BK PRECISION 251x - Manual
BK PRECISION 251x - Manual
BK PRECISION 251x - Manual
2
A CAUTION statement calls attention to an operating procedure, practice,
or condition, which, if not followed correctly, could result in damage to or
destruction of part or all of the product.
NOTE:
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50 – 60 Hz) for AC sine
wave applications and as Vdc for DC applications.
WARNING
WARNING
Disconnect all probes, leads, power adapter, and USB cable before
opening the battery cover.
Do not turn on the instrument while the battery cover is removed.
Use only the included insulated probes, test leads, and power adapter
supplied with the instrument.
Use only the battery pack supplied with the instrument or approved by
us. Do not use other substitutes.
For AC power operation, always connect the adapter to the outlet first
before connecting to the instrument.
Observe all markings and ratings on the instrument before connecting
to the instrument.
While making measurements, verify that the performance ratings of
instruments and accessories that are used meet the correct safety levels
of the instrument.
3
Do not insert metal objects into any of the connectors.
WARNING
Do not use any exposed metal BNC or banana plug connectors with the
instrument. Use only the insulated accessories supplied with the
instrument.
Do not connect the USB cable while the instrument is not in use. Keep
the cable away from all probes and test leads.
Always connect the probes or test leads to the instrument first before
connecting to any other device under test (DUT). To remove,
disconnect them from the DUT first before disconnecting from the
instrument.
Do not operate the instrument in wet environments.
For high voltage testing, keep the BNC connectors and the entire
instrument dry.
Do not apply input voltages above the instrument and the probe’s
voltage rating.
CAUTION
Always cover the DC power input and USB ports on the side of the
instrument with the rubber lid when not in used.
Do not use any abrasive chemicals or volatile liquid when cleaning the
instrument. If using a cloth dampened with water, allow the instrument
to completely dry before reconnecting it.
4
Compliance Statements
Disposal of Old Electrical & Electronic Equipment (Applicable in the
European Union and other European countries with separate collection
systems)
5
CE Declaration of Conformity
The power supply meets the requirements of 2006/95/EC Low Voltage
Directive and 2004/108/EC Electromagnetic Compatibility Directive with the
following standards.
EMC Directive
- EN 61326-1: 2006
- EN 61000-3-2: 2006+A2: 2009
- EN 61000-3-3: 2008
6
Safety Symbols
Oscilloscope Inputs
Maximum input voltage to CH1 and CH2 BNC direct(1:1) - CATII 300V
RMS, CATI 150V RMS
Maximum input voltage to CH1 and CH2 BNC via 10:1 probe PR250SA
(included with models 2515 and 2516) - CATII 1000V RMS, CATIII 600V
RMS
7
Maximum input voltage to CH1 and CH2 BNC via 1X/10X probe PR150SA
(included with models 2511 and 2512) – CATII 300V RMS
Scope Input – CATII 300V RMS
Meter Input – CATII 600V RMS, CATIII 300V RMS
NOTE:
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50 – 60 Hz) for AC sine
wave applications and as Vdc for DC applications.
These apply to models 2515 and 2516 only, which have channel isolation
that allows for floating measurements.
Maximum floating voltage from input CH1 and CH2 to earth ground
– CATII 1000V RMS, CATIII 600V RMS
Maximum floating voltage between CH1 and CH2 reference – CATII
1000V RMS, CATIII 600V RMS
NOTE:
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50 – 60 Hz) for AC sine
wave applications and as Vdc for DC applications.
8
Environmental Conditions
The instrument may be operated in the following environment.
Operating Environment 0 °C to 40 °C
Storage Humidity 0 – 85% R.H.
Storage Environment -20 °C to +70 °C
Pollution degree Pollution degree 2
Measurement Category CAT II, CAT III
9
Table of Contents
Safety Summary .............................................................................................. 2
1 General Information ............................................................................. 16
1.1 Product Overview ........................................................................... 16
1.2 Package Contents ........................................................................... 17
1.3 Front Panel Overview ..................................................................... 18
Front Panel Description ........................................................................ 18
1.4 Top View ......................................................................................... 20
1.5 Rear Panel Overview ...................................................................... 20
1.6 Side Panel Overview ....................................................................... 21
Side Panel Description .......................................................................... 21
1.7 Display Overview ............................................................................ 22
Oscilloscope Display .............................................................................. 22
Digital Multimeter Display .................................................................... 24
Recorder/Trend Plot Display ................................................................. 25
1.8 Protection Fuse .............................................................................. 26
1.9 Isolation (Model 2515/2516) ......................................................... 26
1.10 Probes ......................................................................................... 27
2 Getting Started...................................................................................... 28
2.1 Input Power Requirements ............................................................ 28
External Power Operation .................................................................... 28
Battery Operation ................................................................................. 29
2.2 Tilt Stand......................................................................................... 30
2.3 Power On Device ............................................................................ 30
2.4 Perform Self-Calibration................................................................. 31
2.5 Set to Factory Default .................................................................... 32
2.6 Configure Date and Time ............................................................... 35
2.7 Probe Compensation ...................................................................... 36
3 Using the Oscilloscope .......................................................................... 39
10
3.1 Channel Selection ........................................................................... 39
3.2 Channel Menu ................................................................................ 40
Channel Coupling .................................................................................. 41
Bandwidth Limit .................................................................................... 41
Vertical Scale ......................................................................................... 41
Probe Attenuation Scale ....................................................................... 41
Invert Waveform ................................................................................... 42
Digital Filter ........................................................................................... 42
3.3 Automatic Settings and Run/Stop Mode ....................................... 43
Auto Setting .......................................................................................... 43
Run/Stop Mode ..................................................................................... 44
3.4 Scope Menu.................................................................................... 44
Acquire Menu ....................................................................................... 44
Display Menu ........................................................................................ 45
Math Menu ........................................................................................... 49
Horizontal System ................................................................................. 54
Vertical System ..................................................................................... 60
Reference Waveform Menu ................................................................. 62
3.5 Trigger System ................................................................................ 64
Adjust Trigger Level............................................................................... 64
Edge Trigger .......................................................................................... 65
Pulse Trigger.......................................................................................... 67
Video Trigger ......................................................................................... 69
Slope Trigger ......................................................................................... 72
Alternative Trigger ................................................................................ 74
3.6 Connecting External Memory ........................................................ 75
3.7 Save/Recall Function ...................................................................... 76
Saving and Recalling Setups and Waveforms ....................................... 78
Saving Screen Capture .......................................................................... 83
Saving Waveform Data ......................................................................... 83
11
3.8 Utility Menu.................................................................................... 84
Check System Information .................................................................... 86
Enable/Disable Key Sound .................................................................... 87
Enable/Disable Frequency Counter ...................................................... 87
Set the Language ................................................................................... 88
Updating Firmware ............................................................................... 88
Set Screen Saver.................................................................................... 90
Waveform Record Function .................................................................. 90
Playback Recorded Waveform .............................................................. 93
3.9 Measurement and Cursor Functions.............................................. 94
Measurement Functions ....................................................................... 94
Cursor Functions ................................................................................. 101
4 Using the Digital Multimeter .............................................................. 108
4.1 DC and AC Voltage Measurements .............................................. 109
DC Voltage Measurement ................................................................... 110
AC Voltage Measurements ................................................................. 112
4.2 DC and AC Current Measurements .............................................. 113
DC Current Measurements ................................................................. 114
Make AC Current Measurements ....................................................... 116
4.3 Make Resistance Measurements ................................................. 118
4.4 Make Diode Measurements ......................................................... 120
4.5 Continuity Test ............................................................................. 121
4.6 Make Capacitance Measurements............................................... 123
4.7 Using Trend Plot ........................................................................... 124
5 Using the Recorder ............................................................................. 125
Scope Trend Plot ..................................................................................... 125
Record Data with Trend Plot............................................................... 127
Save Recorded Data to External Memory........................................... 129
Scope Recorder ....................................................................................... 130
