Promass F
Promass F
Promass F
Technical Information
Proline Promass F 100
Coriolis flowmeter
Table of contents
2 Endress+Hauser
Proline Promass F 100
Endress+Hauser 3
Proline Promass F 100
Symbol Meaning
Direct current
Alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
A Reference to page.
Reference to graphic.
Visual inspection.
Symbols in graphics
Symbol Meaning
1. , 2. , 3. , … Series of steps
A, B, C, ... Views
Hazardous area
-
Safe area (non-hazardous area)
.
Flow direction
4 Endress+Hauser
Proline Promass F 100
Fc = 2 · ∆m (ν · ω)
Fc = Coriolis force
∆m = moving mass
ω = rotational velocity
The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity ν in the system and
thus on the mass flow. Instead of a constant rotational velocity ω, the sensor uses oscillation.
In the sensor, two parallel measuring tubes containing flowing fluid oscillate in antiphase, acting like
a tuning fork. The Coriolis forces produced at the measuring tubes cause a phase shift in the tube
oscillations (see illustration):
• At zero flow (when the fluid is at a standstill) the two tubes oscillate in phase (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the
outlet (3).
1 2 3
A0028850
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the
tube oscillations at the inlet and outlet. System balance is ensured by the antiphase oscillation of the
two measuring tubes. The measuring principle operates independently of temperature, pressure,
viscosity, conductivity and flow profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and
thus the density of the oscillating system (comprising measuring tube and fluid) results in a
corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a
function of medium density. The microprocessor utilizes this relationship to obtain a density signal.
Volume measurement
Together with the measured mass flow, this is used to calculate the volume flow.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor
due to temperature effects. This signal corresponds to the process temperature and is also available
as an output signal.
Endress+Hauser 5
Proline Promass F 100
Measuring system The device consists of a transmitter and a sensor. If a device with Modbus RS485 intrinsically safe is
ordered, the Safety Barrier Promass 100 is part of the scope of supply and must be implemented to
operate the device.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
Transmitter
Sensor
6 Endress+Hauser
Proline Promass F 100
Equipment architecture
1
2 3 4 5
10
11
6 7 8
A0016779
Safety IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.
Endress+Hauser 7
Proline Promass F 100
Input
Measured variable Direct measured variables
• Mass flow
• Density
• Temperature
15 ½ 0 to 6 500 0 to 238.9
25 1 0 to 18 000 0 to 661.5
40 1½ 0 to 45 000 0 to 1 654
50 2 0 to 70 000 0 to 2 573
DN x
8 ³⁄₈ 60
15 ½ 80
25 1 90
40 1½ 90
50 2 90
80 3 110
100 4 130
8 Endress+Hauser
Proline Promass F 100
DN x
150 6 200
250 10 200
To calculate the measuring range, use the Applicator sizing tool → 101
Output
Output signal HART current output
Load 0 to 700 Ω
Resolution 0.38 µA
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Pulse output
Endress+Hauser 9
Proline Promass F 100
Frequency output
The range of options increases if the measuring device has one or more
application packages.
Switch output
The range of options increases if the measuring device has one or more
application packages.
PROFIBUS DP
Modbus RS485
Terminating resistor • For device version used in non-hazardous areas or Zone 2/Div. 2: integrated
and can be activated via DIP switches on the transmitter electronics module
• For device version used in intrinsically safe areas: integrated and can be
activated via DIP switches on the Safety Barrier Promass 100
10 Endress+Hauser
Proline Promass F 100
EtherNet/IP
PROFINET
Current output 4 to 20 mA
4 to 20 mA
Failure mode Choose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
Pulse/frequency/switch output
Pulse output
Frequency output
Switch output
PROFIBUS DP
Modbus RS485
EtherNet/IP
Endress+Hauser 11
Proline Promass F 100
PROFINET
Device diagnostics According to "Application Layer protocol for decentralized periphery", Version 2.3
Local display
Interface/protocol
• Via digital communication:
• HART protocol
• PROFIBUS DP
• Modbus RS485
• EtherNet/IP
• PROFINET
• Via service interface
CDI-RJ45 service interface
Web server
Ex connection data These values only apply for the following device version:
Order code for "Output", option M "Modbus RS485", for use in intrinsically safe areas
Safety-related values
Terminal numbers
Unom = DC 24 V Unom = DC 5 V
Umax = AC 260 V Umax = AC 260 V
12 Endress+Hauser
Proline Promass F 100
Uo = 16.24 V
Io = 623 mA
Po = 2.45 W
With IIC 1): Lo = 92.8 µH, Co = 0.433 μF, Lo/Ro = 14.6 μH/Ω
With IIB: Lo= 372 µH, Co = 2.57 μF, Lo/Ro = 58.3 μH/Ω
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
1) The gas group depends on the sensor and nominal diameter ff.
Transmitter
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Low flow cut off The switch points for low flow cut off are user-selectable.
Manufacturer ID 0x11
Endress+Hauser 13
Proline Promass F 100
The range of options increases if the measuring device has one or more
application packages.
Heartbeat Technology application package
Additional measured variables are available with the Heartbeat Technology
application package:
• Carrier pipe temperature
• Oscillation amplitude 0
PROFIBUS DP
Manufacturer ID 0x11
14 Endress+Hauser
Proline Promass F 100
Modbus RS485
Endress+Hauser 15
Proline Promass F 100
Data access Each device parameter can be accessed via Modbus RS485.
EtherNet/IP
Manufacturer ID 0x49E
Baud rates Automatic ¹⁰⁄₁₀₀ Mbit with half-duplex and full-duplex detection
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Configuration options for • DIP switches on the electronics module for IP addressing
measuring device • Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• Electronic Data Sheet (EDS) integrated in the measuring device
Configuration of the EtherNet • Speed: 10 MBit, 100 MBit, auto (factory setting)
interface • Duplex: half-duplex, full-duplex, auto (factory setting)
Configuration of the device • DIP switches on the electronics module for IP addressing (last octet)
address • DHCP
• Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
16 Endress+Hauser
Proline Promass F 100
Fix Input
O → T configuration: 0x66 64
T → O configuration: 0x64 44
O → T configuration: 0x66 64
T → O configuration: 0x64 44
O → T configuration: 0xC7 -
T → O configuration: 0x64 44
O → T configuration: 0xC7 -
T → O configuration: 0x64 44
Configurable Input
O → T configuration: 0x66 64
T → O configuration: 0x65 88
O → T configuration: 0x66 64
T → O configuration: 0x65 88
O → T configuration: 0xC7 -
T → O configuration: 0x65 88
O → T configuration: 0xC7 -
T → O configuration: 0x65 88
Endress+Hauser 17
Proline Promass F 100
Fix Output
Configuration
Configuration Assembly Only the most common configurations are listed below.
