Operating Instructions Smartec CLD18: Conductivity Measuring System
Operating Instructions Smartec CLD18: Conductivity Measuring System
Operating Instructions Smartec CLD18: Conductivity Measuring System
17
71354508
Products Solutions Services
Operating Instructions
Smartec CLD18
Conductivity measuring system
Smartec CLD18 Table of contents
Table of contents
1 Document information . . . . . . . . . 4 9 Diagnostics and
1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 troubleshooting . . . . . . . . . . . . . . . 34
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . 4 9.1 Trouble shooting instructions . . . . . . . . . 34
1.3 Symbols on device . . . . . . . . . . . . . . . . . . . . 5 9.2 Trouble shooting instructions . . . . . . . . . 34
9.3 Diagnostic messages . . . . . . . . . . . . . . . . 35
2 Basic safety instructions . . . . . . . 5
2.1 Requirements for the personnel . . . . . . . . 5 10 Maintenance . . . . . . . . . . . . . . . . . . 38
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . 5 10.1 Cleaning the housing . . . . . . . . . . . . . . . . 38
2.3 Occupational safety . . . . . . . . . . . . . . . . . . 5
2.4 Operational safety . . . . . . . . . . . . . . . . . . . 7
2.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . 7
11 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . 39
11.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
11.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3 Incoming acceptance and
product identification . . . . . . . . . . 8 12 Accessories . . . . . . . . . . . . . . . . . . . . 40
3.1 Incoming acceptance . . . . . . . . . . . . . . . . . 8 12.1 Calibration solutions . . . . . . . . . . . . . . . . 40
3.2 Product identification . . . . . . . . . . . . . . . . . 8
3.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . 9
3.4 Certificates and approvals . . . . . . . . . . . . 10
13 Technical data . . . . . . . . . . . . . . . . . 40
5 Installation . . . . . . . . . . . . . . . . . . . . 12
5.1 Installation conditions . . . . . . . . . . . . . . . 12
5.2 Mounting the compact device . . . . . . . . . 17
5.3 Post-installation check . . . . . . . . . . . . . . . 17
6 Electrical connection . . . . . . . . . . 18
6.1 Electrical connection of the
transmitter . . . . . . . . . . . . . . . . . . . . . . . . 18
6.2 Ensuring the degree of protection . . . . . 21
6.3 Post-connection check . . . . . . . . . . . . . . . 21
7 Operation options . . . . . . . . . . . . . 22
7.1 Operating keys . . . . . . . . . . . . . . . . . . . . . 23
7.2 Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8 Commissioning . . . . . . . . . . . . . . . . 25
8.1 Switching on the device . . . . . . . . . . . . . . 25
8.2 Display settings (Display menu) . . . . . . . 25
8.3 Device configuration (Setup menu) . . . . 26
8.4 Extended setup (Extended Setup
menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8.5 Calibration (Calibration menu) . . . . . . . . 31
Endress+Hauser 3
Document information Smartec CLD18
1 Document information
1.1 Warnings
Symbol Meaning
Permitted or recommended
Reference to page
Reference to graphic
Result of a step
4 Endress+Hauser
Smartec CLD18 Basic safety instructions
Symbol Meaning
NOTICE
Non-designated use
Incorrect measurements, malfunctions and even measuring point failure could result
‣ Only use the product in accordance with the product specifications.
‣ Pay particular attention to the technical data on the nameplate.
Endress+Hauser 5
Basic safety instructions Smartec CLD18
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable European standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
6 Endress+Hauser
Smartec CLD18 Basic safety instructions
Endress+Hauser 7
Incoming acceptance and product identification Smartec CLD18
8 Endress+Hauser
Smartec CLD18 Incoming acceptance and product identification
Product page
www.endress.com/CLD18
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
Endress+Hauser 9
Incoming acceptance and product identification Smartec CLD18
3.4.2 Hygiene
FDA
All materials in contact with the product are FDA-listed materials (apart from the PVC process
connections).
EHEDG
Certified cleanability according to EHEDG Type EL Class I.
When using the sensor in hygienic applications, please note that the cleanability of the
sensor also depends on the way the sensor is installed. To install the sensor in a pipe, use
the appropriate and EHEDG-certified flow vessels for the particular process connection.
