4008 S Hemodialysis Device Service Manual: Software Version: 11.40 and Higher Edition: 6/06.12 Part No.: M47 959 1
4008 S Hemodialysis Device Service Manual: Software Version: 11.40 and Higher Edition: 6/06.12 Part No.: M47 959 1
4008 S Hemodialysis Device Service Manual: Software Version: 11.40 and Higher Edition: 6/06.12 Part No.: M47 959 1
Hemodialysis device
Service Manual
Software version: 11.40 and higher
Edition: 6/06.12
Part no.: M47 959 1
0123
Table of contents
1 Index
2 Important information
2.1 How to use the Service Manual................................................................................................ 2-1
2.2 Significance of the warning...................................................................................................... 2-1
2.3 Significance of the note ............................................................................................................ 2-2
2.4 Precautions for working on the device ................................................................................... 2-2
2.5 Technical documentation ......................................................................................................... 2-3
2.6 Warnings .................................................................................................................................... 2-3
2.6.1 Warnings relating to hygiene ....................................................................................................... 2-3
2.6.2 Warnings relating to electricity..................................................................................................... 2-3
2.7 Addresses .................................................................................................................................. 2-4
3 Installation
3.1 Important initial start-up information ...................................................................................... 3-1
3.2 Important information for use in a domestic environment.................................................... 3-2
3.3 Initial start-up report ................................................................................................................. 3-2
3.4 Explanations for the initial start-up report.............................................................................. 3-6
3.5 Transportation of hemodialysis unit ..................................................................................... 3-28
4 Specifications
4.1 Dimensions and weight ............................................................................................................ 4-1
4.2 Type label (identification of the device) .................................................................................. 4-2
4.3 Operating environment ............................................................................................................. 4-3
4.4 Electrical safety ......................................................................................................................... 4-3
4.5 Electrical supply ........................................................................................................................ 4-4
4.6 Fuses .......................................................................................................................................... 4-4
4.7 Supply mains ............................................................................................................................. 4-5
4.8 Guidance and manufacturer’s declaration for EMC (IEC 60601-1-2) .................................... 4-7
4.8.1 Electromagnetic emissions.......................................................................................................... 4-7
4.8.2 Electromagnetic immunity ........................................................................................................... 4-7
4.8.3 Recommended separation distances between portable and mobile RF telecommunication devices
and the device ............................................................................................................................. 4-9
6 TSC / maintenance
6.1 Important information ............................................................................................................... 6-1
6.2 Accessories and supplies ........................................................................................................ 6-2
6.3 TSC / MA test report .................................................................................................................. 6-2
6.4 Explanations on the test report – TSC / MA
TSC, explanations on test report ............................................................................................. 6-6
6.5 Test report – TSC..................................................................................................................... 6-32
7 Error messages
7.1 T1 Test ........................................................................................................................................ 7-1
7.1.1 Prerequisites for starting and running the test............................................................................. 7-1
7.1.2 Bypass test .................................................................................................................................. 7-1
7.1.3 Opt. detector test ......................................................................................................................... 7-2
7.1.4 Blood system test ........................................................................................................................ 7-3
7.1.5 Ven. pressure system test ........................................................................................................... 7-5
7.1.6 Air detector test ........................................................................................................................... 7-6
7.1.7 Display test .................................................................................................................................. 7-8
7.1.8 Art. pressure system test............................................................................................................. 7-8
7.1.9 Battery test .................................................................................................................................. 7-8
7.1.10 Blood leak test ............................................................................................................................. 7-9
7.1.11 Temperature test ......................................................................................................................... 7-9
7.1.12 Negative pressure test (negative pressure holding test) ........................................................... 7-10
7.1.13 Positive pressure test (positive pressure holding test) .............................................................. 7-11
7.1.14 UF function test ......................................................................................................................... 7-14
7.1.15 Conductivity test ........................................................................................................................ 7-15
7.1.16 Test DIASAFE® plus / HPU Test ............................................................................................... 7-16
7.2 Device error during cleaning programs ................................................................................ 7-18
7.2.1 V84 monitoring .......................................................................................................................... 7-18
7.2.2 PSW (pressure switch) monitoring during free rinsing (only with devices with CDS) ................ 7-19
7.2.3 Rinse section test (check of V91, V99, V100) (only devices with CDS) .................................... 7-22
9 Calibration / adjustment
9.1 CALIBRATION ............................................................................................................................ 9-1
9.1.1 Settings without menu display ..................................................................................................... 9-1
9.1.2 CALIBRATION menu structure.................................................................................................... 9-2
9.1.3 Part 1: Calibrating the arterial pressure ....................................................................................... 9-4
9.1.4 Part 2: Calibrating the pressure in the arterial blood pump ......................................................... 9-5
9.1.5 Without menu display: Setting the blood pump stop alarm.......................................................... 9-6
9.1.6 Part 3: Calibrating the venous pressure ...................................................................................... 9-7
9.1.7 Part 4: Calibrating the venous pressure measurement in the air detector................................... 9-8
9.1.8 Part 5: Calibrating the blood pump rate ....................................................................................... 9-9
9.1.9 Part 5.1: Calibrating the arterial blood pump rate ........................................................................ 9-9
9.1.10 Part 5.2: Calibrating the SN blood pump rate (option)............................................................... 9-10
9.1.11 Without menu display: Calibrating the Single-Needle blood pump (Single-Needle pressure)
(optional).................................................................................................................................... 9-11
9.1.12 Part 6: Measuring the volume of the UF pump in liters.............................................................. 9-12
9.1.13 Part 7: Calibrating the degassing pressure................................................................................ 9-13
9.1.14 Part 8: Calibrating the 300 flow rate .......................................................................................... 9-13
9.1.15 Without menu display: Adjusting the current increasing pulse .................................................. 9-14
9.1.16 Alternative method of adjusting the current increase (if an oscilloscope is not available) ......... 9-14
9.1.17 Part 9: Calibrating the 500 flow rate .......................................................................................... 9-15
9.1.18 Part 10: Calibrating the 800 flow rate ........................................................................................ 9-16
9.1.19 Part 11: Calibrating the dialysate temperature .......................................................................... 9-16
9.1.20 Part 11.1: Setting the dialysate temperature ............................................................................. 9-17
9.1.21 Part 11.2: Testing the dialysate temperature............................................................................. 9-18
9.1.22 Part 11.3: Testing the dialysate temperature for OCM .............................................................. 9-19
9.1.23 Part 12: Calibrating the mixing system ...................................................................................... 9-20
9.1.24 Part 12.1: Running in the membrane pumps ............................................................................. 9-20
9.1.25 Part 12.2: Determining the balancing chamber volume............................................................. 9-21
9.1.26 Part 12.3: Calibrating the concentrate pump stroke .................................................................. 9-21
9.1.27 Part 12.4: Measuring the volume of the concentrate pump in liters........................................... 9-22
9.1.28 Part 12.5: Calibrating the bicarbonate pump stroke .................................................................. 9-23
9.1.29 Part 12.6: Measuring the volume of the bicarbonate pump in liters........................................... 9-24
9.1.30 Part 12.7: Testing the concentrate and bicarbonate volumes ................................................... 9-25
9.1.31 Part 13: Calibrating the conductivity .......................................................................................... 9-26
9.1.32 Part 13.1: Conductivity setting ................................................................................................... 9-27
9.1.33 Part 13.2: Temperature/conductivity compensation setting....................................................... 9-28
10 Servicing / repair
10.1 Precautions for working on the hemodialysis device.......................................................... 10-1
10.2 Equipment ................................................................................................................................ 10-2
10.3 Component overview .............................................................................................................. 10-3
10.3.1 Monitor....................................................................................................................................... 10-3
10.3.2 Extracorporeal Blood Circuit Module (EBM).............................................................................. 10-4
10.3.3 Hydraulics rear .......................................................................................................................... 10-5
10.3.4 Hydraulics, lateral view from the left.......................................................................................... 10-6
10.3.5 Hydraulics, lateral view from the right........................................................................................ 10-7
10.3.6 Hydraulics legend ...................................................................................................................... 10-8
10.3.7 Assignment of the hydraulic processing unit (HPU) .................................................................. 10-9
10.4 Assembly of components..................................................................................................... 10-10
10.4.1 Self-cutting screws .................................................................................................................. 10-10
10.4.2 Torques ................................................................................................................................... 10-10
10.5 Housing and cart ................................................................................................................... 10-11
10.5.1 Tilting the device...................................................................................................................... 10-11
10.5.2 Brake rollers ............................................................................................................................ 10-11
10.5.3 Brakes ..................................................................................................................................... 10-12
10.5.4 Shunt interlock......................................................................................................................... 10-13
10.5.4.1 Shunt interlock complete .......................................................................................... 10-13
10.5.4.2 Microswitch............................................................................................................... 10-14
11 Functional description
11.1 Functional description ............................................................................................................ 11-1
11.1.1 T1 test description ..................................................................................................................... 11-1
11.1.1.2 T1 test flow diagram, parallel run ............................................................................... 11-3
11.1.1.3 Overview of the individual test sections ..................................................................... 11-4
11.1.1.6 Blood system test ....................................................................................................... 11-7
11.1.1.7 Venous pressure system test ..................................................................................... 11-8
11.1.1.10Arterial pressure system test.................................................................................... 11-11
11.1.1.16UF function test ........................................................................................................ 11-17
11.1.2 Tests during cleaning programs .............................................................................................. 11-20
11.1.2.1 PSW (pressure switch) monitoring during free rinsing
(only with devices with CDS).................................................................................... 11-20
11.1.2.2 Rinse section test (check of V91, V99, V100) (only devices with CDS)................... 11-21
11.1.2.4 V39 Test ................................................................................................................... 11-23
11.1.3 Description of the modules ...................................................................................................... 11-24
11.1.3.1 Arterial blood pump .................................................................................................. 11-24
11.1.3.2 Single-Needle blood pump (optional) ....................................................................... 11-24
11.1.3.3 Heparin pump........................................................................................................... 11-25
11.1.3.4 Air Detector .............................................................................................................. 11-26
11.1.3.5 Optical Detector ....................................................................................................... 11-27
11.1.3.6 Venous pressure measurement ............................................................................... 11-27
11.1.4 Functional description of the hydraulic unit ............................................................................. 11-28
11.1.4.1 Hydraulic flow diagram ............................................................................................. 11-28
11.1.4.2 Description of the hydraulic unit ............................................................................... 11-30
11.1.4.3 Functional principle of the balancing chamber ......................................................... 11-32
11.1.4.4 CDS - Central Delivery System (option)................................................................... 11-35
11.2 Block diagrams / wiring diagrams / P.C.B. layouts ............................................................ 11-43
11.2.1 4008 S block diagram.............................................................................................................. 11-43
11.2.2 Voltage supply block diagram.................................................................................................. 11-44
11.2.3 Monitor block diagram ............................................................................................................. 11-45
11.2.4 HPU block diagram (hydraulic processing unit)....................................................................... 11-46
11.2.5 HPU wiring diagram (hydraulic processing unit)...................................................................... 11-47
11.2.6 CAN communication connection diagram ............................................................................... 11-49
11.2.7 P.C.B. LP 450-2 Air detector control (LD) ............................................................................... 11-50
11.2.8 P.C.B. LP 493 Blood leak detector .......................................................................................... 11-52
11.2.9 P.C.B. LP 624 pump control, arterial BP / Single-Needle BP (option)..................................... 11-53
11.2.10 P.C.B. LP 630 motherboard .................................................................................................... 11-54
11.2.11 P.C.B. LP 632 CPU 2 .............................................................................................................. 11-55
11.2.12 P.C.B. LP 633-5 Input board ................................................................................................... 11-57
11.2.13 P.C.B. LP 634 Output board.................................................................................................... 11-59
12 Appendix
1 Index
The source specifications in the index (e.g., 11-3) refer to the page identification in the footer. In this
example, the index entry can be found in chapter 11 on page 3.
2 Important information
Purpose This document is intended for service technicians and is to be used for
first studies (to acquire a basic knowledge) and for reference purposes
(for TSC, maintenance and repair). The document, however, does not
replace the training courses offered by the manufacturer.
Identification The document can be identified by the following information on the title
page and on the labels, if any:
– Edition of the document
– Part number of the document
Organization of the To facilitate the use of documents from Fresenius Medical Care, the
chapters organization of the chapters has been standardized in all manuals.
There may therefore be chapters within this document without any
content. Chapters without content are identified.
Illustrations The illustrations used in the documents may differ from the original if
this does not have any influence on the function.
Warning
Advises the operator that failure to observe this information can result
in personal injury.
Note
Advises the operator that failure to observe this information can:
– cause damage to the equipment.
– result in a specific function not being executed at all or not being
executed correctly.
Authorized persons
Warning
Commissioning steps, extensions, adjustments, calibrations,
maintenance measures, modifications or repairs may only be carried
out by the manufacturer or persons authorized by him.
Test equipment and The activities described in the document require the availability of the
accessories necessary technical measuring equipment and accessories.
ESD precautions When repairing the device and when replacing spare parts, observe the
applicable ESD precautions.
To be observed after A T1 test and a check of the electrical safety must be performed after
working on the device working on the device.
Warning
When the device is returned to use, check that the pressure of the water
supply meets the prescribed minimum pressure.
2.6 Warnings
Warning
Risk of infection
To protect the patient from infections, disinfect the device each time you
repaired or commissioned the device and each time you carried out
TSC or maintenance measures.
Always run a disinfection after having worked on the device in any way!
Warning
Risk of injury caused by electrical voltage.
Touching live parts will cause an electric shock.
Disconnect the power plug before opening the device. Actuating the
On / Off switch stops operation of the device, but does not
disconnect the device from the supply voltage!
2.7 Addresses
Local service
3 Installation
For initial start-up only The following information is intended only for the initial start-up. This
information is not applicable for recommissioning devices that have
been removed from service or have temporarily been taken out of
service.
Tester's qualification The checks must only be carried out by persons, who, based on their
training, knowledge and on-hand experience, are qualified to carry out
such checks in a proper manner. Furthermore, the persons, who carry
out the checks, may not be subjected to outside instructions with
regards to these checks.
Warning
Commissioning steps, extensions, adjustments, calibrations,
maintenance measures, modifications or repairs may only be carried
out by the manufacturer or persons authorized by him.
Test equipment and The activities described in the document require the availability of the
accessories necessary technical measuring equipment and accessories.
ESD precautions When repairing the device and when replacing spare parts, observe the
applicable ESD precautions.
TSC/ MA intervals The TSC/MA procedures on this device are to be performed after
24 months.
Warning
The power supply's protective conductor must comply with the
IEC 60364-7-710 standard. The following checks must also be made by
suitably-qualified personnel:
– Check the power supply's protective conductor circuit.
The manufacturer points out that local laws take precedence over the
normative standards mentioned above.
Fixed location The location of the hemodialysis device is "fixed" by its connection to
the distinctive power socket (i.e. it is only ever connected to this socket).
Only one distinctive power socket may be set up on any one domestic
premises.
Changing the location in a Where medical care is provided in a domestic environment (home
domestic environment dialysis), the location of the hemodialysis device is fixed. If, following
this fixed-location installation, the hemodialysis device needs to be
moved to a different location, then the requirements must be met as
specified in this chapter.
Remarks:
Remarks:
Warning
On completion of these procedures it is imperative to run a disinfection.
Re 1 Preparation
Operating condition: Off
Re 2 Visual inspection
Operating condition: Off
Re 2.1 Labels and identifications are present and legible (with illustrations
below).
1 2 x label "Type label complete device“
Check device specification:
Serial number
Equipment code
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
25
26 Bed-Side-Link (option):
2 x label: „LAN“ and „CardBox“
26
26
27 Country-specific labels:
27.1 Additional label for the People's Republic of China:
1 x label "CCC" (to the right of the "Type label complete device“ on
the outside)
1 x label "China RoHS" (to the right of the "Type label complete
device“ on the outside)
27.1
27.1
27.2
Re 2.2 Mechanical condition permits further safe use. There are no signs of
damage or contamination affecting proper function of the device.
To be checked:
– Tubings and cuff of BPM (option)
Note
DIP switch 6 (array 1) is provided for service purposes / trouble-
shooting only and must be set to position OFF for dialysis mode.
S1 S2 S3 LP1631
!
O
N
1
2
3
4
5
6
7
8
S1 S2 S3 LP1631
!
O
N
1
2
3
4
5
6
7
8
S1 S2 S3 LP1631
!
O
N
1
2
3
4
5
6
7
8
Check of the DIP switches P.C.B. LP 632 (CPU 2):
Warning
DIP switch 3 (array 1) allows skipping test runs that are requested by
the device.
If the switch is set such that a test run can be skipped, then it must be
noted that the operator has the possibility of evading the automatic test
of the safety systems.
The person requesting such a setting is responsible for this procedure.
S1 S2 LP632
!
O
N
1
2
3
4
5
6
7
8
S1 S2 LP632
!
O
N
1
2
3
4
5
6
7
8
Re 3.4 V91, V99, V100 are functional and tight (test not applicable with
activated CDS or with activated rinse section test).
Operating condition: Off
– (heater rod not active!) Press the Cleaning key. Rinse and reset the
Service switch to OFF (down) after having completed the rinse
program.
Re 3.6 Connect the DIASAFE® plus filter and run the "filter change" program.
Operating condition: Cleaning / filter change
– Use the "filter change" program item in the cleaning menu and start
the disinfection program requested by the device.
Re 4 Check of hydraulics
Re 4.1 Water inlet pressure (reduced): 1.2 bar (+0.2 bar / –0.3 bar)
Operating condition: Service mode / CALIBRATION
Re 4.4 Relief pressure of balancing chamber at 800 ml/min: 2.2 bar (±0.05 bar)
Operating condition: Service mode / CALIBRATION
Conf
key
After
approx. 3 s
Conf
key
Note
The accuracy of the pressure gauge used must correspond to that of the
UMED or HMED.
The accuracy of the pressure gauge used must at least correspond to
the following values:
-1 to 2 bar ±5 mbar
2 to 8 bar ±20 mbar
Re 8.1 Make setup settings in the service program and transfer data to the
dialysis device.
Operating condition: Service mode / CALIBRATION
Conf
key
Using a syringe, set the
calib. sn. BP-RATE pressure at the P-SN connector Conf (B)P–Rate=550 ml/min Esc
according to the set key key
stroke volume (see table).
Set a rate of
≥550 ml/min*
SN. BP-Rate CHECK Conf (B)P–Rate=550 ml/min on the SN blood pump.
key
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the (+/–) keys.
Check the upper switching points according to the table below. The
upper switching point depends on the stroke volume.
The lower switching point is fixed (75 mmHg).
Stroke volume 10 15 20 25 30 35 40 45 50
(ml)
Switching 110 130 150 172 195 219 244 270 299
point (mmHg)
± 7 mmHg
or
Direct measurement according to fig. 4
Basic conditions:
– Measurement of the protective earth resistance performed.
– Perform the measurement with the hemodialysis device being at
operating temperature.
– Dialysate:
Dialysis temperature: ≥ 36.5 °C
Dialysate flow: ≥ 300 ml/min
Conductivity: ≥ 13 mS/cm
– When performing a direct measurement, the following precautions
also must be observed:
The device must be insulated when installed.
All external connections must have been removed from the device.
The line voltage during the measurement will be recorded, as well as
the maximum device leakage current, scaled to the nominal voltage of
the power supply. Maximum device leakage current: 500 µA.
Example:
Line voltage during measurement: 225 V
Device leakage current:
mains polarity 1: 180 µA
mains polarity 2: 120 µA
Maximum value of both mains polarities: 180 µA
Nominal voltage of power supply: 230 V
Scaled to nominal voltage 184 µA
(180 µA: 225 V x 230 V = 184 µA
Device leakage current < 500 µA: OK
Additional requirements:
– If the value scaled to the nominal voltage is higher than 90 % of the
admissible alarm limit (= 450 µA), the last measured value or the first
measured value must additionally be considered for the rating.
– If the device leakage current has considerably increased since the
last measurement or has continuously increased since the first
measurement (slow deterioration of the insulation), or if the sum
composed of the current value plus the difference since the last
measurement is > 500 µA, the measurement has not been passed.
Example 1:
Device leakage current: 470 µA
Last measured value: 450 µA
470 + (470 – 450) = 470 + 20 = 490 (-> successfully passed!)
Example 2:
Device leakage current: 470 µA
Last measured value: 390 µA
470 + (470 – 390) = 470 + 80 = 550 (-> not passed!)
Re 11 Functional test
Re 14 Final check
Re 14.1 Power failure alarm – continuous sound – text displayed: Power Failure
Operating condition: Preparation
Assessing the test The device has been released for further use
(attach inspection label).
It must be ensured that the intended use of the device will not present
a hazard to patients, operators and other third parties.
Within the scope of the overall assessment, the tester must make a
definite decision whether the device may be used or not. The
responsible organization must immediately be informed of any defects
detected.
Date of next inspection:
The next inspection date has to be entered in the report.
The intervals prescribed by the manufacturer must be observed.
Remarks:
Any irregularities which occurred during the assessment are
documented in this section.
Date, signature, stamp:
Assessment of the initial start-up has to be confirmed with date, tester's
signature and stamp.
Warning
– Risk of scalding
Before transporting the device, ensure that there are no corrosive or hot
liquids in the device. Programs, such as disinfection or hot disinfection
(incl. cool down rinse), must have been run through completely.
Warning
During transportation – Lift the hemodialysis device over uneven surfaces (e.g.
cobblestones).
– For vehicular transportation, cushion the hemodialysis device
appropriately and keep it upright if possible. If the hemodialysis
device has to be transported on its side, then lower it gently and
transport it with adequate cushioning on all sides.
After transportation – After moving the device, bring the hemodialysis device back into
service (see Restarting after removing 4008 from Service).
4 Specifications
The type label shown is only an example. The decisive criterion is the
data specified on the type label of the device .
13
12 1
11 2
10
3
9
6 5 4
1 IP type of protection 21
2: Protection against touch and foreign bodies
with a diameter of at least 12.5 mm
1: Protection against ingress of liquids:
drip-proof
2 CE mark
3 Identification of electric and electronic devices
4 Degree of protection against electric shock: type B
5 Relative humidity (Operating conditions)
6 Atmospheric pressure (Operating conditions)
7 Manufacturer with date of manufacture as year digit
8 Operating temperature range
9 Power requirements (voltage / power consumption)
10 Maximum total weight (empty weight + safe working load)
11 Serial number
12 Type identification
13 Equipment code (EC)
4.6 Fuses
Battery 1 x T 3.15 A; power supply unit, fuse in housing foot (rear) / SI5
Warning
When using central venous catheters, the following precautions must
be observed:
1. The dialysis device must be connected to a potential equalization.
2. If additional electro-medical devices are connected to the patient or
they are positioned within close proximity of the patient, it must be
ensured that all leakage currents of these devices (device leakage
currents, housing leakage currents, earth leakage currents and
patient leakage currents) are below the respective limit for CF
applied parts.