• Software write protection
• Mass flow unit
• Mass unit
• Volume flow unit
• Volume unit
• Corrected volume flow unit
• Corrected volume unit
• Density unit
• Reference density unit
• Temperature unit
• Pressure unit
• Length
• Totalizer 1-3:
• Assignment
• Unit
• Operating mode
• Failsafe mode
• Alarm delay
PROFINET
Protocol "Application layer protocol for decentral device periphery and distributed
automation", version 2.3
Conformity class B
Manufacturer ID 0x11
18 Endress+Hauser
Proline Promass F 100
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Configuration options for • DIP switches on the electronics module, for device name assignment (last
measuring device part)
• Manufacturer-specific software (FieldCare, DeviceCare)
• Web browser
• Device master file (GSD), can be read out via the integrated Web server of
the measuring device
Configuration of the device • DIP switches on the electronics module, for device name assignment (last
name part)
• DCP protocol
The range of options increases if the measuring device has one or more
application packages.
Endress+Hauser 19
Proline Promass F 100
The range of options increases if the measuring device has one or more
application packages.
Volume flow
Density
Reference density
Temperature
Electronic temperature
Oscillation frequency
Frequency fluctuation
Oscillation damping
Oscillation frequency
Signal asymmetry
Exciter current
Concentration
20 Endress+Hauser
Proline Promass F 100
Oscillation damping 1
Oscillation frequency 1
Oscillation amplitude 0
Oscillation amplitude 1
Frequency fluctuation 1
Exciter current 1
External temperature 19
Flow override 21
Endress+Hauser 21
Proline Promass F 100
Startup configuration
Startup configuration If startup configuration is enabled, the configuration of the most important
(NSU) device parameters is taken from the automation system and used.
The following configuration is taken from the automation system:
• Management
• Software revision
• Write protection
• System units
• Mass flow
• Mass
• Volume flow
• Volume
• Corrected volume flow
• Corrected volume
• Density
• Reference density
• Temperature
• Pressure
• Concentration application package
• Coefficients A0 to A4
• Coefficients B1 to B3
• Sensor adjustment
• Process parameter
• Damping (flow, density, temperature)
• Flow override
• Low flow cut off
• Assign process variable
• Switch-on/switch-off point
• Pressure shock suppression
• Empty pipe detection
• Assign process variable
• Limit values
• Response time
• Max. damping
• Corrected volume flow calculation
• External reference density
• Fixed reference density
• Reference temperature
• Linear expansion coefficient
• Square expansion coefficient
• Measuring mode
• Medium
• Gas type
• Reference sound velocity
• Temperature coefficient sound velocity
• External compensation
• Pressure compensation
• Pressure value
• External pressure
• Diagnostic settings
• Diagnostic behavior for diverse diagnostic information
22 Endress+Hauser
Proline Promass F 100
Power supply
Terminal assignment Overview: housing version and connection versions
1 2
3 4
A0016770
Transmitter
Endress+Hauser 23
Proline Promass F 100
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Options Device plugs Terminals • Option L: plug M12x1 + thread NPT ½"
A, B → 31 • Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
24 +
_ 3
25
26 +
_ 2
27
1 L+
1
2 L_
A0016888
Terminal number
Order code
Power supply Output 1 Output 2
"Output"
2 (L-) 1 (L+) 27 (–) 26 (+) 25 (–) 24 (+)
24 Endress+Hauser
Proline Promass F 100
Options Device plug Terminals • Option L: plug M12x1 + thread NPT ½"
A, B connectors • Option N: plug M12x1 + coupling M20
→ 31 • Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
26 B
27 A 2
1 L+
1
2 L_
A0022716
Terminal number
Option L DC 24 V B A
Endress+Hauser 25
Proline Promass F 100
Options Device plugs Terminals • Option L: plug M12x1 + thread NPT ½"
A, B → 31 • Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
26 B
2
27 A
1 L+
1
2 L_
A0019528
4 Modbus RS485 terminal assignment, connection version for use in non-hazardous areas and Zone 2/Div.
2
1 Power supply: DC 24 V
2 Modbus RS485
Terminal number
Order code
Power supply Output
"Output"
1 (L+) 2 (L-) 26 (B) 27 (A)
26 Endress+Hauser
Proline Promass F 100
62 B
2
72 A
10 L+
1
20 L_
A0030219
5 Modbus RS485 terminal assignment, connection version for use in intrinsically safe areas (connection via
Safety Barrier Promass 100)
1 Intrinsically safe power supply
2 Modbus RS485
Order code
10 (L+) 20 (L-) 62 (B) 72 (A)
"Output"
Endress+Hauser 27
Proline Promass F 100
Options Device plug Terminals • Option L: plug M12x1 + thread NPT ½"
A, B connectors • Option N: plug M12x1 + coupling M20
→ 32 • Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
1 L+
1
2 L_
A0017054
Terminal number
Order code
Power supply Output
"Output"
2 (L-) 1 (L+) Device plug M12x1
Option N DC 24 V EtherNet/IP
28 Endress+Hauser
Proline Promass F 100
Options Device plug Terminals • Option L: plug M12x1 + thread NPT ½"
A, B connectors • Option N: plug M12x1 + coupling M20
→ 30 • Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
1 L+
1
2 L_
A0017054
Terminal number
Order code
Power supply Output
"Output"
2 (L-) 1 (L+) Device plug M12x1
Option R DC 24 V PROFINET
Endress+Hauser 29
Proline Promass F 100
2 1 26 27
L L+ B A
Power Modbus
supply RS485
1 24VDC
Safe area
Power
Lift panel for
bus termination
Communication
A
Safety Barrier
Promass 100
Hazardous area
A Power Modbus
supply RS485
L L+ B A
20 10 62 72
2
A0030220
Pin assignment, device plug Order codes for the M12x1 connectors, see the "Order code for electrical connection" column:
• 4-20 mA HART, pulse/frequency/switch output → 23
• PROFIBUS DP→ 25
• Modbus RS485 → 26
• EtherNet/IP → 28
• PROFINET→ 29
Supply voltage
For all connection versions except MODBUS RS485 intrinsically safe (device side)
Device plug MODBUS RS485 intrinsically safe with supply voltage → 31
2 Pin Assignment
1 L+ DC 24 V
2 Not assigned
3 1
3 Not assigned
5 4 L- DC 24 V
4
A0029042
5 Grounding/shielding
Coding Plug/socket
A Plug
30 Endress+Hauser
Proline Promass F 100
2 Pin Assignment
Coding Plug/socket
A Socket
•• When
Recommended plug: Binder, series 763, part no. 79 3439 12 05
using the device in a hazardous location, use a suitably certified plug.
PROFIBUS DP
For use in the non-hazardous area and Zone 2/Div. 2.
2 Pin Assignment
1 Not assigned
2 A PROFIBUS DP
1 3
3 Not assigned
5 4 B PROFIBUS DP
4
A0016811
5 Grounding/shielding
Coding Plug/socket
B Socket
•• When
Recommended plug: Binder, series 763, part no. 79 4449 20 05
using the device in a hazardous location, use a suitably certified plug.
