Case v67
Case v67
Case v67
CASE™ V6.7
Software Version 6.73
Operator’s Manual
2060290-201 ENG Revision C
NOTE: The information in this manual only applies to CASE V6.7 systems software version 6.73. It does not
apply to earlier software versions.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
CASE, MAC, MARS, and MUSE are trademarks owned by GE Medical Systems Information Technologies
GmbH, a General Electric Company going to market as GE Healthcare.
CE Marking Information
Compliance
The product CASE bears the CE marking CE 0459, notified body GMED,
indicating its conformity with the provisions of the Council Directive 93/
42/EEC concerning medical devices and fulfills the essential
requirements of Annex I of this directive.
The medical device has been assigned to class IIa as specified in Annex
IX of the Directive 93/42/EEC.
Magnetic and electrical fields are capable of interfering with the proper
performance of the device. For this reason make sure that all external
devices operated in the vicinity of the system cart comply with the
relevant EMC requirements. X-ray equipment, MRI devices, radio
systems, cellular telephones, etc. are possible sources of interference as
they may emit higher levels of electromagnetic radiation. Keep the
system cart away from these devices and verify the performance of CASE
before use.
The safety and effectiveness of this device has been verified against
previously distributed devices. Although all standards applicable to
presently marketed devices may not be appropriate for prior devices (i.e.,
electromagnetic compatibility standards), this device will not impair the
safe and effective use of those previously distributed devices (see
Operator's Manual).
2 Equipment Overview
Revision C CASE i
2060290-201
Viewing and Printing the Operator's Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Online Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
5 Resting ECG
ii CASE Revision C
2060290-201
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
10-Second ECG with Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Operating Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Full Disclosure ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Operating Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
6 Exercise Test
iv CASE Revision C
2060290-201
Audio ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-51
Waterfall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-52
3D Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-53
12SL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-54
Vector Loops . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-55
Report Printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-56
8 Spirometry Test
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Revision C CASE v
2060290-201
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Flow-Volume Measurement - Operating Steps . . . . . . . . . . . . . . .8-9
Bronchodilator Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-12
Spirometry Tests with the LF 501 . . . . . . . . . . . . . . . . . . . . . . . . .8-12
9 File Management
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
11 DICOM Interface
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
vi CASE Revision C
2060290-201
Performing Procedures without the Order List (Unscheduled Procedures) . . . 11-6
12 System Settings
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Before Each Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Technical Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
Disposal at the End of Its Service Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
B Miscellaneous
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-20
Revision C CASE ix
2060290-201
C Order Information
D Technical Description
x CASE Revision C
2060290-201
1 Introduction
Part No./
Date Comment
Revision
Manual Purpose
This manual contains the instructions necessary to operate the
equipment safely and in accordance with its function and intended use.
Intended Audience
This manual is geared for clinical professionals. Clinical professionals
are expected to have working knowledge of medical procedures,
practices, and terminology as required for completing these
examinations.
Illustrations
All illustrations in this manual are provided as examples only. They may
not necessarily reflect your equipment setup or data displayed.
All names appearing in examples and illustrations are fictitious. The use
of any real person's name is purely coincidental.
Safety Information
Intended Use
CASE is intended to be used by trained operators under direct
supervision of a licensed health care practitioner on adult and
pediatric patients.
CASE is designed to acquire, process, record, archive, analyze and
output ECG data (12 and 15 leads) during a period of physiologic
stress or during a resting ECG test and acquire data from ancillary
devices, such as spirometry and ambulatory blood pressure devices.
Furthermore, it provides median morphology recordings and records
ECG in real-time with and without arrhythmia detection.
The arrhythmia detection algorithm of CASE is provided to the user
for the convenience of automatic detection of arrhythmias but does
not provide alarms.
CASE provides the control of external devices. These are typically
treadmills and bicycle ergometers. Additionally, CASE
communicates with centralized electronic/digital storage systems via
data networks.
WARNING
Patient Hazard—Do not use the digital or analog outputs
to control therapeutic devices such as defibrillators.
General Information
This manual is an integral part of the device. It should always be
kept near the device. Close observance of the information given in
the manual is a prerequisite for proper device performance and
correct operation and ensures patient and operator safety. Please
read the manual once in its entirety, because information pertinent
to several chapters is given only once.
To ensure maximum patient safety, interference-free operation and
the specified measuring accuracy, we recommend using only original
GE Healthcare accessories. The user is responsible for application of
accessories from other manufacturers.
The warranty does not cover damage resulting from the use of
unsuitable accessories and consumables from other manufacturers.
GE Healthcare is responsible for the effects on safety, reliability, and
performance of the device, only if:
assembly operations, extensions, readjustments, modifications,
or repairs are carried out by GE Healthcare or by persons
authorized by GE Healthcare.
the electrical installation of the relevant room complies with the
requirements of the appropriate regulations, and
the device is used in accordance with the instructions given in
this manual.
This equipment is protected against the effects of cardiac
defibrillator discharge to ensure recovery as required by test
standards.
Accuracy of the Input Signal Reproduction
Recording speed accuracy is tested using the method described in
IEC 60601-2-51 sub-clause 51.108.4.4. The maximum error is +/-
5 %.
Frequency Response is tested according to IEC 60601-2-51 sub-
clause 51.107.1 conforming to the specifications of sub-clause
51.107.1.1.
Definition
The terms Danger, Warning, and Caution are used throughout this
manual to point out hazards and to designate a degree or level of
seriousness. Familiarize yourself with their definitions and significance.
Danger
Explosion Hazard — The device is not designed for use in
areas of medical locations where an explosion hazard
may occur. An explosion hazard may result from the use
of flammable anesthetics, skin cleansing agents and
disinfectants. Furthermore, the device is suitable for
application in an oxygen-enriched atmosphere only with
certain restrictions. The atmosphere is considered to be
oxygen-enriched when the room air contains more than
25 % of oxygen or nitrous oxide.
Warning
Shock Hazard — Observe the following warnings. Failure
to do so endangers the lives of the patient, the user, and
other persons present.
Warning
Shock Hazard — Observe the following warnings. Failure
to do so endangers the lives of the patient, the user, and
other persons present.
Warning
Shock Hazard — Observe the following warning. Failure
to do so endangers the lives of the patient, the user, and
other persons present.
Warning
Patient Hazard — During exercise tests, a defibrillator
and a pacemaker, both checked for proper functioning,
should be kept at hand.
Caution
Equipment Damage — Before connecting the device to
the power line, check that the voltage and frequency
ratings of your power line match those indicated on the
device label.
Classification
Degree of safety of application in the Not suitable for use in the presence of a
presence of a flammable anesthetic mixture flammable anesthetic mixture with air or
with air or with oxygen or nitrous oxide with oxygen or nitrous oxide.
Equipment Symbols
Note
Some of the symbols may not appear on all equipment.
Signal input
Signal output
Fuse
12V 12 V DC
The number found under this symbol is the date of manufacture in the YYYY-
MM format.
Medical Equipment — Classified with respect to electric shock, fire and
mechanical hazards only in accordance with UL 60601-1/CAN/CSA C22.2
No. 601.1, CAN/CSA C22.2 No. 601-2-25, IEC 60601-2-25, IEC 60601-1-1.
Manufacturer’s identification
Order number
Serial number
Ident number
Lot number
Mass
This symbol indicates that the waste of electrical and electronic equipment
must not be disposed as unsorted municipal waste and must be collected
separately. Please contact an authorized representative of the manufacturer
for information concerning the decommissioning of your equipment.
This symbol indicates that the device was tested and certified according to
the Russian safety standards specified by GOST.
Biocompatibility
All parts of the product and all accessories described in this manual that
come in contact with the patient during the intended use of the product,
fulfill the biocompatibility requirements of the applicable standards. If
you have questions in this matter, please contact GE Healthcare or its
representatives.
Technical Maintenance
For technical data and other detailed technical information, please refer
to the Service Manual.
Legal Notice
Our equipment contains several fields which can be filled in before
performing an ECG. Some of these fields must be filled in before
performing an exam, some are optional and therefore left to the user to
assess whether they are needed to perform the exam. A field RACE is one
of these optional fields. It has been acknowledged by the medical
profession as useful to analyze some pathologies. You should be aware
that, in some jurisdictions, the processing of data revealing an
individual’s racial origin is subject to legal requirements, such as
obtaining the patient’s prior consent. If you elect to collect this type of
data, it is your responsibility to ensure that you comply with all
applicable legal requirements.
Security Patches
Please visit the GE Healthcare Product Security Database regularly and
search it for security patches for your product in order to keep the
operating system up to date and ensure its integrity. Download the
Service Information
Service Requirements
Refer equipment servicing to GE Healthcare authorized service
personnel only. Any unauthorized attempt to repair equipment under
warranty voids that warranty.
Equipment Identification
Every GE Healthcare device has a unique serial number for
identification. The serial number appears on the device label.
D 1 XX 0005 G XX
Month Manufactured Year Manufactured Product Code Product Sequence Number Division Device Characteristics
A = January 1 = 2001 Two-character Manufacturing number One or 2 letters that further
B = February 2 = 2002 product (of total units manufactured.) describe the unit, for
C = March 3 = 2003 descriptor example:
D = April (and so on) P = prototype not conforming
E = May to marketing specification
F = June R = refurbished equipment
G = July S = special product
H = August documented under Specials
J = September part numbers
K = October U = upgraded unit
L = November
M = December
A B C D E F
A Product Code
B Year Manufactured (00-99)
00 = 2000, 01 = 2001, 02 = 2002 (and so on)
C Fiscal Week Manufactured
D Production Sequence Number
E Manufacturing Site
F Miscellaneous Characteristics
CASE System
Front View
H
D
I
E
01A
A Monitor
B Function keypad
C Paper tray
D Console for PC keyboard and mouse
E Power button (non-breaking of AC power)
F Wheel locks
G Thermal writer
H USB drives (2)
I Acquisition module holder
Rear View
02A
Warning
Shock Hazard — Only technical staff is authorized to
open the access panel.
2. Guide the cables through the opening in the access panel and connect
them to the appropriate ports as shown on the reference diagram.
3. Close the panel.
Cable clamps
Reference
diagram
Note
Coil the excess cable for storage as shown above. DO NOT squeeze
the cables when closing the access panel.
Power inlet
Mains power
switch Potential
equalization pin
85A
Cable clamp
Back Panel
1. Connect CASE to the power line and secure the power cord with the
clamp provided.
Note
DO NOT turn the system on yet.
Power button
Side Panel
5. Turn on CASE with the power button on the side panel. The program
will start up automatically and the Windows screen appears.
6. Log on to the system.
7. Double-click the CASE icon.
8. If the local database window appears, click Cancel.
9. The initial CASE screen will appear.
When you see the initial CASE screen (see “Initial Screen” on page 2-16)
and no error message appears, the system is operational.
Note
The screen saver is automatically disabled during ECG
acquisition.
Scan the hard drive for viruses once a week.
Run a functional test on a regular basis (about once a month);
this includes an inspection of the cables and other accessories for
signs of damage.
087A
Warning
Loss of Data — Improper shutdown of the CASE system,
e.g., turning the system off with the mains power switch
before all programs are closed, may damage the
databases.
Note
Turn off CASE with the mains power switch only if it will not be used
for a prolonged period of time. This switch disconnects the system
from mains power.
Function Keypad
The function keypad contains the keys necessary to operate the
ergometer/treadmill, to control the exercise test and print reports.
52A
Note
Press F2 to display a softkey panel at the bottom of exercise test
screens. These softkeys can be used to control the exercise test. The
softkeys are activated by clicking. With ↑ + F2 or Esc you can
remove the softkey panel from the screen.
Symbol Description
Press pretest to initiate the pretest phase and to advance through the
pretest stages (to return to the preceding pretest stage, press longer than
2 seconds).
Press exercise to initiate the exercise phase and to advance through the
individual stages of the exercise phase.
Press recovery to initiate the recovery phase and to advance through the
individual stages of the recovery phase.
Press test end to terminate the test and to initiate the printout of the final
reports.
Symbol Description
Press new test to prepare the system for a new patient/a new test.
Symbol Description
Press start tmill or STOP tmill to start or stop the belt. When restarted
after FAST stop, the treadmill returns to the previous speed and grade.
Desk Light
The CASE exercise testing system work surface and function keypad are
both equipped with lighting.
Note
The desk light turns off automatically after approximately one hour
of keypad inactivity.
29A
There is one set of lead label stickers with AHA labeling and one
set with IEC labeling.
2. Connect the patient cable to the acquisition module.
To remove the cable from the acquisition module, press in the
spring lock located on the side of the cable.
3. Plug the individual lead wires into the acquisition module and
connect the electrode clips to the BNC sockets of the lead wires.
Observe the lead labels: the labels on the cable and on the
acquisition module must be identical.
Having connected the electrode clips to the BNC sockets, turn
them 45° clockwise so that they lock into place.
