COA - Potassium Iodate
COA - Potassium Iodate
COA - Potassium Iodate
Metrological traceability: Directly traceable to the suitable primary standard NIST SRM Potassium
dichromate 136f.
Measurement method: The certified mass fraction was determined by potentiometric titration with
sodium thiosulfate as titration solution. The certified value is based on a
molecular mass M = 213.995 g/mol dried substance.
Intended use: This volumetric standard is intended for standardisation of volumetric solutions
related to the chapter reagents of the Pharmacopoeia.
Instructions for handling The volumetric standard Potassium iodate must be dried at 130 °C for 2 hours
and correct use: before use.
By within-unit homogeneity studies a minimum weigh-in quantity of 100 mg was
determined.
Accreditation: Merck KGaA, Darmstadt, Germany is accredited by the German accreditation
authority DAkkS as registered reference material producer D-RM-15185-01-00
in accordance with ISO 17034 and registered calibration laboratory
D-K-15185-01-00 according to DIN EN ISO/IEC 17025.
Certificate issue date: September 3, 2019
Associated uncertainty:
The associated uncertainty UCRM reported with the certified values is calculated as combined expanded uncertainty
UCRM=k·uCRM in accordance with GUM and EA-4/02, with k=2 as the coverage factor for a 95% coverage probability.
The combined uncertainty uCRM is derived from combination of the squared uncertainty contributions:
u CRM = u 2 Characteri sation + u 2 Homogeneit y + u 2 Stability
ucharacterisation: is the uncertainty in accordance with DIN EN ISO/IEC 17025 which includes the
contributions of the primary reference material and the measuring system.
uhomogeneity: is the between-bottle variation in accordance with ISO 17034. The assessment
of homogeneity is performed by analysis of a representative number of
systematically chosen sample units.
ustability: is the uncertainty obtained from short-term and long-term stability in accordance
with ISO 17034. The stability studies are the basis for the quantification of the
expiry date of this volumetric standard for the unopened bottle.
For more detailed information please read the certification report on our website.
The vibrant M, Supelco, Certipur and Sigma-Aldrich are trademarks of Merck KGaA, Darmstadt, Germany or its affiliates.
Detailed information on trademarks is available via publicly accessible resources.
© 2018 Merck KGaA, Darmstadt, Germany and/or its affiliates. All Rights Reserved.