MODBUS RS485
Device plug for signal transmission with supply voltage (device side), MODBUS RS485 (intrinsically
safe)
2 Pin Assignment
2 A
3 1 Modbus RS485 intrinsically safe
3 B
5 4 L- Supply voltage, intrinsically safe
4
A0029042
5 Grounding/shielding
Coding Plug/socket
A Plug
Endress+Hauser 31
Proline Promass F 100
Device plug for signal transmission (device side), MODBUS RS485 (not intrinsically safe)
For use in the non-hazardous area and Zone 2/Div. 2.
2 Pin Assignment
1 Not assigned
2 A Modbus RS485
1 3
3 Not assigned
5 4 B Modbus RS485
4
A0016811
5 Grounding/shielding
Coding Plug/socket
B Socket
•• When
Recommended plug: Binder, series 763, part no. 79 4449 20 05
using the device in a hazardous location, use a suitably certified plug.
EtherNet/IP
2 Pin Assignment
1 + Tx
2 + Rx
1 3
3 - Tx
4 - Rx
4
A0016812
Coding Plug/socket
D Socket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
PROFINET
2 Pin Assignment
1 + TD +
2 + RD +
1 3
3 - TD –
4 - RD –
4
A0016812
Coding Plug/socket
D Socket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
Supply voltage The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
32 Endress+Hauser
Proline Promass F 100
Transmitter
For device version with communication type:
• HART, PROFIBUS DP, EtherNet/IP: DC 20 to 30 V
• Modbus RS485, device version:
• For use in the non-hazardous area and Zone 2/Div. 2: DC 20 to 30 V
• For use in the intrinsically safe area: power supply via Safety Barrier Promass 100
Maximum
Order code for "Output"
Power consumption
Maximum Maximum
Order code for "Output"
Current consumption switch-on current
Maximum Maximum
Order code for "Output"
Current consumption switch-on current
Power supply failure Depending on the device version, the configuration is retained in the device memoryor in the
pluggable data memory (HistoROM DAT).
Endress+Hauser 33
Proline Promass F 100
A B C
1 2 1 2 3 4
A0016924
• Terminal assignment→ 23
• Pin assignment, device plug→ 30
In the case of device versions with a connector, the transmitter housing does not need to be
opened to connect the signal cable or power supply cable.
Connection examples
1 2 3
4...20 mA
4 5
A0029055
34 Endress+Hauser
Proline Promass F 100
1 2 3 4
4...20 mA
A0028762
Pulse/frequency output
1 2
12345
A0028761
Switch output
1 2
A0028760
Endress+Hauser 35
Proline Promass F 100
PROFIBUS DP
1 2 3
A A
4
B
B
A
4
B
A0028765
13 Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2
1 Control system (e.g. PLC)
2 Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3 Transmitter
If baud rates > 1.5 MBaud an EMC cable entry must be used and the cable shield must continue
as far as the terminal wherever possible.
Modbus RS485
1 2 3
A A
4
B
B
A
4
B
A0028765
14 Connection example for Modbus RS485, non-hazardous area and Zone 2/Div. 2
1 Control system (e.g. PLC)
2 Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications → 39
3 Distribution box
4 Transmitter
36 Endress+Hauser
Proline Promass F 100
1 2 3 4
A A
A A B B
L+ 8
B L+
B
L- L+ L- L-
5 6 7
A0028766
EtherNet/IP
1 2 3 4
A0028767
Endress+Hauser 37
Proline Promass F 100
PROFINET
1 2 3 4
A0016805
HART input
1 2 3 4
4...20 mA
+
5
-
-
+ 6
A0019828
18 Connection example for HART input (burst mode) via current output (active)
1 Cable shield, observe cable specifications
2 Resistor for HART communication (≥ 250 Ω): observe maximum load
3 Connection for HART operating devices
4 Analog display unit
5 Transmitter
6 Sensor for external measured variable
38 Endress+Hauser
Proline Promass F 100
1 2 3 4 5
4...20 mA
+
+ 6
-
-
-
+
-
7
+
4...20 mA
2 3 4 5
A0019830
19 Connection example for HART input (master mode) via current output (active)
1 Automation system with current input (e.g. PLC).
Prerequisite: automation system with HART version 6, HART commands 113 and 114 can be processed.
2 Cable shield, observe cable specifications
3 Resistor for HART communication (≥ 250 Ω): observe maximum load
4 Connection for HART operating devices
5 Analog display unit
6 Transmitter
7 Sensor for external measured variable
Terminals Transmitter
Spring terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Promass 100 safety barrier
Plug-in screw terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Cable entries • Cable gland: M20 × 1.5 with cable Ø 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
• M20
• G ½"
• NPT ½"
Signal cable
Pulse/frequency/switch output
Standard installation cable is sufficient.
Endress+Hauser 39
Proline Promass F 100
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Cable type A
Signal damping Max. 9 dB over the entire length of the cable cross-section
Shield Copper braided shielding or braided shielding with foil shield. When grounding
the cable shield, observe the grounding concept of the plant.
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
Cable type A
Signal damping Max. 9 dB over the entire length of the cable cross-section
Shield Copper braided shielding or braided shielding with foil shield. When grounding
the cable shield, observe the grounding concept of the plant.
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
PROFINET
Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT
5e and CAT 6 are recommended.
For more information on planning and installing PROFINET networks, see: "PROFINET Cabling
and Interconnection Technology", Guideline for PROFINET
Connecting cable between Safety Barrier Promass 100 and measuring device
Cable type Shielded twisted-pair cable with 2x2 wires. When grounding the cable shield,
observe the grounding concept of the plant.
Comply with the maximum cable resistance specifications to ensure the operational reliability
of the measuring device.
40 Endress+Hauser
Proline Promass F 100
The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
0.5 20 70 230
Performance characteristics
Reference operating • Error limits based on ISO 11631
conditions • Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool → 101
Base accuracy
Design fundamentals → 44
Density (liquids)
1) Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
2) Order code for "Application package", option EE "Special density"
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Endress+Hauser 41
Proline Promass F 100
15 ½ 0.200 0.007
25 1 0.540 0.019
40 1½ 2.25 0.083
50 2 3.50 0.129
80 3 9.0 0.330
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
250 2 200 000 220 000 110 000 44 000 22 000 4 400
US units
42 Endress+Hauser
Proline Promass F 100
Accuracy of outputs
The output accuracy must be factored into the measured error if analog outputs are used, but
can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
Current output
Accuracy Max. ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range)
Base repeatability
Design fundamentals → 44
Density (liquids)
±0.00025 g/cm3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Pulse/frequency output
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is
±0.00005 g/cm3 /°C (±0.000025 g/cm3 /°F). Field density calibration is possible.
Endress+Hauser 43
Proline Promass F 100
3
[kg/m ]
18
16
14
12 1
10
8
2
6
4
2
0
-50 0 100 200 300 [°C]
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
Influence of medium The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure pressure and process pressure.
o.r. = of reading
It is possible to compensate for the effect by:
• Reading in the current pressure measured value via the current input.
• Specifying a fixed value for the pressure in the device parameters.
Operating Instructions .