4. Screw the supplied acquisition module holder to the right side of the
cart, then place the acquisition module into the holder.
Caution
Proper Leadwire Connection — Improper connection will
cause inaccuracies in the ECG.
Note
The ECG signal input is isolated and defibrillation-proof only in
conjunction with the CAM-14 acquisition module (type BF).
1. Push the release lever to the right to open the paper compartment
lid.
Caution
The lid has a gas spring and opens automatically.
063A
064A
065A
4. Hold the protruding paper sheet with your right hand and close the
lid with your left hand: you must hear the lid lock into place. Then
the paper will be transported to the next cue hole.
066A
067A
Initial Screen
A
B
C
D
E
F
G
H
06A
Local Database - Click to display the contents of the local database for
B
selection of a patient and test.
C
Archived Data - Click to view patient records archived on a CD-ROM.
Note
For system users not familiar with handling the PC mouse, the basic
functions are explained on the next page.
The system will load the Acrobat Reader program and display the title
page of the operator's manual. Use the controls explained below to
navigate through the document.
Note
Clicking a chapter headline in the operator manual's table of
contents takes you directly to the corresponding chapter. Clicking an
index entry will also display the corresponding page of the manual.
B C
141A
the Print Range (all pages, the current page, or a section of the
document)
the print quality
the number of copies.
3. Click OK to initiate the printout.
Online Help
On many screens you will find a Help button. Click the button to display
a context-sensitive Help window.
Note
The online help implemented in the software provides information
about using the program. For safety information, please refer to the
Operator Manual and all other documents supplied with the product.
Software Options
Many features are available as options which can be unlocked to enhance
the CASE functionality. Following is a list of these optional features. For
details, please contact your Sales Representative.
- functional capacity
ESPI (Ergospirometry)
software package for control of an external ergospirometry
program
The signal acquisition screen will show the applied electrodes and the
signal quality after the right-arm electrode has been applied. When the
right-arm electrode becomes detached, the system behaves as if all
electrodes were disconnected.
WARNING
Strangulation Hazard — Route cables and tubes away
from patient's throat to avoid possible strangulation.
31A
1. Shave any hair from each electrode site and degrease each site with
alcohol.
2. Remove the epidermal skin layer at each site. Use an abrasive pad or
skin prep cream (for example, EPICONT, Part no. 217 083 12).
CAUTION
Skin Lesions — Do not rub the skin vigorously to avoid
inflammation or bleeding.
Applying Electrodes
Lead Labels on the Acquisition Module
CAM-14
AHA Label IEC Label
RL N
RA R
LA L
LL F
V1 C1
V2 C2
V3 C3
V4 C4
V5 C5
V6 C6
A1 A1
A2 A2
A3 A3
A4 A4
CAUTION
Proper Leadwire Connection — Improper connection will
cause inaccuracies in the ECG.
32A
33A
34A
Note
The ECG recorded with the torso placement of the limb lead
electrodes may differ from that recorded with the electrodes on the
limbs. Affected characteristics are the Q-waves and the frontal axes,
whereas ST levels are unlikely to change.
This lead placement is the preferred placement for exercise stress
tests on treadmills or bicycles with patients in the supine or half-
supine position.
For upright bicycle exercise stress tests it is recommended to apply
the limb leads on the back, opposite the limb electrode positions
shown above.
35A
36A
37A
38A
47A
Note
The functions of buttons 1 and 2 on the acquisition module vary with
the test performed:
Resting ECG:
Exercise Test:
For exercise tests, different tasks can be assigned to the buttons (see
“Miscellaneous 1 Tab” on page 12-24). The default functions of the
buttons are:
I
048A
Note
Before running a new test, you are required to select a patient record
from the database or to enter a new patient. Depending on the
system configuration you will be working with the local database of
the system or with the database of the MUSE system. When you
select New Test on the initial screen, a window opens where you
select the patient record and test type or enter a new patient.
5 1 2 4
3
048A
Note
Once you have selected your patient, the acquisition screen for the
new test will appear.
The patient selection window has one additional button: Order List (see
next page).
Note
This is an optional feature.
When you enter a new patient or edit patient information, the
data is first saved to the local database. The data will appear in
the MUSE system only after you have transferred the test to
MUSE and confirmed the data there.
Please refer to section “MUSE Tab” on page 12-55 for
configuration of the system to communicate with MUSE.
F
159A
G
A Requested orders.
B Your location number.
C The list only shows orders for your location .
The list shows orders for all locations .
D The list only shows orders for the selected test type .
The list shows all orders .
E The list only shows orders for the selected patient .
The list shows orders for all patients .
F The list will be updated according to the current search criteria. After changing
the search criteria, you can press the Enter key to start the search.
G The test of the selected order will be started. If you do not change the search
criteria, you can directly press the Enter key to start the search.
048A
2 3 45 9 6 7 10 1 8
Note
The system will automatically capitalize the first letter of the
first and last names.
The patient record cannot be created without an ID number. If at
all possible, also enter the patient's last name, first name, date of
birth and gender to allow the system to perform the necessary
calculations when evaluating the test results.
Note
GE Healthcare is not in a position to guarantee full compatibility of
external programs with CASE.
1. Start CASE.
2. Select patient.
3. Select external program on procedure selection screen (the external
program must already be configured).
4. Perform examination (e.g. Holter ECG).
5. Return to CASE, summary and recording will be assigned to the
patient record.
Note
When the configuration of external programs changes, maintain the
compatibility to ensure that stored tests remain accessible.
051A-R
4 3
The Patient Information tab is open.
1. Verify the patient's name and ID. If the data is incorrect, click
Modify to change.
2. Enter the blood pressure readings or initiate a blood pressure
measurement with Start BP....
3. Enter a designation for the Room (5 characters max.).
4. Select a Location Number or a Location Name (names can be
configured on the MUSE tab, see “MUSE Tab” on page 12-55).
Note
You can assign a Location Name to the Location Number (see section
“MUSE Tab” on page 12-55).
Note
The Patient Information can be assigned to the patient file or only to
the pending test.
Test/Personnel Tab
1 2 3 5 4 6 7
052A-R
Note
The Reason for Test library can be edited (see section “Editing the
Reason for Test Library” on page 12-4).
You enter the names of the physicians and of the ECG technician in
the System Configuration menu (see section “General Tab” on page
12-40).
Medications Tab
2 1 3
053A-R
Note
You can assign drugs to specific medication groups (see section
“Editing Medication Groups, Assigning Drug Names” on page 12-6).
Note
Entering a medication will influence the analysis with the HEART
ECG analysis program. The following medications may affect the
diagnosis:
digitalis antiarrhythmics Ic
diuretics antiarrhythmics III
psychotropics Ca channel antagonists Verapamil
steroids Ca channel antagonists Nifedipin
beta-blockers II nitrates
beta-blockers III ACE
antiarrhythmics Ia alpha-blockers
antiarrhythmics Ib cytostatics
With this modality you can acquire a 10-second resting ECG with
subsequent analysis and/or you can record the full disclosure ECG for a
period of up to 60 minutes.
A B
C
D
2 F
G
H
I
054A-R
O N M L K J
Click to terminate acquisition of the full disclosure ECG and/or to end the
F
test and display the report.
Click to load system setups stored for different system users (see section
G
“Custom Setups, Factory Setup” on page 12-38).
Click to display the resting ECG setup menu (see section “Resting ECG
H
Setup” on page 12-3).
Warning
Misinterpretation — Please bear in mind that filters
might also suppress diagnostically relevant portions of
the signal, as they limit the signal range. Therefore,
filters should only be enabled if necessary.
Operating Steps
1
2
054A-R
Note
When you select ECG Acquisition before the system had time to save
10 seconds of ECG data, a bar graph will inform you of the ongoing
storage process.
The analyzed 10-second ECG appears at the top of the screen. A new tab
provides the corresponding interpretation.
4 5 3
055A-R
Note
Clicking Stop FD/Post Test Review brings up the Test Summary
(see “Test Summary” on page 5-11).
Operating Steps
2a
2b
1
056A-R
2d 2c
Note
Clicking Stop FD/Post Test Review brings up the Test Summary
(see page 5-11).
F
G
H
I
J
K
L
M
057A-R
Click to display a list with more tests of the current patient (local
G
database).
Click to display a list with more tests of the current patient (MUSE
H
database system).
The Modify... buttons allow you to edit or complete the data in the
respective areas.
Note
Changes to the patient information will only affect the current test;
previous tests of the same patient remain unaffected.
ECG Traces
On this screen you see the reference ECG. Depending on the selected
screen format and speed, the waveforms may take up several screens.
A B C D E F
058A-R
Zoom Mode
A zoom factor from 2 to 12 can be selected and the selected segment can
be displayed in all leads.
A B C D E F
059A-R
1 3 2 1 2 3
510A-R
2. Position the cursor on the second measurement point and click. Now
you will see
the second caliper
the measuring line and
the measurement results
Medians
This screen shows the median complexes of all leads as well as the
measurement marks that formed the basis for analysis. You are free to
perform additional measurements, if needed.
A B C D E
511A-R
A Click button for a reanalysis (with the HEART ECG analysis program only, see
next page).
B Click button to display the tabular summary.
C Click to select the gain (amplitude).
D Click to select the speed.
E Click icon to copy the waveform area to the Clipboard.
Reanalysis
Reanalysis allows you to manually change the position of the ST
measurement marks. The reanalysis can only be performed on ECGs
analyzed with the HEART program (see section “Miscellaneous Tab” on
page 12-8).
On the reanalysis screen the medians of all leads are superimposed. The
currently selected lead is highlighted. The measurement marks
determined by the program are shown below the ECG baseline. The
measurement marks above the baseline can be edited. The table in the
top right-hand corner indicates the time intervals.
A B C D E F G H I
512A-R
G Measurement mark.
H Menu for selection of the leads to be displayed.
I Table showing time intervals and position of the cursor in mV.
Performing Reanalysis
1. Click the measurement mark and hold the mouse button depressed.
2. Move the measurement mark to its new position and release the
mouse button.
3. Adjust the other marks in the same way.
4. Click ECG Analysis.
5. Click Reset and ECG Analysis if you wish to restore the previous
analysis results.
Note
The P-onset and P-offset markers may be dashed, e.g. in the presence
of atrial fibrillation. This indicates that the program was unable to
locate the exact beginning and end of the P wave. In this case you can
click the marks and drag them to the correct position; the marks are
now solid instead of dashed. It is possible to restore the marks to the
original position by clicking them again.
2. Holding the mouse button depressed, move the T-offset mark to the
correct position.
3. Click icon to select the leads whose T-offset marks you also wish to
adjust.
Adjusting the T-offset mark on this screen (page 5-19) does not affect
the position of the T-offset mark on the previous screen (page 5-17).
3 1 2
513A-R
Arrhythmia Review
This screen shows one lead of the reference ECG at a time. A bar
diagram below the signal trace indicates, for each individual QRS
complex, the deviation of the RR interval from the mean RR interval.
A B C D E
514A-R
2. Click on the end of the interval. A line appears connecting the two
end points and the corresponding heart rate is displayed.
3. Click the right mouse button to remove the markers and data.
In the presence of atrial fibrillation or flutter:
1. Click Cycles and select the number of cycles over which the HR is to
be determined.
Vector Loops
This screen shows the median complexes of the orthogonal FRANK leads
X, Y and Z as well as the vector loops in three planes.
A B C
515A-R
A B C D E F G H I J
516A-R
2. Click again to mark the end of the segment. A second flag appears.
Printing
1. Click the printer icon. The print setup window opens.
2
3
8
7
5 517A-R
The operating controls basically remain the same. When you click the
printer icon, however, only that page will be printed. Click the zoom icon
again to exit the zoom mode.
1 3 4
527A-R
Report Printout
Click the Print button to print the report. The print setup window opens
where you can temporarily adjust the settings for the printout. Refer to
section “Final Report Tab” on page 12-9 for details on changing the setup
permanently.
1 2 3 1 2
4 5 6 9 8 7
518A-R
Configured Report
10-second ECG (up to 3 different formats, with or without
Interpretation or ACI-TIPI score)
two more formats of the 10-second ECG
Medians
Vector Loops.
2. Select the lead format for the 10-second ECG.
3. Select the number of copies to print.
4. Choose the leads if you selected the Swedish Median H1 format.
5. Select the rhythm leads.
6. Select whether or not the Median H1 report is to include the
measurement marks.
7. Select the gain.
8. Select the printer.
9. Click the Print button to initiate the printout.
A B C D E F G
525A-R
chapter .
E Click icon to confirm the test results.
F Click icons to cut, copy and paste text in the Interpretation area.
G Click the window frame to resize or move the window.
Note
As an alternative you can double-click a text to copy it from left to
right.
4. Select the check box if you wish to copy each text to a new line.
Note
Interpretative statements transferred to MUSE that are part of the
acronym list will be replaced with the appropriate acronyms.
