Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

ASTM Liquid-in-Glass Thermometers: Standard Specification For

Download as pdf or txt
Download as pdf or txt
You are on page 1of 64

Method 9501—Federal Test

Designation: E 1 – 05 Method Standard No. 791b

Standard Specification for


ASTM Liquid-in-Glass Thermometers1
This standard is issued under the fixed designation E 1; the number immediately following the designation indicates the year of original
adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A superscript
epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.

1. Scope responsibility of the user of this standard to establish appro-


1.1 This specification covers liquid-in-glass thermometers priate safety and health practices and determine the applica-
graduated in degrees Celsius or degrees Fahrenheit that are bility of regulatory limitations prior to use.
frequently identified and used in methods under the jurisdiction 2. Referenced Documents
of the various technical committees within ASTM. The various
thermometers specified are listed in Table 1. The inclusion of 2.1 ASTM Standards: 3

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
an IP number in Table 1 indicates, where appearing, that the E 77 Test Method for Inspection and Verification of Ther-
thermometer specification has been jointly agreed upon by the mometers
British Institute of Petroleum (IP) and ASTM. E 344 Terminology Relating to Thermometry and Hydrom-
1.2 This specification also covers adjustable-range etry
enclosed-scale thermometers, graduated in degrees Celsius, E 563 Practice for Preparation and Use of an Ice-Point Bath
which are used in ASTM methods. as a Reference Temperature
1.3 The enclosed-scale thermometers are commonly called E 2251 Specification for Liquid-in-Glass ASTM Thermom-
Beckmann thermometers. They are suitable for measuring eters with Low-Hazard Precision Liquids
small temperature differences not exceeding 6 °C within a 3. Terminology
larger range of temperature. The thermometers are unsuitable
for measuring Celsius- or kelvin-scale temperatures unless they 3.1 Definitions—The definitions given in Terminology
have been compared with standard instruments immediately E 344 apply.
before use. 3.2 Definitions of Terms Specific to This Standard:
1.4 An alphabetic list of the ASTM Thermometers included 3.2.1 adjusting device, n—a section of the instrument used
in this standard is given in Table 2. to adjust the amount of mercury in the bulb and main capillary
to that needed for the intended temperature interval.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
1.5 A list of ASTM Thermometers is given in Table 3 to
facilitate selection according to temperature range, immersion, 3.2.2 bulb length, n—the distance from the bottom of the
and scale-error requirements. bulb to the junction of the bulb and the stem tubing.
3.2.3 contraction chamber, n—an enlargement of the cap-
NOTE 1—For a listing of thermometers recommended for general illary, that will appear below the main scale or between the
laboratory use, the Scientific Apparatus Makers Association Specifications main scale and the auxiliary scale, which serves to reduce its
for General Purpose Glass Laboratory Thermometers may be consulted.2
NOTE 2—It has been found by experience that these ASTM Thermom-
length or to prevent contraction of the liquid column into the
eters, although developed in general for specific tests, may also be found bulb.
suitable for other applications, thus precluding the need for new thermom- 3.2.4 diameter, n—the largest outside dimension of the
eter specifications differing in only minor features. However, it is glass as measured with a ring gage.

Copyright by ASTM Int'l (all rights reserved);


suggested that technical committees contact Subcommittee E20.05 before 3.2.5 expansion chamber, n—an enlargement at the top of
choosing a currently specified thermometer for a new method to be sure the capillary to provide protection against breakage caused by
the thermometer will be suitable for the intended application.
excessive gas pressure.
1.6 This standard does not purport to address all of the 3.2.6 interval error, n—the deviation of the nominal value
safety concerns, if any, associated with its use. It is the of a temperature interval from its true value; either for the total
range (total interval) or for a part of the range (partial interval).
3.2.7 saddle, n—the bottom support of the enclosed scale.
1
This specification is under the jurisdiction of ASTM Committee E20 on
Temperature Measurement and is the direct responsibility of Subcommittee E20.05
on Liquid-in-Glass Thermometers and Hydrometers.
3
Current edition approved Nov. 1, 2005. Published December 2005. Originally For referenced ASTM standards, visit the ASTM website, www.astm.org, or
approved in 1939. Last previous edition approved in 2003 as E 1 – 03a. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
2
Available from SAMA Group of Assocs., 225 Reinekers, Ste. 625, Alexandria, Standards volume information, refer to the standard’s Document Summary page on
VA 23314. the ASTM website.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

Copyright ASTM International 1


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
3.2.8 setting temperature, n—the temperature that yields a Thermometer 99F
reading of zero on the main scale for a given adjustment of the
6.2.3 Ring Top (optional only)—Thermometers 11C and
amount of mercury in the bulb and main capillary.
11F.
3.2.9 thermometric liquid, n—the liquid in a liquid-in-glass
thermometer that indicates the value of temperature. 7. Bulb
3.2.10 top of the thermometer, n—the top of the finished
instrument. 7.1 The bulb shall be made of glass having a viscosity of at
3.2.11 total length, n—the distance from the bottom of the least 1014.6 poises at 490 °C (914 °F) and at least 1013.4 poises
bulb to the top of the finished thermometer, including any at 520 °C (968 °F).
special finish at the top. 7.2 Thermometers made with bulb glasses not meeting the
3.2.12 Other descriptions of terms shall be in accordance minimum properties in 7.1 shall not be subjected to tempera-
with the Terminology section of Test Method E 77. tures above 405 °C (760 °F) or be continuously exposed to
temperatures above 370 °C (700 °F).
Part A—Solid-Stem Thermometers
8. Capillary Clearances
4. Specifications
8.1 The following distances between graduations and the
4.1 The individual thermometers shall conform to the de- bulb, and between graduations and enlargements in the capil-
tailed specifications given in Table 1 and to the general lary, shall be minimum limits for thermometers in this speci-
requirements specified in Sections 5-15.

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
fication.
4.2 Thermometers manufactured to previous revisions of
this specification shall retain the same ASTM status as those NOTE 3—In order for a thermometer to be usable over its entire
meeting current specifications. graduated range, graduation marks should not be placed too close to any
enlargement in the capillary. Insufficient immersion of the thermometric
4.3 The encapsulation (jacketing) of the glass of liquid-in- liquid in the main bulb or capillary enlargement, graduation marks placed
glass thermometers with polyfluorinated hydrocarbons will over parts of the capillary that have been changed by manufacturing
change their performance and physical characteristics, includ- operations, or graduations so close to the top of the thermometer that
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

ing, but not limited to, response time, accuracy, and physical excessive gas pressure results when the thermometric liquid is raised to
dimensions. Therefore, under no circumstances should an this level, may lead to appreciable errors.
encapsulated or otherwise modified ASTM thermometer be 8.1.1 A 13-mm length of unchanged capillary between the
used in performing tests that specify the use of an ASTM bulb and the immersion line or lowest graduation, if the
thermometer. graduation is not above 100 °C (212 °F); a 30-mm length if the
5. Type graduation is above 100 °C (212 °F).
8.1.2 A 5-mm length of unchanged capillary between an
5.1 The thermometers, as specified in Table 1, shall be filled enlargement and the graduation next below, except at the top of
with one of the following liquids: the thermometer.
5.1.1 Mercury, 8.1.3 A 10-mm length of unchanged capillary between an
5.1.2 Mercury thallium eutectic alloy, or enlargement, other than the bulb, and the immersion line or the
5.1.3 Toluene or other suitable liquid colored with a perma- graduation next above, if the graduation is not above 100 °C
nent red dye. (212 °F); a 30-mm length if the graduation is above 100 °C
5.2 The filling above the liquid shall be nitrogen or other (212 °F).
suitable inert gas. 8.1.4 A 10-mm length of unchanged capillary above the
6. Stem highest graduation, if there is an expansion chamber at the top
of the thermometer; a 30-mm length if there is no expansion
6.1 Stem—The stem shall be made of suitable thermometer
chamber. For the purposes of this requirement, “an expansion
tubing and shall have a plain front and enamel back, unless

Copyright by ASTM Int'l (all rights reserved);


chamber” is interpreted as an enlargement at the top end of the
otherwise specified in Table 1.
capillary bore which shall have a capacity equivalent to not less
6.2 Top Finish— The top of all thermometers specified in
than 20 mm of unchanged capillary.
Table 1 shall have a plain rounded finish, except the following
8.2 It is possible to manufacture thermometers that comply
which shall have the top finish indicated below (unless
with the specifications given in Table 1, but do not meet the
indicated as optional):
requirements for capillary clearances given above. In any case,
6.2.1 Glass Button Finish:
the distances given in this section shall be the governing factor.
Thermometers 23C, 24C, and 25C
Under no circumstances shall the scales on thermometers be
6.2.2 Special Finish: placed closer than these minimum distances.
6.2.2.1 Suitable for assembly in a standard 304.8-mm (12-
in.) non-sparking metal armor with open face; in a cup case 9. Graduations and Inscriptions
assembly; or in a flushing case assembly: 9.1 All graduation lines, immersion lines, figures, and
Thermometers 58C, 58F, 59C, 59F, 60C, 60F, 97C, 97F, 98C, 98F, letters shall be clearly defined, suitably colored, and perma-
130C, and 130F
nent. The width and the sharpness of the graduation lines shall
6.2.2.2 Suitable for assembly in a 12-in. non-sparking metal be in accordance with 9.2. The middle of the graduation line
armor with open face: shall be determinable.

Copyright ASTM International 2


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
9.1.1 A suitably etched thermometer with the etched lines 23C, 24C, and 25C
and figures filled with a pigment shall be considered perma-
9.4.4 Precision Thermometers:
nently marked provided it passes the test for permanency of
65F, 66F, 67C, 67F, and 68C
pigment in Test Method E 77.
9.1.2 A thermometer marked by other means shall also be 9.4.5 Tank Thermometer:
considered permanently marked, provided it passes the test for 97F
permanency of pigment in Test Method E 77.
9.4.6 Solidification Point Thermometers:
9.2 Graduation Lines—All graduation lines shall be
100C and 101C
straight, of uniform width, and perpendicular to the axis of the
thermometer. The width of the graduation lines shall be as 9.4.7 Reid Vapor Pressure:
follows: 18C and 18F
9.2.1 Group 1—Maximum line width 0.10 mm; for ther-
9.4.8 Oxidation Stability:
mometers that may read to fractions of a division, often with
22C and 22F
magnifying aids:
Thermometers 14C, 14F, 26C, 28C, 28F, 29C, 29F, 30F, 33C, 9.5 Scale Below Zero—When a scale extends both above
33F, 34C, 34F, 35C, 35F, 44C, 44F, 45C, 45F, 46C, 46F, 47C, 47F, and below 0 °C or 0 °F, the two parts of the scale shall be
48C, 48F, 50F, 51F, 52C, 56C, 56F, 62C, 62F, 63C, 63F, 64C, 64F,
65C, 65F, 66C, 66F, 67C, 67F, 68C, 68F, 69C, 69F, 70C, 70F, 72C,
differentiated by some means. Examples of suitable means are:
72F, 73C, 73F, 74C, 74F, 89C, 90C, 91C, 92C, 93C, 94C, 95C, 96C, 9.5.1 Different pigment colors for the two parts of the scale,

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
100C, 101C, 110C, 110F, 111C, 112C, 113C, 113F, 116C, 117C,
118C, 118F, 119C, 119F, 120C, 121C, 126C, 126F, 127C, 128C,
9.5.2 Different style of numerical characters for the two
128F, 129C, 129F, 132C, 133C, and 137C. parts of the scale, and
9.5.3 Use of minus signs before appropriate numbers below
9.2.2 Group 2—Maximum line width 0.15 mm; for ther-
0 °C or 0 °F.
mometers that may be read to the nearest half division or where
the congestion of scale dictates the use of a scale with moderate
fineness: 10. Special Inscription
Thermometers 1C, 1F, 2C, 2F, 3C, 3F, 5C, 5F, 6C, 6F, 7C, 7F, 8C, 10.1 The special inscription specified in Table 1 shall be
8F, 9C, 9F, 10C, 10F, 11C, 11F, 12C, 12F, 13C, 15C, 15F, 16C, 16F, marked on the thermometer in capital letters and Arabic
17C, 17F, 18C, 18F, 19C, 19F, 20C, 20F, 21C, 21F, 22C, 22F, 23C,
24C, 25C, 36C, 37C, 38C, 39C, 40C, 41C, 42C, 43C, 43F, 49C,
numbers without the use of periods. In addition to the special
54C, 54F, 61C, 61F, 71C, 71F, 82C, 82F, 83C, 83F, 84C, 84F, 85C, inscription prescribed in Table 1, each thermometer shall be
85F, 86C, 86F, 87C, 87F, 99C, 99F, 102C, 103C, 104C, 105C, permanently marked with a unique serial number and the
106C, 107C, 108F, 109F, 114C, 122C, 123C, 124C, 125C, 134C,
135C, 135F, 136C, and 136F. manufacturer’s tradename or mark.
10.2 Engraving Revision Date on ASTM Thermometers—
9.2.3 Group 3—Maximum line width 0.20 mm; for ther- Include year of current revision in ASTM designation (for
mometers with more open scales, usually read to the nearest example, ASTM 1C-99).
division, often times under adverse conditions where a bold
graduation is therefore desired: 11. Permanency of Pigment
Thermometers 27C, 57C, 57F, 58C, 58F, 59C, 59F, 60C, 60F,
75F, 76F, 77F, 78F, 79F, 80F, 81F, 88C, 88F, 97C, 97F, 98C, 98F, 11.1 The test for permanency of pigment shall be performed
130C, and 130F. on any convenient portion of the scale section of the thermom-
9.3 Immersion Line—On partial immersion thermometers eter. The pigment shall not chalk, burn out, or loosen as a result
an immersion line shall be permanently marked on the front of of this test (see Test Method E 77).
the thermometer at the distance above the bottom of the bulb as
specified in Table 1 within a tolerance of 60.5 mm, except for 12. Bulb Stability

Copyright by ASTM Int'l (all rights reserved);


Thermometers 82F to 87F, which shall have no immersion line. 12.1 The test for bulb stability shall be made for the
The immersion inscription shall be written in capital letters and following thermometers in the temperature range specified
abbreviated (for example, 76 mm immersion shall be written below for 24 h. The scale indications after the test shall be
76 MM IMM). within the maximum scale error specified in Table 1. Obser-
9.4 Terminal Numbers—The terminal number shall be in vations of a reference point before and after the test to give a
full when there are one or more numbered graduations between measure of the degree of bulb stability achieved in manufac-
it and the last full number, before the terminal number. This ture. The bulb is considered stable if the change in indications
rule need not necessarily be followed for: of the thermometer in the test is no more than 0.7 (7⁄10) of the
9.4.1 Saybolt Viscosity Thermometers: maximum scale error found in Table 1. Reference should be
17C, 17F, 19C, 19F, 20C, 20F, 21C, 21F, 77F, 78F, 79F, 80F, and made to Test Method E 77.
81F ASTM Test Temperature
Thermometer Number Range
9.4.2 Kinematic Viscosity Thermometers:
28F, 29F, 30F, 44F, 45F, 46F, 47F, 48F, 72F, 73F, 74F, 110F, 118F, 3C, 8C, 10C, 11C, 70C 360 to 370°C
126F, 128F, and 129F 3F, 8F, 10F, 11F, 70F 680 to 700°F
2C, 7C, 69C, 107C 280 to 290°C
9.4.3 Engler Viscosity Thermometers: 2F, 7F, 69F 540 to 560 °F
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 3


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
13. Scale Error 14. Case
13.1 Thermometers shall be verified and calibrated at the 14.1 Each thermometer shall be supplied in a suitable case
temperatures specified in Table 4. Partial immersion thermom- on which shall appear the following marking (except when a
eters shall be calibrated for the emergent stem temperatures transparent case is used): the letters “ASTM,” the thermometer
specified in Table 4. number (33C, 33F, etc.), and the temperature range.
13.1.1 At the time of purchase, the scale errors shall be
within the maximum scale error found in Table 1. The 15. Methods of Verification and Calibration
indications of many high temperature and fractionally gradu-
ated thermometers may change with time and continued use 15.1 Thermometers shall be verified and calibrated at the
due to minute changes in bulb volume. Periodic verification of specified immersion in accordance with Test Method E 77.
these thermometers, either over the entire scale or reverifica- 15.2 For partial immersion thermometers, careful consider-
tion at a reference temperature (ice point or steam point), in ation of emergent stem temperatures shall be observed.
accordance with procedures set forth in Test Method E 77, is 15.2.1 During the manufacture of partial immersion ther-
recommended. mometers, the manufacturer shall calibrate the thermometers
13.2 Due to the application requirements for range and so the indicated temperatures are within the maximum permis-
construction of the following thermometers, it is not practical sible errors found in Table 1 when the emergent stem tempera-
to include reference points such as the ice and steam points. tures found in Table 4 are applied to the readings.
13C, 14C, 14F, 17C, 17F, 18C, 18F, 19C, 19F, 20C, 20F, 21C, 21F, 23C, 24C,

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
26C, 27C, 38C, 49C, 50F, 51F, 56C, 56F, 76F, 77F, 78F, 79F, 80F, 81F, 83C, NOTE 4—To achieve the requirements in 15.2.1, the manufacturer may
83F, 84C, 84F, 87C, 87F, 91C, 92C, 93C, 96C, 98C, 98F, 100C, 101C, 102C, have to measure emergent stem temperatures above its bath, calculate
103C, 104C, 105C, 106C, 107C, 108F, 109F, 111C, 116C, 117C, 122C, 123C, correction factors, and offset its calibrations accordingly. See Test Method
and 124C E 77 for the procedure to correct for emergent stem temperatures.

Copyright by ASTM Int'l (all rights reserved);


--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 4


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
TABLE 1 Specification for ASTM Thermometers
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
All dimensions are in millimeters.
See Table 4 for Verification and Calibration Temperatures.