DN [% o.r./bar] [% o.r./psi]
[mm] [in]
8 3/8 no influence
15 ½ no influence
25 1 no influence
40 1½ –0.003 –0.0002
50 2 –0.008 –0.0006
80 3 –0.009 –0.0006
44 Endress+Hauser
Proline Promass F 100
ZeroPoint
³ BaseAccu ⋅ 100 ± BaseAccu
A0021339
A0021332
ZeroPoint ZeroPoint
< BaseAccu ⋅ 100 MeasValue ⋅ 100
±
A0021333 A0021334
½ ⋅ ZeroPoint
³ BaseRepeat ⋅ 100
A0021340
A0021335
½ ⋅ ZeroPoint ZeroPoint
< BaseRepeat ⋅ 100 ± ½ ⋅ MeasValue ⋅ 100
A0021336 A0021337
E [%]
2.5
2.0
1.5
1.0
0.5
0
0 10 20 30 40 50 60 70 80 90 100 Q [%]
A0028808
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
Mounting location
A0028772
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Endress+Hauser 45
Proline Promass F 100
A0028773
8 ³⁄₈ 6 0.24
15 ½ 10 0.40
25 1 14 0.55
40 1½ 22 0.87
50 2 28 1.10
80 3 50 1.97
100 4 65 2.60
150 6 90 3.54
Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
flow direction (direction of medium flow through the piping).
Orientation Recommendation
A Vertical orientation
A0015591
1)
B Horizontal orientation, transmitter at
top Exceptions:
→ 21, 47
A0015589
46 Endress+Hauser
Proline Promass F 100
Orientation Recommendation
2)
C Horizontal orientation, transmitter at
bottom Exceptions:
→ 21, 47
A0015590
1) Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
2) Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to
the fluid properties.
1 2
A0028774
Inlet and outlet runs No special precautions need to be taken for fittings which create turbulence, such as valves, elbows
or T-pieces, as long as no cavitation occurs → 57.
Endress+Hauser 47
Proline Promass F 100
RUPTURE DISK
2 3 1 2 3
A0028903
For information on the dimensions: see the "Mechanical construction -> Accessories" section
A0016894
22 Minimum distance between additional Safety Barrier Promass 100 or other modules. Engineering unit
mm (in)
48 Endress+Hauser
Proline Promass F 100
Environment
Ambient temperature range Measuring device • –40 to +60 °C (–40 to +140 °F)
• Order code for "Test, certificate", option JM:
–50 to +60 °C (–58 to +140 °F)
‣ If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) (standard version)
–50 to +80 °C (–58 to +176 °F) (Order code for "Test, certificate", option JM)
Endress+Hauser 49
Proline Promass F 100
Process
Medium temperature range
Standard version –50 to +150 °C (–58 to +302 °F) Order code for "Measuring tube
mat., wetted surface", option HA,
SA, SB, SC
Extended temperature version –50 to +240 °C (–58 to +464 °F) Order code for "Measuring tube
mat., wetted surface", option SD, SE,
SF, TH
Ta
A
B
Tm
A0031121
A B A B
Version Ta Tm Ta Tm Ta Tm Ta Tm
Standard version 60 °C (140 °F) 150 °C (302 °F) – – 60 °C (140 °F) 110 °C (230 °F) 55 °C (131 °F) 150 °C (302 °F)
Extended 60 °C (140 °F) 160 °C (320 °F) 55 °C (131 °F) 240 °C (464 °F) 60 °C (140 °F) 110 °C (230 °F) 50 °C (122 °F) 240 °C (464 °F)
temperature
version
Pressure-temperature The following pressure/temperature diagrams apply to all pressure-bearing parts of the device and
ratings not just the process connection. The diagrams show the maximum permissible medium pressure
depending on the specific medium temperature.
Pressure-temperature curves with temperature range +151 to +240 °C (+304 to +464 °F)
exclusively for extended temperature version of measuring devices.
50 Endress+Hauser
Proline Promass F 100
[psi] [bar]
100
1400 PN 100
90
1200 80
1000 70
60 PN 63
800
50
600 40
PN 40
400 30
20
200
10 PN 16
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
100
1400 Class 600
90
1200 80
1000 70
60
800
50
Class 300
600 40
400 30
20
200 Class 150
10
0 0
-50 0 50 100 150 200 250[°C]
Endress+Hauser 51
Proline Promass F 100
[psi] [bar]
100 Class 600
1400
90
1200 80
1000 70
60
800
50
Class 300
600 40
400 30
20
200 Class 150
10
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
1200 70
60 63K
800
50
600 40
40K
400 30
20
200 20K
10
10K
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
50
600 40
400 30 DN 8…40
20
200
10 ≥ DN 50
0 0
-50 0 50 100 150 200 250 [°C]
52 Endress+Hauser
Proline Promass F 100
[psi] [bar]
50
600 40
PN 40
400 30
20
200
10
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
100
1400 Class 600
90
1200 80
1000 70
60
800 Class 300
50
600 40
400 30
Class 150
20
200
10
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
400 30
20
200 20K
10
0 0
-50 0 50 100 150 200 250 [°C]
Endress+Hauser 53
Proline Promass F 100
[psi] [bar]
50
600 40
DN 8…40
400 30
20 ≥ DN 50
200
10
0 0
-50 0 50 100 150 200 250 [°C]
DIN 11851 allows for applications up to +140 °C (+284 °F) if suitable sealing materials are used.
Please take this into account when selecting seals and counterparts, as these components can limit
the pressure and temperature range.
[psi] [bar]
50
600 40
DN 8…40
400 30
20 ≥ DN 50
200
10
0 0
-50 0 50 100 150 200 250 [°C]
[psi] [bar]
400
20
200
10
0 0
-50 0 50 100 150 200 250 [°C]
54 Endress+Hauser
Proline Promass F 100
[psi] [bar]
400
20
200
10
0 0
-50 0 50 100 150 200 250 [°C]
SMS 1145 allows for applications up to 16 bar (232 psi) if suitable sealing materials are used. Please
take this into account when selecting seals and counterparts, as these components can limit the
pressure and temperature range.
VCO
[psi] [bar]
100
1400
90
1200 80
1000 70
60
800
-50 0 50 100 150 200 250 [°C]
Tri-Clamp
[psi] [bar]
400
20
200
10
0 0
-50 0 50 100 150 200 250 [°C]
The clamp connections are suitable up to a maximum pressure of 16 bar (232 psi). Please observe
the operating limits of the clamp and seal used as they can be over 16 bar (232 psi). The clamp and
seal are not included in the scope of supply.
Sensor housing
For standard versions with the temperature range –50 to +150 °C (–58 to +302 °F), the sensor
housing is filled with dry nitrogen gas and protects the electronics and mechanics inside.
For all other temperature versions the sensor housing is filled with dry inert gas.
If a measuring tube fails (e.g. due to process characteristics like corrosive or abrasive fluids), the
fluid will initially be contained by the sensor housing.
