Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Manual Single Gas Detector - Drager Pac 6500

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

Dräger Pac 6x00 / 8x00

Technical manual

i
Contents en

Contents
1 Safety-related information . . . . . . . . . . . . . . . . . . .4
1.1 Basic safety information . . . . . . . . . . . . . . . . . . . . . .4
1.2 Safety information on explosion protection . . . . . . . .4
2 Conventions used in this document . . . . . . . . . . .4
2.1 Definition of alert icons . . . . . . . . . . . . . . . . . . . . . . .4
2.2 Typographical conventions . . . . . . . . . . . . . . . . . . . .4
2.3 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1.1 Gas detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
3.3 Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4.1 Preparations for use . . . . . . . . . . . . . . . . . . . . . . . . .6
4.1.1 Initial start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4.1.2 Switching on the gas detector . . . . . . . . . . . . . . . . . .6
4.1.3 Switching off the gas detector . . . . . . . . . . . . . . . . . .6
4.2 Before entering the workplace . . . . . . . . . . . . . . . . .6
4.2.1 Activating the quick menu . . . . . . . . . . . . . . . . . . . . .6
4.2.2 Opening the quick menu . . . . . . . . . . . . . . . . . . . . . .6
4.2.3 Carrying out a manual bump test . . . . . . . . . . . . . . .6
4.2.4 Information display . . . . . . . . . . . . . . . . . . . . . . . . . .7
4.2.5 Error and warning code displays . . . . . . . . . . . . . . . .7
4.3 During operation . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
5 Calibrating the gas detector . . . . . . . . . . . . . . . . .8
5.1 Opening the maintenance menu . . . . . . . . . . . . . . . .8
5.2 Carrying out a manual fresh air calibration . . . . . . . .8
5.3 Carrying out a manual span calibration . . . . . . . . . .9
5.4 Carrying out manual calibration with the Dräger CC-Vi-
sion Basic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
5.5 Carrying out an automatic calibration with the Dräger
X-dock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . .10
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
7.1 Configuring the gas detector . . . . . . . . . . . . . . . . . .13
7.2 Operating time alarm / end of the operating time . .13
7.3 Data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
7.4 Replacing the sensor . . . . . . . . . . . . . . . . . . . . . . .13
7.5 Replacing the battery . . . . . . . . . . . . . . . . . . . . . . .14
7.6 Replacing the dust and water filter . . . . . . . . . . . . .15
8 Device settings . . . . . . . . . . . . . . . . . . . . . . . . . . .15
8.1 Factory setting . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
8.2 Alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
9 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
10.1 Gas detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
10.2 Technical data of the sensors and the measuring de-
vice settings for Pac 6000/6500 . . . . . . . . . . . . . . .18
10.3 Technical data of the sensors and the measuring de-
vice settings for Pac 8000 . . . . . . . . . . . . . . . . . . . .19
10.3.1 Sensors for Pac 8500 DUAL . . . . . . . . . . . . . . . . . .21
10.3.2 Cross-sensitivities . . . . . . . . . . . . . . . . . . . . . . . . . .22

3
en Safety-related information

1 Safety-related information 2 Conventions used in this


document
1.1 Basic safety information
 Before using this product, carefully read the associated 2.1 Definition of alert icons
instructions for use. This document does not replace the
The following alert messages are used in this document to
instructions for use.
provide and highlight areas of the associated text that require
a greater awareness by the user. A definition of the meaning of
Incorrect calibration each alert message is as follows:
An incorrect calibration leads to incorrect measured values.
► The sensitivity must be checked daily before first using the WARNING
device, with a known concentration of the gas to be Indicates a potentially hazardous situation. If not avoided, it
measured that corresponds to 25 to 50 % of the final could result in death or serious injury.
concentration. The accuracy must amount to 0 to +20 % of
NOTICE
the actual value. The accuracy can be corrected by
Indicates a potentially hazardous situation. If not avoided, it
calibration.
could result in damage to the product or environment.
1.2 Safety information on explosion
protection 2.2 Typographical conventions
Devices or components that are used in potentially explosive
atmospheres and which are tested and approved in This symbol identifies information that make the product
accordance with national, European or international explosion easier to use.
protection guidelines may only be used under the approved
conditions in compliance with the statutory provisions. 2.3 Glossary
Oxygen-enriched atmospheres
Technical term Explanation
Explosion protection is not guaranteed in oxygen-enriched
atmospheres (>21 Vol% O2). Operation signal A periodic optical (green LED) and/or
acoustic signal.
► Remove the device from the potentially explosive
atmosphere. D-Light The D-Light allows the user to check and
Risk of explosion! indicate compliance with certain settings
► Do not open the gas detector in explosion-hazard areas. (e.g. bump test interval). The green LED
flashes for a short period and
superimposes the optical operation
Specific application conditions
signal.
 In certain extreme conditions, open plastic parts and non-
earthed metal parts of the housing may store a combustible
level of electrostatic charge.
 Activities such as carrying the device in a bag or attached
to a belt, operating the button field or cleaning the device
with a damp cloth do not represent a significant
electrostatic danger. However, if a mechanism which
generates static charge, such as repeated rubbing against
clothing, is identified, suitable precautionary measures
must be implemented, e.g. the use of anti-static clothing
and anti-static footwear.

