EcoRay HF-525 PLUS / SIG-40-525

Power ON switch

Electric power status lamp Power OFF switch

Normal: Yellow

2.2.4 EMERGENCY STOP


1. EMERGENCY STOP
(1) Safety devices for patient or operator to disconnect electric power when patient or operator are
placed in the electrical or mechanical emergency situation.
(2) Emergency stop switches are located at the top of the generator and top of the handle bar and if
push this switch and then disconnect electric power, and turn it clockwise electric power will be
supplied.

EMERGENCY STOP EMERGENCY STOP

2.2.5 Turn on the Power ON


1. Introduction of Power ON
(1) To operator the system, push the power ON switch of Console.
(2) After power on, system will diagnosing automatically during 15 seconds.

● Push the power ON switch

● After power ON, Standby lamp lighting with yellow and green.

In normal condition, green lamp lighting after 15 seconds later.

● After power on, system will diagnosing automatically during 15 seconds.

● Display progress at the mA window by 0 to 100.

● In the normal condition, 100 will be displayed at the mA window and could use

X-ray system.

● In abnormal condition, error will be displayed at the mA window before 100 and

then could not use X-ray system.

(Rev. 2.1) 33 Operation Manual


EcoRay HF-525 PLUS / SIG-40-525

2. Procedure of Power ON
(1) Push power ON switch of Console
(2) Standby lamp lighting with yellow and green if system is normal green lamp will be lighting.
(3) ”Set 0 r 1.00” will be displayed at the Console window and if system is normal condition,
“Set 100 r 1.00” display after 15 seconds later.
(4) “Set 100 r 1.00” displayed at the window, could use X-ray system.
(5) If abnormal condition, Error code will be displayed at the Console window and then could not
use X-ray system.

2.2.6 Check System Condition


1. In normal condition

(1) After power ON, check system condition automatically.

(2) After power ON, will be changed to the Radiography mode after
15 seconds later.

2. In abnormal condition

(1) After power ON, check system condition automatically.

(2) During system check condition, error code will be displayed


when system troubled.

(3) If error displayed, turn ON/OFF repeatedly, but error code not
appeared, system must be checked by authorized engineer.

2.2.7 How to Use


1. APR (Anatomical Programming Radiographic) function
(1) APR function will be activated automatically after power ON of console.
(2) Memorized APR data is saved by manufacturer’s experience but operator could change APR data
by need
(3) APR could use without specific procedure.
(4) APR is consists of 4 function switches.
(See next page images)

(Rev. 2.1) 34 Operation Manual


EcoRay HF-525 PLUS / SIG-40-525

1. Select Body size


3. Select Body point

2. Select Exposure method


4. Memory switch

(5) Select Body size

Thin Body size Auxiliary

Middle Body size Obese Body size

(6) Select Exposure method

A-P exposure method APR data memory switch

Lateral exposure method Oblique exposure method

(7) Select Body point


● Body point is consists of 21 anatomical mode and 3 types of exposure method.
● If push switch and then selected and lamp lighting.

2. Adjust exposure condition


(1) If change exposure data and memory it, old exposure data will be deleted and could not use it
again.

APR Function Adjust kV, mA, Time Memory

Select APR point


Memorized new data
(Body size, Exposure method,
Body point)

(2) Adjust Exposure value and display


● kV adjustment: Increase →No. 1, Decrease → No. 2

Fast adjustment → No. 1 + No. 2 or No. 3 + No. 2


1 2 3

● mA adjustment: Increase → No. 1, Decrease → No. 2

1 2
● mAs / Time select: If push No. 3 switch and then selected and selected switch will be
lighting.
● mAs / Time adjustment: Increase → No. 1, Decrease → No. 2
1 2 3