Endress+Hauser 55
Proline Promass F 100
In the event of a tube failure, the pressure level inside the sensor housing will rise according to the
operating process pressure. If the user judges that the sensor housing burst pressure does not
provide an adequate safety margin, the device can be fitted with a rupture disk. This prevents
excessively high pressure from forming inside the sensor housing. Therefore, the use of a rupture
disk is strongly recommended in applications involving high gas pressures, and particularly in
applications in which the process pressure is greater than 2/3 of the sensor housing burst pressure.
If there is a need to drain the leaking medium into a discharge device, the sensor should be fitted
with a rupture disk. Connect the discharge to the additional threaded connection → 78.
If the sensor is to be purged with gas (gas detection), it should be equipped with purge connections.
Do not open the purge connections unless the containment can be filled immediately with a
dry, inert gas. Use only low pressure to purge.
Maximum pressure:
• DN 08 to 150 (3/8 to 6"): 5 bar (72.5 psi)
• DN 250 (10"):
• Medium temperature ≤ 100 °C (212 °F): 5 bar (72.5 psi)
• Medium temperature > 100 °C (212 °F): 3 bar (43.5 psi)
The following sensor housing burst pressures are only valid for standard devices and/or devices
equipped with closed purge connections (not opened/as delivered).
If a device fitted with purge connections (order code for "Sensor option", option CH "Purge
connection") is connected to the purge system, the maximum pressure is determined by the purge
system itself or by the device, depending on which component has the lower pressure classification.
If the device is fitted with a rupture disk (order code for "Sensor option", option CA "Rupture disk"), the
rupture disk trigger pressure is decisive .
The sensor housing burst pressure refers to a typical internal pressure which is reached prior to
mechanical failure of the sensor housing and which was determined during type testing. The
corresponding type test declaration can be ordered with the device (order code for "Additional
approval", option LN "Sensor housing burst pressure, type test").
15 ½ 350 5 070
25 1 280 4 060
40 1½ 260 3 770
50 2 180 2 610
80 3 120 1 740
100 4 95 1 370
150 6 75 1 080
250 10 50 720
Rupture disk To increase the level of safety, a device version with a rupture disk with a trigger pressure of
10 to 15 bar (145 to 217.5 psi)can be used (order code for "Sensor option", option CA "rupture disk").
The use of rupture disks cannot be combined with the separately available heating jacket.
For information on the dimensions: see the "Mechanical construction" section (accessories) → 78
56 Endress+Hauser
Proline Promass F 100
Flow limit Select the nominal diameter by optimizing between the required flow range and permissible pressure
loss.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section → 8
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):
flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
• The flow velocity in the measuring tubes should not exceed half the sound velocity (0.5 Mach).
• The maximum mass flow depends on the density of the gas: formula → 8
To calculate the flow limit, use the Applicator sizing tool → 101
Pressure loss To calculate the pressure loss, use the Applicator sizing tool → 101
Promass F with reduced pressure loss: order code for "Sensor option", option CE "Reduced pressure
loss"
System pressure It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0028777
Thermal insulation In the case of some fluids, it is important to keep the heat radiated from the sensor to the
transmitter to a low level. A wide range of materials can be used for the required insulation.
The following device versions are recommended for versions with thermal insulation:
• Version with extended neck for insulation:
Order code for "Sensor option", option CG with an extended neck length of 105 mm (4.13 in).
• Extended temperature version:
Order code for "Measuring tube material", option SD, SE, SF or TH with an extended neck length of
105 mm (4.13 in).
NOTICE
Electronics overheating on account of thermal insulation!
‣ Recommended orientation: horizontal orientation, transmitter housing pointing downwards.
‣ Do not insulate the transmitter housing .
‣ Maximum permissible temperature at the lower end of the transmitter housing: 80 °C (176 °F)
‣ Thermal insulation with extended neck free: We recommend that you do not insulate the
extended neck in order to ensure optimum dissipation of heat.
A0034391
Endress+Hauser 57
Proline Promass F 100
Heating Some fluids require suitable measures to avoid loss of heat at the sensor.
Heating options
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Heating jackets for the sensors can be ordered as accessories from Endress+Hauser. → 100
NOTICE
Danger of overheating when heating
‣ Ensure that the temperature at the lower end of the transmitter housing does not exceed
80 °C (176 °F).
‣ Ensure that sufficient convection takes place at the transmitter neck.
‣ Ensure that a sufficiently large area of the transmitter neck remains exposed. The uncovered part
serves as a radiator and protects the electronics from overheating and excessive cooling.
‣ When using in potentially explosive atmospheres, observe the information in the device-specific
Ex documentation. For detailed information on the temperature tables, see the separate
document entitled "Safety Instructions" (XA) for the device.
Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
58 Endress+Hauser
Proline Promass F 100
Mechanical construction
Dimensions in SI units Compact version
A G
B C
K
D
L M
A0033787
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
4)
8 148 94 54 75 185 260 136 5.35 70
4)
15 148 94 54 75 185 260 136 8.30 70
4)
25 148 94 54 75 185 260 136 12.0 70
4)
40 148 94 54 105 189.5 294.5 136 17.6 79
4)
50 148 94 54 141 199.5 340.5 136 26.0 99
4)
80 148 94 54 200 219.5 419.5 136 40.5 139
4)
100 148 94 54 254 238 492 136 51.2 176
4)
150 148 94 54 378 259 637 136 68.9 218
4)
250 148 94 54 548 302.5 850.5 136 102.3 305
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
4)
8 137 78 59 75 180 255 134 5.35 70
4)
15 137 78 59 75 180 255 134 8.30 70
4)
25 137 78 59 75 180 255 134 12.0 70
4)
40 137 78 59 105 184.5 289.5 134 17.6 79
4)
50 137 78 59 141 194.5 335.5 134 26.0 99
4)
80 137 78 59 200 214.5 414.5 134 40.5 139
4)
100 137 78 59 254 233 487 134 51.2 176
Endress+Hauser 59
Proline Promass F 100
1) 1)
DN A B C D E 2) 3) F 2) 3) G K L M
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
4)
150 137 78 59 378 254 632 134 68.9 218
4)
250 137 78 59 548 297.5 845.5 134 102.3 305
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
4)
8 124 68 56 75 180 255 112 5.35 70
4)
15 124 68 56 75 180 255 112 8.30 70
4)
25 124 68 56 75 180 255 112 12.0 70
4)
40 124 68 56 105 184.5 289.5 112 17.6 79
4)
50 124 68 56 141 194.5 335.5 112 26.0 99
4)
80 124 68 56 200 214.5 414.5 112 40.5 139
4)
100 124 68 56 254 233 487 112 51.2 176
4)
150 124 68 56 378 254 632 112 68.9 218
4)
250 124 68 56 548 297.5 845.5 112 102.3 305
60 Endress+Hauser
Proline Promass F 100
Flange connections
C
A
B
E
D L
A0015621
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
1) Installation length in accordance with NAMUR recommendation NE 132 optionally available (order code
for "Process connection", option D1N or D5N (with groove))
Flange according to EN 1092-1 (DIN 2501): PN16 with reduction in nominal diameter
1.4404 (F316/F316L
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
Endress+Hauser 61
Proline Promass F 100
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
62 Endress+Hauser
Proline Promass F 100
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 µm
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 µm
Endress+Hauser 63
Proline Promass F 100
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Flange according to ASME B16.5: Class 150 with reduction in nominal diameter
1.4404 (F316/F316L)
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
64 Endress+Hauser
Proline Promass F 100
Flange according to ASME B16.5: Class 300 with reduction in nominal diameter
1.4404 (F316/F316L)
