Equipo PCR PLas Labs 825
Equipo PCR PLas Labs 825
Equipo PCR PLas Labs 825
LIMITED WARRANTY
Plas Labs, Inc., warrants all materials and components used in the manufacture or
assembly of its’ products against defects in materials and workmanship for a period of
twelve (12) months after shipment. In the event of product failure of operation, the
obligation of
Plas Labs under this warranty shall be limited to repairing or replacing at its’ option,
any part of said equipment which upon Plas Labs inspection is determined defec-
tive.
Defective items are to be returned to Plas Labs, Lansing, Michigan. All transporta-
tion charges are to be prepaid.
This warranty does not apply to equipment or parts which fail because of abuse; acci-
dent, alteration, misuse, erosion, improper installation, or improper replacement of a
repaired item.
The buyer assumes all risks for results obtained from these products, whether used
alone or in combination with other items. It is expressly understood that we are not
responsible and will not be held liable for damage and/or injury caused by the use of
our products.
This warranty is in lieu of any other warranty, expressed, or implied, including mer-
chantability of fitness, and of all other obligations or liability whatsoever on Plas
Labs part.
6
C. SCRATCH REMOVERS
Deep scratches should first be sanded with fine grit (400 or finer) wet sandpaper.
Steel Wool (OOOO Finest Grade) is also very helpful. Use the polishing materials
(rubbing compounds), mentioned above for the final stage.
Stainless Steel is resistant to all solvents and detergents. Polishing can be accom-
plished by using fine grades of Steel Wool and/or #707 Scotch Brite Pads (3-M
Corp.). For the final stage, use a polishing spray like “Stainless Steel Magic.”
ALUMINUM COMPONENTS
Again, solvents or detergents may be used for cleaning aluminum. If the aluminum
becomes tarnished, it may be rubbed with any of the many commercial polishes
available.
In all cases, try to avoid the use of abrasives to clean your equipment.
5
MAINTENANCE AND CARE OF
PLAS LABS PRODUCTS
THERMOPLASTIC COMPONENTS
A. CLEANERS
Cleaning thermoplastics is best accomplished with soap or detergent and water solu-
tions. In cases where residues left by these agents is undesirable, special cleaning
solvents may be used. Soaps and detergents (except those of the abrasive type), will
not harm plastics, but several common solvents will.
In general, aromatic and chlorinated hydrocarbons will attack most plastic surfaces.
This applies to all of the plastics used in Plas Labs products.
Examples of these products include (but are not limited to), acetone, ether, gasoline,
lacquer thinner, methyl-ethyl-keytone, methylene chloride, and toluene.
Thermoplastics have a limited resistance to alcohol (all types) but their use is not rec-
ommended.
Dilution of alcohol with water will minimize damage, but the exposure time should
be kept to a minimum. Prolonged contact of plastics to alcohol will cause the plastic
to “craze”. (This is a fine cracking close to the exposed surface.) Crazing severely
reduces the optical qualities and strength of the plastic.
B. POLISHES
While the above cleaning solutions have some polishing capabilities, they will not
remove scratches from plastics. This can only be done with automotive type waxes
or the finer grades of rubbing and polishing compounds. These products should be
specifically for acrylic enamels and lacquer base paint.
4
DAILY OPERATION AND MAINTENANCE R/4-03
this period of time, turn off the U.V. light. Your chamber is ready
for use.
5. Upon completion of your project, remove your materials and
close the corner doors. Reactivate the U.V. lamp for a minimum
fifteen (15) minutes to insure sterility of the interior.
3
GENERAL INFORMATION
#825-UVC/PCR
(Domestic & Export Models)
This chamber has been designed to aid in improving the accuracy of your P.C.R.*
results and general tissue culture procedures. The chance for air-borne contamina-
tion during D.N.A. sequencing is greatly reduced.
This is a “still air enclosure” which contains both fluorescent and U.V. Germicidal
lamps. The U.V. lamp is rated at 254 nm and will effectively kill all micro-organ-
isms that come in direct contact with the U.V. rays. For best results, the U.V. lamp
should be turned on fifteen (15) minutes prior to use and fifteen (15) minutes after
use. Use the automatic timer for this.
The front viewing sash is .375” (9.5 mm)thick acrylic plastic and is effective pro-
tection against against 32P labeled compounds. It affords protection against Beta
rays, but not Gamma rays. The viewing sash is rated for approximately 2,050 hours
of operational durability. If necessary, a replacement sash may be ordered (Catalog
#825-UVC/FS). Changeover is accomplished by removing six (6) screws in the
front sash.
The main body of the unit is not attached to the white base to enable the user to
install large pieces of equipment. The base also includes a formed in place “dam”
to allow easy cleanup of possible spills.
Access to the interior of the chamber is gained by opening the corner doors. If not
needed, these doors may also be removed by raising them off the black plastic
hinges. If a door is opened or removed, electronic sensors turn off power to the
U.V. system. This is a safety feature to prevent accidental U.V. burns.
Two (2) bright white plastic shelves are included, one of which has cut outs for
pipettor storage.
General Information 2
Limited Warranty 6
OPERATION MANUAL
825-UVC P.C.R.* Chamber
(Domestic & Export Models)