Configure Record Options................................................................... 133
12
Start Scope Recording ......................................................................... 133
Playback Scope Recording .................................................................. 134
Meter Trend Plot..................................................................................... 136
Record Data with Trend Plot............................................................... 139
Save Recorded Data to External Memory........................................... 140
6 Remote Communication ..................................................................... 142
7 Troubleshooting Guide ....................................................................... 143
System Message Prompts and Instructions ............................................ 143
Frequency Asked Questions.................................................................... 144
Oscilloscope ........................................................................................ 144
Multimeter .......................................................................................... 145
8 Specifications ...................................................................................... 146
Oscilloscope Specifications ..................................................................... 146
Maximum Input Voltages........................................................................ 150
Multimeter Specifications ....................................................................... 152
Recorder Specifications .......................................................................... 154
General Specifications ............................................................................ 155
11 Maintenance ....................................................................................... 156
Cleaning................................................................................................... 156
SERVICE INFORMATION .............................................................................. 157
LIMITED THREE-YEAR WARRANTY .............................................................. 158
13
List of Figures
14
Figure 38 - USB Drive Removed .................................................................... 76
Figure 39 - Directory Menu 1/2 .................................................................... 79
Figure 40 - Create Directory Name ............................................................... 80
Figure 41 - Save Progress .............................................................................. 81
Figure 42 - Save Successfully......................................................................... 81
Figure 43 - Overwrite Warning ..................................................................... 82
Figure 44 - Subdirectory View ....................................................................... 83
Figure 45 - System Information Display ........................................................ 87
Figure 46 - Frequency Counter Display ......................................................... 88
Figure 47 - Record Frame Indicator .............................................................. 92
Figure 48 - Recording Finished...................................................................... 92
Figure 49 - All Voltage Measurements Display ............................................. 99
Figure 50 - All Time Measurements Display ............................................... 100
Figure 51 - All Delay Measurements Display .............................................. 100
Figure 52 - All Measurements Display ........................................................ 100
Figure 53 - Manual Voltage Cursor ............................................................. 102
Figure 54 - Manual Time Cursor ................................................................. 103
Figure 55 - Track Cursor Display ................................................................. 105
Figure 56 - Auto Cursor Measuring Vpp ..................................................... 106
Figure 57 - Auto Cursor Measuring Period ................................................. 107
Figure 58 - Multimeter Display ................................................................... 108
Figure 59 - Out of Range ............................................................................. 111
Figure 60 - Connection for DC Voltage Measurement ............................... 111
Figure 61 - Connection for AC Voltage Measurement................................ 113
Figure 62 - Connection for Low DC Current Measurement ........................ 115
Figure 63 - Connection for Higher DC Current Measurement .................... 115
Figure 64 - Connection for Low AC Current Measurement ........................ 116
Figure 65 - Connection for Higher AC Current Measurement .................... 117
Figure 66 - Connection for Resistance Measurement ................................ 119
Figure 67 - Connection for Diode Measurement ........................................ 121
Figure 68 - Continuity Test .......................................................................... 121
Figure 69 - Connection for Continuity Test ................................................. 122
Figure 70 - Capacitance Measurement ....................................................... 123
Figure 71 - Connection for Capacitance Measurement .............................. 124
Figure 72 - Scope Recorder External Save Mode ........................................ 133
Figure 73 - Recording Display ..................................................................... 134
Figure 74 - Playback Scope Recording ........................................................ 135
Figure 75 - Meter Trend Plot Display .......................................................... 137
15
1 General Information
Features:
5.7” color TFT LCD Display
Oscilloscope, multimeter, and recorder functionality (trend plot and
waveform recorder)
Fully isolated oscilloscope channels (models 2515 and 2516 only)
and isolated multimeter inputs
6000-count multimeter
Multimeter supports voltage, current, resistance, capacitance,
diode, and continuity measurements
Supports scope and multimeter measurement parameters trend plot
and scope waveform recorder
32 auto measurement functions
Store/recall 2 groups of reference waveforms, 20 groups of common
waveforms, 10 groups of instrument settings
USB interface for remote communication
16
1.2 Package Contents
Please inspect the instrument mechanically and electrically upon receiving
it. Unpack all items from the shipping carton, and check for any obvious
signs of physical damage that may have occurred during transportation.
Report any damage to the shipping agent immediately. Save the original
packing carton for possible future reshipment. Every oscilloscope is shipped
with the following contents:
Verify that all items above are included in the shipping container. If
anything is missing, please contact B&K Precision.
17
1.3 Front Panel Overview
2 26
25
3
4 24
23
5
22
6 21
7 20
19
8 18
9 17
10 16
11 12 13 14 15
5
Recorder function button
Trend plot and waveform recorder
6 Channel 1 button (scope)
18
7 Trigger Menu button
9 Power On/Off
15
Voltage/Resistance/Capacitance input port
(multimeter)
19
26 Menu On/Off button
Battery cover
Kickstand
20
1.6 Side Panel Overview
2 4
21
1.7 Display Overview
Oscilloscope Display
1 2 3 4 5
6
16 7
15 8
9
14 10
11
13 12
Figure 5 - Oscilloscope Display
Trigger status
Armed – Instrument is acquiring pre-trigger data. All triggers
are ignored.
Ready – All pre-trigger data has been acquired and the
instrument is ready to accept a trigger
1
Trig’d – Instrument has triggered and is acquiring post-trigger
data
Stop – Instrument has stopped acquiring waveform data
Auto – Instrument is in auto mode and is acquiring
waveforms without trigger
Scan – Instrument is acquiring and displaying waveform data
continuously. This mode is automatically set for when
22
timebase is slow enough in which requires a longer time to
“scan” to acquire the entire signal on display
2
Displays the position of the present waveform window in internal
memory
3 Horizontal trigger position marker
USB mode indicator
Side USB interface is configured for PC
4
communication
Side USB interface is configured for printer (not available)
Battery indicator
5
- Battery charging
- Battery empty
- Battery full
6 Time and date display
7 Trigger position marker
8 Horizontal trigger position
9 Trigger signal frequency
10 Trigger voltage indicator
11 Menu header/label
12 Horizontal timebase
13 Menu display
Signal coupling
- DC coupling
- AC coupling
14
- GND(ground) coupling
Volts/div indicator
Bandwidth limit On/Off indicator
B - Bandwidth limit ON
15 Channel 2 identifier
16 Channel 1 identifier
23
Digital Multimeter Display
1 2 3 4
5
6
24
Recorder/Trend Plot Display
1 2 3 4
5
6
25
1.8 Protection Fuse
There are two input protection fuses for the current measurement inputs,
labeled mA and 10 A.
26
1.10 Probes
WARNING:
Only use the supplied oscilloscope probes or probes
recommended by B&K Precision. Using probes not
recommended and/or supplied by B&K Precision may
cause a safety hazard, and may also damage the
instrument and void its warranty.
Models 2515 and 2516 come with two touch-protected passive 10X test
probes (PR250SA) with the following specifications:
Models 2511 and 2512 come with two passive 1X/10X probes (PR150SA)
with the following specifications:
27
2 Getting Started
1. Connect the adapter to the external power input on the side of the
instrument first.
2. Then, connect the supplied power cord to the adapter.
3. Connect AC adapter to an electrical outlet.
Adapter specifications:
Input: AC 100-240V, 50/60Hz, 1.2 A
Output: DC 9V, 4 A
Center pin positive, 5.5mm tip
28
WARNING:
Only use the supplied external AC adapter. Using a
different or incorrect type adapter will result in
damage to the instrument and void its warranty.
Battery Operation
The instrument can be powered by an external Li-ion battery pack, rated for
5000 mAh, 7.4 V. The battery must be installed and charged before first
time use for battery operation.
1. Use a screw driver to remove the two screws securing the battery
cover in the rear panel.
2. Connect the 2-pin connector of the battery to the 2-pin input
connector on the lower left corner of the battery compartment.
3. Place the battery pack inside the compartment and place the battery
cover. Fasten the two screws to tighten.
Battery
Battery
1 2 3
29
Charging the Battery
For first time use, fully charge the battery for at least 4 hours with the
instrument turned OFF.
Before charging the battery, follow the instructions above to install the
battery. Then, connect the AC adapter to the instrument. While the
instrument is ON, the battery indicator will show the battery status for
charging and when it’s fully charged (see “1.7 Display Overview”). The
battery also charges when the instrument is OFF.
30
The oscilloscope will display a boot screen shown below for 5-10 seconds
before the main operating display appears.
CAUTION:
Before performing self-calibration, disconnect all
inputs of the instrument. Otherwise, the instrument
may not calibrate correctly or may cause errors.
31
To perform self-calibration, follow these steps:
32
Menu or
Options, Knobs or Buttons Default setup
system
Coupling DC
BW Limit Off
Volts/div Coarse
CH1, CH2
Probe 1X
Invert Off
Filter Off
Volts/div 1.00V
Operation CH1+CH2
CH1 Invert Off
CH2 Invert Off
FFT Operation:
MATH Source CH1
Window Hanning
FFT Zoom 1X
Scale dBVrms
Display Split
Window Main Time Base
Position 0.00 μs
Sec/div 500 μs
HORIZONTAL
Window Zone 50.0 μs
Trigger knob Level
Type Off
Source CH1
CURSOR
Horizontal (voltage) +/-3.2 divs
Vertical (time) +/-5 divs
Three Mode Options Sampling
ACQUIRE Averages 16
Sampling Method Real Time
Type Vectors
DISPLAY Persist off
Grid
33
Intensity 60%
Brightness 40%
Format YT
Menu Display Infinite
Type Setups
SAVE/RECALL Save To Device
Setup No.1
Source CH1
REF REFA Off
REFB Off
Sound On
Frequency Counter On
UTILITY
USB Device Computer
Record Off
Type Edge
Source CH1
Slope Rising
TRIGGER (Edge)
Mode Auto
Coupling DC
Level 0.00V
Type Pulse
Source CH1
Condition =
TRIGGER (Pulse)
Set Pulse Width 1.00ms
Mode Auto
Coupling DC
Type Video
Source CH1
TRIGGER Polarity Normal
(Video) Sync All Lines
Standard NTSC
Mode Auto
Type Slope
TRIGGER Source CH1
(Slope) Condition
Time 1.00ms
34
Mode Auto
Type Alternative
Source CH1
TRIGGER
Mode Edge
(Alternative)
Coupling DC
Slope Rise
To configure:
35
keys to change the day.
