Astme662 2017
Astme662 2017
Astme662 2017
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1. Scope* D2843 Test Method for Density of Smoke from the Burning
1.1 This fire-test-response standard covers determination of or Decomposition of Plastics
the specific optical density of smoke generated by solid E176 Terminology of Fire Standards
materials and assemblies mounted in the vertical position in E662 Test Method for Specific Optical Density of Smoke
thicknesses up to and including 1 in. (25.4 mm). Generated by Solid Materials
1.2 Measurement is made of the attenuation of a light beam 3. Terminology
by smoke (suspended solid or liquid particles) accumulating 3.1 Definitions—For definitions of terms found in this test
within a closed chamber due to nonflaming pyrolytic decom- method refer to Terminology E176.
position and flaming combustion.
4. Summary of Test Method
1.3 Results are expressed in terms of specific optical density
which is derived from a geometrical factor and the measured 4.1 This test method employs an electrically heated radiant-
optical density, a measurement characteristic of the concentra- energy source mounted within an insulated ceramic tube and
tion of smoke. positioned so as to produce an irradiance level of 2.2 Btu/s·ft2
(2.5 W/cm2) averaged over the central 1.5-in. (38.1-mm)
1.4 The values stated in inch-pound units are to be regarded diameter area of a vertically mounted specimen facing the
as standard. The values given in parentheses are mathematical radiant heater. The nominal 3 by 3-in. (76.2 by 76.2-mm)
conversions to SI units that are provided for information only specimen is mounted within a holder which exposes an area
and are not considered standard. measuring 29⁄16 by 29⁄16 in. (65.1 by 65.1 mm). The holder is
1.5 This standard measures and describes the response of able to accommodate specimens up to 1 in. (25.4 mm) thick.
materials, products, or assemblies to heat and flame under This exposure provides the nonflaming condition of the test.
controlled conditions, but does not by itself incorporate all 4.2 For the flaming condition, a six-tube burner is used to
factors required for fire hazard or fire risk assessment of the apply a row of equidistant flamelets across the lower edge of
materials, products or assemblies under actual fire conditions. the exposed specimen area and into the specimen holder
1.6 This standard does not purport to address all of the trough. This application of flame in addition to the specified
safety concerns, if any, associated with its use. It is the irradiance level from the heating element constitutes the
responsibility of the user of this standard to establish appro- flaming combustion exposure.
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use. 4.3 The test specimens are exposed to the flaming and
nonflaming conditions within a closed chamber. A photometric
2. Referenced Documents system with a vertical light path is used to measure the varying
light transmission as smoke accumulates. The light transmit-
2.1 ASTM Standards:2
tance measurements are used to calculate specific optical
density of the smoke generated during the time period to reach
1
This test method is under the jurisdiction of ASTM Committee E05 on Fire
the maximum value.3
Standards and is the direct responsibility of Subcommittee E05.21 on Smoke and
Combustion Products.
5. Significance and Use
Current edition approved Feb. 1, 2017. Published March 2017. Originally 5.1 This test method provides a means for determining the
approved in 1979. Last previous edition approved in 2015 as E662 – 15a. DOI:
specific optical density of the smoke generated by specimens of
10.1520/E0662-17.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
3
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Additional parameters, such as the maximum rate of smoke accumulation, time
Standards volume information, refer to the standard’s Document Summary page on to a fixed optical density level, or a smoke obscuration index provide potentially
the ASTM website. useful information. See Appendix X1.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
developing and maintaining positive pressure during test 7.1.1.1 If the interior wall surfaces become corroded or the
periods, in accordance with 11.12. The air-tightness of the coating starts to peel off, users shall repair the damaged area
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
chamber shall be tested at least one per test day in accordance using any suitable coating material, installed to the coating
with 11.2. manufacturer’s instructions.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
Also shown in Fig. 3 are the spring and rods for retaining the 7.1.5.1 The light source shall be an incandescent lamp
specimen within the holders. operated at a fixed voltage in a circuit powered by a constant-
7.1.4 Framework for Support of Furnace and Specimen voltage transformer. The light source shall be mounted in a
Holder—The furnace and specimen supporting framework sealed and light-tight box. This box shall contain the necessary
shall be constructed essentially in accordance with Fig. 4. optics to provide a collimated light beam passing vertically
7.1.5 Photometric System—The photometric system shall through the chamber. The light source shall be maintained at an
consist of a light source and photodetector, oriented vertically operating voltage required to provide a brightness temperature
to reduce measurement variations resulting from stratification of 2200 6 100°K.
of the smoke generated by materials under test. The system
shall be as shown in Figs. 5 and 6 and include the following:
7.1.5.2 The photodetector shall be a photomultiplier tube, (38.1-mm) aperature on the front and a finned cooler supplied
with an S-4 spectral sensitivity response and a dark current less with compressed air mounted on the rear to maintain a constant
than 10−9 A. A set of nine gelatin compensating filters varying body temperature of 200 6 5°F (93 6 3°C).
