing is securely mounted to the door case unit.
Seal packing
2. Upon completion of mounting [56] switch spacers (2 places) and seal packing, mount [5] display
board with fixing screws (Cross-recessed head tapping screws, nominal 3-L8 × 6 pcs.).
Caution When the display board is fixed with screws, first of all, with the display board hand-tightened,
make sure LED and switch can be definitely pressed from the operation panel side, and then, carry
out final tightening.
Fixing screws
(Cross-recessed head tapping screws,
nominal 3-L8 × 3 pcs.)
Screw tightening torque : 60 to 69 cN・m
Fixing screws
(Cross-recessed head tapping screws, nominal
3-L8 × 3 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-10 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution When the buffer plate is mounted, be sure to apply grease to the buffer spring.
When grease is applied, take utmost care to prevent missing of the buffer spring.
2. After applying a small amount of screw lock (blue) for prevention of screw loosening to insert
sections (2 places), cover the buffer plate unit over the spring with care not to fly or bring down,
lightly pressing from above with a finger, mount fixing screws (Cross recessed head screws with
captive spring washer and plain washer M3-L6 × 2 pcs.).
Caution When buffer plate is mounted, take utmost care not to lose buffer springs.
Fixing screws
(Cross recessed head screws with spring
washer and plain washer M3-L6 × 2 pcs.)
10-2-11 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution y When upper-side hinges or lower-side hinges are mounted, take utmost care to prevent fixing pins
from missing due to fall.
y Never fail to make sure O-rings are securely mounted to hinges.
y Make sure grease is surely applied to sliding portions.
Fixing pins
10-2-12 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution y When the crescent hook unit is mounted, take utmost care not to forget to fit a wave washer.
y Apply grease when fixing pins are attached.
Fixing pins
Wave washer
mounting place
10-2-13 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution When the flexible cable unit is mounted, take utmost care to mount it in the correct direction.
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L6 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-14 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
* In the event that it is difficult to connect the flexible cable unit to the connector, connect the connector
after removing the display board temporarily.
Connector (CN1)
2. Mount the door case unit and the door body unit. When the door case unit and the door body
unit are mounted, install [27] door seal rubber to the upper part of the door body unit and at the
same time, assemble the flexible cable while adjusting its length, and after applying screw lock
(blue) to prevent screws from loosening to insert sections (3 places), fix them with fixing screws
(Cross recessed head screws with captive spring washer and plain washer M3-L8 x 3 pcs.).
When the door case unit and the door body unit are fixed, allow the air-in-line sensor harness
connected to the display board to pass through the hole of the upper air-in-line sensor unit
mounting section.
Fixing screw
(Cross recessed head screws with captive
spring washer and plain washer M3-L8 × 3 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-15 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. Fix the occlusion retainer plate and clamp stand to the door body unit with fixing screws (occlusion
retainer plate: Cross-recessed head tapping screws, nominal 3-L6; clamp stand: Cross-recessed
head tapping screws, nominal 3-L8, Cross recessed head screws with captive spring washer and
plain washer M3-L16). Arrange to allow the clamp stand fixing screws Cross recessed head
screws with captive spring washer and plain washer M3-L16 to come to the right side and
Cross-recessed head tapping screws, nominal 3-L8 to the left side.
Caution In the event that Cross recessed head screws with captive spring washer and plain washer M3-L16
section is mounted with the clamp stand fixing screw, apply screw lock (blue) to prevent screws
from loosening to insert sections (3 places) and fix with fixing screws.
Fixing screw
(Cross-recessed head tapping screws, nominal
3-L8 , Cross recessed head screws with captive
spring washer and plain washer M3-L16)
Screw tightening torque : 60 to 69 cN・m
Fixing screw
(Cross-recessed head tapping screws, nominal 3-L6)
Screw tightening torque : 60 to 69 cN・m
10-2-15 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. Apply silicon adhesive to the air-in-line sensor unit mounting section of door body unit and mount
[18] air-in-line sensor unit with fixing screws (Cross-recessed head tapping screws, nominal 3-L8 ×
2 pcs.). After mounting the air-in-line sensor unit, wipe away excess adhesive.
Caution y When air-in-line sensor unit is mounted, take utmost care to prevent air-in-line harness from being
caught.
y Take utmost care to prevent damage to the head end section of the air-in-line sensor unit.
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L8 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
10-2-16 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive
spring washer and plain washer M3-L10)
Screw tightening torque : 60 to 69 cN・m
10-2-17 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M4-L16 × 2 pcs.)
Screw tightening torque : 80 to 89 cN・m
2. After fixing [3] door unit, fix the retainer plate of [9] flexible cable unit to infusion pump chassis with
fixing screws (Cross-recessed head tapping screws, nominal 3-L8 × 2 pcs.). Do not fully tighten
the fixing screw in this event but keep them hand-tightened (after the main board is mounted, the
flexible cable length must be adjusted while monitoring the door opening amount).
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L8 × 2 pcs.)
10-2-18 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Mounting direction
10-2-19 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
2. Then, mount [38] battery relay harness to the battery box section of [16] infusion pump chassis.
After fixing, connect the inlet harness to the connector (CN1) of [20] switching regulator and the
battery relay harness to the connector (CN2). If these are connected, this operation is not
required.