Copyright ASTM International


Provided by IHS under license with ASTM
No reproduction or networking permitted without license from IHS
Explanatory Notes:
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury
separations and under no circumstances should the thermometer be heated above the highest temperature reading.
B
Toluene or other suitable liquid colored red with a permanent dye shall be used as the actuating liquid.
C
Under certain test conditions, the bulb of the thermometer may be 28 °C (50 °F) above the temperature indicated by the thermometer, and at an indicated
temperature of 371 °C (700 °F) the temperature of the bulb is approaching a critical range in the glass. It is therefore not desirable to use this thermometer under
such conditions at indicated temperatures above 371 °C (700 °F) without checking the ice point.
D
Longest graduation lines at 155 °C, 160 °C, 162 °C, 164 °C, 165 °C, and 170 °C, with arrows at 162 °C and 164 °C.
E
The length of the enlargement, and the distance from the bottom of the enlargement to the bottom of the bulb shall be measured with the test gage shown in
Fig. 1.
F
Long, narrow shape.
G
The test temperature shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H

5
Long, narrow shape; mercury shall be in the chamber at 0 °C (32 °F).
I
The thermometer shall be made to be mounted in a brass ferrule consisting of a tubular bushing 8.0 mm in outside diameter with a flanged head approximately
12 mm in diameter so that the upper extremity of the 8.0 mm diameter is located 90 mm from the bottom of the bulb.
J
To be marked on the glass stem at least 90 mm from the bottom of the bulb.
E 1 – 05

K
Glass button finish, see 6.2.1.
L
Long, narrow shape; mercury shall be near bottom of the chamber at 0 °C.
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-
point reading shall be expressed to the nearest 0.01 °C or 0.02 °F and applied as explained in Test Method E 77, Section 13.
N
Thermometers made to these specifications conform also with the requirements for the titer test thermometer of the American Oil Chemists Society and the
Association of Official Agricultural Chemists, except for the special inscription.
O
Capillary clearances must conform to Section 8.
P
Mercury shall be near middle of chamber at 0 °C.

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Q
The stem may be either the plain front or lens front type. If the thermometer is of the lens front type, the cross section of the stem shall be such that it will pass
through an 8-mm ring gage but will not enter a 5-mm slot gage.
R
A suitable mercury-thallium alloy shall be used as the actuating liquid.
S
The expansion chamber shall be of the long narrow type 10 to 20 mm in length. The length of unchanged capillary between the nearest graduation mark and
the expansion chamber shall be not less than 10 mm.
T
Mercury shall be near the bottom of the chamber at 0 °C.
U

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


The length of unchanged capillary between the nearest graduation mark and contraction chamber shall be not less than 10 mm.
V
Change in correction over any 5 °F interval shall not exceed 0.10 °F.
W
Expansion chamber shall be of the long narrow type and there shall be not less than 10 mm of unchanged capillary between the base of the chamber and the
top graduation.
X
Mercury shall be in the chamber at 32 °F.
Y
Over any interval of 2 °C the change in correction shall not exceed 0.02 °C.
Z
Over any interval of 4 °F the change in correction shall not exceed 0.05 °F.
AA
Special finish, see 6.2.2.
BB
The bulb diameter shall not be more than 0.5 mm greater than the stem.
CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but will not enter a 5.0-mm
slot gage. A minor diameter of 4 mm is permissible provided that the major diameter is not less than 7 mm.
DD
Bulb bottom shall be essentially hemispherical.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
EE
Immersion line shall be omitted.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal
Celsius range.
GG

Copyright ASTM International


The immersion line shall be visible in the case opening after assembly. The immersion shall be measured from the bottom of the bulb rather than from the
bottom of the armor. See 6.2.2.
HH
The stem shall be either the round or lens-front type.
II

Provided by IHS under license with ASTM


Contraction chamber to be long narrow type.
JJ
Over any interval of 1 °C the change in correction shall not exceed 0.01 °C. The correction at the lowest temperature of the nominal range shall not change by
more than 0.02 °C immediately after the thermometer has been heated for 15 min at a temperature 30 °C higher, and allowed to cool naturally in air.
KK
The capillary bore shall be large enough in relation to the bulb to ensure that (without tapping) jumping of the meniscus does not exceed one half of the
smallest scale division, when the temperature is rising at a uniform rate not exceeding 0.05 °C/min.
LL
The thermometer is to be calibrated for 100-mm immersion for the main scale, the ice point is to be calibrated for total immersion.
MM

No reproduction or networking permitted without license from IHS


Bulb shape ellipsoidal (see Fig. 2).
NN
This thermometer may be furnished with an optional ring top. See 6.2.3. Addition of a ring top will increase the total length by an amount equal to the outside
diameter of the ring.
OO
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 7.0 mm ring gage.

ASTM No. 1C-99 1F-99FF 2C-99 2F-99FF 3C-99 3F-99FF


IP No. 62C 73C
Name Partial Immersion Partial Immersion Partial Immersion
Reference Fig. No. 3 3 3
Range −20 to + 150°C 0 to 302°F −5 to + 300°C 20 to 580°F −5 to + 400°CC 20 to 760°FC
For test at
A Immersion, mm 76 76 76
Graduations:

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Subdivisions 1°C 2°F 1°C 2°F 1°C 2°F
Long lines at each 5°C 10°F 5°C 10°F 5°C 10°F
Numbers at each 10°C 20°F 10°C 20°F 10°C 20°F
Scale error, max 0.5°C 1°F 1°C 2°F 1°C to 301°C 2°F to 574°F
1.5°C above 3°F above

6
301°C 574°F
Special inscription ASTM ASTM ASTM
1C-99 or 1F-99 2C-99 or 2F-99 3C-99 or 3F–99
E 1 – 05

76 MM IMM 76 MM IMM 76 MM IMM


Expansion chamber:
A A
Permit heating to 200°C 392°F
B Total length, mm 317 to 327 385 to 395 410 to 420
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.5
D Bulb length, mm 19 to 25 10 to 15 10 to 15
E Bulb OD, mm 5.0 to 6.0 5.0 to 6.0 5.0 to 6.0
Scale location:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Bottom of bulb to line at 0°C 32°F 0°C 32°F 0°C 32°F
F Distance, mm 111 to 118 100 to 110 100 to 110
G Length of graduated portion, mm 170 to 200O 225 to 265O 250 to 290O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


I Distance to bottom, min, mm
J Distance to top, max, mm
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations and under no circumstances should
the thermometer be heated above the highest temperature reading.
B
Under certain test conditions, the bulb of the thermometer may be 28°C (50°F) above the temperature indicated by the thermometer, and at an indicated temperature of 371°C (700°F) the temperature of the
bulb is approaching a critical range in the glass. It is therefore not desirable to use this thermometer under such conditions at indicated temperatures above 371°C (700°F) without checking the ice point.
O
Capillary clearances shall conform to Section 8.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
ASTM No. 5C-86 5F-86FF 6C-86 6F-86FF 7C-86 7F-86FF

No reproduction or networking permitted without license from IHS


IP No. 1C 2C 5C
Name Cloud and Pour Low Cloud and PourB Low Distillation
Reference Fig. No. 3 3 4
Range −38 to + 50°C −36 to + 120°F −80 to + 20°C −112 to + 70°F −2 to + 300°C 30 to 580°F
For test at
A Immersion, mm 108 76 total
Graduations:
Subdivisions 1°C 2°F 1°C 2°F 1°C 2°F
Long lines at each 5°C 10°F 5°C 10°F 5°C 10°F
Numbers at each 10°C 20°F 10°C 20°F 10°C 20°F
Scale error, max 0.5°C 1°F 1°C to − 33°C 2°F to − 28°F 0.5°C to 150°C 1°F to 300°F
2°C below − 33°C 4°F below − 28°F 1°C above 150°C 2°F above 300°F
Special inscription ASTM ASTM ASTM
5C-86 or 5F–86 6C-86 or 6F-86 7C-86 or 7F–86
108 MM IMM 76 MM IMM
Expansion chamber:
A
Permit heating to 100°C 212°F 60°C 140°F
B Total length, mm 225 to 235 225 to 235 380 to 390

7
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 7 to 10 7 to 10 10 to 15
E Bulb OD, mm ≤5.5 and ≥stem ≤5.0 and ≥stem ≤5.0 and ≥stem
Scale location:
E 1 – 05

Bottom of bulb to line at −38°C −36°F −70°C −94°F 0°C 32°F


F Distance, mm 120 to 130 100 to 120 100 to 110
O
G Length of graduated portion, mm 65 to 85 70 to 100O 225 to 255O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
A

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations and under no circumstances should
the thermometer be heated above the highest temperature reading.
B
Toluene or other suitable liquid colored red with a permanent dye shall be used as the actuating liquid.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 8C-86 8F-86FF 9C-86 9F-86FF 10C-86 10F-86FF
IP No. 6C 15C 16C
Name High Distillation Low-Pensky-Martens High-Pensky-Martens

No reproduction or networking permitted without license from IHS


Reference Fig. No. 4 5 5
Range −2 to + 400°CC 30 to 760°FC −5 to + 110°C 20 to 230°F 90 to 370°C 200 to 700°F
For test at
A Immersion, mm total 57 57
Graduations:
Subdivisions 1°C 2°F 0.5°C 1°F 2°C 5°F
Long lines at each 5°C 10°F 1°C and 5°C 5°F 10°C 25°F
Numbers at each 10°C 20°F 5°C 10°F 20°C 50°F
Scale error, max 1°C to 300°C 2°F to 570°F 0.5°C 1°F 1°C to 260°C 2.5°F to 500°F
1.5°C above 300°C 3°F above 570°F 2°C above 260°C 3.5°F above 500°F
Special inscription ASTM ASTM ASTM
8C-86 or 8F-86 9C-86 or 9F–86 10C-86 or 10F–86
57 MM IMM 57 MM IMM
Expansion chamber:
A A
Permit heating to 160°C 320°F
B Total length, mm 380 to 390 285 to 295 285 to 295
C Stem OD, mm 6.0 to 8.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 10 to 15 9 to 13 7 to 10

8
E Bulb OD, mm ≤5.0 and ≥stem ≤5.5 and ≥stem ≤4.5 and ≥stem
Scale location:
Bottom of bulb to line at 0°C 32°F 0°C 32°F 90°C 200°F
F Distance, mm 30 to 40 85 to 95 80 to 90
E 1 – 05

G Length of graduated portion, mm 290 to 330O 140 to 175O 145 to 180O


Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm
Stem enlargement

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm 7.5 to 8.5 7.5 to 8.5
L Length, mm 2.5 to 5.0E 2.5 to 5.0E
M Distance to bottom, mm 64 to 66 64 to 66
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations and under no circumstances should
the thermometer be heated above the highest temperature reading.
C

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Under certain test conditions, the bulb of the thermometer may be 28°C (50°F) above the temperature indicated by the thermometer, and at an indicated temperature of 371°C (700°F) the temperature of the
bulb is approaching a critical range in the glass. It is therefore not desirable to use this thermometer under such conditions at indicated temperatures above 371°C (700°F) without checking the ice point.
E
The length of the enlargement, and the distance from the bottom of the enlargement to the bottom of the bulb shall be measured with the test gage shown in Fig. 1.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
TABLE 1 Continued
FF
ASTM No. 11C-86 11F-86 12C-86 12F-86FF 13C-86

Provided by IHS under license with ASTM


IP No. 28C 64C 64F 47C
Name Cleveland Open Flash Density-Wide Range Loss on Heat
Reference Fig. No. 3 4 9
Range −6 to + 400°CC 20 to 760°FC −20 to + 102°C −5 to + 215°F 155 to 170°C
For test at

No reproduction or networking permitted without license from IHS


A Immersion, mm 25 total total
Graduations:
Subdivisions 2°C 5°F 0.2°C 0.5°F 0.5°C
Long lines at each 10°C 10°F 1°C 1°F 1°CD
Numbers at each 20°C 20°F 2°C 5°F 155°C, 160°C, 165°C,
170°C
Scale error, max 2°C to 260°C 5°F to 500°F 0.15°C 0.25°F 0.5°C
4°C above 260°C 7°F above 500°F
Special inscription ASTM ASTM ASTM
11C-86 or 11F–86 12C-86 or 12F–86 13C-86
25 MM IMM
Expansion chamber:
A
Permit heating to 150°C 300°F 200°C
B Total length, mm 305 to 315NN 415 to 425 150 to 160
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 5.5 to 7.0
D Bulb length, mm 7 to 10 15 to 20 10 to 15
E Bulb OD, mm ≤4.5 and ≥stem bulb size ≥stem size ≤5.0 and ≥stem
Scale location:

9
Bottom of bulb to line at 0°C 32°F −20°C −4°F 155°C
F Distance, mm 45 to 55 35 to 50 50 to 60
G Length of graduated portion, 210 to 240O 305 to 350O 40 to 60O
mm
E 1 – 05

Ice-point scale:
Range
H Bottom of bulb to ice-point,
mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 30F
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm
L Length, mm
M Distance to bottom, mm
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations, and under no
circumstances should the thermometer be heated above the highest temperature reading.
C

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Under certain test conditions, the bulb of the thermometer may be 28°C (50°F) above the temperature indicated by the thermometer, and at an indicated temperature of 371°C (700°F) the
temperature of the bulb is approaching a critical range in the glass. It is therefore not desirable to use this thermometer under such conditions at indicated temperatures above 371°C (700°F) without
checking the ice point.
D
Longest graduation lines at 155°C, 160°C, 162°C, 164°C, 165°C, and 170°C, with arrows at 162°C and 164°C.
F
Long, narrow shape.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.
NN
This thermometer may be furnished with an optional ring top. See 6.2.3. Addition of a ring top will increase the total length by an amount equal to the outside diameter of the ring.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
ASTM No. 14C-86 14F-86FF 15C-86 15F-86FF 16C-86 16F-86FF

No reproduction or networking permitted without license from IHS


IP No. 17C 60C 61C
Name Wax Melting Point Low Softening Point High Softening Point
Reference Fig. No. 7 4 4
Range 38 to 82°C 100 to 180°F −2 to + 80°C 30 to 180°F 30 to 200°C 85 to 392°F
For test at
A Immersion, mm 79 total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.2°C 0.5°F 0.5°C 1°F
Long lines at each 0.5°C 1°F 1°C 1°F 1°C 5°F
Numbers at each 1°C 2°F 2°C 5°F 5°C 10°F
Scale error, max 0.1°C 0.2°F 0.2°C 0.4°F 0.3°C 0.5°F
Special inscription ASTM ASTM ASTM
14C-86 or 14F–86 15C-86 or 15F–86 16C-86 or 16F–86
79 MM IMM
Expansion chamber:
Permit heating to 100°C 212°F 130°C 270°F 250°C 482°F
B Total length, mm 370 to 380 390 to 400 390 to 400

10
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 18 to 28 9 to 14 9 to 14
E Bulb OD, mm 5.0 to 6.0 4.5 to 5.5 4.5 to 5.5
Scale location:
E 1 – 05

Bottom of bulb to line at 40°C 104°F 0°C 32°F 30°C 86°F


F Distance, mm 115 to 125 75 to 90 75 to 90
O
G Length of graduated portion, mm 210 to 240O 245 to 285 245 to 280O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 41
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 17C-86 17F-86FF 18C-86 18F-86FF 19C-86 19F-86FF

No reproduction or networking permitted without license from IHS


IP No. 23C
Name Saybolt Viscosity Reid Vapor Pressure Saybolt Viscosity
Reference Fig. No. 8 8 8
Range 19 to 27°C 66 to 80°F 34 to 42°C 94 to 108°F 49 to 57°C 120 to 134°F
G G G
For test at 21.1 and 25°C 70 and 77°FG 37.8°C 100°FG 50 and 54.4°C 122 and 130°FG
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Long lines at each 0.5°C 1°F 0.5°C 1°F 0.5°C 1°F
Numbers at each 1°C, except 21 2°F 1°C 2°F 1°C, except 54 2°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
17C-86 or 17F-86 18C-86 or 18F-86 19C-86 or 19F-86
Expansion chamber:
Permit heating to 100°C 212°F 100°C 212°F 115°C 240°F
B Total length, mm 270 to 280 270 to 280 270 to 280

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 25 to 35 25 to 35 25 to 35

11
E Bulb OD, mm ≤5.0 and ≥stem ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at 19°C 66°F 34°C 94°F 49°C 120°F
E 1 – 05

F Distance, mm 135 to 150 130 to 150 135 to 150


G Length of graduated portion, mm 67 to 101O 60 to 90O 67 to 101O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60H 60H 60H

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 10.0 8.0 to 10.0 8.0 to 10.0
L Length, mm 4.0 to 7.0 4.0 to 7.0 4.0 to 7.0
M Distance to bottom, mm 112 to 116 112 to 116 112 to 116
G
The test temperatures shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H
Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 20C-86 20F-86FF 21C-86 21F-86FF 22C-86 22F-86FF

No reproduction or networking permitted without license from IHS


IP No. 24C
Name Saybolt Viscosity Saybolt Viscosity Oxidation Stability
Reference Fig. No. 8 8 8
Range 57 to 65°C 134 to 148°F 79 to 87°C 174 to 188°F 95 to 103°C 204 to 218°F
G G G
For test at 60°C 140°FG 82.2°C 180°FG 98.9 and 100°C 210°FG
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Long lines at each 0.5°C 1°F 0.5°C 1°F 0.5°C 1°F
Numbers at each 1°C 2°F 1°C, except 82 2°F 1°C 2°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
20C-86 or 20F-86 21C-86 or 21F–86 22C-86 or 22F–86
Expansion chamber:
Permit heating to 115°C 240°F 140°C 285°F 155°C 310°F
B Total length, mm 270 to 280 270 to 280 270 to 280
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 8.0
D Bulb length, mm 25 to 35 25 to 35 25 to 35

12
E Bulb OD, mm ≤5.0 and ≥stem ≤5.0 and ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at 57°C 134°F 79°C 174°F 95°C 204°F
F Distance, mm 135 to 150 135 to 150 135 to 150
E 1 – 05

G Length of graduated portion, mm 67 to 101O 67 to 101O 70 to 100O


Ice-point scale:
Range
H Bottom of bulb to ice-point,
mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60H 60H 60H

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 10.0 8.0 to 10.0 8.0 to 10.0
L Length, mm 4.0 to 7.0 4.0 to 7.0 4.0 to 7.0
M Distance to bottom, mm 112 to 116 112 to 116 112 to 116
G
The test temperatures shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 23C-86 24C-86 25C-86
IP No.

