高压注射器1 CTA+维修手册
高压注射器1 CTA+维修手册
高压注射器1 CTA+维修手册
901214-001,04 2018-03
Proprietary Information Notice
This document contains information proprietary to Bracco Injeneering S.A.. All
Rights reserved.
No part of this document may be reproduced, transmitted, processed or recorded
by any means or form, electronic, mechanical, photographic or otherwise, nor be
released to any third party without the written consent of Bracco Injeneering S.A..
Copyright 2018 by Bracco Injeneering S.A..
The written and graphic product descriptions in this manual were effective at the
time of printing. Bracco Injeneering S.A. reserves the right to change
specifications and designs, or to discontinue any model without prior notification.
EmpowerCTA, EmpowerSync and IRiSCT are registered trademarks of Bracco
Injeneering S.A.
Fastload and EDA are trademarks of Bracco Injeneering S.A.
Manufacturer:
Distributor/Distributed by (USA)
Table of Contents
PROPRIETARY INFORMATION NOTICE .................................................................. II
READ THIS FIRST!...................................................................................................... 1
SECTION 1: INTRODUCTION .................................................................................. 4
1.1 PURPOSE OF THIS MANUAL ................................................................................. 4
1.2 AUDIENCE .......................................................................................................... 4
1.3 ORGANIZATION ................................................................................................... 4
1.4 RELATED DOCUMENTATION ................................................................................. 5
1.5 MANUAL CONVENTIONS....................................................................................... 5
SECTION 2: SYSTEM DESCRIPTION ..................................................................... 6
2.1 SYSTEM CONFIGURATION .................................................................................... 6
2.1.1 CTA+ Configurations ............................................................................... 7
2.1.1.1 CTA+ INJECTOR WITH REMOTE ................................................................ 7
2.2 MOUNTING SYSTEMS ........................................................................................ 10
2.3 TOOLS AND TEST EQUIPMENT ........................................................................... 10
2.3.1 Toolkit .................................................................................................... 10
2.3.2 Special Test Equipment ........................................................................ 10
2.3.3 Equipment for EDA Functional Testing ................................................. 10
2.4 PHILIPS SCANNER CONNECTION ........................................................................ 11
SECTION 3: PREVENTIVE MAINTENANCE ......................................................... 13
3.1 PREVENTIVE MAINTENANCE PROCEDURES ......................................................... 13
3.1.1 Preventive Maintenance ........................................................................ 13
3.1.2 Service ................................................................................................... 13
3.1.3 Preventive Maintenance Supplies ......................................................... 14
SECTION 4: TROUBLESHOOTING ....................................................................... 15
4.1 ORGANIZATION ................................................................................................. 15
4.2 PRELIMINARY DIAGNOSTIC ................................................................................ 16
4.3 POWER FAILURES ............................................................................................. 17
4.3.1 Injector Fails to Power Up ..................................................................... 17
4.4 COMMUNICATION MESSAGES............................................................................. 22
4.4.1 Incorrectly Configured Com Port Settings ............................................. 22
4.4.1.1 EDA Not Found Message ............................................................................ 22
4.4.2 Check Scanner Cable Message ............................................................ 23
4.5 FAULT CODES................................................................................................... 24
4.5.1 Fault Code Table ................................................................................... 25
4.5.2 Accessing Fault Codes .......................................................................... 38
4.6 OTHER FAULTS ................................................................................................. 38
4.6.1 Injector Head Troubleshooting – Heater Harness ................................. 38
4.6.2 EDA is Out of Range or Cannot Establish Baseline.............................. 39
4.6.3 Syringe Nest Does Not Open or Close ................................................. 40
4.6.4 False Door Message ............................................................................. 41
4.6.5 Syringe Warmer Does Not Warm .......................................................... 41
4.6.6 Contrast/Saline LEDs Malfunction......................................................... 43
4.6.6.1 LEDs Fail to Illuminate ................................................................................. 43
4.6.7 Pendant Switch Malfunction .................................................................. 44
4.6.8 Language Mismatch Failure .................................................................. 45
SECTION 5: REPAIRING AND REPLACING ........................................................ 46
5.1 ORGANIZATION ................................................................................................. 46
The following warnings refer to hazards that can cause serious injury or death. Read this
section carefully.
The sole purpose of the EmpowerCTA®+ Injector System is to
intravenously administer iodinated Contrast solutions followed by
Saline into human patients who are undergoing diagnostic exams for
Computed Tomography. It is not intended for any other use. DO NOT
use this equipment for any application other than its expressed
purpose.
When operating with the EmpowerCTA®+ Injector System, you must
put only Contrast in the Contrast syringe and Saline in the Saline
syringe. Failure to do so may lead to inadequate diagnostic results
and potential injury to the patient.
This equipment is not for use in chemotherapy, drug infusion, or any
other application for which it is not indicated.
The EmpowerCTA®+ Injector System must be used properly to
prevent the risk of an air embolism. Always fill the syringe with the
Injector pointing fully upward. When the syringe has been filled to the
desired volume, all the air should be purged from the syringe and
coiled tubing with the Injector still in the fully vertical position. Failure
to do so may lead to serious injury and /or death.
Never use the slow reverse, fast forward or fast reverse buttons on
the EmpowerCTA®+ Injector System (i.e., the buttons with downward
and upward arrows shown above them), nor the Replace Syringe
button on the Injector System, while the EmpowerCTA®+ Injector
System is connected to the patient.
In the event of a system malfunction, immediately turn off the
EmpowerCTA®+ Injector System and quickly disconnect the patient
from the unit. Confirm that the system malfunction has been repaired
before reconnecting the patient. Turn the system on, arm the Injector
and perform a test injection. If the system shows no further
malfunction, reconnect the patient. If a fault message still exists and
cannot be corrected and/or the Injector is still malfunctioning,
DISCONTINUE USE of the EmpowerCTA®+ Injector System until the
problem is properly identified and solved. Contact your regional
service provider for further assistance.
Exercise extreme caution when setting the flow rate on the Remote
Control and/or the Injector Controller so as not to create an
inadvertently high flow rate injection. PATIENT INJURY CAN
RESULT FROM HIGH FLOW RATE VENOUS INJECTIONS. Be
The following precautions refer to hazards that could result in minor injury or damage to
the CTA+ Injector system or other equipment. Read this section carefully.
Connect the EmpowerCTA®+ Injector System only to an electrical
source of the proper voltage and frequency as specified in Appendix
C of the user manual. If an incorrect voltage is used, the Injector
System may be damaged when it is turned on.
When retracting the syringe plunger with the EmpowerCTA®+ Injector
System during Contrast or Saline filling or after the end of an
injection, do not let a vacuum build in the syringe by leaving the
EmpowerCTA®+ Connecting Tube attached to the end of the syringe.
In addition do not occlude the J-Tube or patient coiled tubing to the
syringe. Failure to let the syringe properly vent when retracting the
syringe plunger can cause the plunger to recoil forward when the
Injector ram reaches the Replace Syringe position.
Dried Contrast may make removal of Luer fittings difficult. Do not let
Contrast dry. Make sure Luer connection is not over tightened.
Rx only (USA): Federal law (USA) restricts this device to sale,
distribution, and use by, or on the order of, a physician.
In the event of a loss of communications between the two systems,
the Injector system will go to Monitor mode. An acknowledgement
message will be displayed to denote the fault occurrence.
The use of accessory equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice
shall include:
Use of the accessory in the patient vicinity.
Evidence that the safety certification of the accessory has been
performed in accordance to the appropriate IEC/EN 60601-1-1
harmonized national standard.
Section 1: Introduction
1.1 PURPOSE OF THIS MANUAL
This service manual is intended for use with the EmpowerCTA®+ Injector System
and provides preventive inspection and service procedures to ensure proper
operation. In addition, this manual provides troubleshooting and repair
procedures to be performed by an authorized service technician, and describes
options and available software.