This means:
10 µA maximum in normal conditions, and 50 µA in "single fault
conditions".
This also applies to patient positioning devices (e.g. patient chairs).
Devices with leakage currents within these limits, but with an
application current exceeding the specified leakage currents
(e.g., on electro stimulators) may not be used. This also applies to
defibrillators, which have no applied part of the CF type.
If all requirements have been fulfilled, these devices may be
operated on the patient or within the reachable area of the patient,
provided they are, like the dialysis device, integrated into the
potential equalization.
If these conditions are not fulfilled, no other electro-medical devices
must be connected to the patient or positioned within the reachable
area of the patient.
In case of doubt, contact the local technician.
all types!
(B / BF / CF)
Housing
Leakage currents
of all types
The patient is 10 times higher than
connected "CF" patient
electrically leakage current!
to earth potential
via the blood and
dialysate.
Voltage dips, short <5 % UT (>95 % dip <5 % UT (>95 % dip After power supply interruptions, the
interruptions and in UT) for 0.5 cycle in UT) for 0.5 cycle battery of the hemodialysis device takes
voltage variations on 40 % UT (60 % dip in 40 % UT (60 % dip in over the supply without delay.
power supply input UT) for 5 cycles UT) for 5 cycles
lines IEC 61000-4-11
70 % UT (30 % dip in 70 % UT (30 % dip in
UT) for 25 cycles UT) for 25 cycles
< 5 % UT (>95 % dip < 5 % UT (>95 % dip
in UT) for 5 s in UT) for 5 s
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be
(50/60 Hz) magnetic at levels characteristic of a typical location
field IEC 61000-4-8 in a typical commercial or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Portable and mobile RF communications
equipment should be used no closer to any
part of the hemodialysis device, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Conducted RF 3 Vrms 3 Vrms d = 1.2 √P
IEC 61000-4-6 150 kHz to 80 MHz for 150 kHz to < 80 MHz
Radiated RF 3 V/m 3 V/m d = 1.2 √P
IEC 61000-4-3 80 MHz to 2.5 GHz for 80 MHz to < 800 MHz
d = 2.4 √P
for 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range
b.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
hemodialysis device is used exceeds the applicable RF compliance level above, the hemodialysis device
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the hemodialysis device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF telecommunication devices and
the hemodialysis device
The hemodialysis device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or user of the hemodialysis device can contribute to avoiding electromagnetic errors,
by adhering to the minimum distance between portable and mobile HF telecommunications devices (transmitters)
and the hemodialysis device – dependent on the exit performance of the communication device, as indicated
below.
Rated maximum output Separation distance according to frequency of transmitter m
power of transmitter 150 kHz to < 80 MHz 80 MHz to < 800 MHz 800 MHz to 2.5 GHz
W d = 1.2 √P d = 1.2 √P d = 2.4 √P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Water inflow rate 1.3 l/min; with an inlet pressure of 1.5 bar
Water inflow line The length of the water inflow line to a fixed permeate connection
(e.g. MSC 08) must not be more than 3 m.
Water quality Only water, which is suitable for hemodialysis, should be used for
operating the hemodialysis device. The quality of the water should
comply with the local regulations
(e.g. European Pharmacopoeia – Ph. Eur.).
Purified water
Water drain The height of the water drain outlet must be 0 to 100 cm over ground
level. The drain outlet must be lower than the dialyzer. An air gap of at
least 2 cm must be maintained at the end of the drain. A dedicated,
separate, unused drain outlet must be provided for each dialysis device.
Never combine drain outlets from multiple devices before the air gap is
reached. The length of the drain outlet to the air gap must not exceed
3 m. Observe any specific national provisions on the drain outlet air
gaps design and backflow prevention measures.
Concentrate supply 0 to –100 mbar; max. suction height 1 m, max. height of fall 0 m
Operating temperature 15 °C to 35 °C
range
Average energy emission Dialysis with a flow of 500 ml/min: 0.48 kWh
to drain Hot disinfection: 0.46 kWh per hot rinse disinfection
Water inlet temperature 15 °C in each case
Average energy emission Dialysis with a flow of 500 ml/min: 0.40 kWh
to environment Hot disinfection: 0.33 kWh per hot rinse disinfection
Ambient temperature 20 °C in each case
Temporary downtime Prior to the downtime, it is recommended to carry out a hot rinse
program without a cooling rinse (PGM 2) and a disinfection prior to
commencing operations.
Antifreeze When storing the hemodialysis device with antifreeze, make sure to use
antifreeze of the following composition:
– 49.875 % water
– 49.875 % glycerin
– 0.25% ClearSurf
Charging the integrated Upon receipt of the hemodialysis device, the integrated batteries must
batteries be charged as follows:
– Connect the device to the external power supply by means of the
power cable.
– Then turn the device on for approx. 10 hours.
If the device is not used, this procedure should be repeated every six
months.
Warning
Additional equipment connected to medical electrical equipment must
comply with the respective IEC or ISO standards
(e.g., (DIN) EN 60950-1:2006 or IEC 60950-1:2006 for data processing
equipment). Furthermore all configurations must comply with the
requirements for medical electrical equipment (clause 16 of the 3Ed. of
IEC 60601-1:2005). Anybody connecting additional equipment to
medical electrical equipment configures a medical system and is
therefore responsible that the system complies with the requirements
for medical electrical equipment. Attention is drawn to the fact that local
laws take priority over the above mentioned requirements. If in doubt,
consult the local service.
Warning
The operator may not touch the patient and the plug connectors or
socket contacts of the device at the same time. Otherwise, there is the
risk of patient leakage current.
Alarm out For the connection of an external alarm indicator (staff call). (Potential-
(alarm output) free alarm output. Alternating contact maximum 24 V / 24 W).
Connector: 5-pin diode plug via a shielded line; shield grounded on
either side.
Audio paused Mute alarm time: adjustable in the SETUP MENU from 1 minute to
2 minutes
(factory setting: 1 minute)
Any new alarm reactivates the silenced audible alarm.
Plastics
ABS
EPDM X
NBR
PA (PA 6.6)
PBT
PBT X
PBT/ABS GF 20
PC X
PC+ABS
PE
PE (soft)
PEEK X
PES X
Polyester
POM
PP X
PP-H X
PP(E)
PPO X
PPS X
PPSU X
PS
PSU
PTFE X
PUR
PVC
PVDF X
Silicone SIK X
Silicon tubing X
TPE
VMQ SI (Q)
Silicone X
TEEE
CM/ EPDM-L
EPDM
Metals / glass
Graphite X
Glass X
Aluminum
Sheet copper
Steel sheet
Aluminum sheet
St37K
1.4300
1.4301
1.4305
1.4401 X
1.4404/1.4435 X
Titanium X
Radox
Lexan (plexiglass)
Oxit 100
Steel 1.1121
MS 58
MS 63
Leather (bag)
Paper
Silpad 400
Ceramics X
Miscellaneous – adhesives
Loctite
Scotch-Weld DP 460
Scotch-Weld V 23
Cyanolit
Elastosil
3 M Scotch Weld
EPXDPU60
Miscellaneous – lacquers
Enamel: Pehapol
Prime coat:
P 81.914 (base)
Miscellaneous – packaging
Ethafoam (polystyrene)
Corrugated cardboard
Cellular rubber
Wooden pallet
Electrical equipment
The Service menus CALIBRATION, DIAGNOSTICS, MISCELLANEOUS, and SETUP MENU can only
be called in service mode.
Warning
Setting the service switch to ON while the treatment or the cleaning
program is in progress will turn off the heater relay.
CALIBRATION, DIAGNOSTICS, MISCELLANEOUS and SETUP MENU are controlled with the following
keys.
Function Key
CALIBRATION
DIAGNOSTICS
MISCELLANEOUS
SETUP MENU
Conf
key
SET ALARM/WARN TIME Conf see Part 1 SOUND I/O-SWITCH Conf see Part 13
key key
SETUP CLEANING PGM Conf see Part 2 SET KEY-CLICK Conf see Part 14
key key
SET CONDUCT. LIMIT Conf see Part 4 SET RINSE-VOLUME Conf see Part 16
key key
INFO SOUND ( C-PGM ) Conf see Part 5 T1-TEST AUTOSTART Conf see Part 17
key key
SET DIAL PARAMETERS Conf see Part 6 VENOUS LIMITS Conf see Part 18
key key
OCM SETTINGS Conf see Part 8 AutoOFF after AutoON Conf see Part 20
key key
AUTOM. SN-START Conf see Part 9 Init. CAMUS-baudrate Conf see Part 21
key key
optional
ACTIV. MONIT_NTC109 Conf see Part 10 SET BPM Conf see Part 22
key key
UF EINSTELLUNGEN Conf see Part 11 STORE DEFAULT VALUES Conf see Part 23
key key
SET STD. PRIME-TIME Conf see Part 12 BACK TO MAIN MENU ? Conf
key key
Conf
key
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Set UF-WARNING-TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Set MUTE-TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Note
When indicating the cleaning times, the default values and the
adjustable range are not indicated, as they depend on the particular
device options.
CLEANING Times
Conf
key
Rinsing TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Hotrinse TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Disinfection TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Hot-Disinf TIME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
CITRO-Mandat-Ri-Time
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
INTEGRATED-HR Time
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
If "NO central-delivery“
is set: (see Part 19: Setting central delivery page 5-23)
SETUP DILUTION
Conf
key
If "central Acid"
is set: (see Part 19: Setting central delivery page 5-23)
SETUP DILUTION
CDS 1+34
Conf
key
+/–
Mute DATA STORED After
key approx. 3 s
canister 1+34 Mute ACKNOWLEDGED CDS 1+44 C
key
Esc
key
+/–
+/–
CDS 1+44 ACF
Esc
key
CDS 1+34
Mute DATA STORED After
Mute key approx. 3 s
canister 1+44 C key ACKNOWLEDGED +/–
Esc
key
CDS 1+44 C
+/–
Esc
key
CDS 1+34
Mute DATA STORED After
Mute key approx. 3 s
canister 1+44 ACF key ACKNOWLEDGED +/–
Esc
key
+/– CDS 1+44 C
Esc
key
If "centr acetate-supply“
is set: (see Part 19: Setting central delivery page 5-23)
SETUP DILUTION
Conf
key
canister 1+34 Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
1+35.83 ( NaCl 20 ) Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
1+35.83 ( NaCl 26 ) Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
1+35.83 ( Belgium ) Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
canister 1+44 C Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
canister 1+44 ACF Mute ACKNOWLEDGED CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
key
Esc
+/– key
VARIABLE SETTING Conf see Part 3.1 CDS 1+34 Mute DATA STORED After
key key approx. 3 s
Esc
+/– key
Esc
key
Warning
The operator or technician is informed about his duty of care to enter the
component parameters and settings correctly or to check them and to
set the CD alarm window to the expected conductivity value.
When using the user adjustable mixing ratio, make sure to use the right
concentrate.
Using a bibag® in combination with the user adjustable mixing ratio is
not allowed.
Only enter approved user adjustable mixing ratios.
SETUP DILUTION
Conf
key
+/–
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Mute
key
ACKNOWLEDGED
After
approx. 3 s
HCO3-[mmol]: 25 Esc
key
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Na(B)[mmol]: 25 Esc
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Conf
key
AdaptedFlow ON/OFF
Conf
key
Conf
key
Optional
AdpFlow factor: 1.9 Esc
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Mute
key
DATA STORED
After
approx. 3 s
CALC.CUMUL.BLOOD-VOL
Conf
key
OCM SETTINGS
Conf
key
OCM MEASUREMENT
Conf
key
Conf
key
Delaytime: 18 s Esc
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
AUTOM. SN-START
Conf
key
UF SETTINGS
Conf
key
+/– Esc
key
UF rate: 750 ml/h
UF goal: 3000 ml
UF time left: 4 h
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
SOUND I/O-SWITCH
Conf
key
SET KEY-CLICK
Conf
key
Note
The BPR/UFR warning may be deactivated only for service work or
when troubleshooting. In dialysis mode, the BPR/UFR warning must
always be enabled, i.e. "ON".
BPR/UFR-WARNING
Conf
key
SET RINSE-VOLUME
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
T1-TEST AUTOSTART
Conf
key
VENOUS LIMITS
Conf
key
ASYM. LIMITS
Conf
key
Conf
key
SET CENTRAL-DELIVERY
Conf
key
Conf
key
Init. CAMUS-baudrate
Conf
key
SET BPM
Conf
key
Preselected Pressure
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Conf
key
Store the
default values
by pressing the
Mute key.
DATA STORED
After
approx. 3 s
5.1.4 DIAGNOSTICS
CALIBRATION
DIAGNOSTICS Conf
key
MISCELLANEOUS
SETUP MENU
In DIAGNOSTICS, all inputs and outputs of the dialysis device can be activated.
Activation refers to CPU1 (P.C.B. LP 1631), CPU2 (P.C.B. LP 632), as well as to the output board (P.C.B.
LP 634) and the input board (P.C.B. LP 633-5).
This program allows technicians to make their own settings for testing error patterns.
The corresponding levels are displayed in the "READ DIGITAL INPUTS" menu via the fields to in
the display as well as the status indicators and the loudspeaker.
The active signal state (may indicate both voltage and no voltage) is indicated by 1111 on the display
(fields to ), by the status indicators being activated and by an audible signal.
The audible signal can be deactivated by pressing the Mute key. If the audible signal is deactivated, the
Mute key flashes to alert the user.
The audible signal allows the operator to evaluate the signal state without having to have a view of the
monitor. This is advantageous if measurements must be taken behind the device (e.g., hydraulics unit).
Note
In DIAGNOSTICS, the signals are listed in the order of their electric
connection, i.e. in latch groups of 8 signals each according to the 8-bit
data bus and according to the latch numbering on the wiring diagram
(e.g., P.C.B. LP 633-5: CS_LATCH0 – CS_LATCH6).
The signals are not grouped by relationship (e.g., all bibag® signals in
successive row). An exception here are the signals for activating the
solenoid valves. If possible, these are listed in the menu by their number
to facilitate finding an individual valve because, usually, several valves
must be activated at the same time for trouble-shooting purposes.
Based on their assignment to latch groups, signals can be easily found
in the menu by means of the wiring diagram - even if the signal name
has changed. Already one known signal within one latch group is
sufficient to find the renamed signal in the menu by counting.
The voltage values specified can differ because of tolerances in the
particular devices.
The CPU1: RD DIGITAL INP menu provides the E:CPU1_KEY_TESTING item. This menu item can be
used to test the keys.
The four arrow keys, the On/Off key and the Conf key are not implemented because their function can
be tested by selecting the appropriate menu.
XXXX
XXXX
XXXX
XXXX
DIAGNOSTICS
Conf
key
READ INPUTS Conf READ ANALOG INPUTS Conf CPU1: RD ANALOG INP. Conf see Part 1
key key key
READ DIGITAL INPUTS Conf CPU1: RD DIGITAL INP Conf see Part 3
key key
Conf
key
back to menu ? Conf
key
WRITE OUTPUTS Conf WRITE ANALOG OUTPUTS Conf CPU1: WR ANALOG OUTP Conf see Part 1
key key key
WRITE DIGIT. OUTPUTS Conf CPU1: WR DIGIT. OUTP Conf see Part 3
key key
Conf
key
back to menu ? Conf
key
Navigation
Menu display Description
Explanation
Part 1
E: CPU1_24V_EM 24V_EMERGENCY
24 V, IC 28/28, ADC 117
(22.5-26 V)
E: CPU1_COND_SIGNAL1 CD display
0-10.8 V, IC 28/4, ADC 0-231
Back to menu ?
Explanation
Part 2
E: CPU2_COND_SIGNAL CD display
0-10.8 V, IC 12/8, ADC 0-251
Back to menu ?
Explanation
All fields (to ) in the display show 0000, status indicators red,
yellow, green off:
Low level applied to latch on P.C.B. LP 633-5
All fields (to ) in the display show 1111, status indicators red,
yellow, green on:
High level applied to latch on P.C.B. LP 633-5
If high level is applied, an audible alarm is active at the same time. This
can be suppressed by pressing the Mute key. In this case, the Mute
LED is lit.
Part 3
E: CPU1_COND_V84 V 84 CD detection
IC 19/2
Test by pulling off/short-circuiting the sensors
E: CPU1_V43 Valve 43
IC 13/2
Test by opening/closing the valve
E: CPU1_V26 Valve 26
IC 13/3
Test by opening/closing the valve
E: CPU1_V24 Valve 24
IC 13/5
Test by opening/closing the valve
E: CPU1_LC11 IC 13/9
E: CPU1_LA32 IC 15/2
E: CPU1_HEP_ALARM IC 16/8
Initiate heparin pump alarm
(e.g., by blocking the slide carriage)
E: CPU1_HDF_ON HDF on
IC 21/3
Press the HDF On/Off key
E: CPU1_CSS_REED IC 21/6
E: CPU1_HEAT_CLK IC 21/7
E: CPU1_BYP_REQ IC 21/8
E: CPU1_CLP_REQ IC 21/9
Back to menu ?
Explanation
All fields (to ) in the display show 0000, status indicators red,
yellow, green off: Low level applied to latch on P.C.B. LP 632
All fields (to ) in the display show 1111, status indicators red,
yellow, green on: High level applied to latch on P.C.B. LP 632
If high level is applied, an audible alarm is active at the same time. This
can be suppressed by pressing the Mute key. In this case, the Mute
LED is lit.
Part 4
E: CPU2_V147 IC 7/3
E: CPU2_V145 IC 7/6
E: CPU2_24V_SW 24 V switch
IC 8/7
Back to menu ?
Explanation
Part 5
Back to menu ?
Explanation
Part 6
Back to menu ?
Explanation
Part 7
A: CPU1_V24 Valve 24
IC 10/19
A: CPU1_V26 Valve 26
IC 10/18
A: CPU1_V30 Valve 30
IC 7/16
A: CPU1_V31 Valve 31
IC 12/19
A: CPU1_V32 Valve 32
IC 12/18
A: CPU1_V33 Valve 33
IC 12/17
A: CPU1_V34 Valve 34
IC 12/16
A: CPU1_V35 Valve 35
IC 12/15
A: CPU1_V36 Valve 36
IC 12/14
A: CPU1_V37 Valve 37
IC 12/13
A: CPU1_V38 Valve 38
IC 12/12
A: CPU1_V41 Valve 41
IC 7/13
(Valve automatically closes after a short time to prevent water from
overflowing)
A: CPU1_V43 Valve 43
IC 7/15
A: CPU1_V84 Valve 84
IC 7/18
Note: Activation of valve V84 must be followed by a rinse program.
A: CPU1_V86 Valve 86
IC 7/17
A: CPU1_V87 Valve 87
IC 10/15
A: CPU1_V91 Valve 91
IC 10/14
Remark: If this menu item is exited (return to "CPU1: WR DIGIT. OUTP."), the
valve is closed.
A: CPU1_V99 Valve 99
IC 7/19
Remark: If this menu item is exited (return to "CPU1: WR DIGIT. OUTP."), the
valve is closed.
A: CPU1_SET_FLOW_ON Flow on
Data word to Gal 23:
0000 0010 (active, V 32 open)
0000 0011 (inactive, V 31, 32 opened)
A: CPU1_EMPTIING_PRG Emptying-program
Data word to Gal 23:
0001 0010 (V 32, 35 open)
Back to menu ?
Explanation
Part 8
A: CPU2_V24_EN V 24 enable
IC 24/16
Preparation: Activate V 24 (CPU1: WR DIGIT. OUTP).
Test: Switch valve on/off with V 24_EN.
A: CPU2_V26 Valve 26
X632/A6
A: CPU2_V43 Valve 43
X632/C5
A: CPU2_NC9 X632/C6
IC 29/13
Back to menu ?
Explanation
Display (field ): Feedback/input (in addition, the 3 status indicator LEDs are lit with 1111)
Display (field ): Activation/output (adjustable by pressing the + / – keys)
Part 9
Back to menu ?
5.1.4.12 HPU
Part 10
HPU
READ ANALOG INPUTS Conf
key
Conf
key
A:WTR_AIR_P
Compressor
back to menu ? Conf 185
key
A:WTR_V_TEST
Test valve
V183
A:WTR_V39
Negative pressure valve
V39
A:WTR_RETENT_V
Retentate valve
V189
A:WTR_V_EVAC_INDI
not
implemented
A:WTR_V_EVAC_SOD
Evacuation valve
V188
Part 11
BPM
Conf
key
Pressure test Conf Start test? Conf Curr. press: xxx mmHg
key key
◄► Esc
key
Leakage test Conf Start test? Preselected Pressure
key
Esc
key
5.1.5 MISCELLANEOUS
CALIBRATION
DIAGNOSTICS
MISCELLANEOUS Conf
key
SETUP MENU
Conf Conf
key key
Part 1
SYSTEM CLOCK
Conf
key
Time Date
XX:XX:XX XX:XX:XXXX Esc
key
Conf
key
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
5.1.5.3 SW-VERSION-NUMBER
Part 2
SW-VERSION-NUMBER
Conf
key
+/–
+/–
+/–
+/–
Optional
BPM-Ver-No.: XX.XX Esc
key
+/–
+/–
+/–
Part 3
BPM
Conf
key
5.1.6 CALIBRATION
For a description of the CALIBRATION service menu, please refer to Chapter Calibration / adjustment
(see chapter 9.1 page 9-1).
Note
DIP switch 6 (array 1) is provided for service purposes / trouble-
shooting only and must be set to position OFF for dialysis mode.