No reproduction or networking permitted without license from IHS


Name Engler ViscosityI Engler ViscosityI Engler ViscosityI
Reference Fig. No. 7K 7K 7K
Range 18 to 28°C 39 to 54°C 95 to 105°C
For test at 25°C 40 and 50°C 100°C
A Immersion, mm 90 90 90
Graduations:
Subdivisions 0.2°C 0.2°C 0.2°C
Long lines at each 1°C 1°C 1°C
Numbers at each 2°C full figures at 25 2°C full figures at 40 and 50 2°C full figures at 100
Scale error, max 0.1°C at 25°C 0.1°C at 40 and 50°C 0.1°C at 100°C
Special inscription ASTM ASTM ASTM
23C-86 24C-86 25C-86
90 MM IMMJ 90 MM IMMJ 90 MM IMMJ
Expansion chamber:
Permit heating to 100°C 105°C 155°C
B Total length, mm 207 to 217 232 to 242 207 to 217
C Stem OD, mm 5.5 to 6.5 5.5 to 6.5 5.5 to 6.5
D Bulb length, mm 13 to 19 13 to 19 13 to 19

13
E Bulb OD, mm 5.5 to 6.5 5.5 to 6.5 5.5 to 6.5
Scale location:
Bottom of bulb to line at 18°C 39°C 95°C
F Distance, mm 108 to 118 108 to 118 108 to 118
E 1 – 05

G Length of graduated portion, mm 42 to 69O 67 to 94O 42 to 69O


Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60I 60I 60I
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm
L Length, mm
M Distance to bottom, mm
I
The thermometer shall be made to be mounted in a brass ferrule consisting of a tubular bushing 8.0 mm in outside diameter with a flanged head approximately 12 mm in diameter so that the upper
extremity of the 8.0 mm diameter is located 90 mm from the bottom of the bulb.
J

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


To be marked on the glass stem at least 90 mm from the bottom of the bulb.
K
Glass button finish, see 6.2.1.
L
Long, narrow shape; mercury shall be near bottom of the chamber at 0°C.
O
Capillary clearances shall conform to Section 8.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 26C-86 27C-86 28C-86 28F-86FF
IP No. 31C

No reproduction or networking permitted without license from IHS


Name Stability Test of Soluble Nitrocellulose Turpentine Distillation Kinematic ViscosityM
Reference Fig. No. 9 7 6
Range 130 to 140°C 147 to 182°C 36.6 to 39.4°C 97.5 to 102.5°F
For test at 134.5°C 37.8°C 100°F
A Immersion, mm total 76 total

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Graduations:
Subdivisions 0.1°C 0.5°C 0.05°C 0.1°F
Long lines at each 0.5°C 1°C 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C and in full at 130, 135, 140 2°C from 148 1°C 1°F
Scale error, max 0.2°C 0.5°C 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
26C-86 27C-86 28C-86 or 28F-86
76 MM IMM
Expansion chamber:
Permit heating to 175°C 230°C 105°C 220°F
B Total length, mm 458 to 468 296 to 306 300 to 310
C Stem OD, mm 6.5 to 8.0 6.0 to 7.0 6.0 to 8.0
D Bulb length, mm 54 to 67 10 to 15 45 to 55

14
E Bulb OD, mm 6.0 to 7.0 4.0 to 5.5 ≥stem
Scale location:
Bottom of bulb to line at 134.5°C 147°C 36.6°C 97.5°F
E 1 – 05

F Distance, mm 320 to 340 100 to 115 145 to 165


G Length of graduated portion, mm 112 to 145O 131 to 166O 40 to 90O
Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 100
J Distance to top, max, mm 100I 40I 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
I
Long, narrow shape; mercury shall be near bottom of the chamber at 0°C.
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 29C-86 29F-86FF 30F-86FF 33C-86 33F-86FF

No reproduction or networking permitted without license from IHS


IP No. 34C 20C
Name Kinematic ViscosityM Kinematic ViscosityM Low Aniline Point
Reference Fig. No. 6 6 3
Range 52.6 to 55.4°C 127.5 to 132.5°F 207.5 to 212.5°F −38 to + 42°C −36.5 to + 107.5°F
For test at 54.4°C 130°F 210°F
A Immersion, mm total total 50
Graduations:
Subdivisions 0.05°C 0.1°F 0.1°F 0.2°C 0.5°F
Long lines at each 0.1 and 0.5°C 0.5 and 1°F 0.5 and 1°F 1°C 1°F
Numbers at each 1°C 1°F 1°F 2°C 5°F
Scale error, max 0.1°C 0.2°F 0.2°F 0.2°C 0.5°F
Special inscription ASTM ASTM ASTM
29C-86 or 29F–86 30F-86 33C-86 or 33F-86
50 MM IMM
Expansion chamber:
Permit heating to 105°C 220°F 266°F 100°C 212°F
B Total length, mm 300 to 310 300 to 310 415 to 425
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 7.5

15
D Bulb length, mm 45 to 55 45 to 55 10 to 20
E Bulb OD, mm ≥stem ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at 52.6°C 127.5°F 207.5°F −35°C −31°F
E 1 – 05

F Distance, mm 145 to 165 145 to 165 100 to 125


G Length of graduated portion, mm 40 to 90O 40 to 90O 240 to 280O
Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 100 100
J Distance to top, max, mm 125 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 34C-86 34F-86FF 35C-86 35F-86FF 36C-86
IP No. 21C 59C
Name Medium Aniline Point High Aniline Point Titer TestN

No reproduction or networking permitted without license from IHS


Reference Fig. No. 3 7 3
Range 25 to 105°C 77 to 221°F 90 to 170°C 194 to 338°F −2 to + 68°C
For test at
A Immersion, mm 50 50 45
Graduations:
Subdivisions 0.2°C 0.5°F 0.2°C 0.5°F 0.2°C
Long lines at each 1°C 1°F 1°C 1°F 1°C
Numbers at each 2°C 5°F 2°C 5°F 2°C
Scale error, max 0.2°C 0.5°F 0.4°C 1.0°F 0.2°C
Special inscription ASTM ASTM ASTM
34C-86 or 34F-86 35C-86 or 35F–86 36C-86
50 MM IMM 50 MM IMM 45 MM IMM
Expansion chamber:
Permit heating to 150°C 302°F 220°C 428°F 85°CO
B Total length, mm 415 to 425 415 to 425 400 to 410
C Stem OD, mm 6.0 to 7.5 6.0 to 7.5 6.0 to 7.0Q
D Bulb length, mm 10 to 20 10 to 20 15 to 25
E Bulb OD, mm ≤5.0 and ≥stem 5.0 to ≥stem ≤5.5 and ≥stem

16
Scale location:
Bottom of bulb to line at 25°C 77°F 90°C 194°F −2°C
F Distance, mm 100 to 115 100 to 115 50 to 60
E 1 – 05

G Length of graduated portion, mm 240 to 280O 240 to 280O 290 to 320O


Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 35P
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm
L Length, mm
M Distance to bottom, mm
N
Thermometers made to these specifications conform also with the requirements for the titer test thermometer of the American Oil Chemists Society and the Association of Official Agricultural Chemists, except
for the special inscription.
O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


P
Mercury shall be near middle of chamber at 0°C.
Q
The stem may be either the plain front or lens front type. If the thermometer is of the lens front type, the cross section of the stem shall be such that it will pass through an 8-mm ring gage but will not enter a
5-mm slot gage.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 37C-86 38C-86 39C-86

No reproduction or networking permitted without license from IHS


IP No. 77C 78C 79C
Name Solvents Distillation Solvents Distillation Solvents Distillation
Reference Fig. No. 3 3 7
Range −2 to + 52°C 24 to 78°C 48 to 102°C
For test at
A Immersion, mm 100 100 100
Graduations:
Subdivisions 0.2°C 0.2°C 0.2°C
Long lines at each 1°C 1°C 1°C
Numbers at each 2°C 2°C 2°C
Scale error, max 0.2°C 0.2°C 0.2°C
Special inscription ASTM ASTM ASTM
37C-86 38C-86 39C-86
100 MM IMM 100 MM IMM 100 MM IMM
Expansion chamber:
Permit heating to 80°C 105°C 130°C
B Total length, mm 390 to 400 390 to 400 390 to 400

17
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 15 to 20 15 to 20 15 to 20
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
E 1 – 05

Bottom of bulb to line at −2°C 24°C 48°C


F Distance, mm 125 to 145 125 to 145 125 to 145
G Length of graduated portion, mm 190 to 235O 190 to 235O 190 to 235O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 35T
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Capillary clearances shall conform to Section 8.
T
Mercury shall be near the bottom of the chamber at 0°C.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 40C-86 41C-86 42C-86

No reproduction or networking permitted without license from IHS


IP No. 80C 81C 82C
Name Solvents Distillation Solvents Distillation Solvents Distillation
Reference Fig. No. 7 7 7
Range 72 to 126°C 98 to 152°C 95 to 255°C
For test at
A Immersion, mm 100 100 100
Graduations:
Subdivisions 0.2°C 0.2°C 0.5°C
Long lines at each 1°C 1°C 1°C
Numbers at each 2°C 2°C 5°C
Scale error, max 0.2°C 0.3°C 1°C
Special inscription ASTM ASTM ASTM
40C-86 41C-86 42C-86
100 MM IMM 100 MM IMM 100 MM IMM
Expansion chamber:
Permit heating to 150°C 180°C 280°C

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
B Total length, mm 390 to 400 390 to 400 390 to 400
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0

18
D Bulb length, mm 15 to 20 15 to 20 15 to 20
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
E 1 – 05

Bottom of bulb to line at


F Distance, mm 72°C 98°C 95°C
G Length of graduated portion, mm 125 to 145 125 to 145 125 to 145
190 to 235O 190 to 235O 190 to 235O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


I Distance to bottom, min, mm
J Distance to top, max, mm 35T 35T 35T
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
T
Mercury shall be near the bottom of the chamber at 0°C.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 43C-86 43F-86FF 44C-86 44F-86FF 45C-86 45F-86FF
IP No. 65C 29C 30C
Name Kinematic ViscosityM,R Kinematic ViscosityM Kinematic ViscosityM

No reproduction or networking permitted without license from IHS


Reference Fig. No. 10 6 6
Range −51.6 to − 34°C −61 to − 29°F 18.6 to 21.4°C 66.5 to 71.5°F 23.6 to 26.4°C 74.5 to 79.5°F
For test at 20°C 68 and 70°F 25°C 77°F
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.05°C 0.1°F 0.05°C 0.1°F
Long lines at each 0.5 and 1°C 1°F 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C starting at −51°C 2°F starting at − 60°F 1°C 1°F 1°C 1°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
43C-86 or 43F-86 44C-86 or 44F-86 45C-86 or 45F–86
MERC-THAL
Expansion chamber:
Permit heating to 105°CS 220°FS 105°C 220°F 105°C 220°F
B Total length, mm 410 to 425 300 to 310 300 to 310
C Stem OD, mm 7.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 30 to 40 45 to 55 45 to 55
E Bulb OD, mm 6.0 to 7.0 ≥stem ≥stem

19
Scale location:
Bottom of bulb to line at −51.6°C −61°F 18.6°C 66.5°F 23.6°C 74.5°F
F Distance, mm 60 to 90 145 to 165 145 to 165
G Length of graduated portion, mm 140 to 225O 40 to 90O 40 to 90O
E 1 – 05

Ice-Point scale:
Range −0.6 to + 0.6°CO 31 to 33°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 290U 100 100
J Distance to top, max, mm 310U 125 125
Stem enlargement:
K OD, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest
0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
R

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


A suitable mercury-thallium alloy shall be used as the actuating liquid.
S
The expansion chamber shall be of the long narrow type 10 to 20 mm in length. The length of unchanged capillary between the nearest graduation mark and the expansion chamber shall be not less than 10
mm.
U
The length of unchanged capillary between the nearest graduation mark and contraction chamber shall be not less than 10 mm.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 46C-86 46F-86FF 47C-86 47F-86FF 48C-86 48F-86FF

No reproduction or networking permitted without license from IHS


IP No. 66C 35C 90C
Name Kinematic ViscosityM Kinematic ViscosityM Kinematic ViscosityM
Reference Fig. No. 6 6 6
Range 48.6 to 51.4°C 119.5 to 124.5°F 58.6 to 61.4°C 137.5 to 142.5°F 80.6 to 83.4°C 177.5 to 182.5°F
For test at 50°C 122°F 60°C 140°F 82.2°C 180°F
A Immersion, mm total total total
Graduations:
Subdivisions 0.05°C 0.1°F 0.05°C 0.1°F 0.05°C 0.1°F
Long lines at each 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C 1°F 1°C 1°F 1°C 1°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
46C-86 or 46F–86 47C-86 or 47F–86 48C-86 or 48F–86
Expansion chamber:
Permit heating to 105°C 220°F 105°C 220°F 105°C 220°F
B Total length, mm 300 to 310 300 to 310 300 to 310
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 45 to 55 45 to 55 45 to 55

20
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at 48.6°C 119.5°F 58.6°C 137.5°F 80.6°C 177.5°F
E 1 – 05

F Distance, mm 145 to 165 145 to 165 145 to 165


G Length of graduated portion, mm 40 to 90O 40 to 90O 40 to 90O
Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 100 100 100
J Distance to top, max, mm 125 125 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest
0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 49C-86 50F-86FF 51F-86FF
IP No.

No reproduction or networking permitted without license from IHS


Name Stormer Viscosity Gas Calorimeter Inlet Gas Calorimeter Outlet
Reference Fig. No. 7 9 9
Range 20 to 70°C 54 to 101°F 69 to 116°F
For test at
A Immersion, mm 65 total total
Graduations:
Subdivisions 0.2°C 0.1°F 0.1°F
Long lines at each 1°C 0.5°F 0.5°F
Numbers at each 2°C 1°F 1°F
Scale error, max 0.2°C 0.2°FV 0.2°FV
Special inscription ASTM ASTM ASTM
49C-86 50F-86 51F-86
65 MM IMM
Expansion chamber:
Permit heating to 100°C 150°F 212°F
B Total length, mm 300 to 310 463 to 473 463 to 473
C Stem OD, mm 5.5 to 6.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 15 to 30 25 to 30 25 to 30

21
E Bulb OD, mm ≤5.0 and ≥stem 6.5 to 7.0 6.5 to 7.0
Scale location:
Bottom of bulb to line at 20°C 54°F 69°F
E 1 – 05

F Distance, mm 70 to 80 100 to 120 100 to 120


G Length of graduated portion, mm 165 to 200O 292 to 338O 292 to 338O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 50P 43X 43X

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.
P
Mercury shall be near middle of chamber at 0°C.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


V
Change in correction over any 5°F interval shall not exceed 0.10°F.
X
Mercury shall be in the chamber at 32°F.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 52C-86 54C-86 54F-86FF 56C-86 56F-86FF
IP No. 18C
Name Butadiene Boiling Point Range Congealing Point Bomb Calorimeter

No reproduction or networking permitted without license from IHS


Reference Fig. No. 4 4 9
Range −10 to + 5°C 20 to 100.6°C 68 to 213°F 19 to 35°C 66 to 95°F
For test at
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°C 0.5°F 0.02°C 0.05°F
Long lines at each 0.5°C 1°C 1°F 0.1°C 0.1 and
0.5°F
Numbers at each 1°C 2°C 5°F 0.2°C 1°F
Scale error, max 0.1°C 0.2°C 0.5°F 0.10°CY 0.20°FZ
Special inscription ASTM ASTM ASTM
52C-86 54C–86 or 54F-86 56C–86 or 56F-86
Expansion chamber:
Permit heating to 100°CW 110°C 230°F 66°C 150°F
B Total length, mm 157 to 167 305 to 315 570 to 600
C Stem OD, mm 6.0 to 6.5 6.0 to 8.0 7.0 to 8.0
D Bulb length, mm 9 to 13 10 to 12 35 to 55
E Bulb OD, mm 5.5 to ≥stem 4.5 to 6.0MM 7.0 to 8.0BB

22
Scale location:
Bottom of bulb to line at −10°C 20°C 68°F 19°C 66°F
F Distance, mm 28 to 36 60 to 70 165 to 187
E 1 – 05

G Length of graduated portion, mm 70 to 100O 170 to 215O 323 to 385O


Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 76
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.
W
Expansion chamber shall be of the long narrow type and there shall be not less than 10 mm of unchanged capillary between the base of the chamber and the top graduation.
Y
Over any interval of 2°C the change in correction shall not exceed 0.02°C.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Z
Over any interval of 4°F the change in correction shall not exceed 0.05°F.
BB
The bulb diameter shall not be more than 0.5 mm greater than the stem.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.
MM
Bulb shape ellipsoidal (see Fig. 2).

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 57C-86 57F-86FF 58C-86 58F-86FF 59C-86 59F-86FF
IP No.