1.2 AUDIENCE
This manual is written for qualified technical personnel familiar with the operation
and design of the Empower Injector System and who have been authorized to
service the system.
1.3 ORGANIZATION
This manual is organized into the following sections:
The WARNING symbol alerts you to a hazard that may result in equipment damage or
personal injury.
75 Ft. Power
Communications 9 Ft. USB
Injector Cable Option Power Cable Remote
Head Supply Control
Acceptable Power Communications Cable lengths are 50ft, 50+12ft extension, or 75ft.
Longer lengths of cable will cause voltage drop and may cause system failure.
®+
Figure 2-2. EmpowerCTA Injector (Floor Mount Shown)
Service Keyboard
Connection
Service Keyboard
Connection
Service Keyboard
Connection
2.3.1 TOOLKIT
A toolkit is available from the manufacturer that contains specialized tools and
test fixtures to be used in conjunction with the service technicians commonly
available tools to perform preventative maintenance and repair to the Injector.
3.1.2 SERVICE
Preventive Maintenance and Service must be performed only by an authorized
technician who is trained in the use, operation, and servicing of the
EmpowerCTA®+ Injector System. Unauthorized service may result in personal
injury and damage to the system, and void of warranty.
Failure to observe cleaning instructions may result in equipment damage and create an
electrical hazard that can result in personal injury to the technologist and patient. When
®+
not in use, the EmpowerCTA Injector System should be stored in a safe place that
avoids the possibility of spills or other accidental damage. It should be placed in an area
away from dust, extreme or quickly changing temperatures. Follow proper personal
safety precautions when working around contaminated equipment or bodily fluids.
The Field Test Form must be completed immediately after system installation or when
®+
any service activity is performed on a EmpowerCTA Injector System component.
Section 4: Troubleshooting
4.1 ORGANIZATION
This section provides troubleshooting procedures for detecting and identifying
malfunctions in the EmpowerCTA®+ Injector System. It may be necessary to open
parts of a system for purposes of performing additional diagnostics or replacing
faulty components. Refer to Section 5: Repairing and Replacing for
disassembly instructions. Before troubleshooting and replacing components,
check all cable connections.
The troubleshooting procedures are divided into the following categories (see
Table 4-1):
Table 4-1: Troubleshooting Procedures
Troubleshooting
Description of Problem Section
Procedures
28VDC
GND
Pin 3, GND
Pin 4, GND
PIN 6, 28VDC
PIN 5, 28VDC
Acceptable Power Communications Cable lengths are 50ft, 50+12ft extension, or 75ft.
Additional lengths of cable will cause voltage drop and may cause system failure.
f. Reconnect Power Communications Cable
b. Check voltage between red wire (28 VDC, pin 4) and black wire
(GND, pin 1). See Figure 4-3
GND
28VDC (Brown)
(Red)
GND 28VDC
(Black)) (Yellow)
b. Turn on Injector
b. Turn on Injector
b. Turn on Injector
8. Check cables going through the Pivot Junction to the Controller Board
a. Individually disconnect cables passing from the Controller Board
to the Interconnect Board in the Injector housing
The Power Supply contains two fast blow 6.3 Amp, 250VAC fuses (5 x 20 mm). These
fuses are protected by a current limiting circuit. When current draw exceeds the
specified amount, the Power Supply will go into current limiting and shut down. It will
return to normal operation when it senses the current draw is within specification. Due
to the current limiting feature of the Power Supply, if a faulty fuse is found try replacing
the fuse. If the fuse fails again, replace the Power Supply.
a. Remove the outer Power Supply casing to access the fuses
located on the circuit board
This section applies only to systems that have the EDA installed. If the Injector
Controller and the Remote Control displays EDA Not Found, perform the
following:
1. Power the Injector Controller On and Off to clear the message
2. If cycling power does not clear the message, replace the EDA
Module. See Section 5.7.1: Replacing the EDA Module
3. If the message still occurs, replace the Injector to EDA Cable
4. If the message still occurs, replace the Single Board Computer/
Controller Circuit Board. See Section 5.5.2: Replacing the Single
Board Computer/Controller Circuit Board Assembly
When an EDA fault occurs (applies only to systems with the optional EDA),
the Injector Controller displays:
EDA INOPERABLE
FAULT No. nnn
The Remote Control displays messages in two areas. In the EDA area, it
displays EDA INOPERABLE; in the message area, it displays:
EDA INOPERABLE
FAULT No. nnn
Press EDA Softkey to Continue
Re-power Injector after Injection
or Call for Service
Table 4-2: Fault Codes lists all software fault codes displayed by the system and
provides repair actions for each code.
The software codes are identified as follows:
100 series - Remote
200 series – Injector
300 series – EDA (applies only to systems with the EDA)
If the Injector Controller or Remote Control displays a fault message but does not
display a numerical fault code, refer to Section 4.4 Communication Messages.
The Field Test Form must be completed immediately after system installation or when
®+
any service activity is performed on an EmpowerCTA Injector system component.
Fault
Definition Troubleshooting
No.
This error will most likely occur during a
software installation.
If there is a configuration file error (or the
file is corrupted) it will result in 202 fault.
The system can recover from this fault on
a power cycle. However, the calibration
files will get replaced by the default
values.
Fault
Definition Troubleshooting
No.
This fault will only occur during a power-
up due to a Hardware Voltage Error.
Fault
Definition Troubleshooting
No.
This fault will occur if the FPGA watchdog
circuitry is failing.
Fault
Definition Troubleshooting
No.
System will report Error-208 if any of
the following conditions are
encountered:
- There is a problem enabling the
saline/contrast motor
- Problem reading the FPGA
Revision
- Problem reading the ADC count
from the FPGA
- Enabling the watchdog, or
- Watchdog reset or watchdog set
threshold
- Problem configuring the FPGA
Quadrature
- Problem with the reading the
contrast/saline Quadrature
encoder
208 FPGA Function Failure - Problem setting the DAC Value
- Problem resetting a single bit to
the FPGA
- Problem reading the digital I/O
- FPGA watchdog cannot be kicked
Fault
Definition Troubleshooting
No.
Burn-In File contains information for the
software to perform Burn-In Test. Error-
211 Burn In File Corruption 211 can occur due to a corrupted Burn-in
File. Replace the USB Drive (600779-
001) to address this issue.
1. Check if a syringe plunger is
advanced beyond the replace syringe
position when the syringe door is
open. If it is, power off the injector and
manually retract the plunger to the
replace syringe position using the
hand knob, close the syringe door,
212 Door Open Fault and repower the injector.
2. Check for the magnet in the Door
housing.
3. Check for proper syringe nest door
operation
4. Clean Detent Rod Assembly
5. Replace Faulty Door Sensor
1. Confirm language on Injector and
Remote
2. Disconnect communications cable
from Remote Control
225 Language Mismatch Failure
3. Switch language on Remote to match
the Injector
4. Reconnect communications cable and
confirm error has been removed
This error can occur due to a failure in the
voltage to the Linear Potentiometer
Transducer. The Linear Potentiometer
has its own Power Supply on the
Controller board and an error in this
voltage can result in this error.
226 Linear Transducer Voltage Failure
1. Check the connections between the
Linear Potentiometer Transducer and
the Controller Board
2. See 5.5.2: Replacing the Single
Board Computer/Controller Circuit
Board Assembly
Fault
Definition Troubleshooting
No.
Software checks for motor movement at
the moment when the “ARM” button is
pressed.
Fault
Definition Troubleshooting
No.
This Error can occur due to the motor
encoder constant for Saline or Contrast
motor is out of the expected range.
