Elisys Uno: - User Manual
Elisys Uno: - User Manual
Elisys Uno: - User Manual
| User Manual
REF 17350/1
REVISION LIST OF THE MANUAL
Rev. /DATE. REVISION DESCRIPTION
01/2006-06 First Edition
02/2008-02 Adaptation to new corporate design
03/2009-05 Performance check kit added
04/2011-09 Correction dimensions
05/2022-05 Reworked - IVDR IP added, QR Code for software and settings added
06/2022-07 IVDR minor text adjustment
SYSTEM VERSION
COPYRIGHT
Copyright 2022, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in
writing. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7
2 INTRODUCTION 9
2.1 INTENDED PURPOSE ELISYS UNO 9
2.2 DETAILED DESCRIPTION 9
3 INSTALLATION 11
3.1 INSTRUMENT SETUP 11
3.2 INSTRUMENT CHECK OUT 16
7 ASSAY EDITOR 73
7.1 MAIN MENU AND TOOLBAR 74
7.1.1 Assay 75
7.2 CREATING ELISA ASSAYS 82
7.2.1 Absorbance mode 83
7.2.2 Point-to-point mode 83
7.2.3 Regression mode 84
7.2.4 Cubic spline (constrained) mode 86
7.2.5 Cut-off standard mode 86
7.2.6 Cut-off standard mode 88
7.2.7 Dose response mode 88
7.2.8 Polynomial modes 88
7.2.9 Parameter logistic 89
7.2.10 %Absorbance 89
7.3 QC CRITERIA 89
7.4 ASSAY STEPS 91
7.4.1 Select step 91
7.4.2 Edit step 102
CONTENTS
9 TROUBLESHOOTING 127
9.1 FLAGS AND ERROR MESSAGES 127
9.1.1 Flags 127
9.1.2 Error messages 129
9.1.3 Rackfile – Troubleshooting 135
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read the
”GENERAL SAFETY WARNINGS”, describing the appropriate operating procedures.
Please contact your HUMAN authorised local Technical Service in the event of
instrument failure or other difficulties with the instrument.
Figure 1
Biological hazard symbol
2 INTRODUCTION
ELISYS UNO may also be used in production processes involving micro volume
dispensing, diluting, incubating and reading.
3 INSTALLATION
How to unpack, set up, and check out your ELISYS UNO instrument.
ELISYS UNO is carefully packaged in a custom-made container to assure its safe
arrival. If upon receipt the outer packaging is damaged report damage to your
freight carrier immediately.
2. Connect one end of the drain tubing to the drain outlet located on the bot-
tom of the instrument. Push the tubing over the fitting as far as possible,
ideally flush with the plastic nut. The waste drain tubing may be routed
! Note: Do not allow the end of
the tubing to rest in the drain
container below the expected lev-
through an access hole on the lab bench or routed to the front or back of the el of the waste liquid.
instrument as desired towards the waste drain container.
12
3. At the edge of the table or above a lab table access hole, cut the drain tube
and connect it to the barb tee, as shown in Figure 2. One of the other ends
of the barb tee should connect to the remaining drain tube, leading down-
wards to the waste drain container.
Figure 2
Figure 3
The upright barb will act as an air vent as shown in Figure 4 so that the waste
fluid will not develop an air pocket preventing the fluid from flowing into the
waste container below.
Figure 4
4. Place the waste drain container at a level lower than the instrument to allow
the fluid to flow into the waste drain container. A 2l bottle should be suffi-
cient for drainage (waste container is not supplied with the instrument). Cut
the end of the drain tube diagonally as shown in figure (below) and place the
diagonally cut end into the drainage container. The diagonally cut tube pre-
vents the tubing from resting on the bottom of the waste container, which
may cause the waste fluid to back up into the instrument.
Figure 5
Figure 6
Connectors
6. Put de-ionised water into the bottle marked Rinse. Put wash buffer provided
into the bottle marked Wash for ELISA tests. Leave the Waste bottle empty.
Check that each bottle cap is securely fastened and that no sensor wires are
! Note: The hydrophobic filter
on the waste bottle is de-
signed to protect the pump from
crossed. Note that the Waste bottle has short sensor leads in order to detect liquid and may become clogged
when waste is nearly full. Rinse and Wash bottles have long leads in order to when wet. Arrange the tubing so
sense when bottles are nearly empty. that the filter hangs down below
the connecter on the side of the
7. Fill the Prime bottle with fresh, clean de-Ionised water. This should be done ELISYS UNO to prevent clogging.
each day - This water enters the precision calibrated syringe pump and
therefore must be very pure to avoid damage and prolong the life of these
components. Prime bottle installed as shown in Figure 7.
Figure 7
Install prime bottle
14
8. Using the USB cable provided, connect the computer’s USB A port to the
ELISYS UNO USB B port (see Figure 8).
Figure 8
Back of instrument
3
9. Turn on the computer and insert the installation USB stick. The installer
should automatically start, if not, select Run from the Windows® Start Menu,
run USB stick drive:\setup, and follow the prompts for installing the program.
Execute the ELISYS UNO program after installation. Enter Admin as the user
ID and Admin as the password.
Figure 9
Figure 10
Figure 11
Figure 12
Figure 13
Figure 14
Figure 15
10. Connect the power cord to the instrument, then to an approved power
source. It is strongly advised that a UPS (Uninterruptible Power Supply) be
used to avoid power interruptions to the ELISYS UNO and to the computer.
11. Place a rack onto each of the two rack holders. The default positioning places
the reagent rack on the left and the Sample rack on the right. Place the plate
loaded with microwells into the plate carrier at the right.
12. The program default is to use COM1 for communication with the instrument.
If connecting the instrument to a different port, go to the Settings Menu and
select Software. Select a communications port and click OK.
Figure 16
16
If the instruments power comes on, but these actions do not occur and the beep
sound continues, there is a problem with the communications setup. Check the
USB cable connections and COM port settings.
At this time, the wash arm is raised from its shipping position. Install the wash
head (found in the accessories carton) to the wash arm using the two attached
thumbscrews. The luer fittings must face outward (towards the user), and the
colour-coded fittings on the tubing should be fastened finger-tight to the wash
head. See Figure 17.
Figure 17
Wash head installation
Each of the two racks and the plate move independently toward the front and
back of the instrument. Commonly referred to as a reagent rack, a sample rack,
and a Reaction plate. However, reagents can be placed in the sample rack, or
two racks can be used to perform pre-dilutions. Each rack has an arrangement
of holes or grooves configured to hold different types of tubes, bottles, micro
tubes, microwells, and other containers. Racks are identified in the software in
order to tell the instrument which configuration is to be used. They are also
displayed graphically.