No reproduction or networking permitted without license from IHS


Name Tag Closed Tester Low Range Tank Tank
Reference Fig. No. 5 4AA 4AA
Range −20 to + 50°C −4 to + 122°F −34 to + 49°C −30 to + 120°F −18 to + 82°C 0 to 180°F
For test at
A Immersion, mm 57 total total
Graduations:
Subdivisions 0.5°C 1°F 0.5°C 1°F 0.5°C 1°F
Long lines at each 1°C 5°F 1°C 5°F 1°C 5°F
Numbers at each 5°C 10°F 5°C 10°F 5°C 10°F
Scale error, max 0.5°C 1°F 0.3°C 0.5°F 0.3°C 0.5°F
Special inscription ASTM ASTM ASTM
57C-86 or 57F-86 58C-86 or 58F–86 59C-86 or 59F-86

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
57 MM IMM
Expansion chamber:
Permit heating to 100°C 212°F 100°C 212°F 100°C 212°F
B Total length, mm 282 to 292 300 to 305 300 to 305
CC CC
C Stem OD, mm 6.0 to 7.0
D Bulb length, mm 9 to 13 15 to 25 15 to 25

23
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at −20°C −4°F −34°C −30°F −18°C 0°F
F Distance, mm 75 to 90 75 to 90 65 to 80
E 1 – 05

G Length of graduated portion, mm 143 to 177O 165 to 205O 175 to 210O


Ice-point scale
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm 7.5 to 8.5
L Length, mm 2.5 to 5.0E
M Distance to bottom, mm 64 to 66
E
The length of the enlargement, and the distance from the bottom of the enlargement to the bottom of the bulb shall be measured with the test gage shown in Fig. 1.
O
Capillary clearances shall conform to Section 8.
AA

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Special finish, see 6.2.2.
CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but will not enter a 5.0-mm slot gage. A minor diameter of 4 mm is
permissible provided that the major diameter is not less than 7 mm.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 60C-86 60F-86FF 61C-86 61F-86FF 62C-86 62F-86FF
IP No. 63C
Name Tank Petrolatum Melting Point Precision

No reproduction or networking permitted without license from IHS


Reference Fig. No. 4AA 7 4
Range 77 to 260°C 170 to 500°F 32 to 127°C 90 to 260°F −38 to + 2°C −36 to + 35°F
For test at
A Immersion, mm total 79 total
Graduations:
Subdivisions 1°C 2°F 0.2°C 0.5°F 0.1°C 0.2°F
Long lines at each 5°C 10°F 1°C 1°F 0.5°C 1°F
Numbers at each 10°C 20°F 2°C 5°F 1°C 2°F
Scale error, max 0.5°C 1°F 0.2°C 0.5°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
60C-86 or 60F-86 61C-86 or 61F-86 62C-86 or 62F-86
79 MM IMM
Expansion chamber:
A
Permit heating to 150°C 300°F 55°C 130°F
B Total length, mm 300 to 305 375 to 385 374 to 384
CC
C Stem OD, mm 6.0 to 8.0 7.0 to 8.0
D Bulb length, mm 15 to 25 18 to 28DD 25 to 35
E Bulb OD, mm ≥stem 5.0 to 6.0 6.0 to 7.0

24
Scale location:
Bottom of bulb to line at 77°C 170°F 32°C 90°F −35°C −31°F
F Distance, mm 105 to 120 105 to 115 77 to 98
G Length of graduated portion, mm 135 to 170O 200 to 240O 239 to 289O
E 1 – 05

Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 41
Stem enlargement:
K OD, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


L Length, mm
M Distance to bottom, mm
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations and under no circumstances should
the thermometer be heated above the highest temperature reading.
O
Capillary clearances shall conform to Section 8.
AA

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Special finish, see 6.2.2.
CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but will not enter a 5.0-mm slot gage. A minor diameter of 4 mm is
permissible provided that the major diameter is not less than 7 mm.
DD
Bulb bottom shall be essentially hemispherical.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
FF

No reproduction or networking permitted without license from IHS


ASTM No. 63C-86 63F-86 64C-86 64F-86FF 65C-86 65F-86FF
IP No.
Name Precision Precision Precision
Reference Fig. No. 4 6 6
Range −8 to + 32°C 18 to 89°F 25 to 55°C 77 to 131°F 50 to 80°C 122 to 176°F
For test at
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Long lines at each 0.5°C 1°F 0.5°C 1°F 0.5°C 1°F
Numbers at each 1°C 2°F 1°C 2°F 1°C 2°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
63C-86 or 63F–86 64C-86 or 64F-86 65C-86 or 65F–86
Expansion chamber:
Permit heating to 80°C 175°F 105°C 220°F 130°C 265°F
B Total length, mm 374 to 384 374 to 384 374 to 384
C Stem OD, mm 7.0 to 8.0 7.0 to 8.0 7.0 to 8.0

25
D Bulb length, mm 25 to 35 25 to 35 25 to 35
E Bulb OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
Scale location:
E 1 – 05

Bottom of bulb to line at −5°C 23°F 25°C 77°F 50°C 122°F


F Distance, mm 77 to 98 115 to 135 115 to 135
G Length of graduated portion, mm 239 to 289O 189 to 229O 189 to 229O
Ice-point scale:
Range −0.5 to + 0.5°CO 31 to 33°FO −0.5 to + 0.5°CO 31 to 33°FO
H Bottom of bulb to ice-point, mm 60 to 70 60 to 70
Contraction chamber:
I Distance to bottom, min, mm 80 80

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 100 100
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
FF

No reproduction or networking permitted without license from IHS


ASTM No. 66C-86 66F-86 67C-86 67F-86FF 68C-86 68F-86FF
IP
Name Precision Precision Precision
Reference Fig. No. 6 6 6
Range 75 to 105°C 167 to 221°F 95 to 155°C 203 to 311°F 145 to 205°C 293 to 401°F
For test at
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.2°F 0.2°C 0.5°F 0.2°C 0.5°F
Long lines at each 0.5°C 1°F 1°C 1°F 1°C 1°F
Numbers at each 1°C 2°F 2°C 5°F 2°C 5°F
Scale error, max 0.1°C 0.2°F 0.2°C 0.5°F 0.2°C 0.5°F
Special inscription ASTM ASTM ASTM
66C-86 or 66F-86 67C-86 or 67F-86 68C-86 or 68F-86
Expansion chamber:
Permit heating to 155°C 310°F 180°C 355°F 230°C 445°F
B Total length, mm 374 to 384 374 to 384 374 to 384
C Stem OD, mm 7.0 to 8.0 7.0 to 8.0 7.0 to 8.0

26
D Bulb length, mm 25 to 35 10 to 20 10 to 20
E Bulb OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
Scale location:
E 1 – 05

Bottom of bulb to line at 75°C 167°F 95°C 203°F 145°C 293°F


F Distance, mm 115 to 135 115 to 135 115 to 135
G Length of graduated portion, mm 189 to 229O 189 to 229O 189 to 229O
Ice-point scale:
Range −0.5 to + 0.5°CO 31 to 33°FO −1 to + 1°CO 30 to 34°FO −1 to + 1°CO 30 to 34°FO
H Bottom of bulb to ice-point, mm 60 to 70 45 to 55 45 to 55
Contraction chamber:
I Distance to bottom, min, mm 80 65 65

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 100 85 85
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 69C-86 69F-86FF 70C-86 70F-86FF 71C-86 71F-86FF
IP No. 72C

No reproduction or networking permitted without license from IHS


Name Precision Precision Oil in Wax
Reference Fig. No. 6 6 3
C C
Range 195 to 305°C 383 to 581°F 295 to 405°C 563 to 761°F −37 to + 21°C −35 to + 70°F
For test at
A Immersion, mm total total 76
Graduations:
Subdivisions 0.5°C 1°F 0.5°C 1°F 0.5°C 1°F
Long lines at each 1°C 5°F 1°C 5°F 1°C 5°F
Numbers at each 5°C 10°F 5°C 10°F 5°C 10°F
Scale error, max 0.5°C 1°F 0.5°C 1°F 0.2°C 0.5°F
Special inscription ASTM ASTM ASTM

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
69C-86 or 69F-86 70C-86 or 70F-86 71C-86 or 71F-86
76 MM IMM
Expansion chamber:
A
Permit heating to 330°C 625°F 105°C 220°F
B Total length, mm 374 to 384 374 to 384 350 to 360
C Stem OD, mm 6.5 to 8.0 6.5 to 8.0 6.0 to 8.0
D Bulb length, mm 10 to 27 10 to 27 15 to 20

27
E Bulb OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
Scale location:
Bottom of bulb to line at 195°C 383°F 295°C 563°F −37°C −35°F
F Distance, mm 125 to 145 125 to 145 170 to 185
E 1 – 05

G Length of graduated portion, mm 179 to 219O 179 to 219O 105 to 140O


Ice-point scale:
Range −2 to + 2°CO 27 to 37°FO −2 to + 2°CO 27 to 37°FO
H Bottom of bulb to ice-point, mm 52 to 62 52 to 62
Contraction chamber:
I Distance to bottom, min, mm 72 72
J Distance to top, max, mm 95 95
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm
L Length, mm
M Distance to bottom, mm
A
An expansion chamber is provided for relief of gas pressure to avoid distortion of the bulb at higher temperatures. It is not for the purpose of joining mercury separations and under no circumstances should
the thermometer be heated above the highest temperature reading.
C

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Under certain test conditions, the bulb of the thermometer may be 28°C (50°F) above the temperature indicated by the thermometer, and at an indicated temperature of 371°C (700°F) the temperature of the
bulb is approaching a critical range in the glass. It is therefore not desirable to use this thermometer under such conditions at indicated temperatures above 371°C (700°F) without checking the ice point.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 72C-86 72F-86FF 73C-86 73F-86FF 74C-86 74F-86FF
IP No. 67C 68C 69C

No reproduction or networking permitted without license from IHS


Name Kinematic ViscosityM Kinematic ViscosityM,R Kinematic ViscosityM,R
Reference Fig. No. 10 10 10
Range −19.4 to − 16.6°C −2.5 to + 2.5°F −41.4 to − 38.6°C −42.5 to − 37.5°F −55.4 to − 52.6°C −67.5 to − 62.5°F
For test at −17.8°C 0°F −40°C −40°F −53.9°C −65°F
A Immersion, mm total total total
Graduations: 0.05°C 0.1°F 0.05°C 0.1°F 0.05°C 0.1°F
Long lines at each 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C 1°F 1°C 1°F 1°C 1°F
Scale error, max 0.1°C 0.2°F 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
72C-86 or 72F-86 73C-86 or 73F-86 74C-86 or 74F-86
MERC-THAL MERC-THAL
Expansion chamber:
Permit heating to 105°C 220°F 105°C 220°F 105°C 220°F
B Total length, mm 300 to 310 300 to 310 300 to 310
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 45 to 55 45 to 55 45 to 55
E Bulb OD, mm ≥stem ≥stem ≥stem

28
Scale location:
Bottom of bulb to line at −19.4°C −2.5°F −41.4°C −42.5°F −55.4°C −67.5°F
F Distance, mm 80 to 110 80 to 110 80 to 110
E 1 – 05

G Length of graduated portion, mm 40 to 90O 40 to 90O 40 to 90O


Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.3 to + 0.3°CO 31.5 to 32.5°FO

H Bottom of bulb to ice-point, mm


Contraction chamber:
I Distance to bottom, min, mm 180 180 180
J Distance to top, max, mm 205 205 205

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
R
A suitable mercury-thallium alloy shall be used as the actuating liquid.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 75F-86FF 76F-86FF 77F-86FF
IP No.

No reproduction or networking permitted without license from IHS


Name Coolant (Antifreeze) Freezing Point Coolant (Antifreeze) Freezing PointR Saybolt Viscosity
Reference Fig. No. 3 3 8
Range −35 to + 35°F −65 to + 5°F 245 to 265°F
For test at 250°FG
A Immersion, mm 100 100 total
Graduations:
Subdivisions 0.5°F 0.5°F 0.5°F
Long lines at each 1°F 1°F 1°F
Numbers at each 5°F 5°F 5°F
Scale error, max 0.5°F 1°F 0.5°F
Special inscription ASTM ASTM ASTM
75F-86 76F-86 77F-86
100 MM IMM 100 MM IMM
MERC-THAL
Expansion chamber:
Permit heating to 140°F 140°F 315°F
B Total length, mm 403 to 413 403 to 413 270 to 280
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0

29
D Bulb length, mm 20 to 30 20 to 30 25 to 35
E Bulb OD, mm 5.0 to 6.0 5.0 to 6.0 <5.0 and >stem
Scale location:
Bottom of bulb to line at −35°F −65°F 245°F
E 1 – 05

F Distance, mm 200 to 230 200 to 230 135 to 150


G Length of graduated portion, mm 116 to 162O 116 to 162O 67 to 101O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60H

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 10.0
L Length, mm 4.0 to 7.0
M Distance to bottom, mm 112 to 116
G
The test temperature shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).
O
Capillary clearances shall conform to Section 8.
R
A suitable mercury-thallium alloy shall be used as the actuating liquid.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 78F-86FF 79F-86FF 80F-86FF

No reproduction or networking permitted without license from IHS


IP No.
Name Saybolt Viscosity Saybolt Viscosity Saybolt Viscosity
Reference Fig. No. 8 8 8
Range 295 to 315°F 345 to 365°F 395 to 415°F
For test at 300°FG 350°FG 400°FG
A Immersion, mm total total total
Graduations:
Subdivisions 0.5°F 0.5°F 0.5°F
Long lines at each 1°F 1°F 1°F
Numbers at each 5°F 5°F 5°F
Scale error, max 0.5°F 0.5°F 0.5°F
Special inscription ASTM ASTM ASTM
78F-86 79F-86 80F-86
Expansion chamber:
Permit heating to 365°F 415°F 465°F
B Total length, mm 270 to 280 270 to 280 270 to 280
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 25 to 35 25 to 35 25 to 35

30
E Bulb OD, mm ≤5.0 and ≥stem ≤5.0 and ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at 295°F 345°F 395°F
E 1 – 05

F Distance, mm 135 to 150 135 to 150 135 to 150


G Length of graduated portion, mm 67 to 101O 67 to 101O 67 to 101O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60H 60H 60H

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 10.0 8.0 to 10.0 8.0 to 10.0
L Length, mm 4.0 to 7.0 4.0 to 7.0 4.0 to 7.0
M Distance to bottom, mm 112 to 116 112 to 116 112 to 116
G
The test temperature shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H
Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 81F-86FF 82C-86 82F-86FF 83C-03 83F-03FF
IP No.

No reproduction or networking permitted without license from IHS


Name Saybolt Viscosity Fuel Rating Engine Fuel Rating Air-Low
Reference Fig. No. 8 11 11
Range 445 to 465°F −15 to + 105°C 0 to 220°F 15 to 70°C 60 to 160°F
G
For test at 450°FG 51.7°C and 65.6°C 125°Fand
150°FG
A Immersion, mm total 30EE 40EE
Graduations:
Subdivisions 0.5°F 1°C 2°F 1°C 1°F
Long lines at each 1°F 5°C 10°F 5°C 5°F
Numbers at each 5°F 10°C 20°F 10°C 10°F
Scale error, max 0.5°F 1°C 2°F 1°C 2°F
Special inscription ASTM ASTM ASTM
81F-86 82C-86 or 82F-86 83C-03 or 83F-03
30 MM IMM 40 MM IMM
Expansion chamber:
Permit heating to 515°F 125°C 260°F 100°C 215°F
B Total length, mm 270 to 280 159 to 165 168 to 174
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0

31
D Bulb length, mm 25 to 35 6 to 11 6 to 11
E Bulb OD, mm ≤5.0 and ≥stem 5.0 to 6.5 5.0 to 6.5
Scale location:
Bottom of bulb to line at 445°F −15°C 0°F 15°C 60°F
E 1 – 05

F Distance, mm 135 to 150 62 to 70 71 to 78


G Length of graduated portion, mm 67 to 101O 65 to 81O 64 to 81O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 60H

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 10.0 8.0 to 9.0 8.0 to 9.0
L Length, mm 4.0 to 7.0
M Distance to bottom, mm 112 to 116 28 to 32 38 to 42
G
The test temperature shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).
O
Capillary clearances shall conform to Section 8.
EE
Immersion line shall be omitted.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 84C-86 84F-86FF 85C-86 85F-86FF 86C-86 86F-86FF

No reproduction or networking permitted without license from IHS


IP No.
Name Fuel Rating, Orifice Tank Fuel Rating, Surge Fuel Rating, Mix
Reference Fig. No. 11 11 8
Range 25 to 80°C 75 to 175°F 40 to 150°C 100 to 300°F 95 to 175°C 200 to 350°F
G G
For test at 51.7°C 125°FG 107.2°C 225°FG 104.4°C and 220°F and
148.9°CG 300°FG
A Immersion, mm 249EE 181EE 35EE
Graduations:
Subdivisions 1°C 1°F 1°C 2°F 1°C 2°F
Long lines at each 5°C 5°F 5°C 10°F 5°C 10°F
Numbers at each 10°C 10°F 10°C 20°F 10°C 20°F
Scale error, max 1°C 2°F 1°C 2°F 1°C 2°F
Special inscription ASTM ASTM ASTM
84C-86 or 84F-86 85C-86 or 85F-86 86C-86 or 86F-86
249 MM IMM 181 MM IMM 35 MM IMM
Expansion chamber:
Permit heating to 100°C 215°F 170°C 340°F 195°C 390°F
B Total length, mm 378 to 387 305 to 314 164 to 170

32
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 6 to 11 6 to 11 6 to 11
E Bulb OD, mm 5.0 to 6.5 5.0 to 6.5 5.0 to 6.5
Scale location:
E 1 – 05

Bottom of bulb to line at 25°C 75°F 40°C 100°F 95°C 200°F


F Distance, mm 284 to 292 213 to 221 67 to 75
G Length of graduated portion, mm 62 to 79O 54 to 77O 58 to 81O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 22
Stem enlargement:
K OD, mm 8.0 to 9.0 8.0 to 9.0 8.0 to 9.0
L Length, mm
M Distance to bottom, mm 247 to 251 179 to 183 33 to 37
O

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Capillary clearances shall conform to Section 8.
EE
Immersion line shall be omitted.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
TABLE 1 Continued
ASTM No. 87C-86 87F-86FF 88C-86 88F-86FF 89C-86