To enable or disable the date and time display, toggle in the menu
to change Display to On or Off. Date and time will be displayed in the
upper right corner, as illustrated below.
When you connect the probe to an arbitrary channel for the first time,
please make the following adjustment to make the probe match with the
channels. Probes without compensation or compensation warp may lead to
36
imprecise or false measurements. You can perform the adjustments
manually to match your probe with the input channels.
1. Set the probe attenuation option in the channel 1 menu to 10X by
pressing and press F4 to select Probe until it shows 10X.
2. Connect the BNC side of the probe to channel 1 input on the top of the
instrument.
37
4. Connect the probe tip to the 3V port (1 kHz, 3Vpp Square Wave) and
the reference ground clip to the ground port on the probe
compensation connector.
7. Adjust the trim on the probe or repeat all the operations above if
necessary.
38
PR250SA
PR150SA
39
3.2 Channel Menu
Menu
Setting Description
Option
DC passes both AC and DC
DC
components of the input signals.
AC blocks the DC component of the
Coupling
AC input signals and attenuates signals
below 10 Hz.
GND GND disconnects the input signal.
Limit the bandwidth above 20 MHz to
On reduce display noise; filters the signals
BW Limit
Off to reduce noise and other unwanted
high frequency components.
Change the range of voltage by 1-2-5
Coarse
sequence.
V/div
Fine control of the volt/div scale
Fine
settings.
Set to match the type of probe
1X, 5X, 10X, 50X,
Probe 100X, 500X, 1000X
attenuation you are using to ensure
correct vertical scale display.
Enter the second page of CH1/CH2
Next Page Page1/2
menu.
40
Table 3 - Channel Menu 2/2
Menu
Setting Description
Option
On
Invert Enable/Disable waveform invert function.
Off
Filter Enter the FILTER menu.
Set waveform vertical position and trigger level to
To Zero
zero.
Next
Page 2/2 Return to the first page of CH1/CH2 menu.
Page
Channel Coupling
Both channel 1 and channel 2 coupling can be set to DC, AC, or GND
(ground). Press in the channel menu to toggle the settings.
Bandwidth Limit
Vertical Scale
The vertical scale control adjustment can be set for coarse or fine control.
Press the button in the channel menu to toggle between Coarse
and Fine.
41
in the channel menu to select and change the Probe setting. As you change
this setting, the volt/div scale indicator will also change based on the
attenuation factor selected.
Invert Waveform
The oscilloscope can display the inverted signal of the input waveform on
channel 1 and channel 2. To enable this, press from the channel
menu to go to the second menu page, and press to toggle Invert
between On and Off.
Digital Filter
There are four types of digital filters available and can be used to apply to
the input waveform. To access the digital filter menu, go to page two of the
channel menu and press to enter the filter menu. Below is a table
of the filter menu structure.
42
3.3 Automatic Settings and Run/Stop Mode
Auto Setting
When measuring an unknown signal, the auto setting function can be used
to allow the oscilloscope to automatically adjust the vertical and horizontal
scale, range, and trigger to display the signal.
43
Run/Stop Mode
In continuous run mode, the signal going into channel 1 and/or 2 will
constantly be triggered regardless of the position of the trigger level. At the
top, the indicator will display “Trig’d” to indicate that the signal is triggered.
In stopped mode, the oscilloscope will stop triggering the signals going into
channel 1 and/or 2. At this point, users can zoom or pan to view the entire
waveform acquired from the last trigger.
To access the scope menu, press the orange colored key. There are
five submenus: Acquire, Display, Math, Horizon, and Ref.
Acquire Menu
Menu
Setting Description
Option
44
Sampling Sample and display most waveforms accurately.
Acquisition Peak Detect Detect burr and reduce fake wave phenomena.
Display Menu
45
Table 6 - Display Menu 1/2
Menu
Setting Description
Option
YT format displays the vertical voltage in
YT
relation to time (horizontal scale).
Format XY format displays a dot each time a
XY sample is acquired on channel 1 and
channel 2.
Normal Set to normal mode.
Screen
Inverted Set to invert color display mode.
Display grids and axes on the screen.
Grid Turn off the grids.
Turn off the grids and axes.
46
Menu 2sec, 5sec, 10sec,
Set display time of menu on the screen.
Display 20sec, Infinite
Next Page Page 2/2 Return to the first page of DISPLAY menu.
for grid brightness in the Display menu page 1. Then, use the or
Below is an example of using XY mode with two sine waves for channel 1
and 2 with 90 degree offset.
Figure 16 - XY Mode
47
Screen Invert: The colors of the screen can be inverted. Toggle the
button in the Display menu 2 to switch between normal and
inverted display. Below is an illustration of the inverted screen.
Grid Display: The display of the grid can be changed to show grid and axes,
axes only, or no grid or axes. The below screenshots illustrate the options.
This can be changed by toggling in the Display menu 2.
48
Figure 19 - Axes Only Display
Math Menu
49
Table 8 - Math Menu 1/2
50
Figure 21 - Math Waveform Display
FFT
When FFT is selected, additional menu options are available. See table
below for the FFT menu structure.
51
(see table
Window Selects FFT windowing.
below)
FFT Zoom 1X, 2X, 5X, 10X Zoom function of FFT waveform.
52
On
Turn On/Off the FFT waveform.
Off
When using the FFT function, first select the channel source by toggling the
button from the FFT menu 1 to change Source between CH1 and
CH2.
Use the button from the FFT menu 1 to select the Window to use.
Descriptions of each window are provided in Table 11 - FFT Window
Options above.
Use the button in the FFT menu 1 to adjust the FFT Zoom. You can
also use the horizontal timebase keys to change the FFT Zoom.
The FFT waveform can be displayed as part of the full screen or as a split
screen from channel 1 and/or 2. Below illustrates the differences.
The vertical position of the FFT waveform can be adjusted using channel 1
vertical position keys (if Source is CH1) or channel 2 vertical position keys
(if Source is CH2) . The position can always be set back to 0 by going into
FFT menu 2 and pressing to select ToZero.
53
1
2 4
Horizontal System
54
Figure 23 - Horizontal Position Reference Marker
This position can be adjusted by using the keys to move left (<) or
right (>). If the left or right position key is pressed down for > 3 seconds,
the position change will be coarse. The scale factor is indicated as marked
by the illustration below.
The position with center reference to 0 will be displayed as you change the
position, as illustrated below:
55
Figure 25 - Horizontal Position Indicator
The horizontal scale or timebase scale factor can be changed using the
Delay Scan
The instrument has a delay scan function that can be enabled to zoom in on
a portion of the channel 1 and/or channel 2 waveforms.
56
Figure 27 - Delay Scan
The blue shaded areas mark the masked out area and the area in between
the blue shaded area is the zoomed portion, shown at the bottom in the
split screen.
Deep Memory
57
4. Now, go back to Horizon menu and change MemDepth to Long
Mem. Go back to Acquire menu afterwards and note the Sa Rate is
now 100.0MSa.
5. The differences can be visually observed by pressing to stop
the waveform at 100 µs. Then adjust the horizontal scale factor to a
slower timebase until the entire waveform captured can be shown
on the display.
6. Observe with both MemDepth set to Normal and Long Mem.
7. Below is an example between the two captured at 100 µs timebase.
58
Note: Different timebase settings will result in different sampling rates,
both with normal and long memory depth configured in MemDepth.
Selecting a timebase lower than 50 ms will automatically set the
instrument into scan mode. Deep memory is not supported in scan mode.
For timebase selections faster than 25 µs, the deep memory function will
have an adverse affect and will result in fewer sample points than if deep
memory is disabled.