2 3 4
526A-R
1. Click Compare.
A window opens listing all the patient's resting ECGs.
2. Select an ECG.
3. Click OK to clear the window.
The comparison screen with the two 10-second ECGs appears.
10-Second ECG
A B C D E F G
519A-R
Zoom Mode –
A B
520A-R
A Click scroll box to move the segment of procedure 2 to the right and left.
B Click scroll box to move the segment of procedure 2 up and down.
Medians
Click the Medians tab to display the screen for comparison of the two
median complexes. The scroll boxes also allow you to move the median of
procedure 2 up and down and to the right and left.
Interpretation
Click the Interpretation tab to display the screen for comparison of the
two interpretations.
A B C D
521A-R
Note
Software option ECGH is required for use of this functionality.
1. Click Compare.
A B C D E F
522A-R
A Acquisition date.
B Click to select the gain.
C Click to select the leads.
D Click icon to activate the zoom function.
E Click icon to print the screen.
F Click to select and deselect the ST values.
Zoom Mode
Follow these steps to enable the zoom mode:
A B C D E
523A-R
A Acquisition date.
B Click to select the lead.
C Click icon to terminate the zoom function.
D Click icon to print the median.
E Click button to display the 3D graph.
3D Graph
Click the 3D graph button to display a three-dimensional representation
of the medians from all selected procedures. The scroll boxes allow you to
rotate the illustration horizontally and vertically.
A B C D E
524A-R
Patient Education
Before the test you should give your patient some information about
what will happen in the course of the test. Click the Patient Education
button on the pre-acquisition screen to display a description of what the
patient can expect to happen before, during, and after the exercise test.
051A-S
6
The Patient Information tab is open.
1. Verify the patient's name and ID. If the data is incorrect, click
Modify to change.
2. Select the protocol for the next test (refer to section “Protocol Editor
Tab” on page 12-19 for information on changing the default protocol).
3. Confirm or change the Max. predicted HR for the exercise test (refer
to section “Miscellaneous 1 Tab” on page 12-24 for information on
changing the calculation method).
4. Enter the Target HR (%) of the Max. predicted HR (refer to section
“Miscellaneous 1 Tab” on page 12-24 for information on changing the
percentage permanently).
5. Enter a designation for the Room (5 characters max.).
6. Select a Location Number or a Location Name (names can be
configured on the MUSE tab, see “MUSE Tab” on page 12-55).
Note
You can assign a Location Name to the Location number (see section
Note
The Patient Information can be assigned to the patient file or only to
the pending test.
Note
Before initiating TWA analysis check that all TWA settings are
correct (see “TWA Tab” on page 12-30) and that the electrodes are
properly applied according to the instructions given in chapter
“Preparing the Patient” on page 3-1.
Test/Personnel Tab
1 2 3 4 6 5 7 8
052A-S
Note
The Reason for Test library can be edited (see section “Miscellaneous
2 Tab” on page 12-26).
You can define two Extra Questions as needed (see section “Entering
Extra Questions” on page 12-28).
You enter the names of the physicians and of the technician in the
System Configuration (see section “General Tab” on page 12-40).
Medications Tab
2 1 3
053A-S
Note
You can edit the libraries (see section “Editing the Comments, Test
Types, Reason for Test, Reasons for Termination, Medications
Library” on page 12-26).
The risk calculations are a prediction of group incidence and may not
apply to individual patients. Refer to the American Heart Association
publication "Instructions for Using Coronary Risk and Stroke Prediction
Charts and Worksheet" (1990) for details.
The system automates the manual calculation of risk. You can edit the
values calculated by the system. Edited values are marked with an
asteriks *.
2 2 1 3 3 4
94A
2 2 1 3 4
95A
1. Click the AHA Stroke Risk or the AHA CHD Risk tab.
2. Enter the factors that influence both coronary AND stroke risks.
3. Enter the factors that influence stroke OR coronary risk.
4. Click Stroke Risk or Coronary Heart Disease Risk.
A B C
D
E
F
G
H
I
J
K
056A-S
U T S R Q P O N M L
L Heart rate.
M System messages window.
N Indicates AC line filter and arrhythmia recording enabled.
O Enable or disable FRF or Cubic Spline algorithm, see note below.
P Select filter frequency (current ECG).
Q Select screen gain and speed.
R Click to change selected leads.
S Lead label, click the label with the right mouse button to display an up/
down arrow for adjustment of the ECG baseline.
T Select writer gain and speed (for ECG recordings in real-time only).
U Click icon to freeze and release the ECG waveforms (except lead at the
top).
Warning
Misinterpretation — Please bear in mind that muscle
filters (20 Hz, 40 Hz, 100 Hz) might also suppress
diagnostically relevant portions of the signal, as they
limit the signal range. Therefore, muscle filters should
only be enabled if necessary.
Note
Cubic Spline and FRF are algorithms that significantly reduce
artifact and baseline fluctuations without adversely affecting the
ECG signal as conventional filters do. Therefore, the system allows
you to enable and disable the algorithms before or during the test, as
needed.
Note
During the exercise test, a new protocol selection with Select Protocol
will become effective in the next phase.
Note
Pacemaker Patients — It is very important that you indicate on the
first database screen, pictured below, whether your patient has a
PACEMAKER implanted. A check indicates that the patient DOES
have a pacemaker.
FRF ON
Operating Steps
1. Check the applied electrodes and verify the signal quality in all
leads.
Note
Once you initiate the pretest phase you can no longer change the
test settings.
6. Press pretest to initiate the test. The pretest screen will appear and
the pretest phase begins.
57A-S
Note
The acquisition screen has the same type of information in the
pretest, exercise and recovery phases.
F Selected protocol, exercise clock - starts with the exercise phase and
indicates its duration.
G Current phase, phase clock - each new phase (pretest, exercise,
recovery) resets the clock to zero.
H Current stage, stage clock - restarts with each new stage.
I Treadmill: belt speed and grade
bicycle ergometer: pedal speed (RPM) and load
METS (metabolic equivalents) see “METS Formula (Metabolic
Equivalents)” on page 6-22
RPP (Rate Pressure Product).
85A-S
86A-S
Pretest Phase
Overview
The pretest phase might consist of stages configured in each protocol.
Commonly used stages are:
Supine
Standing
Hyperventilation
Warm-up
The first QRS complex of the exercise phase is saved as the baseline
complex.
Operating Steps
1. Wait until the message Median update ceased... Relearn disappears,
the baseline ST measurement appears and the blood pressure has
been taken.
2. Press
a. the 12 lead button to initiate a recording on the thermal writer
or laser printer
b. the12SL button to acquire and analyze a baseline ECG
c. the medians button to print a baseline ECG on the thermal
writer or laser printer.
3. Press pretest to advance to the next stage.
Note
When the TWA algorithm is enabled, ensure that the pretest phase is
long enough (approx. 1 minute for update factor of 1/8).
Exercise Phase
Overview
The selected protocol will control the treadmill or bicycle ergometer.
When you enter the exercise phase:
the belt speed and grade or the bicycle load will change according to
the selected protocol
the exercise clock will start (phase and stage clocks restart at zero),
and
the system will start saving test data. This data can be viewed on the
tabs at the top of the screen (see “ST/Arrhy Window” on page 6-18 to
“Interpretation Window” on page 6-26).
Operating Steps
Treadmill
1. Press the exercise button to enter the exercise phase.
Warning
Fall Hazard — Your patient should wait until the
treadmill belt is moving before stepping onto the belt.
Otherwise severe injury could result from a fall. For the
same reason, avoid rapid changes in belt speed.
Note
When the TWA algorithm is enabled, ensure that the patient does
not run until the heart rate exceeds the TWA heart rate limit
(default value: 125 BPM, see “TWA Tab” on page 12-30).
Bicycle
The exercise test starts automatically when the patient starts pedalling.
Manual Control
Note
In the pretest phase, Shift + pretest or Shift + F1 allow you to return
to the previous stage.
Note
If a bicycle ergometer test is initiated with the exercise button, the
clock will start even when the patient has not yet begun to pedal.
Note
When the patient reaches the target HR, the HR reading turns red
and an audio signal sounds (if configured, see section “Miscellaneous
1 Tab” on page 12-24).
Note
Using the treadmill speed and grade buttons during a protocol puts
the treadmill in manual control. The system is no longer controlled
automatically by the protocol. The system will enter a new stage and
the phase name appearing in the Tabular Summary will be Manual.
The system chooses the two leads with the greatest amplitude for
arrhythmia analysis (click E to select another lead). The system starts
saving arrhythmias at the beginning of the pretest (20 arrhythmias
max., following preset priorities, see section “Arrhythmia Review” on
page 6-45).
A B C D E F G H
58A-S
K I
A Click the lead label to display the median complexes of other leads. The message
Scan indicates that the system has chosen the lead with the most significant ST
depression. (Leads aVR, aVL, V1 are not considered.)
B Click the tab to display the ST/Arrhy window.
C Click Superimposition to display the median complexes from all leads
superimposed (click again to restore normal presentation).
D Click Zoom to change the size of the median complex.
E Click the lead label to open a window for manual selection of the arrhythmia lead
(the system automatically selects the 2 leads with the greatest amplitudes).
F Arrhythmia label.
G Time at which arrhythmia occurred in the indicated phase.
H Number of ventricular ectopics per minute.
I ST measurements.
K Click arrows to adjust the E, J and J+x reference points (only in "manual" mode,
see section “ST/Medians / 12SL Tab” on page 12-29).
Note
Significant arrhythmias are displayed with red waveforms.
Medians Window
The Medians window displays the current median complex. The system
saves new median beats at the time intervals selected in the Protocol
A B C
59A-S
E D
forward
E Median label (the complex is updated at the time interval selected in the Protocol
Editor).
Waterfall Window
This window displays a graphic color representation of the median beat
changes in the course of the test (updated at 30 second intervals).
A B C D E
60A-S
B C A
61A-S
Treadmill
(speed x 26.8 x 0.1) + (grade/100 x 1.8 x speed x 26.8) + 3.5
METS =
3.5
speed = miles per hour
grade = percent
grade = percent
Bicycle Ergometer
weight = kg
Between two stages, the METS value will be interpolated. The correct
METS value of a stage is reached after a stage time of 60 seconds (default
value, can be changed, see section “Miscellaneous 1 Tab” on page 12-24).
Trends Window
The Trends window displays graphic trends of different parameters. The
trends are updated in 30 second intervals.
A B C
62A-S
forward
A B C
63A-S
forward
12SL Window
The 12SL window displays the 12SL resting ECG measurements and
interpretation acquired by pressing 12SL during preacquisition, pretest
and recovery.
A B C
64A-S
D E F
65A-S
Note
The vector loop is saved with the 12SL report only if the message
Median update ceased... Relearn has disappeared from the ST/Arrhy
window at the time the 12SL report is stored.
Note
The system stores up to five 12SL reports.
Interpretation Window
The Interpretation window allows the physician to enter an
interpretation of the test.
A B
66A-S
D C C E F
67A-S
H G
Page 1 of the median complexes will appear in the lower part of the
screen for reference.
A B C
D
E
F
G
68A-S
forward
Recovery Phase
Overview
Press the recovery button to advance to the recovery phase.
The clock begins timing the recovery phase. In recovery, the treadmill
speed and grade or the bicycle load will change according to the protocol
configuration.
2. Click Yes to confirm the test end. The Interpretation window appears
with the patient's ECG waveforms below.
Note
After you press the test end button, the following happens:
4 1 3 5 3 2 6 7
67A-S
Operating Steps
1. Enter or select the Reasons for Termination (multiple selections are
permitted).
69A-S
Note
Real-time reports can be printed with the following buttons on the
function keypad:
12 lead
medians
rhythm
8. Click the test end menu tabs to review and edit the data.
9. Click Print or press test end to print the final report (see “Report
Printout” on page 6-56).
10. Click Post Test Review to review the stored test information. This will
end the display of the patient ECG waveforms.
Note
If you do not want to go to Post Test Review:
F
G
H
I
J
K
L
70A-S
Click to display a list with more tests of the current patient (local
G
database).
Click to display a list with more tests of the current patient (MUSE
H
database system).
The Modify... buttons allow you to edit or complete the data in the
respective areas.
Tabular Summary
The Tabular Summary window displays test data in three tabular
formats:
stage format: 1 line per stage, or
detailed format: 1 line minimum per 30 seconds
all entries
A B C D
71A-S
Note
Only the stage format of the Tabular Summary is transferred to the
MUSE system.
METS Formula
See “METS Formula (Metabolic Equivalents)” on page 6-22 for details of
the METS formulas.
Between two stages, the METS value will be interpolated. The correct
METS value of a stage is reached after a stage time of 60 seconds (default
Graphic Trends
Overview
A B
C
D
72A-S
forward
Note
The system starts collecting trend data at the beginning of the
exercise phase.