S1 S2 S3 LP1631
O
N
1
2
3
4
5
6
7
8
1 2 Maximum UF rate
ON ON 1000 ml/h
OFF ON 2000 ml/h
ON OFF 3000 ml/h
OFF OFF 4000 ml/h
3
ON
OFF not assigned
4
ON
OFF not assigned
5
ON
OFF not assigned
6 CRC/RAM Test
ON Skip
OFF Go
7 Heater rod
ON 1300 W (at 100 to 127 V)
OFF 1600 W (at 220 to 240 V)
S1 S2 S3 LP1631
O
N
1
2
3
4
5
6
7
8
DIP switch / Function
Position
1 CAL mode
ON Mode 0
OFF Mode 1
4 Remote control
ON Device with remote control
OFF Device without remote control
6 COMMCO
ON Special protocol
OFF Default protocol
S1 S2 S3 LP1631
O
N
1
2
3
4
5
6
7
8
DIP switch / Function
Position
1 2 3 4 5 6 Language setting
8
ON
OFF not assigned
Warning
DIP switch 3 (array 1) allows skipping test runs that are requested by
the device. If the switch is set such that a test run can be skipped, then
it must be noted that the operator has the possibility of evading the
automatic test of the safety systems. The person requesting such a
setting is responsible for this procedure.
S1 S2 LP632
O
N
1
2
3
4
5
6
7
8
1
ON
OFF not assigned
2 T1 Test
ON Serial sequence
OFF Parallel sequence
3 T1 Test
ON Skip
OFF Mandatory
4 Test Service
ON ON (single test steps can be selected; dialysis mode not possible)
OFF OFF (automatic T1 test)
5 Cyclic PHT
ON Every 2 minutes and events display (Service)
OFF Every 12.5 minutes, alarm output only with cyclic PHT alarm
6 Cyclic PHT
ON ON
OFF OFF
7 Air detector
ON Not permitted
OFF with P.C.B. LP 450-2
8
ON
OFF not assigned
S1 S2 LP632
O
N
1
2
3
4
5
6
7
8
DIP switch / Function
Position
1 DIASAFE® plus
ON ON
OFF OFF
2
ON
OFF not assigned
3
ON
OFF not assigned
4 Hydraulics unit
ON With HPU
OFF Not permitted
5 V39 test
ON OFF
OFF ON
7
ON
OFF not assigned
8
ON
OFF not assigned
S1 LP634
O
N
1
2
3
4
DIP switch / Function
Position
1 2 3 4
OFF OFF OFF OFF I2C-SW mode (software-selectable alarm system)
ON OFF OFF OFF new alarm system
OFF ON OFF OFF old alarm system
Preparation Switch the device off, dismount the monitor rear, pull out the monitor.
Perform software update Concerning CPU 2, the update is still performed by replacing the
EPROM.
With software version 11.00 or higher, the update of CPU 1 and
MDC-II (monitor board) is performed using the SD-card:
1. P.C.B. LP 1631 (CPU 1): field 3, set DIP switch 7 to ON.
2. P.C.B. LP 1631: insert SD card
3. Turn the device on
4. The update is performed automatically.
P.C.B. LP 1631: While CPU 1 is updated, LED D1 is flashing. As
soon as the update of CPU 1 is complete, D1 will be illuminated.
Then D3 flashes until the update of the MDC-II board is complete.
As soon as the update is complete, D3 will be illuminated.
The flashing frequency of the LEDs D1 and D3 is approx. 2 Hz in
case of an error approx. 6 Hz). In the event of an error, the LEDs D4
to D8 serve for evaluating the error (see Error code table
page 5-70).
5. Turn the device off
6. Re-set DIP switch 7 to OFF
7. The SD-card may be removed or remain inserted
Quick Guide – The Quick Guide is located on the CD for the PC Service Software
PC service software 4008 4008. The Service Software is installed on a PC (e.g., the technician's
laptop). In addition, you also have the option – e.g., for a new installation
– of accessing the Quick Guide directly from the CD without installing
the Service Software.
Accessing the Quick Guide Three options for accessing the Quick Guide are described below:
1. Precondition: PC Service Software 4008 is installed.
With the Service Software 4008 application open, use "Help" from
the menu bar to access the Quick Guide, as follows:
→ "Help“ → "Help Topics“ (1)
or:
→ „Help“ → „Context Help“ (2) (see Fig. below)
1
2
6 TSC / maintenance
Checks This chapter includes the Technical Safety Checks (TSC) and the
maintenance procedures (MA) to be performed.
Technical Safety Checks (TSC) must be carried out every 2 years
(24 months).
Performance of the Technical Safety Checks must be recorded in the
Medical Device Register.
The maintenance procedures (MA) are a recommendation of the
manufacturer. The maintenance procedures have to be carried out after
2 years (24 months) at the latest and serve to prevent malfunctions.
The explanations on the TSC / MA test report
are applicable to the TSC / MA test report and for the TSC test report.
Tester's qualification The initial start-up must be performed by the service of the
manufacturer or a person authorized by them.
The checks must only be carried out by persons, who, based on their
training, knowledge and on-hand experience, are qualified to carry out
such checks in a proper manner. Furthermore, the persons, who carry
out the checks, may not be subjected to outside instructions with
regards to these checks.
Warning
Commissioning steps, extensions, adjustments, calibrations,
maintenance measures, modifications or repairs may only be carried
out by the manufacturer or persons authorized by him.
Test equipment and The activities described in the document require the availability of the
accessories necessary technical measuring equipment and accessories.
Warning
When the device is returned to use, check that the pressure of the water
supply meets the prescribed minimum pressure.
ESD precautions When repairing the device and when replacing spare parts, observe the
applicable ESD precautions.
The following articles are required for carrying out the TSC / MA.
Unisilkon grease, L 250 L, (10-g tube) Use only in places described in the
explanations.
Battery CR 1225, Lithium (3 V/42 mAh) P.C.B. LP 1631 Replace every 4 years.
Solenoid valve adapter fitting, one port V84 Replace every 2 years, only if ,
Puristeril 340, Puristeril plus are used.
Remarks:
Remarks:
Warning
On completion of these procedures it is imperative to run a disinfection.
Re 1 Visual inspections
Operating condition: Off
Re 1.1 The battery fuse accessible (see illustration below) (1) from the outside
corresponds to the indicated value.
Alternative to checking the fuse: If the seal is undamaged, the battery
fuse does not have to be checked.
Re 1.2 Labels and identifications are present and legible (with illustrations
below).
1 2 x label "Type label complete device“
Check device specification:
Serial number
Equipment code
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
25
26 Bed-Side-Link (option):
2 x label: „LAN“ and „CardBox“
26
26
27 Country-specific labels:
27.1 Additional label for the People's Republic of China:
1 x label "CCC" (to the right of the "Type label complete device“ on
the outside)
1 x label "China RoHS" (to the right of the "Type label complete
device“ on the outside)
27.1
27.1
27.2
Re 1.3 Mechanical condition permits further safe use. There are no signs of
damage or contamination affecting proper function of the device.
To be checked:
– Tubings and cuff of BPM (option)
Re 1.4 Rotor(s) cleaned. Rotor(s) on the line roller pump(s) show no signs of
damage and are fully functional. Color coding of rotor(s) checked.
4 2
3
3
Re 2 Preventive measures
Operating condition: Off
Re 2.1 Replace the sealing plunger in the concentrate and in the bicarbonate
suction tube and grease it with silicone compound. If necessary, replace
the rivets of the suction tubes.
Re 2.2 Check the detent rubber seal of the rinse chambers for proper
functioning.
Re 2.8 Check the sampling valve (V116) in the dialysate circuit tubing for
proper function and replace the complete valve, if necessary.
– Battery (2)
Re 3 General checks
S1 S2 S3 LP1631
O
N
1
2
3
4
5
6
7
8
– Without Central Delivery System:
P.C.B. LP 1631: Array 2, switch 7 is on ON
S1 S2 S3 LP1631
!
O
N
1
2
3
4
5
6
7
8
– With Central Delivery System:
P.C.B. LP 1631: Array 2, switch 8 is on ON
(switch 7 is no longer relevant!)
S1 S2 S3 LP1631
!
O
N
1
2
3
4
5
6
7
8
S1 S2 LP632
!
O
N
1
2
3
4
5
6
7
8
S1 S2 LP632
!
O
N
1
2
3
4
5
6
7
8
Re 4 DIASAFE® plus
Re 4.2 Replace hydrophobic filter (F111) and hydrophobic filter / test valve
(F184).
Operating condition: Off
Re 5 bibag®
Re 6 OCM
Re 7 Check of hydraulics
All pressures must be checked with undampened pressure gauges!
Re 7.2 Check the loading pressure of the balancing chamber and, if required,
correct it.
Operating condition: Service mode / CALIBRATION
Connect a pressure gauge to the pressure side of the degassing pump
(measuring point B in HU). The pressure must be 1.45 bar ±0.05 bar.
Re 7.4 Check balancing chamber relief pressure at a flow of 800 ml/min (relief
valve A78).
Conf
key
ACKNOWLEDGED
After
approx. 3 s
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
Re 8.2 Measure the volume of the concentrate pump in liters or compare with
an appropriate reference device. If necessary, make settings according
to calibration instructions.
Re 8.3 Measure the volume of the bicarbonate pump in liters or compare with
an appropriate reference device.
If necessary, make settings according to calibration instructions.
Operating condition: Service mode / CALIBRATION
Re 9 Extracorporeal components
CALIB. (B)-PUMP-RATE
Conf
key
Conf
key
Using a syringe, set the
calib. sn. BP-RATE pressure at the P-SN connector Conf (B)P–Rate=550 ml/min Esc
according to the set key key
stroke volume (see table).
Set a rate of
≥550 ml/min*
SN. BP-Rate CHECK Conf (B)P–Rate=550 ml/min on the SN blood pump.
key
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the (+/–) keys.
Check the upper switching points according to the table below. The
upper switching point depends on the stroke volume.
The lower switching point is fixed (75 mmHg).
Stroke volume 10 15 20 25 30 35 40 45 50
(ml)
Switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
Re 10 Dialysis mode
DIALYSATE Pressure
Conf
key
Use the potentiometer P3/LP 633 to set Switch off the flow.
value indicated by the external reference
instrument on the alpha display. Leave the UF pump running,
until approx. 500 mmHg are
indicated by the external
reference instrument.
Repeat
the procedure Enter the value indicated by the
until 0 and –500 external reference instrument
mmHg correspond Not by pressing the +/– keys.
to the external OK
comparison Mute
instrument. key
Mute
key
DATA STORED
After
approx. 3 s
– Manually open the venous occlusion clamp, hold it and insert the
tubing.
– Release the venous occlusion clamp. The venous occlusion clamp
closes.
– Use the syringe to generate a pressure of approx. 2 bar (see
Checking the venous occlusion clamp page 9-50).
– The pressure may not drop by more than 0.1 bar within 3 minutes.
Re 11 BPM (option)
Re 11.1 Check that the tube connector is correctly attached to the device.
Operating condition: Off
Re 11.2 Check the attachment of the internal blood pressure module, of the
printed circuit boards and all cable connections.
Operating condition: Off
or
Direct measurement according to fig. 4:
Basic conditions:
– Measurement of the protective earth resistance performed.
– Perform the measurement in dialysis or preparation mode with the
device at operating temperature.
– Dialysate:
Dialysis temperature: ≥ 36.5 °C
Dialysate flow: ≥ 300 ml/min
Conductivity: ≥ 13 mS/cm
– When performing a direct measurement, the following precautions
also must be observed:
The device must be insulated when installed.
All external connections must have been removed from the device.
The line voltage during the measurement will be recorded, as well as
the maximum device leakage current of both mains polarities, scaled to
the nominal voltage of the power supply. Maximum device leakage
current: 500 µA
Example:
Line voltage during measurement: 225 V
Device leakage current:
mains polarity 1: 180 µA
mains polarity 2: 120 µA
Maximum value of both mains polarities: 180 µA
Nominal voltage of power supply: 230 V
Scaled to nominal voltage 184 µA
(180 µA: 225 V x 230 V = 184 µA
Device leakage current < 500 µA: OK
Additional requirements:
1) If the value scaled to the nominal voltage is higher than 90 % of the
admissible alarm limit (= 450 µA), the last measured value or the first
measured value must additionally be considered for the rating.
2) If the device leakage current has considerably increased since the
last measurement or has continuously increased since the first
measurement (slow deterioration of the insulation), or if the sum
composed of the current value plus the difference since the last
measurement is > 500 µA, the measurement has not been passed.
Example 1):
Device leakage current: 470 µA
Last measured value: 450 µA
470 + (470 – 450) = 470 + 20 = 490 (-> successfully passed!)
Example 2):
Device leakage current: 470 µA
Last measured value: 390 µA
470 + (470 – 390) = 470 + 80 = 550 (-> not passed!)
Re 13 Functional test
Assessing the test The device has been released for further use
(Attach inspection label).
It must be ensured that the intended use of the device will not present
a hazard to patients, operators and other third parties.
Within the scope of the overall assessment, the tester must make a
definite decision whether the device may be used or not. The
responsible organization must immediately be informed of any defects
detected.
Date of next inspection:
The next inspection date has to be entered in the report.
The intervals prescribed by the manufacturer must be observed.
Remarks:
Any irregularities which occurred during the assessment are
documented in this section.
Date, signature, stamp
Assessment of the check must be confirmed with the date, the signature
of the person performing the check and a stamp.
TSC report numbering Numbers, which are not listed here, are not included in the TSC. They
are part of the Maintenance procedures.
Remarks:
Remarks:
Warning
On completion of these procedures it is imperative to run a disinfection.
7 Error messages
7.1 T1 Test
Dialines not conn? The dialysate lines are not in the shunt interlock.
Connect Conc. Line The concentrate connector is in the rinse chamber, or concentrate is not
Wrong conc. supply connected at all. The error message depends on the Central Delivery System
preselected in the SETUP MENU.
Flow alarm Dialysate inlet or dialysate outlet tube kinked, malfunctions in the hydraulics.
F04 Bypass The extended bypass cannot be correctly switched by CPU2 (V24 = off,
V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) X632/A4 not 24 V
– Acknowledgement (V26, X637/C2) X632/A6 not 0 V
– Acknowledgement (V24B, X637/C23) X632/A5 not 24 V
F05 Bypass The extended bypass cannot be correctly switched off by CPU2 (V24 = on,
V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) X632/A4 not 0 V
– Acknowledgement (V26, X637/C2) X632/A6 not 24 V
– Acknowledgement (V24B, X637/C23) X632/A5 not 0 V
F06 Bypass CPU1 fails to set the temperature control to hot rinse.
– Control line (HOTRINSE, X634R/C24) X639/A20 not 12 V
– Acknowledgement (HOTRINSE, X634R/C24) X632/A26 not 12 V
F07 Bypass The extended bypass cannot be correctly switched by CPU1 (V24 = off,
V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) X632/A4 not 24 V
– Acknowledgement (V26, X637/C2) X632/A6 not 0 V
– Acknowledgement (V24B, X637/C23) X632/A5 not 24 V
F09 Bypass The extended bypass cannot be correctly switched off by CPU1 (V24 = on,
V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) X632/A4 not 0 V
– Acknowledgement (V26, X637/C2) X632/A6 not 24 V
– Acknowledgement (V24B, X637/C23) X632/A5 not 0 V
F01 opt. Detector CPU1 interprets the optical detector in a different way than does CPU2.
– Acknowledgement (OD_OUT, X633L/C7) X632/A30 and the digital input
of P.C.B. LP 633-5 measure different levels
F04 opt. Detector CPU2 recognizes that the optical detector senses opaque fluid (required
because of the test in the cleaning program).
– Acknowledgement X632/A30 not 12 V.
– AD28 defective
F09 Blood System Acknowledgement that CPU2 recognizes that the arterial blood pump is inactive
(BP not running).
– Acknowledgement (BPSB_ART, X348a/6) X632/A11 not 12 V.
– Control line (BPSST_ART, X634L/B14) X348a/1 not 12 V or (BPST_ART,
X634L/A14) X348a/3 not 12 V
F10 Blood System Acknowledgement that CPU1 recognizes that the arterial blood pump is inactive
(BP not running).
– Acknowledgement (BPSB_ART, X348a/6) X633/A11 not 12 V
– Control line (BPSST_ART, X634L/B14) X348a/1 not 12 V or (BPST_ART,
X634L/A14) X348a/3 not 12 V
– Level is raised during the T1 test
F11 Blood System The arterial blood pump cannot be stopped by CPU1.
CPU2 recognizes that the arterial blood pump remains active.
– Control line (BPSST_ART, X634L/B14) X348a/1 not 0 V as well as
(BPST_ART, X634L/A14) X348a/3 not 0 V
– Acknowledgement (BPSB_ART, X348a/6) X632/A11 not 0 V.
– The level is raised during the T1 test, or the up/down key on the air detector
is blocked and the level is constantly raised.
F12 Blood System The arterial blood pump cannot be stopped by CPU1.
CPU1 recognizes that the arterial blood pump remains active.
– Control line (BPSST_A, X634L/B14) X348a/1 not 0 V as well as
(BPST_ART, X634L/A14) X348a/3 not 0 V
– Acknowledgement (BPSB_ART, X348a/6) X633/A11 not 0 V
F19 Blood System Check of BPUS signal (CPU1 P.C.B. LP 1631 via LP 633-5)
At the beginning of the test step a maximum of 40 s may pass until rotation has
stopped. If the blood pump is being activated, the rotation stop alarm must have
been cleared.
– Acknowledgement line (BPUS, X348A/8) X633L/ A12 not 0 V
– Acknowledgement line (BPUS, X348A/8) X633L/ A12 not 12 V
F01 Venous CPU1 (input board) shows a venous zero point deviation of more than
±12 mmHg (60 s).
– Control (VENT_VALVE, X634R/C18) X351/1 of the vent valve in the LD is
defective
– Acknowledgement (P_VEN, X351/4) X633L/B5 that the voltage value is
outside the zero point tolerance
– P-venous has not been calibrated
F02 Venous CPU2 shows a venous zero point deviation of more than ±12 mmHg (60 s).
– Control (VENT_VALVE, X634R/C18) X351/1 of the vent valve in the LD is
defective
– Acknowledgement (P_VEN, X351/4) X632L/C17 that the voltage value is
outside the zero point tolerance
– P-venous has not been calibrated
F03 Venous With detuning in positive direction, the achieved change in the venous display is
less than 100 mmHg (7 s).
– The test detuning is defective (PV_DET, X632/C18) X351/2
– Acknowledgement (P_VEN, X351/4) X633L/B5, the change in voltage is
too low
– P-venous has not been calibrated
F04 Venous The deviation in the measured value between CPU1 and CPU2 is higher than
±12 mmHg (if Pven > 100 mmHg).
– Acknowledgement (P_VEN, X351/4) X633L/B5 and X632/C17 measure
different voltage values
– P-venous has not been calibrated
F01 Air Detector CPU1 interprets the air detector signal in a different way than does CPU2.
– Acknowledgements (LDA1, X351/14) X632/C13 and X633L/C10
recognize different signal levels
F02 Air Detector The air detector alarm is not recognized by CPU2.
– Acknowledgement (LDA1, X351/14) X632/C13 not 0 V
– Transmission weakening (LDSA, X632/C16) X351/10 not 12 V
F03 Air Detector Air detector clamps acknowledgement (CPU2) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) X632/C14 not 24 V
– Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F04 Air Detector Air detector clamps acknowledgement (CPU1) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) X633L/C13 not 24 V
– Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F05 Air Detector The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgement (BL_AL, X634L/C15) X632/C21 not 12 V
F06 Air Detector Closing of the air detector clamp via the CPU2 control line was not possible.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 0 V
– Acknowledgement (LDA2, X351/6) X632/C14 not 0 V
F07 Air Detector Opening of the air detector clamp via the CPU2 control line was not possible.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
– Acknowledgement (LDA2, X351/6) X632/C14 not 24 V
F08 Air Detector Closing of the air detector clamp via the CPU1 control line was not possible, or
CPU2 acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634/C14) X351/8 not 0 V.
– Acknowledgement (LDA2, X351/6) X632/C14 not 0 V
F09 Air Detector Closing of the air detector clamp via the CPU1 control line was not possible, or
CPU1 acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634/C14) X351/8 not 0 V.
– Acknowledgement (LDA2, X351/6) X633L/C13 not 0 V
F11 Air Detector Air detector clamps acknowledgement (CPU2) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) X632/C14 not 24 V
– Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F12 Air Detector Air detector clamps acknowledgement (CPU1) activated (clamp closed).
– Acknowledgement (LDA2, X351/6) X633L/C13 not 24 V
– Clamp control (CLP_CTL, X634/C14) X351/8 not 12 V.
– Clamp control (CLP_CTL, X632/C10) X351/8 not 12 V
F13 Air Detector The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgement (BL_AL, X634L/C15) X632/C21 not 12 V
F14 Air Detector Raise level key on the air detector is constantly active.
– Acknowledgement (LEVEL_UP, X351/3) X632/C11 not 0 V
F15 Air Detector Acknowledgement of the supply voltage for the ultrasonic output stage not
between 6.5 and 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 X633L/25A jumper to X633L/B7) not 12 V
– Relay on AD28 failed to drop.
F16 Air Detector Acknowledgement of the supply voltage for the ultrasonic output stage after
3 seconds not > 14.5 V.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 X633L/ 25A jumper to X633L/B7) not voltage
of 16 V / 24 V
– Relay on AD28 is not controlled
– No 10 Hz signal at ALARM_REST (X351/12)
F17 Air Detector Acknowledgement of the supply voltage for the ultrasonic output stage not
between 6.5–13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 X633L/25A jumper to X633L/ B7) not 12 V
– Relay on AD28 failed to drop.
F01 Display CPU1 failed to start the display test within 5 sec.
– The “test started” information transmitted via the serial interface is missing
F02 Display CPU1 failed to complete the display test within 120 sec.
– The “Test passed” information transmitted via the serial interface is missing
F01 Arterial With detuning in negative direction, the change achieved on the arterial display
is less than 100 mmHg (2 sec).
– Acknowledgement (P_ART, X348A/7) X633L/B12, insufficient voltage
change
– Test detuning defective (PA_DET, X632/A17) X348A/9
F02 Arterial With detuning in positive direction, the achieved change in the arterial display is
less than 100 mmHg (2 s).
– Acknowledgement (P_ART, X348A/7) X633L/B12, insufficient voltage
change
– Test detuning defective (PA_DET, X632/A17) X348A/9
F01 Battery CPU1 failed to complete the battery test within 5 sec.
– The “Test passed” information transmitted via the serial interface is missing
F02 Battery The battery charge is insufficient for emitting an audible alarm over 1 min
(maybe no battery connected).
– The battery voltage (U_ACCU, ...) X633L/B21 dropped below 17.6 V.
– Acknowledgement (U_ACCU, ...) X633L/B21 of the battery voltage
defective.