from 0.1 to 0.9 neutral density are mounted one or more as 7.1.6.1 As an option to the air-cooled radiometer, a water-
required in the optical measuring system to correct for differ- cooled heat flux meter is suitable for use in measuring the heat
ences in the luminous sensitivity of the photomultiplier tube. flux. The heat flux meter shall consist of a Schmidt-Boelter
These filters also provide correction for light source or photo- (thermopile) sensor approximately 1.0 in. (25 mm) in diameter
multiplier aging and reduction in light transmission, through mounted in a specimen holder. The specimen holder shall
discolored or abraded optical windows. An additional criterion include the millboard described in 8.3.3.2, with a hole in the
for selection of photomultiplier tubes requires a minimum center to accommodate the meter. The meter shall be mounted
sensitivity equivalent to that required to give a full scale such that the sensing surface is flush with the millboard. The
reading with only the No. 5 compensating filter in the light meter shall have an operating range of 0-4.4 Btu/s·ft2 (0-5.0
path. A light-tight box located directly opposite the light source W/cm2) and an accuracy of within 63 %.
shall be provided to mount the photodetector housing and the 7.1.7 Thermocouple—A thermocouple shall be fixed to the
associated optics. A glass window shall be used to isolate the center of the inner surface of the wall opposite the door.
photodetector and its optics from the chamber atmosphere. 7.1.8 Output Instrumentation—The outputs of the radiom-
7.1.5.3 In addition to the above compensating filter, a eter shall be measured using a potentiometer and the results
neutral density range extender filter permitting the system to recorded. The photodetector output shall be measured with a
measure to Optical Density 6 is incorporated in the commercial potentiometer or other suitable instrument capable of measure-
version of the smoke density chamber. The accuracy of ment over the range of the apparatus. See Annex A1.
read-outs in the range above Ds 528 is affected by the excessive 7.1.9 Sensor for Chamber Pressure Measurements —A
light scattering present in such heavy smoke concentration. pressure sensor (for example, a manometer or pressure trans-
Where Ds values over 500 are measured, it is necessary to ducer) with a range up to 6 in. (152 mm) of water (1.5 kPa)
provide a chamber window cover to prevent room light from shall be provided to monitor chamber pressure and leakage.
being scattered into the photomultiplier, thereby providing an The pressure measurement point shall be through a gas-
incorrect higher transmission value. sampling port in the chamber.
7.1.6 Radiometer—The radiometer for standardizing the 7.1.10 Chamber Pressure Relief System—A simple water
output of the radiant heat furnace shall be of the circular foil column or relief valve shall be provided to permit control of
type, the operation of which was described by Gardon.6 The chamber pressure (see A2.8).
construction of the radiometer shall be as shown in Fig. 7. It
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
above the lower opening of the specimen holder and 1⁄4 6 1⁄32 position during nonflaming exposures. The fuel shall be
in. (6.4 6 0.8 mm) away from the face of the specimen surface. propane having a 95 % purity or better. Filtered oil-free air and
Provision shall be made to rotate or move the burner out of propane shall be fed through calibrated flowmeters and needle
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
immediately removed while still warm and viscous. One way adhesive is intended for field application of a finish material or
to correct or prevent this situation, is for the user to prepare a substrate, the prescribed type of adhesive and the spreading
set of six tempered spring steel wires each approximately 31⁄2 rate recommended for the assembly of test specimen shall be
in. (89 mm) long fabricated from 30-gage (0.014 in.) wire, with used and reported.
one end crimped or brazed to a knob to facilitate handling and 8.3.1.1 Finish materials, including sheet laminates, tiles,
to prevent possible loss of the wire by complete insertion. fabrics, and others secured to a substrate material with
When a burner tiplet becomes clogged as indicated by flame adhesive, and composite materials not attached to a substrate,
extinguishment and inability to relight or by a distorted flame have the potential to be subject to delamination, cracking,
shape, thus invalidating the test, insert one of the wires and peeling, or other separations affecting their smoke generation.
work it through several times to clear the obstruction. Imme- To evaluate these effects, it is often necessary to perform
diately upon removal of the burner from the chamber while still supplementary tests on a scored (split) exposed surface, or on
warm, insert all six wires in a like manner but leave them in interior layers or surfaces. When supplementary tests are
place until the next time the burner is used. Where residues and conducted for this purpose, the manner of performing such
clogging persist, prepare a suitable solvent bath so as to supplementary tests, and the test results, shall be included in
immerse the complete burner and use the wires to loosen any the report with the conventional test results.
hardened residue. Because of the construction, it is impossible 8.3.2 For comparative tests of finish materials without a
to service the individual burner tiplets from the opposite normal substrate or core, and for screening purposes only, the
direction, but because of ratio of diameters any obstruction following procedures shall be employed:
pushed through the small diameter tiplets is likely to readily 8.3.2.1 Rigid or semirigid sheet materials shall be tested by
drop through the large diameter body tubing. Since most of the standard procedure regardless of thickness.
these solvents are hazardous, take proper precautions for 8.3.2.2 In the absence of a specified assembly system,
handling and protection of personnel. If flammable solvents are paints, adhesives, etc., intended for application to combustible
used, take care to ensure that “hot” burners are not immersed base materials, shall be applied to the smooth face of 1⁄4-in.