4
Description en

3 Description
3.1 Product overview
3.1.1 Gas detector

2 3
1 4 8 (4x)

5 9
7 6

00333742.eps

1 Alarm LEDs 6 [OK] button


2 Operation signal/ 7 [▼] button
D-Light
3 Gas inlet 8 Screw (4x)
4 Horn 9 IR interface
5 Display

3.1.2 Display

1 2 3 4

9 8 7 6 5

00733742.eps

1 Key word symbol 6 Span calibration


2 Error symbol 7 Fresh air adjustment
3 Information symbol 8 TWA/STEL
4 Battery charge status 9 Peak concentration
5 Measurement unit

3.2 Intended use


The Dräger Pac 6x00/8x00 is a gas detector and is used to
measure and alert the user of gas concentrations in the
ambient air.

3.3 Approvals
A copy of the name plate and the declaration of conformity are
provided in the enclosed supplementary documentation (order
no. 90 33 741).
The name plate on the gas detector must not be concealed.

5
en Use

4 Use After the gas detector has been switched on, the current
measured value is shown in the display.
4.1 Preparations for use Check whether the warning [!] appears. If it is displayed, we
recommend performing a bump test, see Section 4.2.3 on
4.1.1 Initial start-up Page 6.
The gas detector remains in deep sleep mode upon delivery 4.2.1 Activating the quick menu
and must be activated during the initial start-up.
The quick menu can be activated using the Dräger CC-Vision
1. Hold down the [▼] button for approx. 3 s. Basic PC software.
This activates the gas detector.
Up to two preferred functions can be saved in the quick menu.
4.1.2 Switching on the gas detector The following functions are available:
1. Hold down the [OK] button for approx. 3 s.  Fresh air adjustment
 Bump test
The following is displayed or activated:
 Display elements, LEDs, alarm signal and vibration alarm 4.2.2 Opening the quick menu
 Self-test
Prerequisite:
 Software version and gas name
 The quick menu is activated.
 Alarm thresholds A1 and A2
 Time to next calibration (configurable) To open the quick menu:
 Time to expiration of the bump test interval (configurable) 1. Press the [▼] button three times within three seconds. A
double signal tone sounds.
Before every use, check whether the display elements and 2. Depending on the configuration, the fresh air calibration
information are displayed correctly. symbol or the bump test signal will flash.
3. Press the [OK] button to start the feature or press the [▼]
A warm-up phase takes place when first switching on the gas button to switch to the next feature or to return to
detector (duration depends on the sensor type; see the sensor measuring mode (depending on the configuration).
data sheet).
4.1.3 Switching off the gas detector The quick menu closes automatically after 60 s of inactivity.
 Hold down both keys for approx. 3 s, until shut-down is
complete. 4.2.3 Carrying out a manual bump test

4.2 Before entering the workplace WARNING


Danger to health! Do not inhale the test gas.
WARNING ► Pay attention to the hazard information in the respective
Serious damage to health! safety data sheets.
An incorrect calibration can lead to incorrect measurement
results, which may result in serious damage to health. Two modes can be selected for a bump test. Use the Dräger
► Before performing safety measurements, check the CC-Vision Basic PC software to set the mode.
calibration by way of a bump test, adjust as necessary,  Quick bump test (test for alarm triggering)
 Checks whether the pre-alarm threshold (A1) is
and check all alarm elements. If national regulations exist,
the bump test must be performed in accordance with exceeded (or not reached in the case of O2).
 Checks whether the concentration remains above the
these regulations.
alarm for a specified period of time.
 Checks whether the test duration remains below a

WARNING specified maximum time period.


Incorrect measurement results!  Advanced bump test (test for accuracy)
 Checks whether the defined test gas concentration is
The gas inlet opening is fitted with a filter that protects against
reached within a specified tolerance. The tolerance is
dust and water. Contamination may change the qualities of the
defined with a default value for each gas in the gas
dust and water filter.
detector, but can be adjusted by the user as needed.
► Do not damage the filter. Immediately replace damaged or
 Checks whether the test gas concentration remains
blocked filters.
within the tolerance window for a specified time period.
 Checks whether the test duration remains below a
To ensure correct function:
specified maximum time period.
 Do no cover the gas inlet opening.
 Place the gas detector on clothing near the mouth. The bump test can be performed as follows:
 At temperatures below -20 °C, there may be deviations of  Manual bump test (with quick menu)
>10 % of the measured value if the corresponding sensor  Bump test with X-dock (see instructions for use for the
was calibrated at room temperature. Dräger recommends Dräger X-dock)
calibration at the primary application temperature if  Bump test with the Dräger Bump Test Station (see the quick
measurements will be taken at very low temperatures. This reference guide on the Dräger Bump Test Station)
enables the highest possible measurement accuracy.