The incubator plate/ well can be set to heat to 25°C, 37°C, or remain at ambient
room temperature. The plate/ well will heat to 25°C providing the ambient room
temperature is below 25°C (it should be noted that the option of heating the
plate/ well to 25°C should only be used when the ambient room temperature is
consistently below 20°C).
When the probe carries a reagent to an incubated reaction plate, the temperature-
controlled coil can be set to pre-warm the liquid before dispensing. Reagent
racks can be loaded and unloaded with bottles from run to run. The location
of each reagent is indicated using a colour-coded computer screen. Alternately,
preferred reagent rack setups can be stored in panels. For convenience, multiple
pre-loaded racks can be stored in the refrigerator ready to load and use.
When taking an optical reading, the reaction plate automatically positions itself
under the 4-channel optical system. Four lamps are aligned to simultaneously
shine down through four wells. A filter wheel with eight filters rotates constantly
below the plate. The filter wheel is designed so that four filters align with the
four lit wells for absorbance readings. Depending on the setup, reports may be
displayed or printed to create permanent lab records and physician reports.
18
Dimensions (W x D x H)
Instrument without 102 x 55 x 47 cm
any components:
Space required for 170x 80x 80 cm
routine use
Packaging 63 x 105 x 78 cm
Weight Gross 70.0kg, Net 35.0 kg
Washing
Wash head 8-probe, automatic prime and rinse
Programs Create and run user programmable protocols (aspirate,
dispense, soak). Can wash wells for re-use as applicable
Reading
Optical design Reads absorbance in four simultaneous channels; NIST
traceable calibration; user selects monochromatic or
bichromatic results
Light source Tungsten-Xenon lamp
4 position filter 405, 450, 490, 630
wheel
Interference filters Long life, hard coat, ion-assisted deposition, +/- 2 nm,
10 nm typical half band pass
Linear range -0.2 to 3.0 A
Photometric ± (1% of the reading +0.005 A from 0 to 1.5 A)
accuracy ± (2% of the reading +0.005 A from 1.5 to 3.0 A)
Software
Format USB drive or via Web download
Operating systems Windows(R) 7 and 10
Minimum system Processor 1Ghz, 1 GB Ram, USB-B Port, 800x600
Recommended Core i3 2,2Ghz, 8GB Ram, 512GB SSD, USB-B
system
Secondary menu Create/ edit protocols, import/ export data, etc., Con-
options trol, Run, Setup
Calculation modes Absorbance, single standard, factor, fixed time kinetics,
kinetics by standard or factor, multi-calibrator point-
to-point, linear regressions, log-logit, cubic spline, and
nonlinear regressions (curve fit).
Self monitoring Lamp, bottle volume, filters, pressure, vacuum, me-
modes chanical function, and more
QC options Store control data, print Levey-Jennings or QC range
plots, calculate SDs
USB port Standard USB A to B, connection via COM port auto-
matically established
20
Power
Voltage range 100- 250 VAC
Frequency range 50- 60 Hz
Power maximum 160 W
There are three security access levels: Administrator, Manager and Operator.
See Table 1 for security levels.
- Login as different user: Displays the Login screen and allows you to login
with a different name and password.
- Log out: Logs current user out of the system.
- Create new user: Creates a new user with ID, password, and security level.
The manager can create users, but will only be able to give them operator
level security.
- Remove user: Administrators can remove any user. Managers can only re-
move operators.
- Change password: Any security level can use this feature. Enter the old pass-
word followed by the new password.
- Who is logged in: Security access window appears to show who is logged in
and their access level (see Figure 19).
Figure 19
Security access window
- View log file: Opens a log of every user name that has logged on, including a
time and date stamp and a running count of entries.
5.2.1 Alignment
Figure 20
Under the Utilities menu,
Utilities menu
select Alignment
Use the arrow buttons to position the probe correctly. The probe tip should be
centred about 3 mm above the rear pin, as shown:
Figure 22
1 Side-to-side (blue) 1
2
3
Figure 23
To check new alignment, select Test. When finished click Save, then click Next
to continue with Step 2 of 2.
Use the arrow keys to move the probe until it almost reaches the bottom of the
sample cup. When the correct depth has been reached click Save and then the
Close button.
26
Use the arrows to move the probe into the correct position.
The probe tip should be centred in well H01 and almost touching the bottom.
Press Test to confirm. When finished, click Save and then the Close button.
Figure 29
Press the button. Adjust probe position with
Wash Cup
When finished, click Test to check new alignment. If aligned, click Save, then
Close.
- Use the top set of three buttons to select the High dispense height, and the
bottom set of three buttons to select the Low dispense height - the up and
down arrows will adjust the probe up and down, the button will
move the probe to its actual position setting.
- Low is normally used for dispensing small volumes such as serum to a pre-
dilution.
- High is normally used for dispensing a large volume of diluent (these set-
tings are specified when programming the assay in Assay Editor
- Low should be set inside the tube to eliminate splashing and loss of sample;
- High should be set as high as possible to thoroughly mix the sample/diluent,
but not high enough to cause reagent foaming.
Test with your particular reagents for appropriate dispense heights. Click Save
on each to save the new settings, then click Close.
Figure 32
Adjust probe dispense height
Press the Test Current Voltages button to view the filter voltage readings for all
four channels at each of the eight wavelengths (if nothing happens, wait for the
automatic lamp warm-up time).
Figure 34
Current voltages test results
All filter voltages should be between 2.00 and 10.00 (see Figure 34). Lower
readings might indicate that a lamp is misaligned or partially blocked. The data
storage buttons on this window are for troubleshooting and are only used when
instructed to do so by service engineers. Click Close to exit.
If abnormally low voltage is detected, the user will be prompted to mark the
wells read by that lamp as unavailable. If not able to immediately replace the
lamp at that time, allow the software to mark the wells as unavailable until it
can be replaced.
30
Figure 35
Select Channel Blank...
from Utilities menu
This will allow the user to blank the four photometer channels on a wetted
optically-clear solution (example blanking solution included, 0.5 N NaOH with
100 µl Triton X-100/ l). The ELISYS UNO software will show which wells it will be
using for channel blanks (preferably, use a new set of wells), and then prompt to
put Blanking Solution in position 1 of Rack 1 (never use plain water).
Figure 36
Load the wetted blanking
material into position
The ELISYS UNO will then pipette the four blank wells, read, and store them,
then do a confirmatory reading. The absorbance values of the readings will
be displayed and should be 0 ± 0.0050 A. These blanks will automatically be
subtracted from absorbance readings. See Figure 37; notice at the bottom of the
window the Result is displayed.