until cooled to room temperature. (6.4-mm) thick tempered hardboard, nominal density 50 to 60
lb/ft3 (800 to 960 kg/m3), using recommended (or practical)
8. Test Specimens application techniques and coverage rates. Tests shall also be
8.1 Size—The test specimens shall be 3 by 3, +0, −0.03 in.
(76.2 by 76.2, +0, −0.8 mm) by the intended installation 7
The boldface numbers in parentheses refer to the list of references at the end of
thickness up to and including 1 in. (25.4 mm). Materials this standard.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
conducted on the hardboard alone, and these values shall be 8.3.3.1 All specimens shall be covered across the back,
recorded as supplemental to the measured values for the along the edges, and over the front surface periphery with a
composite specimen. single sheet of aluminum foil (0.001 6 0.0005 in. or approxi-
8.3.2.3 Paints, adhesives, etc., intended for application to mately 0.04 mm) with the dull side in contact with the
noncombustible substrate materials, shall be applied to the specimen. Care shall be taken not to puncture the foil or
smooth face of 1⁄4-in. (6.4-mm) thick inorganic reinforced introduce unnecessary wrinkles during the wrapping operation.
cement board, nominally 110 6 10 lb/ft3 (1762 6 160 kg/m3) Fold in such a way so as to minimize losses of melted material
in density, using recommended (or practical) application tech- at the bottom of the holder. Excess foil along the front edges
niques and coverage rates. shall be trimmed off after mounting. A flap of foil shall be cut
8.3.2.4 If fabrics or thin flexible films tend to shrink, to and bent forward at the spout to permit flow from melting
bunch, to blister, or to pull out from under the specimen holder specimens.
during the test, the three test specimens shall be stapled with its 8.3.3.2 All specimens shall be backed with a sheet of 1⁄2-in.
aluminum foil wrapper to the inorganic insulation millboard (12.7-mm) thick inorganic insulation millboard. The specimen
backing. Five wire staples,8 approximately 1⁄2 by 1⁄4 by 0.02 in. and its backing shall be secured with the spring and retaining
(12.7 by 6.3 by 0.5 mm),9 shall be positioned horizontally at rod. A modified C-shape retaining rod or similar device shall be
the center, and at the center of the four quadrants.8 used with specimens from 5⁄8 to 1 in. (16 to 25 mm) thick. Do
8.3.2.5 Electrical and Optical Fiber Cables—For test speci- not deform compressible specimens below their normal thick-
mens of electrical or optical fiber cables up to 1 in. in diameter, ness.
cut the cables to 3 + 0, −0.03 in. (76.2 + 0 − 0.8 mm) lengths
and insert enough pieces in the specimen holder to fill it, 9. Conditioning
arranged vertically. Wrap a sheet of 1⁄2-in. (12.7-mm) thick 9.1 Pre-dry specimens for 24 h at 140 6 5°F (60 6 3°C) and
inorganic insulation millboard with aluminum foil and place it then condition to equilibrium (constant weight) at an ambient
behind the wires as a backing board before inserting the spring temperature of 73 6 5°F (23 6 3°C) and a relative humidity of
and retaining rod. 50 6 5 % (see 6.7).
8.3.3 Specimen Mounting:
9.2 While in the conditioning chamber, specimens shall be
supported in racks so that air has access to all surfaces.
8
Bostitch B8, Swingline 888, Acc. Monarch No 1 FT, or similar, have been Forced-air movement in the conditioning chamber will assist in
found satisfactory. accelerating the conditioning process.
11.11 Record the light transmittance and the corresponding 12.1 Calculate specific optical density, Ds, at any given time
time either as a continuous plot with a multirange recorder or as follows:
at time intervals no greater than 30 s with a multirange meter D s 5 G @ log10 ~ 100/T ! 1F #
readout. Make and note the necessary full-scale range changes
where:
in decade steps.
G = V/AL,
11.11.1 The photometer used with this instrument shall have V = volume of the closed chamber, ft3 (or m3),
an accuracy of 63 % or better of the maximum reading on any A = exposed area of the specimen, ft2 (or m2),
range. As such, the percentage error of a given reading L = length of the light path through the smoke, ft (or m),
becomes progressively worse at the lower portion of the scale. T = percent light transmittance as read from the light-
Avoid light transmittance on scale readings less than 10 by sensing instrument, and
making the appropriate decade range change. F = depends on the following:
11.11.2 Some chambers are equipped with a switch that not (1) If the movable filter (see 7.1.5.3) is in the light path at
only incorporates ranges of 100, 10, 1, and 0.1 but also ranges the time that T is being measured, F = 0, and T is the actual
of 30, 3, and 0.3. With such an instrument the greatest accuracy percent transmittance.
would be achieved in light transmittance readings by making a (2) If the filter has been moved out of the light path (see
range change in these intermediate ranges when the light 7.1.5.3 and 11.15) at the time that T is being measured, F = the
transmittance reading reaches 30 on the 0-to-100 meter scale or known optical density of the filter (see A1.1.4), and T is an
10 on the 0-to-33 scale. apparent percent transmittance.
(3) If the optical system is not equipped with a movable
11.12 Observe the increase in chamber pressure with the
filter in accordance with 7.1.5.3, F = 0, and T is the actual
manometer described in 7.1.9. Use regulator (see A2.8) to
percent transmittance.
maintain the pressure in the range of 4 6 2 in. (100 6 50 mm)
of water during most of the test. If negative pressure develops 12.1.1 For an instrument constructed in accordance with
after very intense specimen flaming, open the inlet vent slightly this standard, corrections for the volume of the furnace
to equalize the pressure. As a result of pressure rise, adjust the assembly and the volume included in the door recess are
fuel and air valves during the flaming test to maintain constant generally less than 1 %. As such, G = 132.
flow rate. 12.1.2 A table for Ds versus actual percent light transmit-
tance is given in Appendix X2. The Ds values above 528 are
11.13 Record any observations pertinent to the burning and
based on an assumed optical density of 2.00 for the movable
smoke generating properties of the material under test, in
filter.
accordance with 13.1.6 and 13.1.7.