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Endress+Hauser 65
Proline Promass F 100
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
8 1) 95 70 4 × ⌀15 14 15 370
15 95 70 4 × ⌀15 14 15 404
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
66 Endress+Hauser
Proline Promass F 100
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
Endress+Hauser 67
Proline Promass F 100
X X
C
A
B
E
D
L
A0015627
38 Detail X: Asymmetrical process connection; the part shown in blue is provided by the supplier.
Flange DIN11864-2 Form A, for pipe according to DIN11866 series A, flat with notch
1.4404 (316/316L)
Order code for "Process connection", option KCS
DN A B C D E L
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
8 54 37 4 × ⌀9 10 10 387
15 59 42 4 × ⌀9 10 16 418
25 70 53 4 × ⌀9 10 26 454
40 82 65 4 × ⌀9 10 38 560
50 94 77 4 × ⌀9 10 50 720
3A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE or
Ramax = 0.38 µm: order code for "Measuring tube material", option SC, SF
Ramax = 0.38 µm electropolished: order code for "Measuring tube material", option BC
68 Endress+Hauser
Proline Promass F 100
C
A
B
E
F
D L
A0022221
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
D2C)
2) DN 8 with DN 15 flanges as standard
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
Endress+Hauser 69
Proline Promass F 100
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
2) DN 8 with DN 15 flanges as standard
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
ABC)
2) DN 8 with DN 15 flanges as standard
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
ACC)
2) DN 8 with DN 15 flanges as standard
70 Endress+Hauser
Proline Promass F 100
DN A B C D E F L Ldiff 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
8 2) 95 70 4 × ⌀ 15 14 51 15 370 0
15 95 70 4 × ⌀ 15 14 51 15 404 0
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
NEC)
2) DN 8 with DN 15 flanges as standard
Endress+Hauser 71
Proline Promass F 100
Clamp connections
Tri-Clamp
B
L
A0015625
DN Clamp A B L
[mm] [in] [mm] [mm] [mm]
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE or
Ramax = 0.38 µm: order code for "Measuring tube material", option SC, SF
Ramax = 0.38 µm electropolished: order code for "Measuring tube material", option BC
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE or
Ramax = 0.38 µm: order code for "Measuring tube material", option SC, SF
Ramax = 0.38 µm electropolished: order code for "Measuring tube material", option BC
72 Endress+Hauser
Proline Promass F 100
Threaded couplings
A
B
L
A0015628
DN A B L
[mm] [in] [mm] [mm]
8 Rd 34 × ¹⁄₈ 16 367
15 Rd 34 × ¹⁄₈ 16 398
25 Rd 52 × ¹⁄₆ 26 434
40 Rd 65 × ¹⁄₆ 38 560
50 Rd 78 × ¹⁄₆ 50 720
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE
DN A B L
[mm] [in] [mm] [mm]
8 Rd 28 × ¹⁄₈ 10 367
15 Rd 34 × ¹⁄₈ 16 398
25 Rd 52 × ¹⁄₈ 26 434
40 Rd 65 × ¹⁄₆ 38 560
50 Rd 78 × ¹⁄₆ 50 720
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE or
Ramax = 0.38 µm: order code for "Measuring tube material", option SC, SF
Ramax = 0.38 µm electropolished: order code for "Measuring tube material", option BC
Endress+Hauser 73
Proline Promass F 100
DN A B L
[mm] [in] [mm] [mm]
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE
74 Endress+Hauser
Proline Promass F 100
A
B
L
A0015623
DN A 1) B L
[mm] [mm] [mm] [mm]
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 0.76 µm: order code for "Measuring tube material", option SB, SE or
Ramax = 0.38 µm: order code for "Measuring tube material", option SC, SF
Ramax = 0.38 µm electropolished: order code for "Measuring tube material", option BC
Endress+Hauser 75
Proline Promass F 100
VCO
B
L
A0015624
8-VCO-4 (½")
1.4404 (316/316L)
Order code for "Process connection", option CVS
DN A B L
[mm] [in] [mm] [mm]
8 AF 1 10.2 390
12-VCO-4 (¾")
1.4404 (316/316L)
Order code for "Process connection", option CWS
DN A B L
[mm] [in] [mm] [mm]
15 AF 1½ 15.7 430
76 Endress+Hauser
Proline Promass F 100
C
B D
A0016777
A B C D
Endress+Hauser 77
Proline Promass F 100
Accessories
1 1;2 1 1;2
45°
A
L L A
½" NPT
½" NPT
1 2
A0028914
39
1 Connection nipple for purge connections: order code for "Sensor options", option CH "Purge connection"
2 Connection nipple with rupture disk: order code for "Sensor option", option CA "Rupture disk"
DN A L
8 62 216
15 62 220
25 62 260
40 67 310
50 79 452
80 101 560
78 Endress+Hauser
Proline Promass F 100
A G
B C
K
D
L M
A0033787
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
4)
³⁄₈ 5.83 3.70 2.13 2.95 7.28 10.24 5.35 0.211 2.76
4)
½ 5.83 3.70 2.13 2.95 7.28 10.24 5.35 0.33 2.76
4)
1 5.83 3.70 2.13 2.95 7.28 10.24 5.35 0.47 2.76
4)
1½ 5.83 3.70 2.13 4.13 7.46 11.59 5.35 0.69 3.11
4)
2 5.83 3.70 2.13 5.55 7.85 13.41 5.35 1.02 3.90
4)
3 5.83 3.70 2.13 7.87 8.64 16.52 5.35 1.59 5.47
4)
4 5.83 3.70 2.13 10 9.37 19.37 5.35 2.02 6.93
4)
6 5.83 3.70 2.13 14.88 10.2 25.08 5.35 2.71 8.58
4)
10 5.83 3.70 2.13 21.57 11.91 33.48 5.35 4.03 12.01
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
4)
³⁄₈ 5.39 3.07 2.32 2.95 7.09 10.04 5.28 0.211 2.76
4)
½ 5.39 3.07 2.32 2.95 7.09 10.04 5.28 0.33 2.76
4)
1 5.39 3.07 2.32 2.95 7.09 10.04 5.28 0.47 2.76
4)
1½ 5.39 3.07 2.32 4.13 7.26 11.4 5.28 0.69 3.11
4)
2 5.39 3.07 2.32 5.55 7.66 13.21 5.28 1.02 3.90
4)
3 5.39 3.07 2.32 7.87 8.44 16.32 5.28 1.59 5.47
4)
4 5.39 3.07 2.32 10 9.17 19.17 5.28 2.02 6.93
Endress+Hauser 79
Proline Promass F 100
DN A 1) B C D F 2) 3) F 2) 3) G K L M
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
4)
6 5.39 3.07 2.32 14.88 10 24.88 5.28 2.71 8.58
4)
10 5.39 3.07 2.32 21.57 11.71 33.29 5.28 4.03 12.01
[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]
4)
³⁄₈ 4.88 2.68 2.20 2.95 7.09 10.04 4.41 0.21 2.76
4)
½ 4.88 2.68 2.20 2.95 7.09 10.04 4.41 0.33 2.76
4)
1 4.88 2.68 2.20 2.95 7.09 10.04 4.41 0.47 2.76
4)
1½ 4.88 2.68 2.20 4.13 7.26 11.4 4.41 0.69 3.11