Figure 37
Reset channel blank results
If one of the values is not between 0 ± 0.0050, follow the instructions and try
doing channel blanks again. Do make sure there are clean wells in the positions
indicated to run channel blanks again.
32
Figure 38
Figure 39
Notepad file
This file can be saved and/ or printed or filed for future reference.
Figure 41
1 2 3 4 5 6 7 8 9
Figure 42
Layout tab
2 3 4 5
This window displays the current status of the instrument including the
temperatures. Also shown are the currently loaded racks and plate. The software
automatically keeps track of which wells in the reaction plate have been used.
Table 2
Feature: Description: Item No.:
Status The temperatures of the probe/coil and reaction plate 1
window are continually displayed in the ELISYS UNO Status Win-
dow. The temperature reading of the probe coil and
plate are only accurate near 37 °C.
At normal ambient temperatures, the reported temper-
atures are not accurate and often the displayed temper-
ature of the probe coil and plate do not agree and may
report higher than the actual temperature. When plate
heating is turned off, so is probe coil heating. When run-
ning room temperature assays, disregard the displayed
temperature values.
Calibration tab
Figure 43
Calibration tab
1. Choose assay: Use the drop-down menu to choose one of the assays from
the list. Press the Print Report button, to print for the selected assay.
2. Add calibration test information: Lists the calibrators and controls by name,
copies required, and whether or not they are valid.
Selections:
- Select the Curve option to insert a blank and a calibrator for the assay into
the work list. Select multiple times to add more copies of each. This button
is not active if the user is running an assay that does not require calibrators.
! Note: The user may also se-
lect calibrators, controls, and
blanks individually.
- Select the Control option to add all of the controls specified in the assay.
Press the button multiple times to add multiple copies of the controls.
36
3. View results: View the calibration and control results in this area.
5. Activate selected button: To edit the curve, check the curve records (choose
part of curve records which look good to user), then click the Activate Select-
ed button. This button will be enabled once the software calculates (based
on time, logic, math, etc.) a valid curve. After activating new curve records,
the current curve is changed. The software will look at the test list to recal-
culate all finished tests of this assay.
6. Discard recent: Upon accepting the new adjusted curve with the settings in
Strategy Settings set to Auto, the option to ‘discard recent’ is no longer avail-
able (reference Accept Curve and Control Results Menu in chapter 6.1.1.2
ELISA strategy.) However, if Strategy Settings remain defaulted to ‘manual’,
the user has the option to ‘discard recent’ and return to the previous curve.
With Strategy Settings set to Manual, the new concentration values will
display, however, the original values remain, allowing the user to ‘discard
recent’.
7. Print preview: Preview calibration and control results before printing statis-
tics such as %CV, %Dif, and mean values are also shown.
8. Print report
Figure 44
Sample tab
This tab can be used to set up a quick and easy work list.
1. Enter sample IDs: Press the Add Numerical ID button to enter samples by
number. After pressing the button, follow the instructions in the pop-up.
Press Add Patient ID to choose a patient from the Patient Database. Refer-
! Note: The ‘drag and drop’ fea-
ture allows reagents and sam-
ple racks to be changed within
ence 6.2.2.8 Choose patient record for more information. their rack location and between
racks as well.
2. Choose tests: Click on one or more of the patient ID’s on the left. With the
patient(s) highlighted, choose the assay or assays to run with the chosen
patients. Click on the test and press Add test. Click on the same test again
to deselect it.
3. Work list: Lists the IDs and assays assigned to each and requested copies. To
add more copies to a test, highlight that row and use the hidden pull down
to change the number. To remove a patient, click the row and press the Re-
move button. If you click Calibrate, all registered calibrators are loaded to
the worklist as well (not visible in this list). You can then print the work list
and/or press the Request button. When you choose Request and confirmed
the schedule and the wash bottle loading, the layout tab opens. Verify the
location of all samples and press Start Run when done loading the racks. The
test will now begin.
38
1 2
Figure 45
Test list tab
3 4
2 3
Figure 46
4 1 5 5 6 7 8
By default, the Report tab shows the information from the most recent test run.
In Figure 46, the Report tab shown is displaying information after running for
example TSH, T3 and T4.
Results are organised based on the time they are run. Select all or individual
tests using the checkboxes.
1. Select Sorting option from drop down menu to change the way your results
are displayed.
2. Select the History checkbox to begin searching for results by date, substance
type, name/ID or test name.
40
Past results may also be searched for this way. Input as much search criteria
as possible for more precise search results.
• Search by Date: Use the drop down menus to select the dates to search
for results.
• By Name/ ID: Search for results for a specific patient.
• By Assay: Use this menu to search for results from a specific assay.
3. Select the magnifying glass to display the kinetic trend (absorbance vs. sec-
onds) of the selected sample.
4. Select All: Check this box to select all Names/IDs. Only selected results will
be printed.
5. Print: Print all selected Results. Print Preview shows a preview of the print-
out.
6. Export: Export selected reports to a text file (*.txt), MS Excel file (*.xls), or
XML file (*.xml). Save for future reference.
7. Remove: Remove all selected Results.
8. Retest (add to Sample Tab): Add the selected items to the sample tab to be
retested.
This section will introduce you to the items located in the ELISYS UNO Manager’s
main menu.
Figure 47
Management menu options
Option Description
Initialize Establish or re-establish communication between the soft-
ware and the instrument without restarting the software.
Pause engine Pauses the process that your ELISYS UNO is doing at that
time. Resume allows ELISYS UNO to resume the previous
task from the point it paused.
Reload assay Use this feature after editing or creating new assays with
files Assay Editor. The new or edited assays are then added to
the list of available assays in ELISYS UNO Manager.
Change Change the temperature for specific tests. For instance,
temperature when running the Performance Check, change to Room
Temperature.
Change rack1 Change the rack style
Change rack 2 Select another rack type for sample tubes.
Multi-plate test Used when running a plate that contains long incubation
times. Other tests, even entirely different modes, can be
run by loading a different plate. There is a reminder to re-
load the plates when time is almost expired.
Communication Used for service purposes ONLY
window
Calibration event See Chapter 6.2.3 Calibrators for more information.
LIS Import Read the information that follows this table explaining the
LIS Import Option and examples of import and export files.
Exit Exit the ELISYS UNO Manager Software.
42
- Import process:
• Select Management LIS Import
• Choose a file containing LIS requests.
• Requested tests are added automatically to request list in the Sample
tab.
- Export process:
• In the Report tab, user will select records and press Export button.