6
Use en

Prerequisites for the manual bump test:  Short-term exposure limit (STEL1), not for Pac 6000)
 The gas detector is switched on.  Duration of use (Pac 6000 always, Pac 6500 and 8x00
 A suitable test gas cylinder is available, e.g. a test gas depending on the configuration)
cylinder (order no. 68 11 130) with the following mixed gas  Device ID
ratios: 50 ppm CO, 15 ppm H2S, 2.5 Vol% CH4, 18 Vol% To display the information:
O2
1. Press the [OK] button in measuring mode.
To perform a manual bump test: This displays the peak concentration and the symbol for
1. Prepare a Dräger test gas cylinder. The volume flow must peak concentration.
be 0.5 L/min and the gas concentration must be higher 2. Press the [OK] button to display the information
than the alarm threshold concentration that is to be tested. consecutively. The information display closes after the
2. Connect the gas detector and the test gas cylinder to the device ID.
calibration adapter.
4.2.5 Error and warning code displays
3. Press the [▼] button three times within three seconds to
open bump test mode (if configured). A signal tone sounds. If a warning, an error or a notice is pending, the error or notice
The notice symbol starts flashing. symbol flashes and a three-digit error code appears.
4. Press the [OK] button to start the bump test. To display the error and warning codes:
5. Open the test gas cylinder valve to let test gas flow over the 1. Press the [OK] button.
sensor. If there are several pending error or warning codes, the next
6. If the gas concentration exceeds the alarm threshold A1 or error or warning code can be displayed by pressing the [OK]
A2, the corresponding alarm triggers. button.
 Bump test failed: The gas detector switches to error mode
and displays an error. The error symbol flashes, and an 4.3 During operation
error code is displayed until the error is confirmed.
Subsequently, instead of the measured value, the - – –
WARNING
display and the error symbol appear. In this case, repeat
the bump test or calibrate the gas detector. Danger to life and/or risk of explosion!
 Bump test passed: OK is displayed until the concentration The following alarms indicate a danger to life and/or explosion
is below A1. hazard:
 A2 alarm
The result of the bump test (passed or failed) is saved in the
 STEL or TWA alarm
data logger (see Section 7.1 on Page 13).
 Device error
4.2.4 Information display ► Immediately leave the hazard area.
Different information can be displayed when the device is on
and off. The continuous operation of the gas detector is indicated by an
optical and/or acoustic operation signal that is emitted in a 60-
second cycle (configurable using the Dräger CC-Vision Basic
The information display closes automatically after 3 s of PC software).
inactivity.
The respective measured value is displayed as an alternating
When the gas detector is off: display for dual sensors.
The following information is displayed:
 Gas name, full scale deflection and measurement module. The operation signal must be turned on for measurements
 Duration of use (Pac 6000 always, Pac 6500 and 8x00 in accordance with EN 45544 (CO, H2S) or in accordance
depending on the configuration) with EN 50104 (O2).
 Device ID
If the allowable measuring range is exceeded or a negative
To display the information: zero drift occurs, the following message appears in the display:
1. Hold down the [▼] button for approx. 1 s while the device (concentration too high) or (negative drift).
is off. The measuring channels do not need to be checked after a
The gas name, full scale deflection and measurement brief exceeding of the measuring range by the EC measuring
module are displayed. channels (up to one hour) (this does not apply when using
2. Press the [▼] button to display the information DrägerSensor XXS CO H2-CP).
consecutively. The information display closes after the
device ID. In the event of an alarm, the corresponding displays, including
the visual, audible and vibration alarms, are activated, see
When the gas detector is on: Section 8.2 on Page 16.
The following information can be displayed depending on the  Press the [▼] button to illuminate the display.
configuration:
 Error codes
 Peak concentration
1)
 Time-weighted average (TWA , not for Pac 6000)

1) configurable

7
en Calibrating the gas detector

5 Calibrating the gas detector 5.2 Carrying out a manual fresh air
calibration
WARNING To improve accuracy, a fresh air calibration can be carried out
Danger to health! if there is a zero deviation.
Do not inhale the test gas. Observe the hazard warnings on Observe the following notices for the calibration:
the corresponding Safety Data Sheets and the instructions for  During fresh air calibrations, the zero point of all sensors
use of the gas detector! Observe national regulations when (except the DrägerSensors XXS O2) is set to 0. For the
defining calibration intervals. XXS O2, the reading is set to 20.9 Vol%.
 The DrägerSensor XXS O3 must be calibrated with a
suitable zero gas which is free from carbon dioxide / ozone
For Pac 8500, two measuring channels are used for the (e.g. N2).
calibration.
To carry out the fresh air calibration:
 Calibration must be performed by qualified personnel after 1. Open the maintenance menu, see Section 5.1 on Page 8.
a failed bump test or after a defined calibration interval (see 2. Press the [OK] button while the fresh air calibration symbol
EU standard EN 50073). is flashing. The measured value flashes.
 Recommended calibration interval for the O2, H2S and CO 3. Press the [OK] button to confirm the fresh air calibration or
sensors: 6 months. Calibration intervals for other gases: the [▼] button to cancel the fresh air calibration.
see the instructions for use of the respective The gas detector returns to measuring mode, or the span
DrägerSensors. calibration symbol flashes (depending on the
Calibration can be carried out as follows: configuration).
 Manual calibration or calibration using the Dräger  Fresh air calibration passed: A short double tone sounds,
CC-Vision Basic and OK / gas name is shown in an alternating display.
 Automatic calibration with the Dräger X-dock (see the Press the [OK] button to return to measuring mode.
instructions for use for the Dräger X-dock)  Fresh air calibration failed: A long single tone sounds.

 Automatic calibration with the Dräger Bump Test Station – – – is displayed instead of the measured value. The error
(see the quick reference guide on the Dräger Bump Test symbol and the fresh air calibration symbol are displayed.
Station) In this case, repeat the fresh air calibration or calibrate the
gas detector.
5.1 Opening the maintenance menu

The gas detector automatically returns to measuring mode


if no button is pressed in the menu for 1 minute (with the
exception of the span calibration menus, where the menu
remains active for 10 minutes without any activity).

The maintenance menu is password protected.


Factory setting: 001

1. Hold down the [▼] button for 5 seconds.


2. Enter the password. Press the [▼] button to change the
value of the flashing position. Press the [OK] button to
confirm the value. Repeat this procedure to define the next
two values. After the last confirmation with the [OK] button,
the entire password flashes.
3. Press the [OK] button to confirm the password or the [▼]
button to cancel the password.
4. If the correct password is entered, a signal tone sounds
and the fresh air calibration symbol flashes (configurable).
5. Press the [OK] button to open the fresh air calibration or the
[▼] button to switch to the span calibration.
6. Press the [OK] button to open the span calibration or the
[▼] button to switch back to measuring mode.

