Experiment 08 Finalx
Experiment 08 Finalx
Experiment 08 Finalx
Reagents:
1. Potassium chromate indicator
2. Silver nitrate solution (0.0141 N)
Apparatus:
1. Burette
2. Measuring cylinder
3. Beaker
4. Dropper
5. Stirrer
\
Procedure:
1. Take 50 mL of the sample in a beaker and add 5 drops (about 1 mL) of
potassium chromate indicator to it.
2. Add standard (0.0141 N) silver nitrate solution to the sample from a
burette, a few drops at a time, with constant stirring until the first
permanent reddish color appears. This can be determined by comparison
with distilled water blank. Record the mL of silver nitrate used.
3. If more than 7 or 8 mL of silver nitrate solution are required, the entire
procedure should be repeated using a smaller sample diluted to 50 ml
with distilled water.
Assignment
2. Why it is necessary to dilute sample if end point does not obtained even
after adding more than 7 or 8 mL silver nitrate to the original sample?
Ans: If the sample has a high concentration of chloride ion, then it is not
possible to detect the end point after adding more then 7/8 ml silver nitrate to
the original sample. So then, it is necessary to dilute the sample and use a
portion of the sample to obtain the end point.
Ans: The end point of titration cannot be detected visually unless an indicator
capable of demonstrating the presence of excess Ag+ is present. The indicator
normally used is potassium chromate, which supplies chromate ions. As the
concentration of CI- ions becomes exhausted, the silver ion concentration
increases and a reddish brown precipitate of silver chromate is formed.
This is taken as evidence that all chloride has been precipitated. Since an excess
Ag+ is needed to produce a visible amount of Ag2CrO4 , the indicator error is
subtracted from all titrations.
The indicator error or blank varies somewhat with the ability of individuals to
detect a noticeable color change. The usual range is 0.2 to 0.4 mL of titrant. An
error of 0.2 mL will be used in the class.