• In the Save-As Type drop down list, the user will select LIS files (*.lis). User
types in a file name, and a folder, for the file and presses Save button.
A carriage return is used to indicate the end of the line (220 is the maximum).
Record Layout Specifications Minimum Requirements
Figure 48
Figure 49 Figure 50
Import file example (.LIS file extension) Export file example (.LIS file extension)
44
Option Description
Start of Day See Chapter 6.2.1 Run start of day for more information.
End of Day See Chapter 6.4 End of day for more information.
Wash Probe Select Wash the Probe and reset syringe positions.
Prime Syringe Primes the fluid handling system with the liquid in the
Prime bottle.
Prime Washer - Primes the wash system with the solution in the Wash
Wash Bottle bottle.
Prime Washer - Primes the wash system with the solution in the Rinse
Rinse Bottle bottle.
Temperature On / Enable and disable temperature control.
Temperature Off
Read Wells... Enables selection of wells to be read; selection of filters
(primary and differential); number of reads; and to use,
or not use, a stored blank. Select the Clear button to re-
move all of the results. Select Export to send the results
to a text (.txt) file.
Custom Script... Allows scripts to be run consisting of a series of com-
mands. Recommended for advanced users.
5.8 QC tracking
Enables Controls and Calibrators to be tracked using a Levy-Jennings chart. In
past versions, QC tracking could be accessed at the assay level through the QC
tab. With ELISYS UNO this feature is accessed by selecting QC Tracking from QC
on the main menu.
Figure 52
Tracking menu page
Figure 53
Tracking example
46
Enter the assay, control, and lot number information in the drop down menus
to select the control that you wish to review. The curve can be viewed by
absorbance or by concentration results.
Points in the Levey-Jennings Chart can be edited. Unselect the points which shall
not be displayed in the graph. Make sure to click Save QC Points afterwards.
5.8.1 Sample DB
See Chapter 6.2.2 Sample database setup for details.
5.8.2 Settings
See Chapter 6 Running ELISYS UNO for details on software settings.
5.8.2.1 Software
See Chapter 6 Running ELISYS UNO for details on software settings.
Or Assay specific calibration templates (have the preference over assay mode
templates).
Figure 54
48
5.8.2.3 Language
Choose between available languages. Restart the software after changing a
language. The instrument does not need to be restarted.
5.8.3 Utilities
5.8.3.1 Alignment
See Chapter 5.2 Alignment setup.
It is necessary to supply an empty sample vial for [PNP] to be used when self
test is running. [SOL1] is intended to refill [SOL2] if required.
Figure 55
Figure 56
Self test
To begin the test the user selects Self Test from the Utilities menu. The user then
places the bottle containing 23ml of [SOL1] in position 1 of the reagent rack and
puts an aliquot of the [PNP] in a sample vial placed in position 1 of the sample
rack. After loading the solutions and inserting a clean plate into the instrument,
the Start button is pressed and the self test is performed. The Test Progress bar
will display the progress of the test. After the self test completes the instrument
displays Pass/ Fail results on the screen. Clicking the Details button will display
the complete self test report.
Pressing the Details button once the Self Test is completed will open the Self
Test log in Notepad.
50
The Self Test log in Notepad displays the date, time, status of Lamps, Blank,
Pipetting Accuracy, Photometer/ Plate Positioning, Temperature, Blank Abs,
Filter Volts, etc.
Figure 57
Type CV Criteria
Large (18 μl sample) > 2%
Medium (10 μl sample) > 2%
Small (2 μl sample) > 3%
Please note:
- Verify start and ending temperatures are room temperature;
- All blanks should be between -0.005 and +0.005;
- If not, repeat setting blanks;
- Filter Voltages must be between 2 and 10;
- Make sure the plate position has passed.
If running the Self Test yields poor results, attempt the following steps:
1. If small (2 μl) CV% results are within specifications but medium (10 μl) and
large (18 μl) are out of range, failure is likely not with pipetting accuracy but
rather with photometer/ plate positioning.
! Note: [PNP] is temperature
sensitive. Make sure the in-
strument temperature is set to
2. If small (2 μl) CV% results are not within specifications, check the following: OFF and the temperature has sta-
• Check for diluter system leaks bilized. [PNP] irritates eyes, skin
• Check syringes for freedom of movement and mucous membranes. Upon
• Check that probe tip is not partially bent contact, rinse thoroughly with co-
• Rerun channel blanks pious amounts of water and con-
• Recheck photometer alignment. If this is adjusted, check plate X/Y parms. sult a doctor.
• Recheck probe alignments
Figure 58
52
Figure 59
Select Pack options from the drop down menu. Options include packing assay
files, root files, lot# files, and rack files. An individual file may be selected by
clicking in the check box next to the file name, or all files may be selected by
clicking on the Select all checkbox.
Figure 60
Figure 61
The screen will display a message when the pack file operation has completed.
Figure 62
The packed file will be located in the directory with the file name shown in the
To display box on the screen.
54
5.8.4 Security
See Chapter 5.1.2 Password security and logging in for more information.
Figure 63
The About selection will open a window and display the version of the ELISYS
UNO Manager installed in your instrument.
Figure 64
Figure 66
Software settings
56
You may choose to keep the records of numerical ID records from 1 to 10 years;
choose to keep sample IDs from 1 to 30 days. However, whatever preferences
you choose, all sample test records are backed up monthly to an archive
database. The communication port can be changed. The default temperatures
are displayed for ELISA mode. Press OK when finished.
Option
Sample Database
Figure 67 Description
Customize sample database field
names, uncheck field name from
the list displayed if you do not want
to use it.
Option
EIA Strategy
Description Figure 68
Set parameters for how to run the
tests and how to handle the results
in ELISA mode.
Save reagent
Sacrifices some instrument per-
formance in exchange for reagent
economy.
Option
Report Appearance
Figure 69 Description
Specify your preferences on the re-
port appearance.
Option
Report Output
Figure 70 Description
Specify optional data to print for
different report types. You can also
specify if report should be printed
jointly.
The sample handling system will be primed with deionised water. The washer (if
applicable) will be primed with wash solution and the lamps will turn on. After
running the Start of Day program, visually check the sample handling tubing
! Note: Observe the fluid han-
dling system and ensure there
are no leaks.
and the syringe for the presence of any bubbles.
If bubbles are present, repeat the Start of Day program, tapping the tubing
where the bubbles are present. If this does not eliminate the trapped bubbles,
perform the Weekly alcohol cleaning procedure, see Chapter 6.2.1.1.