8
Calibrating the gas detector en

5.3 Carrying out a manual span calibration 5.4 Carrying out manual calibration with the
 Prepare the test gas cylinder, connect the test gas cylinder Dräger CC-Vision Basic
to the calibration adapter and connect the calibration To calibrate the gas detector using the Dräger CC-Vision Basic
adapter to the gas detector. PC software:
1. Open the maintenance menu, see Section 5.1 on Page 8. 1. Connect the gas detector to a PC using the communication
2. Press the [OK] button while the span calibration symbol is module.
flashing. The configured test gas concentration is 2. The calibration can be carried out using the Dräger
displayed. The test gas concentration can be used or CC-Vision Basic PC software. For further information, see
adjusted according to the concentration in the test gas the Dräger CC-Vision Basic online help.
cylinder.
3. Press the [▼] button to change the configured test gas
concentration. The first position flashed. Press the [▼] A calibration interval can be set using the “Configurable
button to change the value of the flashing position. Press operating time” (in days) feature, see Section 7.1 on
the [OK] button to confirm the value. Repeat this procedure Page 13.
to define the next three values. After the last confirmation
with the [OK] button, the test gas concentration is 5.5 Carrying out an automatic calibration
complete. with the Dräger X-dock
4. Open the test gas cylinder valve to let test gas flow over the
sensor (flow: 0.5 L/min). The gas detector can be automatically calibrated using the
5. Wait until the displayed measured value is stable (after at Dräger X-dock, see the Dräger X-dock instructions for use.
least 120 seconds).
6. Press the [OK] button to start the span calibration. The
concentration display flashes.
7. As soon as the measured value displays a stable
concentration, press the [OK] button to confirm the span
calibration or the [▼] button to cancel the span calibration.
 Span calibration passed: A short double tone sounds, and
OK / gas name is shown in an alternating display. Press
the [OK] button to return to measuring mode.
 Span calibration failed: A tone sounds.
– – – is displayed instead of the measured value. The error
symbol and the span calibration symbol are displayed. In
this case, repeat the span calibration.

To test the measured value configuration times, allow t90


test gas to flow over the calibration adapter on the gas
detector. Check the results in accordance with the
information in the table (see the supplementary
documentation, order no. 90 33 741) up to a reading of
90 % of the final display.

For Pac 8500, the span calibration is carried out


consecutively for the various gasses.

9
en Troubleshooting

6 Troubleshooting
An error message is displayed in the event of instrument faults. The number that appears below the message is used for service
functions. If the fault remains after turning the device on and off several times, contact the service of Dräger.
Error code Cause Remedy
101 Life span of the gas detector expired. Use a new Pac gas detector.
102 The user's service life counter has elapsed. Reset the service life counter with the Dräger
CC-Vision Basic PC software.
103 Gas detector faulty. Contact the service of Dräger.
104 Program code checksum error. Contact the service of Dräger.
105 Bump test interval expired. Carry out bump test, see see Section 4.2.3 on
Page 6.
106 Calibration interval expired. Carry out a span calibration.
107 Software error Contact the service of Dräger.
109 General error. For example, menu functions cannot Check the configuration, view the error code (e.g. on
be carried out. the information display or with the Dräger CC-Vision
Basic PC software)
111 Alarm element test failed: Alarm lamp. Repeat the alarm element test with X-dock.
112 Alarm element test failed: Alarm horn. Repeat the alarm element test with X-dock.
113 Alarm element test failed: Vibration motor. Repeat the alarm element test with X-dock.
114 Parameter check failed. Correct the parameters and repeat the test with
X-dock.
115 Gas detector deactivated by X-dock. Activate the gas detector with X-dock.
116 Wrong software version. Update the software.
117 User parameters implausible. Check and adjust the configuration of the user
parameters.
131 Gas detector faulty. Contact the service of Dräger.
132 Parameter checksum error. Carry out a sensor inauguration using the Dräger
CC-Vision Basic PC software.
133 Wrong software version for the options board Update the software.
134 Hardware error. Contact the service of Dräger.

201 No valid zero adjustment of DrägerSensor XXS Carry out a fresh air calibration, see Section 5.2 on
measuring channel 1. Page 8.
202 No valid span calibration of DrägerSensor XXS Carry out a span calibration, see Section 5.2 on
measuring channel 1. Page 8.
203 The measured value of DrägerSensor XXS Carry out a fresh air calibration, see Section 5.2 on
measuring channel 1 is in the negative range. Page 8.
204 DrägerSensor XXS measuring channel 1 is faulty or Check DrägerSensor XXS measuring channel 1, see
not connected. Section 7.4 on Page 13.
205 Error during the bump test of DrägerSensor XXS Repeat the bump test and calibrate or replace
measuring channel 1. DrägerSensor XXS measuring channel 1 if
necessary, see Section 7.4 on Page 13.
206 Faulty filter test. Repeat the filter test with X-dock.
207 Faulty rise time test. Repeat the rise test with X-dock.
208 User parameters of the sensor implausible. Check and adjust the configuration of the sensor
parameters.
212 Calibration interval of DrägerSensor XXS measuring Carry out a calibration, see Section 5 on Page 8.
channel 1 expired.
225 Compensation channel error on DrägerSensor XXS Carry out a calibration, see Section 5 on Page 8.
measuring channel 1.

10
Troubleshooting en

Error code Cause Remedy


227 Sensor hardware on the gas detector faulty. Contact the service of Dräger.
228 Compensation channel calibration error on Carry out a calibration, see Section 5 on Page 8.
DrägerSensor XXS measuring channel 1.