12.2 Calculate the maximum specific optical density, Dm,
11.14 Continue the test for a period of 3 min after a
using the equation in 12.1 with a light transmittance corre-
minimum light transmittance value is reached or after an
sponding to the minimum level reached during the test.
exposure of 20 min, whichever occurs first.
12.2.1 Similarly, calculate Dc using the Tc value.
11.14.1 Optionally, the test shall be permitted to be con-
12.2.2 Calculate Dm (corrected) as follows:
ducted for periods in excess of 20 min at the request of the test
D m ~ corr! 5 D m 2 D c
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
sponsor.
11.15 If transmittance falls below 0.01 %, the chamber 12.3 For systems without “dark current” cancellation or
window shall be covered with an opaque screen to avoid “blank adjust” provisions, a correction shall be made for any
possible light-scattering effects from room light. Also any light transmittance reading, T, approaching the dark current
supplementary optical filter in the photometer system shall be value, Td. Calculate the corrected light transmittance, T', as
removed or displaced in order to extend the measuring range. follows:
If the potential exists for extraneous light to reflect into the 12T T 2 Td
T' 5 1 2 5
photometer during removal of the filter, turn the high voltage 1 2 Td 1 2 Td
Flexible polyurethane foam, high resiliency, 1⁄2-in. 29.6 23.6 47.3 80.7
Rigid polyisocyanurate foam, 1⁄2-in. 6.2 13.4 10.0 38.4
NBS SRM 1007aB 6.5 7.6 10.4 23.0
A
Precision statements for polystyrene sheet and fiber glass-reinforced polyester sheet are not given because the Dm values fell outside the range of the instrument.
B
The average Dm value obtained by 20 laboratories testing 3 samples each (60 samples) was 433.
accordingly. It is reasonable to expect that this version of the 14.1.4.2 Reproducibility, R2—The critical difference within
method will provide better precision than that tabulated. which two averages of three specimens each, obtained by two
14.1.2 The precision statements in these tables are ex- different operators, using different instruments in different
pressed as a percentage of the average Dm of each material and laboratories, can be expected to lie 95 % of the time because of
are based on only the validated results (see Section 3) from the the random variations within and between laboratories.
three replicates submitted to each laboratory.
14.1.3 Coeffıcient of Variation—The ratio of either the 14.2 Bias—The bias is unknown because the value of
“within laboratory” or “between laboratories” standard devia- specific optical density obtained in this procedure is defined
tion to the overall average Dm value for the material, expressed only in terms of this test method.
as a percent.
14.1.4 Relative Precision: 15. Keywords
14.1.4.1 Repeatability, R1—The critical difference within 15.1 fire; fire-test response standard; smoke; smoke cham-
which two averages of three specimens each, obtained on the ber; smoke density; smoke obscuration; solids; specific optical
same material by a single operator using the same instrument, density
can be expected to lie 95 % of the time because of random
variation within a laboratory.
ANNEXES
(Mandatory Information)
A1.1 Photometric System the optical beam may be of importance in cases where light
A1.1.1 A properly used photometer of the type described in scatter takes place, as often occurs in smoke aerosols. Because
this document is an inherently linear device provided that of this, the following instructions are included for use in cases
linear electronic measuring and recording equipment has been where the photometer beam needs to be realigned following
used. The linearity of absorption measurements is not depen- replacement of the light source or some accidental misalign-
dent upon critical beam collimation; however, collimation of ment.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
A1.1.2.2 Attach the templet with transparent tape to, and instrument shall have an accuracy of 63 % of the maximum
centered on, the upper optical window. With the optical system reading on any range. It involves a spectral band quite similar
in its normal operational mode, observe the projected image on to that corresponding to human vision. This is defined by the
the templet. A properly aligned beam will completely fill the operating condition of the lamp source and the spectral
2-in. (51-mm) circle with some spill-over. Because of the sensitivity of the photodetector. Since no precise control is
filament, the pattern will not be a perfect circle. If the pattern maintained over the size of this spectral band, it would be
is too large or too small, the lower lens will require adjustment. necessary, if accurate calibration were to be attempted, to make
Remove the cover from the light source enclosure. If the use of filters with constant transmission over a spectral band of
pattern is not centered, it will require repositioning of the light at least 350 to 750 nm. Such filters are not readily available.
source or slight readjustment of the lens mount in its track. One Because of this and the inherent linearity of a properly
way to optimize the lens position is by slight adjustment until constructed photometer and measuring circuit, it is not recom-
mended that the test method user attempt precise calibration of
the maximum photometer reading is obtained, whereupon it is
the instrument over its operating range. The following rough
locked. Replace the enclosure cover, making sure that all
calibration procedure is, however, recommended as a means to
screws have been tightly seated.
ensure that no gross failure of the photometric measuring
A1.1.2.3 Switch off the photometer and remove the cover system has occurred:
from the roof-mounted optical enclosure. Remove the compen- A1.1.3.1 Complete alignment as in A1.1.2.