Actions to be applied:
CALIB_MOTOR_ENCODER_FAILURE Note: Before replacing components do
231 the following
Motor encoder calibration failure a. Check if the linear potentiometers
are not contaminated with
contrast or saline media. Clean
those if required.
b. Check the Motor/Encoder and
Linear Potentiometers cables
between the Controller and
Interconnect Boards.
c. Recalibrate the Linear
Potentiometers.
Fault
Definition Troubleshooting
No.
Fault
Definition Troubleshooting
No.
This fault will only happen after a Power
Up. Voltage failure to any of the following
signals after Power Up can result in this
fault code:
i. 28V
ii. 5V Analog
iii. 12V
iv. 5V Digital
v. 3.3V
vi. 1.2V mixed
vii. 1.2V core
Fault
Definition Troubleshooting
No.
Fault
Definition Troubleshooting
No.
Fault
Definition Troubleshooting
No.
The EDA reports an internal hardware
310
fault, the failure of the Zener Diode 2.5V
The EDA reports an internal hardware
311 fault, the failure of the ROM CRC
background
The EDA reports an internal hardware
312
fault, the failure of the Digital voltage 5V
The EDA reports an internal hardware
313
fault, the failure of the Analog voltage 5V
During the EDA power up diagnostics,
314
the Microprocessor ROM failed
The EDA reports an internal hardware
315 fault, the failure of the Isolated voltage
3.28
During the ARM sequence, the system
316 reported that the EDA resistance (261
ohms) failed
During the ARM sequence, the system
317 reported that the EDA clip assembly is
broken or miswired
Loss of communication with the EDA.
318
EDA Crystal Timer failure
During the ARM sequence, the system
319
failed the Oscillator Off Test
During the ARM sequence, the system
320
failed K3 relay Function Test
During the ARM sequence, the system
321
relays failed to open
The Heater turns on when a Contrast syringe is loaded and initialized. It remains on for
one hour and then shuts off until a new syringe is initialized.
These messages show EDA status during normal operation, and do not necessarily
reflect system problems. A problem may exist only if the message is not cleared after
technologist intervention.
1. Power down Injector using switch on the Injector Controller and
unplug Power Supply cord from wall outlet
2. Ensure the following conditions are met and retest after each step:
a. There are no damaged parts of the EDA Clip (look for cracks and
frayed wiring)
3. If the problem still exists, do the following and retest after each step:
a. Clean the gold contacts on the EDA Clip using a cotton swab and
alcohol
It is possible to remove the bowed side-mount external retaining ring (E-Style) without
removing the Injector Actuator, but the side-mount external retaining ring is easier to
remove with the mechanism removed.
1. Remove the Injector Actuator (see Figure 4-6) from the Injector Head
(see 5.5.9: Replacing/Installing the EmpowerCTA®+ Actuator
Subassembly)
2. Inspect the Locking Detent Rod
3. Make certain that the bowed side-mount external retaining ring and
flat washer on the Locking Detent Rod are in place and undamaged
4. If necessary, replace the bowed side-mount external retaining ring
5. Otherwise, replace the Detent Rod (see Section 5.5.13: Replacing
EmpowerCTA®+ Locking Detent Rod Assembly)
The bowed side-mount external retaining ring is located on the Detent Rod between the
Anti-Rotation Bracket and the Motor and Bearing Mounting Plate. It is curved with the
concave side (curving inward) facing the Anti-Rotation Bracket.
The warmer works in a cyclical fashion, heating and cooling, with a certain amount of
hysteresis between these events. If you grip the warmer while it is in the “cooling”
phase, it may not feel warm.
If the Syringe Warmer fails to get warm, perform the following:
1. Power down the Injector Head
2. Disconnect the Syringe Warmer from the Injector Head. Use the
digital voltmeter to make the following resistance measurements on
the Syringe Warmer’s connector (see Figure 4-7)
a. The resistance between pin #3 and pin #4 should be between 9 to
12 kohms (thermistor)
b. The resistance between pin #1 and pin #2 should be 120 to
160 ohms (heater)
> If either or both measurements fail, replace the Syringe
Warmer
> If both measurements pass, proceed to the next step
Pin 2
Pin 1
Pin 4 Pin 3
The Field Test Form (FM0455) must be completed immediately after system installation
®+
or when any service activity is performed on an EmpowerCTA Injector system
component.
Each procedure is structured as follows:
Title – identifies the component to be replaced
Required Tools – identifies the tools needed for the procedure
Removal Procedure – step-by-step instructions for removing the
component
Replacement Procedure – step-by-step instructions for replacing the
component and any test requirements
Lethal voltages are present. Always power down the Injector and the Remote Control
and unplug their AC power cords before beginning disassembly. Failure to comply may
result in personal injury, death, or equipment damage.
Always power down the Injector and Remote Control before connecting or
disconnecting wires and cables.
Electrostatic Sensitive Devices are present. Wear a wrist grounding strap and place
components on an anti-static pad. Failure to comply may result in equipment damage
due to electrostatic discharge.
Table 5-1 identifies the repair and replacement procedures provided in the
following sections.
Table 5-1. Repair and Replace Procedures
The Pendant and Fan Cables can be reversed, because they fit either connector.
For correct installation, the Fan Cable is connected near the heat sink, and the
Pendant Cable is connected near the battery.
h. Connect the 2-pin connector (Pendant Port Cable Assembly) and
the 4-pin connector (Power Switch Assembly) to the back of the
circuit board (See Figure 5-3. Cable Connections to the
Controller Circuit Board)
Fan Connector J6
Power Switch
Connector P3
Speaker
Connector J2
Pendant Switch
Connector P7
4. Install Phillips screw, flat washer and lock washer to cable clamp that
is securing Injector Head Pigtail Cable using a #2 Phillips hex key
wrench and torque to 12 in-lb
5. Install phillips screw, flat washer and lock washer to cable clamp that
is securing EDA and ground wire
a. Tighten the PHP screws using a #2 Phillips Bit and torque to
12 in-lb.
6. Install the Injector Head Enclosure using a 3/32 hex key wrench and
torque to 64 in-ozs
7. Install Wire Hood (P/N 204148-001) using 4 screws (P/N 200142-
002)
a. Tighten the screws using a 5/64 hex key wrench and torque to
8 in-lb. (See Figure 5-4. Injector Controller Cable Cover)
c. If applicable, unplug the EDA Clip from the EDA module. Remove
three Phillips screws securing the bracket to the right side of the
Injector Controller (when facing Injector Controller)
Sudden release of the Spring J-arm may result in personal injury. When making height
limit, tension adjustments or removing the Injector, hold the Spring J-arm and gently
guide it to its highest position. If needed, adjust maximum height in order to get to a
comfortable working position.
When working with the Articulating J-arm, be aware of surroundings and inadvertent
contact with other equipment.
6. Remove the Injector J-arm with the Injector as follows:
a. Collar (A) on the Spring J-arm of the Articulating J-arm is secured
to sleeve (B) with screw (C) located opposite the semi-lunar
groove. Remove phillips screw phillips screw (C) (See Figure 5-6.