301 No valid zero adjustment of DrägerSensor XXS Carry out a fresh air calibration, see Section 5.2 on
measuring channel 2. Page 8.
302 No valid span calibration of DrägerSensor XXS Carry out a span calibration, see Section 5.2 on
measuring channel 2. Page 8.
303 The measured value of DrägerSensor XXS Carry out a fresh air calibration, see Section 5.2 on
measuring channel 2 is in the negative range. Page 8.
304 DrägerSensor XXS measuring channel 2 is faulty or Check DrägerSensor XXS measuring channel 2, see
not connected. Section 7.4 on Page 13.
305 Error during the bump test of DrägerSensor XXS Repeat the bump test and calibrate or replace
measuring channel 2. DrägerSensor XXS measuring channel 2 if
necessary, see Section 7.4 on Page 13.
306 Faulty filter test. Repeat the filter test with X-dock.
307 Faulty rise time test. Repeat the rise test with X-dock.
308 User parameters of the sensor implausible. Check and adjust the configuration of the sensor
parameters.
312 Calibration interval for measuring channel 2 expired. Carry out a calibration, see Section 5 on Page 8.
325 Compensation channel error on DrägerSensor XXS Carry out a calibration, see Section 5 on Page 8.
measuring channel 2.
327 Gas detector faulty. Contact the service of Dräger.
328 Compensation channel calibration error on Carry out a calibration, see Section 5 on Page 8.
DrägerSensor XXS measuring channel 2.

Warning code Cause Remedy


151 Dräger service life expired. Use a new Pac gas detector.
152 The user's service life counter will elapse soon. Reset the service life counter with the Dräger
CC-Vision Basic PC software.
153 Data memory 90% full. Read the data memory and clear it at the next
opportunity.
154 Data memory full. Read the data memory and clear it.
155 Bump test interval expired. Carry out bump test, see see Section 4.2.3 on
Page 6.
159 General warning pending. Menu functions cannot View the notice code on the information display and
be carried out (e.g. notice indicating a warming up confirm it if necessary.
sensor).
160 Date and time incorrect, e.g. after battery change. Set the date and time with the Dräger CC-Vision
Basic PC software.

251 DrägerSensor XXS measuring channel 1 warming Wait until warm-up time is complete.
up (warm-up phase 1).
252 DrägerSensor XXS measuring channel 1 warming Wait until warm-up time is complete.
up (warm-up phase 2).
253 Measuring channel 1 concentration has drifted into Carry out a fresh air calibration for DrägerSensor XXS
the negative range. measuring channel 1, see Section 5.2 on Page 8.
254 The temperature is too high. Operate the gas detector within the permissible
temperature range.

11
en Troubleshooting

Warning code Cause Remedy


255 The temperature is too low. Operate the gas detector within the permissible
temperature range.
256 Calibration interval for DrägerSensor XXS Carry out a span calibration for DrägerSensor XXS
measuring channel 1 expired. measuring channel 1, see Section 5.3 on Page 9
275 Compensation channel calibration interval Carry out a calibration for DrägerSensor XXS
measuring channel 1, see Section 5 on Page 8.
276 Calibration necessary because of over-gassing of Carry out a calibration for DrägerSensor XXS
the compensation channel. measuring channel 1, see Section 5 on Page 8.

351 DrägerSensor XXS measuring channel 2 warming Wait until warm-up time is complete.
up (warm-up phase 1).
352 DrägerSensor XXS measuring channel 2 warming Wait until warm-up time is complete.
up (warm-up phase 2).
353 Measuring channel 2 concentration has drifted into Carry out a fresh air calibration for DrägerSensor XXS
the negative range. measuring channel 2, see Section 5.2 on Page 8.
354 The temperature is too high. Operate the gas detector within the permissible
temperature range.
355 The temperature is too low. Operate the gas detector within the permissible
temperature range.
356 Calibration interval for DrägerSensor XXS Carry out a span calibration for DrägerSensor XXS
measuring channel 2 expired. measuring channel 2, see Section 5.3 on Page 9

12
Maintenance en

7 Maintenance 7.1 Configuring the gas detector


The following maintenance instructions must be carefully read, The gas detector can be configured using the Dräger
understood and followed to prevent flammable or combustible CC-Vision Basic PC software. For further information, see the
atmospheres from igniting and to ensure that the intrinsic Dräger CC-Vision Basic online help.
safety of the gas detector is not impaired. To configure the gas detector using the Dräger CC-Vision
Basic PC software:
WARNING 1. Use the communication module (order no. 83 18 587) to
Incorrect measurement! connect the gas detector to a PC or the Dräger X-dock
► A bump test and/or calibration must be performed each maintenance station.
time the gas detector is opened. This includes any battery 2. The configuration can be carried out using the Dräger
change as well as each sensor replacement in the gas CC-Vision Basic PC software. For further information, see
detector. the Dräger CC-Vision Basic online help or the instructions
for use of the Dräger X-dock maintenance station.

NOTICE 7.2 Operating time alarm / end of the


Damage to components! operating time
The gas detector contains components at risk of charging. An individual operating time can be configured using the
Before opening the gas detector, make sure that the operator Dräger CC-Vision Basic PC software to, for example, set a
is earthed in order to prevent damage to the gas detector. For “calibration date”, an “inspection date”, a “switch-off date”, an
example, earthing may be ensured by an ESD workplace “operating time alarm” etc.
(electro static discharge). If an operating time is configured, a warning period starts
before the end of the installed operating time. After the gas
detector is turned on, the remaining operating time flashes
NOTICE during this warning period, e.g. “30” / “d” ( in days).
Damage to the gas detector!
This operating time alarm triggers at 10 % of the configured
When replacing the battery or the sensor, make sure that no
operating time (maximum 30 days before the end of the
components are damaged or short-circuited.
operating time).
► Do not use sharp objects to remove the battery or the
sensor. 7.3 Data logger
The gas detector is equipped with a data logger. The data
WARNING logger saves events and the average concentration which are
Replacing components may impair the intrinsic safety of the stored for the duration of a period of time which can be variably
gas detector. The following maintenance instructions must be configured using the Dräger Gas Vision PC software or the
carefully read, understood and followed to prevent flammable Dräger CC-Vision Basic PC software. Depending on the
or combustible atmospheres from igniting and to ensure that number of changing measured values, the data logger runs for
the intrinsic safety of the gas detector is not impaired. up to four weeks at an interval of one minute (configurable). If
► When replacing the battery or a sensor, do not damage or the data logger memory is full, the data logger overwrites the
short-circuit any components. Do not use any sharp oldest data.
objects to remove the battery or the sensor.
7.4 Replacing the sensor