sating filter holder from the lens mount and observe the A1.1.3.2 With the photometer beam blocked, determine that
converging beam of light. A properly focused and aligned the instrument shows zero transmission on all the normal
beam will form a small intense spot at the disk aperture of the photometer ranges without removal of the range extension
photomultiplier housing projecting into the roof of the enclo- filter from the photometer head.
sure. If the beam is misaligned or not properly focused, loosen A1.1.3.3 Measure the transmission of a neutral density filter
the lens mount screws very slightly and carefully refocus. of nominal optical density of 3.0 which has been previously
Tighten the screws and recheck the light spot. Remount the calibrated in another smoke density photometer. The two
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
TABLE A1.1 Correction Factors for Range Extension Filter ND2 Neutral Density Filter Removal Correction FactorsA
Meter Correction Optical density of Meter Correction Optical density of
indication, factor neutral density filter, indication, factor neutral density filter,
%T Cf log Po/P=D %T Cf log Po/P=D
31 –27.4 1.79 51 +1.1 2.01
32 –25.6 1.81 52 +2.2 2.02
33 –23.8 1.82 53 +3.3 2.025
34 –22.1 1.83 54 +4.4 2.03
35 –20.4 1.845 55 +5.5 2.04
36 –18.8 1.86 56 +6.5 2.05
37 –17.3 1.87 57 +7.5 2.06
38 –15.7 1.88 58 +8.5 2.064
39 –14.2 1.89 59 +9.5 2.07
40 –12.8 1.90 60 +10.5 2.08
41 –11.4 1.91 61 +11.4 2.086
42 –10.0 1.92 62 +12.3 2.09
43 –8.6 1.93 63 +13.2 2.10
44 –7.3 1.94 64 +14.2 2.107
45 –6.0 1.95 65 +15.0 2.114
46 –4.8 1.96 66 +15.9 2.12
47 –3.5 1.97 67 +16.8 2.13
48 –2.3 1.98 68 +17.6 2.135
49 –1.2 1.99 69 +18.5 2.14
50 0.0 2.00 70 +19.3 2.146
A
Corrections are to be applied to the Ds values equivalent to the 0.01 to 0.001 %T and 0.001 to 0.00001 %T values only.
set to the “1-scale.” Adjust the micrometer knob to give an furnace in the same position as in A1.2.2. Immediately
exact mid-scale reading (0.5 % transmittance). Rotate the thereafter, obtain a short (5 to 15-s) record of the temperature
range switch back to the“ 100-scale” and move the range rise of the disk. Determine this temperature rise of the
extension filter out of the optical path. Observe the meter calorimeter by measuring the electrical output of the thermo-
reading. If the meter reading is 50 % T, the value of the filter couple attached to the back of the disk, employing a recording
is exactly optical density 2.0 and the preprinted conversion potentiometer operating at a fast chart speed (1 in./s; 25 mm/s).
tables, Appendix X2, are suitable for direct use. If the meter Remove the calorimeter and allow it to cool back to room
temperature.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
NOTE A1.1—As an example of the procedure proposed, it is possible to this case the value is 8.8 mV.
simplify the equation for the radiant heat absorbed by a particular copper
disk, as follows:ASSUME, AS AN EXAMPLE: A1.2.8 Under normal continuous use conditions, the radi-
ometer shall be calibrated at least once every three months.
m = 29.78 g
Ag = 11.40 cm2
Annual recalibrations shall be required in all cases.
n = 4 A1.2.9 The blackened face of the radiometer shall be
Ah = 0.008 cm2
inspected frequently. In case the coating is blistered, cracked,
Then: discolored, or broken, the coating shall be removed, the face of
An = 11.37 cm2 the radiometer cleaned, and a new coating applied. In this case,
C = 0.0927 cal·g−1·°C−1 the recoated radiometer shall be recalibrated before being used.
K = 4.184 cal·g−1·°C−1
0.040 mV·°C−1 A1.2.10 The copper disk standard shall be carefully handled
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
k =
a = 0.98 when in use, and protected from surface contamination and
mechanical abuse when stored. If the blackened face shows
11
alterations as in A1.2.9 the coating shall be removed and the
Nextel velvet 101-C10 provides a radiation absorption characteristic (a) of
face cleaned. The disk shall then be reweighed and recoated
0.98. Nextel velvet 101-C10 and its replacement, Solar Absorber Coating ECP-
2200, are no longer manufactered by 3M Company. Nextel is a registered trademark and any appropriate corrections made in the calibration
of the 3M Company. constant, G, before it is used again.