60
Figure 72
When the cycle is complete, replace the bottle containing 70% Isopropyl Alcohol
with the prime bottle containing fresh deionised water and repeat the Prime
Syringe procedure.
2. When first opened, the Sample Database window will not have any records
to display.
Figure 73
3. Select the Edit button to enter Sample ID, Family Name, Given Name, etc.
This will be helpful as search criteria when searching the database. Press the
Done button when finished.
4. An instructional dialogue box will open.
• Navigate through the fields on the screen by using the Tab or arrow keys
on the keyboard.
• Create a Sample ID by typing in the Sample ID field.
! Note: The Edit button toggles
to Done
Figure 75
Figure 76
2. Modify the information and then press the Done button to save the changes.
Figure 78
Figure 79
64
Figure 80
2. Click on File and select Save. Close the window when finished.
Figure 81
Search result
displayed
Figure 82
6.2.3 Calibrators
Before Samples can be read, Calibrators must be registered and read. Open the
Calibration tab and choose “TSH calibrator A” as an example from the Choose
Assay drop down menu. Entering the lot numbers for calibrators is optional,
since the calibrator concentrations are set in the Assay Editor.
Lot # Registration
1. Open Lot Registration from the top menu.
Figure 83
Choose the corresponding name
3
2
66
2. Click the Substance in the left column you would like to register.
3. Click on Register New Lot#
Figure 84
Enter Lot # and optional
description
Concentration range
Figure 85
Entering concentration range
Figure 86
Select curve
CC
NC
PC
If you wish to save this as a Work List for validating your calibrators in the future,
press Save, choose a location, enter a name, and press OK.
68
1 1 2 3
Figure 87
Verify reagent, calibrator and
controls are properly loaded
Figure 88
Test List tab shows the status of
the current run
When ELISYS UNO is finished reading, the Calibration Event window opens. This
window shows whether the new curves and controls were accepted. If the curve
is not accepted, this procedure must be repeated.
The controls will be accepted even if they read out of range if the Control Properties
are set to Warn and Continue. To check your Control Properties or change this
setting, see Chapter 7 Assay editor. If you do not want to see this window every
time, click the Do not automatically show up box in the bottom left corner of the
window (see Figure 89 Calibration event). If checked, the Calibration Event can
only be viewed by selecting it from the Routines window or by pressing the F8
key on your keyboard at any time.
Figure 89
Calibration event
When selected, Calibration Event will display the blank, calibrator, and controls
valid for 7 days. After they expire, you will have to re-run the Calibrators to use
that assay. You may run more frequent recalibrations and control checks as often
as you like. To change the time valid, you must edit the assay using Assay Editor.
70
Press Add Numerical ID. Type in the First ID and the total number of patients
(see Chapter 5.3.2 Sample tab). Click OK to add the IDs.
Use loading positions to correctly load the racks (refer to Figure 92 Layout tab
with load positions).
1 2 3
Figure 92
Layout tab with load position
1 1 2 3
Figure 93
Test List tab opens
After the tests are complete, you must Accept, Delete or Rerun the tests. Tests
must be accepted before they can be viewed in the Report tab. Click the Report
tab to get reports on all of the results (see Chapter 5.3.4 Report tab).
72
7 ASSAY EDITOR
Assay Editor works in conjunction with ELISYS UNO Manager. Although all
HUMAN Methods are included and pre-programmed in the ELISYS UNO Manager
software it is possible to add new methods by choosing Launch Assay Editor in
the Utilities menu.
Figure 94
Welcome screen
Figure 95
User software menu
Access the Assay Editor via the shortcut, or click Launch Assay Editor in the
Utilities drop down menu within the user software.
Figure 96
Working screen
7.1.1 Assay
New ELISA: Clears all entries and allows building of a new Assay.
Open ELISA: Opens an existing Assay.
7.1.1.1 View
Figure 97
View menu
7.1.1.2 Substances
Use this menu to choose one of the Substances and manage the list. Managing
the list refers to adding: New, Edit or Delete Current Substances.
Figure 98
Substances menu
Figure 99
Manage reagents
76
7.1.1.3 Panels
This feature allows you to set up and save panels of assays to be run as a group.
Use New EIA to create a new Panel Assay list. Select an assay from All Assays and
add it to the Panel list by pressing the arrow buttons. Assays can belong to more
than one panel. Click Open EIA to open an existing Panel Assay list.
Figure 102
Panel menu
Figure 103
Select an existing or newly
created panel
7.1.1.4 Indices
Sometimes a test result is not a measurement, but a calculation made from
other measurements. To create an equation that calculates a test result, select
New ELISA.
When making a Patient Report, the Index is reported only if there was a valid
test result for that patient for each of the Assays required to calculate the Index.
1 3 4 5 7 8
Figure 104
Create a new index
2 6
78
7.1.1.5 Security
Figure 105
Security menu
Table 3
Security level Administrator Manager Operator
Disable security Y N N
Enable security Y N N
Create manager Y N N
Create operator Y Y N
Remove manager Y N N
Remove operator Y Y N
Open assay Y Y Y
Save assay Y Y N
Open panel Y Y Y
Save panel Y Y N
Open index Y Y Y
Save index Y Y N
Create substance Y Y N
Edit substance Y Y N
Delete substance Y Y N
Login as Different User: Shows the login screen and allows you to login with a
different name and password.
Logout: You are logged out and the Login screen appears and waits for next user.
Press Cancel to quit Assay Editor.
Create New User: Create a new user with ID, password, and security level. The
manager can create users, but will only be able to give them operator level
security.
Remove User: Administrators can remove any user. Managers can only remove
operators.
Change Password: Any level can use this feature. Enter your old password
followed by the password you want to change it to.
Who Is Logged In: Security access window appears to show who is logged in and
their security access.
Figure 106
Who Is Logged In
View Log File: Displays a text file listing logins, anything that has been modified,
date, time and by whom.
80
7.1.1.6 Settings
Figure 107
Settings
Figure 108
Printout Font selection
Figure 109
Installed Filters (may vary)
7.1.1.7 Icons
To see the function each icon has, place the mouse cursor over the icon in Assay
Editor. A description will appear.
Save: Saves the current Assay under the name entered in the “As-
say Definition” Section.
See Chapter 7.4 for more information.
Help: Opens the Help window. See Chapter 5.8.5 for details.
82
Figure 110
For ELISA Assays
Figure 111
Assay Definition
section
Figure 112
Figure 113
Assay definition
Figure 114
When using the Logit calculation, the 0 calibrator is not plotted as part of the
curve; it is used only for the calculations.
Figure 116
Figure 117
COV = X*mean(NC) + F
- Use this equation if only a negative control is used to determine the COV.