59
5 µs
2.5 µs
1 µs
250 MSa/s 250 MSa/s
500 ns
250 MSa/s
250 ns
500 MSa/s
100 ns
50 ns
25 ns 250 MSa/s
10 ns 500 MSa/s
5 ns 1 GSa/s 500 MSa/s
2.5 ns (100
MHz
models)
Vertical System
The vertical position of channel 1 and channel 2 are marked on the left side
of the display, as illustrated below:
60
To move the vertical position up or down, use the left vertical position keys
(for CH1) or right vertical position keys (for CH2) . If the up or down key
is pressed down for > 3 seconds, the position change will be coarse. The
vertical position relative to center 0 reference will be indicated when
changed, as illustrated below:
The vertical scale factor can be adjusted by pressing the left volts/div range
keys (for CH1) or the right volts/div range keys (for CH2) . To increase
the scale, press V. To decrease, press mV.
By default, the volts/div range keys will change the vertical scale factor in 1-
2-5 increments. Users can select fine control so that the change will be
61
Figure 32 - Fine Control Volts/Div
The instrument can save and display two reference waveforms from both
channel 1 and/or channel 2. A reference waveform copies the source
waveforms’ vertical position, vertical scale factor, horizontal position, and
horizontal scale factor.
To access the Reference menu, press and to select Ref.
Menu Description
Setting
Option
CH1/CH1 Off Choose the channel to save as a reference
Source
CH2/CH2 Off waveform.
Ref A Choose to save or recall reference A or
Ref B reference B waveform.
Save the source waveform to the selected
Save
reference (A or B).
62
On Toggles displaying the reference waveform.
Ref A/Ref B
Off
Note: Under Source, CH1 Off and CH2 Off will be displayed if both channel
1 and 2 are turned Off.
63
3.5 Trigger System
There are five types of trigger functions supported by the instrument: edge,
pulse, video, slope, and alternative.
The trigger level position is marked by T on the right of the display. When
changed, the screen will temporarily display the trigger level line across the
window with the trigger position indicator in the bottom left corner, as
illustrated below:
Note: Some options within the trigger menu share the same arrow keys
used to adjust the trigger level. In which case, exit the trigger menu by
pressing first, then the arrow keys can be used to change the
trigger level.
64
Edge Trigger
Menu
Setting Description
Option
Trigger on the rising or falling edge of the input
Type Edge
signal.
CH1
Source Set CH1 or CH2 as the trigger source.
CH2
Trigger on rising edge of the trigger signal.
Trigger on falling edge of the trigger signal.
Slope
Trigger on rising and falling edge of the trigger
signal.
Use this mode to let the acquisition free-run in
the absence of a valid trigger; This mode allows
Auto
an un-triggered, scanning waveform at 100
ms/div or slower time base settings.
Mode Use this mode when you want to see the
Normal triggered waveforms only; the scope will not
acquire waveform until it’s been triggered.
The setup detects a trigger and acquires
Single
waveform, then stops to give a signal capture.
Set Up Enter the Edge Trigger Setup Menu.
65
Table 16 - Edge Trigger Setup Menu
Menu
Setting Description
Option
DC Passes all components of the signal
Blocks DC components and attenuates
AC
signals below 170 Hz.
Attenuates the high-frequency
Coupling HF Reject
components above 140 kHz.
Blocks the DC component and attenuates
LF Reject the low-frequency components below 7
kHz.
Adjusts the hold off time.
Hold off
Hold off
Reset hold off time to 100 ns.
Reset
Return Return the Edge Trigger Menu.
Hold Off: To adjust the hold off time, go to the Edge Trigger Setup Menu
Hold off time can be reset to 100 ns by pressing to select Hold Off
Reset.
66
Figure 35 - Hold Off Time Indicator
Pulse Trigger
Menu
Setting Description
Option
Type Pulse Select the pulse trigger function.
CH1 Set CH1 or CH2 as the trigger
Source
CH2 source.
67
(Positive pulse width
less than pulse width set)
(Positive pulse width
larger than pulse width
setting)
(Positive pulse width
equal to pulse width
setting) Select the pulse trigger
When (Negative pulse conditions; it triggers as soon as
width less than pulse width condition is met.
setting)
(Negative pulse
width larger than pulse
width setting)
(Negative pulse
width equal to pulse width
setting)
Set 20.0 ns - 10.0 s Set the pulse width for pulse
Width trigger condition.
Next Enter the second page of the
Page 1/2
Page Pulse Trigger Menu.
Menu
Setting Description
Option
Select the pulse to trigger the pulse match the
Type Pulse
trigger condition.
Auto
Select the type of triggering; Normal mode is
Mode Normal
best for most Pulse Width trigger applications.
Single
Enter the Pulse Trigger Setup Menu. See
Setup
Table 16 for details.
Return to the first page of the Pulse Trigger
Next Page Page 2/2
Menu.
68
To set the pulse width for the pulse trigger condition, press in the
Note: Trigger level cannot be changed while the Pulse Trigger Menu is
active because the SetWidth option shares the same arrow keys to make
changes.
Video Trigger
Menu
Setting Description
Option
69
Select Video trigger. To trigger NTSC, PAL and
Type Video
SECAM video signals, set coupling to AC.
CH1
Source Set CH1 or CH2 as the trigger source.
CH2
Normal trigger on the negative edge of the sync
(Normal)
pulse.
Polarity
Inverted trigger on the positive edge of the sync
(Inverted) pulse.
All Lines
Line Num
Sync Select appropriate video sync.
Odd Field
Even Field
Next
Page 1/2 Enter the second page of Video Trigger Menu.
Page
Sync: The bottom right of the display indicates the Sync setting the
instrument is configured to.
All Lines The video trigger will sync on all lines of the
signal.
Line Num: To change the line number for video trigger, use the or
Note: Trigger level cannot be changed while the Video Trigger Menu is
active because the Line Num Sync option shares the same arrow keys to
make changes.
70
Table 20 - Video Trigger Menu 2/2
Menu
Setting Description
Option
When you select type video and select
Type Video AC coupling, you could trigger to a
NTSC, PAL and SECAM video signal.
NTSC, Select the video standard for sync and line
Standard
PAL/SECAM number count.
Use this mode to let the acquisition free-
run in the absence of a valid trigger. This
Auto mode allows an un-triggered, scanning
waveform at 100 ms/div or slower time
base settings.
Mode
Use this mode when you want to see only
triggered waveforms; the oscilloscope does
Normal
not display a waveform until after the first
trigger.
Single Use this mode to run a single capture.
Enter the Video Trigger Setup Menu. See
Setup
Table 16 for details.
Return the first page of Video Trigger
Next Page Page 2/2
Menu.
Menu
Setting Description
Option
Adjusts the hold off time.
Hold off
Hold off
Reset hold off time to 100 ns.
Reset
71
Return Return to the Video Trigger Menu.
Slope Trigger
Slope trigger can be used to trigger off a rising or falling edge with specified
slope timing conditions and vertical boundaries. It is similar to Edge trigger
but with more options users can setup to seek for specific slope conditions
from the selected source.
Menu
Setting Description
Option
Trigger on positive slope or
Type Slope
negative slope.
CH1
Source Select trigger source.
CH2
72
Use the arrow keys to set
Time Set time slope time. The time range is
20 ns – 10 s.
Enter the second page of the
Next Page Page 1/2
Slope Trigger Menu.
Menu
Setting Description
Option
Type Slope Select slope trigger.
(Set high trigger level
position of rising edge)
(Set low trigger level
position of rising edge)
(Set high and low
Selects the high and low trigger
trigger level position of rising
level that can be adjusted for
edge)
Vertical slope trigger. Use this to set
(Set high trigger level
the boundaries for the slope
position of falling edge)
trigger.
(Set low trigger level
position of falling edge)
(Set high and low
trigger level position of
falling edge)
Use this mode to let the
acquisition free-run in the
absence of a valid trigger; This
Auto mode allows an un-triggered,
scanning waveform at 100
Mode ms/div or slower time base
settings.
Use this mode when you want
to see only triggered
Normal
waveforms; the oscilloscope
does not display a waveform
73
until after the first trigger.
Use this mode to run a single
Single
capture.
Enter the Slope Trigger Setup
Setup
Menu. See Table 16 for details.
Return to the first page of the
Next Page Page 2/2
Slope Trigger Menu.
Alternative Trigger
Alternative trigger will alternate the trigger between two channels to allow
triggering both channels. Trigger types can be set to Edge, Pulse, Video, and
Slope. Trigger information will be displayed at the bottom right of the
screen.
Menu
Setting Description
Option
CH1
Source Select trigger source.
CH2
Edge
Pulse
Mode Select trigger type for the alternative trigger.
Video
Slope
See
74
If Mode is Edge, see Edge Trigger section.
Next
If Mode is Pulse, see Pulse Trigger section.
Page/Set
If Mode is Video, see Edge Trigger section.
up
If Mode is Slope, see Slope Trigger section.
Menu
Setting Description
Option
To connect an external USB flash drive, plug the drive into the USB host port
on the side of the instrument as labeled in “Figure 4” in section “1.6 Side
Panel Overview”.