No reproduction or networking permitted without license from IHS


IP No.
Name Fuel Rating Coolant Vegetable Oil Flash Solidification Point
Reference Fig. No. 8 5 3
Range 150 to 205°C 300 to 400°F 10 to 200°C 50 to 392°F −20 to + 10°C
G
For test at 190.6°C 375°FG
A Immersion, mm 40EE 57 76
Graduations:
Subdivisions 1°C 1°F 1°C 2°F 0.1°C
Long lines at each 5°C 5°F 5°C 10°F 0.5°C
Numbers at each 10°C 10°F 10°C 20°F 1.0°C
Scale error, max 1°C 2°F 1°C 2°F 0.1°C
Special inscription ASTM ASTM ASTM
87C-86 or 87F-86 88C-86 or 88F-86 89C-86
40 MM IMM 57 MM IMM 76 MM IMM
Expansion chamber:
Permit heating to 225°C 440°F 250°C 490°F 60°C
B Total length, mm 169 to 175 282 to 292 365 to 375
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0

33
D Bulb length, mm 6 to 11 9 to 13 18 to 28
E Bulb OD, mm 5.0 to 6.5 ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at 150°C 300°F 10°C 50°F −20°C
E 1 – 05

F Distance, mm 72 to 80 75 to 90 116 to 130


G Length of graduated portion, mm 59 to 82O 143 to 177O 185 to 219O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 21

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 9.0 7.5 to 8.5E
L Length, mm 2.5 to 5.0
M Distance to bottom, mm 38 to 42 64 to 66
E
The length of the enlargement, and the distance from the bottom of the enlargement to the bottom of the bulb shall be measured with the test gage shown in Fig. 1.
O

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Capillary clearances shall conform to Section 8.
EE
Immersion line shall be omitted.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued

No reproduction or networking permitted without license from IHS


ASTM No. 90C-86 91C-86 92C-86
IP No.
Name Solidification Point Solidification Point Solidification Point
Reference Fig. No. 3 7 7
Range 0 to 30°C 20 to 50°C 40 to 70°C

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
For test at
A Immersion, mm 76 76 76
Graduations:
Subdivisions 0.1°C 0.1°C 0.1°C
Long lines at each 0.5°C 0.5°C 0.5°C
Numbers at each 1.0°C 1.0°C 1.0°C
Scale error, max 0.1°C 0.1°C 0.1°C
Special inscription ASTM ASTM ASTM
90C-86 91C-86 92C-86
76 MM IMM 76 MM IMM 76 MM IMM
Expansion chamber:
Permit heating to 80°C 100°C 110°C
B Total length, mm 365 to 375 365 to 375 365 to 375

34
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 18 to 28 18 to 28 18 to 28
E Bulb OD, mm ≤5.0 and ≥stem ≤5.0 and ≥stem ≤5.0 and ≥stem
E 1 – 05

Scale location:
Bottom of bulb to line at 0°C 20°C 40°C
F Distance, mm 116 to 130 116 to 130 116 to 130
G Length of graduated portion, mm 185 to 219O 185 to 219O 185 to 219O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


I Distance to bottom, min, mm
J Distance to top, max, mm 41 41
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued

No reproduction or networking permitted without license from IHS


ASTM No. 93C-86 94C-86 95C-86
IP No.
Name Solidification Point Solidification Point Solidification Point
Reference Fig. No. 7 7 7
Range 60 to 90°C 80 to 110°C 100 to 130°C
For test at
A Immersion, mm 76 76 76
Graduations:
Subdivisions 0.1°C 0.1°C 0.1°C
Long lines at each 0.5°C 0.5°C 0.5°C
Numbers at each 1.0°C 1.0°C 1.0°C
Scale error, max 0.1°C 0.1°C 0.2°C
Special inscription ASTM ASTM ASTM
93C-86 94C-86 95C-86
76 MM IMM 76 MM IMM 76 MM IMM
Expansion chamber:
Permit heating to 130°C 150°C 170°C
B Total length, mm 365 to 375 365 to 375 365 to 375

35
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 18 to 28 18 to 28 18 to 28
E Bulb OD, mm ≤5.0 and ≥stem ≤5.0 and ≥stem ≤5.0 and ≥stem
E 1 – 05

Scale location:
Bottom of bulb to line at 60°C 80°C 100°C
F Distance, mm 116 to 130 116 to 130 116 to 130
G Length of graduated portion, mm 185 to 219O 185 to 219O 185 to 219O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


I Distance to bottom, min, mm
J Distance to top, max, mm 41 41 41
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 96C-86 97C-86 97F-86FF 98C-86 98F-86FF
IP No.

No reproduction or networking permitted without license from IHS


Name Solidification Point Tank Tank
Reference Fig. No. 7 4AA 4AA
Range 120 to 150°C −18 to + 49°C 0 to 120°F 16 to 82°C 60 to 180°F
For test at
A Immersion, mm 76 total total
Graduations:
Subdivisions 0.1°C 0.5°C 1°F 0.5°C 1°F
Long lines at each 0.5°C 1°C 5°F 1°C 5°F
Numbers at each 1.0°C 5°C 10°F 5°C 10°F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Scale error, max 0.2°C 0.3°C 0.5°F 0.3°C 0.5°F
Special inscription ASTM ASTM ASTM
96C-86 97C-86 or 97F-86 98C-86 or 98F-86
76 MM 1MM
Expansion chamber:
Permit heating to 190°C 100°C 212°F 100°C 212°F
B Total length, mm 365 to 375 300 to 305 300 to 305
CC CC
C Stem OD, mm 6.0 to 7.0
D Bulb length, mm 18 to 28 15 to 25 15 to 25

36
E Bulb OD, mm ≤5.0 and ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at 120°C −18°C 0°F 16°C 60°F
E 1 – 05

F Distance, mm 116 to 130 65 to 80 65 to 80


G Length of graduated portion, mm 185 to 219O 175 to 210O 175 to 210O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 41

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.
AA
Special finish, see 6.2.2.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but will not enter a 5.0-mm slot gage. A minor diameter of 4 mm is
permissible provided that the major diameter is not less than 7 mm.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 99C-92 99F-86FF 100C-86 101C-86
IP No.
Name Weathering TestR Solidification Point Solidification Point
Reference Fig. No. 3AA 7 7

No reproduction or networking permitted without license from IHS


Range −50 to + 5°C −58 to + 41°F 145 to 205°C 195 to 305°C
For test at
GG
A Immersion, mm 35 76 76
Graduations:
Subdivisions 0.2°C 0.5°F 0.2°C 0.5°C
Long lines at each 1°C 1°F 1°C 1°C
Numbers at each 2°C 5°F 2°C 5°C
Scale error, max 0.2°C 0.4°F 0.4°C 1.0°C
Special inscription ASTM ASTM ASTM
99C-92 or 99F–86 100C-86 101C–86
35 MM IMM 76 MM IMM 76 MM IMM
MERC-THAL

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Expansion chamber:
Permit heating to Elongated type Elongated type 150°F 250°C 330°C
66°C
B Total length, mm 299 to 305 365 to 375 365 to 375
C Stem OD, mm 6.5 to 8.0HH 6.0 to 7.0 6.0 to 7.0
D Bulb length, mm 15 to 20 18 to 28 18 to 28

37
E Bulb OD, mm 6.0 to 7.0 ≤5.0 and ≥stem ≤5.0 and ≥stem
Scale location:
Bottom of bulb to line at −50°C −58°F 145°C 195°C
E 1 – 05

F Distance, mm 50 to 75 116 to 130 116 to 130


G Length of graduated portion, mm 169 to 215O 185 to 219O 185 to 219O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
I Contraction chamber:
Distance to bottom, min, mm
J Distance to top, max, mm 41 41

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K Stem enlargement:
OD, mm
L Length, mm
M Distance to bottom, mm

O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


R
A suitable mecury-thallium alloy shall be used as the actuating liquid.
AA
Special finish, see 6.2.2.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.
GG
The immersion line shall be visible in the case opening after assembly. The immersion shall be measured from the bottom of the bulb rather than from the bottom of the armor. See 6.2.2.
HH
The stem shall be either the round or lens-front type.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 102C-86 103C-86 104C-86

No reproduction or networking permitted without license from IHS


IP No. 83C 84C 85C

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Name Solvents Distillation Solvents Distillation Solvents Distillation
Reference Fig. No. 7 7 7
Range 123 to 177°C 148 to 202°C 173 to 227°C
For test at
A Immersion, mm 100 100 100
Graduations:
Subdivisions 0.2°C 0.2°C 0.2°C
Long lines at each 1°C 1°C 1°C
Numbers at each 2°C 2°C 2°C
Scale error, max 0.3°C 0.4°C 0.4°C
Special inscription ASTM ASTM ASTM
102C-86 103C-86 104C-86
100 MM IMM 100 MM IMM 100 MM IMM
Expansion chamber:
Permit heating to 200°C 225°C 250°C
B Total length, mm 390 to 400 390 to 400 390 to 400

38
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 15 to 20 15 to 20 15 to 20
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
E 1 – 05

F Bottom of bulb to line at 123°C 148°C 173°C


G Distance, mm 125 to 145 125 to 145 125 to 145
Length of graduated portion, mm 190 to 235O 190 to 235O 190 to 235O
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 35T 35T 35T
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


T
Mercury shall be near the bottom of the chamber at 0°C.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 105C-86 106C-86 107C-86

No reproduction or networking permitted without license from IHS


IP No. 86C 87C 88C
Name Solvents Distillation Solvents Distillation Solvents Distillation
Reference Fig. No. 7 7 7
Range 198 to 252°C 223 to 277°C 248 to 302°C
For test at
A Immersion, mm 100 100 100
Graduations:
Subdivisions 0.2°C 0.2°C 0.2°C
Long lines at each 1°C 1°C 1°C
Numbers at each 2°C 2°C 2°C
Scale error, max 0.6°C 0.8°C 1.0°C
Special inscription ASTM ASTM ASTM
105C-86 106C-86 107C-86
100 MM IMM 100 MM IMM 100 MM IMM
Expansion chamber:
Permit heating to 275°C 300°C 325°C
B Total length, mm 390 to 400 390 to 400 390 to 400

39
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 15 to 20 15 to 20 15 to 20
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
E 1 – 05

Bottom of bulb to line at 198°C 223°C 248°C


F Distance, mm 125 to 145 125 to 145 125 to 145
G Length of graduated portion, mm 190 to 235O 190 to 235O 190 to 235O
Ice-point scale:
Range

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 35T 35T 35T
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
O
Capillary clearances shall conform to Section 8.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


T
Mercury shall be near the bottom of the chamber at 0°C.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 108F-86FF 109F-86FF 110C-86 110F-86FF
IP No. 93C

No reproduction or networking permitted without license from IHS


Name Saybolt Viscosity Saybolt Viscosity Kinematic ViscosityM
Reference Fig. No. 8 8 6
Range 270 to 290°F 320 to 340°F 133.6 to 136.4°C 272.5 to 277.5°F
For test at 275°FG 325°FG 135°C 275°F
A Immersion, mm total total total
Graduations:
Subdivisions 0.5°F 0.5°F 0.05°C 0.1°F
Long lines at each 1°F 1°F 0.1 and 0.5°C 0.5 and 1°F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Numbers at each 5°F 5°F 1°C 1°F
Scale error, max 0.5°F 0.5°F 0.15°C 0.3°F
Special inscription ASTM ASTM ASTM
108F-86 109F-86 110C-86 or 110F-86
Expansion chamber:
Permit heating to 340°F 390°F 170°C 350°F
B Total length, mm 270 to 280 270 to 280 300 to 310
C Stem OD, mm 6.0 to 7.0 6.0 to 7.0 6.0 to 8.0
D Bulb length, mm 25 to 35 25 to 35 45 to 55
E Bulb OD, mm ≤5.0 and ≥stem ≤5.0 and ≥stem ≥stem

40
Scale location:
Bottom of bulb to line at 270°F 320°F 133.6°C 272.5°F
F Distance, mm 135 to 150 135 to 150 160 to 180
E 1 – 05

G Length of graduated portion, mm 67 to 101O 67 to 101O 40 to 90O


Ice-point scale:
Range –0.3 to +0.6°C 31.5 to 33.0°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 100
J Distance to top, max, mm 60H 60H 125
Stem enlargement:

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


K OD, mm 8.0 to 10.0 8.0 to 10.0
L Length, mm 4.0 to 7.0 4.0 to 7.0
M Distance to bottom, mm 112 to 116 112 to 116
G
The test temperature shall be indicated by an arrow whether the graduation corresponding to that point is numbered or not.
H
Long, narrow shape; mercury shall be in the chamber at 0°C (32°F).
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


reading shall be expressed to the nearest 0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius
range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 111C-86 112C-86 113C-86 113F-86FF

No reproduction or networking permitted without license from IHS


IP No. 89C
Name Tar Acids Distillation Solidification Point of Benzene Softening Point (Bitumen) Wide Range
Reference Fig. No. 7 6 4
Range 170 to 250°C 4 to 6°C −1 to + 175°C 30 to 350°F
For test at 5.4°C
A Immersion, mm 100 total total
Graduations:
Subdivisions 0.2°C 0.02°C 0.5°C 1°F
Long lines at each 1°C 0.1°C 1°C 5°F
Numbers at each 2°C 0.2°C 5°C 10°F
Scale error, max 0.4°C to 225°C 0.04°C 0.5°C 1°F
0.6°C above 225°C
Special inscription ASTM ASTM ASTM
111C-86 112C-86 113C-86 or 113F-86
100 MM IMM
Expansion chamber:
Permit heating to 275°C 50°C 225°C 440°F
B Total length, mm 390 to 400 210 to 220 400 to 410

41
C Stem OD, mm 6.0 to 7.0 6.0 to 7.5 6.0 to 8.0
D Bulb length, mm 10 to 15 25 to 35 10 to 15
E Bulb OD, mm 6.0 to 7.0 6.0 to ≥stem 4.5 to 5.5
Scale location:
E 1 – 05

Bottom of bulb to line at 170°C 4°C 0°C 32°F


F Distance, mm 115 to 135 100 to 115 80 to 90
G Length of graduated portion, mm 200 to 245O 45 to 75O 250 to 290O
Ice-point scale:
Range −0.2 to + 0.2°CO
H Bottom of bulb to ice-point, mm 60 to 70
Contraction chamber:
I Distance to bottom, min, mm 80

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 35 90II

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
II
Contraction chamber to be long narrow type.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 114C-86 115C-86 116C-86 117C-86
IP No. 14C

No reproduction or networking permitted without license from IHS


Name Aviation Fuel Freezing PointB See Table 5 Bomb Calorimeter Bomb Calorimeter
Reference Fig. No. 4 9 and 14 9 and 14
Range −80 to + 20°C 18.9 to 25.1°C 23.9 to 30.1°C
For test at
A Immersion, mm total total total
Graduations:
Subdivisions 0.5°C 0.01°C 0.01°C
Long lines at each 1°C 0.05°C 0.05°C
Numbers at each 5°C 0.1°C 0.1°C
Scale error, max 1°C 0.1°CJJ 0.1°CJJ
Special inscription ASTM ASTM ASTM
114C-86 116C-86 117C-86
Expansion chamber:
Permit heating to 45°C 60°C 60°C
B Total length, mm 295 to 305 604 to 614 604 to 614
C Stem OD, mm 6.0 to 8.0 7.0 to 8.2 7.0 to 8.2
KK KK
D Bulb length, mm 8 to 16
E Bulb OD, mm ≥stem 7.0 to 8.2BB 7.0 to 8.2BB

42
Scale location:
Bottom of bulb to line at −80°C 19°C 24°C
F Distance, mm 50 to 70 220 to 240 220 to 240
G Length of graduated portion, mm 170 to 210O 300 to 350O 300 to 350O
E 1 – 05

Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 110 110
Stem enlargement:
K OD, mm

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


L Length, mm
M Distance to bottom, mm
B
Toluene or other suitable liquid colored red with a permanent dye shall be used as the actuating liquid.
O
Capillary clearances shall conform to Section 8.
BB
The bulb diameter shall not be more than 0.5 mm greater than the stem.
JJ

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


Over any interval of 1°C the change in correction shall not exceed 0.01°C. The correction at the lowest temperature of the nominal range shall not change by more than 0.02°C immediately after the
thermometer has been heated for 15 min at a temperature 30°C higher, and allowed to cool naturally in air.
KK
The capillary bore shall be large enough in relation to the bulb to ensure that (without tapping) jumping of the meniscus does not exceed one half of the smallest scale division, when the temperature is
rising at a uniform rate not exceeding 0.05°C/min.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 118C-86 118F-86FF 119C-86 119F-86FF 120C-86
IP No. 92C
Name Kinematic ViscosityM Coolant (Antifreeze) Kinematic

No reproduction or networking permitted without license from IHS


Freezing Point ViscosityM
Reference Fig. No. 6 10 6
Range 28.6 to 31.4°C 83.5 to 88.5°F −38.3 to − 30°C −37 to − 22°F 38.6 to 41.4°C
For test at 30°C 86°F 40°C
A Immersion, mm total 100LL total
Graduations:
Subdivisions 0.05°C 0.1°F 0.1°C 0.2°F 0.05°C
Long lines at each 0.1 and 0.5°C 0.5 and 1°F 0.5°C 1°F 0.1 and 0.5°C
Numbers at each 1°C 1°F 1°C 2°F 1°C
Scale error, max 0.1°C 0.2°F 0.2°C 0.4°F 0.1°C

Special inscription ASTM ASTM ASTM


118C-86 or 118F-86 119C-86 or 119F-86 120C-86
100 MM IMM
Expansion chamber:
Permit heating to 105°C 220°F 50°C 122°F 105°C
B Total length, mm 300 to 310 415 to 425 300 to 310
C Stem OD, mm 6.0 to 8.0 6.0 to 7.0 6.0 to 8.0

43
D Bulb length, mm 45 to 55 25 to 35 45 to 55
E Bulb OD, mm ≥stem 5.0 to ≥stem ≥stem
Scale location:
E 1 – 05

Bottom of bulb to line at 28.6°C 83.5°F −38.3°C −37°F 38.6°C


F Distance, mm 145 to 165 200 to 230 145 to 165
G Length of graduated portion, mm 40 to 90O 60 to 110O 40 to 90O
Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO −0.5 to + 0.5°CO 31 to 33°FO −0.3 to + 0.3°CO
H Bottom of bulb to ice-point, mm 370 to 390
Contraction chamber:
I Distance to bottom, min, mm 100 325 100