F01 Blood Leak Blood leak channel and dimness not in alarm-free condition during the T1 test.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgement (BLL, X637/A18) X633L/B10 voltage value within the
alarm tolerances (<3 V / ≥8 V)
– Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value within
the alarm tolerances (<1.5 V / ≥8 V)
– DAC_BLL or DAC_DIM not within the tolerances (check calibration)
F02 Blood Leak The blood leak alarm / dimness alarm is not recognized during test detuning.
– Acknowledgement (BLL, X637/A18) X633L/B10 voltage value not within
the alarm tolerances (≥3 V)
– Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value not
within the alarm tolerances (≥1.5 V)
– Test detuning (BLL_DET, X632/A25) X633L/B27 not 5 V
– Calibration of DAC_BLL or DAC_DIM is too high
– Detuning (DAC_DIM, X634R/A11) X633L/C3 impossible
– Dimness calibration is set to potentiometer calibration
(BR6 from pos. 1/2 to 2/3)
F03 Blood Leak After test detuning, the blood leak channel and dimness fail to enter the alarm-
free state.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgement (BLL, X637/A18) X633L/B10 voltage value within the
alarm tolerances (<3 V / ≥ 8 V)
– Test detuning (BLL_DET, X632/A25) X633L/B27 not 0 V
– Acknowledgement (BLL_DIM, X637A/21) X633L/B11 voltage value within
the alarm tolerances (<1.5 V / ≥8 V)
– DAC_BLL or DAC_DIM not within the tolerances (check calibration)
F02 Temperature The actual temperature is less than 35.0 °C (test running time > 15 minutes).
– Calibrate the temperature
– The heater rod has failed
– Acknowledgement (T_DIAL1, X633L/B16) X632/A24, voltage got stuck.
F05 Temperature Detuning in positive direction not higher than 3 °C (10 sec).
– Acknowledgement (T_DIAL1, X633L/B16) X632/A24 change in voltage
insufficient.
– Detuning (T_DETADJ, X632/A23) X633R/C21 insufficient
F07 Temperature The test release is missing (max. test running time is 10 minutes).
– Run-time problem (software).
F08 Temperature CPU1 failed to transmit a bibag® status message within 3 sec.
– Run-time problem (software).
F10 Temperature bibag® temperature display outside of measuring range (15 to 45 °C)
– Acknowledgement (NTC_BIB, X633R/C15) ADW on P.C.B. LP 633-5
F01 neg. Pressure During the start phase a negative pressure of more than 450 mmHg has
developed (max. test running time 120 sec).
– The hydraulic system is contaminated,
– the air separation pump started running
F02 neg. Pressure Setting the dialysate pressure to the test pressure (–300 mmHg to –450 mmHg)
was not possible (max. test running time 120 sec).
Upon repetition of measurement, the range was extended from –260 mmHg to
–490 mmHg.
– Leakage in the hydraulic system
– The UF pump is defective
F03 neg. Pressure The working point (116 digits) of the differential amplifier cannot be set correctly
(max. test running time 120 sec).
– Pressure variations are too large
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29 is defective
– CI signal is missing (P.C.B. LP 632 X632/B22).
F04 neg. Pressure Completion of pressure measurement was not possible (max. test running time
120 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29 is defective
F05 neg. Pressure The air separation pump started running during the measurement phase.
– Acknowledgement (ACKN_ASP, X634L/B10) X632/A19 not 0 V.
– ASP has been interrupted electrically
F06 neg. Pressure Negative pressure holding test failed. The dialysate pressure drop exceeds
±40 mmHg (related to ten balancing chamber switching cycles).
– Leakage in the hydraulic system
F07 neg. Pressure Current increasing pulses were not recognized (min. 2 x).
– No 5 V balancing chamber pulses (CI. X634R/A23) X632/B22
F01 pos. Pressure The mandatory filling program of CPU1 has not been completed (10 sec).
– The solenoid valve V43 is not closed
F25 pos. Pressure No pressure increase above 150 mmHg (change in pressure) after valve
switching.
– Control signals of V24 and V24B mistaken for each other
– Leakage in the external system (shunt interlock, dialysate lines, etc.)
F26 pos. Pressure No pressure compensation after opening of V43 (–125 mmHg to 55 mmHg).
– V24 got stuck (mechanically open)
– V43 not open
– V26 leaky
F27 pos. Pressure No pressure compensation after opening of V43 (–125 mmHg to 55 mmHg).
– V24 got stuck (open mechanically)
– V43 not open
– V189 (retentate valve) leaking
F02 pos. Pressure The loading pressure cannot be measured via the solenoid valve V26 in the
hydraulic system
(P-Dial. < 600 mmHg, 15 sec).
– Solenoid valve V26 mechanically not open
– Solenoid valve V43 mechanically not closed
The balancing chamber is switched to passage during this test sequence. V24,
V24B and V43 are closed; V26 is open.
F03 pos. Pressure The hydraulic system cannot be deaerated via the solenoid valve V43; the zero
point of –125 to 55 mmHg has not been reached (15 sec).
– Solenoid valve V26 mechanically not closed
– Solenoid valve V43 mechanically not open
– Zero point outside the –125 to 55 mmHg range.
The balancing chamber is switched to passage during this test sequence. V24,
V24B and V26 are closed; V43 is open.
F04 pos. Pressure The first working point (220 digits) of the differential amplifier cannot be set.
– Pressure variations are too large
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29 is defective
F05 pos. Pressure Test detuning results in a change in the measuring range of more than
95 mmHg (60 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29, change in voltage too
large
– Detuning defective (P_DETADJ, X632/C20) X633R/C22
– The balancing chamber valve V36 or V38 (drain valve) is leaky
F06 pos. Pressure Test detuning results in a change in the measuring range of less than 85 mmHg
(60 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29, change in voltage
insufficient
– Detuning defective (P_DETADJ, X632/C20) X633R/C22
– V26 leaky
F07 pos. Pressure After detuning in the test there is a difference (P. diff > ±9 mmHg). between the
display and the differential amplifier.
– The voltage divider R23/R9 or the operational amplifier IC2 is defective
– The operational amplifier IC1/IC3 is defective
– The balancing chamber valve V36 or V38 (drain valve) is leaky
F08 pos. Pressure Test detuning results in a change in the measuring range of more than
400 mmHg (20 sec).
– The operational amplifier (IC2) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29, change in voltage too
large
– Detuning defective (P_DETADJ, X632/C20) X633R/C22
F09 pos. Pressure Test detuning results in a change in the measuring range of less than
350 mmHg (20 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29, change in voltage
insufficient
– Detuning defective (DIAL_DET_ADJ, X632/C20) X633R/C22
F10 pos. Pressure The second working point (116 digits) of the differential amplifier cannot be set
correctly.
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
F11 pos. Pressure Change in the dialysate pressure after closing of the solenoid valve V43 (zero
point change from –20 mmHg to +80 mmHg within 15 sec).
– The solenoid valve V24B is not closed
– The balancing chamber valve V36 or V38 (drain valve) is leaky
The balancing chamber is switched to passage during this test sequence. V43,
V24B and V26 are closed; V24 is open.
F12 pos. Pressure The loading pressure cannot be measured via the solenoid valves V24 and
V24B in the hydraulic system (P-Dial. < 600 mmHg, 15 sec).
– Solenoid valve V24 or V24B mechanically not open
The balancing chamber is switched to passage during this test sequence. V43
and V26 are closed; V24 and V24B are open.
F13 pos. Pressure The hydraulic system cannot be deaerated via the solenoid valve V43; (P-Dial.
not equal to –125 to 55 mmHg, 20 sec).
– The solenoid valve V24 is not closed
– V43 neither opens electrically nor mechanically
The balancing chamber is switched to passage during this test sequence. V24
and V26 are closed; V24B and V43 are open.
F14 pos. Pressure Zero point change after closing of solenoid valve V43 (20 sec).
Standard: P-Dial. not equal to –125 to 55 mmHg.
The balancing chamber is switched to passage during this test sequence. V24,
V26 and V43 are closed; V24B is open.
F15 pos. Pressure The loading pressure is below 780 mmHg ± 30mmHg (10 sec).
– The loading pressure is too low
F16 pos. Pressure During the start phase, the pressure dropped below 620 mmHg
(measuring tolerance: ±30 mmHg, max. test running time 120 sec).
– Major leakage in the hydraulic system
– The UF pump spring is defective
– The loading pressure is too low
– The air separation pump fails to occlude
– Relief valve A78 or V43 is leaky
F17 pos. Pressure During the start phase, it was not possible to reduce the dialysate pressure to a
value below 760 mmHg (measuring tolerance: ±30 mmHg, test running time
120 sec).
– The loading pressure is too high
– The UF pump is defective
F18 pos. Pressure The working point (116 digits) of the differential amplifier cannot be set correctly
(test running time 120 sec).
– The pressure variations in the system are too large
F19 pos. Pressure Completion of pressure measurement was not possible (max. test running time
120 sec).
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
– Acknowledgement (P_DIAL, X633L/B6) X632/A29 is defective
F20 pos. Pressure Positive pressure holding test failed. A pressure drop by more than
±80 mmHg/min was detected in the hydraulic system during flow-off.
– Leakage in the hydraulic system
– The UF pump spring is defective
– The air separation pump fails to occlude
– Relief valve leaking
– V84 leaking
F21 pos. Pressure The dialysate pressure cannot be set to a value between 460 and 760 mmHg
±30 mmHg (10 sec).
– The heat exchanger is defective
– Problem in the hydraulic system
F22 pos. Pressure The air separation pump is not running during the test phase (2 sec).
– Control line (AIR_SEP+/A22) ASP/... not 24 V
– Control line (AIR_SEP–/C22) ASP/... not 0 V
– Acknowledgement (ACKN_ASP, X634L/B10) X632/A19 not 12 V.
F23 pos. Pressure Pressure drop in the hydraulic system during the measuring phase (8 s).
Change larger than +4 digits or larger than -8 digits.
– Leakage in the pump segment of the air separation pump
– Leakage in the heat exchanger
– Acknowledgement (P_DIAL, X633L/B6) X632/A29, change in voltage too
large
F28 pos. Pressure ASP functional test (running and delivery test)
– ASP line segment is occluded
– ASP line segment has been incorrectly inserted (check direction of delivery)
– ASP is not running (electrically or mechanically).
– V87 electrically or mechanically closed
F01 UF Function The pause between the strokes of the UF pump 1 was shorter than 220 ms.
Correct volume delivery is not ensured due to too short a return.
– The pump rate emitted by CPU1 is too high
F02 UF Function The pulse time for the UF pump 1 is shorter than 180 ms. Correct volume
delivery is not ensured due to too short an emission time.
– The monoflop on P.C.B. LP 634 is defective (IC42/R82/C47)
F03 UF Function The pulse time for the UF pump 1 is longer than 500 ms. A maximum rate of
5000 ml/h is not possible.
– The monoflop on P.C.B. LP 634 is defective (IC42/R82/C47)
F07 UF Function The change in pressure after a stroke is less than 20 mmHg.
– The UF pump 1 is mechanically defective
– Acknowledgement (UF_P1_CTL, X632/C27) X634R/A24, no LOW pulses
F09 UF Function Dialysate pressure is outside the measuring range (15 s).
– UF pressure transducer defective.
– The D-A converter (IC11) on P.C.B. LP 632 is defective
– The operational amplifier (IC1/IC3) on P.C.B. LP 632 is defective
F01 Conductivity The conductivity failed to be within the scale limits or to stabilize within
10 minutes (±0.1 mS/10 sec).
– Concentrate is not connected
– Acknowledgement (COND_SIG, X633L/B8) X632/A22, voltage outside
the measuring range or unstable
F02 Conductivity Detuning in positive direction not more than 0.5 mS (10 sec).
– Acknowledgement (COND_SIG, X633L/B8) X632/A22 insufficient
– Detuning (COND_DET, X632/A21) X633L/B31 insufficient
F03 Conductivity Detuning in negative direction not more than 0.5 mS (10 sec).
– Acknowledgement (COND_SIG, X633L/B8) X632/A22 insufficient
– Detuning (COND_DET, X632/A21) X633L/B31 insufficient
F05 Conductivity CPU1 failed to transmit a bibag® status message within 3 sec.
– Run-time problem (software).
F08 Conductivity CPU1 fails to increase the working point (when the conductivity is < 40 mS/cm
uncompensated) for the bibag® conductivity by > 5 digits.
– Detuning (HOT_RINSE, X634R/C24 X633R/A16) not 12 V
– P.C.B. LP 633-5 T2 or IC26 defective
F01 DIASAFE plus Present options and DIP switch settings do not match.
F01 HPU
CPU1 system status (MST), HPU status and DIP switch/array 2 changed during
the test running time.
– DIASAFE plus:
CPU2: Array 2, DIP switch 2 not OFF
CPU2: Array 2, DIP switch 3 not OFF
– MST transmitted by CPU1 not matching with the set DIP switch of array 2.
– DIP switch/array 2 changed during the test running time.
– HPU logged off
F34 DIASAFE plus Pressure holding test failed to be passed. Max. number of treatments
exceeded?
– Diasafe filter membranes leaking/worn.
F02 DIASAFE plus Dialysate outlet pressure (DA1) outside the permissible range (10 s).
F02 HPU Test range for DA1 –25 mmHg² P_dial ² 55 mmHg
– Acknowledgement DA1 (P_DIAL, X633L/B6) X632/A29
– Acknowledgement DA2 (see HPU diagram)
F03 DIASAFE plus Cross comparison of both pressure transducers (DA1 / DA2) is outside the
F03 HPU acceptable tolerance (10 s).
P(DA2) == P(DA1) ±20 mmHg
– Acknowledgement DA1 (P_DIAL, X633L/B6) X632/A29
– Acknowledgement DA2 (see HPU diagram)
F41 DIASAFE plus The test valve V183 is leaking. Pressure increase in the system of DP(DA2) >
F41 HPU 30 mmHg within 4 s.
– V183 open, contaminated, or mechanically defective
– HPU, output stage etc. defective
F42 DIASAFE plus No pressure increase of DP(DA2) > 200 mmHg within 4 s after opening the test
F42 HPU valve V183 in the system.
– V183 fails to open, mechanically defective
– Compressor P185 defective, not running
– HPU, V183 and/or compressor P185 output stage etc. defective
F43 DIASAFE plus The lower pressure test range of DP(DA2) > 300 mmHg failed to be achieved
F43 HPU within 1 s after closing the test valve V183.
– HPU, output stage etc. defective
F44 DIASAFE plus The upper pressure test range of DP(DA2) 750 mmHg was exceeded within 4 s
F44 HPU after closing of the test valve V183.
– HPU, output stage etc. defective
F04 DIASAFE plus The air separation pump P97 is running although valve V43 is closed.
– HPU, output stage etc. defective
F26 DIASAFE plus Insufficient test pressure (P < 750 mmHg) in the system.
– Hydraulics system leaking
F27 DIASAFE plus After the valve V189 opened, the pressure drop in the system was insufficient
(DP < –70 mmHg).
– Valve V189 electrically or mechanically not open
– Diasafe filter strongly contaminated
– Filter before/after V43 strongly contaminated
F28 DIASAFE plus Pressure increase in the system fails to exceed P > 760 mmHg.
– Diasafe filter membrane leaking (major leakage)
– No Diasafe filter installed
F29 DIASAFE plus Pressure holding test failed to be passed. Excess pressure drop within a
measurement time of 30 s (DP > –10 mmHg).
– Diasafe filter membrane leaking
F30 DIASAFE plus During the pressure holding test valve V189 was closed (according to electronic
acknowledgement).
– Valve control failed
F34 DIASAFE plus See error message between F01 and F02 DIASAFE plus
F41 DIASAFE plus See error message between F01 and F02 DIASAFE plus
F42 DIASAFE plus See error message between F03 and F04 DIASAFE plus
F43 DIASAFE plus See error message between F03 and F04 DIASAFE plus
F44 DIASAFE plus See error message between F03 and F04 DIASAFE plus
F01, F02 and F03 cause the V84 monitoring flag to be set. I.e. after one
of these error messages has occurred, Bergström or ISO-UF dialysis is
no longer possible, since it is not possible to switch the flow off. The V84
malfunction can be eliminated by correctly performing Dis I – IV.
Another possibility of correcting the problem can be found in the service
mode (by a service technician only) in the NOVRAM menu item (clear
V84 malfunction).
Rinse Failure F05 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I – V.
It was impossible to open the pressure switch for PSW_104 (S124)
(bicarbonate).
– Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Concentrate: X633L/A20
Rinse Failure F06 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I – V.
It was impossible to open the pressure switch for PSW_102 (S123)
(concentrate).
– Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Concentrate: X633L/A20
Rinse Failure F07 Rinse-free program, Dis, HDIS, or mandatory rinse in Dis I to V.
Pressure drop during the monitoring phase on PSW_104 (S124) (bicarbonate)
or pressure build-up impossible.
– Check switching point of pressure switch
– Check loading pressure
(possibly splinter or contamination in orifice 151, remove and purge the
tubing from both ends)
– Check negative pressure and orifice (89)
(for this purpose, remove and purge the tubing from both ends)
– Check CDS valve (104)
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Bicarbonate: X633L/A19
– Cartridge filter (F210) upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be distinguished by
different adapters.
Rinse Failure F08 Rinse-free program, Dis, HDIS, or mandatory rinse in Dis I – V.
Pressure drop during the monitoring phase on PSW_102 (S123) (concentrate)
or pressure build-up impossible.
– Check switching point of pressure switch
– Check loading pressure
(possibly splinter or contamination in orifice 151, remove and purge the
tubing from both ends)
– Check negative pressure and orifice (89)
(for this purpose, remove and purge the tubing from both ends)
– Check the check valve (117) and filter (119).
– Check CDS valve (102)
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Concentrate: X633L/A20
– Cartridge filter (F210) upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be distinguished by
different adapters.
Rinse Failure F09 Five minutes before the end of the mandatory rinse in Dis I – V.
The pressure switch PSW_104 (S124) (bicarbonate) or PSW_102 (S123)
(concentrate) did not open after pressure reduction.
See Rinse Failure F12.
Rinse Failure F12 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I – V.
The pressure switch for PSW_104 (S124) (bicarbonate) and for PSW_102
(S123) (concentrate) could not be opened.
– Membrane pumps fail to run
– V102 or V104 fails to open
– Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Bicarbonate: X633L/A19
Rinse Failure F13 Rinse-free program with following Dis or HDIS or mandatory rinse as individual
program in Dis I – V.
Pressure drop during the monitoring phase on PSW_104 (S124) (bicarbonate)
and PSW_102 (S123) (concentrate) or pressure build-up impossible.
– Check switching point of pressure switch
– Check loading pressure
(possibly splinter or contamination in orifice 151, remove and purge the
tubing from both ends)
– Check negative pressure and orifice (89)
(for this purpose, remove and purge the tubing from both ends)
– Check the check valve (117) and filter (119).
– Check CDS valve (102/104)
– Verify tightness of CDS path
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
Bicarbonate: X633L/A19
Concentrate: X633L/A20
– Cartridge filter (F210) upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be distinguished by
different adapters.
In case of F07, F08 and F13, the DO NOT SWITCH OFF !! message
can, in addition, be alternately displayed.
However, this message is displayed only if a mandatory rinse program
is requested, since the concentrate and bicarbonate lines still have to
be emptied before the device is switched off.
7.2.3 Rinse section test (check of V91, V99, V100) (only devices with CDS)
Rinse Failure F11 Three minutes before the end of the mandatory rinse in Dis I – V.
The pressure switch PSW_102 (S123) (concentrate) did not open after pressure
reduction.
– Pressure on distribution piping > 500 mbar (according to specification, the
permissible pressure is max. 500 mbar). Pressure peaks on distribution
piping: Frequently occurs in distribution pipings with user points if e.g.,
several patients are disconnected simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired= 700 mbar ±20 mbar
– Membrane pumps fail to run
– V102 fails to open electrically or mechanically
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
X633L/A20
V91/V100 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
V91 or V100 cannot be opened.
– V91 or V 100 fail to open electrically:
P.C.B. LP 634: V91 = X634L/A12; V100 = X634L/C13
– V 91 or V 100 mechanically not open:
Check filter (148) before V100, or valves clogged
– V99 constantly open (electrically P.C.B. LP 634: X634L/B12 or
mechanically)
– V 102 not open
– Pressure switch for PSW_102 (S 123) fails to switch
V99 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
V99 cannot be opened.
– V 99 fails to open electrically:
P.C.B. LP 634: X634L/B12
– V 99 fails to open mechanically:
Check filter (149) before V99, or V99 clogged
– Pressure switch for PSW_102 (S123) fails to open
V130 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
V130 cannot be opened.
– V130 electrically defective:
P.C.B. LP 634: X634L/A4
– V130 mechanically defective or clogged
– Pressure switch for PSW_102 (S123) fails to open
– Check the tubing arrangement for the bicarbonate suction line and the
bibag® block
Rinse Failure F14 Shortly before the end of the mandatory rinse in Dis I – V.
Rinse section test not completed correctly. Possibly caused by flow problems.
7.2.4 Rinse section test (check of V91 and V98) (only devices without CDS)
Rinse Failure F14 Three minutes before the end of the mandatory rinse in Dis I – V.
It was not possible to readjust the flow to 750 ml/min ±50 ml/min. V91 defective.
V91 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
After V91 has opened, a flow > 950 ml/min failed to develop. V91 or valve V98
defective.
Rinse Failure F14 Shortly before the end of the mandatory rinse in Dis I – V.
Rinse section test not completed correctly. Possibly caused by flow problems.
7.2.5 Rinse section test (check of V91, V99, V100, V130) (only devices without CDS)
Rinse Failure F15 Three minutes before the end of the mandatory rinse in Dis I – V.
DS (bibag® pressure switch 134) could not be opened at the beginning of the
test.
– Check pressure switch: Switching point: desired value: 130 mbar +30 mbar
– Suction error of bicarbonate pump
– V91 constantly electrically or mechanically open
– V99/100 constantly electrically or mechanically closed
V91 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
It is impossible to build up pressure on DS (bibag® pressure switch 134) via V91.
– Pressure switch fails to close mechanically: check switching point
– V91 fails to open electrically: P.C.B. LP 634: X634L/A12
– V91 fails to open mechanically (possibly clogged)
– V130 electrically not closed: P.C.B. LP 634: X634L/A4
– V130 fails to close mechanically (possibly clogged)
– bibag® connector leaking (check O rings)
– Sealing on the bicarbonate suction tube leaking
– Check acknowledgement of pressure switch on P.C.B. LP 633-5:
X633L/A8
– V99 constantly electrically or mechanically open
V100 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
It is impossible to build up pressure on DS (bibag® pressure switch 134) via
V100.