Articulating J-arm)
7. Remove three hole-plugs from the Injector then remove the Phillips
screws from those holes
8. Carefully remove the Injector from the Injector J-arm
The Injector Mounting J-arm has two keyways. Note which keyway was used for
mounting the Injector.
b. Insert the Injector’s cables into the short leg of the Injector J-arm
c. Continue feeding the cables. Pull the Injector Cable and its ground
wire/cable through the top hole of Injector J-arm and the EDA
Cable through the J-arm’s lower hole
b. Remove screw (C). Lower collar (A) back to its original position,
and secure it with the phillips screw phillips screw (C)
c. Verify that the Injector Mounting J-arm rotates freely within the
sleeve
b. Secure the Pendant Switch bracket to the left side of the Injector
Controller (when facing Injector Controller) with three Phillips
screws. Plug the Pendant Switch into the bottom of the Injector
Controller and rest it in the bracket
c. Secure the EDA Clip bracket to the right side of the Injector
Controller (when facing Injector Controller) with three Phillips
screws. Plug the EDA Clip into the EDA module and rest it in the
bracket
b. Secure the Pendant Switch bracket to the left side of the Injector
Controller (when facing Injector Controller) with three Phillips
screws. Plug the Pendant Switch into the bottom of the Injector
Controller and rest it in the bracket
c. Secure the EDA Clip bracket to the right side of the Injector
Controller (when facing Injector Controller) with three Phillips
screws. Plug the EDA Clip into the EDA module and rest it in the
bracket
2. Ensure that all casters on the floor stand base are locked
3. Lift shroud
4. Remove the cable’s two strain relief
5. Disconnect the Power Communications Cable from the Injector Cable
and push the pigtail into the pole
6. If necessary, remove the Heater, Pendant Switch and associated
bracket, and the EDA Clip and bracket (if applicable) as follows:
a. Unplug the Heater from the Injector Head. If it is mounted on a
syringe, remove the Heater and the syringe
4. Remove the Phillips screws from the Injector Head’s Rear Enclosure
(see Figure 5-11. Removing the Rear Enclosure)
4. Disconnect the 6-pin connector of the Injector Pigtail Cable from the
Controller Board and cable clamp of the Injector Pigtail Cable from the
Controller Assembly Mount
2. Push the end of the Injector Pigtail Cable with the metal connecter
down through the top of the Pivot Junction Assembly and pull it
through the bottom of the Pivot Junction Assembly
4. Insert the end of the Injector Pigtail Cable that has the metal
connector through the top of the pole or J-arm
6. Place the Injector Assembly onto the pole or J-arm while pushing the
Injector Pigtail Cable and EDA Cable down through the pole or J-arm
7. Pull the Injector Pigtail Cable out through the hole at the bottom of the
pole or through the hole on the side of the J-arm. The EDA Cable
comes out of the first hole
8. Connect the end of the Injector Pigtail Cable with the metal connector
to the metal connector on the Injector Power Communications Cable
When installing EDA Harness Cable and/or Injector Pigtail Cable, keep them both on one
side of the Pivot Junction, opposite all other cables. This will make it easier to remove
the cables.
When installing EDA Harness Cable and/or Injector Pigtail Cable, keep them both on one
side of the Pivot Junction, opposite all other cables. This will make it easier to remove
the cables.
3. About 4 inches from the right end, secure a cable tie (202403-001) to
cause the ribbon cables to wrap around the motor and heater cables.
Cut off excess tie. Do not include 301004-001 (See Figure 5-15.
Cables thru Pivot Junction)
5. Install the Injector Assembly to the Pole or J-arm and slide the Injector
Pigtail Cable and EDA Harness Cable through the Pivot Junction
Assembly
6. If Articulating Arm, follow Section 5.3.4: Installing Injector to J-
Arm, or If Floor Stand, follow Section 5.3.5: Installing Injector to
Floor Stand
Fully rotate Injector to the left and right while ensuring all interconnect cables do not
become disconnected. Verify interconnect cables do not bind causing stress on the
connectors.
The Field Test Form (FM0455) must be completed immediately after system installation
®+
or when any service activity is performed on an EmpowerCTA Injector system
component.
®+
Touchscreen Calibration is required when the EmpowerCTA Injector System detects a
Touchscreen Calibration Failure (Fault #230), LCD Touchscreen replacement and/or
replacement of a Single Board Computer/Controller Circuit Board Assembly.
The following procedures are for removing and installing the LCD Touchscreen
from the CTA+ Controller circuit board.
9. Gently lift the single board computer/controller board away from the
screen retainer while unplugging the Touchscreen ribbon cable from
J10 (See Figure 5-19. Single Board Computer)
10. Remove the four BHCS screws (200142-001) holding the Screen
Retainer (204144-002) onto the Front Housing with a 5/64 hex key
wrench
11. Gently lift the screen retainer away from the Front Housing while
carefully sliding the 18-wire LVDS Cable (P/N 301317-001) and USB
PCA (301333-001) cable from the opening on the screen retainer.
(See Figure 5-20. Screen Retainer)
5. Connect the Video Cable back to the connector by sliding the cable
into the slots of the ferrites
6. Attach the 18 wire LVDS Video Cable (301317-001)
7. Apply one gasket (P/N 301423-001) starting with the piece with a
notch taken out of it. This is applied near the 18 wire LVDS cable.
Remove plastic backing from the adhesive one section at a time (See
MI Gasket)
8. Make sure screen gasket (206715-001) is laid flat in its groove and
clean (See Figure 5-23. Screen Gasket)
10. Route the 18 wire LVDS cable and USB cable through the openings in
the Screen Retainer (P/N 204144-002)
11. Install Screen Retainer onto the front housing, using 4 screws (P/N
200142-001)
a. Tighten screws using a 5/64 hex key wrench and torque to 8
in-lb.
12. If required, install USB circuit board (P/N 301333-001) to bottom of
controller front housing using 2 screws (P/N 200140-001)
a. Tighten screws using a 1/16 hex key wrench and torque to 64
in-oz.
13. While placing the single board computer/controller board in the screen
retainer,
a. Connect the LVDS cable to J11
b. Connect the Touchscreen cable to J10
c. Connect the USB Cable to P7
d. Connect the 2-wire LCD Screen Cable to J7. Route cable
underneath board
The Single Board Computer is mounted on the Controller circuit board. When field
servicing, both boards must be replaced. Do not attempt to separate them; they must be
replaced as a set.
When replacing the Single Board Computer and Controller Circuit Board (701320-002),
the System Software will need to be programmed using USB Memory Stick 701503-001.
See Section 6.2: Upgrading Injector Software for instructions.
®+
Touchscreen Calibration is required when the EmpowerCTA Injector System detects a
Touchscreen Calibration Failure (Fault #230), LCD Touchscreen replacement and/or
replacement of a Single Board Computer/Controller Circuit Board Assembly.
Attempting to separate the Single Board Computer and Controller circuit board may
result in damage to both boards.
The Single Board Computer is programmed with the pressure profile data (m and b
constants) that are unique to each system. Before removing the Single Board Computer,
retrieve the pressure profile from the current Single Board Computer and use them later
to program into the new Single Board Computer using the procedure in Section 7.5:
Retrieving and Programming Pressure Profile Constants
.
After replacing the Single Board Computer/Controller Circuit Board Assembly, it is
necessary to recalibrate both Linear Potentiometers. See Section 5.5.12
Figure 5-27. Fan Speaker Power Switch & Pendant Cable Connections
2. Uninstall Fan (P/N 301348-001) from EMI shield (P/N 204145-001) by
removing the 4 screws P/N 200140-002 with 1/16 hex key wrench
(See Figure 5-28. Fan and Speaker)
The thread locking fluid should be allowed to penetrate the threads for at least 1 minute
before removing the excess with a cotton swab.
4. Follow Section 5.3.2: Injector Controller Assembly procedure
5. Verify fan is running quietly, adjust fan mounting screws if required
The thread locking fluid should be allowed to penetrate the threads for at least (1)
minute before removing the excess with a cotton swab.
4. Connect ground cable (301374-001), and route connectors from rear
housing through opening in EMI shield near the speaker
5. Attach ground cable to threaded hole between fan and speaker
6. Follow Section 5.3.2: Injector Controller Assembly procedure
1. Insert the wires of the Power Switch Assembly into the Injector
Controller’s Rear Enclosure through the enclosure’s opening (see
Figure 5-30. Power Switch)
Orientation of switch is critical. Orient the switch so that the “O” is on the bottom as
displayed above.
2. Push the switch into the opening until 2 clicks are heard
3. Thread the Pendant Switch, Power Supply Cable and Ground Cable
through the slot in the EMI Shield.