3-A
Certified according to 3-A Standard 74- ("3-A Sanitary Standards for Sensor and Sensor
Fittings and Connections Used on Milk and Milk Products Equipment").
EC Regulation No. 1935/2004
The sensor meets the requirements of EC Regulation No. 1935/2004 on materials and articles
intended to come into contact with food.
10 Endress+Hauser
Smartec CLD18 Product description
4 Product description
5
4
3
2
A0019184
1 Elements
1 Temperature sensor
2 Process connection
3 Leakage bore (offset by 90° in relation to the flow direction)
4 Removable housing cover
5 Window for display
6 Cable glands (M12)
7 Flow opening of sensor
Endress+Hauser 11
Installation Smartec CLD18
5 Installation
5.1 Installation conditions
5.1.1 Installation instructions
For a 3-A compliant installation, the following must be noted:
After the instrument is installed its hygienic integrity shall be maintained. The
instrument shall be installed with the leakage detection at the lowest point of the
assembly. Furthermore all process connections must be 3-A compliant.
Orientations
The sensor must be completely immersed in the medium. Avoid air bubbles in the area of the
sensor.
> 1 m (3.3 ft)
A0017691
If the flow direction changes (after pipe bends), turbulence in the medium can result.
Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend.
12 Endress+Hauser
Smartec CLD18 Installation
The product should flow along the hole of the sensor (see the arrows on the housing). The
symmetrical measuring channel allows flow in both directions.
a
If the distance to the wall is smaller, the installation factor
increases for electrically insulating pipes (f > 1) and decreases A0018962
1.05
2
1.00
1
0.95
0 10 20 30 40 a [mm]
Endress+Hauser 13
Installation Smartec CLD18
28 (1.10) 85 (3.35)
78 (3.07)
127 (5.00)
A 23 A
(0.91)
69 (2.72)
(2.10)
(0.24)
53.3
6
112 (4.41)
28 (1.10) 85 (3.35)
(2.89)
123 (4.84)
73.5
5.2
(0.20)
B B
(2.10)
53.3
(2.72)
69
20.5
(0.81)
114.5 (4.51)
12.3 (0.49)
136.5 (5.37)
123 (4.84)
32.8 (1.29)
17.1 (0.67)
C D
101.7 (4.00)
86 (3.39)
A0018942
14 Endress+Hauser
Smartec CLD18 Installation
51 (2.01)
Ø 48.3
(1.90)
Ø 44.3
(1.74)
A0019302
Endress+Hauser 15
Installation Smartec CLD18
(1.22)
31
(1.50)
(1.61)
Ø 41
Ø 38
A0022166
16 Endress+Hauser
Smartec CLD18 Installation
(2.26)
57.5
Ø 40 (1.58)
Ø 50 (1.97)
A0024073
8 Installation in DN 40 pipe with coupling nut 2¼" PVC process connection, dimensions in mm
(inch)
1. When installing, align the compact device in such a way that the medium flows through
the flow opening of the sensor in the direction of medium flow. Use the arrow on the
nameplate to help you align the device.
2. Tighten the flange.
Endress+Hauser 17
Electrical connection Smartec CLD18
6 Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
mA
IOut1 MRS
Lf
mA
IOut2 L
X
Alarm Out
A0033106
9 Electrical connection
18 Endress+Hauser
Smartec CLD18 Electrical connection
L+
Out+
MRS+
IOut2+
+
IOut1+
L-
L-
Out-
MRS-
IOut2-
IOut1-
L
2
A0029684
10 Terminal assignment
NOTICE
Removing the electronics box will destroy the sensor connection!
‣ The electronics box must not be removed under any circumstances.
‣ Do not open the cover on the electronics box.
The recommended cable cross-section for the connecting cables is 0.5 mm2. The
maximum cable cross-section is 1.0 mm2.
Endress+Hauser 19
Electrical connection Smartec CLD18
4 3
1 2
A0033108
5
4
6
7 3
8
1 2
A0033109
20 Endress+Hauser
Smartec CLD18 Electrical connection
Are the transmitter and cables free from damage on the outside? Visual inspection
Are the signal cables correctly connected as per the wiring diagram?