– V100 fails to open electrically: P.C.B. LP 634: X634L/C13
– V100 fails to open mechanically (possibly clogged)
– V91 constantly electrically or mechanically open
– Concentrate pump fails to pump
– Filter (148) clogged
– Pressure switch fails to open
Rinse Failure F16 Three minutes before the end of the mandatory rinse in Dis I – V.
DS (bibag® pressure switch 134) cannot be closed. V99 or V130 is leaking.
– V91 fails to open electrically or mechanically
– V99 constantly electrically or mechanically open
– V130 constantly electrically or mechanically open
– Sealing on the concentrate suction tube leaking
– Pressure switch fails to close
V99 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
DS (bibag® pressure switch 134) cannot be opened. V99 does not open.
– V99 fails to open electrically or mechanically
– V100 fails to open electrically or mechanically
– Pressure switch fails to open
– V91 electrically or mechanically open
– Filter (149) before V99 clogged
Rinse Failure F17 Three minutes before the end of the mandatory rinse in Dis I – V.
DS (bibag® pressure switch 134) cannot be closed.
– V91 fails to open electrically or mechanically
– V130 electrically or mechanically open
– V100 electrically or mechanically open
– Pressure switch fails to close
V130 Failure Three minutes before the end of the mandatory rinse in Dis I – V.
DS (bibag® pressure switch 134) cannot be opened.
– V130 fails to open electrically or mechanically
– Pressure switch fails to open
– Check the tubing arrangement for the bicarbonate suction line and the
bibag® block
– Bicarbonate line squeezed at strain relief
– Narrowing in the reducer on the bibag® connector
Rinse Failure F 20 Impossible to close the pressure switch (134) via V91/100
– V91 fails to open electrically or mechanically
– V130/V188 electrically or mechanically open
– Pressure switch fails to close
V188 Failure The pressure on pressure switch (134) cannot be reduced via V188.
– V188 fails to open electrically or mechanically
– Pressure switch fails to open
– Check the tubing arrangement for the concentrate suction line and the air
separation block
– Concentrate line squeezed at strain relief
Rinse Failure F14 Shortly before the end of the mandatory rinse in Dis I – V.
Rinse section test not completed correctly. Possibly caused by flow problems.
V39 Failure On opening V39 a difference in pressure (averaged value V39 open – averaged
value V39 closed) is detected on the dialysate pressure transducer (182):
bibag® device: < 20 mmHg
– V39 fails to open / close electrically or mechanically (possibly hydraulic
processing unit defective)
– It is impossible to re-adjust the degassing pump (P.C.B. LP 634).
– V91, V99, V100 fails to open electrically or mechanically
– Dialysate pressure transducer (182) defective or not calibrated (possibly
HPU P.C.B. LP 941 defective)
– Filter 210 (before degassing orifice) clogged
7.2.7 Further messages which may be displayed before or during a cleaning program
Shunt cover open Start of a cleaning program or during a cleaning program in RI I to II, HR I to III,
Dis I to V.
The shunt interlock is not closed.
Conc line not conn Start of a cleaning program in RI I to II, HR I to III, Dis I to V, or end of the
disinfectant suction phase in Dis V.
The concentrate suction tube is not connected to the rinse chamber.
Reconnect the concentrate suction tube to the rinse chamber.
Bic line not conn Start of a cleaning program in RI I to II, HR I to III, Dis I to V, or end of the
disinfectant suction phase in Dis V.
The bicarbonate suction tube is not connected to the rinse chamber.
Reconnect the bicarbonate suction tube to the rinse chamber.
Rinse after Disinf. Selection of a cleaning program, although a mandatory rinse has been
requested in HR.
A disinfection program has been stopped and subsequently a rinsing or hot
rinsing program started.
BIBAG cover open Upon start of a cleaning program in RI I to II, HR I to III, Dis I to V.
The bibag® connector is not closed (flap not closed).
Heater Error During the CDS rinsing phase at the end of a hot rinsing program or a hot
disinfection program in CDS: HR I to III, Dis II to IV.
The heater signal (P.C.B. LP 633-5: X633R/A26) is not changing for > 40 sec.
Watchdog Error This error message can only be displayed shortly after turning power on.
Turn the device off and back on again.
Check the WD relay and components. Check CPU2/CPU1.
Check the plug connectors on the monitor.
NTC 109 switched off No valid value has been filed during start in the NOVRAM. The temperature
difference between NTC109 and NTC3 is too high.
Switch off NTC 109 in the SETUP MENU, or recalibrate the temperature.
Cyclic PHT F03 IC1 or IC3 on P.C.B. LP 632 is defective, or device leakage
Cyclic PHT F04 It was not possible to complete the test within a specific time interval
Shunt Cover open – P.C.B. LP 633-5 C24 (100 nF) temporarily short-circuited
(temporarily) – Shunt interlock defective (check switches)
UF1 volume - Error Failure to pass the test for an UF pump. The fill volume for the secondary air
separator is outside the tolerance of 100 ml ±4 ml.
Possible cause:
– The UF pump fails to deliver correctly (not calibrated or mechanical defect)
– If the test result is > 104 ml, the problem can also be caused by air coming
from a poorly deaerated dialyzer
F327 UF failure Pause between two UF1 pump strokes less than 220 ms.
Possible cause:
– CPU1 defective
F328 UF failure Pulse time of one UF1 pump stroke less than 180 ms.
Possible cause:
– Controlling monoflop on LP 634 defective
F329 UF failure Pulse time of one UF1 pump stroke exceeds 500 ms.
Possible cause:
– Controlling monoflop on LP 634 defective
F331 UF failure Theoretical/actual rate of the UF1 pump deviates by more than ±10 %.
Possible cause:
– System error
F332 UF failure UF1 pump stopped for more than the maximum time period.
Possible cause:
– Controlling output stage on LP 634 defective
– UF pump interruption
– System error
F333 UF failure Volume changes by more than 10 ml during prescribed standstill (only
monitored if OD is dark).
Possible cause:
– System error
F350 UF failure A difference of more than 100 ml between the CPU1 and the CPU2 volume is
detected during an UF data transfer after turning the UF unit on.
Possible cause:
– System error
F351 UF failure CPU2 could not detect plausibility of the CPU1 UF parameters.
Possible cause:
– System error
F361 UF failure CPU1 sent the UF parameter set to CPU2 and has not received a release from
CPU2 after a timeout of 30 seconds.
Possible cause:
– System error
F363 UF failure CPU2 did repeatedly not receive a complete UF parameter set.
Possible cause:
– System error
F364 UF failure UF1 volume change although the UF goal has already been reached.
Possible cause:
– System error
HPU Error F00 The HPU logs off with index STATUS_ER; no bit is set in the error bit field.
– Problem on P.C.B. LP 941
– Problem on CAN distributor board
– Problem on P.C.B. LP 763
– Problem on P.C.B. LP 630
HPU Error F01 The cyclic communication has failed for more than 2 seconds.
– System error
HPU Error F07 The HPU was logged off by the monitor. Will not be displayed since CPU1 has
already stopped the communication.
– System error
HPU Error F15 Error in the HPU software. Valves are activated incorrectly.
– System error
HPU Error F98 Proceeding to the T1 test is not allowed after restart.
– System error
BPM: Measurement value Blood pressure cuff not correctly connected. Leakage in pressure tubing. Alarm
outside the alarm limits limits not adjusted to the patient.
BPM: Cuff cannot be inflated. CPU-P1 pressure < 25 mmHg 15 s after the start of a measurement
BPM: Cuff pressure lower CPU-P1 pressure < 40 % of the desired pressure
than preselected pressure
E72 BPM error Pressure difference between CPU-P1 and CPU-P2 on 1st stage > 20 mmHg
(Remove the cuff from the
patient!)
F83 BPM error CPU-P1 pressure zero point drift after T1 test
(Remove the cuff from the
patient!)
E84 BPM error CPU-P2 pressure zero point drift after T1 test
(Remove the cuff from the
patient!)
E86 BPM error CPU-P2 pressure > 300 mmHg for more than 15 s.
(Remove the cuff from the
patient!)
E90 BPM error CPU-P2 pressure > 330 mmHg for more than 300 ms.
(Remove the cuff from the
patient!)
E91 BPM error CPU-P2 pressure > 15 mmHg for more than 180 s.
(Remove the cuff from the
patient!)
E93 BPM error CPU-P1 reference voltage deviation. Valid range: 1.4-1.6 V
(Remove the cuff from the
patient!)
E94 BPM error Deviation +12V pump voltage. Valid range: 11-13 V
(Remove the cuff from the
patient!)
E96 BPM error Deviation of the calibration parameters between RAM and EEPROM
(Remove the cuff from the
patient!)
E98 BPM error Pressure change during measurement ≤ 3 mmHg / 5 milliseconds within
(Remove the cuff from the 25 seconds. Monitoring will then be active in order to detect a pressure stage.
patient!)
E99 BPM error Pressure difference between CPU-P1 and CPU-P2 on 1st stage > 20 mmHg
(Remove the cuff from the
patient!)
Warning
The precision of the measuring equipment used during calibration plays
an important role for the accuracy of the OCM measurement.
The measuring equipment used for calibrating the conductivity must
have a precision of 0.05 mS/cm in a temperature range of 35 °C to
39 °C.
Printer module
9 Calibration / adjustment
9.1 CALIBRATION
Note
If detected before starting calibration on the hydraulics unit, potential
deposits must be removed by running an appropriate disinfection
program.
CALIBRATION Conf
key
DIAGNOSTICS
MISCELLANEOUS
SETUP MENU
Setting the blood pump stop alarm (see chapter 9.1.5 page 9-6)
Calibrating the Single-Needle blood pump (see chapter 9.1.11 page 9-11)
(Single-Needle pressure) (optional)
Adjusting the current increasing pulse (see chapter 9.1.15 page 9-14)
Setting the Hall sensors in the heparin pump (see chapter 9.1.46 page 9-38)
CALIBRATION
Conf
key
Navigation
Menu display Description
Conf
key
Mute
key
Art. PRESS. CHECK Conf
key
ACKNOWLEDGED
art.press: ±0 mmHg
Conf After
back to menu ? key approx. 3 s
Activate various pressure
values. The values on the
display, the art. pressure art.gain.: +210 mmHg
display and the reference
instrument must be identical.
Check whether the scale Apply ≥210 mmHg to art.
limits can be reached. pressure transducer
Tolerance: ±10 mmHg (reference instrument!).
Mute
key
Note
The accuracy of the pressure gauge used must correspond to that of the
UMED or HMED.
The accuracy of the pressure gauge used must at least correspond to
the following values:
–1 to 2 bar: ±5 mbar
2 to 8 bar: ±20 mbar
Conf
key
Acknowledge by pressing
the Start/Stop key on the
blood pump module
Acknowledge by pressing
the Start/Stop key on the
blood pump module
9.1.5 Without menu display: Setting the blood pump stop alarm
Conf
key
Mute
key
Ven. PRESS. CHECK Conf
key
ACKNOWLEDGED
ven.press: ±0 mmHg
Conf after
back to menu ? key approx. 3 s
Activate various pressure
values. The values on the
display, the venous pressure ven.gain.: +500 mmHg
display and the external
reference instrument must
be identical. Check whether Connect the external
the scale limits can be reference instrument
reached.
Tolerance: ±10 mmHg
Apply ≤ 500 mmHg to the
Esc venous pressure transducer.
key Read the external reference
instrument and set the value on the
display by pressing the +/– keys.
Mute
key
Note:
If calibration of this DATA STORED
function is not possible,
the “CAL. VEN. P_MODULE”
mode must first after
be performed approx. 3 s
in the air detector.
Note
The accuracy of the pressure gauge used must correspond to that of the
UMED or HMED. The accuracy of the pressure gauge used must at
least correspond to the following values:
-1 to 2 bar ±5 mbar
2 to 8 bar ±20 mbar
9.1.7 Part 4: Calibrating the venous pressure measurement in the air detector
Conf
key
ven.press: ±0 mmHg
Esc
key
Note:
Check zero point and slope;
if necessary, repeat the procedure.
Note:
When adjusting the
air detector,
execute the
CAL. VENOUS PRESSURE
menu item.
Potentiometer 3
Potentiometer 4
CALIB. (B)-PUMP-RATE
Conf
key
Conf
key
Adjust a rate of
≥ 550 ml/min* on the
art. BP-Rate CHECK Conf (B)P–Rate=550 ml/min art. blood pump.
key
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the (+/–) keys.
Conf
key
Using a syringe, set the
calib. sn. BP-RATE pressure at the P-SN connector Conf (B)P–Rate=550 ml/min Esc
according to the set key key
stroke volume (see table).
Set a rate of
≥550 ml/min*
SN. BP-Rate CHECK Conf (B)P–Rate=550 ml/min on the SN blood pump.
key
Note: Set the line diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min represents a default value. It can be changed using the (+/–) keys.
Upper switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
2. Press the and (+ / –) keys to set the stroke volume and confirm with Start/Stop.
Acknowledge by pressing
the Start/Stop key on the
SN blood pump.
Upper switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
Note
If necessary, change the Single-Needle stroke volume
(see chapter 9.1.10 page 9-10):
– Simultaneously press the and Start/Stop keys.
– Press the and (+ / –) keys to set the stroke volume and confirm
with Start/Stop.
Conf
key
ACKNOWLEDGED
After
approx. 3 s
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Also refer to degassing pump pressure (see chapter 9.2.2 page 9-42).
DATA STORED
After
approx. 3 s
Note
If the 300 / 500 / 800 flow rates cannot be set or flow alarm problems
are incurred after "Calibrate flow", maybe the setting of the current
increasing pulse must be changed.
Note
The flow selected first is accompanied by the DIASAFE-filling act.
message which is displayed for the duration of 17 balancing chamber
switchings.
MP1: 1 V/cm
GND
MP8: 2 V/cm
9.1.16 Alternative method of adjusting the current increase (if an oscilloscope is not
available)
Note
After having adjusted the current increasing pulse, check and, if
necessary, readjust the 300 / 500 / 800 flow rate settings.
DATA STORED
After
approx. 3 s
DATA STORED
After
approx. 3 s
CALIB. TEMPERATURE
Conf
key
Note
Accuracy of the measuring instrument to be connected externally:
±0.2 °C.
Temp. ADJUSTMENT
adj. temp to 37 °C
Conf
key
XXXX ADC digits
adj. temp to 37 °C Esc
key
XXXX DAC digits
Connect the external reference instrument.
Place the CONC suction tube into the CONC container. XXXX ADC digits
Adjust the digital value
in display field by
pressing the +/– keys.
System in bypass
NTC3, 12 bit
NTC3 temperature
NTC4, 12 bit
Dynamic measuring range switching
Note
The hydraulic unit must be installed, the rear wall closed and the
DIASAFE® plus must be covered.
Check TEMPERATURE
Use ◄► to select
the Temperature
setting field
XXX.X °C
set temp = XX.X °C
Note
The hydraulic unit must be installed, the rear wall closed and the
DIASAFE® plus must be covered.
OCM TEMP.-Check
OCM TEMP.-Check
Conf
key
XXXX ADC digits
Open the shunt cover
to switch the system
to the bypass mode
XXXX °C
act temp = 37.0 °C
XXXX ADC digits
Use ◄► to select
the Temperature
setting field XX.XX °C
Note
The hydraulic unit must be installed, the rear wall closed and the
DIASAFE® plus must be covered.
CAL. MIXING-SYSTEM
Conf
key
The membrane pumps must be run in to adjust the concentrate and the bicarbonate pump to operating
temperature before their volume is measured in liters.
The concentrate tubes (bicarbonate / concentrate) are located in a container filled with water.
Run-In PUMPS
Conf
key
Mute
key
After 9 s
Mute
key
Conf
key
e.g.:
BC-Volume = 30.19 ml
Mute
key
DATA STORED
After
approx. 3 s
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
cal. CONC.-PUMP-VOL
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Note
Measuring cylinder accuracy: ±0.5 %.
9.1.27 Part 12.4: Measuring the volume of the concentrate pump in liters
Conf
key
Mute
key
After 9 s
Mute
key
ACKNOWLEDGED Esc
key
After
approx. 3 s
Mute
key
Notes:
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
calib. BIC.-PUMP-VOL
Conf
key
Mute
key
DATA STORED
After
approx. 3 s
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
9.1.29 Part 12.6: Measuring the volume of the bicarbonate pump in liters
Conf
key
Mute
key
After 9 s
Mute
key
ACKNOWLEDGED Esc
key
After
approx. 3 s
Mute
key
Notes:
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
Conf
key
Mute
key
After 9 s
pulse-amount = 50 Esc
key
Mute
key
ACKNOWLEDGED Esc
key
After
approx. 3 s
Mute
key
Notes:
Note
Measuring cylinder accuracy: ±0.5 %.
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
CALIB. CONDUCTIVITY
Conf
key
CONDUCTIVITY Set
Conf
key
Mute Mute
key key
ACKNOWLEDGED ACKNOWLEDGED
After After
approx. 3 s approx. 3 s
Mute Mute
key key
After After
approx. 3 s approx. 3 s
XXXX Steps
(concentrate pump)
XXXX Steps
(bicarbonate pump)
conc. set = ±0 %
If no key is
pressed for
act temp = XX.XX °C
approx. 4 s
Use +/– to reduce/increase XX.XX mS/cm
the concentrate pump volume
until the conductivity is
inside the alarm limits
13.9 – 14.5 mS/cm XX.XX mS/cm
conc. set = +X %
XX.XX mS/cm
Mute
key
XX.XX mS/cm
ACKNOWLEDGED
Alpha Temperature
display setting
After
approx. 3 s
After
approx. 3 s
Mute
key
DATA STORED
After
approx. 3 s
Mute
key
DATA STORED
After
approx. 3 s
Note
If the value is outside ±25, the conductivity measuring system of the
dialysis device must be checked.
CONDUCTIVITY Check
Conf
act. cond: XX.X mS/cm
key
After
Steps for bicarbonate pump
approx. 3 s
CD cell 110; 12 bit / 8 bit switching
act.cond: XX.X mS/cm
Verify the conductivity.
Check:
Reference meter, display,
LED bargraph display.
Repeat the
procedure with
different values.
Esc
key
OCM COND.-Check
Conf
OCM Cond.-Check
key
OCM Cond.-Check
XXXX ADC digits (OCM)
Use ◄► to select
the Concentrate
XX.XX mS/cm
setting field.
Repeat the
procedure with
different values.
Esc
key
Conf
act temp = XX.XX °C
key
XX.XX mS/cm
act temp = XX.XX °C
Note
The temperature / conductivity compensation test allows checking the
two conductivity cells (7, 110) against each other.
The tolerance of the two conductivity cells may not exceed 0.05 mS/cm.
If this tolerance is exceeded, the OCM pulse must be calibrated.
Conf
key
Note
The accuracy of the pressure gauge used must correspond to that of the
UMED or HMED.
The accuracy of the pressure gauge used must at least correspond to
the following values:
-1 to 2 bar ±5 mbar
2 to 8 bar ±20 mbar
DIALYSATE Pressure
Conf
key
Use the potentiometer P3/LP 633 to set Switch off the flow.
the value indicated by the external reference
instrument on the alpha display. Leave the UF pump running
until approx. –500 mmHg are
indicated by the external
reference instrument.
Repeat
the procedure Enter the value indicated by the
until 0 and –500 external reference instrument
mmHg correspond Not by pressing the +/– keys.
to the external OK
comparison Mute
instrument. key
Mute
key
DATA STORED
After
approx. 3 s
TMP-Check
Conf
key
Esc
key
Note:
Observe the venous pressure!
Note:
Switch the flow on and off
again once in a while,
to maintain the operating
temperature of the pressure
transducer.
PDIAL2 press-check
CALIBRATE BLD
Conf
key
After
approx. 3 s
Conf
key
Mute
key
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Mute
key
DATA STORED
After
approx. 3 s
**
BIBAG Cond.-Check Conf act.cond: XX.XmS/cm Esc
key key
Conf
key
Audible alarm
Are you sure ? Esc
key
Mute
key
ACKNOWLEDGED
After
approx. 3 s
9.1.45 Part 18: Initializing the NOVRAM, clearing the mandatory rinse, erasing a V84
malfunction
Audible alarm
Are you sure ? Esc
key
Mute
key
ACKNOWLEDGED
After
approx. 3 s
Esc
RESET Filterchange ? key
Mute
key
ACKNOWLEDGED
NOVRAM
After
Conf approx. 3 s
key
Audible alarm
Reset MAND. RINSE Conf Are you sure ? Esc
key key
Mute
Audible alarm key
Reset V84 Conf Are you sure ? Esc
key key
ACKNOWLEDGED
Mute
key
Conf After
back to menu ? key approx. 3 s
ACKNOWLEDGED
BACK TO MAIN MENU ?
After
approx. 3 s
9.1.46 Without menu display: Setting the Hall sensors in the heparin pump
Setting Hall sensor 1 – Briefly press the key (syringe plunger holder moves to its upper
end position)
– Disconnect Hall sensor 2 (remove plug)
– Insert syringe (e.g., type BD Plastipak 20 ml)
– Key the key pressed (until the syringe plunger holder has reached
its lower end position).
– Adjust Hall sensor 1, so that the the bottom of the plunger and the
end of the syringe's scale (0 ml) are on the same level.
The must not be any mechanical noise. If the error message E55 occurs
during the Hall sensor adjustment, the message can be acknowledged
by pressing the Start/Stop key. If Hall sensor 1 was adjusted, the error
message E55 must not occur anymore.
Each time Hall sensor 1 was adjusted, the complete setting for Hall
sensor 1 have to be repeated.
– After adjusting Hall sensor 1, Hall sensor 2 has to be re-connected.
– Briefly press the key (syringe plunger holder moves to its upper
end position).
Setting Hall sensor 2 – Press and hold the key until the syringe plunger holder is approx.
2 cm before its lower end position.
– Manually turn the threaded spindle in delivery direction (one scale
line on the syringe).
– Key the key pressed (until the syringe plunger holder has reached
its lower end position).
– Adjust Hall sensor 2 so that the bottom of the plunger is pressed
against the bottom of the syringe.
The must not be any mechanical noise. If the error message E55 occurs
during the Hall sensor adjustment, the message can be acknowledged
by pressing the Start/Stop key. If Hall sensor 2 was adjusted, the error
message E55 must not occur anymore.