WARNING WARNING
Incorrect measurement results! Risk of explosion! Do not replace the sensor in explosion-
If a battery or sensor replacement is carried out incorrectly, this hazard areas. Replacing components may impair the intrinsic
may cause incorrect measurements. safety of the gas detector.
► Carry out a bump test and/or calibration each time the gas The following maintenance instructions must be carefully read,
detector is opened. understood and followed to prevent flammable or combustible
atmospheres from igniting and to ensure that the intrinsic
safety of the gas detector is not impaired.
► When replacing the sensor, make sure that no
components are damaged or short-circuited. Do not use
sharp objects to remove the sensor.

13
en Maintenance

7.5 Replacing the battery


If the gas detector can no longer be calibrated, the sensor
must be replaced!
WARNING
Risk of explosion!
Only use the DrägerSensor XXS with the same part ► Only the lithium battery type LBT 01** (order no.
number! 83 26 856) may be used.
► Do not remove or replace batteries in potentially explosive
1. Connect the gas detector to a PC using the communication areas. Do no throw used batteries into fire or try to open
module. them by force. Dispose of batteries in accordance with the
2. In the CC-Vision Basic PC software1), open the “Sensor national provisions.
replacement assistant” feature and follow the instructions.
a. Switch off the gas detector. The battery is part of the Ex approval.
b. Loosen the four screws on the rear part of the housing.
c. Open the front part of the housing.
d. Remove the battery (optional).
e. Remove the sensor.
f. Insert the new sensor and make a note of the sensor
code printed on the sensor.
g. If the battery was removed: Wait for 5 seconds, insert
new battery. The battery runtime is reset when the
35 Ncm
battery is inserted.
h. Place the front part of the housing on the gas detector
and re-tighten the four screws on the rear part of the
housing. 00533742.eps

3. Sign in the sensor with the previously recorded sensor


1. Switch off the gas detector.
code.
4. After the sensor replacement, the sensor requires a warm- 2. Loosen the four screws on the rear part of the housing.
up phase (see the sensor data sheet). The displayed 3. Open the front part of the housing, remove battery and wait
for 5 s.
concentration flashes until the warm-up phase is complete.
5. After the sensor replacement and the warm-up phase, the 4. Insert the new battery. Make sure the battery plug is
gas detector must be calibrated (see Section 5 on Page 8). inserted correctly (see figure). The battery runtime is reset
when the battery is inserted.
5. Place the front part of the housing on the gas detector and
If the sensor code of the new sensor does not match that re-tighten the four screws on the rear part of the housing
of the previous sensor, the new sensor must be signed on (torque: 35 Ncm).
with the CC-Vision Basic PC software as described. Dräger 6. Turn on the gas detector.
recommends signing on the sensor with the Dräger A successful battery replacement is confirmed with a
CC-Vision Basic PC software, even if the sensor code is vibration (5 s) when the gas detector is turned on.
identical to the previous one. Notice 160 is displayed after the start sequence (see
Section 6 on Page 10).
7. Confirm the notice with the [OK] button.
8. After the battery replacement, the sensor requires a warm-
up phase (see the sensor data sheet). The displayed
concentration flashes until the warm-up phase is complete.

1) A free version of the Dräger CC-Vision Basic PC software can be


downloaded at www.draeger.com/software

14
Device settings en

7.6 Replacing the dust and water filter 8 Device settings


Only trained and qualified personnel may change the device
settings.

8.1 Factory setting


The factory settings may differ for customised orders.
Bump test mode Quick bump test
Vibration alarm On
Bump test interval Off
Operation signal On
D-Light On
Switch off Always
Data logger interval 1 min
Usage duration (user) Off

00633742.eps

1. Place the special tool on the dust and water separator.


2. Press the lever on the special tool downward and then
remove the dust and water separator.
3. Insert a new dust and water separator. The dust and water
separator must click into place.

4 units 40 units
Sensor grille:
Order no.: Order no.:
Pac 6xx0 83 26 853 83 26 857
Pac 8xx0 83 26 852 83 26 859
For special gases (ozone, phosgene), the entire front casing,
including the special diaphragm, must be replaced due to the
special diaphragm.
Front casing: Order no.:
Pac 8000 ozone 83 26 851
Pac 8000 phosgene 83 26 854

15
en Device settings

8.2 Alarm settings


Alarm Display Latching Acknowledgeable LED Horn Vibration1)
Alarm 1 A1  
Alarm 2 A2  
STEL STEL  
TWA TWA  
2)
Pre-alarm  
Main alarm3)  
Instrument alarm  
1) The intensity of the vibration depends on the temperature
2) After the initial battery pre-alarm, the battery’s life span amounts to between 1 day and 2 weeks under normal conditions of use. The life span is shorter at low
temperatures and/or in the event of alarms.
3) The gas detector automatically switches off after 10 s.