Are the PE distributor blocks grounded (if present)? Grounding is carried out at the point
of installation.
Endress+Hauser 21
Operation options Smartec CLD18
7 Operation options
1
3
Main µS/cm
- ESC -
_
+ E
4
A0018963
22 Endress+Hauser
Smartec CLD18 Operation options
You can always quit menu items / submenus at the end of the menu via "x Back".
Symbols in the editing mode:
Accept entry.
A0020597 If this symbol is selected, the entry is applied at the position specified by the user, and you quit editing mode.
Reject entry.
If this symbol is selected, the entry is rejected and you quit editing mode. The previously set text remains.
A0020598
Delete backwards.
A0020600
If this symbol is selected, the character to the left of the cursor position is deleted.
Delete all.
if this symbol is selected, the entire entry is deleted.
A0020601
Endress+Hauser 23
Operation options Smartec CLD18
7.2 Menus
The operating functions of the compact measuring device are divided into the following
menus:
Display Settings for the device display: contrast, brightness, time for alternating measured values on the display
* The air set and the correct cell constant have already been configured at the factory for
Smartec CLD18 devices. A sensor calibration is not necessary during commissioning.
24 Endress+Hauser
Smartec CLD18 Commissioning
8 Commissioning
8.1 Switching on the device
Familiarize yourself with the operation of the transmitter before it is first switched on. After
power-up, the device performs a self-test and then goes to the measuring mode.
If you are commissioning the device for the first time, program the setup as described in the
following sections of the Operating Instructions.
Endress+Hauser 25
Commissioning Smartec CLD18
Out1 0/4 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the min. current
0 μS/cm value (0/4 mA) is present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the max. current
0 μS/cm value (20 mA) is present at the transmitter output.
Out2 0/4 mA -50 to 250 °C Use this parameter to enter the measured value at which the min. current
0.0 °C value (0/4 mA) is present at the transmitter output.
Out2 20 mA -50 to 250 °C Use this parameter to enter the measured value at which the max. current
100.0 °C value (20 mA) is present at the transmitter output.
Damping main 0 ... 60 s Damping value for the conductivity measured value
0s
Manual hold Off, On Function for freezing the current and alarm outputs
Hold release 0 to 600 s Prolongs the device hold when the hold condition no longer
0s applies
26 Endress+Hauser
Smartec CLD18 Commissioning
Inst. factor 0.1 to 5.0 The affect of the distance from the wall can be corrected with the
1.0 installation factor (see the "Installation factor" section).
Alpha coeff. 1.0 to 20.0 %/K Coefficient for linear temperature compensation
2.1 %/K
Process check The process check checks the measuring signal for stagnation. An
alarm is triggered if the measuring signal does not change over a
specific period (several measured values).
Duration 1 to 240 min The measured value must change within this time as otherwise
60 min an error message is triggered.
Out1 0/4 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the min.
0 μS/cm current value (0/4 mA) is present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the
0 μS/cm max. current value (20 mA) is present at the transmitter output.
Out2 0/4 mA -50 to 250 °C Use this parameter to enter the measured value at which the min.
0.0 °C current value (0/4 mA) is present at the transmitter output.
Out2 20 mA -50 to 250 °C Use this parameter to enter the measured value at which the
100.0 °C max. current value (20 mA) is present at the transmitter output.
Out1 0/4 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the min.
0 μS/cm current value (0/4 mA) is present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm Use this parameter to enter the measured value at which the
0 μS/cm max. current value (20 mA) is present at the transmitter output.
Out2 0/4 mA -50 to 250 °C Use this parameter to enter the measured value at which the min.
0.0 °C current value (0/4 mA) is present at the transmitter output.
Out2 20 mA -50 to 250 °C Use this parameter to enter the measured value at which the
100.0 °C max. current value (20 mA) is present at the transmitter output.
