Shielding Calculations For Radiotherapy Calculation Examples
Shielding Calculations For Radiotherapy Calculation Examples
Shielding Calculations For Radiotherapy Calculation Examples
General
When planning new premises, keep in mind that, often, the premises are in
use for several decades and the equipment is changed approximately every
ten years. It is recommended to use initial shielding that is slightly thicker than
the required level or at least to reserve space for increasing shielding at a
later time. Part of the shielding may be easily removable for the purposes of
equipment replacement. It is also recommended to reserve the rooms
adjacent to the radiotherapy rooms for maintenance and storage purposes.
STUK can provide more information, as necessary.
The dose that results from radiation practices must not exceed 120 µSv per
week in the supervised area or 6 µSv per week in other areas. Furthermore,
the instantaneous dose rate must not exceed 20 µSv/h in areas where
persons stay or work at all times. The shielding calculations are performed at
the maximum excitation energy and field size of the radiotherapy equipment
unless deviating from the maximum values is justified. The dose rate is
primarily selected based on the highest value to be used. If the workload of
the equipment is not known in advance, the calculations must use a value of
800 Gy/week, at the minimum.
The occupancy factor (T) refers to the use rate of a room. If a room is
occupied during the use of radiotherapy equipment, the occupancy factor is T
= 1, but T may also be lower. In work areas (such as control rooms, offices or
laboratories), waiting rooms and patient rooms, the occupancy factor is T = 1.
In indoor and outdoor premises where no persons remain on a continuous
basis (such as toilets, corridors, streets and storage rooms), the occupancy
factor is at least 0.1. If there are persons at all times in a corridor, the
occupancy value should be
T = 1.
The orientation factor (U) takes into account that the direction of the radiation
beam of the equipment may vary. For isocentric equipment, the factor should
be set to U = 1 down and at least to U = 0.25 in the other possible directions.
If the equipment is intended to be used in a specific direction to a
considerable degree, for example in whole body radiotherapy, the orientation
factor for that direction should be U = 1.
∙
, (1)
∙ ∙ ∙
where Dmax is the planning value (120 μSv or 6 μSv per week). The distance
between the viewing point and radiation source (accelerator’s target) is r. The
distance between the target and isocenter is ro. DI is the planned weekly
Radiation and Nuclear Safety Authority 2 (6)
STO
22.3.2015
If there are persons staying or working at the viewing point at all times, it must
be checked whether the calculated attenuation is also sufficient in terms of the
instantaneous dose rate. The instantaneous dose rate can be calculated
with the formula
∙ ∙
, (2)
where is the dose rate of the radiotherapy equipment in the isocenter and
the other annotations are as above. The orientation and occupancy factors
are not used here.
If, based on the weekly dose, the assessed shielding is not sufficient for
limiting the instantaneous dose rate, the necessary transmission factor shall
be calculated according to the instantaneous dose rate (Formula 2).
The number of the tenth value layers necessary for the shielding (TVL, see
Table 1) is n
1/ . (3)
1 ∙ . (4)
TVL1 refers to the first tenth value layer and TVLe refers to the subsequent
tenth value layers. The TVL values are selected from Table 1 according to the
materials and radiation energy used.
Figure 1. Typical radiotherapy room.
6 ∙ 10 ∙ 5,5
8,3 ∙ 10
875 ∙ 0,25 ∙ 0,1 ∙ 1
Since steel is the first material in the direction of radiation and its TVL1 and
TVLe values are the same, the number of tenth value steel layers ns can be
calculated by dividing the thickness of steel with its TVL value: ns = 13 cm / 11
cm = 1.182. Since concrete is the next material after steel, the number of
tenth value concrete layers nc can be calculated by dividing the thickness of
concrete with its TVLe value: nc = 100 cm / 44 cm = 2.273. The total number of
tenth value layers is ns + nc 3.45. Therefore, approximately 1.6 TVL are still
needed for the target value of 5.08 TVL, which can be achieved with 18 cm of
additional steel.
If there are persons staying in the area at all times, the additional shielding
according to the above calculation is not sufficient because the instantaneous
dose rate in point A is
Radiation and Nuclear Safety Authority 4 (6)
STO
22.3.2015
4 / ∙ ∙ 1 1,1 66
,
5,5
which exceeds the limit value of 20 μSv/h. In order to lower the dose rate, the
steel thickness must be increased by approximately a half tenth value layer,
which results in 24 cm of additional steel instead of the previously calculated
18 cm.
Secondary radiation is generated when the primary beam touches the patient
or the walls of the radiotherapy room. Furthermore, the equipment’s leakage
radiation in all directions must be taken into account. When dimensioning the
shields, the proportion of scattered radiation at an excitation energy of more
than 10 MeV is usually insignificant compared to leakage radiation. However,
the effects of scattered radiation must be taken into account when planning
the access corridor. Table 2 presents scatter factors, and Table 3 presents
tenth value layers for secondary radiation.
Table 2. Scatter factors. The scatter factors presented in the table indicate the relationship
between the scattered radiation dose and the in‐field dose when the field size on the surface of
the scattering object is 400 cm2 and the distance between the examined point and the target is
one metre.
∙
. (5)
∙ , ∙ ∙
The distance between the radiation source and the scattering object is r1, and
the distance between the scattering object and the examined point is r2. The
scatter factor a (from Table 2) is determined based on the scattering angle
and energy. A is the area on the scattering object’s surface that is subjected
to direct radiation. The formula uses a factor of 0.04 because the scatter
factor is determined for a field size of 0.04 m2. Additionally, it must be
examined whether the calculated attenuation is also sufficient in terms of the
instantaneous dose rate.