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


J Distance to top, max, mm 125 350 125
Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.
II
The thermometer is to be calibrated for 100-mm immersion for the main scale, the ice point is to be calibrated for total immersion.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 121C-86 122C-86 123C-86

No reproduction or networking permitted without license from IHS


IP No. 32C 94C 95C
Name Kinematic ViscosityM Brookfield ViscosityR Brookfield Viscosity
Ref. Fig. No. 6 4 4
Range 98.6 to 101.4°C –45 to –35°C –35 to –25°C
For Test at 98.9 and 100°C 150°C
A Immersion, mm total total total
Graduations:
Subdivisions 0.05°C 0.1°C 0.1°C
Long Lines at each 0.1 and 0.5°C 0.5°C 0.5°C
Numbers at each 1°C 1°C 1°C
Scale error, max 0.1°C 0.4°C 0.4°C
Special inscription ASTM ASTM ASTM
121C-86 122C-86 123C-86
MERC-THAL
Expansion chamber:
Permit heating to 130°C 80°C 80°C
B Total length, mm 300 to 310 295 to 305 295 to 305
C Stem OD, mm 6.0 to 8.0 5.5 to 8.0 5.5 to 8.0

44
D Bulb length, mm 45 to 55 30 to 40 30 to 40
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at 98.6°C –45°C –35°C
E 1 – 05

F Distance, mm 145 to 165 100 to 120 100 to 120


G Length of graduated portion, mm 40 to 90O 115 to 165O 115 to 165O
Ice-point scale:
Range –0.3 to +0.3°CO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 100
J Distance to top, max, mm 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.
O
Capillary clearances shall conform to Section 8.
R
A suitable mecury-thallium alloy shall be used as the actuating liquid.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 124C-86 125C-86 126C-86 126F-86FF

No reproduction or networking permitted without license from IHS


IP No. 96C 97C 71C
Name Brookfield Viscosity Brookfield Viscosity Kinematic ViscosityM
Reference Fig. No. 4 4 10
Range −25 to − 15°C −15 to − 5°C −27.4 to − 24.6°C −17.5 to − 12.5°F
For test at −26.1°C −15°F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
A Immersion, mm total total total
Graduations:
Subdivisions 0.1°C 0.1°C 0.05°C 0.1°F
Long lines at each 0.5°C 0.5°C 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C 1°C 1°C 1°F
Scale error, max 0.2°C 0.2°C 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
124C-86 125C-86 126C-86 or 126F-86
Expansion chamber:
Permit heating to 80°C 80°C 105°C 220°F
B Total length, mm 295 to 305 295 to 305 300 to 310
C Stem OD, mm 5.5 to 8.0 5.5 to 8.0 6.0 to 8.0
D Bulb length, mm 30 to 40 30 to 40 45 to 55

45
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at −25°C −15°C −27.4°C −17.5°F
E 1 – 05

F Distance, mm 100 to 120 100 to 120 80 to 110


G Length of graduated portion, mm 115 to 165O 115 to 165O 40 to 90O
Ice-point scale:
Range −0.3 to + 0.3°CO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 180
J Distance to top, max, mm 205

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest
0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 127C-86 128C-86 128F-86FF 129C-86 129F-86FF

No reproduction or networking permitted without license from IHS


IP No. 99C 33C 36C
Name Kinematic ViscosityM Kinematic ViscosityM Kinematic ViscosityM
Ref. Fig. No. 10 4 6
Range −21.4 to − 18.6°C −1.4 to + 1.4°C 29.5 to 34.5°F 91.6 to 94.4°C 197.5 to 202.5°F
For Test at −20°C 0°C 32°F 93.3°C 200°F
A Immersion, mm total total total
Graduations:
Subdivisions 0.05°C 0.05°C 0.1°F 0.05°C 0.1°F
Long Lines at each 0.1 and 0.5°C 0.1 and 0.5°C 0.5 and 1°F 0.1 and 0.5°C 0.5 and 1°F
Numbers at each 1°C 1°C 1°F 1°C 1°F
Scale error, max 0.1°C 0.1°C 0.2°F 0.1°C 0.2°F
Special inscription ASTM ASTM ASTM
127C-86 128C-86 or 128F-86 129C-86 or 129F-86
Expansion chamber:
Permit heating to 105°C 105°C 220°F 120°C 250°F
B Total length, mm 300 to 310 300 to 310 300 to 310
C Stem OD, mm 6.0 to 8.0 6.0 to 8.0 6.0 to 8.0
D Bulb length, mm 45 to 55 45 to 55 45 to 55

46
E Bulb OD, mm ≥stem ≥stem ≥stem
Scale location:

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Bottom of bulb to line at −21.4°C −1.4°C 29.5°F 91.6°C 197.5°F
E 1 – 05

F Distance, mm 80 to 110 145 to 165 145 to 165


G Length of graduated portion, mm 40 to 90O 40 to 90O 40 to 90O
Ice-point scale:
Range −0.3 to + 0.3°CO −0.3 to + 0.3°CO 31.5 to 32.5°FO
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm 180 100
J Distance to top, max, mm 205 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
M
For kinematic viscosity thermometers, the ice-point reading shall be taken within 1 h after being at the test temperature for not less than 3 minutes. The ice-point reading shall be expressed to the nearest
0.01°C or 0.02°F and applied as explained in Test Method E 77, Section 13.

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


O
Capillary clearance shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International


Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 130C-90 130F-90FF 132C-95 133C-01 134C-03
IP No. 102C 37C

No reproduction or networking permitted without license from IHS


Name Tank Kinematic ViscosityM Precision Sludge
Ref. Fig. No. 4AA 6 3 3
Range −7 to + 105°C 20 to 220°F 148.6 to 151.4°C –38 to +2°C 144 to 156°C
For Test at 150°C 150°C
A Immersion, mm total total 76 100
Graduations:
Subdivisions 0.5°C 1°F 0.05°C 0.1°C 0.2°C
Long Lines at each 1°C 5°F 0.1 and 0.5°C 0.5°C 1°C
Numbers at each 5°C 10°F 1°C 1°C 2°C
Scale error, max 0.5°C 1°F 0.20°C 0.1°C 0.2°C
Special inscription ASTM ASTM ASTM ASTM
130C-90 or 130F–90 132C–95 133C-01 134C-03
76 MM IMM 100 MM IMM
Expansion chamber:
Permit heating to 125°C 257°F 185°C 55°C required
B Total length, mm 300 to 305 300 to 310 374 to 384 260 to 280
C Stem OD, mm cc 6.0 to 8.0 7.0 to 8.0 5.0 to 8.0
D Bulb length, mm 15 to 25 45 to 55 25 to 35 10 to 25

47
E Bulb OD, mm ≥stem ≥stem ≥stem ≥stem
Scale location:
Bottom of bulb to line at 0°C 32°F 148.6°C -38°C 150°C
F Distance, mm 76 to 93 160 to 180 115 to 135 170 to 210
E 1 – 05

G Length of graduated portion, mm 175 to 210O 40 to 90°CO 190 to 230O 50 to 80O


Ice-point scale:
Range −0.3 to + 0.6°CO
H Bottom of bulb to ice-point, mm
Contraction chamber: required
I Distance to bottom, min, mm 100
J Distance to top, max, mm 125

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm
L Length, mm
M Distance to bottom, mm
AA
Special finish, see 6.2.2.
CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


will not enter a 5.0-mm slot gage. A minor diameter of 4-mm is permissible provided that the major diameter is not less than 7 mm.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion
corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Copyright ASTM International
Provided by IHS under license with ASTM
TABLE 1 Continued
ASTM No. 135C-03 135F-03FF 136C-03 136F-03FF 137C-03
IP No.

No reproduction or networking permitted without license from IHS


Name Fuel Rating Air-High Aviation Fuel Density Oxidation Cell Test
Ref. Fig. No. 11 4 3
Range 38 to 93°C 100 to 200°F –20 to +60°C –5 to +140°F 80 to 100°C
For Test at 51.7°CG 125°FG
A Immersion, mm 40EE total 76
Graduations:

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
Subdivisions 1°C 1°F 0.2°C 0.5°F 0.1°C
Long Lines at each 5°C 5°F 1°C 1°F 0.5°C
Numbers at each 10°C 10°F 2°C 5°F 2°C
Scale error, max 1°C 2°F 0.15°C 0.25°F 0.1°C
Special inscription ASTM ASTM ASTM
135C-03 or 135F–03 136C-03 or 136F–03 137C-03
40 MM IMM 76 MM IMM
Expansion chamber:
Permit heating to 125°C 250°F 110°C 230°F 130°C
B Total length, mm 168 to 174 285 to 295 250 to 260
OO
C Stem OD, mm 6.0 to 7.0 6.0 to 8.0
D Bulb length, mm 6 to 11 15 to 20 18 to 28

48
E Bulb OD, mm 5.0 to 6.5 bulb size ≥ stem size $6.0 and #stem
Scale location:
Bottom of bulb to line at 38°C 100°F –20°C –4°F 80°C
E 1 – 05

F Distance, mm 71 to 78 35 to 50 90 to 100
G Length of graduated portion, mm 64 to 81O 200 to 230O 115 to 125
Ice-point scale:
Range
H Bottom of bulb to ice-point, mm
Contraction chamber:
I Distance to bottom, min, mm
J Distance to top, max, mm 45

Not for Resale, 05/22/2006 09:01:29 MDT.jsx


Stem enlargement:
K OD, mm 8.0 to 9.0
L Length, mm
M Distance to bottom, mm 38 to 42
AA
Special finish, see 6.2.2.
CC
The stem shall be of the lens front type. The cross section of the stem shall be such that it will pass through a 8.0-mm ring gage but will not enter a 5.0-mm slot gage. A minor

Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane


diameter of 4-mm is permissible provided that the major diameter is not less than 7 mm.
O
Capillary clearances shall conform to Section 8.
FF
For Fahrenheit thermometers, dimension G (length of graduated portion) shall be measured as the length of graduated portion corresponding to the nominal Celsius range.

Copyright by ASTM Int'l (all rights reserved);


Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
E 1 – 05

TABLE 2 Alphabetical List of ASTM Thermometers Covered by Specification E 1

NOTE 1—The specifications appear in numeric sequence in this standard.


--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Thermometer No. Thermometer No.


Thermometer Name Thermometer Name
°C °F °C °F
Aviation fuel density 136C 136F ... ...
Aviation fuel freezing point 114C ... 3C 3F
Bomb calorimeter 56C 56F Petrolatum melting point 61C 61F
116C ... Precision 62C 62F
117C ... 63C 63F
Brookfield Viscosity 122C ... 64C 64F
123C ... 65C 65F
124C ... 66C 66F
125C ... 67C 67F
Butadiene boiling point range 52C ... 68C 68F
Cleveland open flash 11C 11F 69C 69F
Cloud and pour 5C 5F 70C 70F
... ... 133C ...
Cloud and pour, low 6C 6F Reid vapor pressure 18C 18F
Congealing point 54C 54F Saybolt viscosity 17C 17F

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Coolant (antifreeze) freezing point ... 75F 18C 18F
... 76F 19C 19F
119C 119F 20C 20F
Density wide range 12C 12F 21C 21F
Enclosed scale 115C ... 22C 22F
Engler viscosity 23C ... ... 77F
24C ... ... 78F
25C ... ... 79F
Fuel rating 82C 82F ... 80F
83C 83F ... 81F
84C 84F ... 108F
85C 85F ... 109F
... ... Sludge 134C ...
86C 86F Softening point (bitumen) wide-range 113C 113F
87C 87F Solidification point 89C ...
135C 135F ... ...
Gas calorimeter, inlet ... 50F 90C ...
Gas calorimeter, outlet ... 51F 91C ...
General purpose—see partial immersion; precision 92C ...
Gravity 12C 12F 93C ...
High aniline point 35C 35F 94C ...
High distillation 8C 8F 95C ...
High Pensky Martens 10C 10F 96C ...
High softening point 16C 16F 100C ...
Kinematic viscosity 28C 28F 101C ...
29C 29F Solidification point of benzene 112C ...
... 30F Solvents distillation 37C ...
43C 43F 38C ...
44C 44F 39C ...
45C 45F 40C ...
46C 46F 41C ...
47C 47F 42C ...
48C 48F 102C ...

Copyright by ASTM Int'l (all rights reserved);


72C 72F 103C ...
73C 73F 104C ...
74C 74F 105C ...
110C 110F 106C ...
118C 118F 107C ...
120C ... Stability test of soluble nitrocellulose 26C ...
121C ... Stormer viscosity 49C ...
126C 126F Tag closed tester, low range 57C 57F
127C ... Tag closed tester, high range 9C 9F
128C 128F Tank 58C 58F
129C 129F 59C 59F
132C
Loss on heat 13C ... 60C 60F
Low aniline point 33C 33F 97C 97F
Low cloud and pour 6C 6F 98C 98F
Low distillation 7C 7F 130C 130F
Low Pensky Martens 9C 9F Tar acids distillation 111C ...
Low softening point 15C 15F Titer test 36C ...
Medium aniline point 34C 34F Turpentine distillation 27C ...
Oil in wax 71C 71F Vegetable oil flash 88C 88F
Oxidation cell test 137C ... ... ...

Copyright ASTM International 49


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

Thermometer No. Thermometer No.


Thermometer Name Thermometer Name
°C °F °C °F
Oxidation stability 22C 22F Wax melting point 14C 14F
Partial immersion (general use) 1C 1F Weathering test 99C 99F
2C 2F

TABLE 3 List of ASTM Thermometers by Temperature Range

Scale ASTM Scale ASTM


Celsius (Centigrade) Immersion, Fahrenheit Immersion,
Error, Thermometer Error, Thermometer
Range mm Range mm
max Number max Number
Graduated in 0.01°C
18.9 to 25.1°C total 0.1 116C ... ... ... ...
23.9 to 30.1°C total 0.1 117C ... ... ... ...
Graduated in 0.02°C Graduated in 0.05°F
4 to 6°C total 0.04 112C ... ... ... ...
19 to 35°C total 0.10 56C 66 to 95°F total 0.20 56F
Graduated in 0.05°C Graduated in 0.1°F
−55.4 to −52.6°C total 0.1 74C −67.5 to −62.5°F total 0.2 74F

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
−41.4 to −38.6°C total 0.1 73C −42.5 to −37.5°F total 0.2 73F
−27.4 to −24.6°C total 0.1 126C −17.5 to −12.5°F total 0.2 126F
−21.4 to −18.6°C total 0.1 127C ... ... ... ...
−19.4 to −16.6°C total 0.1 72C −2.5to +2.5°F total 0.2 72F
−1.4 to +1.4°C total 0.1 128C 29.5to 34.5°F total 0.2 128F
... ... ... ... 54 to 101°F total 0.2 50F
18.6 to 21.4°C total 0.1 44C 66.5 to 71.5°F total 0.2 44F
... ... ... ... 69 to 116°F total 0.2 51F
23.6 to 26.4°C total 0.1 45C 74.5 to 79.5°F total 0.2 45F
28.6 to 31.4°C total 0.1 118C 83.5 to 88.5°F total 0.2 118F
36.6 to 39.4°C total 0.1 28C 97.5 to 102.5°F total 0.2 28F
38.5 to 41.5°C total 0.1 120C ... ... ... ...
48.6 to 51.4°C total 0.1 46C 119.5 to 124.5°F total 0.2 46F
52.6 to 55.4°C total 0.1 29C 127.5 to 132.5°F total 0.2 29F
58.6 to 61.4°C total 0.1 47C 137.5 to 142.5°F total 0.2 47F
80.6 to 83.4°C total 0.1 48C 177.5 to 182.5°F total 0.2 48F
91.6 to 94.4°C total 0.1 129C 197.5 to 202.5°F total 0.2 129F
... ... ... ... 207.5 to 212.5°F total 0.2 30F
98.6 to 101.4°C total 0.1 121C ... ... ... ...

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
133.6 to 136.4°C total 0.15 110C 272.5 to 277.5°F total 0.3 110F
148.6 to 151.4°C total 0.20 132C
Graduated in 0.1°C Graduated in 0.2°F
−51.6 to −34°C total 0.1 43C −61 to −29°F total 0.2 43F
−45 to −35°C total 0.4 122C ... ... ... ...
−38.3 to −30°C 100 0.2 119C −37 to −22°F 100 0.4 119F
−38 to +2°C total 0.1 62C −36 to +35°F total 0.2 62F
–38 to +2°C 76 mm 0.1 133C
−35 to −25°C total 0.4 123C ... ... ... ...
−25 to −15°C total 0.2 124C ... ... ... ...
−15 to −5°C total 0.2 125C ... ... ... ...
−20 to +10°C 76 0.1 89C ... ... ... ...
−10 to +5°C total 0.1 52C ... ... ... ...

Copyright by ASTM Int'l (all rights reserved);


−8 to +32°C total 0.1 63C 18 to 89°F total 0.2 63F
0 to 30°C 76 0.1 90C ... ... ... ...
19 to 27°C total 0.1 17C 66 to 80°F total 0.2 17F
20 to 50°C 76 0.1 91C ... ... ... ...
25 to 55°C total 0.1 64C 77 to 131°F total 0.2 64F
34 to 42°C total 0.1 18C 94 to 108°F total 0.2 18F
38 to 82°C 79 0.1 14C 100 to 180°F 79 0.2 14F
40 to 70°C 76 0.1 92C ... ... ... ...
49 to 57°C total 0.1 19C 120 to 134°F total 0.2 19F
50 to 80°C total 0.1 65C 122 to 176°F total 0.2 65F
57 to 65°C total 0.1 20C 134 to 148°F total 0.2 20F
60 to 90°C 76 0.1 93C ... ... ... ...
75 to 105°C total 0.1 66C 167 to 221°F total 0.2 66F
79 to 87°C total 0.1 21C 174 to 188°F total 0.2 21F
80 to 100°C 76 0.1 137C ... ... ... ...
80 to 110°C 76 0.1 94C ... ... ... ...
95 to 103°C total 0.1 22C 204 to 218°F total 0.2 22F
100 to 130°C 76 0.2 95C ... ... ... ...
120 to 150°C 76 0.2 96C ... ... ... ...
130 to 140°C total 0.2 26C ... ... ... ...