Each time Hall sensor 2 was adjusted, the complete setting for Hall
sensor 2 have to be repeated.
– Briefly press the key (syringe plunger holder moves to its upper
end position).
Hall sensor 1
Hall sensor 2
Note
Measuring equipment for measurement points in the hydraulic unit:
UMED, HMED or pressure gauge
with a measuring range of –1 to +2.2 bar.
Condition Flow on
V41
A
A61
Note
If the pressure of the degassing pump was changed, make sure to
check the loading pressure and readjust, if necessary.
9.2.2.1 Calibration of the negative degassing pressure on installation sites situated at higher altitudes
Note
If the loading pressure was changed, make sure to check the degassing
pump pressure and readjust, if necessary.
A65
B
A78
Note
When using scales as a measuring instrument, please ensure that
concentrate is not connected.
2
3
Measuring equipment UMED, HMED or pressure gauge (e.g., 0–1 bar, accuracy ±1 %) and
syringe. Measurement setup see the following figures.
Check / adjustment Connect the measuring equipment as illustrated in the diagram below.
– Select the DIAGNOSTICS service menu, continue with:
READ INPUTS
READ DIGITAL INPUTS
CPU1: RD DIGITAL INP
E: CPU1_PSW_V102
– Activate the audible alarm by pressing the (Alarm) Tone Mute key
(depressurized: alarm on).
– If this menu option has been selected, the solenoid valve 102 is
closed.
– Use the syringe to build up a pressure of 0.7 bar.
– Use an artery forceps to clamp the line ((1)), so that a pressure of
0.7 bar continues to act on the pressure switch.
– Verify the switching point by means of the audible alarm. Desired
value:
Pressure switch (Envec): 0.68–0.72 bar.
General notes on the The calibration described below can be performed by means of a
bibag® pressure multimeter or the DIAGNOSTICS service menu.
transducer
To read the corresponding signal, you must select the menu option
E: CPU1_BIBAG_PSW in the diagnostic program.
The calibration of the pressure switch is not possible. The pressure
switch is set to a fixed position and can be checked in the
DIAGNOSTICS service menu.
The audible signal can be activated by pressing the Tone Mute key.
Selection in the 1. The DIAGNOSTICS service menu can only be called in service
DIAGNOSTICS service mode.
menu
– The device must be switched off.
– Set the Service switch to ON (up).
– Turn the device on.
Wait briefly, until the Service mode screen is displayed.
– Use the and keys to select DIAGNOSTICS.
Press the Confirm key to confirm the selection.
2. In the DIAGNOSTICS service menu, continue as follows:
– Use the and keys to select DIAGNOSTICS.
Press the Confirm key to confirm the selection.
– Use the and keys to select READ INPUTS.
Press the Confirm key to confirm the selection.
– Use the and keys to select READ DIGITAL INPUTS.
Press the Confirm key to confirm the selection.
– Use the and keys to select CPU1: RD DIGITAL INP.
Press the Confirm key to confirm the selection.
– Use the and keys to select E: CPU1_BIBAG_PSW.
Press the Confirm key to confirm the selection.
– The active levels 0000 or 1111 are shown on the display.
Note
In the menu item E: CPU1_BIBAG_PSW, the active level is identified
by 1111.
The active level of the circuit output is LOW (< 1 V).
Fig.: 450-2
Potentiometer P1
LED D5
Potentiometer P2
LED D10
Potentiometer P5
Jumper J1
P.C.B.
Luer-Lock adapter
Fluid level
Warning
For a calibration of the air detector, the device must be in the Service
mode. The ambient temperature should be between 15 °C and 35 °C.
Warning
The calibration using the set for the air detector calibration is only valid
for devices running with Fresenius tubing systems.
Observe the “use by” date!
9.3.1.1 Adjustment using the set for air detector calibration (see adjustment instructions)
Measuring equipment UMED, HMED or pressure gauge, adjusting block and checking
(reference) block
Calibrating the air detector The device must be in the Service mode,
the jumper J1 / P.C.B. LP 450-2 must be set to the calibration position.
– Fill the spherical recesses of the adjusting block with grease and use
a spattle to remove any excess grease so that only the recesses are
completely filled with grease.
– Place the greased adjusting block into the drip chamber holder (with
the bevelled edges first). It must be ensured that the ultrasonic
sensors click correctly into place in the spherical recesses of the
block. The adjusting block must not touch the holder wall, but must
hang freely between the sensors.
– Turn potentiometer 1 and potentiometer 2 on P.C.B. LP 450-2
clockwise, until the LED D5 and LED D10 on P.C.B. LP 450-2 are
dark.
Slowly (attention: time constant) turn potentiometer 1
counterclockwise, until LED D5 lights.
Slowly (attention: time constant) turn potentiometer 2
counterclockwise, until LED D10 lights (see figure LP 450-2).
– Set jumper J1 / P.C.B. LP 450-2 to the operation position.
LED D5 and LED D10 must both be dark.
– Take the adjusting block out of the drip chamber holder.
Remove residual grease using lint-free cloth and a permissible
disinfectant.
Checking the air detector – Fill the spherical recesses of the test block with grease and use a
spattle to remove any excess grease so that only the recesses are
completely filled with grease.
– Place the greased test block into the drip chamber holder. It must be
ensured that the ultrasonic sensors are correctly positioned in the
spherical recesses of the test block. The test block must not touch
the holder wall, but must hang freely between the sensors.
– LED D5 and LED D10 on P.C.B. LP 450-2 must both be illuminated.
If one or both LEDs are not illuminated, the calibration must be
repeated.
– Take the test block out of the drip chamber holder.
Remove residual grease using lint-free cloth and a permissible
disinfectant.
– In addition, the venous occlusion clamp has to be checked and the
optical detector has to be calibrated and checked (see
chapter 9.3.1.2 page 9-53).
9.3.1.4 Alternative: Adjustment without the set for air detector calibration
Calibrating the air detector The device must be in the Service mode,
the jumper J1 / P.C.B. LP 450-2 must be set to the calibration position.
– Insert the bubble catcher into the air detector (ultrasonic detector);
for the moment, do not insert the line into the venous occlusion
clamp.
– Fill the bubble catcher. The fluid level must be set to approx. 10 mm
above the top edge of the sensor holder.
– Turn potentiometer 1 and potentiometer 2 on P.C.B. LP 450-2
clockwise, until the LED D5 and LED D10 on P.C.B. LP 450-2 are
dark.
Slowly (attention: time constant) turn potentiometer 1
counterclockwise, until LED D5 lights.
Slowly (attention: time constant) turn potentiometer 2
counterclockwise, until LED D10 lights (see figure LP 450-2).
– After completion of the calibration procedure, set the jumper J1 /
P.C.B. LP 450-2 back to the operation position.
Checking the air detector – Lower the fluid level in the bubble catcher:
An alarm must be emitted. LED D5 and LED D10 light.
If one or both LEDs are not illuminated, the calibration procedure
must be repeated.
– Raise the fluid level in the bubble catcher:
It must be possible to clear the alarm; both LEDs must be off.
– Remove the bubble catcher from the air detector.
– In addition, the venous occlusion clamp has to be checked and the
optical detector has to be calibrated and checked (see
chapter 9.3.1.2 page 9-53).
10 Servicing / repair
To be observed after A disinfection, a T1 test and a check of the electrical safety must be
working on the performed after working on the hemodialysis device.
hemodialysis device
To be observed after After a disinfection program has been stopped or if the device is to be
stopping a disinfection preserved, the hemodialysis device must be disconnected from the
program water supply after a maximum of 3 days.
Warning
When the device is returned to use, check that the pressure of the water
supply meets the prescribed minimum pressure.
10.2 Equipment
Unisilkon grease
Anti-tamper fixing
(low-/medium-/high-strength screwlock)
10.3.1 Monitor
10
9
8 2
7 3
6 5
1 2 3 4 5
19
18
6
17
16 8
15 9
14 10
13 12 11
V102
S123
P185 H54a
V100*
F76
V91
H89 V117
V24
S182 F210
F119
B A65
H151
V26 S9
P21
C V41
P22a
A61
F73
A78 V24b D P29 A
002
003
V92
H54 H95
V112
S12
S134
H66 V30
H90b
S202
CD132
H201
V188
F74
H77
P22
P97a
H94
H88
F148
F149
H68
V35
V36
H90a P21a
V189
V84
H205
P25 P23
X B2 X C1 X A7 X1
X S1 P-Dial 2
Pumpen / pumps Ventile / valves Sensoren / sensors CAN
X S2 N.C.
X 16
N.C. X R1 OCM_NTC4 X2 NTC2 X S3 N.C.
N.C.
Conc. P. Bic. P.
X 15 OCM_LF1 10 X3 NTC3 X S4 N.C.
CO1 BI1
X 14 NTC109 X5 FL-Switch X 17 X S5 Niveau Konz.
Kabelbaum / cable harness X 13 LF-Bibag 132 X6 Niveau 6 RJ-45 X S6 NTC308
X A5 X A1 X A2 X 12 Bibag NTC133 X7 LF 7 X S7 LF309
X 11 Bibag (LP1628) X8 P-Dial 1 (9) X R3 Bibag SW
X A6 X A4 X A3
X 10 ZKV PSW X9 Reedk. 10 X B1 X R2 V89 Sens.
BLD
LP 941 WTR / HPU
10.4.2 Torques
Luer-Lock: 2.3 Nm
10.5.3 Brakes
10.5.4.2 Microswitch
1 Pull off the cable from the monitor (1) and the
2 cable from the motherboard (2), disconnect
the heater rod cable (4), disconnect the EBM
3 power supply (3), pull off or unscrew the
ground cables, remove 2 screws, and pull the
power supply unit out of the housing to the
rear.
After replacement or repair, check the
voltages in the calibration menu, calibrate the
4 temperature, and check the temperature.
10.6.3 Battery
10.7 Monitor
1
Remove the screws (1) and the cover (2) and
push the monitor to the front.
Note
When replacing the P.C.B.s, the Calibration and test steps after
repair (see chapter 10.10 page 10-51) must be observed.
1 1
10.7.5 Display
10.7.5.5 Backlighting
1
Remove the plug-in boards from the monitor.
Pull off the connector (3) from P.C.B. LP 636.
Press back the detent mechanisms (1) and
remove P.C.B. LP 1131 (2).
Warning
Improper placing of the overflow tube will impede the exit air and
overflow of the hemodialysis device.
Defect in the hydraulics unit
When closing the hydraulics rear wall, always place the overflow tube
through the opening provided. Do not kink or squeeze the tubing inside.
Removal
Installation
Removal
1
2
Installation
Removal
4
– Unscrew the strain relief (1) from the cap
3 (2).
1 – Unscrew the cap (2) from the connector
2
(3).
– Open the connector's strain relief.
– Release the cable wires (4).
– Expose the cables.
Installation
1 2
Pull off the cable from V43 (12) and the level
sensor (1).
Protect the other hydraulic components from
13 leaking fluid. Pull off the tubes and collect the
fluid (see chapter 10.8.3.1 page 10-25).
12
Unscrew 2 screws (10) and remove the
11 3
multifunction block.
10 4
When installing, ensure proper tubing
arrangement.
9
After replacement or repair, run the
degassing test and the fill program.
8 7 6 5
1
2 Pull off the tubes, unscrew the screws (5), lift
3
4 the heat exchanger a little and remove it.
The heat exchanger is operated based on the
counter-current principle:
1 Inlet of chamber A
5
2 Outlet of chamber B
3 Outlet of chamber A
4 Inlet of chamber B
1 1 to valve V30
2 to inlet of heater block H66b
2 4
3 to valve V87
4 to outlet of heater block H66a
After replacement or repair, check the flows.
10.8.7 UF pump
2
3
10.8.9.1 Pump
10.8.9.2 Motor
Solenoid valve with 1 adapter, low kv value V31-38, V43, V86, V89, V100, V102, V183, V188, V130
Warning
Risk of caustic burning!
To prevent any skin contact with the disinfectant, wear protective
gloves.
When working with acidic substances:
Wear goggles!
Observe the safety precautions of the disinfectant used!
In the event of contact with acid:
Eye: Immediately flush with flowing water for 15 minutes.
Skin: Use soap under flowing water for neutralization.
Ingestion: Do not induce vomiting, but have the person affected drink
plenty of still water. Seek medical advice.
Warning
Microbiological risks!
A defect of the disinfection valve can result in disinfectant in the
container becoming diluted with fluid entering it and thus reducing the
disinfecting effect.
The disinfectant container must be replaced following repair.
1
2
1
2 Pull off the reed switch cable from the HPU.
3 Pull off the inlet tubes (6) and rinse tubes (3).
4 Take the suction rods out of the rinse
chamber, disconnect it from the tubes,
remove the screw (4), and pull the tubes
5 inwards and out of the rinse chamber adapter
(2).
6
Disconnect the tube (1) running to the
balancing chamber from the T-piece.
Unscrew the screws (5), lift the rinse chamber
a little and remove it.
When installing, ensure proper tubing
arrangement according to flow diagram.
After replacement or repair, run a rinse
section test.
1
Pull off the cable (1) from P.C.B. LP 1628,
take the cables and tubes out of the holder
(2), and open buckle (3).
2
10.8.15.2 Microswitch
Warning
Risk of infection
Contamination of the hemodialysis device with blood results in a risk of
infection.
The following must be observed to avoid an infection:
– Wear safety gloves.
– Disinfect your hands after removing the gloves.
– Wear surgical mask and goggles if there is a risk of splashes and
aerosoles.
Clean and disinfect the device before coming into contact.
Warning
Risk of infection
Contamination of the hemodialysis device with blood results in a risk of
infection.
If fluid may have passed through the hydrophobic filter, the device must
be checked for contamination after the treatment. If it is contaminated,
the device must be removed from service. Before starting the device
again, all affected parts must be replaced in accordance with the
manufacturer's specifications.
Checking the EBM front Check the housing front in the vicinity of the Luer-lock connectors for
residual blood, fluid and precipitation, especially at the following
positions:
– venous pressure transducer
– Single-Needle pressure transducer (optional)
– arterial pressure measurement unit
Checking the housing If blood or fluid enters into the housing main part or into other
main part components (e.g., P.C.B. LP 450-2, P.C.B. LP 950, BPM, etc.), these
must be cleaned or replaced.
Hydrophobic filter Check hydrophobic filters inside the device to find out whether they are
wetted with fluid.
1
2
10.9.6.3 Drive
10.9.6.4 Mechanics
1
2
3
4
Warning
A disinfection, a T1 test and a check of the electrical safety must be
performed after working on the hemodialysis device, irrespective of the
table below.
Power supply unit (power board, heater Check voltages (calibration menu), calibrate temperature, perform
board) temperature test
P.C.B. LP 630 Motherboard Check jumpers (see P.C.B. LP 630 motherboard page 11-54)
P.C.B. LP 1631 CPU 1 Set the DIP switches, install current software version, initialize
Novram, make setup settings or import data, calibrate, set time
LP 632 CPU 2; snap-hat battery Set DIP switches, import calibration data or make the necessary
calibrations
P.C.B. LP 633-5 Input board Calibrate BP rate, arterial and venous pressure, blood leak, bibag®
values (temperature and conductivity), TMP, temperature,
conductivity and OCM.
Check pressure transducer 1 and 2 (hydraulics).
Check all analog signals.
Check jumpers BR4 (see P.C.B. LP 633-5 layout page 11-58).
P.C.B. LP 634 Output board Set DIP switches, set current increase, flows, pressures (hydraulics
unit), temperature, degassing pump, blood leak, dimness
P.C.B. LP 636 External connectors Jumper J1 must not be connected (see P.C.B. LP 636 external I/O
board page 11-61), check status indicator for proper functioning
P.C.B. LP 763 Multi-interface board Check DIP switches, check BPM (option)
Display, P.C.B. LP 922; MDC-II board; Install current software in MDC-II and set the language version (if
backlighting, touch panel MDC-II board was replaced), display test, key test
P.C.B. LP 950 for heparin pump Set hex switches, calibrate Hall sensors
P.C.B. LP 1131 Status indicators Check status indicator for proper functioning
P.C.B. LP 450-2 for air detector Set jumpers, calibrate air detector sections, optical detector, venous
pressure
(see P.C.B. LP 450-2 Air detector control (LD) page 11-50).
P.C.B. LP 624 Blood pump Set hex switches, calibrate arterial pressure, check line diameter, BP
rate, stroke rate / volume (SN)
P.C.B. LP 1628 Distributor Check bibag® switching states, check BPM (option) for proper
functioning
EBM
BPM module (optional) Check leakage test, pressure test and safety valve
Heparin pump; P.C.B. LP 950; optical Set hex switches, calibrate Hall sensors
sensor
Hydraulics unit
Balancing chamber Measure balancing chamber volume in liters (only if no volume label),
read off balancing chamber volume (only with volume label),
calibrate balancing chamber volume, check flows and conductivity
Flow pump Set flows, check relief pressure, adjust current increase
Evacuation valve V188 Check ventilation after air has been sucked in
11 Functional description
MODULE
T1
TEST
BYPASS
TEST OK no STORAGE
? ERROR NUMBER
TEST
OPT. DETECTOR
TEST OK no STORAGE
? ERROR NUMBER
TEST
BLOOD SYSTEM
TEST OK no STORAGE
? ERROR NUMBER
TEST
VENOUS PRESSURE
TEST OK no STORAGE
? ERROR NUMBER
TEST
AIR DETECTOR
TEST OK no STORAGE
? ERROR NUMBER
TEST
DISPLAY
TEST OK no STORAGE
? ERROR NUMBER
TEST
ARTERIAL PRESSURE
TEST OK no STORAGE
? ERROR NUMBER
RETURN
MODULE
T1
START T1 TEST
TEST OK yes TEST OK yes TEST OK yes TEST OK yes TEST OK yes TEST OK yes
? ? ? ? ? ?
no no no no no no
TEST TEST
VENOUS PRESSURE TEMPERATURE
STORAGE STORAGE
ERROR NUMBER ERROR NUMBER
STORAGE STORAGE
FURTHER yes NEXT INCORRECT
ERROR NUMBER ERROR NUMBER INCORRECT
TEST STEPS TEST STEP
Navigation
Test Test description
X632/A26 HOT_RINSE
Power X639/
Supply A20
4008
X632/A10 H_REL_W X639/A12
X632/A9 EM_H_OFF
F X639/A17
X632/C25 V24_EN
X632/C26 V24B_EN
N
X632/B22 CI
LP 632
X632/A6 V26
6
CPU 2
X632/A6 V26
6
X632/A5 V24B
X632/A4 V24
4 X634R/A18 X634R/C22
X634L/C25
X634L/C12
X634L/A25
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
V24B
V24
V26
LP 922
Display
board V24 V24B V26
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
X632/A30 OD_OUT
X633L/C7
X633L/C8 OD_IN
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P 1631
C PU 1 Mother-
board
X351/5 X351/7
LP 922 Pven
X351/10
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
X632/C10 CLP_CTL
X632/C14 LDA2
X632/C21 BL_AL
LP 632
X632/B11 BPSB_VEN
CPU 2
X632/A11 BPSB_ART
X632/A15 SN_ART
X634L/C15
X632/B27 X632/B28
LP 633 X634L/A14 LP 634
X634L/A15
Input Output
board board
BPST_ART
X633L/ C13 /A13 / X634L/B14 /C14
/B15
X631/A20 X631/A21
X1631/20a X1631/21a
BPSST_A
LP 630 DATA BUS
L P 1631
Mother-
BPSST_V
CPU 1
board
X348/V1
X348/V3
X348/V6
X348a/1
X348a/3
X348a/6
X348a/2
X351/8
X351/6
LP 922 Pven
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Check the lower limit by verifying the venous zero point.
Test the upper limit by detuning the venous pressure unit in positive
direction
(the venous line clamp remains closed while the test is in progress).
LP 632
X632/C16 LDSA
CPU 2
X632/C18 PV_DET
X632/C17 P_VEN
X632/B27 X632/B28
LP 633 LP 634
Input Output
board board
X633L/B5 X634R/C18
VENT_V
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P 1631
CPU 1 Mother-
board
X351/1 X351/2
PVEN
LP 922
Display
board
X351/4 X351/10
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description – Test the air detector by checking the alarm state.
– Turn off the venous line clamp in the air detector module.
X632/C10 CLP_CTL
LP 632
X632/C14 LDA2
CPU 2 X632/C13 LDA1
X632/C16 LDSA
X632/C21 V145
LP 633 LP 634
Input Output
board board
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
X351/10 X351/14
PVEN
LP 922 X351/6
Display
board
X351/8
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
CPU 2
X634R/A16
LP 633 LP 634
X632/ B27 X632/ B28 Input Output
board board
+LS X634L/A13
-LS X634L/B13
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Test the arterial pressure unit by electronically detuning it in positive or
negative direction.
LP 632
CPU 2 X632/A17 PA_DET
X632/B27 X632/B28
LP 633 LP 634
Input Output
board board
X633L/B12 P_ART
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
Mother-
CPU 1
board
X348a/7 X348a/9
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
TESTBATT
LP 632
CPU 2
X634R/C23 X639/A10
LP 633 LP 634
X632/B27 X632/B28 Input Output Power
board board Supply
4008
X633L/B21 X639/A2
X639/A3
X631/A20 X631/A21 X639/A4
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
Mother- U_ACCU
CPU 1
board
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Test the blood leak detector by lowering the capacity of the transmitter
diode.
LP 632
CPU 2
X632/A25 BLL_DET
X632/B27 X632/B28
X633L/B27
LP 633 LP 634
Input X633L/B10 Output
board board
X633L/B9
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
Mother-
CPU 1
board
BLL_TEST
BLL
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Test the upper alarm limit by electronically detuning the temperature
display in positive direction.
Figure
X632/A23 T_DETADJ
X633R/C21 X633L/B16
X634R/C24
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
und Italienversion 1.00
Monokonzentrat
MON_NTC
X639/A20
NTC_BIB
LP 922
Power
Display Supply
board 4008
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Within a defined time period, the actual value of the dialysate pressure
transducer may change within certain limits only.
X632/A19 ACKN_ASP
LP 632
X632/B22 CI
CPU 2
X632/C27 UF_P_CTL
X632/A29 P_DIAL
LP 633 LP 634
Input Output X634L/B10
board board
X633R/C28 X634R/A24
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
UF_P_CTL
+P_DIAL
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Check valves V24, V24b and V26 for proper functioning (mechanically).
Test the TMP pressure unit by electronically detuning it in positive
direction.