4)
2 4.88 2.68 2.20 5.55 7.66 13.21 4.41 1.02 3.90
4)
3 4.88 2.68 2.20 7.87 8.44 16.32 4.41 1.59 5.47
4)
4 4.88 2.68 2.20 10 9.17 19.17 4.41 2.02 6.93
4)
6 4.88 2.68 2.20 14.88 10 24.88 4.41 2.71 8.58
4)
10 4.88 2.68 2.20 21.57 11.71 33.29 4.41 4.03 12.01
80 Endress+Hauser
Proline Promass F 100
Flange connections
C
A
B
E
D L
A0015621
DN A B C D E L
[in] [in] [in] [in] [in] [in] [in]
Flange according to ASME B16.5: Class 150 with reduction in nominal diameter
1.4404 (F316/F316L)
Endress+Hauser 81
Proline Promass F 100
DN A B C D E L
[in] [in] [in] [in] [in] [in] [in]
Flange according to ASME B16.5: Class 300 with reduction in nominal diameter
1.4404 (F316/F316L)
DN A B C D E L
[in] [in] [in] [in] [in] [in] [in]
82 Endress+Hauser
Proline Promass F 100
C
A
B
E
F
D L
A0022221
DN A B C D E F L Ldiff 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
2) DN ³⁄₈" with DN ½" flanges as standard
DN A B C D E F L Ldiff 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in]
Endress+Hauser 83
Proline Promass F 100
DN A B C D E F L Ldiff 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
2) DN ³⁄₈" with DN ½" flanges as standard
DN A B C D E F L Ldiff 1)
[in] [in] [in] [in] [in] [in] [in] [in] [in]
1) Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
2) DN ³⁄₈" with DN ½" flanges as standard
84 Endress+Hauser
Proline Promass F 100
Clamp connections
Tri-Clamp
B
L
A0015625
DN Clamp A B L
[in] [in] [in] [in] [in]
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 30 µin: order code for "Measuring tube material", option SB, SE or
Ramax = 15 µin: order code for "Measuring tube material", option SC, SF
Ramax = 15 µin electropolished: order code for "Measuring tube material", option BC
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 30 µin: order code for "Measuring tube material", option SB, SE or
Ramax = 15 µin: order code for "Measuring tube material", option SC, SF
Ramax = 15 µin electropolished: order code for "Measuring tube material", option BC
Endress+Hauser 85
Proline Promass F 100
Threaded couplings
A
B
L
A0015628
DN A B L
[in] [in] [in] [in]
3-A version available: order code for "Additional approval", option LP in conjunction with
Ramax = 30 µin: order code for "Measuring tube material", option SB, SE
86 Endress+Hauser
Proline Promass F 100
VCO
B
L
A0015624
8-VCO-4 (½")
1.4404 (316/316L)
Order code for "Process connection", option CVS
DN A B L
[in] [in] [in] [in]
12-VCO-4 (¾")
1.4404 (316/316L)
Order code for "Process connection", option CWS
DN A B L
[in] [in] [in] [in]
½ AF 1½ 0.62 16.93
B D
A0016777
A B C D
Endress+Hauser 87
Proline Promass F 100
Accessories
1 1;2 1 1;2
45°
A
L L A
½" NPT
½" NPT
1 2
A0028914
1 Connection nipple for purge connections: order code for "Sensor options", option CH "Purge connection"
2 Connection nipple with rupture disk: order code for "Sensor option", option CA "Rupture disk"
DN A L
½ 2.44 8.66
1 2.44 10.24
1½ 2.64 12.20
2 3.11 17.78
3 3.98 22.0
4 4.72 27.0
6 5.55 34.6
10 7.17 14.96
Weight All values (weight exclusive of packaging material) refer to devices with EN/DIN PN 40 flanges.
Weight specifications including transmitter: order code for "Housing", option A "Compact, aluminum
coated".
Different values due to different transmitter versions:
Weight in SI units
8 9
15 10
25 12
40 17
50 28
80 53
100 94
88 Endress+Hauser
Proline Promass F 100
150 152
250 398
Weight in US units
3/8 20
½ 22
1 26
1½ 37
2 62
3 117
4 207
6 335
10 878
3
A0020640
Endress+Hauser 89
Proline Promass F 100
Adapter for cable entry with internal thread G ½" Nickel-plated brass
Device plug
Sensor housing
The material of the sensor housing depends on the option selected in the order code for
"Measuring tube mat., wetted surface".
"316L
With order code for "Sensor option", option CC
Sensor housing": stainless steel, 1.4404
(316L)
Measuring tubes
• DN 8 to 100 (3/8…4"): stainless steel, 1.4539 (904L);
Manifold: stainless steel, 1.4404 (316/316L)
• DN 150 (6"), DN 250 (10"): stainless steel, 1.4404 (316/316L);
Manifold: stainless steel, 1.4404 (316/316L)
• DN 8 to 250 (3/8 to 10"): Alloy C22, 2.4602 (UNS N06022);
Manifold: Alloy C22, 2.4602 (UNS N06022)
90 Endress+Hauser
Proline Promass F 100
Process connections
• Flanges according to EN 1092-1 (DIN2501) / according to ASME B 16.5 / as per JIS B2220:
• Stainless steel, 1.4404 (F316/F316L)
• Alloy C22, 2.4602 (UNS N06022)
• Lap joint flanges: stainless steel, 1.4301 (F304); wetted parts Alloy C22
• All other process connections:
Stainless steel, 1.4404 (316/316L)
Available process connections→ 91
Seals
Welded process connections without internal seals
Surface roughness All data relate to parts in contact with fluid. The following surface roughness quality can be ordered.
• Not polished
• Ramax = 0.76 µm (30 µin)
• Ramax = 0.38 µm (15 µin)
• Ramax = 0.38 µm (15 µin) electropolished
Operability
Operating concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Quick and safe commissioning
• Individual menus for applications
• Menu guidance with brief explanations of the individual parameter functions
Endress+Hauser 91
Proline Promass F 100
Reliable operation
• Operation in the following languages:
• Via "FieldCare", "DeviceCare" operating tool:
English, German, French, Spanish, Italian, Chinese, Japanese
• Via integrated Web browser(only available for device versions with HART, PROFIBUS DP,
PROFINET and EtherNet/IP):
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Bahasa (Indonesian), Vietnamese, Czech, Swedish, Korean
• Uniform operating philosophy applied to operating tools and Web browser
• If replacing the electronic module, transfer the device configuration via the plug-in memory
(HistoROM DAT) which contains the process and measuring device data and the event logbook. No
need to reconfigure.