COV = Y*mean(PC) + F
- Use this equation if only a positive control is used to determine the COV.
Reverse Cut-off mode: With the Reverse Cut-off option, the samples with
values lower than the cut-off are labelled as positive. If you choose this
option, be careful to follow the < and > signs in the prompts for entering
cut-offs and ranges.
88
This yields a sigmoid curve. The parameter Bottom is the absorbance value at
the bottom of the plateau, Top is the absorbance value at the top of the plateau,
and LogEC50 is the concentration value halfway between the Bottom and
Top. The parameter HillSlope describes the steepness of the curve. When the
Hillslope is less than 1.0 the curve is shallower, when the Hillslope is greater
than 1.0 the curve is steeper.
7.2.10 %Absorbance
These mode used to show absorbance ratio %.
7.3 QC Criteria
Clicking the QC Criteria button brings up the QC Criteria window shown in
Figure 119. From this window one can add new control criteria, delete, edit or
change the checked order of the control criteria defined in the Assay
! Note: There must be substanc-
es (standards, controls, blank)
in the assay before selecting QC
Criteria.
Figure 119
Assay editor – QC
Criteria
90
Figure 120
QC Criteria over-
view
Pressing the New button brings up the window shown in Figure 122 Cut-off QC
general criterion. By clicking on the Substances, Operators, and Numbers you
can enter any QC equation that you would like. This General Formula option is
also accessible from the QC button in assay modes “other than cut-off mode”.
Figure 121
QC Criteria tem-
plate (for other than
Cutoff)
Figure 122
QC Criteria general
criteria (for Cut Off)
Figure 123
Assay steps dialogue box
Figure 125
Add sample step
parameters dialogue
Figure 126
Pre-dilute parameters
Figure 127
94
1 1
A. Main settings
Dilution ratio: Ratio of sample volume to total volume
Total volume: Total predilution volume
Sample volume: Sample volume to be added to the diluent to make the
dilution.
Advanced button (optional): Opens the advanced parameter dialogue,
where the Aspiration Speed and Air Gap can be changed.
Dilute to plate (instead of sample rack): Check this option if you want the
dilution to be sent directly to the well.
Set the volumes manually: Manually specify the volume of Sample and
diluent to be used for the predilution.
Diluent volume: Amount of diluent to be added to a Sample.
B. Aspiration settings
Aspirate separately: If this option is not checked, both the diluent and the
Sample are aspirated in a single step, and then dispensed together to the
vial on the sample rack.
Separation air pocket:
What should be diluted (in addition to samples):
• Dilute standards: Check this option if you want to dilute both samples
and standards.
• Dilute controls: Check this option if you want to dilute both samples
and controls.
• Use diluent for blank: Check this option if you want to use the diluent
for blanking.
C. Dilution mixing in Rack2
Mix by aspirate/ dispense
• Mixes the dilution before loading by repetitive aspiration and dispensing.
• The number of times and the percentage of volume can be set.
96
Figure 130
Select or create a
substance (reagent)
Select Add Reagent from the list. Select an existing Substance or Reagent, or,
create a new one.
Figure 131
Add reagent step parameters
Figure 132
Add reagent step advanced
parameters
Figure 133
Incubate step parameters
Figure 134
Incubate step parameters
7.4.1.4 Read
Read step parameters:
Figure 135
Figure 136
Primary filter: Select a primary filter to be used for readings. There is no default
value. A primary filter must be selected before continuing.
Differential filter: Select a differential filter that should be used for readings, if
any. There is no default value. One of the options from the list must be selected
to continue. One of the options is None. The differential filter cannot be the
same as the primary filter.
Some assay inserts do not specify to use a differential filter, but when running
assays in microplates as in ELISYS UNO, a differential reading filter usually
improves results.
! Note: A differential filter
should always be used if at all
possible when running assays in
ELISYS UNO
It is important to choose a differential wavelength which is at an absorbance
minimum in the absorbance spectrum of your reacted reagent. If your assay
insert does not specify a differential wavelength to use for your reagent and
there is no corresponding ELISYS UNO example to use, contact the reagent
manufacturer for the best wavelength choice.
100
Figure 137
Kinetic read
windows
! Note: Some reading time delays can occur especially with large assay bat-
ches, but the abs/ min rate is calculated based on the actual reading time
interval not the programmed interval.
Figure 138
Figure 139
Figure 140
102
Figure 141
Figure 142
Figure 143
1. Units: Use this drop down menu to select the units of measure for an assay.
Figure 145
Select units
Figure 146
Normal range
Choose this option to set the range, which identifies that a Sample is normal. If
a result is out of normal range, it will be interpreted as High or Low.
4. Positive / Negative
Figure 147
Positive / Negative
Choose this option to set Positive and Negative limits. Results will be interpreted
as Positive or Negative. Samples in a range between the Positive and Negative
limit are interpreted as Equivocal.
Special groups can be defined based on age, gender, or keyword. In Figure 151
Special Group Edit Menu, Adult is defined as a patient who is greater than 18
years old.
To access Special Groups, click on Normal Range (see diagram box 1) in the
Interpretation Settings section of the screen. The Special Groups button (see
diagram box 2) will appear in the lower left hand corner of the screen.
106
The Special Groups Interpretation window will open. A list of defined groups
will display. Defined groups may be added, edited, and deleted.
Figure 150
If Edit Group is chosen, the Special Group Edit window will open. The Group
ID is the group definition. Age range and gender may be selected. Keywords
should be separated by a comma. The keyword characterizes the group. It may
be matched with a keyword in the Sample ID database or entered (for numerical
ID) when assigning tests for a sample. Example: Pregnant, non-smoker.
Figure 151
Special group edit menu
Keywords can also be used to determine the group assignment. When the
keyword is found in the patient database keyword field that patient will be
automatically assigned to the defined group. It is important to note that the text
used to name the Group ID is not searched by the database; the user must enter
the keyword(s) used to define the group in the Keyword text box. Figure 152
Figure 152
Special group edit
using keyword(s)
Refer to the ELISYS UNO Manager section for entering keywords in the Patient
Database. When running samples with numerical IDs the group can be manually
assigned at run time.
7.7 Standards
Standards are used to create the curve for calculating concentration from
absorbance. Standards are available in all calculation modes except Absorbance
and Factor. Assay editor has one standard predefined called Standard.
Figure 153
Add standard
1
1 Add button
2
2 Properties
3 3 Arrow keys
4 Remove
4
5 Curve valid time
5
! Note: The “Add” button is disabled when the chosen assay mode does not
use a Standard. It also becomes disabled when the maximum number of
Standards for a selected mode has been added to the assay.