Copyright ASTM International 50


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 3 Continued
Scale ASTM Scale ASTM
Celsius (Centigrade) Immersion, Fahrenheit Immersion,
Error, Thermometer Error, Thermometer
Range mm Range mm
max Number max Number
Graduated in 0.2°C Graduated in 0.5°F
... ... ... ... −65 to +5°F 100 1 76F
−50 to +5°C 35 0.2 99C −55 to +40°F 35 0.4 99F
−38 to +42°C 50 0.2 33C −36.5 to +107.5°F 50 0.5 33F
... ... ... ... −35 to +35°F 100 0.5 75F
−20 to +60°C total 0.15 136C −5 to +140°F total 0.25 136F
−20 to +102°C total 0.15 12C −5 to +215°F total 0.25 12F
−2 to +52°C 100 0.2 37C ... ... ... ...
−2 to +68°C 45 0.2 36C ... ... ... ...
−2 to +80°C total 0.2 15C 30 to 180°F total 0.4 15F
18 to 28°C 90 0.1 23C ... ... ... ...
20 to 70°C 65 0.2 49C ... ... ... ...
20 to 100.6°C total 0.2 54C 68 to 213°F total 0.5 54F
24 to 78°C 100 0.2 38C ... ... ... ...
25 to 105°C 50 0.2 34C 77 to 221°F 50 0.5 34F
32 to 127°C 79 0.2 61C 90 to 260°F 79 0.5 61F
39 to 54°C 90 0.1 24C ... ... ... ...
48 to 102°C 100 0.2 39C ... ... ... ...

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
72 to 126°C 100 0.2 40C ... ... ... ...
90 to 170°C 50 0.4 35C 194 to 338°F 50 1 35F
95 to 105°C 90 0.1 25C ... ... ... ...
98 to 152°C 100 0.3 41C ... ... ... ...
95 to 155°C total 0.2 67C 203 to 311°F total 0.5 67F
... ... ... ... 245 to 265°F total 0.5 77F
123 to 177°C 100 0.3 102C ... ... ... ...
... ... ... ... 270 to 290°F total 0.5 108F
144 to 156°C 100 0.2 134C 295 to 315°F total 0.5 78F
145 to 205°C total 0.2 68C 293 to 401°F total 0.5 68F
145 to 205°C 76 0.4 100C ... ... ... ...
148 to 202°C 100 0.4 103C ... ... ... ...
... ... ... ... 320 to 340°F total 0.5 109F
A
170 to 250°C 100 111C ... ... ... ...
173 to 227°C 100 0.4 104C ... ... ... ...
... ... ... ... 345 to 365°F total 0.5 79F
198 to 252°C 100 0.6 105C ... ... ... ...
... ... ... ... 395 to 415°F total 0.5 80F
... ... ... ... 445 to 465°F total 0.5 81F
223 to 277°C 100 0.8 106C ... ... ... ...
248 to 302°C 100 1 107C ... ... ... ...
Graduated in 0.5°C Graduated in 1°F
−80 to +20°C total 1 114C ... ... ... ...
−37 to +21°C 76 0.2 71C −35 to +70°F 76 0.5 71F
−34 to +49°C total 0.5 58C −30 to +120°F total 0.5 58F
−20 to +50°C 57 0.5 57C −4 to +122°F 57 1 57F
−18 to +49°C total 0.5 97C 0 to 120°F total 0.5 97F
−18 to +82°C total 0.5 59C 0 to 180°F total 0.5 59F
−7 to +105°C total 0.5 130C 20 to 220°F total 1 130F
−5 to +110°C 57 0.5 9C 20 to 230°F 57 1 9F
−1 to +175°C total 0.5 113C 30 to 350°F total 1 113F
... ... ... ... 60 to 160°F 40 2 83F

Copyright by ASTM Int'l (all rights reserved);


16 to 82°C total 0.5 98C 60 to 180°F total 0.5 98F
... ... ... ... 75 to 175°F 249 2 84F
30 to 200°C total 0.3 16C 85 to 392°F total 0.5 16F
95 to 255°C 100 1 42C 100 to 200°F 40 2 135F
147 to 182°C 76 0.5 27C ... ... ... ...
155 to 170°C total 0.5 13C ... ... ... ...
... ... ... ... 300 to 400°F 40 2 87F
195 to 305°C total 0.5 69C 383 to 581°F total 1 69F
195 to 305°C 76 1 101C ... ... ... ...
295 to 405°C total 0.5 70C 563 to 761°F total 1 70F
Graduated in 1°C Graduated in 2°F
B C
−80 to +20°C 76 6C −112 to +70°F 76 6F
−38 to +50°C 108 0.5 5C −36 to +120°F 108 1 5F
−15 to +105°C 30 1 82C 0 to 220°F 30 2 82F
−20 to +150°C 76 0.5 1C 0 to 302°F 76 1 1F
−5 to +300°C 76 1 2C 20 to 580°F 76 2 2F
D E
−5 to +400°C 76 3C 20 to 760°F 76 3F
F G
−2 to +300°C total 7C 30 to 580°F total 7F
H I
−2 to +400°C total 8C 30 to 760°F total 8F
10 to 200°C 57 1 88C 50 to 392°F 57 2 88F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 51


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 3 Continued
Scale ASTM Scale ASTM
Celsius (Centigrade) Immersion, Fahrenheit Immersion,
Error, Thermometer Error, Thermometer
Range mm Range mm
max Number max Number
15 to 70°C 40 1 83C ... ... ... ...
25 to 80°C 249 1 84C ... ... ... ...
38 to 93°C 40 1 135C ... ... ... ...
40 to 150°C 181 1 85C 100 to 300°F 181 2 85F
77 to 260°C total 1 60C 170 to 500°F total 1 60F
95 to 175°C 35 1 86C 200 to 350°F 35 2 86F
150 to 205°C 40 1 87C ... ... ... ...
Graduated in 2°C Graduated in 5°F
J K
−6 to +400°C 25 11C 20 to 760°F 25 11F
L M
90 to 370°C 57 10C 200 to 700°F 57 10F
A
0.4°C to 225°C; 0.6°C above 225°C.
B
1°C to − 33°C; 2°C below − 33°C.
C
2°F to − 28°F; 4°F below − 28°F.
D
1°C to 301°C; 1.5°C above 301°C.
E
2°F to 574°F; 3°F above 574°F.
F
0.5°C to 150°C; 1°C above 150°C.
G
1°F to 300°F; 2°F above 300°F.

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
H
1°C to 300°C; 1.5°C above 300°C.
I
2°F to 570°F; 3°F above 570°F.
J
2°C to 260°C; 4°C above 260°C.
K
5°F to 500°F; 7°F above 500°F.
L
1°C to 260°C; 2°C above 260°C.
M
2.5°F to 500°F; 3.5°F above 500°F.

TABLE 4 Verification and Calibration TemperaturesA


--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Av Temp. Av Temp. Av Temp. Av Temp.


of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
Thermometer 1C Thermometer 1F Thermometer 2C Thermometer 2F
−20 to +150°C 0 to 302°F −5 to +300°C −20 to +580°F
−20°C 15°C 0°F 60°F 0°C 22°C 32°F 72°F
0°C 22°C 32°F 72°F 75°C 34°C 150°F 90°F
50°C 30°C 122°F 86°F 150°C 35°C 300°F 95°F
100°C 33°C 212°F 91°F 225°C 40°C 450°F 105°F
150°C 36°C 302°F 97°F 300°C 45°C 580°F 114°F
Thermometer 3C Thermometer 3F Thermometer 5C Thermometer 5F
−5 to +400°C 20 to 760°F −38 to +50°C −36 to +120°F
0°C 21°C 32°F 70°F −35°C 21°C −30°F 70°F
100°C 33°C 200°F 90°F 0°C 21°C 32°F 70°F
200°C 39°C 370°F 101°F 50°C 21°C 120°F 70°F
300°C 44°C 540°F 110°F
370°C 54°C 700°F 129°F
Thermometer 6C Thermometer 6F Thermometer 7C Thermometer 7F
−80 to +20°C −112 to +70°F −2 to +300°C 30 to 580°F

Copyright by ASTM Int'l (all rights reserved);


−70°C 21°C −94°F 70°F 0°C 32°F
−35°C 21°C −30°F 70°F 50°C 100°F
0°CB 21°C 32°FB 70°F 100°C 200°F
20°C 21°C 70°F 70°F 150°C 300°F
200°C 400°F
250°C 500°F
300°C 570°F
Thermometer 8C Thermometer 8F Thermometer 9C Thermometer 9F
−2 to +400°C 30 to 760°F −5 to +110°C 20 to 230°F
0°C 32°F 0°C 19°C 32°F 66°F
100°C 200°F 35°C 28°C 100°F 86°F
200°C 370°F 70°C 40°C 160°F 106°F
300°C 540°F 105°C 50°C 220°F 123°F
370°C 700°F
Thermometer 10C Thermometer 10F Thermometer 11C Thermometer 11F
90 to 370°C 200 to 700°F −6 to +400°C 20 to 760°F
100°C 61°C 212°F 141°F 0°C 18°C 32°F 64°F
200°C 71°C 390°F 159°F 100°C 44°C 200°F 110°F
300°C 87°C 570°F 189°F 200°C 64°C 370°F 144°F

Copyright ASTM International 52


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 4 Continued
Av Temp. Av Temp. Av Temp. Av Temp.
of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
370°C 104°C 700°F 220°F 300°C 91°C 540°F 187°F
370°C 115°C 700°F 240°F
Thermometer 12C Thermometer 12F Thermometer 13C Thermometer 14C
−20 to +102°C −5 to +215°F 155 to 170°C 38 to 82°C
−20°C −5°F 155°C 40°C 25°C
−10°C 15°F 163°C 50°C 25°C
0°C 32°F 170°C 60°C 25°C
10°C 60°F 70°C 25°C
20°C 85°F 80°C 25°C
30°C 110°F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
40°C 135°F
50°C 160°F
60°C 185°F
70°C 210°F
80°C
90°C
100°C

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Thermometer 14F Thermometer 15C Thermometer 15F Thermometer 16C
100 to 180°F −2 to +80°C 30 to 180°F 30 to 200°C
100°F 77°F 0°C 32°F 30°C
120°F 77°F 20°C 70°F 60°C
140°F 77°F 40°C 100°F 90°C
160°F 77°F 60°C 140°F 120°C
180°F 77°F 80°C 180°F 150°C
180°C
200°C
Thermometer 16F Thermometer 17C Thermometer 17F Thermometer 18C
85 to 392°F 19 to 27°C 66 to 80°F 34 to 42°C
90°F 21°C 70°F 38°C
140°F 25°C 77°F 41°C
190°F
240°F
290°F
340°F
390°F
Thermometer 18F Thermometer 19C Thermometer 19F Thermometer 20C
94 to 108°F 49 to 57°C 120 to 134°F 57 to 65°C
100°F 50°C 122°F 60°C
107°F 54°C 130°F 64°C
Thermometer 20F Thermometer 21C Thermometer 21F Thermometer 22C
134 to 148°F 79 to 87°C 174 to 188°F 95 to 103°C
140°F 82°C 180°F 99°C
147°F 86°C 187°F 102°C
Thermometer 22F Thermometer 23C Thermometer 24C Thermometer 25C
204 to 218°F 18 to 28°C 39 to 54°C 95 to 105°C

Copyright by ASTM Int'l (all rights reserved);


210°F 20°C 22°C 40°C 35°C 95°C 75°C
212°F 25°C 25°C 50°C 42°C 100°C 64°C
Thermometer 26C Thermometer 27C Thermometer 28C Thermometer 28F
130 to 140°C 147 to 182°C 36.6 to 39.4°C 97.5 to 102.5°F
130°C 155°C 25°C 0°C 32°F
135°C 165°C 25°C 37.8°C 100°F
140°C 175°C 25°C 39°C 102°F
Thermometer 29C Thermometer 29F Thermometer 30F Thermometer 33C
52.6 to 55.4°C 127.5 to 132.5°F 207.5 to 212.5°F −38 to +42°C
0°C 32°F 32°F −35°C 5°C
54.4°C 130°F 210°F −20°C 15°C
55°C 132°F 212°F 0°C 20°C
20°C 25°C
40°C 30°C
Thermometer 33F Thermometer 34C Thermometer 34F Thermometer 35C
−36.5 to +107.5°F 25 to 105°C 77 to 221°F 90 to 170°C
−31°F 41°F 25°C 25°C 77°F 77°F 100°C 70°C

Copyright ASTM International 53


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 4 Continued
Av Temp. Av Temp. Av Temp. Av Temp.
of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
−4°F 59°F 45°C 37°C 113°F 99°F 120°C 63°C
32°F 68°F 65°C 43°C 149°F 109°F 140°C 57°C
68°F 77°F 85°C 45°C 185°F 113°F 160°C 50°C
104°F 86°F 100°C 45°C 212°F 113°F 170°C 47°C
Thermometer 35F Thermometer 36C Thermometer 37C Thermometer 38C
194 to 338°F −2 to +68°C −2 to +52°C 24 to 78°C
212°F 158°F 0°C 25°C 0°C 25°C 25°C 25°C
250°F 145°F 15°C 25°C 15°C 25°C 40°C 25°C
285°F 134°F 30°C 25°C 30°C 25°C 55°C 25°C
320°F 122°F 45°C 25°C 50°C 25°C 75°C 25°C
338°F 116°F 65°C 25°C
Thermometer 39C Thermometer 40C Thermometer 41C Thermometer 42C
48 to 102°C 72 to 126°C 98 to 152°C 95 to 255°C
50°C 30°C 75°C 30°C 100°C 30°C 100°C 30°C
65°C 30°C 90°C 30°C 115°C 33°C 150°C 35°C
80°C 30°C 105°C 30°C 130°C 35°C 200°C 40°C

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
100°C 30°C 125°C 30°C 150°C 35°C 250°C 45°C
Thermometer 43C Thermometer 43F Thermometer 44C Thermometer 44F
−51.6 to −34°C −61 to −29°F 18.6 to 21.4°C 66.5 to 71.5°F
−50°C −60°F 0°C 32°F
−45°C −50°F 20°C 68°F
−40°C −40°F 21°C 70°F
−35°C −30°F
0°C + 32°F
Thermometer 45C Thermometer 45F Thermometer 46C Thermometer 46F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
23.6 to 26.4°C 74.5 to 79.5°F 48.6 to 51.4°C 119.5 to 124.5
0°C 32°F 0°C 32°F
25°C 77°F 50°C 122°F
26°C 79°F 51°C 124°F
Thermometer 47C Thermometer 47F Thermometer 48C Thermometer 48F
58.6 to 61.4°C 137.5 to 142.5°F 80.6 to 83.4°C 177.5 to 182.5°F
0°C 32°F 0°C 32°F
60°C 140°F 82.2°C 180°F
61°C 142°F 83°C 182°F
Thermometer 49C Thermometer 50F Thermometer 51F Thermometer 52C
20 to 70°C 54 to 101°F 69 to 116°F −10 to +5°C
20°C 25°C every 5° from 55°F every 5° from 70°F −10°C
35°C 25°C 0°C
50°C 25°C 5°C
70°C 25°C
Thermometer 54C Thermometer 54F Thermometer 56C Thermometer 56F
20 to 100°C 68 to 213°F 19 to 35°C 66 to 95°F
20°C 70°F every 2° from 19°C every 4° from 66°F and including
50°C 120°F 95°F

Copyright by ASTM Int'l (all rights reserved);


75°C 170°F
100°C 210°F
Thermometer 57C Thermometer 57F Thermometer 58C Thermometer 58F
−20 to +50°C −4 to +122°F −34 to +49°C −30 to +120°F
−20°C 25°C −3°F 77°F −30°C −20°F
0°C 25°C 32°F 77°F 0°C 32°F
25°C 25°C 77°F 77°F 25°C 80°F
50°C 25°C 122°F 77°F 45°C 120°F
Thermometer 59C Thermometer 59F Thermometer 60C Thermometer 60F
−18 to +82°C 0 to 180°F 77 to 260°C 170 to 500°F
0°C 32°F 100°C 212°F
25°C 80°F 175°C 350°F
55°C 130°F 255°C 490°F
80°C 180°F
Thermometer 61C Thermometer 61F Thermometer 62C Thermometer 62F
32 to 127°C 90 to 260°F −38 to +2°C −36 to +35°F
40°C 25°C 100°F 77°F −37°C −35°F

Copyright ASTM International 54


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 4 Continued
Av Temp. Av Temp. Av Temp. Av Temp.
of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
60°C 25°C 150°F 77°F −30°C −15°F
80°C 25°C 200°F 77°F −20°C 0°F
100°C 25°C 250°F 77°F −10°C 15°F
120°C 25°C 0°C 32°F
Thermometer 63C Thermometer 63F Thermometer 64C Thermometer 64F
−8 to +32°C 18 to 89°F 25 to 55°C 77 to 131°F
−7°C 20°F 0°C 32°F
0°C 32°F 25°C 80°F
10°C 50°F 35°C 95°F
20°C 70°F 45°C 115°F
30°C 88°F 55°C 130°F
Thermometer 65C Thermometer 65F Thermometer 66C Thermometer 66F
50 to 80°C 122 to 176°F 75 to 105°C 167 to 221°F
0°C 32°F 0°C 32°F
50°C 125°F 75°C 168°F
60°C 145°F 85°C 185°F