Endress+Hauser 27
Commissioning Smartec CLD18
Please confirm no
no, yes
28 Endress+Hauser
Smartec CLD18 Commissioning
1.05
2
1.00
1
0.95
0 10 20 30 40 a [mm]
14 Relationship between the installation factor (f) and the distance from wall (a)
1 Electrically conductive pipe wall
2 Electrically insulating pipe wall
Endress+Hauser 29
Commissioning Smartec CLD18
Where
k(T) = conductivity at process temperature T
k(T0) = conductivity at process temperature T0
The temperature coefficient depends on both the chemical composition of the solution and on
the temperature, and is between 1 and 5 % per °C. The electrical conductivity of the majority
of diluted saline solutions and natural waters changes in a close-to-linear fashion.
Typical values for the temperature coefficient α:
30 Endress+Hauser
Smartec CLD18 Commissioning
General
A conductivity measuring system is generally calibrated in such a way that the exact cell
constant is determined or checked using suitable calibration solutions. This process is
described in the standards EN 7888 and ASTM D 1125, for example, and the method for
producing a number of calibration solutions is explained.
Endress+Hauser 31
Commissioning Smartec CLD18
32 Endress+Hauser
Smartec CLD18 Commissioning
Endress+Hauser 33
Diagnostics and troubleshooting Smartec CLD18
Simulation
Reset device
No measured value displayed No power supply connected Check the power supply of the device.
Diagnostic message is displayed The list of diagnostic messages is provided in the following section.
34 Endress+Hauser
Smartec CLD18 Diagnostics and troubleshooting
Endress+Hauser 35
Diagnostics and troubleshooting Smartec CLD18
36 Endress+Hauser
Smartec CLD18 Diagnostics and troubleshooting
Endress+Hauser 37
Maintenance Smartec CLD18
10 Maintenance
LWARNING
Risk of injury if medium escapes!
‣ Before performing any maintenance work, make sure that the process pipe is
unpressurized, empty and has been rinsed.
The electronics box does not contain any parts that the user must maintain.
• The cover on the electronics box may be opened only by Endress+Hauser Service staff.
• The electronics box may be removed only by Endress+Hauser Service staff.
NOTICE
Prohibited cleaning agents
Damage to the housing surface or housing seal
‣ Never use concentrated mineral acids or alkaline solutions for cleaning.
‣ Never use organic cleaners such as benzyl alcohol, methanol, methylene chloride, xylene or
concentrated glycerol cleaner.
‣ Never use high-pressure steam for cleaning purposes.
38 Endress+Hauser
Smartec CLD18 Repairs
11 Repairs
The O-ring is defective if medium escapes from the leakage hole. Contact the E+H Service
Department to replace the O-ring.
11.1 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at www.endress.com/support/return-material.
11.2 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.
Endress+Hauser 39
Accessories Smartec CLD18
12 Accessories
The following are the most important accessories available at the time this
documentation was issued. For accessories not listed here, please contact your service or
sales office.
13 Technical data
13.1 Input
13.1.1 Measured variable
Conductivity
Temperature
40 Endress+Hauser
Smartec CLD18 Technical data
Voltage range 0 V to 30 V
Voltage HIGH min. 12 V
Voltage LOW max. 9.0 V
Current consumption at 24 V 30 mA
Undefined voltage range 9.0 to 12 V
13.2 Output
13.2.1 Output signal
13.2.2 Load
Max. 500 Ω
13.2.3 Characteristic
Linear
Endress+Hauser 41
Technical data Smartec CLD18
13.4.3 Repeatability
42 Endress+Hauser
Smartec CLD18 Technical data
13.5 Environment
13.5.1 Ambient temperature range
13.5.3 Humidity
≤ 100 %, condensating
Endress+Hauser 43
Technical data Smartec CLD18
13.6 Process
13.6.1 Process temperature
Stainless steel process connection:
-10 to +110 °C (14 to 230 °F)
Max.130 °C (266 °F) up to 60 minutes
44 Endress+Hauser
Smartec CLD18 Technical data
p (abs.)