For devices with Modbus RS485, the data recovery function is implemented without the plug-in
memory (HistoROM DAT).
Efficient diagnostics increase measurement availability
• Troubleshooting measures can be called up via the operating tools and web browser
• Diverse simulation options
• Status indicated by several light emitting diodes (LEDs) on the electronic module in the housing
compartment
Local display A local display is only available for device versions with the following communication protocols:
HART, PROFIBUS-DP, PROFINET, EtherNet/IP
The local display is only available with the following device order code:
Order code for "Display; operation", option B: 4-line; illuminated, via communication
Display element
• 4-line liquid crystal display with 16 characters per line.
• White background lighting; switches to red in event of device errors.
• Format for displaying measured variables and status variables can be individually configured.
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F). The readability of
the display may be impaired at temperatures outside the temperature range.
4 6
1 2 3 5
A0028747
92 Endress+Hauser
Proline Promass F 100
1 2
4 4 4
A0020903
Star topology
1 2 3
5 5 5
A0032078
Endress+Hauser 93
Proline Promass F 100
Star topology
1 3 2
4 4 4
A0026545
HART
1
A0016926
45 Connection for the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
1 Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2 Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3 Standard Ethernet connecting cable with RJ45 plug
94 Endress+Hauser
Proline Promass F 100
PROFIBUS DP
1
A0021270
EtherNet/IP
1
A0016940
Endress+Hauser 95
Proline Promass F 100
PROFINET
1
A0016940
Modbus RS485
2 1
A0030216
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
96 Endress+Hauser
Proline Promass F 100
Ex approval The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex ia
Category (ATEX) Type of protection
Ex nA
Category (ATEX) Type of protection
CCSAUS
Currently, the following versions for use in hazardous areas are available:
IS (Ex i)
• Class I Division 1 Groups ABCD
• Class II Division 1 Groups EFG and Class III
NI (Ex nA)
Class I Division 2 Groups ABCD
Endress+Hauser 97
Proline Promass F 100
EtherNet/IP certification The measuring device is certified and registered by the ODVA (Open Device Vendor Association).
The measuring system meets all the requirements of the following specifications:
• Certified in accordance with the ODVA Conformance Test
• EtherNet/IP Performance Test
• EtherNet/IP PlugFest compliance
• The device can also be operated with certified devices of other manufacturers (interoperability)
Modbus RS485 certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed
all the test procedures carried out.
Pressure Equipment The devices can be ordered with or without a PED approval. If a device with a PED approval is
Directive required, this must be explicitly stated in the order. For devices with nominal diameters less than or
equal to DN 25 (1"), this is neither possible nor necessary.
• With the identification PED/G1/x (x = category) on the sensor nameplate, Endress+Hauser
confirms conformity with the "Essential Safety Requirements" specified in Appendix I of the
Pressure Equipment Directive 2014/68/EU.
• Devices bearing this marking (PED) are suitable for the following types of medium:
• Media in Group 1 and 2 with a vapor pressure greater than, or smaller and equal
to0.5 bar (7.3 psi)
• Unstable gases
• Devices not bearing this marking (PED) are designed and manufactured according to good
engineering practice. They meet the requirements of Article 4 paragraph 3 of the Pressure
Equipment Directive 2014/68/EU. The range of application is indicated in tables 6 to 9 in Annex II
of the Pressure Equipment Directive 2014/68/EU.
98 Endress+Hauser
Proline Promass F 100
• NAMUR NE 131
Requirements for field devices for standard applications
• NAMUR NE 132
Coriolis mass meter
• NACE MR0103
Materials resistant to sulfide stress cracking in corrosive petroleum refining environments.
• NACE MR0175/ISO 15156-1
Materials for use in H2S-containing Environments in Oil and Gas Production.
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Application packages
Many different application packages are available to enhance the functionality of the device. Such
packages might be needed to address safety aspects or specific application requirements.
The application packages can be ordered with the device or subsequently from Endress+Hauser.
Detailed information on the order code in question is available from your local Endress+Hauser sales
center or on the product page of the Endress+Hauser website: www.endress.com.
Detailed information on the application packages:
Special Documentation for the device → 103
Endress+Hauser 99
Proline Promass F 100
Special density Many applications use density as a key measured value for monitoring quality or
controlling processes. The device measures the density of the fluid as standard and
makes this value available to the control system.
The "Special Density" application package offers high-precision density
measurement over a wide density and temperature range particularly for
applications subject to varying process conditions.
Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Accessories Description
Heating jacket Is used to stabilize the temperature of the fluids in the sensor.
Water, water vapor and other non-corrosive liquids are permitted for use as fluids.
If using oil as a heating medium, please consult with Endress+Hauser.
Heating jackets cannot be used with sensors fitted with a rupture disk.
BA00371F
For details, see "Technical Information" TI00429F and Operating Instructions
Wireless HART adapter Is used for the wireless connection of field devices.
SWA70 The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
100 Endress+Hauser
Proline Promass F 100
Fieldgate FXA320 Gateway for the remote monitoring of connected 4 to 20 mA measuring devices via
a Web browser.
BA00053S
For details, see "Technical Information" TI00025S and Operating Instructions
Fieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART
measuring devices via a Web browser.
Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices and can be used in non-hazardous areas.
Endress+Hauser 101
Proline Promass F 100
Memograph M graphic The Memograph M graphic data manager provides information on all the relevant
data manager measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
BA00247R
For details, see "Technical Information" TI00133R and Operating Instructions
iTEMP The temperature transmitters can be used in all applications and are suitable for
the measurement of gases, steam and liquids. They can be used to read in the
medium temperature.
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer (www.endress.com/deviceviewer): Enter serial number from
nameplate
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
Technical Information
102 Endress+Hauser
Proline Promass F 100
ATEX/IECEx Ex i XA00159D
ATEX/IECEx Ex nA XA01029D
cCSAus IS XA00160D
INMETRO Ex i XA01219D
INMETRO Ex nA XA01220D
NEPSI Ex i XA01249D
NEPSI Ex nA XA01262D
Special Documentation
Installation Instructions
Contents Comment
Installation instructions for spare part sets and accessories Documentation code: specified for each individual accessory .
Registered trademarks
HART®
Registered trademark of the FieldComm Group, Austin, Texas, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
Modbus®
Registered trademark of SCHNEIDER AUTOMATION, INC.
EtherNet/IP™
Trademark of ODVA, Inc.
PROFINET®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
Microsoft®
Registered trademark of the Microsoft Corporation, Redmond, Washington, USA
TRI-CLAMP®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
Endress+Hauser 103
*71505580*
71505580
www.addresses.endress.com