If the USB flash drive is recognized by the instrument, it will display the
following prompt:
75
Figure 37 - USB Drive Detected
When the USB flash drive is removed from the USB host port, the following
prompt will display:
76
Table 26 - Save/Recall Menu
Menu
Setting Description
Option
Setups Select to save instrument settings/setups.
Waveforms Select to save the waveform traces on screen.
Type Picture Select to save the screenshot as a .BMP file.
CSV Select to save the waveform data as a .CSV file.
Factory Select to set instrument to default settings.
Device Select to save the file into internal memory.
Save to
File Select to save the file into external memory.
Choose the position number to save/recall
Setup No.1 to No.20
setups.
Select to save instrument setups or waveform
Save
traces.
Select to recall instrument setups or waveform
Recall
traces.
Menu
Setting Description
Option
Type Picture Select to save the screenshot as a .BMP file.
Save Picture Select to save screenshot to external memory.
Print Picture Print option is not supported.
77
Select to save bitmap image to external
memory. (This option is available when a USB
Save
flash drive is connected and detected by the
instrument)
Menu
Setting Description
Option
Type CSV Select to save the waveform data as a .CSV file.
Displayed Select to save only the waveform data
Data
displayed within the grid.
Depth
Maximum Select to save all captured waveform data.
Para On Choose to save scope parameters in the data
Save Off file.
NOTE:
If Data Depth is set to Maximum, it may take several
minutes to complete the save process. Do not press
any key during the save process.
Instrument setups and waveform traces can be saved into internal and
external memory.
78
To save setups and waveforms, follow the steps below:
1. Press to go into the Save/Recall menu, then select Type.
2. Toggle to choose Setups to save instrument setups, or
choose Waveforms to save waveform traces on display.
3. To save the file into internal memory, press to set Save To
to Device. To save the file to external memory, set it to File and go
to step 5.
4. If Save To is set to Device, toggle to choose the internal
memory location to save the file to. There are total of 20 storage
locations (No.1 – No.20).
5. Now, press to Save and the current instrument settings
will be stored into the configured memory location. If saving to
external memory, the directory explorer will display, as shown
below. Here, you can choose to create a new directory or new file
to save as your file.
79
Figure 40 - Create Directory Name
7. You can set the name using the on screen virtual keypad. Use
and keys to move the cursor position of the
name to left and right respectively. To select a character to insert at
80
Figure 41 - Save Progress
13. If the file name is found to be the same as another file already in the
external memory, the instrument will prompt an overwrite warning.
Press to select Confirm to overwrite the file, or press
to select Cancel.
81
Figure 43 - Overwrite Warning
Note: Instrument setups will be saved as a .SET file and waveform traces
will be saved as a .DAV file. Both of these file types can only be recalled
from within the oscilloscope Save/Recall menu.
82
by pressing . To go back up a folder, just press it again with
cursor selecting UP.
Follow the steps below to save waveform data into external memory:
1. From the Save/Recall menu, toggle to select CSV for Type.
83
2. Then, select Data Depth as Displayed or Maximum. (See “Table 28”
for description of the options)
3. Then, select Para Save as On or Off. Set to On if you want scope
parameters (i.e. timebase, volts/div) to be saved along with the
waveform data.
4. Press to select Save.
5. Similar to saving a screenshot image, setups, and waveform traces
as instructed in previous sections above, you will enter the Directory
menu.
6. Create a new file to save as and the file will be saved to external
memory as a .CSV file.
NOTE:
If Data Depth is set to Maximum, it may take several
minutes to complete the save process. Do not press
any key during the save process.
Menu
Setting Description
Option
System Displays the instrument
Status information.
Enable key sound.
Sound
Disable key sound.
On Turn ON the frequency counter.
Counter
Off Turn OFF the frequency counter.
84
English, Français, 日本語,
한국의, Deutsch, Español,
Language русский, Italiana, Select the interface language.
Português, ال عرب ية, Polish,
简体中文, 繁體中文
Next
Page 1/4 Enter the next page of the menu.
Page
Menu
Setting Description
Option
Do a self calibration to calibrate the
Do Self Cal
channels.
Screen Test Run the screen detect program
Do Self Test
Keyboard Test Run the keyboard detect program
Print Setup Not supported.
Printer Not supported
Select to allow remote communication with
USB Device
Computer application software via the miniUSB port
on the side panel.
Next Page Page 2/4 Enter the third page of the menu.
Menu
Setting Description
Option
Update firmware Select to update firmware.
85
Press this button to enter the Waveform
Record
Record Menu.
Next Page Page 3/4 Enter the fourth page of the menu.
86
Figure 45 - System Information Display
The key press sound can be enabled or disabled from the Utility Menu 1/4.
Press to toggle the Sound option to enable or disable key sound.
87
Figure 46 - Frequency Counter Display
Updating Firmware
88
4. Press to enter the Utility Menu and press to go to
page 3/4 of the menu.
5. Select Update Firmware by pressing .
WARNING:
Do not power off the instrument or press any of the
front panel keys at any time during the update
process. Any power interruption or key press
interruption may damage the instrument or corrupt
the firmware, causing it to malfunction and void the
warranty.
10. Wait for 1-2 minutes for the update to finish. A progress bar on the
display will indicate when update is complete.
11. When finished, a prompt will say “Update Successfully” and will
prompt the user to reboot the instrument.
12. At this point, power off the instrument by holding down for at
least three seconds, then power back on again.
13. To verify that the instrument has updated successfully, press
to go to the Utility menu and press from page 1 of the
menu to select System Status.
89
Set Screen Saver
The instrument has a screen saver mode to save power when the
instrument is idle or not in use. The timer to enable the screen saver can be
configured by going to page 4 of the Utility menu and pressing to
select Screen saver option. Toggle this to change the screen saver timer.
NOTE:
Record function is not available when Alternative
Trigger is selected. Single trigger is also not
supported when using this function.
90
Table 33 - Waveform Record Menu
Menu
Setting Description
Option
Record Set to configure and record.
Mode Replay Set to playback recorded waveform.
Off Turn off waveform record menu.
Source CH1, CH2 Choose recorder source.
Set interval to record waveform. (1 ms
Interval
– 1000 s)
Set the maximum number of frames
End Frame
to record. (1 – 2500)
Start to record
Operate
Stop recording
91
An indicator at the top left of the grid display will indicate the approximate
frame count of the recording.
92
Playback Recorded Waveform
Menu
Setting Description
Option
Mode Play Back Set to playback recorded waveform.
Press to start playback.
Operate
Press to stop playback.
Set to repeat playback.
Play Mode
Set to run playback one time.
Set interval between frames during
Interval
playback.
Next Page Page 1/2 Enter the second page of the menu.
Menu
Setting Description
Option
93
End Frame Set end frame.
Measurement Functions
NOTE:
The measurement function will not make automatic
measurements when the instrument is in SCAN mode.
Time base must be ≤ 50 ms.
By default, the menu options will show five different measurements in the
menu display area. Each of these 5 measurements can be changed to show
different measurements by selecting its corresponding function key
( - ).
94
Table 36 - Measurement Selection Menu
Menu
Setting Description
Option
Select input signal source for
Source CH1,CH2
voltage measurement.
Vpp, Vmax, Vmin, Vamp, Vtop,
Select the type of voltage
Type Vbase, Vavg, Mean, Vrms, Crms,
measurement.
FOV, FPRE, ROV, RPRE
, , , Display the corresponding icon
, , , for the selected voltage
, , , measurement as well as the
, measured value.
Return Return to the Measure menu.
95
Vpp – Voltage peak-to- Mean – Arithmetic mean
peak over the entire waveform
Vmax – Positive peak Vrms – True rms voltage
voltage over the entire waveform
Vmin – Negative peak Crms – rms voltage of the
voltage first cycle of the waveform
Vamp – Vtop and Vbase FOV – Falling edge
difference overshoot voltage
𝑉𝑚𝑖𝑛 − 𝑉𝑏𝑎𝑠𝑒
𝐹𝑂𝑉 =
𝑉𝑎𝑚𝑝
Vtop – Max. voltage FPRE– Falling edge
during measurement preshoot
𝑉𝑚𝑎𝑥 − 𝑉𝑡𝑜𝑝
𝐹𝑃𝑅𝐸 =
𝑉𝑎𝑚𝑝
Vbase – Min. voltage ROV – Rising edge
during measurement overshoot voltage
𝑉𝑚𝑎𝑥 − 𝑉𝑡𝑜𝑝
𝑅𝑂𝑉 =
𝑉𝑎𝑚𝑝
Vavg – Average voltage RPRE – Rising edge
over first cycle preshoot
𝑉𝑚𝑖𝑛 − 𝑉𝑏𝑎𝑠𝑒
𝑅𝑃𝑅𝐸 =
𝑉𝑎𝑚𝑝
Menu
Setting Description
Option
Select input signal source for
Source CH1, CH2
Time measure.