A2.1 Radiant Heat Furnace that it is 11⁄2 6 1⁄32 in. (38.1 6 0.8 mm) away from, parallel to,
A2.1.1 The furnace shown in Fig. 2 has been found to be and centered with respect to the exposed specimen area.
suitable. The dimensions that are shown in Fig. 2 and the A2.3.2 Adjustment screws shall be provided to align the
components to which they refer are critical. Other portions of furnace with reference to the specimen.
the design are optional. The heating element consists of a
A2.3.3 The framework shall have two 3⁄8-in. (9.5-mm)
coiled wire or other suitable electrical heating element capable
diameter transverse rods of stainless steel to accept the guides
of dissipating about 525 W, mounted vertically in a horizontal
of the specimen holder described in 7.1.3. The rods shall
ceramic tube 3 in. (76.2 mm) in inside diameter by 33⁄8 in. (85.7
support the holder so that the exposed specimen area is parallel
mm) in outside diameter by 15⁄8 in. (41.3 mm) long. The tube
to the furnace opening. Spacing stops shall be mounted at both
is bored out at one end to 31⁄32-in. (77.0-mm) inside diameter
ends of each rod to permit quick and accurate lateral position-
and to a depth of 5⁄8 in. (15.9 mm) to accommodate the heating
ing of the specimen holder.
element. A 1⁄16-in. (1.6- mm) insulation paper gasket and two
stainless steel reflectors are mounted behind the heating A2.4 Photometric System
element. A 3⁄8-in. (9.5-mm) insulation millboard disk, provided
with ventilation and lead wire holes, shall be positioned behind A2.4.1 The photometric system shall consist of a light
the heating element and used to center the assembly with source and photosensitive element as defined in 7.1.5. The
respect to the front 3⁄8-in. (9.5-mm) insulation millboard ring system shall be as shown in Figs. 5 and 6. The window in the
by means of a 6-32 stainless steel screw. The adjustment nuts chamber floor through which the light beam passes shall be
on the end of the centering screw shall provide proper spacing provided with a ring-type electric heater mounted on the
of the furnace components. The cavities adjacent to the heating underside of the window out of the light path. The heater
element assembly shall be packed with glass wool. The furnace maintains the minimum window temperature at 125°F (52°C)
assembly shall be housed in a 4-in. (102-mm) outside diameter on the inner surface of the window to minimize smoke
by 0.083-in. (2.1-mm) wall by 41⁄8-in. (105-mm) long stainless condensation. The collimated beam inside the chamber shall
steel tube. Two additional 3⁄8-in. (9.5-mm) insulation board have a path length of 36 6 1⁄8 in. (914 6 3 mm) and a sensing
spacing rings and a rear cover of 3⁄8-in. (9.5-mm) insulation cross section of 11⁄2 6 1⁄8-in. (38 6 3-mm) diameter (see
board shall complete the furnace. The furnace shall be located Annex A1). A typical photomultiplier photometer system will
centrally along the long axis of the chamber with the opening require a high-voltage d-c power supply and a neutral density
facing toward and about 12 in. (305 mm) from the right wall. filter of sufficient optical density to produce a convenient signal
The centerline of the furnace shall be about 73⁄4 in. (195 mm) level for the indicator or recorder. The photometer system used
above the chamber floor. shall be capable of permitting the recording of reliable optical
densities of at least 6.0, corresponding to transmittance values
A2.2 Specimen Holder of 0.0001 % of the incident light (see Appendix X2).
A2.2.1 The specimen holder shall conform in shape and A2.4.1.1 The two optical platforms and their housings shall
dimension to Fig. 3 and be fabricated by bending and brazing be kept in alignment with three metal rods, 1⁄2 in. (12.7 mm) in
(or spot welding) 0.025-in. (0.6-mm) thick stainless steel sheet diameter, fastened securely into 5⁄16-in. (7.9-mm) thick exter-
to provide a 11⁄2-in. (38.1-mm) depth, and to expose a 29⁄16 by nally mounted top and bottom plates and symmetrically
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
29⁄16 in. (65.1 by 65.1-mm) specimen area. As described in arranged about the collimated light beam.
7.1.3, the holder shall have top and bottom guides to permit
A2.5 Radiometer
accurate centering of the exposed specimen area in relation to
the furnace opening. A3 by 3-in. (76.2 by 76.2-mm) sheet of A2.5.1 The 200°F (98°C) body temperature of the radiom-
1⁄2-in. (12.7-mm) inorganic insulation millboard, having a eter shall be monitored with a 100 to 200°F (38 to 100°C)
nominal density of 50 6 10 lb/ft3 (800 6 160 kg/m3) shall be thermometer located as shown in Fig. 7 in a 1⁄2 by 1⁄2 by 11⁄2-in.
used to back the specimen. A spring bent from 0.010 in. (12.7 by 12.7 by 38.1-mm) long brass or copper well drilled to
(approximately 0.25-mm) thick phosphorbronze sheet shall be accept the thermometer with a close fit. The use of silicone
used with a steel retaining rod to securely hold the specimen grease is a way to enhance the probability of providing good
and millboard backing in position during testing. thermal contact. The circular receiving surface of the radiom-
eter shall be spraycoated with an infrared-absorbing black
A2.3 Support of Furnace and Specimen Holder paint. The radiometer shall be calibrated calorimetrically in
A2.3.1 The framework as shown in Fig. 4 shall have welded accordance with the procedure summarized in A1.2.