16
Disposal en

9 Disposal 10 Technical data


This product must not be disposed of as household 10.1 Gas detector
waste. This is indicated by the adjacent symbol. You
can return this product to Dräger free of charge. For Ambient conditions during operation
information please contact the national marketing Temperature Up to -30 °C ... +55 °C
organizations or Dräger. (temperature depending (briefly up to 1 h -40 °C ...
on the sensor) +55 °C)

Batteries must not be disposed of as household waste. Humidity 10 ... 90 % r.h., non-condensing
This is indicated by the adjacent symbol. Dispose of Pressure 700 ... 1300 hPa
batteries at battery collection centers as specified by Ambient conditions during storage:
the applicable regulations.
Temperature 0 ... 40 °C
Humidity 30 ... 80 % r.h., non-condensing
Typ. battery lifetime (under normal conditions):
24 h operation/day, 24 months
1 min alarm/day O2 sensor: 10 months
Dual sensors (without O2):
22 months
Alarm volume approx. 90 dBA at a distance of
30 cm
Dimensions (not including 64 x 84 x 20 mm
clip)
Weight approx. 106 g (113 g with clip)
Degree of protection IP 68

17
10.2 Technical data of the sensors and the measuring device settings for Pac 6000/6500

18
CO-LC H2S-LC O2 SO2
Measuring range 0 ... 1999 ppm 0 ... 100 ppm 2 ... 25 Vol% 0 ... 100 ppm
Calibration gas concentration 50 ppm 20 ppm 18 Vol% 10 ppm
Temperature range, operation -40 ... 50 °C -40 ... 50 °C -40 ... 50 °C -30 ... 50 °C
en Technical data

-40 ... 122 °F -40 ... 122 °F -40 ... 122 °F -22 ... 122 °F
Alarm threshold A1 1) 30 ppm 5 ppm 19 Vol% 2) 1 ppm
Acknowledgeable   - 
Latching - -  -
Alarm threshold A2 1) 60 ppm 10 ppm 23 Vol% 2 ppm
Acknowledgeable - - - -
Latching    
TWA threshold value 1) 3) 30 ppm 10 ppm - 1 ppm
STEL threshold value 1) 3) 60 ppm 10 ppm - 1 ppm
Mean value period 15 min 15 min - 15 min
Measurement accuracy
Zero point:  ±2 ppm  ±0.1 ppm  ±0.2 Vol%  ±0.1 ppm
Sensitivity: [% of the measured value]  ±2 ±5 ±1  ±2
Long-term drift (20 °C)
Zero point:  ±2 ppm/a  ±0.2 ppm/a  ±0.5 Vol%/a  ±1 ppm/a
Sensitivity: [% of measured value/year]  ±3  ±5  ±1  ±2/month
Sensor item number 4) 68 13 210 68 11 525 68 10 881 68 10 885
Sensor data sheet item number 90 33 454 90 23 970 90 23 820 90 23 919
1) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
2) For O2, A1 is the lower alarm threshold for displaying oxygen deficiency.
3) Not applicable for Pac 6000.
4) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage is 0 ... 35 °C (32 ... 95 °F).
10.3 Technical data of the sensors and the measuring device settings for Pac 8000
NH3 PH3 HCN NO NO2-LC CO2
Measuring range 0 ... 300 ppm 0 ... 20 ppm 0 ... 50 ppm 0 ... 200 ppm 0 ... 50 ppm 0 ... 5 Vol%
Calibration gas concentration 50 ppm in N2 0.5 ppm in N2 10 ppm in N2 50 ppm in N2 5 ppm in N2 2.5 Vol% in air
Temperature range, operation -30 ... 50 °C -20 ... 50 °C -20 ... 50 °C -40 ... 50 °C -30 ... 50 °C -20 ... 40 °C
-22 ... 122 °F -4 ... 122 °F -4 ... 122 °F -40 ... 122 °F -22 ... 122 °F -4 ... 104 °F
Alarm threshold A1 1) 50 ppm 0.1 ppm 10 ppm 25 ppm 0.5 ppm 0.5 Vol%
Acknowledgeable      
Latching - - - - - -
Alarm threshold A2 1) 100 ppm 0.2 ppm 20 ppm 50 ppm 1 ppm 1 Vol%
Acknowledgeable - - - - - -
Latching      
TWA threshold value 1) 20 ppm 0.1 ppm 1.9 ppm 25 ppm 0.5 ppm 0.5 Vol%
STEL threshold value 1) 40 ppm 0.1 ppm 3.8 ppm 50 ppm 1 ppm 2 Vol%
Mean value period 15 min 15 min 15 min 15 min 15 min 15 min
Measurement accuracy
Zero point:  ±4 ppm  ±0.02 ppm  ±0.5 ppm  ±0.3 ppm  ±0.02 ppm  ±0.3 Vol%
Sensitivity: [% of the measured value]  ±3  ±2  ±5  ±3  ±3  ±20
Long-term drift (20 °C)
Zero point:  ±5 ppm/a  ±0.05 ppm/a  ±2 ppm/a  ±0.3 ppm/a  ±0.04 ppm/a  ±0.2 Vol%/a
Sensitivity:  ±2  ±2  ±5  ±2  ±2  ±15
[% of measured value/month]
Sensor item number 2) 68 10 888 68 10 886 68 10 887 68 11 545 68 12 600 68 10 889
Sensor data sheet item number 90 23 922 90 23 920 90 23 921 90 33 091 90 33 093 90 23 923
1) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
2) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage is 0 ... 35 °C (32 ... 95 °F).
Technical data en