Period, Freq, +Wid, -Wid, Rise Select the type of time
Type
Time, Fall Time, BWid, +Dut, - measurement.
96
Dut
Display the corresponding icon
, ,
for the selected time
, ,
measurement as well as the
, ,
measured value.
Return Return to the Measure menu.
Menu
Setting Description
option
Select input signal source for
Source CH1, CH2
delay measure.
Type Phase, FRR, FRF, FFR, FFF, Select the type of delay
97
LRR, LRF, LFR, LFF measurement.
Display the corresponding icon
, ,
for the selected delay
, ,
measurement as well as the
, ,
measured value.
Return Return to the Measure menu.
You can enable all voltage measurements, all time measurements, or all
delay measurements to display on the screen. You can also show all 32
total automatic measurements on the display.
98
Table 40 - All Measurement Display Menu
Menu
Setting Description
option
Select input signal source for
Source CH1, CH2
delay measure.
Select to enable/disable
On
Voltage displaying all voltage
Off
measurements.
Select to enable/disable
On
Time displaying all voltage
Off
measurements.
Select to enable/disable
On
Delay displaying all delay
Off
measurements.
Return Return to the Measure menu.
99
Figure 50 - All Time Measurements Display
100
Cursor Functions
There are three cursor modes available: Manual, Track, and Auto.
Manual cursor mode allows users to set either horizontal or vertical cursors
to make measurements from a selected signal source.
From the Cursor menu, set Mode to Manual. Select the Type of cursor for
the measurement, and select Source.
Note: MATH, REFA, and REFB are not selectable as Source unless they are
enabled. Refer to “3.4 Scope Menu” to enable them.
Menu
Setting Description
Option
Mode Manual Set to manual cursor mode.
Use cursors to measure voltage
Voltage
Type parameters.
Time Use cursors to measure time parameters.
Source CH1, CH2, Choose the signal source to measure with
101
MATH, REFA, cursors.
REFB
5
4 1
3
1 Cursor A
2 Cursor B
3 Cursor A position
4 Cursor B position
5 Voltage difference between Cursor A and Cursor B
102
Time Type – Time cursors are used to make cursor measurements. Two
vertical cursor lines will appear on the display, and their positions and
differences will be indicated in the upper left corner of the display.
6
5
1
4
3
1 Cursor B
2 Cursor A
3 Cursor A time position
4 Cursor B time position
5 Frequency between Cursor A and Cursor B
6 Time difference between Cursor A and Cursor B
103
Track Cursor Mode
Track cursor mode allows users to setup two cursors with both time and
voltage tracked from selected signal sources.
From the Cursor menu, set Mode to Track. Select the source for Cursor A
and the source for Cursor B by toggling and respectively.
Menu
Setting Description
Option
Mode Track Set to track cursor mode.
CH1, CH2,
Cursor A Select signal source for cursor A to measure.
NONE
CH1, CH2,
Cursor B Select signal source for cursor B to measure.
NONE
104
9
8
7
6
5
4
3
2
1
1 Cursor B
2 Cursor A
3 Voltage difference between Cursor A and Cursor B
4 Frequency between Cursor A and Cursor B
5 Time difference between Cursor A and Cursor B
6 Cursor B voltage position
105
In this mode, the cursor(s) will be displayed when selecting automatic
measurements.
Select any of the five measurements displayed in the menu display area,
then select any measurement parameter types.
Selecting Vpp as the voltage measurement Type will show the following
display, with two cursors shown on display to illustrate their positions to
which the measurements are being made.
Selecting Period as the time measurement Type will show the following
display, with two cursors shown on display to illustrate their positions to
which the measurements are being made.
106
Figure 57 - Auto Cursor Measuring Period
107
4 Using the Digital Multimeter
The handheld digital storage oscilloscope has a built-in multimeter that can
measure DC and AC voltage, DC and AC current, resistance, diode,
continuity, and capacitance.
108
Note: In the digital multimeter mode, only the function
keys ( - ), , and
CAUTION:
Always use only the included test leads to make
measurements. Verify that test leads are in
proper working conditions and that they meet the
specified electrical ratings and input requirements
of the instrument.
109
Table 43 – DC/AC Voltage Measurement Menu
DC Voltage Measurement
key to increase the range. At the highest range, pressing this will
automatically go back to the lowest range.
110
Figure 59 - Out of Range
CAUTION:
Always connect the test leads to the instrument
inputs first before probing the DUT to avoid
potential shock hazard.
10 A mA COM V.Ω.C
- +
DC Voltage
111
WARNING:
Never connect more than 1000V DC across the
input terminals.
AC Voltage Measurements
key to increase the range. At the highest range, pressing this will
automatically go back to the lowest range.
CAUTION:
Always connect the test leads to the instrument
inputs first before probing the DUT to avoid
potential shock hazard.
112
10 A mA COM V.Ω.C
AC Voltage
WARNING:
Never connect more than 750V AC across the
input terminals.
DC Current Measurements
key to increase the range. At the highest range, pressing this will
automatically go back to the lowest range.
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
114
10 A mA COM V.Ω.C
A
+ -
DC Current
Figure 62 - Connection for Low DC Current Measurement
WARNING:
Do not connect more than 600 mA DC current
across the mA input terminal or the protection
fuse will trip. Never input 10 A or more current
across the terminal or the instrument will be
damaged and warranty will be void.
10 A mA COM V.Ω.C
A
+ -
DC Current
Figure 63 - Connection for Higher DC Current Measurement
115
WARNING:
Do not connect more than 10 A DC current across
the 10A input terminal or the protection fuse will
be tripped.
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
10 A mA COM V.Ω.C
A
+ -
DC Current
Figure 64 - Connection for Low AC Current Measurement
116
WARNING:
Do not connect more than 600 mA AC current
across the mA input terminal or the protection
fuse will be tripped. Never input 10 A or more
current across the terminal or the instrument will
be damaged and warranty will be void.
10 A mA COM V.Ω.C
A
+ -
DC Current
Figure 65 - Connection for Higher AC Current Measurement
WARNING:
Do not connect more than 10 A DC current across
the 10A input terminal or the protection fuse will
be tripped.
117
4.3 Make Resistance Measurements
key to increase the range. At the highest range, pressing this will
automatically go back to the lowest range.
Resistance ranges: 600 Ω, 6 kΩ, 60 kΩ, 600 kΩ, 6 MΩ, and 60 MΩ.
118
NOTE: When the measurement unit is changed, range
will automatically change to Auto ranging.
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
10 A mA COM V.Ω.C
Resistance
WARNING:
Do not apply more than 1000 VDC across the
terminals or they will be damaged.
119
4.4 Make Diode Measurements
The instrument can measure the forward voltage of general purpose
diodes.
WARNING:
Fully discharge the capacitor before connecting it
to any of the inputs or it may damage the
instrument.
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
120
10 A mA COM V.Ω.C
Diode
WARNING:
Do not apply more than 1000 VDC across the
terminals or they will be damaged.
121
WARNING:
Fully discharge the capacitor before connecting it
to any of the inputs or it may damage the
instrument.
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
10 A mA COM V.Ω.C
Continuity
WARNING:
Do not apply more than 1000 VDC across the
terminals or they will be damaged.
122
4.6 Make Capacitance Measurements
WARNING:
Fully discharge the capacitor before connecting it
to any of the inputs or it may damage the
instrument.
123
CAUTION:
Always connect the test leads to the instrument
inputs first before connecting the DUT to avoid
potential shock hazard.
NOTE:
The capacitance measurement function is used for
measuring general purpose electrolytic capacitors.
10 A mA COM V.Ω.C
Capacitor
WARNING:
Do not apply more than 1000 VDC across the
terminals or they will be damaged.
124
5 Using the Recorder
There are three categories under the recorder function: Scope Trend Plot,
Scope Recorder, and Meter Trend Plot.
NOTE:
The oscilloscope cannot be in SCAN mode when
using scope trend plot. Before selecting the scope
trend plot function, check that the oscilloscope is
not in SCAN mode which will be indicated at the
top left of the display.
NOTE:
When in scope trend plot mode, only the function
keys and the mode keys ( , ,
)are active.
125
From the Recorder menu, press to select Scope Trend Plot. The
scope trend plot screen and menu will be displayed.
126
Record Data with Trend Plot
For time measurement parameters, the menu and selections are the
same as in “Table 38 - Time Measurement Parameters Menu”.
127
to selection NO or press to select YES. Selecting
NO will cancel your changes and return to the Scope Trend Plot
menu.
NOTE:
If YES is selected, all data that has been recorded
previously will be deleted, including all data in
Param B.
Pause Recording
Restart Recording
The display settings on the trend plot can be changed. Users can select
viewing the data up to the minute, or select viewing all data in a
compressed format.
