Sensor CLS12 CLS13
Sensor CLS12 CLS13
Sensor CLS12 CLS13
16
71340911
Products Solutions Services
Operating Instructions
Condumax CLS12/CLS13
Conductivity sensors for high-temperature
applications
0
Table of contents
Table of contents
1 Document information . . . . . . . . . 3
1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4 Installation . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Mounting the sensor . . . . . . . . . . . . . . . . . 9
4.2 Post-installation check . . . . . . . . . . . . . . . . 9
5 Electrical connection . . . . . . . . . . . 9
5.1 Connection conditions . . . . . . . . . . . . . . . 10
5.2 Connecting the sensor . . . . . . . . . . . . . . . 11
5.3 Ensuring the degree of protection . . . . . 11
5.4 Post-connection check . . . . . . . . . . . . . . . 11
6 Commissioning . . . . . . . . . . . . . . . . 12
7 Maintenance . . . . . . . . . . . . . . . . . . 12
8 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . 13
8.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
9 Technical data . . . . . . . . . . . . . . . . . 14
10 EU Declaration of
conformity . . . . . . . . . . . . . . . . . . . . . 17
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2 Endress+Hauser
Condumax CLS12/CLS13 Document information
1 Document information
1.1 Warnings
1.2 Symbols
Symbol Meaning
Permitted or recommended
Reference to page
Reference to graphic
Result of a step
Endress+Hauser 3
Basic safety instructions Condumax CLS12/CLS13
4 Endress+Hauser
Condumax CLS12/CLS13 Basic safety instructions
2. Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Label the damaged product as defective.
3. If faults cannot be rectified:
Take the products out of operation and safeguard them to ensure that they are not
operated inadvertently.
Endress+Hauser 5
Basic safety instructions Condumax CLS12/CLS13
Temperature classes
Name Type Medium temp. Ta for temperature class (Tn) Cat.
x1 x2 x3 x4
The following connection values are safety limits which may not be exceeded when connecting
to the transmitter:
Parameters Connection data
6 Endress+Hauser
Condumax CLS12/CLS13 Incoming acceptance and product identification
Endress+Hauser 7
Incoming acceptance and product identification Condumax CLS12/CLS13
3. Enter the order code from the nameplate into the search field, and then select "Show
details".
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
Declaration of conformity
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EU directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
8 Endress+Hauser
Condumax CLS12/CLS13 Installation
4 Installation
4.1 Mounting the sensor
‣ Mount the sensor directly on a process connection that matches the version. Alternatively
you can install the sensor in a flow assembly CLA751.
Ensure that the electrodes are fully immersed in the medium during measurement. If the
sensor is being used in the ultrapure water range, you must work under air-evacuated
conditions. Otherwise the CO2 in the air can dissolve in the water and its (weak)
dissociation can increase the conductivity by up to 3 µS/cm.
5 Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
Endress+Hauser 9
Electrical connection Condumax CLS12/CLS13
EX
Zone 2
IIC 3G
EX
IIC 2G
Zone 1
Ex ia Ga Ex ia Ga CPS41 CPS441
CPS42
CPS11
CPS12
Zone 0
IIC 1G
CPS71
CPS72
CPS91
CPS13
CPS64
CPS471
CPS491 CLS50 CLS54 CLS12 CLS13 CLS15 CLS16 CLS21
A0031175
10 Endress+Hauser
Condumax CLS12/CLS13 Electrical connection
Cable Sensor
GY Plug
GN/YE
GN 3
Pt 100
WH
Outer shield
} YE } 1
(see wiring diagram BN n.c.
of transmitter)
A0024205-EN
A VMB junction box and another CYK71 cable are required for the cable extension.
Are the outside of the sensor, assembly, cable undamaged? Visual inspection
Is a sufficient length of the cable cores stripped, and is it positioned Check the fit (by pulling gently)
in the terminal correctly?
Are all cable entries mounted, tightened and leak-tight? For lateral cable entries, make sure the cables
loop downwards to allow water to drip off
Are all cable entries installed downwards or mounted laterally?
Endress+Hauser 11
Commissioning Condumax CLS12/CLS13
6 Commissioning
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check that the cleaning medium (e.g. water or
air) is connected correctly.