[psi] [bar]
188.5 13
1
130.5 9
2
87 6
14.5 1
50 60 110 130 T [°C]
15 Pressure-temperature ratings
1 Stainless steel process connection
2 PVC process connection
A Process temperature increased briefly (max. 60 minutes)
Endress+Hauser 45
Technical data Smartec CLD18
13.7.2 Weight
13.7.3 Materials
46 Endress+Hauser
Smartec CLD18 Index
Index
A Menus . . . . . . . . . . . . . . . . . . . . . . . . 24
Accessories . . . . . . . . . . . . . . . . . . . . . 40 MRS . . . . . . . . . . . . . . . . . . . . . . . . . 31
Advanced Setup . . . . . . . . . . . . . . . . . . 26
Air set . . . . . . . . . . . . . . . . . . . . . . . . 33
N
Application examples . . . . . . . . . . . . . . . 15 Nameplate . . . . . . . . . . . . . . . . . . . . . . 8
C O
Calibration . . . . . . . . . . . . . . . . . . . . . 31 Occupational safety . . . . . . . . . . . . . . . . . 5
Cell constant . . . . . . . . . . . . . . . . . . . . 31 Operating keys . . . . . . . . . . . . . . . . . . . 23
Certificates and approvals . . . . . . . . . . . . . 10 Operation . . . . . . . . . . . . . . . . . . . . . . 22
Cleaning the housing . . . . . . . . . . . . . . . 38 Operational safety . . . . . . . . . . . . . . . . . . 7
Commissioning . . . . . . . . . . . . . . . . . . . 25 Orientations . . . . . . . . . . . . . . . . . . . . . 12
D P
Declaration of Conformity . . . . . . . . . . . . . 10 Post-connection check . . . . . . . . . . . . . . . 21
Designated use . . . . . . . . . . . . . . . . . . . . 5 Post-installation check . . . . . . . . . . . . . . 17
Device configuration . . . . . . . . . . . . . . . . 26 Power up . . . . . . . . . . . . . . . . . . . . . . 25
Device diagnosis . . . . . . . . . . . . . . . . . . 34 Product description . . . . . . . . . . . . . . . . 11
Diagnostic messages . . . . . . . . . . . . . . . . 35 Product identification . . . . . . . . . . . . . . . . 8
Diagnostics . . . . . . . . . . . . . . . . . . . . . 34 Product page . . . . . . . . . . . . . . . . . . . . . 9
Display settings . . . . . . . . . . . . . . . . . . . 25 Product safety . . . . . . . . . . . . . . . . . . . . 7
Disposal . . . . . . . . . . . . . . . . . . . . . . . 39
R
E Repairs . . . . . . . . . . . . . . . . . . . . . . . . 39
Electrical connection . . . . . . . . . . . . . . . . 18 Residual coupling . . . . . . . . . . . . . . . . . . 33
Ensuring the degree of protection . . . . . . . . 21 Return . . . . . . . . . . . . . . . . . . . . . . . . 39
I S
Incoming acceptance . . . . . . . . . . . . . . . . 8 Safety instructions . . . . . . . . . . . . . . . . . . 5
Installation . . . . . . . . . . . . . . . . . . . 12, 17 Scope of delivery . . . . . . . . . . . . . . . . . . . 9
Installation conditions . . . . . . . . . . . . . . . 12 Symbols . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation examples . . . . . . . . . . . . . . . 15
T
Installation factor . . . . . . . . . . . . . . . . . 29
Technical data . . . . . . . . . . . . . . . . . . . 40
Interpreting the order code . . . . . . . . . . . . . 9
Temperature compensation . . . . . . . . . . . . 30
IT security measures . . . . . . . . . . . . . . . . . 7
Trouble shooting instructions . . . . . . . . . . 34
M Troubleshooting . . . . . . . . . . . . . . . . . . 34
Maintenance . . . . . . . . . . . . . . . . . . . . 38
W
Manufacturer's address . . . . . . . . . . . . . . . 9
Warnings . . . . . . . . . . . . . . . . . . . . . . . 4
Measuring range switching . . . . . . . . . . . . 31
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 18
Menu . . . . . . . . . . . . . . . . . . . . . . . . . 26
Calibration . . . . . . . . . . . . . . . . . . 31
Diagnostics . . . . . . . . . . . . . . . . . . 34
Display . . . . . . . . . . . . . . . . . . . . . 25
Setup . . . . . . . . . . . . . . . . . . . . . . 26
Endress+Hauser 47
*71354508*
71354508
www.addresses.endress.com