To toggle between the two display modes, go to the second page of the
Scope Trend Plot menu and press to toggle between Normal and
View All.
128
Record Data by Manual Trigger
When set to On, the plot will stop recording. Instead, the button
will function as a manual trigger button. Each subsequent press will trigger
one measurement recording to the plot (both Param A and Param B).
NOTE:
In this mode, the button will be used as a
trigger instead of a means to enter the Recorder
menu. To exit out of Scope Trend Plot, press
to select Return. All previously recorded
data will be lost if you exit. A confirmation
message will ask if you want to exit.
The recorded data on the scope trend plot is stored temporarily. All
recorded data will be lost if the recording gets reset, exit out of the scope
trend plot mode, or power OFF the instrument.
However, the data can be saved onto an external USB flash drive.
To save:
1. Go to the second page of the Scope Trend Plot menu and select
CSV by pressing .
2. The Directory menu will be displayed.
129
3. Follow the same instructions from section “3.7 Save/Recall
Function” to create a directory and/or a new file, then select Load
from the menu to save all recorded data into a new file on the
external USB flash drive.
NOTE:
Depending on the amount of data recorded, the
saving process may take a while. Please wait and
do not interrupt the instrument while it is saving
data. When finished, a message will prompt to
notify the file has been saved successfully.
To exit out of the Directory menu, go to the second page and select
Return.
Scope Recorder
The scope recorder allows users to record the oscilloscope’s channel input
signals for a long period of time. This function is available when the scope is
configured with slower timebase settings.
130
NOTE:
The Scope Recorder function is only available when
the oscilloscope is in SCAN mode. SCAN mode is
automatically enabled when the oscilloscope
timebase is set between 100 ms to 50 s.
Before entering the Scope Recorder menu, verify that the oscilloscope is
operating in SCAN mode. Press to enter the oscilloscope
operation mode and set SCAN mode. SCAN mode is available only for
timebase 100 ms to 50 s.
Menu
Setting Description
Option
Start Begin recording CH1 and/or CH2 waveforms.
Replay Select to replay the recorded waveform.
131
Copy recording saved in internal memory to
Copy
external USB flash drive.
Save Memory Copy recording to internal memory.
mode USB Key Copy recording to external USB.
Return Return to the Scope Recorder menu.
Menu
Setting Description
Option
Record and replay channel waveform with full
Full screen
screen.
Viewer Record and replay channel waveform with a
Split split screen. The top half will display CH1 and
the bottom half will display CH2.
Continuously record data. When memory is
continuous
Record full, the oldest data will be overwritten.
Single Stop recording data when memory is full.
When replaying, the screen waveform updates
Replay By Point
every dot from left to right.
132
When replaying, the screen waveform updates
By frame the whole screen according to the time of
sampling every frame of data.
Return Return to the Scope Recorder menu.
Press to set the Viewer for Full Screen or Split view. Then, set
Record to continuous or Single by pressing .
If USB key is selected for Save Mode, File Name will be shown in the menu
with a file name starting with BKXXXXX where XXXXX will be an auto
increment number starting with 00001. If no external USB flash drive is
detected, File Name will indicate No U disk.
133
Press to select Start. The instrument will start scope recording.
The top left will indicate the record time, which tracks the time length of
the recording. The remaining time is also indicated to track how much
longer the instrument can record before memory is full.
To playback the recorded waveforms, from the Scope Recorder menu, press
to select Record. Then press to select Replay. The
waveform will begin to replay on the display.
134
Figure 74 - Playback Scope Recording
When the replay is finished, a message will say “This is the end of the
waveform”.
To view different portions of the recording, use the Previous and Next
menu options by pressing or respectively to go backward
or forward.
135
Meter Trend Plot
Meter Trend Plot allows storing of measurement data and plotting it into a
graph when in multimeter mode. The data can be stored into external
memory using a USB flash drive connected to the USB host port.
NOTE:
The measurement function (i.e. DCV, ACV, Diode,
etc.) that the meter trend plot will store and plot
depends on the function configured within the
multimeter mode. Therefore, it is important to
set up the multimeter first with the measurement
function you want to use with trend plot.
By default, the trend plot will continuously store and plot data when you
enter the Meter Trend Plot menu.
136
1 2 3 4
1
Recording/Trend Plot time and percentage of temporary storage
memory used.
2 Measured data value
3 Measurement function
4 Battery indicator
5 Range mode indicator
6 Measured data plot
7 Current measured value and time
137
Menu
Setting Description
Option
Trend Plot Restart Restart the trend plot recording.
10Sa/s, 5Sa/s,
2Sa/s, 1Sa/s, Set the sampling rate of the trend plot
Sa Rate
0.5Sa/s, recording.
0.2Sa/s
Display the trend plot recorded data up to
Normal
the minute.
View All Display all trend plot recorded data.
Run Record data automatically.
Stop Stop recording data.
Enter the second page of the Meter Trend
Next Page Page1/2
Plot menu.
138
Record Data with Trend Plot
2. Select the sampling rate at which you want to record and plot the
measurement data.
Pause Recording
Restart Recording
The display settings on the trend plot can be changed. Users can select
viewing the data up to the minute, or select viewing all data in a
compressed format.
139
Record Data by Manual Trigger
When set to On, the plot will stop recording. Instead, the button
will function as a manual trigger button. Each subsequent press will trigger
one measurement recording to the plot.
NOTE:
In this mode, the button will be used as a
trigger instead of entering the Recorder menu. To
exit out of Meter Trend Plot, press to
select Return in the second page of Meter Trend
Plot menu. All previously recorded data will be
lost if you exit. A confirmation message will ask if
you want to exit.
The recorded data on the meter trend plot is stored temporarily. All
recorded data will be lost if the recording gets reset, exit out of the meter
trend plot mode, or power OFF the instrument.
However, the data can be saved onto an external USB flash drive.
To save:
1. Go to the second page of the Meter Trend Plot menu and select CSV
by pressing .
2. The Directory menu will be displayed.
140
3. Follow the same instructions from section 3.7 “Save/Recall
Function” to create a directory and/or new file, then select Load
from the menu to save all recorded data into a new file on the
external USB flash drive.
NOTE:
Depending on the amount of data recorded, the
saving process may take a while. Please wait and
do not interrupt the instrument while it’s saving.
When finished, a message will be displayed to
notify the file has been saved successfully.
To exit out of the Directory menu, go to the second page and select
Return.
141
6 Remote Communication
The 2510 series digital storage oscilloscope comes with application software
which provides most of the controls that emulate the front panel of the
instrument. The miniUSB device port on the side of the instrument, when
not used for probe compensation, can be used to connect to a computer
using a miniUSB to USB type A cable and allow remote communication via
the software.
142
7 Troubleshooting Guide
143
File successfully saved.
Ready Data Success!
Setup data or waveform data from the internal storage of the
oscilloscope or USB flash drive was read successfully.
Record Wave Success!
This message will appear when you finish recording waveforms.
Oscilloscope
1. I pressed the auto button, but it says no signal. Is it malfunctioning?
The auto function does not work for all waveforms. Usually,
waveforms that are periodic can be detected.
144
Multimeter
1. After I change a menu selection, how do I set the changes and close
the menu window?
145
8 Specifications
Oscilloscope Specifications
All specifications apply to measurements with a 10X probe. To verify that
the oscilloscope meets specifications, the oscilloscope must first meet the
following conditions:
The oscilloscope must have been operating continuously for 30
minutes within the specified operating temperature.
If the operating temperature changes by more than 5° C, you must
perform the Do Self Cal operation, accessible through the Utility
menu.
146
Max. BNC Input
CATII 300 V RMS from BNC signal to BNC shell
Voltage
Max. input (4)1x/10x CATII 300 V RMS (PR150SA)
Voltage for (5)(6)10x CATIII 600 V RMS, CATII 1000 V RMS (PR250SA)
probe
2mV - 200mV : ±1.6V 5mV - 200mV : ±1.6V
Channel Voltage
206mV - 10V : ±40V 206mV - 10V : ±40V
Offset Range
10.2V - 100V : ±400V 10.2V - 100V : ±400V
Bandwidth Limit 20MHz (-3dB)
Horizontal Scan 5.0nS/div - 2.5nS/div - 5.0nS/div - 2.5nS/div -
Range 50S/div 50S/div 50S/div 50S/div
Timebase
± 50 ppm measured over 1 ms interval
Accuracy
Input Coupling AC, DC, GND
Input 1 MΩ +/- 2 % || 18 pF ± 3 1 MΩ +/- 2 % || 16 pF ± 3
Impedance pF pF
Probe
Attenuation
1X, 5X, 10X, 50X, 100X, 500X, 1000X
Selectable
Factors
Vertical and
Vertically or horizontally expand or compress a live or
Horizontal
stopped waveform.
Zoom
I/O Interface
USB host port support USB flash drives (FAT format).