With the dialysate flow off, apply positive pressure to the balancing
system. The actual value of the dialysate pressure transducer is now
monitored within a defined time period.
Test the pump segment of P97.
X632/C5 V43
X632/A19 ACKN_ASP
X632/C27 UF_P_CTRL
X632/C25 V24_EN
X632/C26 V24B_EN
LP 632
X632/A4 V24
CPU 2 X632/A6 V26
X634R/C22
X634R/A18
X634R/A24
X632/C20 P_DETADJ
X634L/B10
X632/A29 P_DIAL
X633L/ X633R/
X632/B27 X632/B28 B6 A18
. X634L/ X634L/A6
LP 633 A25 LP 634 X634L/A8
X634L/A5
Input Output
X634L/A7
board board
. X634L/
C25
X634L/C22
X634L/A22
X633R/
C28
X634L/
X634L/
A-C23
A-C28
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
Mother-
CPU 1
board
V43
DEGAS+P2
AIR_SEP+
AIR_SEP-
+P_DIAL
UF_P1
V31
V35
V26
V24
LP 922
V31 V33
Display
board
P97
ASP
V26 V24
V35 V37
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
X632/C7 UF_P2
X632/A7 UF_P1
LP 632
X632/C28 UF_P_EN
CPU 2
X632/C27 UF_P_CTL X634R/A24
X632/B24 UF_P2CTL
X632/A29 P_DIAL
X633L/ X634R / X 634R/
X632/B27 X632/B28 B6 C11 A22
LP 633 LP 634
Input Output
board board
X1631/20a X1631/21a
LP 630 DATA BUS
L P 1631
.
CPU 1
Mother-
board
.
UF_P2 (nur bei 4008 HDF)
UF_P1
+P_DIAL
LP 922
Display
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Test the alarm limits by electronically detuning the conductivity either by
+5 % or by –5 %.
X632/A26 HOTRINSE
LP 632
CPU 2 X632/A21 COND_DET
X632/A22 COND_SIG
7b
LP 633 LP 634
X633R/A16
Input Output
board board
X633R/
C17
X633R/C27
X631/A20 X631/A21
X1631/20a X1631/21a
LP 630 DATA BUS
L P1631
CPU 1 Mother-
board
COND_C132
COND_BIB
COND_C1
LP 922
Display (bibag-LF-Zelle)
board
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Test description Volume test of the inner capillaries and pressure holding test.
X 632/B 5 V _DSA FE
X632/C 25 V 24_E N
X 632/C 26 V 24B_EN
L P632
CPU 2 X 632/A 6 V 26
X634R/A18
X634R/C22
X 632/A 29 P_DIA L
X633L/
X 632/B27 X 632/B28 B6
X 634L /A -C 28
L P 633 L P634
X 634L /C 5
I nput Output
X 634L /C 7
board
. X 634L /C 25
board
X 634L /A 7
.
X633R/
C 28
X634L/
X634L/
A-C23
A-C30
X631/A20 X631/A21
X1631/20a X1631/21a
L P630 DATA B US
L P1631
CPU 1 M other-
board
DEGAS+P1
V 112
FLOW+P1
+P_DI AL
V36
UF_P1
V35
V26
L P922
V 32
Display
board
V 26
V 35
V 35 V 36
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
11.1.2.1 PSW (pressure switch) monitoring during free rinsing (only with devices with CDS)
START
no V104 opens
PSW 104
Pressure decrease
open
by membrane pumps
yes
no V102 opens
PSW 102
Pressure decrease
open
by membrane pumps
yes
11.1.2.2 Rinse section test (check of V91, V99, V100) (only devices with CDS)
Mandatory no V91/99/100/102
rinse time open for 900 ms
≤ 3 min V104 closed
yes
yes no
11.1.2.3 Rinse section test (check of V91, V99, V100, V130) (only devices without CDS)
PSW 134 is used for the test evaluation. The following requirements
must be met for carrying out the rinse section test:
– The test is only carried out in mandatory rinse mode and during the
last 3 minutes.
– P.C.B. LP 1631 (CPU 1): Field 2, DIP switch 7 to ON
Mandatory no
rinse time
≤ 3 min
yes
yes yes
yes yes
Pressure build-up
V91/100 open
no Rinse Failure
Pressure switch
F20
closed
Program stopped
Pressure reduction
via V99
no V188 Failure
Pressure switch
Program stopped
open
The following requirements must be met for carrying out the V39 test:
– The test is only carried out in mandatory rinse mode and during the
last minute.
– P.C.B. LP 632 (CPU 2): Field 2, DIP switch 5 to OFF
Mandatory rinse
≤ 1 min
Wait for 10 s
Measure for 5 s:
Mean pressure value
dav1
Open: V39;
Wait for 5 s
Measure for 5 s:
Mean pressure value
dav2
no
no Test failed
DAC degas. pump Pressure compensation
= 200 ? „V39 Failure“
yes
Stroke volume 10 15 20 25 30 35 40 45 50
(ml)
Switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±
7 mmHg
Warning
Do not change the position of the hex switch during operation.
Warning
The syringes classified as approved among the syringes indicated
below were successfully tested with the device.
If syringes that are classified as not approved in the table below are
used, it is the responsibility of the responsible organization to ensure
correct function of the device.
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the hemodialysis device resulting from the use of non-approved or
unsuitable consumables or accessories.
0 20 ml BD approved
3 10 ml BD not approved
4 30 ml BD not approved
5 50 ml BD not approved
B 20 ml Fresenius approved
C not assigned –
D not assigned –
E not assigned –
F not assigned –
Air entering the patient's extracorporeal blood circuit may cause an air
embolism. In order to catch limited amounts of air and to separate
accompanying air bubbles, the venous blood line is expanded (venous
drip chamber). A major task of the air detector is to monitor the filling
level in the venous drip chamber.
The protection system against air infusion uses the method of ultrasonic
transmission. Ultrasonic converters are attached on either side of the
venous bubble catcher. At periodic intervals of approx. 90 ms, a
transmitting resonator generates attenuated ultrasonic vibrations at a
natural resonance of approx. 90 kHz, which are absorbed by a receiving
resonator. The amplitude of the signal received is dependent upon the
medium between the converters. Its value is at its minimum with the
bubble catcher empty (air) and at its maximum with bubble-free fluids.
The amplitude decreases with increasing air content (foam). The signal
path is fail-safe up to and including the receiving resonator, i.e. the
failure of any component always leads to a smaller amplitude and, thus,
to an alarm. Starting at the receiving resonator, the signal voltage is
always sent onto two independent receiver paths. As soon as the signal
is too weak, one of these receiver paths causes the blood pump to stop
and the other the venous line clamp to close.
The and keys are used to both raise and lower the blood level in
the venous bubble catcher. As long as the key is pressed, the venous
line clamp closes. The deaeration valve in the air detector module
opens, and the blood level rises. The blood pump runs at reduced
speed (100 ml/min). As long as the key is pressed, the venous line
clamp remains open. The deaeration valve in the air detector module
opens, the ventilation pump is running, and the blood level sinks. The
blood pump runs at the preselected speed.
V116
S8 P185
F63
V26
A V30
F76 CD110 S182
PT4
V41 P29
V31 V32 V33 V34 CD7
P97
V92 F114
V43
H54 PT3 PT109
S5 V39 S6
PT2
H88
H68
V189
H66 H89
H66a H66b H66c H88a H88b H88c
V35 V36 V37 V38 S9
H151 C A78
F210 85
H77 S115 V84 S115
P21
A65
H94 H95 B
V99
F148
S10 S12
F149 F74
H V100 H
90a 90b H205
V91 P22
121 F119 V117 S123
V102
P23 P25
S134
PT133
S138 S137
CD132
H136
H201 S202 S204 H203
V188 V130
H94 H95
F71 F72
As soon as the water inlet valve V41 opens, the water flows through the
pressure reducing valve A61 into the water inlet chamber H66a of the
heater block H66 and across the heat exchanger H77 into the heater
rod chamber H66b.
The concentrate pump P23 admixes the concentrate to the inflowing
water per balancing chamber phase.
The vent tubing prevents pressure from building up in chambers H66b
and H66c. In the HOT RINSE mode, the developing vapor can escape
through the vent tubing.
While it is rising, the fluid is warmed up to the preset temperature by the
heater H54. The heater is controlled by the temperature sensor PT2.
From chamber H66b, the dialysate flows into the chamber H66c.
Incorporated in this chamber is the float switch S5, which controls the
solenoid valve V41, thus ensuring the correct fluid level.
The degassing pump P29 draws in the dialysate via the degassing
orifice H89. This generates a negative pressure of approx. 0.8 bar.
In the lines and the following degassing chamber H88a, the dialysate is
degassed to a level which is sufficient for hemodialysis.
Via the degassing pump P29, dialysate and released air are directed
tangentially into the primary air separator H88c, where air bubbles and
the airless dialysate are separated. The air accumulates at the top of the
chamber H88c. Then, together with the recirculation flow and via the
loading pressure valve A65 as well as the chamber H66c, the air
escapes into the atmosphere.
Chamber H88c is provided with a separating disc (standard hydraulics
only), which serves to prevent bicarbonate, if added, from being
recirculated via the heater rod chamber H66b.
At the bottom of chamber H88c, the degassed dialysate is pressed out
and into the balancing chamber H68 by means of the loading pressure.
Together with the eight solenoid valves V31 to V38, the balancing
chamber H68 constitutes the balancing system. Each of the two
sections of the balancing chamber comprises of two compartments
separated by an elastic membrane each. Hence, there are two
chambers with four spaces:
– F1 and F2: fresh solution
– A1 and A2: used solution (waste)
As soon as one of the chambers (A1 or A2) is filled with dialysate, the
solenoid valves are reversed in groups of four. The valves are reversed
by the electronic evaluation of the current rise pulse of the drive motor
of the flow pump P21; this pulse is created upon membrane abutment.
Within the filling phase, F1 or F2 is filled with fresh dialysate by means
of the loading pressure. In order to obtain a continuous flow, a second
chamber is switched parallel to the first chamber. The second chamber
is operated at an inverse sequence.
Each time the chamber is changed over (maximum deflection of the
membrane), all valves are closed for approx. 100 ms (dead time).
From the balancing chamber H68, the dialysate flows through the
conductivity cell CD7 with integrated temperature sensor PT3. The
measured conductivity values are indicated on the monitor in ms/cm,
related to 25 °C.
The temperature sensor PT3 has the following functions:
– Temperature compensation of the conductivity display
– Indication of the dialysate temperature
Should the actual values (temperature or conductivity) of the dialysate
exceed or fall below the limit settings, the bypass valve V26 opens, and
the dialyzer inlet valve V24 is closed. The device is now in the bypass
mode. The dialysate is discharged into the drain not via the dialyzer, but
via the secondary air separator H88b and the balancing chamber H68.
If the actual conductivity and temperature values of the dialysate are
within the set limits, the dialyzer inlet valve V24 opens. The bypass
valve V26 is closed. The dialysate flows to the dialyzer.
After the dialyzer, the dialysate which is now loaded with the urinary
excreted substances flows via the filter F73, the dialyzer outlet
valve V24b and the blood leak detector S8 into the secondary air
separator H88b. The secondary air separator H88b comprises of the
pressure transducer S9 and the level sensor S6.
A warning will be emitted by the blood leak detector S8 in case of blood
losses of 0.5 ml per minute, for a hematocrit of 0.25.
Together with the venous return pressure, the signal of the pressure
transducer S9 is evaluated and indicated on the monitor as TMP. The
fluid level in the secondary air separator H88b is monitored by the level
sensor S6. The secondary air separator H88b ensures that no air can
enter the balancing chamber H68. Any presence of air bubbles in the
balancing chamber H68 would cause balancing errors.
The dialysate is pressed through the flow pump P21 into the balancing
chamber H68. As mentioned above, the balancing chamber valves are
reversed by the current rise pulses of the drive motor of the flow
pump P21. Using the speed of the flow pump P21, the dialysate flow
can be adjusted in the dialysis program: 300, 500 and 800 ml/min. In the
cleaning programs, the flow of the dialysate is fixed.
The relief valve A78 is used to limit the pressure of the flow pump P21
before the balancing chamber to approx. 2 bar.
After the balancing chamber H68, the dialysate flows though the outlet
valve V30, the heat exchanger H77 and the drain valve V87 into the
drain.
The recirculation valve V86 and the drain valve V87 serve to recirculate
fluid during the hot rinsing and disinfection programs.
Standard program
1. cycle:
V30
H68
F1 A1 F2 A2
P21
2. cycle:
V30
H68
F1 A1 F2 A2
P21
This system ensures that equal amounts of fluid enter and exit the
dialyzer. This results in an exact balancing of the dialysate and, in
conjunction with the UF pump P22, a controlled volumetric
ultrafiltration.
As soon as the fluid level in the secondary air separator H88b has
dropped below the level sensor S6, this sensor activates the air
separation pump P97. Should the fluid level not have reached the level
sensor S6 within a given time period, the FILL PROGRAM is started.
Note
In order to detect the fluid level, the level sensor S6 requires a fluid with
a certain minimum conductivity, which is definitely achieved in all
dialysis programs. Separation of air is only required for the dialysis
programs. In all other programs, the air separation pump P97 and the
fill valve V43 are force-actuated.
1. cycle:
V30
V43
H68
F1 A1 F2 A2
S6
V35 V36 V37 V38
S6
H88b
P21
S9
V84
2. cycle:
V30
V43
H68
F1 A1 F2 A2
S6
V35 V36 V37 V38
S6
H88b
P21
S9
V84
If not enough air was separated and the fluid level is still below the level
sensor S6, the FILL PROGRAM is activated. The fill valve V43 opens.
The air can escape into the drain.
1. cycle:
Chamber F1 is filled. This forces the fluid from chamber A1 into
chamber A2. The fluid is then forced by chamber A2 via the dialyzer and
into the secondary air separator H88b.
2. cycle:
Chamber F2 is filled. This forces the fluid from chamber A2 into
chamber A1. The fluid is then forced by chamber A1 via the dialyzer and
into the secondary air separator H88b.
Note
Repeated activation of the fill program during treatment indicates a
defect (leakages).
End of dialysis
Dialysis
PGM Program
R Rinse
R endl. Rinse endless
F Free rinsing
HR Hot rinse
C Cooling rinse
D Disinfection
D(F) Cleaning
Disinfectant is sucked in from the front
(concentrate suction tube)!
HDIS Hot disinfection
M Mandatory rinse
IHR Integrated hot rinse
The rinse chamber is emptied at the end of the set program for approx.
1 min.
The time specifications are based on the factory setting. Shorter or
longer program times can be set at any time with the SETUP MENU
(see chapter 5.1.3 page 5-2).
Rinse
PGM 1: –R–
T/°C
37
33
t/min
Rinse chamber
Rinse evacuation
10 strokes each
Start End
T/°C
37
33
t/min
Rinse endless
Start End
(abortion of program)
Hot rinse
PGM 1: –F–HR–C–
T/°C
80
37
34,5
t/min
Heating Hot rinsing Cooling rinse Rinse chamber
Timing from 80 °C down to 34.5 °C evacuation
Rinsing clean 10 strokes each
PGM 2: –F–HR–
T/°C
80
37
t/min
Heating Hot rinsing Rinse chamber
Timing from 80 °C evacuation
Rinsing clean 10 strokes each
PGM 3: –IHR–
T/°C
80
t/min
Rinse chamber
Integrated hot rinsing evacuation
10 strokes each
Start End
PGM 3: –IHR–C–
T/°C
approx.
80
approx.
35
t/min
Rinse chamber
Heating Integrated hot rinsing Cooling rinse evacuation
10 strokes each
Disinfection (cleaning)
PGM 1: –F–D–M–
T/°C
37
33
t/min
Rinse chamber
Rinsing clean Prep.* Disinfection Mandatory Rinse evacuation
10 strokes each
3–10 min approx. 10–20 min (Setup) 15–30 min (Setup) approx. 1 min
(Setup) 1 min
Start End
PGM 2: –F–HDIS–M–
T/°C
80
37
34,5
t/min
Heating Rinse chamber
Hot disinfection Mandatory Rinse evacuation
Rinsing clean Prep.* 10 strokes each
approx. 6 min 10–20 min (Setup) 15–30 min (Setup) approx. 1 min
3–10 min approx. approx.
(Setup) 1 min 4 min
Start End
PGM 3: –F–D–M–HR–
T/°C
80
37
3–10 min approx. 10–20 min approx. 6 min 15–30 min (Setup) approx. 1 min
(Setup) 1 min (Setup)
Start 15–30 min approx. End
(Setup) 4 min
Prep.*: preparation phase:
Heater off.
Set the level of the float switch chamber below the lower switching point of the float switch
by 1 balancing chamber changeover and 4 UF pump strokes.
Aspiration of disinfectant for 50 UF-pump strokes.
PGM 4: –F–HDIS–M–HR–
T/°C
80
37
t/min
Heating Hot Heating Rinse chamber
disinfection Hot rinsing evacuation
Rinsing clean Prep.* Mandatory Rinse 10 strokes each
approx. 6 min 10–20 min approx. 6 min 15–30 min (Setup) approx. 1 min
(Setup)
3–10 min approx. approx. 15–30 min approx.
(Setup) 1 min 4 min (Setup) 4 min
Start End
PGM 5: –F–D(F)–M–
T/°C
37
33
t/min
Rinse chamber
Rinsing clean Prep.* Disinfection Mandatory Rinse evacuation
10 strokes each
3–10 min approx. 10–20 min (Setup) 15–30 min (Setup) approx. 1 min
(Setup) 1 min
Start End
6/06.12
interface Supply Akku
+24V_SW 4008 Battery
11-43
Chapter 11: Functional description
RAM
Speicher
RS232 CPU
RS232
11.2.3 Monitor block diagram
RAM
RS232 CPU
SM-EN
Motherboard LP 630
U-ACCU
24V_US
6/06.12
Adress-/Databus
MDC-II-board TFT-Display
Display board
LP 922 ROM
RS232
Video
RAM DRAM
11-45
Chapter 11: Functional description
BPM (Option)
LP 1135
LP 1627 Anzeige / Display
X_BP_ART X_BP_SN X_HEP X_LD
X_BPM_MON XS
LP 1131
Ampel / Kurzschluß- LP 1628
Trafic X 189 X 189 ST 4 X 154 Verteiler / X__21/47
teil /
Diagnose / LP 624 LP 624 LP 950 LP 450-2 Shunt Distribution
X1 Diagnosis BP Art BP SN Heparinpumpe / Luftdetektor / interlock
X_BPM
EBM X 348 X 348 ST 1 X 351 X_CAN_MON X_WTR X_CAN_WTR
X 632
X CAN
X ACCU
Akku /
Accu
LP 633 LP 634
X PWR
Input Output
SM-EN
board board
Netzteil
4008 /
X RS
Power
Supply LP 763
4008 LP 1631 Multi
CPU 1 Interface
6/06.12
X 634R
X 63Z
X 639A
X 633R
X 631
Board
11.2.5 HPU wiring diagram (hydraulic processing unit)
X 635
LP 630
Motherboard
LP 922
X B2 X C1 X A7 X1
Display X S1 P-Dial 2
board Pumpen Ventile Sensoren CAN
X S2 NC
N.C.
X R1 OCM_NTC4 X2 NTC2 X 16 X S3 NC
Conc. P. Bic. P.
X 15 OCM_LF110 X3 NTC3 X 17 X S4 NC
CO1 BI1
X 14 NTC109 X5 FL-Switch X S5 Niveau Konz.
X 13 LF-Bibag 132 X6 Niveau 6 X S6 NTC308
X A5 X A1 X A2
X 12 Bibag NTC133 X7 LF 7 X S7 LF309
X A6 X A4 X A3 X 11 Bibag X8 P-Dial 1 (9) X R3 Bibag SW
X 10 ZKV PSW X9 Reedk. 10 X B1 X R2 V89 Sens.
LP 941
Blutleckdetektor /
Wasserteilrechner / Hydraulic Processing Unit
Blood Leak Detector
Chapter 11: Functional description
11-47
Chapter 11: Functional description
LP 1631
US-Sender LP 450-2
US-Transmitter
X156 20 +24V
19 +24V
18 0 (24V)
X155 0 (24V)
17
16 +12V
US-Empfänger 0 (12V)
15
US-Receiver
14 LDA1 = L
1 13 Niveau senken / Level down = H
2 X154 Dialyse Start / Dialysis start = H
3 12
11 n.c.
Niveau heben/senken
X351 LDSA = H
Level up/down 10
9 Potenzialausgleich / Ground
1
2 8 Klemme u lamp closed = L
LP 248 X160 ODSA = H
3 7
4 6 LDA2 = H
Potenzialausgleich
Ground
13.05.09
11-54
8 8
2 2 2 2 2 X350 2 X351 2
1
XDIAG 1
X637A 1
X348A 1
X348V 1 1 1
XSHUNT
C8
C9
1
2
C10
C11
C6
C7
J5
C4
C5
C15
C17
7 C18 C13 C16 C14 7
C12
FU6
FU5
FU4
32
32
32
C
C
C
B
A
B
A
B
A
P
P
2
1
XACDA 1
X637B
O
O
O
1 2
Fig.: P.C.B. LP 630 layout
X636 2
N
N
N
3 4 7 8 J3
J2
M
1 2 5 6
L
L
1 Codierbrücken:
XHDF 2
K
K
6 Standard (4008): J2=7-8 J3=1-2 6
J
J
C19
4008-ADS: J2=5-6 J3=2-3
I
I
1
H
H
Chapter 11: Functional description
X632
X633L
X634L
G
G
1
C20
X637C 2
XACCU
F
XPWR
E
E
B5 A5
D
D
C21 B4 A4
Anschlußkabel
C
C
C
C24 2
C22 B2 A2 1 XCAN
B
eingelötet
B
B
5 C25 5
C23 B1 A1
A
A
C26
11.2.10 P.C.B. LP 630 motherboard
R3
R2
R1
1
1
1
FU8
DR1
FU9
SCHUTZVERMERK NACH
DIN 34 BEACHTEN.