10-2-20 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. Connect the inlet harness to the connector (CN1) of AC inlet unit from the infusion pump bottom
side.
Connector (CN1)
Wiring
(infusion pump
side surface)
Wiring
(infusion pump
bottom side)
10-2-20 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Main board
fixing pawl
Fixing screw
(Cross-recessed head tapping screws,
nominal 3-L6 × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
2. Connect connectors (CN2, 7, 9) from each sensor, etc. of main board [6].
Connectors (CN2, 7, 9)
10-2-21 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
3. To connect the flexible cable, connect the flexible cable to the flexible cable connector (CN-1) with
the door open, and fully tighten the flexible cable retainer plate fixing screws (Cross-recessed
head tapping screws, nominal 3-L8 × 2 pcs.) with the flexible cable flexed between main board and
space on the door side. After fully tightening, open and close the door once or twice, and make
sure the door is smoothly opened and closed without problem. After making sure, apply a small
amount of screw lock (blue) to prevent screws from loosening to fixing screws.
Fixing screw
(Cross-recessed head tapping screws, nominal
3-L8 × 2 pcs.)
Screw tightening torque : 20 to 29 cN・m
Connectors
(CN1)
10-2-21 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Connectors
(CN1)
2. Mount [25] rotation sensor board to the infusion pump chassis by fixing screws (Cross-recessed
head tapping screws, nominal 3-L6, Cross recessed head screws with captive spring washer and
plain washer M3-L5). After mounting the board, apply a small amount of screw lock (blue) to
prevent screws from loosening to fixing screws.
Fixing screw
(Cross recessed head screws with captive spring
washer and plain washer M3-L5)
Screw tightening torque : 60 to 69 cN・m
Fixing screw
(Cross-recessed head tapping screws, nominal 3-L6)
Screw tightening torque : 60 to 69 cN・m
10-2-22 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Fixing screws
(Cross-recessed head tapping screws, nominal 3-L10
× 2 pcs.; inner clip washer × 2 pcs.)
Screw tightening torque : 60 to 69 cN・m
Caution When mounting GND conductor, take utmost care to prevent failure to mount inner clip washers.
10-2-23 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
10-2-24 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
When the self-diagnosis mode is achieved and adjustments are carried out:
1. Set the infusion pump to the self-diagnosis mode (turn ON DIPSW2-4P) and turn on the power
supply.
2. Upon completion of self-check, press the “flow rate selector” switch twice and make sure [2000] is
displayed on the delivery-limit/volume delivered display and A/D value [***] on the flow rate
display.
3. Rotate VR1 of the occlusion detection board from the infusion pump bottom, and set the A/D value
of the flow rate display to “135 to 175” (this operation is shown only for a good-rule-of-thumb).
Bottom side of
infusion pump
VR1
Press the occlusion plungers with jigs (1.8 mm/2.8 mm), measure them three times, respectively,
and calculate the mean value.
y 1.8 mm jig: AVG1 (mean value of three measurements)
y 2.8 mm jig: AVG2 (mean value of three measurements)
In such event, [51] ≤ (AVG1 - AVG2) ≤ [57]
• When the value exceeds [57], rotate VR1 clockwise to decrease the excess of the reading.
• When the value does not reach [51], rotate VR1 counterclockwise to increase the shortfall of
the reading.
4. When adjustment of the occlusion A/D value is completed, open the door, turn on the power
supply at no load (condition in which no load is applied to the occlusion plunger), and hold the
“Quick-delivery” switch depressed for about 3 seconds. The occlusion A/D value (at no load) is
stored in memory.
10-2-25 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
5. See “Section 6.2.5 Occlusion inspection method, set the occlusion inspection jig to the pump main
body, and press the “START” switch.
6. Make sure the pressure is detected in the range from 60 to 120 kPa (0.61 to 1.22 kgf/cm2). In
addition, make sure the A/D value varies more than “95” when the plunger is pressed to the
extreme end of the stroke.
7. When the occlusion pressure was detected without acceptable range, return to Paragraph 3 and
carry out readjustment. In addition, make sure the plunger operates smoothly.
8. After adjustment, return the rotary switch setting (0-9) to the original setting after the infusion pump
returns to the normal mode.
10-2-25 (2)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
A/D value with water tube mounted A/D value with empty tube mounted
10-2-26 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
10-2-27 (1)
SY14-9B002
Applicable Lot
Lot No. 0607 –
Prepared in Feb., 2007
Caution When mounting, take utmost care to prevent breakage of drip-proof cap.
Fixing pawl
2. Put [1] case unit on the chassis from the back side and fix with fixing screws (Cross recessed head
screws with captive spring washer and plain washer M3-L8 × 3 pcs.).
Fixing screws
(Cross recessed head screws with captive spring
washer and plain washer M3-L8 × 3 pcs.)
Screw tightening torque : 60 to 69 cN・m
* See “Section 7.1 Replacement procedure” for the method of mounting the battery and installing the
drip-sensor.
10-2-28 (1)