942A-S
943A-S
1 2 3
949A-S
Note
Dashed lines appear in case of noisy signal and/or when the heart
rate is too high (see TWA Heart Rate Limit in Chapter 12, “System
Settings” under “TWA Tab” on page 12-30).
A B C
D
E
F
G
H
J 73A-S
Note
Reanalyze the sample cardiac cycles as described under “Reanalysis”
on page 6-41.
I Click icon to activate the zoom and reanalysis function (see next page).
J Click Reset to original to discard the results of the reanalysis and return to the
original measurements.
Reanalysis
Reanalysis allows you to manually change the position of the ST
measurement marks. On the reanalysis screen the medians of all leads
are superimposed. The currently selected lead is highlighted. The
measurement marks determined by the system are indicated. The marks
above the baseline can be edited.
To activate reanalysis
Click the Zoom icon on the Sample Cardiac Cycles screen
position it over a complex and click the left mouse button
A B C
F
G
74A-S
Performing Reanalysis
1. Click a measurement mark above the baseline and hold the mouse
button depressed.
2. Move the mark to its new position and release the mouse button.
3. Adjust the other marks in the same way.
4. Select the complexes to reanalyze.
5. Click the Reanalyze button.
6. Click ...Reset and Reanalyze... to restore the original analysis results.
(Reset to original restores the original results for all complexes of the
entire exercise test).
2. Position the cursor on the second measurement point and click. Now
you will see
the second caliper
the measuring line
the measurement results
3. Click the right mouse button to remove the measurement points.
A
B
C
D
E
F
H 938A-S
ECG Strips
This screen displays 10-second ECG strips
stored during the last 10 seconds of each stage
acquired manually, or
acquired automatically as configured in the Protocol Editor.
G A C D B E F G H
75A-S
Note
You can perform measurements as described under “Performing
Measurements on the Median Complex” on page 6-42.
Arrhythmia Review
The system will save a 10-second ECG strip each time an arrhythmia
occurs and when the Event Marker Full Disclosure button is clicked on
the acquisition screen (to a maximum of 20). When the event memory is
full, the strips will be overwritten according to preset priorities.
A H B C E D F G H I
A B C D E F G H
76A-S
Note
You can perform measurements as described under “Performing
Measurements on the Median Complex” on page 6-42.
- learn phase3 L
1. significant arrhythmias
2. The arrhythmia analysis algorithm will not output these events in the presence of
supraventricular arrhythmia (e.g. atrial fibrillation).
3. not in arrhythmia window, no arrhythmia strip (only full disclosure ECG)
A B C D E F
G
H
I
J
K
L
M
N
O
77A-S
2. Click again to mark the end of the segment. A second flag appears.
Printing
1. Click the printer icon. The print setup window opens.
2
3
8
7
5 517A-R
3. Click the Format-1 option button, if you wish to print only one lead.
4. Select the lead.
5. Click the Format-2 option button, if you wish to print multiple leads,
and select the leads.
6. This box indicates the number of pages to print; check paper supply
before printing.
7. Click the Save button to save the print setup.
8. Initiate the printout or close the window with Cancel.
Note
To save or print a 12-lead ECG, enable the zoom function as
described below. With button B you are then able to save or print the
ECG strip.
The operating controls basically remain the same. When you click the
printer icon, however, only that page will be printed.
A
B
78A-S
Audio ECG
The audio ECG allows the medical staff to assess the exercise test
quickly and easily (playback not possible with CASE systems because
CASE does not have the required output media).
The ECG is played back at 60 times the original speed, so that you can
listen to a 10-minute recording in 10 seconds.
Also refer to the Audio ECG Physician’s Guide, Part No. 2051167-001.
Waterfall
The Waterfall display is a graphic color representation of the median
beat changes in the course of the test (updated at 30 second intervals).
A B C D E F GH
79A-S
3D Graph
A B C
84A-S
12SL
This screen shows the measurements from the 12SL ECG analysis.
A B C DE F
80A-S
A Analysis results.
B Click to change the gain.
C Click to change the speed.
D Click to display the 12SL Report.
E Click to select another 12SL ECG.
F Click to view the corresponding Vector Loops (see next page).
Vector Loops
This screen shows the median complexes of the orthogonal FRANK leads
X, Y and Z as well as the vector loops in three planes.
A B C D
81A-S
A Median complex.
B Click torso to toggle between right and left sagittal planes.
C Click to change the gain.
D Click to return to the 12SL Report window.
Note
You can perform measurements as described in section “Measuring
the ECG” on page 5-15.
The vector loop is saved with the 12SL report only if the message
Median update ceased... Relearn has disappeared from the ST/Arrhy
window at the time the 12SL report is stored.
Report Printout
Click the Print button to print the report. The print setup window opens
where you can temporarily adjust the settings for the printout. Refer to
section “Final Report Tab” on page 12-23 for details on changing the
setup permanently.
1 2 3 6 4 1 5 8 7
82A-S
Note
Select the ECG Grid option if you print on plain paper.
Report Formats
In-Test Reports
Format Description
12SL Interpretative resting ECG with analytical statements on morphology, rhythm, and
conduction. Measurements of rate, intervals and axes. Available before the test, in pretest
and recovery. Formats include: 4 x 2.5; 4 x 2.5 +1, 2 x 5; H2 (5 s ECG at a speed of
50 mm/s x 6) and H1 (6-lead ECG at 12.5 mm/s, Medians at 50 mm/s).
3, 6, 12 or 15-Lead Report Presents raw ECG data in various formats: 4 x 2.5 s ECG; 4 x 2.5 +1 (2.5 seconds of raw
data plus 1 continuous 10 second recording of continuous rhythm); 2 x 5 (5 seconds of raw
data); 1 x 10 (10 seconds of raw data); H2 format (5 seconds of raw data at a speed of
50 mm/s x 6).
Rhythm Report A continuous, real-time recording of raw ECG data: 3, 6, 12 leads. Leads for arrhythmia
report correspond to leads on the screen (not with laser printer).
Arrhythmia Report Automatic 10-second documentation of arrhythmias showing 2.5 seconds of raw data prior
to ectopic beat. Leads for arrhythmia report correspond to leads on the screen.
Comparative Medians Report Baseline and current medians followed by 2.5 seconds of real-time rhythm in 12 leads. ST
level and ST slope reported for each lead. Also comes in H1 Medians format of 12
medians at 50 mm/s + 6 leads (selectable) of 10 seconds of rhythm at 12.5 mm/s.
Linked Medians Report 4 x 2.5 with rhythm lead. 3 channels of median beat which are linked together at the
current heart rate with a fourth 10-second channel of raw, real-time rhythm.
Note
The formats may differ slightly when the Swedish Reports are
selected.
Final Reports
Format Description
Configured Report User-configured portrait-style final report. User formats narrative text and selects data
fields to create report template. Configured Report automatically merges text and data
according to templates selected. For details on the Configured Report, refer to section
“Report Editor” on page 12-10.
Tabular Summary Tabular Summary of test by stage including time, speed, grade, workload, MET level,
heart rate, blood pressure, comments, with optional selections for double product (RPP),
ST level, SpO2 and user-configured entries. Can also be printed as Detailed Summary in
minimum of 30 second intervals. For instructions on Tabular Summary configuration, see
section “Configuring the Tabular Summary” on page 12-14.
Graded Exercise Summary Report One page overview of entire test with median morphologies for 12 leads from baseline and
either max. ST depression or peak exercise. For max. ST depression, report only prints
when a minimum of -0.5 mm of ST depression occurs in one of the following leads: I, II, III,
aVF, V2 to V6 (V1, aVR, aVL excluded; for ST elevation, -aVR is excluded).
Selected Medians Report Records median morphologies at baseline, max. ST depression, peak exercise and test
end for 12 leads. For max. ST medians, column only prints when a minimum of -0.5 mm of
ST depression occurs in one of the following leads: I, II, III, aVF, V2-V6. (V1, aVR, aVL
excluded; for ST elevation, -aVR is excluded).
Sample Cardiac Cycles Records median morphologies for 12 leads at user-defined intervals. Intervals defined in
Protocol Editor for each protocol by configuration of Store Median interval; 25 or 50 mm/s.
Trend Reports Records plots of PVCs, heart rate and blood pressure or any custom combination of
parameters (Trends: Custom). Also produces trend report of ST level and slope versus
time (Trends: ST Level/Slope, Trends: TWA). For details on the Trend Graphs, refer to
section “Configuring Trends” on page 12-15.
ST/HR Hysteresis A two-dimensional representation of ST level versus heart rate in exercise and recovery
phases.
ST/HR Slope Records linear regression of heart rate-adjusted slope for all leads, plus median
morphology of lead with highest slope.
ECG Strips Report of all 12 leads (also medians and recall stored in 12-lead format) stored during the
tests (manual or automatic). User can select all or some of the 12 leads to be printed.
Arrhythmia Review Report of 20 stored arrhythmias showing 2.5 seconds of raw data prior to the ectopic beat
through 2.5 seconds after the last. User can select all or some arrhythmias to be printed.
Vector Loops Report showing Y, X, Z leads and three-plane vector loop display of horizontal, frontal and
sagittal planes using Frank X, Y, Z lead set. Derived from associated 12SL report.
Sample TWA Cycles TWA medians report (see “Sample TWA Cycles” on page 6-43).
Swedish Median Report Special medians format for Sweden. Only if Swedish Reports are enabled.
1. Click Remote View on the initial screen to view the remote exercise
test screen.
2. Click the Setup button and enter the PC names in the left column of
the window.
3. Click the arrow buttons to enter the corresponding IP address
automatically. If the system does not find the IP address, you can
enter the address manually.
4. Click OK to clear the window.
5. Click the Select Station button and select the station.
A
B
C
83A-S
F E D
Caution
Patient Hazard — Disconnect TONOPORT from the
patient when it is connected to the PC.
Note
Check that the correct port for the ambulatory blood pressure
monitor is configured (see section “Devices Tab” on page 12-
53).
2. Wait for the time to be displayed, then connect the monitor to the PC.
B
C
D
E
F
G
H
I
181A-BP
E Click to delete the data stored in the ambulatory blood pressure monitor.
Operating Steps
1. Turn the monitor off and on again.
2. Wait for the time to be displayed, then connect the monitor to the PC.
3. Click Connect.
A B C D E F
182A-BP
4. Click the Clear Data button to delete data stored in the monitor (with
TONOPORT V devices, version 2.0 and later, this command will also
clear the patient ID).
Note
If the stored data has not been transferred to the system,
download it as described in the next section.
6. Select the protocol and click Programming. You can choose between
two protocols. Refer to section “Editing the BP Protocol” on page 12-
32 for information on configuring the protocols.
Note
If you are using TONOPORT V devices, version 2.0 and later,
check the displayed patient data to avoid data mismatch.
2. Wait for the time to be displayed, then connect the monitor to the PC.
3. Select the patient.
Note
If the patient ID has been stored in TONOPORT V devices of
version 2.0 and later, you are not required to select the patient.
Simply click Select and you will see the acquisition screen.
4. Click Connect.
5. Click Download Data. After the download, the Test Summary tab
appears.
Note
Do not forget to delete the data stored in the monitor. This is to
ensure that the monitor does not contain data from a previous
test when used on a new patient.
Test Summary
A B C D E
F
G
H
I
J
K
L
183A-BP
Click to display a list with more tests of the current patient (local
G
database).
Click to display a list with more tests of the current patient (MUSE
H
database system).
The Modify... buttons allow you to edit or complete the data in the
respective areas, as described below:
Patient Information
The Patient Information window opens. You can edit or complete the
data. Any changes you make will be assigned to this procedure, not to the
patient.
Note
The Patient Information can be assigned to the patient file or only to
the pending test.
184A-BP
Test Information
The Test Information window opens.
You can edit the data (the changes you make will be assigned to this
procedure, not to the patient) or display the other tabs by clicking the tab
headers.
185A-BP
Test/Personnel
1 2 4 3 5 6
052A-R
2. Enter the Reason for Test or select one from the list box. Multiple
selections are permitted.
3. Type the physicians' and technician's names or select them from the
list boxes.
4. Type any Comment about the test.
5. Answer the Extra Questions.
6. Enter an Order Number. This number will appear on the printed
reports and on the Test Summary.
Note
The Reason for Test library can be edited (see section “Miscellaneous
Tab” on page 12-33).
You enter the names of the physicians and of the technician in the
System Configuration (see section “General Tab” on page 12-40).
Medications
2 1 3
053A-R
Note
You can edit the medication library (see section “Editing the Reason
for Test, Medication Library” on page 12-36).
Measurement Results
In this window you can temporarily modify the statistics intervals and
the blood pressure limits for this report.
A B C D
193A-BP
A Select to perform the statistical analysis with the Statistics Intervals from
the ambulatory BP device; otherwise the intervals set in the system will be
used (see section “Editing the BP Protocol” on page 12-32). The check box is
accessible only when 1 or 2 BP periods have been selected at TONOPORT V.
B Select box to change the statistical night time.