A
A
A
C
C
C
32
C
2
32
32
32
B
A
P
XRS
P
1
C3
C2
C1
O
J1
N
2
M
M
J4
1
2
4 1 4
L
L
L
Jumper J1,4,5 offen/open
K
K
K
K
J
J
J
J
I
I
I
R4
H
H
H
X63Z
X633R
X634R
X631
G
G
G
X639A
G
1
F
F
F
F
XB1 2
E
E
E
D
D
3 3
C
B
A
1
1
1
1
2
FU2
FU3
FU7
X635
FU1
1
2 2
DATUM NAME
BLATT
BE 22.10.03 OTT
1 e neues LT-O GE 22.10.03 Schg MOTHER-BOARD 1 /1 1
d XB1;XC2;XU1neu/LT-N 02.10.09 Schg LAYOUT LP630O
c 9006 unter FU/1/2/4/5/9 21.09.09 Schg
b neues LT-M 03.02.09 Schg
SM-EN
6/06.12
Crystal-controlled Adress 8 KByte
Reference U-REF 10V oscillator decoder NOVRAM U-REF
voltage 11 MHz 128 KByte 10 V
5 V / 10 V EPROM
U-REF 5V 2 KByte
SRAM
ANALOG
Fig.: P.C.B. LP 632 block diagram
11.2.11 P.C.B. LP 632 CPU 2
80C535
Micro-
Control bus
controller 8-
16-
DIGITAL DIGITAL
inputs outputs
ADRESSES
SM-EN
2x8 8-
DIP Data bus POWER
SWITCH outputs
DATA
(BUFFER)
6/06.12
Tx
serial interface
24- Rx 8x
DIGITAL Status
inputs LED's
(BUFFER)
8-
DIGITAL
outputs
PORT
(DIRECT)
Control bus
Data bus
Chapter 11: Functional description
11-55
Chapter 11: Functional description
analog digital
BUFFER 6
BUFFER 0
P_V102 CSS_REED FL_SWITCH_+5V
5V/12V
SP ARE CI
P_VEN SP ARE ABG_BYP
SP ARE ABG_ON
BPR_VEN SP ARE MF ABG_ALARM
ADC 0
BPR_AR T
V43
VEN_BPR_SET V26
V24B
BUFFER 7
BUFFER 1
AR T_BPR_SET V24
LD_SERIAL
UF_P1
Uref LDA2
SUB_W_P
SP ARE
U_ACCU REED_BIC
BIBAG
P_V104 REED_RINSE
BUFFER 2
BIBAG_C
24V_EM PSW_V102
PSW_V104
P_AR T PWR_OFF
ADC 1
HEP_ON
BLL_DIM
SP ARE
BLL/BLL_OLD SW_ON_OFF
PWR_F AIL
COND_SIGNAL BUFFER 3 SHUNT_OUTP
SHUNT_INP
Uref SHUNT
SER V_EN
LEV_SIGNAL
SP ARE SN
ADKS
TEMP_DIAL2 BPSB_AR T
BUFFER 4
BPUS_AR T
COND_SIGNAL2 BPSB_VEN
BPUS_VEN
BIBAG_TEMP HEP_ALARM
ADC 2
BIB_LEVEL
BIBAG_COND
EXT_ALARM
SP ARE SER VICE_MODE
LEVEL UP
BUFFER 5
Microprocessor Interface
DATEN–BUS
STEUER–BUS
ADR–BUS
CPU 1
18.10.10
Latch +AIRSEP_P
UF_P_CTRL UF_P1
UF_P_EN
UF_P2_CTRL UF_P2
EN2
V31 V35
V32 V36
Latch GAL
V33 V37
V34 V38
EN3
Latch CONC_P
Micro-
controller
EN4
Latch BIC_P
&
EN_CPU2
EN6 V24
& V24 V 1
V24B V24b V102
Steuerbus / Latch
Adress- V26 V126
Controlbus V87 V130
decodier-
Adressbus / ung /
Adressbus Adress-
decoding EN7
V30 V86
Latch V41 V
V43 V104
Datenbus / Bus V84
Data bus treiber /
Bus
driver
EN8
SN_EN HDF_LOG1
Latch SNST PROG_LOG2
CPU_OFF VENT_V AL VE
V_ADS
EN9
CLR_ALARM CPU_A U
HOT_RINSE HEA T_OFF
Latch
TEST_BA TT STAFF_CALL
EN10
BPST_ART BPSST_VEN
BPSST_ART HDF_SEL
Latch
CLAMP_CTRL BPST_VEN
Latch GAL
Micro-
controller
Oszillator /
Oscillator
EN1 1
Zähler /
Latch FLOW_P
Counter
D/ A DEGAS_P
TEMP_SET DAC_BLL
DAC_DIM DAC_
TEMP_AD DAC_P_ZE
U_ref STEUER_EP BIBAG_TEMP_A
STEUER_FP
EN12
V_Z V1
V_Z V2
Latch
V
FREE_OUTP
9006
R114 BR1
F MP8 MP7 MP6 MP5 MP4 MP3 MP2 MP1 1 F
C66 C65
IC1
IC2
IC3
IC5
P1
R17
R19
DIP-Switch S1
IC16
R14 R15
R21
C1 C2 C3 R3 C5 R4
O
N
R23 1 2 3 4
S1
1
2
3
4
SO1
IC24
R150
R60
OFF OFF OFF OFF I²C-SW-MODE
IC15
C76
Fig.: P.C.B. LP 634 layout
R27
C17
C24
ON OFF OFF OFF AKUT / ACUTE
NTC1
D11
Q1
C19
C22
R38
9001 OFF ON OFF OFF STANDARD 4008
R12
C14
C29
C58
R11 R149
C9
R82 R65 ON ON OFF OFF ART-AS ab SW 1.10
IC9
E R25 R44 R62 E
R125
IC25
R46 R61 R127 C77
C72
C67
R52 C31
C68
C47
C45
R51
Chapter 11: Functional description
R126
C71 C15 IC32 R33 IC36 C25
IC22
+
C38
IC26
R42R49
R111
C32
D5 R50
C70
D51 SO3
R59
IC18
R54 R75 D50 C20 D13
C18
R47
IC42
IC40
R143R147R146 R145
C51
C44
IC30 C40 C37
R144
R53
C59
C43
C64 D12 R120 R45 R39
R40
C99
R119 R121 R16 R24R22 C35
C21
9002
R20
R55
R136
C95
IC8 IC27
R91
R37
R58
1
R31
Q2
9004
IC31
IC35
DR5
R115 R28 R35 R137
D36 R18
IC51
IC39
C23
C98
IC12
R117
R128
R116
9005
C53
R13
R10
C46 R99 C39 D6 R56
+
SO2 R34
9011 C34 C63 C10 R2 R131 C73 R36 C12
IC17
IC23
C16
C69
D D
C55 C13
IC4
IC10
IC54
IC14
R43 R48
IC29
C57
IC55
C54
C97
IC38 C8 R7 C4 R1
C36
C41
R66
R64
D32
R69
R72
R71
C42
R107
DR1 DR4 R68 R67 R63
IC7
R41
IC13
R140
IC34
1
IC41
R102 C33 R139
C11
IC50 R138
C48
D49
D48
R94
R93
D31
R26
+
R9
C7 R108 C74 C75
R6
T1
R80
R86
R89
R100 R97 R90 C49 R118 R8 C96 C62
R92
R74
+ C60 D34
+ C61 D35
IC6
IC11
IC52
IC21
D19
C C
T3
C94
9002
R164
IC46
R161
C6
R130
DR3
R104
R101
9003
R5
R167
R172
R113 R112
R178
R175
R98
R190
R191
R189
R187
R188
R186
R184
R185
R124
R155
R77
R76
D52
R174
R153
T4
IC20
R129
IC43
IC44
R163
R162
T16 T15 T13 T12 T6 T5
D4
D7
T9
24
C86
R177
13
R169
D41
D40
D42
D37
D29
9012 25 12
R135
C93 Q3
R105
R106
R109
D45 D28 D10 D2 D15 D22 IC45 R179 C92 IC33
C87
D16
1 C89 R170 36 1
R132
C84
C90
R166
R165
D21
48
37
T8
C85
IC28
SCHUTZVERMERK NACH
DIN 34 BEACHTEN.
T17 T18 T14 T7 T2 T10 T11 D17 C81
D9
D8
IC37 R160
T19
D24
D30
D43
D27
D33
D23
R154 R168
R79
IC19
D25
D26
R83 R81
R73
R32
C27
C26
R30
R29
R95
C79 C82
C80
C83
+ C78 R158
R70 C28 R134 R133
C91
C88 R159
D20
D18
R110 R78
C50
R181
R180
C C
R183
R182
B B
32 B 1 32 1
A A
X634L X634R
SM-EN
h IC43,44,46 neu LT-P LP634 M46141
ÄNDERUNG DATUM NAME Fresenius Medical Care A3
1 2 3 4 5 6 7 8
6/06.12
1 2 3 4 5 6 7 8
F F
ST3
ST2 S1
ST1
ST4 ST5
SM-EN
R3
C29
C32
C31
RL1 RL2 T1 2
IC1 X1131
BR1
R4
D1
R1 1
R2
1 J1
C1
6/06.12
D4 D7
C21
C22
IC2
C6
C5
C7
C8
C10
C11
C18
C17
C16
C15
C14
T2 T3
P1
D3
D2
C27
C26
D6 D5 D8
1
X636 2 1 1 ST6 1
SCHUTZVERMERK NACH
DIN 34 BEACHTEN.
ST7 ST8
B B
11-61
Chapter 11: Functional description
05.12.08
12.Nov.07
F F
E E
Chapter 11: Functional description
T11
R59 2 ST5 1 C37
TD7
1
T2 2 C4 DR13 R36
1 BR2
C16
ST2
R28
R1
D3
D7
D8
1
R49
R52
T1
R48
R44
R5
R17 R31 R6
R43
C15
ZD2
R12
R46
DR12
C17 C18
C9
2
1
DR23 R16 R15 R47
C26
C32 R34
DR22 R18 R2 C25
R19
R53
R20
D1 DR21
C1
R30 DR2 DR19
R4
C8
C19
R29 T4 T6 T8 T10
R8
DR18
R11
R21
R23
IC5 C10
R9
Q1
C3
DR17
R33
R32
C47
C2
C21 DR20 IC1 IC8 IC13
R63
C41
R45
R50
C48 DR30
IC12
R14
T3 T5 T7 T9
IC14
DR31 IC2
C31 C28
D2 C33
C5
ST4
D11
C49
IC17
R42
DR14
C7
R62 R41
R7
D DR3 D
R13
R10
R22
IC4
DR25 C6 C12 C13 C14
C20 DR28 DR29
C22
DR8
DR9
DR1 DR27
C11
C27 2 DR24 T12
C24 ST1 DR26
1 DR33
DR32 SW1 IC11
1
IC3
R27
R58
C29
1
R3
C34
1
Q2
C23
R24
11.2.19 P.C.B. LP 950 Control board (HEP)
ST6
DR10
C30
DR11
DR34
DR15
DR16
DR6
DR5
DR7
DR4
C45 C43 C46 C42
M47254 4008E/B/H/S/ARrTplus/Art/Prometheus
M46988 4008S V10
M46169 MFT
SM-EN
1 2 3 4 5 6 7 8
6/06.12
11-67 6/06.12 SM-EN 4008 S (Version V10) Fresenius Medical Care
COPYING OF THIS DOCUMENT, AND GIVING IT TO OTHERS AND THE
USE OR COMMUNICATION OF THE CONTENTS THEREOF ARE FORBIDDEN
SCHUTZVERMERK NACH WITHOUT EXPRESS AUTHORITY. OFFENDERS ARE LIABLE TO THE PAYMENT
DIN 34 BEACHTEN. OF DAMAGES. ALL RIGHTS ARE RESERVED IN THE EVENT OF THE GRAND
OF A PATENT OR THE REGISTRATION OF A UTILITY MODEL OR DESIGN.
F
A
B
E
C
D
1
1
2
2
D18
D6
D30
D31 D19 D7
3
3
D32 D20 D8
D33 D21 D9
ÄNDERUNG
R1
R2
4
4
DATUM
R8
R9
D11
D23
D35
NAME
BE
GE
R13
R12
5
DATUM
LAYOUT
14.10.08
03.09.08
Bottomside not fitted.
Unterseite nicht bestückt.
Ott
NAME
D37 D25 D13
R5
R3
LP1131E
D38
D27
6
6
D39 D15
R11
R10
D16
D40
D29
D41
D17
LP1131 SMD
1
R6
R7
X1
D28
7
7
Ampel
Traffic Light
M40876
8
8
/
A4
BP
BLATT
F
A
B
E
C
D
Fig.: P.C.B. LP 1131 layout
11.2.20 P.C.B. LP 1131 Status indicators
Chapter 11: Functional description
6/06.12 SM-EN 4008 S (Version V10) Fresenius Medical Care 11-68
COPYING OF THIS DOCUMENT, AND GIVING IT TO OTHERS AND THE
USE OR COMMUNICATION OF THE CONTENTS THEREOF ARE FORBIDDEN
SCHUTZVERMERK NACH WITHOUT EXPRESS AUTHORITY. OFFENDERS ARE LIABLE TO THE PAYMENT
DIN 34 BEACHTEN. OF DAMAGES. ALL RIGHTS ARE RESERVED IN THE EVENT OF THE GRAND
OF A PATENT OR THE REGISTRATION OF A UTILITY MODEL OR DESIGN.
F
A
B
E
C
D
1
1
2
2
C34
29
21
C8 R26 C50
30
20
C3 R25
IC17
1
IC4
C1
R111
C6
4
14
C19
5
13
3
3
C33
IC7
Q2
IC1
C48
C49
R32
C5
C51 C25
60 41
11 1
f neues LT-C
b neues LT-A
e neues LT-B
61 40
ÄNDERUNG
12 44
C7
IC16
C24
IC5
IC3
c V1,V2;P1 geändert
a C44,C45 geändert.
22 34
C22
R112
4
4
23 33
80 21
Q1
1 20 C4
R101 R45
C10
R43 R31
1
DATUM
C23 C9 D11 X3
C55 2
кмͨймͨкл
R3
R30
06.04.06 Wa
23.04.09 Rei
27.07.09 Rei
d Leiterb.verstärkt/C64 neu11.12.09 GAn
19.07.10 GAn
R14 C21
NAME
R21 IC18
D6
R8 R20 R2 T2
T3 C56
R39
GE R40 IC2
R13
R42
R9 R38 R37
R27
T1
IC12 R23
C14
C41
R41
F4
R48
C26
C47
IC19
D4
C27 R7
P1
5
5
C43
R10
DATUM
LAYOUT
C58 IC15
1
C17
T6
C28
C40
R49
L3
C31
R6 R5 R35
T7
C15
IC13 R28
26.11.04 Wa
D7
R19
R15
R33
NAME
D3 R47
C59
C44
IC9 D2
26.11.04 Gerhard
C64
IC11
C53
LP1147C
R36
V2
R16
C32 R17
1
C39
T5
L4
C37
C38
C42
IC6
C46
D8
L2
IC14
C13
V1
R18
C30
C16
1
6
6
IC10 IC8
1 D1 D5
R46
L1
C45
C62
C54
C63
C60
F3
F1
C36 1 1
X2 X1 2
C61
LP1147
7
7
BPM 5008
MAINBOARD
M31640
8
8
A4
BP
BLATT
F
A
B
E
C
D
Fig.: P.C.B. LP 1147 layout
11.2.21 P.C.B. LP 1147 BPM (optional)
Chapter 11: Functional description
11-69 6/06.12 SM-EN 4008 S (Version V10) Fresenius Medical Care
COPYING OF THIS DOCUMENT, AND GIVING IT TO OTHERS AND THE
USE OR COMMUNICATION OF THE CONTENTS THEREOF ARE FORBIDDEN
WITHOUT EXPRESS AUTHORITY. OFFENDERS ARE LIABLE TO THE PAYMENT
SCHUTZVERMERK NACH OF DAMAGES. ALL RIGHTS ARE RESERVED IN THE EVENT OF THE GRAND
DIN 34 BEACHTEN. OF A PATENT OR THE REGISTRATION OF A UTILITY MODEL OR DESIGN.
F
A
B
E
C
D
1
1
R12
T4
T1
R9
2
2
LD7
T2
R10
T3
R1
R5
R7
R2
R6
R3
R8
R4
R11
3
3
DP9
DP8
1
2
DP7
R28
T8
LD6
LD5
T5
X_BP_ART
R25
T6
4
4
R26
T7
R17
R21
R23
R18
R22
R19
R24
R20
R27
R13
R15
R14
R16
5
5
DP6
ÄNDERUNG
DP5
DP4
R34
R40
R36
DATUM
R42
R35
R41
LD3
LD4
R43
R37
NAME
R44
R38
R45
BE
GE
R39
R47
R48
R46
6
6
DATUM
LAYOUT
20.08.08
Ott
NAME
1 X_LD
LP1627
X_BP_SN
7
7
DP3
LP1627
DP2
DP1
R29
R31
R32
R30
Display
2 2
Anzeige
LD1
LD2
1 1
X1 X2
M40292
8
8
/
A3
BP
BLATT
F
A
B
E
C
D
Fig.: P.C.B. LP 1627 layout
11.2.22 P.C.B. LP 1627 Display , EBM SN (optional)
Chapter 11: Functional description
Chapter 11: Functional description
A B C D
5 5
WITHOUT EXPRESS AUTHORITY. OFFENDERS ARE LIABLE TO THE PAYMENT
OF DAMAGES. ALL RIGHTS ARE RESERVED IN THE EVENT OF THE GRAND
USE OR COMMUNICATION OF THE CONTENTS THEREOF ARE FORBIDDEN
4 4
2 +24V_EM 2
X_CAN_MON
1 1
PWR_GND X_BPM_MON
R1 R2 MP2 C2
DR1 R14
2
1
R15
R9
MP1 R16
R8
X_CAN_WTR
R11
T2
R12
2 R13
R3
T1
1 R7
R6
X_CAN2
C10
IC4
MP3 MP4
C9
3 C8 3
C3
1
2
C7
C5
IC3
C4
2 C6
X_BPM
X_CAN3
C1
IC2 IC1
SCHUTZVERMERK NACH
1 1 1
X_WTR X_S21/47 XS
DIN 34 BEACHTEN.
2 2
1 1
A B C D
DATUM NAME
BE 24.06.08 Ott Verteilerplatine BLATT
/
GE 05.12.08 Scheuring
LAYOUT LP1628
Distributionboard
b C3-C6 war 1µ0/16V
a C5,C6 gedreht
04.05.09
05.12.08 Schg LP1628 M40286 BP
ÄNDERUNG DATUM NAME Fresenius Medical Care A4
A
B
E
C
D
1
1
R12
T4
T1
R9
2
2
T2
R10
T3
R1
R5
R7
R2
R6
R3
R8
R4
R11
3
3
LD7
1
2
X_BP_ART
4
4
R34
R40
DP9
R36
R42
R35
R41
R43
DP8
R37
R44
R38
R45
R39
DP7
R47
R48
R46
LD6
LD5
5
5
ÄNDERUNG
DATUM
NAME
BE
GE
6
6
DATUM
LAYOUT
23.04.08
Ott
NAME
1 X_LD
LP1629
7
DP3
LP1629
DP2
R13
R14
R16
DP1
R15
2 2
LD1
LD2
1 1
X1 X2
Display EBM
Anzeige EBM
M40289
8
8
/
A3
BP
BLATT
F
A
B
E
C
D
Fig.: P.C.B. LP 1629 layout
11.2.24 P.C.B. LP 1629 EBM display
Chapter 11: Functional description
Chapter 11: Functional description
2 MByte 32 kByte
ROM NOVRAM
Digital
I/O
Digital
I/O
DIP switch
array 2
Digital
I/O
SD card Digital
I/O
DIP switch
array 3
LEDs
d
f
c
b
e
1
2
3
4
5
ÄNDERUNG
IC3 hinzu LT-C
SPI hinzu LT-B
DATUM
08.12.09
23.03.10
J1,R160-162 hinzu LT-D28.05.10
X1/USB Pin's geänd. 28.05.10
28.01.11
R101 D1
C29
R102 D2
BAH1
A
32 B
C31
C
DR3
C20-1
NAME
Schg
Schg
Schg
Schg
Schg
IC17 R103 D3
IC11
R51 R53
R45
R84
DR6
R104 D4
R52 R54
BE
R83
GE
X_LAN
C17
MP14
C37-1
C30 R76
C81
D5
R30
IC8
R105
IC3
C80
C54
R31
R159
D9
C21 C23 C48 R77
C55
R46 R106 D6
R55
8
IC19
R57
R107 D7
C1
IC10
IC12
DATUM
R143
R138
LAYOUT
DR2 DR1 C8 R108 D8
IC7
14.06.09
14.04.07
R142
C12
R35 R49 IC16
T1
C53
B
B
R28 R56
R141
C51C52
C15 R60
13
Ott
C34
C2 + C14 R61 C78
R62
NAME
C16
C38
R58
R78
IC9
C3 R63
IC13IC15
IC2
R154
C13 R135 R79
C76
9 1 2
LP1631E
C70
C4 + DR4
SM-EN
C69 +
Scheuring
C5 C68 C71
R140
C61
C77
R1 DR7 R126 C66 C64
24
R139
13
R150
R129
R128
R2
DR12 DR13 1
R127
R156
C19 C57 25 12
C67 C75
R3 C60 R124 R130 R151 R81
R4
Q2
R162 R125 R149
14
C73
R82
IC21
C6 C58 R155
C72
R147
IC22
R122
IC1
3 4 5 6 7 8 11
48
37
10
C9
R136
R137
R123 R152
12
R5 R133 R132
IC4 C32
R158
R85
6/06.12
C35 O
208
157
1
C39
C42
156 R86 N
1
C50 R67 2
R33
R43
R25
R41
R87 3
C62
R20 R88 4
R131
5
S1
R157 R89
R116
R7 6
1
2
R148 C43
R90
R32
R34
R42
R44
R18 R26 7
R9 R91 8
C
C
R11 R17 R92
C10 R69 O
R163
R93 1
LP1631
1
R12 R14 IC5 R94 2 N
R70 3
R95
IC14
R71 4
R96
J1
5
S2
C44
R29
C27
R8 R6 R27 C56 R97 6
R72 R98 7
4
R15
C22
R50
C24 C25 C26 C28
R16 R99 8
X_JTAG
Q1
R19 2 N
R21 R110 3
105
R36 R111
52
4
DR11
S3
CPU-1
53
R74 R112 5
CPU-1
104
C33
1
1 6
DR9
C36 C40 C41 R75 R113 7
R114 8
C65 X_USB
R118
R119
R115
C63
D13
R121
R144
DR10
R66 R117R120
M39459
D
A4
BP
BLATT
1
2
3
4
5
Chapter 11: Functional description
11-73
Chapter 11: Functional description
12 Appendix
Chapter without content