to it a 5-in. (127-mm) outside diameter, 1⁄4-in. (6.4-mm) wall,
A2.6 Chamber Wall Thermocouple
2-in. (50.8-mm) long horizontally oriented steel tube to support
the radiant heat furnace described in 7.1.2. This support tube A2.6.1 A thermocouple suitable for measuring a tempera-
shall have provision to accurately align the furnace opening so ture of 35°C shall be mounted with its junction secured to the
A2.7.1 The multiple burner shall have six tubes with con- A2.8 Chamber Pressure Regulator
struction details as shown in Fig. 3 (Note A2.1). The six tubes
shall be made from 1⁄8-in. (3.2-mm) outside diameter by A2.8.1 A suitable pressure regulator consists of an open,
0.031-in (0.8-mm) wall stainless steel tubing. All tubes shall be water-filled bottle and a length of flexible tubing, one end of
swaged at the tip to reduce the opening diameter to 0.055 in. which is connected to a sampling port on the top of the
(1.4 mm). The manifold section of the burner shall consist of chamber. The other end of the tubing is inserted 4 in. (100 mm)
1⁄4-in. (6.4-mm) outside diameter by 0.035-in. (0.9-mm) wall below the water surface. The bottle shall be located at or below
stainless steel tubing. The other end of the manifold is attached the floor level of the chamber to avoid back-siphoning.
to a fitting in the chamber floor.
APPENDIXES
(Nonmandatory Information)
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
X2. TABULAR CONVERSION OF PERCENT TRANSMITTANCE TO SPECIFIC OPTICAL DENSITY WHEN G = 132.
X4. COMMENTARY
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
(1) radiant heating in the absence of ignition and (2) an open test method might be applied.
FIG. X4.1 Diagonal Survey of Irradiance at Specimen During Operation at Nominal 2.5 W/cm2
terial placed in the trough, only a small quantity of smoke following a test. In addition, the current version requires
developed when another material was placed in the trough. reporting Dm (corr). The reporting of Dm as a preferred
This did not seem, however, to be too different from that measurement result is based on the following facts:
performance which might be expected from the same materials X4.5.1.1 The deposit remaining following a test represents a
in another fire exposure, and thus there does not seem to be any part of the smoke produced. Thus, it seems irrational to
reason to ban from the test thermoplastic materials that melt or subtract this unless it can be shown that the deposit results from
drip into the trough. late accumulation following a peak smoke reading. The pro-
cedures of the test method seem to make this unlikely.
X4.4 Precision X4.5.1.2 Experience has shown that the determination of Tc
used eventually to calculate D m (corr), is subject to variations
X4.4.1 In any method, one of the important considerations
in operator technique during the chamber venting procedure.
is the degree to which it, when applied to a given material, will
X4.5.1.3 The introduction of the correction, while not in
yield constant results. Since this test results in destruction of
itself a significant technical problem, suggests a technical
the specimen, the results of any test to determine precision are
sophistication that simply is not justified on the basis of
affected not only by the random errors that might be inherent
intended use of the data. The effect of these facts was noticed
in the procedure but also by any variation in the properties of
during analysis of the round-robin experimental data. The
the replicate specimens. Thus, in studying the degree to which
results were found to be more consistent for the uncorrected
experimental results can be repeated within a given laboratory,
data (Dm).
it is desirable to use a material from which specimens of
uniform composition and dimensional characteristics can be X4.6 Limitations on Application of Smoke Measurement
prepared. Data
X4.4.1.1 This fact was recognized in planning the large X4.6.1 The smoke problems that develop during unwanted
interlaboratory study of precision of the measurement method. fires have been recognized for many years. Fire fighters are
In spite of this, some of the experimental variability observed faced with it daily in their work. However, three problems have
was undoubtedly related to variations in the replicate speci- tended to prevent application of standards limiting the accept-
mens. In at least one instance, variation in thickness as great as ability of materials or products on the basis of smoke produc-
20 % was observed. To assist in identifying variability resulting tion: (1) the extent to which the smoke measurement assesses
from this cause, requirements for weighing specimens have the smoke hazard (12, 13); (2) the lack of a well-defined
now been included as a part of the test procedure. measurement method which could be shown to provide a
X4.4.2 Various changes were made in the test method technically valid means for smoke characterization; and (3)
description as adopted as compared to the description used to most materials or products, when burning, release large quan-
advise for the round-robin test conducted. These included: (1) tities of smoke, and there have been only limited ways of
running additional samples when the results of three specimens reducing smoke production.
are highly variable, (2) maintenance of pilot burner, (3) X4.6.1.1 The first problem still exists, although as indicated
deletion of data that are inconsistent with the equipment, and in X4.2.2.1 there is a valid basis for consideration that this type
(4) improved calibration and alignment procedures. These of smoke measurement does in fact serve to measure the light
changes are such that the precision data given should be attenuating properties of the smoke.
assumed to be conservative as they relate to the test method X4.6.1.2 The second problem has been partially alleviated
adopted. Better precision would be expected if another labo- with the development of the smoke chamber. However, it must
ratory round-robin test should be conducted. be recognized that only two of a wide range of fire exposure
X4.4.2.1 When studying the results reported by the various conditions are simulated by the test method. Thus any rank
laboratories participating in the round-robin study, it was ordering of materials by the test must be recognized as only
realized that the test method draft given to the laboratories to based on the fire exposure conditions applied and, in fact, the
follow failed to contain a section describing conditions under test method develops different rankings depending upon
X5.1 In the 1970s the Office of Standard Reference X5.1.2 The last certificate produced by NIST regarding
Materials, National Bureau of Standards (NBS), Gaithersburg, SRM 1006D described the material as single-layer sheets of
MD, developed two materials, intended for use as standard paper, principally α cellulose. Each sheet was approximately
reference materials for this instrument: a single layer of 172 by 254 by 1.23 mm thick. The NIST certificate further
alpha-cellulose (cotton linters) paper SRM 1006 and a plastic stated that the certified value and expanded uncertainty for
sheet, SRM 1007a. Subsequently, the National Institute of maximum specific optical density of SRM 1006D, at a single-
Standards and Technology (NIST), successor to the National layer thickness was:
Bureau of Standards, developed a new plastic sheet, made of an D m 5 210 6 18 ~ without correction for window deposit !