C Select box to change the statistical wake-up time.
D Click buttons to change the blood pressure limits.
A B C D E
186A-BP
chapter .
D Click icon to confirm the test results.
E Click icons to cut, copy and paste text in the Interpretation area.
Note
As an alternative you can double-click a text to copy it from left to
right.
Graphics
The Graphics screen shows all BP values acquired over the monitoring
period. The blood pressure values are shown at the top, the pulse
pressure values (systolic pressure minus diastolic pressure) below. You
can show or hide the pulse pressure values (see “Ambulatory Blood
Pressure Measurement Setup” on page 12-31).
You can choose between display formats 1, 2 and 3. This format can be
preset in the System Settings and is saved as a user-specific
configuration. The selected format will also be used for printouts.
For measurements taken with TONOPORT IVa, the ECG strips are also
displayed.
A B C D
187A-BP
Tabular Summary
Note
Individual measurements in the Tabular Summary can be
suppressed by double-clicking or by clicking with the right mouse
button; they can be restored in the same way. Suppressed
measurement data are excluded from the statistical analysis.
188A-BP
Hourly Averages
The diagram shows the Hourly Averages of all measurements taken
during the monitoring period. Click Tabular to view the same values in
tabular format.
189A-BP
Statistics Summary
The Statistics Summary shows one histogram each for the systolic values
(left) and the diastolic values (right) as well as the frequency distribution
given as a percentage.
The table below indicates the maximum, minimum and mean values as
well as the standard deviations.
The percentage of readings that exceeded the set limits is shown at the
bottom.
Similar report pages exist for the day time, the night time and the wake-
up time.
190A-BP
Report Printout
Click the Print button to print the report. The print setup window opens
where you can temporarily adjust the settings for the printout. Refer to
section “Miscellaneous Tab” on page 12-33 for details on changing the
setup permanently.
1 2 1 3 2 4 5
191A-BP
Sensors
SpiroSoft respiration flow sensor for FVC measurements
LF 501 respiration flow sensor for VC and FVC measurements
For information on setup and operation of the sensors, refer to the
separate operator manuals.
Check whether the system is set up for the correct sensor (see section
“Devices Tab” on page 12-53) and connect the sensor to the configured
port.
101A-SP
3 2
The Patient Information tab is open.
1. Verify the patient's name and ID. If the data is incorrect, click
Modify to change.
Note
Depending on the selected reference value equation, the program
requires the gender, height, weight, date of birth, and race data for
calculation of the spirometry reference value (see section “Spirometry
Reference Value Equations, Interpretation Modes, Measurements”
on page B-13).
You can assign a location name to the location number (see section
“MUSE Tab” on page 12-55).
Test/Personnel Tab
1 2 3 5 4 6 7
52A-R
Note
The Reason for Test library can be edited (see section “Editing the
Reason for Test, Medication Library” on page 12-36).
You enter the names of the physicians and of the technician in the
System Configuration (see section “General Tab” on page 12-40).
Medications
2 1 3
53A-R
Note
You can edit the medication library (see section “Editing the Reason
for Test, Medication Library” on page 12-36).
With the ATS mode (American Thoracic Society), the program will first
determine the best expiratory and inspiratory curves of a test, based on
the largest sum of FVC + FEV1 or FIVC + FIV1. Then the system will
determine the test acceptability, using the ATS criteria (see
“Acceptability Criteria” on page 8-10). The best three tests are sorted
according to the largest sum of FVC + FEV1 and stored (see ATS
Summary table). In the next step, the program will determine whether
the best two tests meet the ATS reproducibility criteria (see
“Reproducibility Criteria” on page 8-10). The displayed overall
measurement reflects the flow-volume loop of the best test. The FVC and
FEV1 values are the best values from all tests; similar to the flow-volume
loop, the other values are taken from the best test.
210A-SP
Envelope ATS
Note
For flow-volume measurements, the patient's cooperation is
essential.
The ambient conditions should be checked every day before
performing spirometry tests; the room temperature is the factor
with the most impact on the test results (see section “Acquisition
Tab” on page 12-35).
Refer to section “Acquisition Tab” on page 12-35 for information
on selecting the Measurement Mode, the Interpretation Mode, or
the Equation for Reference Value Calculation or on entering the
PT Code.
A B C D E F G
H
I
J
K
L
M
N
O
P
201A-SP
S R
Click to load system setups stored for different system users (see section
M
“Custom Setups, Factory Setup” on page 12-38).
Click to display the spirometry setup menu (see section “Spirometry
N
Setup” on page 12-35).
Warning
Incorrect Measurements, Risk of Infection — The
pneumotach including its mouth piece is designed for
single use. Use a new pneumotach for each patient.
Check that the PT code on the pneumotach is the same as
the PT code (R) shown on the acquisition screen.
Clean the nose clip after each patient and attach new
foam pads.
2. Attach nose clip to ensure that the entire respiratory volume passes
through the sensor.
3. During the measurement, the patient should sit straight and hold
the SpiroSoft sensor in a horizontal position.
4. Click Start Test, then click OK to confirm that the patient is not yet
breathing through the sensor.
7. Click Stop Test (beep). The system also emits the beep when
SpiroSoft terminates the measurement automatically.
Note
The flow-volume measurements in the ATS mode must meet specific
criteria:
Acceptability Criteria
Reproducibility Criteria
The flow-volume loop of the current test and of the best test are
displayed. The table shows the measuring values (Act.), the reference
values and the percentage deviation of the current test.
A B C D E F
202A-SP
Bronchodilator Test
The bronchodilator test always consists of two measurements: one before
and one after medication. After medication, measure the dilatation as
follows:
When you select the Post Test Review, you have another opportunity to
select a pre-dilation test for comparison with the current test.
VC Measurement
1. Explain the test maneuver to the patient.
2. Attach nose clip to ensure that the entire respiratory volume passes
through the sensor.
3. Click Start Test and ask your patient to perform the following
maneuver when the orange indicator on the sensor lights up:
Note
The maneuver must be concluded within 20 seconds of clicking
Start Test.
The program terminates the measurement when it does not
detect a flow reversal within 4 s of two identified inversion
points, or after 5 full breaths.
The curve window shows the volume-time curve (expiration only) and the
table indicates the measuring values and the percentage deviation from
the reference values.
FVC Measurement
1. On the acquisition screen click FVC Measure..
A B C D E F
G
H
I
J
K
L
M
N
203A-SP
Click to display a list with more tests of the current patient (local
H
database).
Click to display a list with more tests of the current patient (MUSE
I
database system).
The Modify... buttons allow you to edit or complete the data in the
respective areas, as described below:
Patient Information
The Patient Information window opens. You can only enter or correct
data that does not affect the calculation of reference values.
184A-BP
Test Information
The Test Information window opens (see “Entering Test Information” on
page 8-4).
Note
The Test Information is not assigned to the patient, but to the test
data.
FVC Measurement
Click the FVC Measure. tab to display the acquisition screen with the
corresponding flow-volume loop and the measuring values. The status
line indicates the ambient conditions set at the time of the measurement.
A B C D E
206A-SP
chapter .
D Click icon to confirm the test results.
E Click icons to cut, copy and paste text in the Interpretation area.
Note
As an alternative you can double-click text to copy it from left to
right.
Report Printout
Click the Print button to print the report. The print setup window opens
where you can temporarily adjust the settings for the printout. Refer to
section “Miscellaneous Tab” on page 12-36 for details on changing the
setup permanently.
1 3 2 4 5
207A-SP
Configured Report
Standard Page with Flow-Volume Loop, Volume-Time Curve,
Measurements
Interpretation (appears on the Standard Page)
2. Select the number of copies to print.
3. Select a format for the Configured Report (click Modify... to display
the Report Editor, see section “Report Editor” on page 12-10).
4. Select the printer.
5. Click the Print button to initiate the printout.
1. Click Compare.
A window opens listing all the patient's spirograms.
2. Select a test.
3. Click OK to clear the window.
A B C
204A-SP
A B
209A-SP
Overview
All patient records are saved to the local database. The patient records
can be
Click Local Database to open the window with the names of all stored
patients.
A B C D E F G H I J
K
L
M
N
O
P
Q
R
S
101A
1. Select all in the Patient list box, if you would like to see all stored
patient records and not just those of the selected patient.
Note
Patient records marked with an asterisk * are stored on an external
medium.
You can select multiple patient records, holding the Shift key
depressed.
You can edit the information. Changes are saved automatically to the
local database when you exit the screen. To print the displayed patient
record, click the Print button.
Note
You can also display a patient record by double-clicking.
When you have selected the option Delete local test data after transfer
to MUSE on the MUSE tab (see section “MUSE Tab” on page 12-55),
all test data will be automatically deleted from the local database.
Only the patient demographics remain stored.
The Combined Report can be edited and printed, but it cannot be saved.
2. Enter the MUSE password, if prompted. The MUSE web page will
appear.
3. Select Display with Frame or Display without Frame.
4. Submit query or identify the patient record to be retrieved by patient
ID or name.
5. Display the individual documents of the patient record.
6. Print reports using the Acrobat Reader printer tools on screen (tool
bar directly above the report page).
DO NOT use the Internet Explorer printer tools (toolbar at the top of
the screen).
Save to Archive
To save the patient record(s) to a SD card, click the Save to Archive
button.
Note
Use SD cards ONLY to transfer data from MAC ECG systems.
Note
This button is enabled only if the data transfer to the MUSE
system has been selected on the MUSE tab (see “MUSE Tab” on
page 12-55).
Data sent to the MUSE system can be edited only when the
program is configured accordingly (see “MUSE Tab” on page 12-
55).
Data sent to the MUSE system can be viewed and printed with
the MUSE Browser. Please note: Select the Landscape format
before each printout.
Transfer to MUSE
To transfer the patient record(s) to a MUSE system, click the Transfer to
MUSE button.
Note
This button is enabled only if the data transfer to the MUSE
system has been enabled on the MUSE tab (see section “MUSE
Tab” on page 12-55).
Data sent to the MUSE system can be edited only when the
program is configured accordingly (see “MUSE Tab” on page 12-
55).
Data sent to the MUSE system can be viewed and printed with
the MUSE Browser. Please note: Select the Landscape format
before each printout.
Send
Click the Send button to send the patient record(s) via modem or save
them to a storage medium.
Export
Click the Export button to export patient record(s) using different file
formats.
A B C D E F
102A
Settings
A B C D E F
102A
Note
If you do not use the PDF995 PDF printer driver to export data in the
PDF format, you must set the export folder in the corresponding PDF
writer application.
If you wish to work temporarily with the offline (local) database (house
call), you can copy the patient list (without any examinations) to the local
database. This can be done manually or automatically.
Manual
1. Connect the PC to the data network.
2. Click System Configuration to display the system configuration
menu.
3. Click Database...
4. In the Database Setup window select use offline database.
162A
Now you can disconnect the PC from the data network and continue
working with the local database as usual, i.e., you can select a patient
and perform a test or admit a new patient to perform a test.
When you have finished working with the local database, reconnect the
PC to the network and deselect the use offline database check box: new
tests as well as new patients for whom tests exist will be copied to the
database on the server.
Automatic
If you select the automatic switch to offline database check box in the
Database Setup window, an additional button will be displayed when you
shut down the program: Use Offline Database.
163A
164A
When you terminate the program with this button, the patient list will
be copied to the local database. Continue working with your PC as usual.
When you have finished working with the local database, exit the
program, reconnect the PC to the network and start the program. New
tests as well as new patients for whom tests exist will automatically be
copied to the database on the server. A message informs you of the copy
procedure.
Overview
The program can be used in combination with the following
administrative tools:
Description
The CASE program supports communication with administrative tools
via the following interfaces:
Note
Please refer to the separate descriptions supplied.
After selecting the patient, you choose the test type. This will
automatically transfer you to CASE which you operate as usual. If you
wish to select another patient, you will have to exit the examination
program and choose the patient in the administrative tool.
After ending the test and terminating the examination program, the
most important data, but no curves, will be transferred to the
administrative tool.
The patient is admitted to the administrative tool with last name, first
name, patient ID, date of birth, height and weight.
Note
GE Healthcare is not in a position to guarantee full compatibility of
administrative tool with CASE in all situations.
Overview
DICOM (Digital Imaging and Communications in Medicine) is an open
data exchange standard for medical information.
DICOM standardizes both the file format for data storage as well as the
communications protocol for the exchange of these data.
Note
A valid DICOM option code will activate the DICOM interface as well
as the DICOM tab.
104A
An order list query is automatically sent via the network. These are the
default search criteria:
Note
It is also possible to view the order list without selecting a patient
first. An order list query is automatically sent via the network. These
are the default search criteria:
3. If required, you can changed these default settings. Then click Find
to start the query.
4. Select the order from the list. More details can be found below the
order list.
5. Click Start Procedure.
The procedure with the patient data of the selected order will start. For
exercise stress tests, a test protocol assigned to the order will
automatically be selected (see Field Service Manual).
