FROG SERIES WEIGHING INDICATOR
USER’S MANUAL
(LED, LCD BATTERY VRSION)
第二页
Please read the manual carefully before operation!
第三页 目录
第四页 型号定义
Model is defined as following:
Frog □ □ □ 0 - □ □ □
Bit 7, panel light type, “1” for normal; “2” for bright
Bit 3, battery type, “1” for battery “2” for Linth “3” for dry
Bit 2, display, “E” for LED display; “C” for LCD display
Note: When Bit 2 is “E” with LED display, value defined as following:
1 for white; 2 for green; 3 for yellow green
4 for yellow; 5 for amber; 6 for red
7 for blue; 8 for red-green-yellow; 9 for red blue ribbon
When Bit 2 is “C” with LCD display, color for backlight is defined as following:
1 for white; 2 for green; 3 for yellow green
4 for yellow; 5 for amber; 6 for red
7 for blue;
8 for seven color(red,green,yellow,blue,ribbon,blue,white (selectable)
Example 1: Model FrogE120-117 stands for Frog series, Led display, with rechargeable
battery, water proof, normal precision, with normal panel brightness, display color blue
Example 2: Model FrogC120-118 stands for Frog series, Lcd display, with rechargeable
battery, water proof, normal precision, with normal panel brightness, backlight color is seven-
color
第五页 主要参数
Chapter One Specification
1, Verified counts N=3000;
2, Sampling speed: 10 times/sec
3, Load cell sensitivity range: 1.5~3mV/V
4, Load cell connection number: 1~4 at 350 ohm
5, Division: 1/2/5/10/20/50
6, Display: 6 bits LED, 9 status indications(LED version)
6 bits LCD, 13 status indications(LCD version)
7 Communication interface(optional): RS232, serial sending, able to connect to scoreboard
or PC
8 Power supply: AC 100~240V/50~60Hz adatper
DC 6V/4AH(when voltage is below 5.4V, it will indicate)
9, Charge currency: 1A
10, Working temperature range: -10~40℃
11,Transportation temperature range: -25~55℃
第六页(显示与键盘)
一、Chapter Two Front View
Front View: LED version
Keyboards
[Zero] key: To zero within the manual zero range, and current weight will be zero;
[Tare] key: To tare and net indication will be on, and it will display net weight;
[Accum] key: To accumulate and get accumulation times/accumulation value;
[Count] key: To enter into counting function;
[Keep] key: To get peak hold function;
[Enter] key: To confirm and set up for function;
[ON/OFF] key: To turn on when it if off and to turn off when it is on;
二、Chapter Three Interface(传感器连接,加图)
Load cell interface
1, The load cell connector is 7-pin, round, each pin is defined as following:
2, If 4-wire cable is adopted, then short connect +E and +S, -E and –S;
3, Pin definition is as following:
Pin 1: -IN
Pin 2: +IN
Pin 3: +E
Pin 4: +S
Pin 5:-S
Pin 6: -E
Pin 7: Shield
Calibration(标定设置)
Press [on/off] key to turn on the indicator, then press [span] button which is at the back of
indicator until it displays the software version, it will display [SET-CL]. There are four types of
parameters, they are [SET-CL][SET-FU][SET-GR][SET-SP], you could press [hold] key to
switch among the four types of parameter. The definition and operation for each type of
parameter is defined as following:
[SET-CL]
When indicator displays [SET-CL], press [tare] key, it will initialize the parameter as following:
A: Default metrology parameter:
FU1-1(KG);FU2-1(2%); FU3-1(10%); FU4-1(0.5d); FU5-1(FAST);FU6-1(0.5d);FU7-1(0.25s)
B: Default gravity setup parameter:
The default gravity for both calibration place and use place is 908035;
C: Default calibration data clear (needs to re-calibrate);
D: Default user’s setting for different version:
LED, single display version
FNC1-1(No); FNC2-1(No);FNC3-1(No);FNC4(); FNC5-3(Bright);FNC6();FNC7();FNC8();
FNC9(); FNCa-1(. For decimal point);FNCb-1();FNCc();FNCd-1();
FNCe-1(9600);FNCf-1();FNCg(); FNCh-1(indicate after stable); LO(0); UP(0);
E=x for division set up Press [hold] key to select, press [enter] key for next;
Px.xxx for decimal point set up Press [hold] key to select, press [enter] key for next;
FSCALE for full capacity set up Press [enter] key to set up for full capacity, it displays
xxxxxx, press[→][←][↑] to modify the value, press [enter] key for next;
ZERO for zero point calibration Press [zero] key, it will display from 5 to 1 to calibrate zero
point;
LOAD for full capacity calibration Load some weights according to the full capacity, press
[enter] key for confirm, it displays xxxxxx, press [→][←][↑] key to input the value that is in
accordance with the weight value, after stable, press [tare] key, it displays from 5 to 1 to
calibrate full capacity point;
END4, pres [enter] key, it will display [SET-SP], press [hold] key to select other parameters,
press [span] button to back to weighing status.
Inner code display
Press [on/off] key to turn on the indicator, press [zero] key before it displays 999999, it will
display inner code, press [enter] key to back to weighing status.