4. Reattach the EMI Shield and Rear Enclosure with 4 screws using 5/64
hex key wrench to Injector
5. Reattach Ground Wire to the thread hole between fan and speaker
using a 5/64 hex key wrench
6. Follow Section 5.3.2: Injector Controller Assembly procedure
There is a sealing washer that is part of the assembly and cannot be removed.
1. Slide the EmpowerCTA®+ Actuator Assembly into the top cavity of the
Front Enclosure
2. Carefully thread the Door Sensor Cable under the Interconnect Board.
3. Thread Contrast Potentiometer under the Interconnect Board, and
remove Interconnect Board, if necessary. (See Figure 5-34. Contrast
Door Cable & Potentiometer Cable)
®+
Figure 5-35. EmpowerCTA Actuator Assembly
5. Perform the following Interconnect Board connections:
a. Plug the Contrast LED Cable connector into connector marked
Contrast LED J9
b. Plug the Contrast Door Sensor Cable into connector marked
Contrast Door J11
c. Plug the Contrast Motor Cable into connector marked Motor &
Encoder Contrast J1
d. Plug the Contrast Linear Potentiometer Cable into connector
marked Linear Position Contrast J5
6. Connect the braided Ground Cable (301002-001) to the Contrast Door
Assembly using 1 screw (202404-001) and 1 external tooth lock
washer (202510-001). Tighten using a #2 Phillips screwdriver and
torque to 12 in-lb. (See Figure 5-36. Ground/Door Cable
Connections)
7. Secure cables with nylon cable ties and trim excess length (see
Figure 5-46. Cutting Interconnect Circuit Board Cable Ties)
1. Position the Motor in the Injector assembly so that the timing belt
reaches around the Motor Pulleys (motor shaft is offset)
2. Use the 5/32" hex key wrench to secure the Motor with two socket
head cap screws (see Figure 5-40. Motor – Removal and
Replacement) torque to 40 in-lb.
3. Clean shaft with isopropyl alcohol
4. Place Loctite Adhesive 401(500081-001) flats of the motor shaft
5. Quickly install timing pulley onto motor shaft
6. Use the external snap ring pliers to install the external retaining ring
7. Install the Timing Belt by placing it on the small motor pulley and
gently sliding it onto the large pulley while turning the hand knob
8. Plug the Motor & Encoder Cable (P/N 301323-001) into the
Interconnect circuit board connector marked J1 (Contrast) or J2
(Saline) and put tie wrap on while cutting any excess
9. Follow Section 5.3.8: Installing the Injector Head’s Rear
Enclosure procedure
10. Set the Pressure Limiting Constants (see Section 7.4: Setting
Pressure Profile Constants)
b. Pull out the Linear Potentiometer Cable Assembly. It has one flat
washer and one spring washer (see Figure 5-41. Removing
Snap Ring / Linear Potentiometer Cable Assembly)
Reversing the Wiper orientation will prevent the linear pot from being properly inserted
and will cause equipment damage.
b. Use the internal snap ring pliers to install the snap ring that
secures the Wiper in the Anti-Rotation Bracket. Make certain the
snap ring is properly seated inside the groove in the Anti-Rotation
Bracket
b. Use the internal snap ring pliers to install the retaining ring on top
of the Motor and Bearing Mounting Plate (see Figure 5-41.
Removing Snap Ring / Linear Potentiometer Cable Assembly)
Perform the Linear Potentiometer Adjustment only on the Linear Potentiometer that has
been replaced.
The Linear Potentiometer is removed before removing the Motor and Bearing Mounting
Plate as a precaution to avoid damaging the Linear Potentiometer.
4. Remove 3 socket head screws and 3 lock washers holding the Motor
and Bearing Mounting Plate from the Actuator sub-assembly with a
5/32" hex key wrench
5. Remove snap ring and washer from end of Detent Rod Assembly with
needle nose
6. Loosen, but do not remove the two screws on the anti-rotation bracket
that is on the ball screw assembly using a 1/8" hex key wrench
7. Remove the 2 screws holding the Detent Rod in the Plastic Linear
Sensor Support Mount using a 7/64 hex wrench. (See Figure 5-45.
Plastic Linear Sensor Support Mount)
b. Insert the Locking Detent into the bore in the Door Housing
Assembly and secure the Sensor Support Mount with two socket
head cap screws and split lock washers. Tighten the screws with
the 7/64" hex key wrench to 65 in-oz.
c. Install the flat side-mount external retaining ring, flat washer, and
compression spring on the Locking Detent Rod
2. Install the Ball Screw Assembly. Align the small hole in the Anti-
Rotation Bracket with the Locking Detent Rod and slide the Ball
Screw Assembly into the Door Housing Assembly
3. Install the flat washer, and then the bowed side-mount external
retaining ring on the Locking Detent Rod. The bowed side-mount
external retaining ring must fit into the groove on the rod, and its
concave side (inward curve) must face the Anti-Rotation Bracket
4. Install the Linear Potentiometer (see Section 5.5.12.2: Installing the
Linear Potentiometer)
5. Secure the Motor and Bearing Plate using 3 Socket Head Screws and
3 Lock Washers (202431-001). Tighten using a 5/32 Hex Bit to 40 in-
lbs.
6. Install the Ball Screw Timing Pulley as follows:
a. Hold onto the threaded shaft of the Ball Screw Assembly and pull
into the motor mount as far as possible
b. Position the pulley hub to face the Motor and Bearing Plate, and
slide the pulley over the shaft. The pulley hub will fit into the
opening in the plate. The groove in the shaft must be visible on
the opposite side of the pulley (see Figure 5-44. Removing the
Pulley)
c. Slide the washer over the shaft and screw lock nut onto the shaft
(see Figure 5-44. Removing the Pulley)
d. Use the strap wrench to hold the Ball Screw Pulley in position
while tightening the lock nut with the 7/16" open end wrench to 30
in-lbs. Use caution so as not to bend the shaft
3. Clean any excess Loctite 401 from lens, and allow it to dry before
proceeding
4. Seal along the edges of LED assembly to form a moisture-proof seal
between the LED and the Front Enclosure
5. Install the Interconnect circuit board (see Section 5.5.14: Replacing
the EmpowerCTA®+ Interconnect Circuit Board)
6. Install the Actuator Assembly (see Section 5.5.9:
Replacing/Installing the EmpowerCTA®+ Actuator Subassembly)
7. Install the Injector Head’s Rear Enclosure (see Section 5.3.8:
Installing the Injector Head’s Rear Enclosure)
The Field Test Form (FM0455) must be completed immediately after system installation
®+
or when any service activity is performed on a EmpowerCTA Injector system
component.
Use ESD handling practices any time static sensitive components are present.
Use ESD handling practices any time static sensitive components are present.
5. Press the Administrator icon on the Windows Main screen, and type
the administrator password, then press ENTER. See Figure 5-55.
Windows Main Screen
4. Press the Computer Name tab. Press the Change button. See
Figure 5-57. System Properties Window
Change window to restart the computer and save the changes. See
Figure 5-59. Window Sequence to Restart Computer and Save
Changes
Double click on the icon, and wait for the software to start. Excessive clicking on the
icon may cause the system to behave abnormally when it returns to the Stop screen.
4. Select Setup on the EmpowerCTA®+ Stop screen, then select
Shutdown on the Setup screen.
5. Turn Controller back on
The Field Test Form (FM0455) must be completed immediately after system installation
®+
or when any service activity is performed on a EmpowerCTA Injector System
component.
