Operating Instructions
Cleanfit CPA450
Retractable assembly for 12mm sensors for pH/ORP and
oxygen measurement
Cleanfit CPA450 Table of contents
Table of contents
4 Installation . . . . . . . . . . . . . . . . . . . . . . . 10
4.1 Installation conditions . . . . . . . . . . . . . . . . . . 10
4.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.3 Post-installation check . . . . . . . . . . . . . . . . . . 18
5 Operation options . . . . . . . . . . . . . . . . . 19
5.1 Initial commissioning . . . . . . . . . . . . . . . . . . . 19
5.2 Operating elements . . . . . . . . . . . . . . . . . . . . 19
5.3 Operating the assembly . . . . . . . . . . . . . . . . . 20
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . 22
6.1 Cleaning the assembly . . . . . . . . . . . . . . . . . . 22
6.2 Cleaning agent . . . . . . . . . . . . . . . . . . . . . . . 23
6.3 Replacing seals . . . . . . . . . . . . . . . . . . . . . . . 24
7 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7.1 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 28
7.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8 Accessories . . . . . . . . . . . . . . . . . . . . . . . 32
8.1 Accessory kits . . . . . . . . . . . . . . . . . . . . . . . . 32
8.2 Welding socket . . . . . . . . . . . . . . . . . . . . . . . 32
8.3 Safety kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.4 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.5 Connection accessories . . . . . . . . . . . . . . . . . . 34
9 Technical data . . . . . . . . . . . . . . . . . . . . 36
9.1 Environment . . . . . . . . . . . . . . . . . . . . . . . . . 36
9.2 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Endress+Hauser 3
About this document Cleanfit CPA450
1.1 Warnings
Structure of information Meaning
Permitted or recommended
Reference to page
Reference to graphic
Result of a step
4 Endress+Hauser
Cleanfit CPA450 Basic safety instructions
Endress+Hauser 5
Basic safety instructions Cleanfit CPA450
During operation:
‣ If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
6 Endress+Hauser
Cleanfit CPA450 Incoming acceptance and product identification
Endress+Hauser 7
Incoming acceptance and product identification Cleanfit CPA450
3.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Order code
• Serial number
• Permitted pressure
• Permitted temperature
Product page
www.endress.com/cpa450
8 Endress+Hauser
Cleanfit CPA450 Incoming acceptance and product identification
3.4.1 CE/PED
The assembly has been manufactured according to good engineering practice as per Article
4, Paragraph 3 of the Pressure Equipment Directive 2014/68/EU and is therefore not
required to bear the CE label.
Endress+Hauser 9
Installation Cleanfit CPA450
4 Installation
A B
X
X
Y
Ø 25 (0.98)
Y
135 (5.31)
Z
Ø 21.2 (0.83)
Z
M23 + 1.5
Ø 27 (1.06)
(1.81)
46
C D E
89
89 (3.5)
(3.5)
X
Y
Y
Y
F, G
X
X
Z
A0037726
10 Endress+Hauser
Cleanfit CPA450 Installation
B CPA450-*C*** 100 (3.94) NPT 1¼" external 558 (21.97) 220 (9.06)
250 (9.84) 708 (27.87) 370 (14.9)
700 (27.5) 1158 (45.59) 820 (32.6)
C CPA450-*D*** 100 (3.94) Flange DN32(as per DIN EN 558 (21.97) 225 (8.86)
250 (9.84) 1092-1) 708 (27.87) 375 (14.76)
700 (27.5) 1158 (45.59) 825 (32.48)
C CPA450-*E*** 100 (3.94) Flange ANSI 1¼" (as per ASME 558 (21.97) 225 (8.86)
250 (9.84) B16.5) 708 (27.87) 375 (14.76)
700 (27.5) 1158 (45.59) 825 (32.48)
D CPA450-*H*** 100 (3.94) NPT 1¼" external 558 (21.97) 130 (5.12)
250 (9.84) 708 (27.87) 280 (11.2)
700 (27.5) 1158 (45.59) 730 (28.7)
E CPA450-*I*** 100 (3.94) Flange DN32 (as per DIN EN 558 (21.97) 92 (3.62)
250 (9.84) 1092-1) 708 (27.87) 242 (9.53)
700 (27.5) 1158 (45.59) 792 (31.18)
E CPA450-*K*** 100 (3.94) Flange ANSI 1¼" (as per ASME 558 (21.97) 82 (3.23)
250 (9.84) B16.5) 708 (27.87) 232 (9.13)
700 (27.5) 1158 (45.59) 782 (30.79)
C CPA450-*N*** 100 (3.94) Flange ANSI 2" (as per ASME B16.5) 558 (21.97) 225 (8.86)
and 250 (9.84) 708 (27.87) 375 (14.76)
CPA450-*R*** 700 (27.5) 1158 (45.59) 825 (32.48)
Suitable sensors
The following sensors are suitable for installation in the assembly:
• Digital sensors with Memosens technology, length 120 mm (4.72")
• pH/ORP glass electrodes, length 120 mm (4.72")
• ISFET sensors: Only the ISFET sensors specified in the "Accessories" section can be
installed.