108
Figure 154
Select or create a standard
Figure 155
Standard selected
2. Standard properties
Figure 156
Standard properties
2.1 Advanced button (optional): This should only be used if you need to set up
different parameters than in the Add Sample step.
Figure 157
Advanced button
Needs different volume: Check this box if the Standard needs a volume that is
different that the volume defined in the Add Sample step. When checked, the
Volume field is enabled.
Needs different advanced parameters (optional): Check this box if the Standard
needs advanced parameter(s) different than those defined in the Add Sample
step. When checked, the Advanced button is enabled.
Figure 158
Advanced parameters
Air gap: Amount of air (in microliters) that will separate the substance being
aspirated from the de-ionized (DiH20) water aspirated. It separates prime bottle
substance from the aspirated substance. Default values are automatically
selected based on the aspiration volume.
Pre-warm time: Some tests require you to pre-warm the standard after
aspiration, before it is added to the well. The pre-warm time can be set here.
Dispense height: There are two different levels of dispensing you can choose
from: High or Low.
3. Arrow buttons: Select a Standard in the list to the left and click the arrow for
the direction in which to move it in the list.
4. Remove: Select a Standard from the list on the left and click Remove to elim-
inate it from the list.
5. Curve valid time:
Figure 159
Curve info menu
Set the amount of time, in days, hours, or both, that the Standard Curve should
remain valid. Default is (7) days.
Figure 160
If the calibration is expired and tests are requested within the Sample menu of
the user software, calibrators will be loaded automatically to the Test list.
7.8 Blank
Figure 161
Blank
Using a Blank is optional in all modes except Standard mode. To use a Blank,
check the box next to Blank Used (Figure 161).
Blank Properties: Set up the Absorbance Range, Out of Range Action, and the
Valid Time for the Blank (Figure 162).
Figure 162
Blank properties
112
Valid time: Blank result valid time in days and hours. Default value is 7 days.
7.9 Controls
Figure 163
Controls menu
1
Controls are used to validate tests (see Figure 163). Assay editor has two controls
predefined: Normal Control and Abnormal Control. New controls can be defined
if needed.
Figure 164
Select or create controls
2. Properties
Figure 165
Control properties
Valid Time: Blank result valid time in days and hours. Default value is 7 days.
Advanced button (optional): This should only be used if you need to set up
different parameters than in the Add Sample step.
Figure 166
Need different parameters
than add sample step
114
Needs different volume: Check this box if the standard needs a volume that is
different that the volume defined in the Add Sample step. When checked, the
Volume field is enabled.
Needs different advanced parameters (optional): Check this box if the standard
needs advanced parameter(s) different than those defined in the Add Sample
step. When checked, the Advanced button is enabled.
Figure 167
Advanced aspiration
and dispense
3. Arrow buttons: Select a Standard in the list to the left and click the arrow for
the direction in which to move it in the list (reference Figure 130).
4. Remove: Select a Standard from the list on the left and click Remove to elim-
inate it from the list (see Figure 130).
8 QUICK GUIDE
If defined Special Interpretation Groups are in the Assay(s) being run, the
Interpretation Group window (Figure 168 Interpretation group) is displayed;
otherwise the Work Schedule (Figure 169 Work scheduler) is displayed. Using
Special Groups Interpretation, unique normal ranges are used for different
Group definitions.
Figure 168
Interpretation
group
116
When multiple Assays are run simultaneously, the software will automatically
generate the best run order based on the shortest total completion time. The
desired run order can be changed by selecting an alternate schedule. The
software will delay the start of certain assays so that conflicts do not occur.
Figure 169
Work scheduler
4. After reviewing the work schedule, click on OK. The Wash bottle selection
window (Figure 170 Select wash bottle) will be displayed. Select whether the
Assays will use the Wash or the Rinse bottle for the plate washing steps.
Click the OK button.
Figure 170
Select wash bottle
5. Place all the needed reagents and samples into the correct rack locations,
place the proper-coated ELISA strips into a plastic strip tray, and click Start
Run. The assay processing will begin. The time remaining for completion of
Figure 171
Test list
Right clicking on any of the Sample rows will display the Step Log (see Figure
172) for that sample. This can be helpful for verifying that all the Assay events
have occurred reasonably on time. For example: If a reagent runs out during
a test and has to be refilled, the timing in this window can help determine
whether to accept a result or delete it.
Figure 172
Step log
118
Check the Assay processing status by clicking on View Schedule button. This will
display an approximated time line of the Assay processing (see Figure 173 ELISA
work schedule).
Figure 173
ELISA work schedule
When Assays are complete the curve can be accepted (see Figure 174 Statistics).
Patient concentrations will be calculated. If controls are run in an Assay, the
Statistics can be displayed. Select the Controls with the left mouse button and
hold down the Ctrl key.
Figure 174
Statistics
Figure 175
Unedited curve
Clicking on the Test List tab will display the patient results (sample shown
below.)
Figure 176
Un-edited results
This data shows that the first copy of standard 5 appears incorrect. By checking
the boxes to the left of each standard, a corrected curve can be generated.
120
Figure 177
TSH results, edited
Figure 178
TSH results, activated
You can now choose to accept or delete any of the sample results by clicking on
the buttons. In Figure 179 Accepted results, all samples are selected.
Figure 179
Accept results
Figure 180
TSH Curve 0% Factor
122
The resulting concentration values for the samples run are shown below in
Figure 181.
Figure 181
TSH results
In Figure 182 TSH Curve -10% Factor below the calibration curve is adjusted by
-10%. Factor adjustments occur immediately and do not require clicking on the
Activate Selected button.
Figure 182
TSH curve -10% factor
The new concentration values are displayed in the Sample tab as shown below
in Figure 183 TSH Results:
Figure 183
TSH results
The percent adjust feature can be used when stored curves are used and there is
evidence that the current run has resulted in increased or decreased absorbance
levels than what is expected.
! Note: Decreasing the standard
curve by a percentage will re-
sult in increasing the sample con-
centration values.
C. Adjusting curves by running less than all the calibrator values:
Less than all the calibrators used in an assay can be run and the stored curve can
be adjusted accordingly. The adjustment factor will be calculated based on the
average percent change of all the new calibrators run compared to their stored
absorbance values.
The new curve will be generated from the new calibrator(s) absorbance values
which are currently run and the adjusted absorbance values of the remaining
calibrators from the stored curve. This feature can be used to control for changes
in reagent activity when using stored curves.