Once the test has been successfully terminated (or after it has been
analyzed, see Field Service Manual), the data will be sent to the
archiving system.
When the data were successfully archived (this may take more or less
time, depending on the archiving system used), the exam is labelled
Archived (DCM) in the examination list. Depending on the configuration,
it may not be possible to edit the archived exam.
109A
H G F E
A The content of these fields depends on the program configuration.
B Search criterion for the workstation: find orders for a specific workstation or find
orders for all workstations.
C Search criterion date: find orders for the selected time period or find orders
irrespective of a time period.
D Search criterion patient name/patient ID: find orders for patients with a given
last name or patient ID or find orders for all patients.
E Click to find orders that match the specified search criteria.
F Click to quit the dialog window.
G Click to reject the order, stating the reason.
H Click to initiate the procedure of the selected order.
Acquisition Tab
The Acquisition tab with the acquisition settings is already on top.
A B C D E F G H I J K L M N
91A-R
insert chapters
insert lines
delete chapters
92A-R
A B C D
Note
Assigning the correct names to the groups is the responsibility of the
physician.
2 4 2 4
93A-R
You can change the lead sequences according to your needs and save
them under the same name or under a new name. You may also create
new lead sequences.
Note
The Standard and Cabrera lead sequences cannot be modified.
1 3 2 4 5 6 7
94A-R
Miscellaneous Tab
A B C D E F G
95A-R
1 2 3 4 5 6 7 8 9 10
96A-R
Configured Report
10-second ECG (up to 3 different formats, with or without
Interpretation)
Medians
Vector Loops.
2. Select the screen format for the 10-second ECG. You can choose two
more formats.
Report Editor
The Configured Report Editor allows you to create a maximum of 10
templates (e.g. letter to referring physician). Two templates (Reports 1
and 2) are preconfigured and cannot be changed.
Click Modify... on the Final Report tab to display the Configured Report
Editor window.
You can enter text and insert report items from a list (e.g., patient name,
heart rate). At the time the report is printed the program will replace
these report items with the actual data.
1 2 10 4 11 5 3 6 7 12 8 9
97A-R
A B
944A
10. Click button to save the template. To do so, enter the template name
first. This name will allow you to select the template later on when
configuring the printed report. If edited, Reports 1 and 2 must be
saved under a new name.
11. Click to delete the template.
12. Close the Configured Report window.
Writer Tab
The Writer tab with the writer settings is already on top.
A B C D E F G H I J K L M N
98A-S
Screen Tab
Click the Screen tab to open the window.
A B C D E F G H I K
99A-S
910A-S
Note
You can enter the values for columns 1 through 5 either directly in
the Tabular Summary or use the comment key.
Configuring Trends
Click Configure Trends on the “Screen Tab” on page 12-13 to display the
configuration window.
1 2 3
911A-S
912A-S
move
bigger font
smaller font
3. Click the appropriate item in the menu to change the font size.
4. Click Move to choose another location for the item. A box appears
around the selected item.
5. Click in the box and, holding the mouse button down, move it to its
new position.
Note
You can define different setups for bicycle ergometers and treadmills.
You can change the lead sequences according to your needs and save
them under the same name or under a new name. You may also define
different lead sequences for the screen display and the printouts.
Note
The first two lead sequences are fixed and cannot be modified.
3 4 1 2 5 6 7 8
913A-S
A B D E F C D E F
914A-S
Note
You can select protocols for pharmacological tests (e.g.
Dobutamine) at D.
A B C D E F G H I J
915A-S
Note
The 12 lead and Median report formats are selected on the Writer
tab.
The medians saved with Store Median are included in the Medians
report.
A Phase name.
B Selected protocol.
C Stage name.
D Select or deselect a ramping phase (= continuous speed/load increase;
selectable only when phase name, e.g. BRUCE, is highlighted; separately
adjustable for each phase).
Caution
Bruise Hazard — DO NOT specify BP intervals shorter
than 2 minutes. The BP readings may be incorrect and
the tissue might be damaged.
4 5 2 6 8 7
915A-S
2. Click the New Protocol button to display the window with all protocol
templates.
3. Select a template from the list.
Note
New protocols can be created only when the displayed protocol is
blank or was deleted.
1 6 2 3 4 1 4 1 4 1 5 4
916A-S
Configured Report
Graded Exercise Summary Report
Selected Medians Report
Tabular Summary (stage/detail)
Trend Reports (only the selected graphs will be printed)
Exercise Test Interpretation (XTI)
Sample Cardiac Cycles
ECG Strips/Arrhythmia Review:
All: prints all stored strips or arrhythmias;
Selected: prints ECG strips or arrhythmias from the post test
review.
12SL/Vector Loops
Waterfall
Sample TWA Cycles
with measurement marks: median complexes with ST
measurement marks.
2. Select a report format.
3. Click Modify... to edit the report or create a new Configured Report
(see “Report Editor” on page 12-10).
4. Select the number of copies.
5. Select the gain.
6. Click button to print the exercise test settings.
Miscellaneous 1 Tab
Click the Miscellaneous 1 tab to open the window.
A B C D E F G H I J K
917A-S
I Select the Max. Load Interpolation function (only possible when the exercise test
is performed on a bicycle ergometer). The maximum load is the load of the last
completed stage plus the duration of the last stage x load increase/stage time.
Example:
load of the last completed stage: 125 Watts
last stage aborted after 1 minute
load increase: 25 Watts
stage time: 2 minutes
125 W + 1 x 25/2 = 137.5 W
This value is also used to determine the percentage of the target load.
J Select Save full disclosure ECG, then choose the phase when storage of the full
disclosure ECG should start, the sampling rate and whether or not you want to
record the audio ECG.
K Select the METS Interpolation Time.
Miscellaneous 2 Tab
Click the Miscellaneous 2 tab to open the window.
918A-S
Editing the Comments, Test Types, Reason for Test, Reasons for Termination,
Medications Library
1. Click Modify.... to open the edit window.
Note
ST/HR Index is the maximal change in ST depression as a function of
change in heart rate during recovery.
Note
You can edit the eight preset topics in the window on the left. Do not
edit the topics, if data will be sent to MUSE. Additional statements
can be entered in the window on the right. To do so, click in the blank
space below the last statement.
1 2 3 4
919A-S
1 2 3
920A-S
1. Click in the text box for the first question and type a text (20
characters max.).
A B C D E F G H
922A-S
TWA Tab
Click the TWA tab to open the window.
During the exercise test the TWA algorithm analyzes the data in the
background. In each of the selected leads, the algorithm calculates the
even and odd medians, the TWA value and the noise limit. The data is
saved at intervals of 15 seconds. Verify the TWA settings on this tab
before the test.
A B C D E F
923A-S
A Select the TWA Heart Rate Limit (TWA values whose HR exceeds this limit are
excluded from calculation of the maximum TWA value and are identified with a
question mark on the display).
B Select the TWA Noise Limit (TWA values whose noise value exceeds this limit
are excluded from calculation of the maximum TWA value and are identified
with a question mark on the display).
C Heart rate: when first reached, the corresponding TWA value is stored
separately.
D Click down arrow to select the update factor.
E Select the check box to enable TWA analysis at test begin, or deselect.
F Click to reset the values to the defaults.
Note
During the test, TWA data records are saved at given points in time.
Each TWA data record provides the following information:
Acquisition Tab
The Acquisition tab is already on top.
A B C D E F G H I J K
192A-BP
2 3 1 4 5 6 7 8 9 10
194A-BP
Miscellaneous Tab
Click the Miscellaneous tab to open the window.
A B C D E F G H I J K
195A-BP
insert chapters
insert lines
delete chapters
Spirometry Setup
The settings specific to the spirometry modality can be entered either
from the spirometry acquisition screen where you click Setup to display
the setup menu, or from the General tab of the System Configuration (see
“General Tab” on page 12-40).
Acquisition Tab
The Acquisition tab is already on top.
A B CD E F GH I J K
205A-SP
Note
Exiting the spirometry setup menu with OK confirms all data and
the PT code. At the same time, date and time in the PT Code line on
the acquisition screen will be updated.
Miscellaneous Tab
Click the Miscellaneous tab to open the window.
A B C D E F G H I J K
208A-SP
insert chapters
insert lines
delete chapters
You access the setup menu by clicking the Custom Setup button on the
acquisition or pre-acquisition screen, or from the the General tab of the
System Configuration (see “General Tab” on page 12-40).
A B C
924A
Save
1. Click the Save button.
2. In the Save Setup window, select the target drive from the list.
Hard drive
Disk drive A
3. Select a line in the Setup window A.
4. In the Save As box, enter a name.
5. Click Save....
Note
With this command you save the setups of all modalities, not just of
the selected modality.
Load or Delete
1. Select the drive.
3. Click Load to load the setup, or click Delete to delete the setup.
Note
Be sure to save the current setup BEFORE loading one of the stored
setups, or your modifications will be lost.
Exercise Tests
Remote View
Spirometry Tests
ambient conditions
System Configuration
System Configuration
Click the System Configuration button on the initial screen to display the
menu.
General Tab
A B C D E F G
H
I
J
K
L
M
N
P
Q
R
S 925A
A Test Patient: Select check box to run a test using simulated patient data.
Select the function each time before starting a simulated test.
WARNING
Interpretation Hazard—Select the simulated
patient data for demonstration purposes only
to avoid misinterpretation of the data.
D Trace Background: Click to select the background for the signal traces.
ECG Grid: Select the grid format for the displayed ECG.
Report ECG Grid: Select the grid format for the printed ECG.
Screen Diagonal: Select the size of your PC monitor
(custom, see “Adjusting the Screen Size for Widescreen Monitors” on page
12-42).
E Enter the physician and technician IDs. When sending data to a MUSE
system, be sure to match User ID numbers assigned at the MUSE system to
User ID numbers assigned at the CASE unit.
F Settings for temporary use of the offline (local) database (see “Working
Temporarily with the Local Database” on page 9-12).
Click to enter a percentage value for the minimum free hard drive storage
capacity. You will be alerted when the capacity drops below this value.
G User List: Click to define the users, groups and privileges for individuals
working with the system (see next page).
Modify Password: Click to modify the current user's password (not with
Windows user list).
H Click to display the assignment list (available only when unassigned tests
exist, see “Assignment List” on page 12-47).
I Click to transfer resting ECGs (see “Receiving Resting ECGs from Other ECG
Devices” on page 12-45).
J Click to copy patient records to another medium. The patient records remain
stored in the local database (see “Copying Patient Records” on page 12-48).
K Click to archive patient records not currently used on an external medium (see
“Archiving Patient Records” on page 12-49).
L Click to compress patient records (see “Compressing Patient Records” on
page 12-50).
M Click to change the system date and time (see “Changing System Date and
Time” on page 12-50).
N Click to perform a statistical analysis of all tests performed with the system
(see “Procedure Statistics” on page 12-51).
O Click to display the Custom Setup screens for
Resting ECG (see “Resting ECG Setup” on page 12-3)
Exercise Test (see “Exercise Test Setup” on page 12-12)
Spirometry (see “Spirometry Setup” on page 12-35)
Ambulatory Blood Pressure (see “Ambulatory Blood Pressure Measurement
Setup” on page 12-31)
Export (see “Settings” on page 9-11)
Printer (see “Printer Settings” on page 12-52)
Custom Setup (see “Custom Setups, Factory Setup” on page 12-38).
P Click to access the service screen (requires Service password).
Q Click to save or load the System Configuration.
R Click to quit the program.
S Click to print the System Configuration.
Note
If you cannot remember the supervisor password, please contact the
GE Healthcare Technical Support or your local Service
Representative.
2. Click .
3. Adjust the height of the displayed window to 10 cm, using a ruler.
4. Adjust the width in the same way.
5. Clear the window.
3 4 5
945A
User Administration
User List (G on the General tab, on page 12-40) displays the list of all
registered system users if you select the Enable Password Function
check box in the preceding window. You can edit user data with Modify...
(select name first), delete a user with Delete... or register a new user with
New....
If you enable Use Windows User List, the user list of the specified
Windows domain will appear.
Note
Note down the administrator's user name and assign all privileges to
the administrator.
Call up the capture settings from the User List. The selected events will
be captured. The records are saved to the Event Log. You invoke the
Event Log with Start ->Programs ->General Administration ->Event Log.
Select the capture item in the Event Log. The source for all entries from
this application is CARDIO. These details are shown after double-
clicking the entry:
event category
user
cart/device number
patient ID
test type
test date
test time
Note
Only the System Administrator is authorized to perform these
administrative tasks.
4 5 1
2
3
926A
1. Enter the user name and ID number. If you enter the last name
before the first name, insert a comma in between. Otherwise the
name entered first becomes the first name and the name entered
second becomes the last name, when the user is registered in the
MUSE system. Please note that MUSE accepts first names with a
maximum length of 10 characters, last names must not be longer
than 16 characters. When sending data to a MUSE system, be sure to
match User ID numbers assigned at the MUSE system to User ID
numbers assigned at the CASE unit.