LWARNING
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
‣ Before applying compressed air to an assembly with cleaning facility, make sure the
connections are correctly fitted.
‣ Do not install the assembly in the process if you cannot make the correct connection
reliably.
7 Maintenance
LCAUTION
Corrosive chemicals
Danger of chemical burns to the eyes and skin. Danger of damage to clothing and equipment
‣ It is absolutely essential to protect the eyes and hands properly when working with acids,
bases and organic solvents!
‣ Wear protective goggles and safety gloves.
‣ Clean away splashes on clothes and other objects to prevent any damage.
‣ Pay particular attention to the information provided in the safety data sheets for the
chemicals used.
LWARNING
Hydrofluoric acid and mineral acids
Risk of serious or fatal injury from caustic burns
‣ Wear protective goggles to protect your eyes.
‣ Wear protective gloves and appropriate protective clothing.
‣ Avoid all contact with the eyes, mouth and skin.
‣ If using hydrofluoric acid, only use plastic vessels.
LWARNING
Thiocarbamide
Harmful if swallowed. Limited evidence of carcinogenicity. Possible risk of harm to the unborn
child. Dangerous for the environment with long-term effects.
‣ Wear protective goggles, protective gloves and appropriate protective clothing.
‣ Avoid all contact with the eyes, mouth and skin.
‣ Avoid releases into the environment.
12 Endress+Hauser
Condumax CLS12/CLS13 Repair
Clean away fouling on the sensor as follows depending on the type of fouling:
1. Oily and greasy films:
Clean with grease remover, e.g. alcohol, as well as hot water and (alkaline) agents
containing surfactants (e.g. dishwashing detergent).
2. Lime, cyanide and metal hydroxide buildup and low solubility organic buildup:
Dissolve buildup with diluted hydrochloric acid (3 %) and then rinse thoroughly with
plenty of clear water.
3. Sulfidic buildup (from flue gas desulfurization or sewage treatment plants):
Use a mixture of hydrochloric acid (3 %) and thiocarbamide (commercially available)
and then rinse thoroughly with plenty of clear water.
4. Buildup containing proteins (e.g. food industry):
Use a mixture of hydrochloric acid (0.5 %) and pepsin (commercially available) and then
rinse thoroughly with plenty of clear water.
5. Readily soluble biological buildup:
Rinse with pressurized water.
After cleaning or regeneration, you must rinse the sensor thoroughly with water .
8 Repair
8.1 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at www.endress.com/support/return-material.
8.2 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.
Endress+Hauser 13
Technical data Condumax CLS12/CLS13
9 Technical data
9.1 Input
9.1.1 Measured values
• Conductivity
• Temperature
9.2 Environment
9.2.1 Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
14 Endress+Hauser
Condumax CLS12/CLS13 Technical data
9.3 Process
9.3.1 Process temperature
CLS12
-20 to 160 ˚C (-4 to 320 ˚F)
CLS13
-20 to 250 °C (-4 to 480 °F)
CLS12
CLS13
1 to 41 bar (15 to 595 psi), absolute
CLS12
p (abs.)
[psi] [bar]
595 41
435 30
290 20
102 7
T
[°C]
-20 0 30 70 100 160
[°F]
-4 32 212 320
A0032120-EN
Endress+Hauser 15
Technical data Condumax CLS12/CLS13
CLS13
p (abs.)