1. Locate the EDA harness and ground wire/cable inside the pole (floor
mounting) or the Injector J-arm (overhead mounting) and pull out a
few inches
2. Pull the connector through the EDA grommet, and insert into the
pole’s/J-arm’s opening
3. Remove the phillips screw phillips screw and lock washer from the
ground post in back of the EDA module
4. Place the ground wire/cable lug on the module’s ground post. Replace
the lock washer and phillips screw phillips screw and tighten
5. Connect the EDA’s harness to the back of the EDA module above the
ground post
6. Feed the excess EDA harness and ground wire/cable back into the
pole/J-arm until the module is flush with the pole/J-arm. Ensure EDA
module is mounted so that the cable connector is on the bottom
7. Attach the EDA module to the Injector J-arm or pole using the module
bracket and 4 Phillips screws (see Figure 5-62. Dismounting the
EDA)
When installing the Power Supply, ensure the following: (1) The Power Communications
Cable can be routed from the Injector to the Power Supply. (2) The USB cable can be
routed from the Remote Control to the Power Supply. (3) The Power Supply’s power
cord (3 meters/approximately 9 ft.) can reach the wall outlet.
8. Remove the Injector Actuator and place on a clean work area (see
Section 5.5.9: Replacing/Installing the EmpowerCTA®+ Actuator
Subassembly)
9. Remove the CTA+ Interconnect Circuit board (See 5.5.14.1:
Removing the CTA+ Interconnect Circuit Board)
10. Remove the Cables through the Pivot Junction assembly
11. Remove the friction retainer plug using #14 Spanner Driver. (See
Figure 5-65. Friction Retainer Plug)
12. With the retainer plug removed, remove the tapered spring and friction
plug. Discard the retainer plug, tapered spring, and friction plug. (See
Figure 5-66. Friction Plug)
13. Remove the pivot insert cap with a #14 spanner to access the locknut.
(See Figure 5-67. Pivot Insert Cap)
14. Remove the locknut with a 7/16” socket wrench. Remove and discard
the nylon washer and pivot insert. The Pivot Junction is now loose
and can be removed. (See Figure 5-68. Pivot Junction Locknut)
16. Remove the nylon washer from the Injector Pivot and discard it. (See
Figure 5-70. Nylon Washer for Injector Pivot)
10. . Secure EMI shield and Rear Housing to Mounting Block using 4
(200142-003). 5/64 Hex to 8 in-lbs. (See Figure 5-80. EMI Shield
and Rear Housing)
13. Route cables thru the Pivot Junction. Follow 5.4.3: Removing
Cables That Go Through The Pivot Junction Procedure
14. Install EDA Harness Cable. Follow 5.4.2: Removing and Replacing
EDA Harness Cable Procedure
15. Install Injector Pigtail Cable. Follow 5.4.1.2: Replacing the Injector
Pigtail Cable: Procedure
16. Install the Injector Head’s Rear Enclosure (see 5.3.8: Installing the
Injector Head’s Rear Enclosure)
17. Follow 5.3.2: Injector Controller Assembly procedure
19. If Applicable, Install EDA (See 5.7.1.2: Installing the EDA Module)
21. Make sure the Injector Head can rotate side to side with some
resistance. It should not drop freely when rotated.
Do not reboot or restart any equipment after downloading the event log and removing
the memory stick.
The memory stick must be non-bootable. Reformatting a bootable memory stick does
not work.
7. Connect the JTAG Cable from Xilinx Platform Cable USB II to the J5
connector on controller board (See Figure 6-6. JTAG Cable
Connection)
9. Connect USB cable from Xilinx Platform Cable USB II to the laptop
computer
10. Verify the LED on the Xilinx Platform Cable USB II is orange. (See
Figure 6-7. Orange LED – Xilinx Platform Cable USB II)
11. Plug Power Supply cord back to the wall outlet and Power up the
injector using switch on the Injector Controller
12. Wait for Injector to cycle to the main screen
13. Verify the LED on the Xilinx Platform Cable USB II is green. If not,
verify connection to J5 and the laptop computer. If connection issue
cannot be resolved, the system controller board cannot be
programmed. (See Figure 6-8. Green LED – Xilinx Platform Cable
USB II)
16. An “Error Message” may appear when the Project File is opened for
the first time. Click on OK. (See Figure 6-10. Error Message)
17. Expand “Boundary Scan” under “iMPACT Flows” and verify under
Boundary Scan <File Name>.bit is displayed
where <File Name> is the current released software file name. e.g.:
400160-001_Rev01.bit (See Figure 6-11. Boundary Scan)
18. Verify IPF file is correctly by letting the mouse pointer hover over the
FLASH icon. The Text Box pop-up should contain the following: (See
Figure 6-12. IPF File Verification)
Part: W25Q80BV
File: <File Name>.mcs
where <File Name> is the current released software file name. e.g.:
400160-001_Rev01.mcs
Type: SPI PROM
19. Right click on the FLASH icon then select Edit Attached Flash
Properties and change Data Width: to 4 (if not already set to 4) and
then Click OK (See Figure 6-13. Edit Flash Properties)
20. Double click on the “Erase” line located in the “iMPACT Processes”
window. (See Figure 6-14. Erase)
21. Injector will produce a continuous beep sound till this operation is
successful and may display an error (e.g. “SYSTEM FAULT NO. 234”
or “SYSTEM FAULT NO. 208”) on the Touch Screen
22. Wait for the erase function to finish and verify the “Erase Succeeded”
message appears. The message box will be blue for a successful
erase. (See Figure 6-15. Erase Succeeded)
23. The continuous beep will stop after the Operation is successful but the
Touch Screen will continue to display error (e.g. “SYSTEM FAULT
NO. 234” or “SYSTEM FAULT NO. 208”)
24. Double click on the Program line located in the “iMPACT Processes”
window. (See Figure 6-16. Program)
25. Injector will produce a continuous beep sound till this operation is
successful and will continue to display error (e.g. “SYSTEM FAULT
NO. 234” or “SYSTEM FAULT NO. 208”) on the Touch Screen
26. Wait for the programming function to finish and verify the
Programming Succeeded message appears. The message box will be
blue for successful programming. (See Figure 6-17. Program
Succeeded)
27. The continuous beep will stop after the Operation is successful but the
Touch Screen will continue to display error (e.g. “SYSTEM FAULT
NO. 234” or “SYSTEM FAULT NO. 208”)
28. Verify the correct program has been flashed into the Xilinx by double
clicking on the ‘Get Device Checksum’ line located in the iMPACT
Processes window. (See Figure 6-18. Get Device Checksum)
29. Injector will produce a continuous beep and will continue to display
error (e.g. “SYSTEM FAULT NO. 234” or “SYSTEM FAULT NO. 208”)
on the Touch Screen
30. Wait for the checksum process to complete (about 1 minute) and
verify the Checksum Succeeded message appears. The message box
will be blue for successful Checksum. (See Figure 6-19. Checksum
Succeeded)
31. Injector will continue to produce a continuous beep and will continue
to display error (e.g. “SYSTEM FAULT NO. 234” or “SYSTEM FAULT
NO. 208”) on the Touch Screen
32. Click in the Console window and scroll to the bottom of the Console
log file. (See Figure 6-20. Console Log)
38. Install Wire Hood (P/N 204148-001) using 4 screws (P/N 200142-
002). Tighten the screws using a 5/64 hex key wrench and torque to 8
in-lb
3. After Remote Control has booted up and is displaying the Stop Mode
Main Screen, press the SETUP button.
12. Delete the existing remote.mdb and logging.mdb from the directory.
13. Paste the file remote.mdb and logging.mdb to the folder from the
memory stick.
20. Verify the previous custom injection programs are available under the
program tab.
21. Shut down remote within the Empower Application and restart the
remote to prevent rebooting to Windows.
Do not allow the Plunger Head to advance above this level. Failure to stop the
Injector Plunger may result in equipment damage.