Note
Do not use any special characters in your passwords. The minimum
length of a password is 6 characters.
Note
If an ECG is received without patient ID (not even blanks), it will
automatically appear in the list for manual assignment with the
special patient ID “000000“.
4
927A
Click Receive Data from ECG Device (I on the General tab) to display the
setup window.
Note
Sometimes, when a patient ID has not been entered, ECG recorders
may send a series of blanks instead of a patient ID.
2. Select this option button to receive data from a storage medium and
specify the directory.
3. Select this option button to receive MAC ECGs from a USB or
memory card (SD card) and specify the directory. Then insert the
USB or memory card adapter into the appropriate drive.
Note
Insert the SD card with the bevelled edge pointing to the front right
(A). Be sure to use USB SD card adapter in the correct drive (B).
6. Push the Copy key on the ECG recorder or send stored ECGs (see
separate Operator Manuals that come with the ECG recorder).
Note
It is possible to simultaneously send the ECGs from up to 4 ECG
recorders via modem to the system. To utilize this feature, an extra
PC is required as the communications server. This server is part of a
network through which the ECG can be routed to the system or to a
MUSE system database. For details, please contact GE Healthcare
Technical Support or your local Service Representative.
Assignment List
In the assignment list, you will find all tests that were not automatically
assigned to a patient record. This list appears whenever incoming ECGs
cannot be assigned, or you can display it manually from the General tab
(see “General Tab” on page 12-40).
C B A
940A
Click Copy Patient Records (J on the General tab) to display the setup
window.
2 1 4 3
928A
Do not copy the data to the root directory (e.g. d:\), create a subdirectory
(e.g. d:\data\).
If you want to view an archived patient record, the system will indicate
the storage medium. Archived patient records can be transferred to the
MUSE database system only after they have been restored.
Click Archive Patient Records (K on the General tab) to display the setup
window.
6 1 2
929A
1. Enter a date: patient records older than this date will be archived.
Note
When working in a network environment, quit the program before
archiving or compressing patient records. Run a backup of all data
before archiving or compressing them. Please note that each file to
archive must be smaller than the capacity of the storage medium.
1. Enter a date: patient records older than this date will be compressed.
2. Click the Compress... button.
Note
When working in a network environment, quit the program before
copying or compressing patient records. Run a backup of all data
before copying or compressing them. Please note that the size of each
copied file must be less than the capacity of the storage medium.
A B F C D E
930A
Note
Access to the Date/Time Properties dialog is restricted to authorized
system users.
Procedure Statistics
Click Procedure Statistics (N on the General tab) to display the setup
window.
3 2 1 4
932A
Printer Settings
Click Printer (O on the General tab, page 12-40) to display the setup
window.
4 3 2 1 5
946A
Devices Tab
Click the Devices tab to open the window.
A B C D E F G
933A
A Select check box, then select a port for the metabolic cart.
B Click to select the BP monitor model and the port.
C Click to select the treadmill model and the port.
D Click to select the spirometry device type and the port (none indicates that the
modality is inactive).
E Click to select the SpO2 monitor model and the port.
F Click to select the ergometer model and the port.
G Click to select the ambulatory BP device and the port (none indicates that the
modality is inactive).
H Enter or select the path to the Holter ECG program.
I Select/deselect an external program for spirometry, enter or select the path.
J Specify the TTL output signal characteristics.
K Select up to 4 analog signals (output A, see “Connecting the System
Components” on page 2-5). The selections become effective only after you
have started the exercise test mode.
Modem Tab
Click the Modem tab to open the window.
A B C D E
934A
A Select modem port and baud rate (same baud rate at sending and receiving
units).
B Select the modem type.
C Field for entry of the initialization string (entered automatically for MultiTech
and ELSA modems).
initialization string for other modem types (e.g. MultiTech 56K):
AT&FM0\X3V0S0=1
D Field for entry of the dial mode (pulse/touch tone dialing).
E Field for entry of the Hang-up Prefix and Hang-up Suffix.
Warning
Patient Hazard — All modems connected to the system
must meet the requirements of IEC 60950 or UL1950.
The specific regulations valid in your country must also
be observed. The modem must be installed within the
medical location, but not in the patient vicinity.
Note
For Dial Mode, Hang-up Prefix and Hang-up Suffix, please refer to
the user manual of your modem.
MUSE Tab
Click the MUSE tab to open the window.
Note
CASE is compatible with the following versions of the MUSE system:
Pre 5C, 5C to 7.1 or 7.2 and above.
N O
A
B
C
D P
E
Q
F
G
H R
J S
T
K U
L
M
935A
A Select to retrieve patient data and orders from the MUSE system.
B Enter information required for communication with the MUSE web system.
C Select, if SSL is configured at the MUSE Web server.
D Box for entry of the port number.
E MUSE exercise test lists (comments, reason for test, interpretation, etc.) are
used: Yes/No. Click Synchronize Lists to load the lists from the MUSE Web
server.
F Select when data is not transferred to a MUSE system.
G Select to transfer data to a MUSE system via disk.
H Select to transfer data to a MUSE system via a network. Enter MUSE FTP
server information (MUSE version 5A and above).
Note
Use the GE Healthcare network installation kit to ensure all patient
isolation requirements are met.
3 5 1 4
936A
1. Click Modify....
2. When asked Change software option. Are you sure?, click Yes to clear
the window.
3. Enter the option code in the appropriate field, or
4. Click Load to load the option code from disk.
5. To enable an option for a limited period of time, click the icon and
enter the expiration date (if not yet displayed).
Note
If you are working with the simulated test data Test Patient you
can activate all options with the code number "123".
Note
After selecting the language, first load the factory settings to set all
text libraries to the new language (Custom Setup — Factory Settings
— Load).
EMR Tab
Click the EMR tab to open the window (with XEMR option only).
A B C D E F
941A
A B C D E F
947A
D Click to open a dialog to add shared directories. You select a folder for which
you have permission to save data. Normally, a user will be created and the
rights needed to save data are assigned to this user. On the cart, this user is
used to log onto the system.
After this, the folder is regularly checked and existing data records are
transferred.
E Click to open a dialog to add inbound connections (see below).
F Click to save the settings.
A B C D E
948A
A Click to select the target directory (the existing directory where the incoming
data are stored will be added here). The connectivity server will monitor this
directory and the data received will be downloaded.
B Click to select the type of connection: Serial or Modem.
C Click to select the modem type (see “Modem Tab” on page 12-54).
D Click to select the port for the connectivity server. The server will receive data
via this port (usually in conjunction with serial/IP software which provides
virtual COM ports and is able to map TCP/IP (LAN) links to COM ports).
E Click to select the baud rate. For modems, a maximum baud rate of 57600 is
supported, for serial connections, a maximum baud rate of 115200.
DICOM Tab
Note
A valid DICOM option code will activate the DICOM interface as well
as the DICOM tab.
A Cleaning and
Maintenance
Warning
Shock Hazard — Disconnect the power cord from the wall
outlet before cleaning or disinfecting the system.
Use a moist cloth to wipe the surface clean; do not allow liquids to
enter the system. All cleaning agents and disinfectants commonly
used in hospitals and containing up to 70 % alcohol are suitable. If
liquids have entered the system, notify service to have the system
inspected for damage before it is used again.
DO NOT use disinfectants with a phenol base or peroxide compounds
to disinfect the external surfaces.
Cables
Disconnect the cable from the system before cleaning or disinfection.
To disconnect cables, always pull on the plug, not on the cable.
Rub the cables clean with soap water, use a disinfectant for
disinfection. All cleaning agents and disinfectants commonly used in
hospitals and containing up to 70 % alcohol are suitable. Do not
immerse the cables in liquids.
Monitor
For cleaning and disinfection of the monitor housing, refer to the
information given above under „Equipment Surface“.
Clean the LCD monitor surface with a dry, lint-free, soft cloth. If
required, moisten the cloth with water or with the combination
cleaners for housing and screen Alklar, Sidolin manufactured by
Henkel.
Manufacturer:
Henkel KG a.A.
http://www.henkel.com
Maintenance
Before Each Use
Before each use, visually inspect the equipment, the cables and the
electrodes for signs of mechanical damage.
Technical Inspections
For safety, the devices require regular maintenance. To ensure
functional and operational safety of this system, Technical Inspections
should be carried out on an annual basis.
B Miscellaneous
Bicycle Ergometers
Connect bicycle ergometers to the COM ports.
Model Configuration
Ergoline 900/900 L/900 EL Ergoline 900
eBike basic, eBike comfort, eBike L, eBike
eBike EL
Plug the round connector of the connecting cable into the J10/RS 232
port of the ergometer and the other end into one of the COM ports of the
system.
Treadmills
Connect treadmills to the COM ports.
Model Configuration
Treadmills without digital communication internal treadmill
interface
GE Healthcare treadmills
T-2000 T2000
T-2100 T2100
Model Configuration
integrated in the ergometer integrated in the ergometer
no BP monitor none
(with prompt to take BP
measurement)
SunTech Tango+ SunTech
Bosotron Bosotron
The BP monitor must be ready for operation before the exercise test is
activated. Otherwise the system will not recognize it and prompt the
user to take manual BP measurements.
Note
If the systolic pressure is expected to exceed 200 mmHg (26.7 kPa),
select a higher deflation rate at the BP monitor to expedite the
measurement. The Bosotron 2 will terminate the measurement after
approx. 90 seconds.
SpO2 Monitor
Connect SpO2 monitors to the COM ports.
Model Configuration
Masimo Radical-7 Masimo Radical-7
SunTech Tango+ with SpO2 in BP monitor
option
ECG Recorder
Connect MAC 1200 (V5.21 or later) to a COM port, using a serial cable.
Connect MAC 600, MAC 800, MAC 1600, MAC 3500, MAC 5000, MAC
5500 to LAN, serial port (RS232), modem.
Modem
Connect the modem to a COM port, using a serial cable.
Laser Printer
Warning
Shock Hazard — In the patient vicinity, the printer
installed on the system cart must meet the requirements
of IEC 60601, or it must be a GE Healthcare modified
printer with a connected protective earth conductor
(potential equalization). Observe the installation
instructions supplied with the printer.
Note
For printer selection information, please refer to “System Settings”
on page 12-1.
Warning
Patient Hazard — As stipulated in IEC 60601-1, printers
operated in the patient environment must be connected
via an isolation transformer or to a potential equalization
system (protective earth conductor).
Caution
The ambient conditions for the CASE system are
different to those of most laser printers. Ensure ambient
conditions suitable for the laser printer.
Application Tips
General Application
When operating the system in a network environment, periodically scan
the system for viruses. The anti-virus software should be updated on a
weekly basis. Customers are responsible for installation and
maintenance of anti-virus software.
Note
The Cubic Spline and the FRF algorithm cause a signal delay of
approx. 2 seconds; therefore they should be disabled for
adjustment of the pacemaker parameters.
If several adverse conditions exist at once, the possibility of pacer
pulses being interpreted as QRS complexes should be considered.
At the same time, however, QRS complexes might be suppressed
in certain situations. For safety, always watch pacemaker
patients closely.
The Cubic Spline and the FRF algorithm cause a signal delay of approx.
2 seconds; therefore they should be disabled if the patient has to be
defibrillated while the ECG is being recorded (“Acquisition Tab” on page
12-3).
Note
If excessive DC voltages are present at the electrode, then a message
will appear indicating a Lead Off condition.
Note
Maximum ST deviation refers to the amount of exercise-induced
ST segment deviation, i.e., the largest elevation or depression
after baseline ST levels have been subtracted.
Note
The user must verify and possibly correct the system-calculated
value.
References
Study the following publications for a correct understanding of the Duke
Treadmill Score:
Mark, D.B.; Shaw, L.; Harrell, F. E.; Lee, K. L.; Bengton, J.R.;
McCants, C.B.; Califf, R.M.; Pryor, D.B.: Prognostic value of a
treadmill exercise score in outpatients with suspected coronary
artery disease. The New England Journal of Medicine. 325 (12): 849 -
853 (Sept. 1991)
Mark, D.B.; Hatky, M.A.; Harrell, F.E.; Lee, K.L.; Bengton, J.R.;
Califf, R.M.; Pryor, D.B. Exercise treadmill score for predicting
prognosis in coronary artery disease. Annals of Internal Medicine.
106 (6): 793 - 800 (June 1987)
Risk Factors
According to the "Instructions of Using Coronary Risk and Stroke Risk
Prediction Charts and Worksheet" proposed by the American Heart
Association (AHA) (see also Physician’ s Guide to GE Stress Systems) the
system automatically calculates the risk factors for coronary heart
disease and myocardial infarction. If an input value is unknown or out-
of-range, the system uses the following risk factor values:
Coronary Heart Disease Risk Factor Values for Unknown or Out-of range Parameters
Male Stroke Risk Factor Values for Unknown or Out-of range Parameters
Female Stroke Risk Factor Values for Unknown or Out-of range Parameters
The reference values are height and gender-related and are defined for
patients aged 19 to 70 years.