• Oxygen sensors, length 120 mm (4.72")
Orientation
The permitted orientation of the assembly depends on the sensor used:
• Digital sensors with Memosens technology, pH/ORP glass electrodes:
Install the assembly at an angle of at least 15° to the horizontal → 2, 12.
• ISFET sensors:
For ISFET sensors, there are basically no restrictions regarding orientation. The
installation angle should be from 0 to 180°.
• All other sensors:
Pay attention to the information in the relevant TI.
Endress+Hauser 11
Installation Cleanfit CPA450
A B
15° 15°
A0011679
2 Orientations
A Glass sensors: 15 ° to the horizontal
B ISFET sensors: 0 to 180° recommended
Insert the immersion assembly into the vessel or pipe to a depth that will ensure that
medium continuously washes around the electrode, even at the minimum level.
LCAUTION
There is a risk of injury if used without a ball valve. due to the risk of medium
escaping.
‣ If used without a ball valve, the process must be stopped prior to dismantling the
immersion tube or replacing the sensor.
A0010209 A0010210
3 Measuring mode (ball valve is open): 4 Service position (ball valve is closed): assembly is
assembly is retracted extended for electrode replacement, calibration,
rinsing
A Top edge of adapter
12 Endress+Hauser
Cleanfit CPA450 Installation
4.2 Installation
LWARNING
If process medium and cleaning medium escape, there is a risk of injury due to high
pressure, high temperatures or chemicals.
‣ Wear protective gloves, protective goggles and protective clothing.
‣ Mount the assembly only if vessels or pipes are empty and unpressurized.
‣ Before exposing the assembly to the process pressure, verify that all connections are
sealed.
3 2
A0038005
Endress+Hauser 13
Installation Cleanfit CPA450
LDANGER
Non-compliance with the safety instructions can lead to injury or death.
‣ The safety instructions must be read and followed.
‣ Install the safety equipment only if the assembly has been removed.
The safety equipment is not a substitute for the standard retainer of the assembly.
A0014680
Preparation
10
11
8 12
6
5
4
3
2
1
13
A0014681
14 Endress+Hauser
Cleanfit CPA450 Installation
2 Safety kit (lower part) 9 Safety kit (upper part with handle)
3 Sensor holder with protection guard 10 Cable protector
4 Locking ring (metal) 11 Handle on immersion tube
5 Coupling nut (black) 12 Safety kit (upper part with handle and locking pin)
6 Safety kit (compression spring) 13 Ball valve with handle
7 Immersion tube
1. Move the assembly to the measuring position, and mark the immersion tube approx.
20 mm (0.8") above the coupling nut (black) (item 5).
This is where the upper part of the safety kit (item 9) will be installed.
2. Use the hook wrench to open the coupling nut (black) by ¼ to ½ turn.
3. Pull the immersion tube out as far as the stop (service position).
4. Screw on the locking ring (item 4), and disconnect the assembly from the service
chamber (item 1).
The side with the chamfered edge points towards the ball valve. Secure the lower part
with a torque of 3.5 Nm (2.6 lbf ft).