2. At the Remote Control, press the SETS 0 ML button. The button will
briefly disappear while the linear potentiometer position is programed
into the memory of the single board computer. Wait for the button to
reappear before continuing
Setting the Replace Syringe Position:
1. Set the Replace Syringe position as follows:
a. Press and hold the FAST RETRACT key on the Injector
Controller.
b. Watch as the syringe plunger retracts until it stops automatically in
down position.
Do not allow the Plunger Head to retract in the bottommost position longer than 2
seconds. Failure to stop the Injector Plunger may result in equipment damage.
c. At the Remote Control, press the SETS REPLACE SYRINGE
button. The button will briefly disappear while the linear
potentiometer position is programed into the memory of the single
board computer. Wait for the button to reappear before continuing
2. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed (see Figure 6-1. Stop Mode Main)
The following procedures are used to set the m and b profile constants when they are
unknown. A test fixture with a calibrated pressure gauge is required to calculate the
pressure profile constants.
The constants are trim values. The b constant is used to adjust the LOW
pressure value displayed on the Injector display to match the fluid pressure
measured by the calibrated pressure gauge on the syringe test fixture. The m
constant is for the HIGH pressure value.
There is a m and b value for the Contrast side of the Injector head and a
independent m and b value for the Saline side of the Injector. The two sides of
the Injector do not interact. The constant equates motor current to injection
pressure thru the lead screw assembly.
The default value for both the m and b constants is 120. Typical values for the m
and b constants are between 105 and 135, but may be outside this range.
The acceptable pressure reading is 10psi plus 10% of the measured pressure
value. For Instance:
At 50psi on the calibrated gauge, the Injector could display between
35 and 65psi.
At 150 psi, the Injector could display between 125 and 175psi.
At 325 psi, the Injector could display between 283 and 367psi
Always try to adjust the m and b values to get the displayed pressure value as
close as possible to the calibrated pressure gauges.
It is recommended to use newer syringes for calibration testing for some models.
Protective eye-wear should be worn for all pressure testing.
DO NOT over-tighten the thumbscrew on the bracket around the syringe. A slight
pressure is all that is required to keep the fixture in place. Too much pressure can
distort the sidewalls of the syringe and is a potential safety hazard. Not all fixtures have
a thumb screw. Newer fixtures slip freely over the syringe.
Carefully turn the syringe to engage the luer fitting on the test fixture; do not strip the
syringe threads.
b. Turn on power to the Injector Controller and Remote
2. Do the following:
a. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
b. Turn the Injector Head upright and refill the syringe with a J-tube
and water bottle. DO NOT try to refill through the coiled tubing or
with the needle valve partially closed
3. Connect the coiled tubing to the end of Syringe Pressure Test Fixture.
4. In the Remote Control, select Phase 1 and enter the following
parameters:
5. Phase No./Media = 1C, Flow Rate = 1.0 ml/sec, Volume = 150 ml
6. Pressure Limit = 50 psi
7. Tilt the Injector Head downward into the Run position and press ARM
and RUN on the Remote Control
8. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 50 psi. Verify that the Injector display reading is
between 35 and 65 psi. If the pressure is within the specified limits,
skip to Step 10
9. If the pressure is outside the specified limits, perform the following:
a. Press the SETUP button on the screen. The Setup Screen is
displayed
h. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed
i. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
j. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done.
c. Tilt the Injector Head downward into the Run position and press
ARM and RUN on the Remote Control
14. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 325 psi. Verify that the Injector display reading is
between 283 and 367 psi. If the pressure is within the specified limits,
skip to Step 16
15. If the pressure is outside the specified limits, perform the following:
a. Press the SETUP button on the screen. The Setup Screen is
displayed
h. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
c. Tilt the Injector Head downward into the Run position and press
ARM and RUN on the Remote Control
20. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 150 psi. Verify that the Injector display reading is
between 125 and 175 psi. If the pressure is within the specified limits,
skip to Step 22
21. If the pressure is outside the specified limits, perform the following:
a. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
22. Remove the Syringe Pressure Test Fixture and coiled tubing from the
Contrast Injector and proceed to setting the constants for the Saline
Injector in Section 7.5: Retrieving and Programming Pressure
Profile Constants. Record new m & b values on Field Test Form
comments field
1. Do the following:
a. Install the Syringe Pressure Test Fixture 600414-001 onto new
Saline syringe in the Injector. Turn on power to the Injector
Controller and Remote Control
It is recommended to use newer syringes for calibration testing for some models.
Protective eye-wear should be worn for all pressure testing.
DO NOT over-tighten the thumbscrew on the bracket around the syringe. A slight
pressure is all that is required to keep the fixture in place. Too much pressure can
distort the sidewalls of the syringe and is a potential safety hazard. Not all fixtures have
a thumb screw. Newer fixtures slip freely over the syringe.
When turning the syringe to meet with the luer fitting on the test fixture, care should be
taken to not strip the syringe threads.
b. If necessary, turn on power to the Injector Controller and Remote
2. Do the following:
a. Press the CONTRAST on the Injector Controller and Initialize the
Contrast Injector. Retract the Contrast RAM to the 190 ml position
c. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
d. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed
3. Connect the coiled tubing to the end of Syringe Pressure Test Fixture.
5. Tilt the Injector Head downward into the Run position and press ARM
and RUN on the Remote Control
6. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 50 psi during the saline phase. Verify that the
Injector display is between 35 and 65 psi. If the pressure is within the
specified limits, skip to Step 8
7. If the pressure is outside the specified limits, perform the following:
a. Press the SETUP button on the screen. The Setup Screen is
displayed
h. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed
i. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
j. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
9. Do the following:
a. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
c. Tilt the Injector Head downward into the Run position and press
ARM and RUN on the Remote Control
10. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 325 psi during the saline phase. Verify that the
Injector display is between 283 and 367 psi. If the pressure is within
the specified limits, skip to Step 12
11. If the pressure is outside the specified limits, perform the following:
a. Press the SETUP button on the screen. The Setup Screen is
displayed
h. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed.
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
c. Tilt the Injector Head downward into the Run position and press
ARM and RUN on the Remote Control
14. Slowly close the needle valve on the Syringe Pressure Test Fixture
until the pressure measured on the gauge reaches an equilibrium
pressure reading of 150 psi during the saline phase. Verify that the
Injector display is between 125 and 175 psi. If the pressure is within
the specified limits, skip to Step 16
15. If the pressure is outside the specified limits, perform the following:
a. Fully open the needle valve on the Syringe Pressure Test Fixture
by rotating the knob counter-clockwise
b. If necessary, turn the Injector Head upright and refill the syringe
with a J-tube and water bottle. DO NOT try to refill through the
coiled tubing or with the needle valve partially closed. Replace the
coiled tubing when done
If possible, use the procedures in this section to retrieve the pressure profile data (m
and b constants).
The Single Board Computer is programmed with the pressure profile data (m and b
constants) that are unique to each system. Before removing the Single Board Computer,
retrieve the pressure profile from the current Single Board Computer and use them later
to program into the new Single Board Computer.
If you do not press the “Send to Injector” button, the value will not load into the Injector.
6. Press the RETURN button. The Service Screen is displayed
7. Press the SALINE PRESSURE PROFILE button on the screen. The
Saline Injector Pressure Profile Setting Screen showing the m and b
constants are displayed. If the values shown are correct, skip to Step
9
8. Enter the Saline m and b constants as follows:
a. Press m Constant. A keypad is displayed. (see Figure 7-7.
Contrast Keyboard Screen)
b. Key in the correct m value and press ENTER
c. Press b Constant. A keypad is displayed
d. Key in the correct b value and press ENTER. (see Figure 7-8.
Saline Keyboard Screen)
e. Press the SEND TO INJECTOR button
9. Repeatedly press the RETURN button on the screen until the Stop
Mode Main Screen is displayed (see Figure 6-1. Stop Mode Main)
Double click on the icon, and wait for the software to start. Excessive clicking on the
icon may cause the system to behave abnormally when it returns to the Stop screen.