19
20
Cl2 OV 1) OV-A 1) Ozone Phosgene
Measuring range 0 ... 20 ppm 0 ... 200 ppm 0 ... 200 ppm 0 ... 10 ppm 0 ... 10 ppm
Calibration gas concentration 5 ppm in N2 20 ppm in N2 20 ppm in N2 0.5 ... 9 ppm O3 3.8 ... 9 ppm
Temperature range, operation -30 ... 50 °C -20 ... 50 °C -20 ... 50 °C -20 ... 50 °C -20 ... 35 °C
en Technical data

-22 ... 122 °F -4 ... 122 °F -4 ... 122 °F -4 ... 122 °F -4 ... 99 °F
Alarm threshold A1 2) 0.5 ppm 10 ppm 10 ppm 0.1 ppm 0.1 ppm
Acknowledgeable     
Latching - - - - -
Alarm threshold A2 1) 1 ppm 20 ppm 20 ppm 0.2 ppm 0.2 ppm
Acknowledgeable - - - - -
Latching     
TWA threshold value 1) 0.5 ppm - - 0.1 ppm 0.1 ppm
STEL threshold value 1) 0.5 ppm - - 0.1 ppm 0.1 ppm
Mean value period 15 min - - 15 min 15 min
Measurement accuracy
Zero point:  ±0.05 ppm  ±3 ppm  ±5 ppm  ±0.02 ppm  ±0.01 ppm
Sensitivity: [% of the measured value]  ±2  ±5  ±20  ±3  ±10
Long-term drift (20 °C)
Zero point:  ±0.2 ppm/a  ±5 ppm/a  ±5 ppm/a  ±0.02 ppm/a  ±0.2 ppm/a
Sensitivity: [% of measured value/month]  ±2  ±2  ±3  ±2  ±2
Sensor item number 3) 68 10 890 68 11 530 68 11 535 68 11 540 68 12 005
Sensor data sheet item number 90 23 924 90 23 994 90 23 995 90 33 259 90 23 924
1) Only for ethylene oxide.
2) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
3) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage is 0 ... 35 °C (32 ... 95 °F).
10.3.1 Sensors for Pac 8500 DUAL
H2S LC / CO LC O2 / CO-LC CO H2-CP
Measuring range 0 ... 100 ppm H2S 0 ... 25 Vol% O2 0 ... 2000 ppm
0 ... 2000 ppm CO 0 ... 2000 ppm CO
Calibration gas concentration 5 ... 90 ppm H2S 12 ... 20 Vol% O2 1000 ppm H2
20 ... 450 ppm CO 20 ... 1800 ppm CO 20 ... 1800 ppm CO
Temperature range, operation -40 ... 50 °C -40 ... 50 °C -40 ... 50 °C
-40 ... 122 °F -40 ... 122 °F -40 ... 122 °F
Alarm threshold A1 1) 5 ppm H2S 19 O2 30 ppm
30 ppm CO 30 CO
Acknowledgeable  O2: - 
CO: 
Latching - O2:  -
CO: -
Alarm threshold A2 1) 10 ppm H2S 23 O2 60 ppm
60 ppm CO 60 CO
Acknowledgeable - O2: - -
CO: -
Latching  O2:  
CO: 
TWA threshold value 1) 5 ppm H2S - -
30 ppm CO 30 ppm CO 30 ppm CO
STEL threshold value 1) 10 ppm H2S - -
60 ppm CO 60 ppm CO 60 ppm CO
Mean value period 15 min 15 min 15 min
Measurement accuracy
Zero point: H2S: 0.4 ppm O2 ±0.4 Vol%  ±6 ppm
CO: 2 ppm CO:  ±2 ppm
Sensitivity: [% of the measured value] H2S:  ±5 O2 ±1  ±2
CO:  ±2 CO:  ±2
Long-term drift (20 °C)
Zero point: H2S:  ±0.2 ppm/a O2 ±0.5 Vol%/a  ±2 ppm/a
CO:  ±2 ppm/a CO:  ±2 ppm/a
Sensitivity: [% of measured value/year] H2S:  ±5 O2 ±1  ±1/month
CO:  ±3 CO:  ±3
Sensor item number 2) 68 13 280 68 13 275 68 11 950
Sensor data sheet item number 90 33 511 90 33 510 90 23 924
1) Observe special, customer-requested settings. Device parameters can be changed with the Dräger CC-Vision Basic PC software.
2) The sensors have a limited life span. Excessive storage periods compromise the operational life time of the sensors. The adequate temperature range for storage
is 0 ... 35 °C (32 ... 95 °F).
Technical data en

21
en Technical data

10.3.2 Cross-sensitivities

Cross-sensitivity factors 4) CO-LC H2S O2


Acetylene 2 No effect  -0.5
Ammonia No effect No effect No effect
Carbon dioxide No effect No effect  -0.04
Carbon monoxide No effect No effect  0.2
Chlorine  0.05  -0.2 No effect
Ethane No effect No effect  -0.2
Ethanol No effect No effect No effect
Ethylene No effect No effect  -1
Hydrogen  0.35 No effect  -1.5
Hydrogen chloride No effect No effect No effect
Hydrogen cyanide No effect No effect No effect
Hydrogen sulphide  0.03 No effect
Methane No effect No effect No effect
Nitrogen dioxide  0.05  -0.25 No effect
Nitric oxide  0.2  0.03 No effect
Propane No effect No effect No effect
Sulphur dioxide  0.04  0.1 No effect

22
Dräger Safety AG & Co. KGaA
Revalstraße 1
23560 Lübeck, Germany
Tel +49 451 882 0
Fax +49 451 882 20 80
www.draeger.com

90 33 742 - TH 4623.700
© Dräger Safety AG & Co. KGaA
Edition 01 - July 2017
Subject to alteration

You might also like