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
70°C 160°F 95°C 200°F
80°C 175°F 105°C 220°F
Thermometer 67C Thermometer 67F Thermometer 68C Thermometer 68F
95 to 155°C 203 to 311°F 145 to 205°C 293 to 401°F
0°C 32°F 0°C 32°F
100°C 205°F 150°C 300°F
110°C 240°F 170°C 340°F
130°C 275°F 190°C 370°F
150°C 310°F 205°C 400°F
Thermometer 69C Thermometer 69F Thermometer 70C Thermometer 70F
195 to 305°C 383 to 581°F 295 to 405°C 563 to 761°F
0°C 32°F 0°C 32°F
200°C 400°F 300°C 570°F
235°C 460°F 335°C 640°F
270°C 520°F 370°C 700°F
305°C 580°F 400°C 760°F
Thermometer 71C Thermometer 71F Thermometer 72C Thermometer 72F
−37 to +21°C −35 to +70°F −19.4 to −16.6°C −2.5 to +2.5°F
−35°C 21°C −30°F 70°F −19°C −2°F
−18°C 21°C 0°F 70°F −17.8°C 0°F
0°C 21°C 32°F 70°F 0°C 32°F
20°C 21°C 70°F 70°F
Thermometer 73C Thermometer 73F Thermometer 74C Thermometer 74F
−41.4 to −38.6°C −42.5 to −37.5°F −55.4 to −52.6°C −67.5 to −62.5°F
−41°C −42°F −55°C −67°F
−40°C −40°F −53.9°C −65°F
0°C 32°F 0°C 32°F
Thermometer 75F Thermometer 76F Thermometer 77F Thermometer 78F

Copyright by ASTM Int'l (all rights reserved);


−35 to +35°F −65 to +5°F 245 to 265°F 295 to 315°F
−35°F 55°F −65°F 25°F 250°F 300°F
0°F 65°F −30°F 55°F 260°F 310°F
32°F 75°F + 5°F 75°F
Thermometer 79F Thermometer 80F Thermometer 81F Thermometer 82C
345 to 365°F 395 to 415°F 445 to 465°F −15 to +105°C
350°F 400°F 450°F 0°C 71°C
360°F 410°F 460°F 50°C 71°C
100°C 71°C
Thermometer 82F Thermometer 83C Thermometer 83F Thermometer 84C
0 to 220°F 15 to 70°C 60 to 160°F 25 to 80°C
32°F 160°F 25°C 35°C 85°F 95°F 30°C 27°C
100°F 160°F 70°C 35°C 135°F 95°F 80°C 27°C
200°F 160°F
Thermometer 84F Thermometer 85C Thermometer 85F Thermometer 86C
75 to 175°F 40 to 150°C 100 to 300°F 95 to 175°C

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 55


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 4 Continued
Av Temp. Av Temp. Av Temp. Av Temp.
of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
100°F 80°F 50°C 54°C 150°F 130°F 100°C 99°C
150°F 80°F 150°C 54°C 250°F 130°F 175°C 99°C
Thermometer 86F Thermometer 87C Thermometer 87F Thermometer 88C
200 to 350°F 150 to 205°C 300 to 400°F 10 to 200°C
225°F 210°F 160°C 132°C 300°F 270°F 40°C 48°C
325°F 210°F 200°C 132°C 400°F 270°F 100°C 56°C
150°C 62°C
200°C 68°C
Thermometer 88F Thermometer 89C Thermometer 90C Thermometer 91C
50 to 392°F −20 to +10°C 0 to 30°C 20 to 50°C
110°F 119°F −20°C 15°C 0°C 20°C 20°C 25°C
212°F 132°F −10°C 15°C 10°C 20°C 30°C 25°C
300°F 143°F 0°C 15°C 20°C 20°C 40°C 25°C
392°F 154°F 10°C 15°C 30°C 20°C 50°C 25°C
Thermometer 92C Thermometer 93C Thermometer 94C Thermometer 95C

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
40 to 70°C 60 to 90°C 80 to 110°C 100 to 130°C
40°C 25°C 60°C 30°C 80°C 30°C 100°C 35°C
50°C 25°C 70°C 30°C 90°C 30°C 110°C 35°C
60°C 25°C 80°C 30°C 100°C 30°C 120°C 35°C
70°C 25°C 90°C 30°C 110°C 30°C 130°C 35°C
Thermometer 96C Thermometer 97C Thermometer 97F Thermometer 98C
120 to 150°C −18 to +49°C 0 to 120°F 16 to 82°C
120°C 35°C −15°C 0°F 20°C
130°C 35°C 0°C 32°F 40°C
140°C 35°C 20°C 70°F 60°C
150°C 35°C 45°C 110°F 80°C
Thermometer 98F Thermometer 99C Thermometer 99F Thermometer 100C
60 to 180°F −50 to +5°C −58 to +41°F 145 to 205°C
60°F −46°C −23°C −50°F −10°F 145°C 40°C
100°F −32°C −23°C −25°F −10°F 165°C 40°C
140°F −18°C −23°C 0°F −10°F 185°C 40°C
180°F 0°C −23°C 32°F −10°F 205°C 40°C
Thermometer 101C Thermometer 102C Thermometer 103C Thermometer 104C
195 to 305°C 123 to 177°C 148 to 202°C 173 to 227°C
200°C 40°C 125°C 35°C 150°C 35°C 175°C 34°C
250°C 40°C 140°C 35°C 165°C 35°C 190°C 38°C
300°C 40°C 155°C 35°C 180°C 35°C 205°C 40°C
175°C 35°C 200°C 35°C 225°C 40°C
Thermometer 105C Thermometer 106C Thermometer 107C Thermometer 108F
198 to 252°C 223 to 277°C 248 to 302°C 270 to 290°F
200°C 40°C 225°C 40°C 250°C 45°C 275°F
215°C 40°C 240°C 40°C 265°C 45°C 285°F
230°C 40°C 255°C 41°C 280°C 45°C
250°C 40°C 275°C 46°C 300°C 45°C

Copyright by ASTM Int'l (all rights reserved);


Thermometer 109F Thermometer 110C Thermometer 110F Thermometer 111C
320 to 340°F 133.6 to 136.4°C 272.5 to 277.5°F 170 to 250°C
325°F 0°C 32°F 170°C 35°C
335°F 135°C 275°F 200°C 40°C
136°C 277°F 250°C 45°C
Thermometer 112C Thermometer 113C Thermometer 113F Thermometer 114C
4 to 6°C −1 to +175°C 30 to 350°F −80 to +20°C
0°C 0°C 32°F −75°C
4°C 50°C 122°F −60°C
5°C 100°C 212°F −40°C
6°C 150°C 302°F 0°C
175°C 347°F
Thermometer 116C Thermometer 117C Thermometer 118C Thermometer 118F
18.9 to 25.1°C 23.9 to 30.1°C 28.6 to 31.4°C 83.5 to 88.5°F
every 1°C every 1°C 0°C 32°F
from 19.0°C from 24.0°C 30°C 86°F
31°C 88°F

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 56


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

TABLE 4 Continued
Av Temp. Av Temp. Av Temp. Av Temp.
of Emergent of Emergent of Emergent of Emergent
Temperature Temperature Temperature Temperature
Mercury Mercury Mercury Mercury
Column Column Column Column
Thermometer 119C Thermometer 119F Thermometer 120C Thermometer 121C
−38.3 to −30°C −37 to −22°F 38.6 to 41.4°C 98.6 to 101.4°C
−38°C 18°C −36°F 64°F 0°C 0°C
−30°C 20°C −22°F 68°F 40°C 100°C
0°C 0°C + 32°F 32°F 41°C 101°C
Thermometer 122C Thermometer 123C Thermometer 124C Thermometer 125C
−45 to −35°C −35 to −25°C −25 to −15°C −15 to −5°C
−45°C −35°C −25°C −15°C
−40°C −30°C −20°C −10°C
−35°C −25°C −15°C −5°C
Thermometer 126C Thermometer 126F Thermometer 127C Thermometer 128C
−27.4 to −24.6°C −17.5 to −12.5°F −21.4 to −18.6°C −1.4 to +1.4°C
−27°C −17°F −21°C 0°C
−26.1°C −15°F −20°C 1°C
0°C 32°F 0°C

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
Thermometer 128F Thermometer 129C Thermometer 129F Thermometer 130C
29.5 to 34.5°F 91.6 to 94.4°C 197.5 to 202.5°F −7 to +105°C
32°F 0°C 32°F 0°C
34°F 93.3°C 200°F 35°C
94°C 202°F 70°C
105°C
Thermometer 130F Thermometer 132C Thermometer 133C Thermometer 134C
20 to 220°F 148.6 to 151.4°C –38 to +2°C 144 to 156°C
32°F 0°C –36 25°C 145°C 40°C
100°F 150°C –30 25°C 150°C 40°C
160°F 151°C –24 25°C 155°C 40°C
220°F –18 25°C
–12 25°C
–6 25°C
0 25°C
Thermometer 135C Thermometer 135F Thermometer 136C Thermometer 136F
38 to 93°C 100 to 200°F –20 to +60°C –5 to +140°F
50°C 35°C 125°F 95°F –20°C –5°F
90°C 35°C 195°F 95°F –10°C 15°F
0°C 32°F
10°C 60°F
20°C 85°F
30°C 110°F
40°C 135°F
50°C
60°C
Thermometer 137C
80 to 100°C
80°C 30°C

Copyright by ASTM Int'l (all rights reserved);


90°C 30°C
100°C 30°C
A
For verification and calibration of total immersion thermometers see Test Method E 77.
For Thermometers 1C, 1F, 2C, 2F, 3C, 3F, 23C, 24C, and 25C the emergent column temperatures correspond on the average with those attained in verification using
equipment such as that described in Test Method E 77.
For all other thermometers listed, the emergent column temperatures are those attained when using the thermometers in the test equipment for which the thermometers
were originally designed. In some cases these temperatures are markedly different from those realized during verification. Also in some instances, such as Thermometers
35C and 35F, the values may not seem reasonable. Analysis of the factors affecting emergent column temperatures in use will provide the explanation for such apparent
inconsistencies.
B
This thermometer is especially sensitive to changes in emergent column temperatures because organic liquids are used as the thermometer fluid. As a means of
avoiding errors due to this cause the ice point should be taken at total immersion and the correction calculated for partial immersion by using the emergent stem correction
formula given in Test Method E 77.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 57


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
Part B—Enclosed-Scale Thermometer 20.1.4 The meniscus shall be distorted as little as possible
by striae or impurities in the glass.
16. Specifications
21. Vacuum
16.1 The thermometers shall conform to the detailed speci-
fications given in Table 5 and the requirements given in 21.1 The capillary above the mercury shall be evacuated,
Sections 17-24. and the vacuum shall be such that no difficulty is experienced
in rejoining the column after setting.
TABLE 5 Specifications for Enclosed Scale Thermometers
22. Construction
ASTM Number 115C–86
Range (nominal), °C 5 or 6 22.1 Shape—The thermometer shall be straight and have
Graduation interval, °C 0.01 insofar as practical circular cross sections.
Adjustable range, °C –10 to 120
Total length, max, mm 640
22.2 Top Finish—The top of the thermometer shall be
Length of main scale/° (graduated length), mm 44 6 4 closed by a suitable cap.
Distance from bottom of bulb to lowest 22.3 Scale—The graduated scale shall be of opal glass and
numbered graduation line, mm 200 to 240
Diameter of sheath (top portion) max, mm 16
shall be securely fastened to prevent relative displacement
External bulb diameter, max, mm 11 between the capillary and scale, and in such a way that it can
Bulb length, mm 30 to 50 freely expand in length.
Bulb shoulder length, max, mm 11
Bulk—dimension/bore ratio, max be/d = 15
22.4 Capillary Tube—The capillary tube shall be transpar-

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
(see 9.2.3) ent, its inside shall be smooth, and its cross-sectional area in
the scale portion shall not show variations from the average
17. Type greater than 5 %. The bore shall be large enough in relation to
17.1 The thermometers shall be of the enclosed-scale the dimensions of the thermometer bulb to ensure that (without
mercury-in-glass type with adjustable range. tapping), the jumping of the meniscus does not exceed one half
of the smallest scale division when the temperature is rising at
18. Temperature Scale a uniform rate not exceeding 0.05°C/min.
18.1 Temperature differences indicated by the thermometers NOTE 5—According to Hall and Leaver, this condition can be achieved
shall be in terms of the International Temperature Scale of or bettered by keeping the ratio be/d < 15.6 The external pressure
1990 (ITS-90)4 as adopted by the General Conference on coefficient is defined by the expression
Weights and Measures. be 5 0.52 De 2/~D e 2– Di 2! (1)

19. Immersion where:


19.1 To obtain the full precision of the thermometer for d = bore of the capillary, mm,
be = external pressure coefficient, m·deg C/cm,
difference measurements, the ambient (stem) temperature and
De = external diameter of the thermometer bulb, and
immersion depth must be the same in all measurements. Di = internal diameter of the thermometer bulb.
19.2 The thermometers are for use in a vertical position. 22.5 Adjusting Device—For adjusting the amount of mer-
19.3 In order to measure Celsius- or kelvin-scale tempera- cury in the bulb and main capillary to the intended range, an
tures after the setting temperature has been determined at a enlargement shall be provided above the measuring capillary to
given immersion depth by comparison with a standard, the serve as a reservoir for the separated mercury. Adjustment shall
thermometer must be used at the same immersion for all main be possible without difficulty. To facilitate adjustment of the
scale readings obtained at this setting. range an auxiliary scale may be provided. Two designs of
adjusting devices are indicated in Fig. 12. Operation of the
20. Glass
adjusting device shall require a temperature of approximately,

Copyright by ASTM Int'l (all rights reserved);


20.1 The bulb and sheath of the thermometer should pref- but not less than, 35°C above the setting point.
erably be made of the same type of thermometric glass. This 22.6 Dimensions—The dimensions shall be as given in
glass shall be selected so that the finished thermometer meets Table 5 and Fig. 12.
the following requirements: 22.7 Graduation and Figuring:
20.1.1 The bulb glass shall be stabilized by suitable heat 22.7.1 The range of graduation of the thermometer shall be
treatment to ensure that the accuracy requirements of Section as specified in Table 5.
23 can be met, 22.7.2 The graduation lines shall be clearly defined and of
20.1.2 Strain in the glass shall be reduced to a level uniform thickness which should in no case exceed 0.05 mm.
sufficient to minimize the possibility of fracture due to thermal The lines shall be at right angles to the axis of the thermometer.
or mechanical shock, 22.7.3 The thermometer shall be graduated and figured for
20.1.3 The accuracy of the reading shall not be impaired by use in a vertical position. The arrangement of the graduation
devitrifying during manufacture, and lines shall be as specified in Fig. 13. Each 0.1-deg line shall be
a long line; each 0.05-deg line shall have a length of two thirds
of the long line; and each 0.01-deg line shall have a length of
4
“The International Temperature Scale of 1990,” Amended Edition of 1990, one third of the long line. All graduation lines shall be readily
Metrologia, Vol 27, No. 1, 1990, pp. 3–15. visible from both sides of the capillary.
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 58


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
22.7.4 The scale shall be extended beyond the lowest long 23.2 The stability of the thermometer shall be such that the
graduation line by from two to five scale divisions. zero reading does not change by more than 0.01°C in 24 h
22.7.5 Figuring shall be as shown in Fig. 13. The lowest when left at the setting temperature after the bulb of the
long graduation line in upward reading thermometers, and the thermometer has been heated for at least 15 min to a tempera-
highest long graduation line in downward reading thermom- ture 35°C above the setting point.
eters shall be figured “0.” Each full degree shall be figured with
24. Inscriptions
large numbers. Every second intermediate one tenth of a degree
shall be figured with small numbers on the right-hand side of 24.1 The following inscriptions shall be permanently and
the scale. legibly marked on the thermometer:
24.1.1 Temperature scale (abbreviations such as “°C” or
23. Interval Error “C”),
24.1.2 Bulb glass type (the glass may be identified by a
23.1 The interval error of each partial interval of a range of colored stripe or stripes or by an inscription on the thermom-
0.5°C shall not exceed 60.01°C. The maximum error for any eter),
larger interval shall not exceed 0.02°C. These values refer to 24.1.3 Identification number (manufacturer’s serial num-
the adjustment of the thermometer to “0” at 20°C and at a ber),
temperature of the emergent stem of 20°C. 24.1.4 Manufacturer’s tradename or mark, and
NOTE 6—The first part of this requirement essentially requires that 24.1.5 ASTM 115C.

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
calibration be performed at 0.5°C intervals. The need for this is indicated
25. Keywords
by findings at the National Institute of Standards and Technology that
calibration every 40 to 50 divisions is required for reliable interpolation to 25.1 bulb; liquid-in-glass thermometers; standard specifica-
one or two tenths of a division. tion; stem; temperature; thermometer

Copyright by ASTM Int'l (all rights reserved);

FIG. 1 Test Gage for Checking Enlargements on Thermometers


9C, 9F, 10C, 10F, 57C, 57F, 88C, and 88F
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 59


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

FIG. 2 Bulb of ASTM Congealing Point Thermometers 54C and


54F

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
FIG. 3

FIG. 4

Copyright by ASTM Int'l (all rights reserved);


FIG. 5

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 60


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

FIG. 6

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
FIG. 7

FIG. 8

FIG. 9

Copyright by ASTM Int'l (all rights reserved);

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright ASTM International 61


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

FIG. 10

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
FIG. 11

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Copyright by ASTM Int'l (all rights reserved);

Copyright ASTM International 62


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05
--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
NOTE 1—All dimensions are in millimetres.
FIG. 12 Enclosed Scale Adjustable Range Thermometers

Copyright by ASTM Int'l (all rights reserved);


FIG. 13 Graduation and Figuring

Copyright ASTM International 63


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx
E 1 – 05

FIG. 14 Schemes for Graduation and Figuring

Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Dec 20 14:17:01 EST 2005
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

--```,```,,,````,,,,,,,```,,`,-`-`,,`,,`,`,,`---
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

Copyright by ASTM Int'l (all rights reserved);

Copyright ASTM International 64


Provided by IHS under license with ASTM Licensee=Total (Martinique)/5940911012, User=Largeteau, Christiane
No reproduction or networking permitted without license from IHS Not for Resale, 05/22/2006 09:01:29 MDT.jsx

You might also like