13. Select Setup on the EmpowerCTA®+ Stop screen, then select
Shutdown on the Setup screen.
14. Turn Controller back on
15. Select various function buttons on the Stop screen, and make sure
the function respond correctly.
16. If the touchscreen fails to respond correctly, repeat calibration. If the
problem persists, contact your regional service provider for further
assistance
Figure 7-14 shows the Articulating J-arm used with a legacy injector; at the time of this
manual’s release, the most current version was not available. Height and Tension
Adjustment information in this section, however, is still applicable.
Sudden release of the Articulating J-arm may result in personal injury. When making
height limit, tension adjustments or removing the Injector, hold the Articulating J-arm
and gently guide it to its highest position. If needed, adjust maximum height in order to
get to a comfortable working position.
When working with the Articulating J-arm, be aware of surroundings and inadvertent
contact with other equipment.
1. Raise and hold the Spring J-arm above the Horizontal J-arm
2. Insert a 5 mm Allen key, as shown in Figure 7-15, and turn clockwise
to decrease tension or counterclockwise to increase tension
3. Test and readjust if necessary. Hold onto the J-arm while bringing it
back to its resting position
Figure 7-15. Height and Tension Adjustments for the Articulating J-arm
Description One way stopcock with fixed male luer – 1050 psi
Used for over pressure testing
®+
EmpowerCTA Injector Head Assembly (Part 1)
Table B- 1. Injector Head Assembly (Part 1)
Item
Part Number Description Quantity
Number
1 204143-002 HOUSING FRONT, CTA 1
2 206715-001 GASKET, CONTROL PANEL SCREEN, HT-800 1
3 701303-001 LCD WITH TOUCHSCREEN 1
4 301383-001 FERRITE, LCD DRIVER 2
5 301423-001 EMI GASKET, DIE-CUT FOAM 1
6 204144-002 RETAINER, SCREEN 1
7 301338-001 BEACON LED, PCA 1
8 200142-001 BHCS, 6-32UNC X ¼,STAI 11
9 201693-001 CABLE, SPEAKER 1
10 701320-002 EMPOWERCTA CONTROLLER, KONTRON ASSEMBLY 1
11 204429-001 LICENSE, WINDOW CE 1
12 301337-001 CABLE, BEACON LED 1
13 301317-001 CABLE, LVDS VIDEO 1
14 201811-001 SCREW, 2-56 X .188 PAN HEAD TORX 4
15 301333-001 USB PCA 1
16 200140-001 BHCS, 4-40UNC X ¼ 2
Item
Part Number Description Quantity
Number
17 301348-001 FAN 412 FM EBMPAPST 1
18 200142-003 BHCS, 6-32UNC X ½” 4
19 204145-001 SHIELD, EMI 1
20 204146-002 HOUSING REAR, CTA 1
21 204285-001 BEACON LENS 1
22 204321-001 BEACON CLIP 1
23 301335-001 CABLE, POWER SWITCH 1
24 200139-007 SCREW, SHCS 4-40 X 1” SS 10
25 301332-001 CABLE, PENDANT CONNECTOR 1
26 301374-001 CABLE, PENDANT GROUND 1
27 204319-001 MOUNTING BLOCK 1
28 202417-001 SCR, 10-32 X .625, SHC, SS 2
29 204148-002 WIRE HOOD, CTA 1
30 205711-001 CLAMP, LOOP, .188 DIA 2
31 202429-001 WASHER, FLAT #6 SS OD 5/16 2
32 701085-001 EMPOWR PIVOT JUNCTION SUB ASSEMBLY 1
33 500080-001 DOW CORNING 200 FLUID 1000 CPS. AR
34 202436-001 WASHER, 5/16ID X 0.75 X .02,TEFLON 1
35 202464-001 PIVOT INSERT 1
36 202465-001 PIVOT INSERT CAP 1
37 202468-001 PLUG, FRICTION MECHANISM 1
38 202517-001 TAPERED SPRING, CTA 1
39 202467-001 RETAINER PLUG, FRICT MECHANISM 1
40 202403-001 CABLE TIE, .091” X 4” LG 1
43 202413-001 SCR, PHP 6-32 X .375, SS 2
55 202435-001 WASHER, ¼ID X 0.472 X .06, NYLON 1
62 900912-064 LABEL, REF 1
63 901117-001 MADE IN U.S.A. LABEL 1
64 900807-003 LABEL, S/N BING 1
65 901078-001 CTA MACHINERY LABEL 1
66 901096-001 LABEL, EMPOWER SOFTWARE VERSION BLANK 1
67 202509-001 Copper Foil Tape AR
68 202427-001 NUT, NYLOC ¼-20 1
69 301322-002 CONTRAST CABLE, INTERCONNECT BOARD SIGNAL 1
70 301415-002 SALINE CABLE, INTERCONNECT BOARD SIGNAL 1
71 205580-002 GROMMET EDGING, ROHS COMPLIANT, SIZE 99 AR
ALT 71 205580-001 GROMMET EDGING, ROHS COMPLIANT, SIZE 36 AR
TRI-P PHILLIPS THREAD-FORMING SCREW 18-8 SS, 4-
72 204522-001
20 Thread, ¼” 1
73 200142-002 SCREW, 6-32 x 3/8” SST 4
75 200176-001 TAPE, ADHESIVE, DOUBLE SIDE 2
76 200140-002 BHCS, 4-40 UNC X 1/2 SS 4
78 202406-001 WASHER, LOCK #6 SPLIT SS 2
A 500076-001 LOCTITE 290 AR
B 500021-001 LOCTITE 242 AR
N/A 400178-002 Apollo IH SW 9.01H AR
®+
EmpowerCTA Injector Head Assembly (Part 2)
®+
Table B-2. Actuator Sub-Assembly (EmpowerCTA )
Part Number Description Quantity per Unit of
Measure
Code
202470-001 RADIAL DOUBLE ROW BEARING 1 EA
202513-001 EMP CTA MOTOR & BRNG MTG 1 EA
PLATE
202472-001 RETAINING RING, BEARING INT. 1 EA
301000-002 EMPOWER LINEAR POT ASSY 1 EA
202444-001 ANTI ROT AND WIPER CARRIER 1 EA
BKT
700992-001 BALLSCREW ASSEMBLY 1 EA
202422-001 SET SCREW, CONT PT, 1/4-28X.50 2 EA
202443-001 SUPPORT ROD 3 EA
700991-001 DOOR HOUSING ASSEMBLY 1 EA
202437-001 GASKET, 3/32 THK X 3/16" W 0.50 FT
202448-001 LOCKING DETENT ROD 1 EA
202447-001 LOCKING DETENT 1 EA
202478-001 COMP.SPRING, DETENT ROD 1 EA
LARGE
202445-001 DETENT & LINEAR SENSOR SUPT 1 EA
MT
202418-001 SCR, 6-32 X .75, SHC, SS 2 EA
202430-001 WSR, LOCK, #6, HIGH COLLAR, SS 2 EA
202480-001 SIDE-MOUNT EXTERNAL 1 EA
RETAINING RING, (DETENT ROD)
202408-001 WSR, FLAT, #8, SS 2 EA
202479-001 BOWED SIDE-MOUNT EXTERNAL 1 EA
RETAINING RING (DETENT ROD)
202438-001 PLUNGER HEAD 1 EA
202417-001 SCR, 10-32 X .625, SHC, SS 3 EA
202431-001 WASHER LOCK #10 5 EA
202442-001 TIMING PULLEY, BALL SCREW, 1 EA
Accessories
800687-003 017412 Power/COM cable, 75 feet
Spare Parts
301127-001 016163 USB Communications cable, 9 feet