[psi] [bar]
595 41
435 30
290 20
145 10
T
[°C]
-20 20 60 100 140 180 220 250
[°F]
-4 68 212 356 482
A0032124-EN
CLS12
Approx. 1.4 kg (3.1 lbs) depending on version
CLS13
Approx. 1.7 kg (3.75 lbs) depending on version
9.4.2 Materials
CLS12
Electrodes Stainless steel 1.4571 (AISI 316Ti)
Process connection Stainless steel 1.4571 (AISI 316Ti)
Terminal head Die-cast aluminum
Insulator body EPDM, PEEK
CLS13
Electrodes Stainless steel 1.4571 (AISI 316Ti)
Process connection Stainless steel 1.4571 (AISI 316Ti)
Terminal head Die-cast aluminum
Dissipator Die-cast aluminum
Seals FFKM
Insulator body Ceramic
16 Endress+Hauser
Condumax CLS12/CLS13 EU Declaration of conformity
10 EU Declaration of conformity
Endress+Hauser 17
Index
Index
A O
Ambient temperature range . . . . . . . . . . . 14 Occupational safety . . . . . . . . . . . . . . . . . 4
Operational safety . . . . . . . . . . . . . . . . . . 4
C
Cell constant . . . . . . . . . . . . . . . . . . . . 14 P
Certification body . . . . . . . . . . . . . . . . . . 8 Pressure-temperature ratings . . . . . . . . . . 15
Check Process . . . . . . . . . . . . . . . . . . . . . . . . 15
Connection . . . . . . . . . . . . . . . . . . 11 Process connection . . . . . . . . . . . . . . . . . 17
Installation . . . . . . . . . . . . . . . . . . . 9 Process pressure . . . . . . . . . . . . . . . . . . 15
Connection Process temperature . . . . . . . . . . . . . . . . 15
Check . . . . . . . . . . . . . . . . . . . . . 11 Product identification . . . . . . . . . . . . . . . . 7
Ensuring the degree of protection . . . . . 11 Product safety . . . . . . . . . . . . . . . . . . . . 5
Connection conditions . . . . . . . . . . . . . . . 10
R
D Repair . . . . . . . . . . . . . . . . . . . . . . . . 13
Declaration of conformity . . . . . . . . . . . 8, 17 Return . . . . . . . . . . . . . . . . . . . . . . . . 13
Degree of protection
Ensuring . . . . . . . . . . . . . . . . . . . . 11 S
Technical data . . . . . . . . . . . . . . . . 14 Safety
Designated use . . . . . . . . . . . . . . . . . . . . 4 Electrical equipment in hazardous areas . . 5
Disposal . . . . . . . . . . . . . . . . . . . . . . . 13 Occupational safety . . . . . . . . . . . . . . 4
Operation . . . . . . . . . . . . . . . . . . . . 4
E Product . . . . . . . . . . . . . . . . . . . . . 5
Electrical connection . . . . . . . . . . . . . . . . 9 Safety instructions . . . . . . . . . . . . . . . . . . 4
Environment . . . . . . . . . . . . . . . . . . . . 14 Scope of delivery . . . . . . . . . . . . . . . . . . . 8
EU Declaration of conformity . . . . . . . . . . . 17 Sensor
Cleaning . . . . . . . . . . . . . . . . . . . . 12
H Connecting . . . . . . . . . . . . . . . . . . 11
Hazardous area approvals . . . . . . . . . . . . . 8 Mounting . . . . . . . . . . . . . . . . . . . . 9
Hazardous areas . . . . . . . . . . . . . . . . . . . 5 State of the art . . . . . . . . . . . . . . . . . . . . 5
Storage temperature . . . . . . . . . . . . . . . . 14
I Symbols . . . . . . . . . . . . . . . . . . . . . . . . 3
Incoming acceptance . . . . . . . . . . . . . . . . 7
Installation T
Check . . . . . . . . . . . . . . . . . . . . . . 9 Technical data
Sensor . . . . . . . . . . . . . . . . . . . . . . 9 Environment . . . . . . . . . . . . . . . . . 14
Input . . . . . . . . . . . . . . . . . . . . . . 14
M Mechanical construction . . . . . . . . . . 16
Manufacturer inspection certificate . . . . . . . . 8 Process . . . . . . . . . . . . . . . . . . . . . 15
Materials . . . . . . . . . . . . . . . . . . . . . . 16 Temperature compensation . . . . . . . . . . . . 14
Measured values . . . . . . . . . . . . . . . . . . 14 Temperature/pressure ratings . . . . . . . . . . 15
Measuring ranges . . . . . . . . . . . . . . . . . 14
U
N Use . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Nameplate . . . . . . . . . . . . . . . . . . . . . . 7
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . 3
18 Endress+Hauser
Index
Weight . . . . . . . . . . . . . . . . . . . . . . . . 16
Endress+Hauser 19