USB
USB device port for PC connectivity
Acquisition Modes
Sampling Display sample data only
Peak Detect Capture the maximum and minimum values of a signal
Waveform averaged, selectable from 4, 16, 32, 64, 128,
Average
256
Trigger System
Edge, Pulse Width, Video*, Slope, Alternative
147
Trigger Source CH1, CH2
Trigger Level
CH1, CH2: ± 6 divisions from center of display
Range
Trigger Pre-trigger: Memory depth/ 2* sampling
Displacement Delay Trigger: 268.04 div
Holdoff Range 100 ns – 1.5 s
Pulse Width Trigger Modes: Positive Pulse (>,<, =), Negative Pulse (>,
Trigger <, =)
Slope Trigger Positive slope (>, <, =), Negative slope (>, <, =)
Time: 20 ns - 10 s
Alternate CH1 trigger type: Edge, Pulse, Video, Slope
Trigger CH2 trigger type: Edge, Pulse, Video, Slope
Hardware Frequency Counter
Reading 1 Hz
Resolution
Range DC couple, 10 Hz to MAX bandwidth
Signal Types Satisfying all trigger signals (except pulse width trigger
and video trigger)
Waveform Math and Measure
Math Operation Add, Subtract, Multiply, Divide, FFT
FFT Window mode: Hanning, Hamming, Blackman,
Rectangular
Sampling points: 1024
Measure Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean, Crms,
Vrms, ROVShoot, FOVShoot, RPREShoot, FPREShoot,
Rise, Fall, Freq, Prd, +Wid, -Wid, +Dut, -Dut, BWid, Phas,
FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF
Cursors
Types Voltage, Time
Measurements ΔV, ΔT, 1/ΔT (frequency)
Display System
Display 5.7” Color TFT, 320 x 234 resolution, 64K color
Display Contrast 150:1
(Typical)
Backlight 300 nits
Intensity
(Typical)
148
Wave Display 8 x 12 div
Range
Wave Display Dots, Vector
Mode
Persistence Off, 1 sec, 2 sec, 5 sec, Infinite
Menu Display 2 sec, 5 sec, 10 sec, 20 sec, Infinite
Screen-Saver Off, 1 min., 2 min., 5 min., 10 min., 15 min., 30 min., 1 hr,
2 hr, 5 hr
Waveform Sin(x)/x, Linear
Interpolation
Measure
Main, Window zoom, Scan, X-Y
Display Modes
X-Y Sampling Support 25 kSa/s - 250 MSa/s sampling rate (1-2.5-5
Frequency order)
Color Mode Normal, Invert
(1) Half channel operation means that only Ch1 or Ch2 is active.
(2) When sampling rate is 1 GSa/s. For sampling rate ≤ 500 MSa/s, the
maximum memory depth is 20 kpts.
(3) Available when sampling rate is < 500 MSa/s and maximum data depth
mode is enabled.
(4) Probe included with models 2511 and 2512 only.
(5) Probe included with models 2515 and 2516 only.
(6) Refer to the respective probe’s manual for more information on the
specification.
149
Maximum Input Voltages
Maximum Input Voltage vs. Frequency
The chart below shows the maximum input voltage vs. frequency.
150
Maximum Input Voltage between Scope Reference and Scope Reference
to ground
The chart below shows the maximum input voltage between scope
reference and scope reference to ground for models 2515 and 2516 only.
151
Multimeter Specifications
All specifications are based on operating at temperatures 23 ± 5°C
and relative humidity < 75%.
Accuracy is based on ± (% of reading + offset)
Input Impedance 10 MΩ
Max. Input
Voltage between CAT I, CAT II 600 V RMS
Multimeter input CAT III 300 V RMS
reference and
ground
Voltage/Resistance/Capacitance port: DC 1000 V, AC
Max. input
750 V RMS
Voltage for
Multimeter input Current port (mA): AC 250 V, 600 mA
ports Current port (10A): AC 250 V, 10 A
DC Voltage
Range Resolution Accuracy
60.00 mV 10 µV ± (1 % + 15 digit)
600.0 mV 100 µV
6.000 V 1 mV
60.00 V 10 mV ± (1 % + 5 digit)
600.0 V 100 mV
1000 V 1V
152
(1)ACVoltage
Range Resolution Accuracy
60.00 mV 10 µV ± (1 % + 15 digit)
600.0 mV 100 µV
6.000 V 1 mV
60.00 V 10 mV ± (1 % + 5 digit)
600.0 V 100 mV
750 V 1V
1. For frequency range 20 Hz to 1 kHz.
Resistance
Range Resolution Accuracy
600.0 Ω 0.1 Ω
6.000 kΩ 1Ω
60.00 kΩ 10 Ω
± (1 % + 5 digit)
600.0 kΩ 100 Ω
6.000 MΩ 1 kΩ
60.00 MΩ 10 kΩ
Capacitance
Range Resolution Accuracy
40.00 nF 10 pF ± (3 % + 10 digit) For > 5 nF
400.0 nF 100 pF
4.000 µF 1 nF
± (4 % + 5 digit)
40.00 µF 10 nF
400.0 µF 100 nF
153
(2)(3)DCand AC Current
Range Resolution Accuracy
60.00 mA 10 µA
± (1 % + 5 digit)
600.0 mA 100 µA
6.000 A 1 mA
± (1.5 % + 5 digit)
10.00 A 10 mA
2. For 10 A terminal, > 6 A DC or AC rms for 10 seconds ON and 15 minutes
OFF.
3. For AC current ranges, frequency is verified for 20 Hz to 1 kHz.
Recorder Specifications
Scope Recorder
Display Mode Full view, Normal
Single Channel: 7M points
Max. Record Length
Dual Channel: 3.5 M points
Number of Channels 2
Maximum Record Size 4GB, 3000 hours
to External Storage
154
General Specifications
Environmental and Safety
Temperature Operating: 32 °F to 104 °F (0 °C to +40 °C)
Not operating: -4 °F to 158 °F (-20 °C to +70 °C)
Humidity Operating: 85% RH, 104 °F (40 °C), 24 hours
Altitude Operating: 9,842.5 ft (3,000 m)
Electromagnetic EMC Directive 2004/108/EC,
Compatibility EN61326:2006
Safety Low voltage directive 2006/95/EC, EN61010-1:2001
General
Storage 2 reference waveforms, 20 setups, 10 waveforms
Memory
AC adapter Input: 100-240 VAC, 50/60 Hz
Power Output: 9V DC, 4 A
Requirements
Battery 5000 mAh, 7.4 VDC
Charge Time Approx. 4 hrs
Dimensions (W 6.42” x 10.21” x 2.10” (163.2 x 259.5 x 53.3 mm)
x H x D)
Weight Approx. 3.4 lbs (1.54 kg) including battery
155
11 Maintenance
Do not expose the LCD display to direct sunlight for long periods of time.
Cleaning
If the instrument requires cleaning, disconnect it from all power sources
and clean only with a mild detergent and water. Be sure the instrument is
completely dry before reconnecting it to any power source.
156
SERVICE INFORMATION
Warranty Service: Please go the support and service section on our website www.bkprecision.com to
obtain an RMA #. Return the product in the original packaging with proof of purchase to the address
below. Clearly state on the RMA the performance problem and return any leads, probes, connectors
and accessories that you are using with the device.
Non-Warranty Service: Please go the support and service section on our website
www.bkprecision.com to obtain an RMA #. Return the product in the original packaging to the
address below. Clearly state on the RMA the performance problem and return any leads, probes,
connectors and accessories that you are using with the device. Customers not on an open account
must include payment in the form of a money order or credit card. For the most current repair
charges please refer to the service and support section on our website.
Return all merchandise to B&K Precision Corp. with pre-paid shipping. The flat-rate repair charge for
Non-Warranty Service does not include return shipping. Return shipping to locations in North
America is included for Warranty Service. For overnight shipments and non-North American shipping
fees please contact B&K Precision Corp.
Include with the returned instrument your complete return shipping address, contact name, phone
number and description of problem.
LIMITED THREE-YEAR WARRANTY
B&K Precision Corp. warrants to the original purchaser that its products and the component parts
thereof, will be free from defects in workmanship and materials for a period of three years from date
of purchase.
B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or
component parts. Returned product must be accompanied by proof of the purchase date in the form
of a sales receipt.
To help us better serve you, please complete the warranty registration for your new instrument via
our website www.bkprecision.com
Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a
result of unauthorized alterations or repairs. The warranty is void if the serial number is altered,
defaced or removed.
B&K Precision Corp. shall not be liable for any consequential damages, including without limitation
damages resulting from loss of use. Some states do not allow limitations of incidental or
consequential damages. So the above limitation or exclusion may not apply to you.
This warranty gives you specific rights and you may have other rights, which vary from state-to-state.