Endress+Hauser 15
Installation Cleanfit CPA450
A0010207
1. Connect the rinse water line to the rinsing nozzle provided. The three rinsing nozzles
at the assembly are identical (G¼ for 316L material - NPT ¼" for Alloy C22 and
titanium).
2. Operate the rinse water connection of the assembly with a water pressure from 2 bar
to max. 6 bar (29 to 87 psi).
16 Endress+Hauser
Cleanfit CPA450 Installation
15
14
13 3
12 4
5
6
11
10 9
A0010212
Endress+Hauser 17
Installation Cleanfit CPA450
LCAUTION
The immersion tube may move suddenly and cause injuries.
‣ Hold the immersion tube with one hand.
Preparing the assembly
1. Use the hook wrench to carefully open the coupling nut (black) (item 14) by ¼ to ½
turn.
2. Release the cable protector (item 1).
3. Pull the immersion tube (item 2) out as far as it will go.
4. Close the ball valve (item. 9).
18 Endress+Hauser
Cleanfit CPA450 Operation options
5 Operation options
LWARNING
Risk of injury from high pressure, high temperature or chemical hazards if process
medium escapes.
‣ Wear protective gloves, protective goggles and protective clothing.
‣ Mount the assembly only if vessels or pipes are empty and unpressurized.
‣ Before subjecting the assembly to the process pressure, verify that all connections are
sealed.
A0010207
10 Operating elements
Endress+Hauser 19
Operation options Cleanfit CPA450
• Lever (item 8)
For opening or closing the ball valve.
• Immersion tube (item 7)
The desired orientation of the sensor is configured by turning the immersion tube.
• Stop/sensor holder (item 6)
When moving the assembly to the service position, pull the immersion tube out as far as
this stop.
When using the safety kit, screw the coupling nut (black) onto the immersion tube
until it is hand-tight.
LCAUTION
The immersion tube may move suddenly and cause injuries.
‣ Hold the immersion tube with one hand.
1. Use the hook wrench to carefully open the coupling nut (black) by ¼ to ½ turn.
2. Pull the immersion tube out as far as the stop (service position).
3. Close the ball valve.
The lever is horizontal.
20 Endress+Hauser
Cleanfit CPA450 Operation options
Unlocking operation
1. Release the coupling nut (black) using the hook wrench.
2. Push the handles towards the ball valve, turning them clockwise.
The locking pin is unlocked.
3. Move the assembly to the service position.
4. Close the ball valve.
Endress+Hauser 21
Maintenance Cleanfit CPA450
6 Maintenance
LWARNING
Risk of injury due to escaping medium, high pressure and high temperature.
‣ Before each maintenance task, ensure that the process pipe or vessel is unpressurized,
empty and rinsed.
‣ Move the assembly to the service position, and close the ball valve.
All parts in contact with the medium, such as the sensor and the sensor guide, must be
cleaned regularly.
1. Remove the sensor in the reverse order to installation.
2. Remove light dirt and fouling with suitable cleaning solutions. (→ 23
3. Remove heavy soiling using a soft brush and a suitable cleaning agent.
4. For very persistent dirt, soak the parts in a cleaning solution. Then clean the parts
with a brush.
22 Endress+Hauser
Cleanfit CPA450 Maintenance
Endress+Hauser 23
Maintenance Cleanfit CPA450
24 Endress+Hauser
Cleanfit CPA450 Maintenance
3
4
1
1
A0010204
2
3
A0010205
Endress+Hauser 25
Maintenance Cleanfit CPA450
Replacing the flat seal of the adapter for flange connections (optional)
A0010206
26 Endress+Hauser
Cleanfit CPA450 Repair
7 Repair
LWARNING
Danger resulting from improper repair!
‣ Any damage to the assembly that compromises pressure safety must be repaired only
by authorized and qualified personnel.
‣ Following each repair and maintenance task, check the assembly for leaks using
appropriate procedures. Following this, the assembly must again comply with the
specifications in the technical data.
‣ Replace all other damaged components immediately.