acrylonitrile-butadiene-styrene copolymer, SRM 1007b. The (X5.1)
calibration sheets issued with these materials demonstrated a
variability with thickness, so that recommendations were made D m corr. 5 193 6 20 (X5.2)
The NIST certificate explained that the certified values and
to assess the thickness to within 0.0005 in. (0.013 mm). uncertainties were the result of 30 tests on representative
X5.1.1 In 2011, NIST discontinued the production and sale single-layer thickness specimens of the SRM lot, where a
of SRM 1006D, the last version produced of the standard single sheet approximately 76.2 by 76.2 by 1.23 mm thick
was used for each test. The series of 30 tests resulted in ex-
cellulosic reference material for the non-flaming exposure panded uncertainties for 1006D of Dm and Dm corr. of 18
condition in Test Method E662.12 and 20, respectively.
X5.1.3 In the absence of SRM 1006D, no NIST Standard
Reference Material for the non-flaming exposure condition in
12
https://www-s.nist.gov/srmors/view_detail.cfm?srm=1006D Test Method E662 exists.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
13
The sole source of supply of the material known to the committee at this time
is Fire Testing Technology Ltd., Charlwoods Road, East Grinstead, West Sussex,
15
RH19 2HL, United Kingdom. If you are aware of alternative suppliers, please The sole source of supply of the material known to the committee at this time
provide this information to ASTM International Headquarters. Your comments will is Schneller LLC, 6200 49th St. North, Pinellas Park, FL 33781, USA. If you are
receive careful consideration at a meeting of the responsible technical committee,1 aware of alternative suppliers, please provide this information to ASTM Interna-
which you may attend. tional Headquarters. Your comments will receive careful consideration at a meeting
14
Refer to https://www-s.nist.gov/srmors/view_detail.cfm?srm=1007B. of the responsible technical committee,1 which you may attend.
REFERENCES
(1) Hirschler, Marcelo M., “Smoke Results from a Set of Over 100 (7) ASTM Committee E05.2 “Interlaboratory Test Data for the Smoke
Carpets in the NBS Smoke Chamber. Statistical Analysis and Inves- Chamber Test Method,” RR: E-5-1002, 1976.
tigation of Affecting Factors,” Fire and Materials, Vol 16, No. 3, (8) Gaskill, J. R.,“ Smoke Development in Polymers During Pyrolysis or
1992, pp. 127–33. Combustion.” Journal of Fire and Flammability, JFFLA, July 1970,
(2) Gross, D., Loftus, J. J., and Robertson, A. F., “Method for Measuring pp. 183–216.
Smoke from Burning Materials,” Symposium on Fire Test Methods— (9) Chien, W. P., and Seader, J. D., “Prediction of Specific Optical Density
Restraint and Smoke, ASTM STP 422 Am. Soc. Testing Mats., 1966, for a Smoke Obscuration in an NBS Smoke Density Chamber,” Fire
p. 166. Technology, FITCA, Vol. 11, August 1975, pp. 206–217.
(3) Gross, D., Loftus, J. J., Lee, T. G., and Gray, V. E., “Smoke and Gases (10) Bartosic, A. J., Rarig, F. J., “Evaluation of the XP2 Smoke Density
Produced by Burning Aircraft Interior Materials,” NBS Bldg. Sci.
Chamber,” Symposium on Fire Test Methods—Restraint and Smoke,
Series, BSS 18, U.S. Government Printing Office, Washington, D. C.,
ASTM STP 422, Am. Soc. Testing Mats., 1966, p. 106.
February 1969.
(11) Robertson, A. F.,“ Estimating Smoke Production During Building
(4) Brenden, J. J.,“Usefulness of a New Method for Smoke Yield from
Species and Panel Products,” Forest Products Journal, FPJ0A, Vol 21, Fires,” Fire Technology, FITCA, Vol. 11, May 1975, pp. 80–94.
1971, pp. 23–8. (12) “The Control of Smoke in Building Fires. A State of the Art Review,”
(5) Lee, T. G., “Interlaboratory Evaluation of Smoke Density Chamber,” Materials Research and Standards, Vol 11, No. 4, 1971, pp. 16–24.
NBS Tech. Note 708, December 1971. (13) Committee E-5 “A Report on Smoke Test Methods,” Standardization
(6) Lee, T. G., “The Smoke Density Chamber Method for Evaluating the News, August 1976, pp. 18–26.
Potential Smoke Generation of Materials,” NBS Tech. Note 757,
January 1973.
SUMMARY OF CHANGES
The following changes have been made to this test method since the last issue, E662–15a (approved Nov. 1,
2015).
(1) The phrase “for no apparent reason” was removed from (2) Section 1.4 was removed and subsequent sections renum-
10.1.1. bered.
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
--`,,,```,``,,```,``,,,,`,`,,,,,-`-`,,`,,`,`,,`---