Note
Patients between 19 and 25 years of age are assumed to be 25.
A range for height is not defined.
The reference values are height and gender-related and are defined for
patients aged 5 to 18 years.
Note
Height range from 107 cm to 182 cm.
For patients exceeding the range limits, the limit values will be
used.
NHANESIII
This reference value equation was published 1999 (AM J RESPIR CRIT
CARE MED 1999; 159:179-187).
The reference values are height and gender-related and are defined for
patients aged 8 to 80 years from the ethnic groups Caucasian (Race
selection: Caucasian), African American (Race selection: Black), Mexican
American (Race selection: Hispanic).
Note
Height range from 110 cm to 200 cm.
Age range from 8 to 80 years.
For patients exceeding the range limits, the limit values will be
used.
The equation defines the following reference values: FEV1, FEV6, FVC,
PEF, MMEF, FEV1/FEV6, FEV1/FVC.
Knudson
This reference value equation was published 1983 (KNUDSON et al. Am
Rev Respir Dis 127 (1983) 725-734).
The reference values are height and gender-related and are defined for
patients aged 5 to 90 years from the ethnic groups Caucasian (Race
selection: Caucasian), African American (Race selection: Black).
Note
Age range from 5 to 90 years.
For patients exceeding the range limits, the limit values will be
used.
A range for height is not defined.
Forche
This reference value equation was published 1994 (Sollwerte für die
Lungenfunktion; Arbeitskreis für die Standardisierung der
Österreichischen Gesellschaft für Lungenerkrankungen und
Tuberkulose 1994).
The reference values are height and gender-related and are defined for
patients aged 5 to 90 years, for adolescents the values are also weight-
related.
Note
Age range from 5 to 90 years.
For patients exceeding the range limits, the limit values will be
used.
A range for height is not defined.
Finland
The equation combines the reference value developed by VILJANEN and
WANNE.
The reference values are height and gender-related and are defined for
patients aged 18 to 65 years.
Note
Age range from 18 to 65 years.
For patients older than 65 years, the age is set to 65.
A range for height is not defined.
The reference values by WANNE are height and gender-related and are
defined for patients aged 6 to 17 years.
Note
Age range from 6 to 17 years.
For patients younger than 6 years, the age is set to 6.
A range for height is not defined.
The equation defines the following reference values: MEF50, PEF, FVC,
FEV1, FEV1/FVC, VC (same as FIVC).
Brazil
The Brazil equation is based on the reference values by PEREIRA and
MALLOZI.
The reference values are height and gender-related and are defined for
men aged 25 to 78 years and women aged 20 to 78 years.
Note
Age range 25 to 78 years for men and 20 to 78 years for women.
For patients older than the upper limit, the upper limit will be
used.
A range for height is not defined.
The equation defines the following reference values: PEF, FVC, FEV1,
FEV1/FVC, MMEF.
The reference values are height and gender-related and are defined for
men aged 6 to 24 years and women aged 6 to 19 years.
Note
Age range 6 to 24 years for men and 6 to 19 years for women.
For patients younger than the lower limit, the lower limit will be
used.
A range for height is not defined.
The equation defines the following reference values: PEF, FVC, FEV1,
FEV1/FVC, MMEF.
Interpretation Modes
Note
The authors of the reference values developed the interpretation
modes for the reference value equations. Other reference value
equations may lead to different interpretations.
Europe
An interpretation mode on the basis of the ECCS reference values. The
flow-volume loop is analyzed according to these criteria:
> 70 % normal*
Criterion FVC Act. < xx % FVC Ref. and (VC MAX/(2 x MEF50) < 0.8)
> 70 % normal*
Expiratory stenoses
211A
protocol TCP/IP
Troubleshooting
Problem Remedy
Send exercise test Check network address of the remote PC on your local PC and
data not possible correct, if necessary.
Check option code and correct, if necessary.
Enable the Remote View function.
Exercise test Not all required electrodes are applied. If you wish to initiate the
analysis does not analysis all the same, click the Relearn medians button.
start
Incorrect heart rate The program may have selected two unsuitable leads. Manually
or arrhythmia select two good leads (see section “Miscellaneous 1 Tab” on
results page 12-24).
Displayed data are The screen diagonal entered in the System Configuration
incomplete. (section “General Tab” on page 12-40) is incorrect.
For screens below 17 inches: the Windows taskbar is always on
top (Windows settings - Taskbar).
Wrong screen resolution in Windows. Please use Small Fonts.
System does not Incorrect ergometer model/port selected (section “Devices Tab”
identify ergometer. on page 12-53).
Check connection cable and connectors.
Turn ergometer off and on again. Restart CASE.
Problem Remedy
USB drive not Restart system.
found.
The ECG grid is not Set up Acrobat Distiller as follows
visible in the PDF Acrobat V6.0:
file. Start -> Settings -> Printer -> Adobe PDF -> menu Printer ->
Document Defaults -> menu Adobe PDF Settings -> disable do
not send fonts to distiller;
menu Paper/Quality -> button Advanced -> Graphic and select
the field Download True Type Font as Softfont.
Acrobat V5.0:
Start -> Settings -> Printer -> Acrobat Distiller -> menu Printer ->
Document Defaults -> menu Adobe PDF Settings -> disable do
not send fonts to distiller;
menu Paper/Quality -> button Advanced -> Graphic and select
the field Download True Type Font as Softfont.
If these steps do not correct the problem, change the grid format:
System Configuration -> General tab -> Printer -> deselect
Dotted line -> select grid format. Tip: If the grid lines appear too
strong, select a shade of gray rather than black.
Message Not Free storage capacity on the hard drive where your database is
enough free stored, e.g., by compressing exercise test data or by deleting
memory on hard examinations. Also ensure that there is enough free storage
drive! capacity on the hard drive where your operating system is
located and in the temporary directories (100 MB minimum).
Patient record In a network environment, the currently open patient record may
cannot be opened be locked following the crash of the client or server (e.g. due to
and is reported to power failure). Please contact Customer Service for advice.
be locked.
Taskbar popup: If you cannot delete a print job using the button Delete print jobs
This document within the CASE application, delete the print job from the print
failed to print. queue (Start - Printers and Faxes - double-click Marquette
Thermal - right-click the print job - left-click Cancel to delete the
print job).
Note
You can send a screenshot to GE Healthcare.
Using the Print key on the keyboard, copy the screen to the
Clipboard.
Open WordPad by selecting Start -> Programs -> Accessories.
Insert the screen from the Clipboard with Edit -> Paste.
Select the page orientation, using File -> Print setup ->
Landscape.
Print the page and fax it to GE Healthcare or save the page as a
WordPad document and send it via e-mail.
Remote Service
In conjunction with the InSiteExC application, CASE is a system with
remote service capability. For InSiteExC installation instructions, please
refer to the CASE Field Service Manual. You must have access to the
internet to be able to use the remote service capability.
Warning
Patient Hazard — Disconnect the patient from CASE
during remote servicing.
Virus Protection
Any anti-virus software installed on the CASE must be compatible with
the CASE operating system.
Warning
Patient Hazard — Disconnect the patient from the
system when the modem is connected.
When installing the fax software from the original media, you can
configure the fax port as the standard printer. With this setting, every
print command is sent to the fax program. Enter frequently used fax
numbers in the Telephone Directory. Select the high printer resolution
and the landscape format. If the same image is transferred more than
twice between two PCs, you must select the portrait orientation after the
second transfer.
Sending a Fax
1. Display the screen to print.
2. Select the fax as the printer (e.g. fax printer CAPTURE FAX BVZP).
4. Select Send. The status window and the audible signal from the
modem indicate the transfer.
Note
All modems connected to the system must meet the requirements of
IEC 60950 or UL1950. Furthermore, the regulations valid in the
respective countries must be observed. The modem must be set up
within the medical location, but not in the patient vicinity.
Warning
Patient Hazard — Disconnect the patient from the
system when the modem is connected.
C Order Information
Order Information
Subject to change. Always refer to latest list of accessories.
Accessories
Peripheral Devices
2079551-001 22-inch EIZO wide-screen LCD monitor
Cables
405535-001 Power cord, USA
D Technical Description
Item Description
Instrument type Unity-enhanced cardiac stress testing system with 14 channel acquisition and
programmable lead configurations. Celeron-based and hard drive storage
delivers local and MUSE database access.
ST measurements ST amplitudes, slope, integral, index, ST/HR slope, ST/HR loops, ST/HR
index up to 15 leads
ECG interpretation (optional) 12SL Adult and Pediatric ECG Analysis Program
Item Description
Technology Active, “Type BF” floating isolated powered 14 channel acquisition module
with built-in lead-fail detection and lead prep impedance measurement
Skew 62.5 µs
Item Description
Noise < 15 µV peak-to-peak noise over 0.01 to 150 Hz (-3 dB) bandwidth
High pass filter 0.01 Hz (or 0.05 Hz, special use) with DC offset control
Table 3. Display
Item Description
Type LCD
Item Description
Monitored leads 3, 6, or 15
Displayed vital signs data Heart rate, target heart rate, blood pressure, exercise clock, stage clock,
(configurable) phase clock, protocol, speed, grade, Watts, METS, RPP, and SpO2
Table 4. Writer
Item Description
Resolution Horizontal 1000 lines/s x 200 dpi dedicated local printing, 200 x 200 dpi
generic printing
Table 6. Physical
Item Description
Height 169 cm (66 ½ inches) with monitor attached (height adjustable, extend and retract,
rotates vertical and horizontal for optimum viewing)
Width 62 cm (24 in.) with paper tray removed
Depth 75 cm (29 in.)
Weight 66.6 kg (147.5 lbs.) without monitor and KISS pump
Interfaces included Acquisition module
Keyboard (PS/2) and dedicated stress keypad (USB)
Mouse (PS/2)
Built-in thermal printer (USB)
6 USB ports
Full Duplex IEEE 802.3 10 Base-T, 100 BaseTX and 1000 BaseT* compatible
through RJ45, MUSE compatible
6 Serial ports (COM 1-6); treadmill, BP, ergometer, SpO2
4 Analog and 1 TTL (trigger) output; analog ergometer, camera synch., etc.
Display Port for monitor
External DVD/CD-RW Drive (DVD to USB) (ordered separately)
External CD-RW Drive (DVD to USB) (ordered separately)
External SD Card Reader (SD to USB)
Table 7. Safety
Item Description
Certification UL 60601-1 classified
UL classified for CAN/CSA C22.2 No. 601.1
CB Certified for IEC 60601-1 / UL Certification
CE marking for Council Directive 93/42/EEC concerning Medical Devices
Meets applicable IEC 60601-2-25 & IEC 60601-2-51 requirements for
ECG safety & performance
Meets IEC 60601-1-2 requirements for Electromagnetic Compatibility/
TUV Rheinland
Type of protection against Class 1
electrical shock
Degree of protection against Ordinary
ingress of liquids
Index
D
A
Danger (Definition) 6
Acquisition Module Buttons 13 Default Printer, Set 6
Acronyms 27, 28 Device
Administrative tools, using the program in con- characteristics 15
junction with 1 DICOM interface 1
AHA CHD Risk 8 Disposal 4
AHA Stroke Risk 8 Drugs, Assign to Medication Groups 6
Ambulatory Blood Pressure Measurement Setup Resting ECG 5
Acquisition Tab 31
Miscellaneous Tab 33
Ambulatory Blood Pressure Monitors 3 E
Anti-virus software 22
Arrhythmia Review 45 ECG Data Windows
Resting ECG 20 12SL analysis 25
ATS, Measurement Mode 7, 10 ECG Strips 24
Audio ECG 51 Interpretation 26
ST/Arrhy 18
Tabular Summary 21
B Trends 23
ECG Measurement 15
Bicycle Ergometers, Connect 3 ECG Recorder, Connect 5
Biocompatibility 13 ECG Recording (Exercise Test) 17
Blood Pressure Monitor, Connect 4 ECG Recording During Defibrillation 7
BP Protocol Configuration 32 ECG Signal Quality Check
Bronchodilator Test 12 Exercise Tests 9
Resting ECG 6
ECG Strips 44
C Edit Patient Records 6
Electrode Application 3
Calculating Patient Risk 7 CM5, CC5, CH Auxiliary Lead Electrode
CASE system, technical specifications 3 Placement 12
Caution (Definition) 6 CM5, CC5, ML Auxiliary Lead Electrode
Classification 11 Placement 11
Cleaning, Equipment Surface, Cables 3 FRANK X, Y, Z Lead Electrode Placement 9
N Race Information 13
Real-time Report Printout 30
New Patient, Enter 7 Reanalysis
Exercise Tests 41
Resting ECG 17
Reason for Test, Edit Library