Endress+Hauser 27
Repair Cleanfit CPA450
7 2
15
14 9
11 3
11 4
12
10
13
A0037952
11 Spare parts
28 Endress+Hauser
Cleanfit CPA450 Repair
Immersion tube 100 mm, 316L, cable gland; old version up to 71069820
01/11
Immersion tube, 250 mm, 316L, cable gland; old version up to 51501521
01/11
Immersion tube, 700 mm, 316L, cable gland; old version up to 51501522
01/11
Immersion tube, 100 mm, 316L, with stop, without handle; 71128830
version from 02/11
Immersion tube, 250 mm, 316L, with stop, without handle; 71128831
version from 02/11
Immersion tube, 700 mm, 316L, with stop, without handle; 71128832
version from 02/11
2 Immersion tube, 100 mm, Alloy C22, with stop, without handle; 71128833
version from 02/11
Immersion tube, 250 mm, Alloy C22, with stop, without handle; 71128834
version from 02/11
Immersion tube, 700 mm, Alloy C22, with stop, without handle; 71128836
version from 02/11
Immersion tube, 100 mm, titanium, with stop, without handle; 71128837
version from 02/11
Immersion tube, 250 mm, titanium, with stop, without handle; 71128838
version from 02/11
Immersion tube, 700 mm, titanium, with stop, without handle; 71128839
version from 02/11
4 Cable connector
Flanged sleeve, locking ring, stainless steel 316L; EPDM O-rings 51501536
Flanged sleeve, locking ring, stainless steel 316L; Viton O-rings 51501537
Flanged sleeve, locking ring, stainless steel 316L; Kalrez O-rings 71028947
8
Flanged sleeve, locking ring, Alloy C22; Kalrez O-rings; version 71128841
from 02/2011
Endress+Hauser 29
Repair Cleanfit CPA450
30 Endress+Hauser
Cleanfit CPA450 Repair
7.2 Return
The product must be returned if repairs or a factory calibration are required, or if the
wrong product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
7.3 Disposal
The device contains electronic components. The product must be disposed of as electronic
waste.
Endress+Hauser 31
Accessories Cleanfit CPA450
8 Accessories
The following are the most important accessories available at the time this documentation
was issued.
‣ For accessories not listed here, please contact your Service or Sales Center.
32 Endress+Hauser
Cleanfit CPA450 Accessories
8.4 Sensors
Memosens CPS97D
• ISFET sensor for pH measurement with long-term stability in media with high dirt loads
• Product Configurator on the product page: www.endress.com/cps97d
Technical Information TI01405C
Endress+Hauser 33
Accessories Cleanfit CPA450
Memosens COS81D
• Sterilizable, optical sensor for dissolved oxygen
• With Memosens technology
• Product Configurator on the product page: www.endress.com/cos81d
Technical Information TI01201C
34 Endress+Hauser
Cleanfit CPA450 Accessories
Endress+Hauser 35
Technical data Cleanfit CPA450
9 Technical data
9.1 Environment
9.2 Process
Pressure/temperature
p [psi] p [bar]
ratings
A
174 12
145 10
116 8
87 6 B
58 4
29 2
130 T [°C]
-15 0 20 40 60 80 100 120 140
266 T [°F]
5 32 68 104 140 176 212 248 284
A0038118-EN
12 Pressure/temperature ratings
A Maximum process pressure (static), only for fully installed assembly
B Maximum pressure for moving the assembly (functional)
Observe the maximum permitted process temperature and process pressure of the
sensor.
36 Endress+Hauser
Cleanfit CPA450 Technical data
Endress+Hauser 37
Index Cleanfit CPA450
Index
A Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Rinse water connection . . . . . . . . . . . . . . . . . . . . . . . 16
Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Welding socket . . . . . . . . . . . . . . . . . . . . . . . . . . 32 S
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
C Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cleaning agent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Cleaning interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Cleaning the assembly . . . . . . . . . . . . . . . . . . . . . . . . 22 Suitable sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Connection Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Rinse water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
T
D Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
E
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 W
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
I
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 13
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . 10
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 37
Mounting instructions . . . . . . . . . . . . . . . . . . . . . . . . 11
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
O
O-rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
P
Post-installation check . . . . . . . . . . . . . . . . . . . . . . . 18
Pressure surges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Product identification . . . . . . . . . . . . . . . . . . . . . . . . . 7
R
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Replacement
Damaged parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
38 Endress+Hauser