124
In the Figure 184 TSH Adjusted with Standard 5, TSH Standard 5 is used to adjust
the stored curve shown in Figure 180 TSH Curve 0% Factor. Instead of requesting
an entire new curve, Standard 5 is requested individually.
Figure 184
TSH Adjusted with Standard 5
Figure 185
TSH adjusted curve
Once activated the original absorbance values remain, but the new calculated
concentration values of the standards are displayed.
Figure 186
Activated TSH adjusted curve
Even though the samples show decreased absorbance values similar to the one
adjustment standard run, the resulting concentration values for the samples
are the same as when they were run with the original complete standard curve
(see Figure 181 TSH results).
126
Figure 187
TSH Standard 5 Results
9 TROUBLESHOOTING
9.1.1 Flags
Possible insufficient aspiration:
This flag may appear:
- If incorrect bottle sizes are used in the reagent rack.
- If reagent bottles are filled past the neck.
Volume calculation:
ELISYS UNO automatically detects liquid surfaces and makes approximate
volume calculations based upon the diameters of the rack cut-outs and the
distance between the detected surface and the bottom of the rack. Hanging
! Note: Conical containers and
containers narrower than the
rack openings will always result
containers that do not rest on the bottom of the rack will introduce error to the in volume calculation errors.
volume approximation if the container bottom level is not set to a new value.
These types of containers can only be used provided the new container bottom
level is properly set using Rack2 setup (see Chapter 3.1 ‘Instrument setup‘).
Sample volumes are usually so small that the vessel configuration is not a factor.
Reagent handling, however, can be adversely affected by significantly tapered or
under-sized vessels.
This is why: the probe detects the liquid surface at the start then ELISYS UNO
calculates the proper insertion depth for the probe so that the probe tip will
remain just below the liquid surface when the aspiration is finished.
128
010 Diluter not acknowledging Check the cable that plugs in the back
of the diluter
011 CSI/O Inactive Unable to communicate with coproc-
essor.
013 Timeout waiting for coprocessor Make sure that the serial cable is
message firmly connected to both the com-
puter and the instrument. Use the ca-
ble (and adapter, if needed) provided
with the instrument.
Make sure the ELISYS UNO is pow-
ered up.
014 Diluter not responding See Error Code 010
015 Timeout waiting for completion See Error Code 013
of last coprocessor command
016 Check reagent/sample level! Check levels
018 Probe sensor malfunction A problem exists with fluid sensing
circuitry.
019 Parameter checksum error See Error Code 013
020 Probe jammed while trying to Make sure you have the proper size
detect the liquid surface bottle (not too small or tapered).
Check that the cap has been removed
from the bottle.
Try adding more reagents. Check the
probe depth setting for the rack in In-
strument Setup.
022 Large syringe stroke error Check the assay programs to make
sure that the 2.5 ml maximum has
not been exceeded.
Select the Routines tab from the
ELISYS UNO Manager menu, and click
on the Wash Probe option to reset
the syringe position.
504 Plate X axis is jammed Check for physical obstructions
blocking the path of the plate in the
X direction.
505 Plate Y axis is jammed Check for physical obstructions
blocking the path of the plate in the
Y direction.
Figure 189
523 Channel blanks are not valid!! Under Settings, select Alignment,
click on Channel Blanks option. Run
Channel Blanks using the blanking
material provided with ELISYS UNO
and new clean and clear-flat bottom
microwells. Follow the instructions
as prompted.
If the information are not accessible via internet, they can be obtained free of
charge from your local distributor.
If you installed software or settings are not the latest one, please contact your
local distributor.
138
Fuse: For continued protection against risk of fire, replace fuse only with one
of the specified type and current rating. Disconnect equipment from supply
before replacing fuse.
- Read instructions: Please take the time to read this manual carefully before
using this instrument. Review the following safety precautions to avoid in-
jury and prevent damage to this instrument or any products connected to
it. To avoid potential hazards, use this instrument only as specified. For best
results, familiarise yourself with the instrument and its capabilities before
attempting any clinical diagnostic tests. Refer any questions to your instru-
ment service provider.
- Servicing: There are no user-serviceable parts inside the instrument. Refer
servicing to qualified service personnel. Use only factory-authorised parts.
Failure to do so may void the warranty.
- Wear protective apparel: Many diagnostic assays utilise materials that are
potential biohazards. Always wear protective apparel and eye protection
while using this instrument. Always operate this instrument with the aer-
osol shield lowered.
- Follow operating instructions: Do not use this instrument in a manner not
specified by the manual, or the protection provided by the instrument may
be impaired.
- Use proper power cord: Use only the power cord specified for this product
and certified for the country of use.
- Ground the product: This product is grounded through the grounding con-
ductor of the power cord. To avoid electric shock, the grounding conduc-
tor must be connected to earth ground. An optional method is to attach a
ground strap from the external grounding terminal on the rear panel of the
instrument to a suitable ground such as a grounded pipe or some metal sur-
face to earth ground.
- Observe all terminal ratings: To avoid fire or shock hazard, observe all rat-
ings and markings on the instrument. Consult this manual for further rat-
ings information before making connections to the instrument.
- Install as directed: Install the instrument on a sturdy, level surface capable
of safely supporting the instrument’s weight (45 kg). The mounting surface
should be free of vibrations. The instrument does not require fastening to
the bench top.
Danger
Causes severe skin burns and eye damage.
H314
142
P310
Immediately call a POISON CENTRE/ doctor.
P405
Store locked up.
P501
Dispose of contents/container in accordance with local/ regional/ national/
international regulations.
- Do not operate the instrument if the wash head probes are damaged.
- If the waste bottle is overturned during operation, set it upright. When the
run has finished, check that the filter has not become wet and replace it if
necessary. If the hydrophobic filter becomes wet due an overturned waste
bottle, it will be blocked. Continued use of the instrument with a blocked
filter will impair washer effectiveness and may result in damage to the in-
strument.
- Do not fill reagent bottles past the neck. Doing so may cause the system to
inadvertently aspirate air.
- Do not fill wash or rinse bottles into the neck to prevent fluid from entering
the pressure tube.
- The wash and rinse bottles are pressurised during normal operation and the
waste bottle is under vacuum.
- Do not remove bottle caps or tubing connections while the bottles are pres-
surised/ evacuated.
- Turn the instrument off or click Pause Engine in the Management menu be-
fore adding more solution, changing bottles or connecting tubing.
The quality of washing often affects the validity of test results. To assure
adequate washing follow these precautions:
- Perform End of the day to clean probe with bleach and wash head with H2O.
- Handle and store the wash head carefully to prevent damage.
- Use the prime cycle before each wash.
144