SW Update Manual R10.6.2
SW Update Manual R10.6.2
SW Update Manual R10.6.2
2
Intera MR
XW8000/W8000 Host
XW6000/W6000 Reconstructor
PROPRIETARY NOTICE
This manual and the information contained therein is proprietary to Philips and shall not be reproduced, copied in whole or
part, adapted, modified, or disseminated without the written permission of Philips.
Use of this manual and the information contained therein is strictly reserved for Philips engineers or persons licensed by
Philips. Use of the manual by unauthorised persons is strictly prohibited.
Printed in Best, the Netherlands. 4522-210-27552
2006-12-19
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 2 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
Check http://incenter.medical.philips.com/:
• R10.6.2 Service Packs.
3. z • R10.6.2 Security Fixes. - -
Check http://incenter.medical.philips.com/:
4. z For required FCO’s (e.g. FCO 781 00133 and FCO 781 00171), and - -
Newsletters with respect to MR.
(Service library Æ Magnetic Resonance Æ FCO/Newsletters)
Check http://incenter.medical.philips.com/:
5. z • Latest FCO’s, Newsletters with respect to - -
EasyVision/ViewForum, EasyAccess and EasyLink products.
Check http://incenter.medical.philips.com/:
• Attributes for the DICOM printer(s).
6. z - -
(Service library Æ Medical IT Æ Interoperability Knowledgebase Æ
Printer Validation Æ DICOM Attributes)
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 3 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
R10.3/R10.4/R10.6.1
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
Obtain Host Proxy settings NetForum <MR User> (z) (see Figure 7-2
10. z 7.2 109
Proxy server settings for NetForum)
See
SIM
13. z Perform Save R10.3/R10.4/R10.6.1 ASW-Configuration on MO-disk. R10.3/R -
10.4/R1
0.6.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 4 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
R10.3/R10.4/R10.6.1
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
Reconnect Paper-printer.
22. z
Ackowledge to install: ”DOT4 USB printing support”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 5 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
R10.3/R10.4/R10.6.1
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
29. z Adjust Host “Hospital Connection” settings (see step 9 for values) 5.1.1 47
30. z Adjust Host Computer and Workgroup name (see step 9 for values) 5.1.2 53
32. z - -
33. z Inform Firewall Administrator for access NetForum site. 7.2 109
1
This setting should not be modified.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 6 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
R10.3/R10.4/R10.6.1
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
39. z Disable Knee-Foot coil when not used (see Coil Administration) 4.7 41
43. z Adjust Host Proxy settings NetForum <MR User> (z). 7.2 109
45. z Optionally adjust Language set-up <MR User> (z). (Default: English) 7.4 113
48. z Adjust Paper-printer settings (see Installation of a Paper Printer). 8.2 125
51. z Check connections with DICOM nodes (see DICOM Verify z) 6.3.5 100
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 7 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
R10.3/R10.4/R10.6.1
DESCRIPTION
SECTION
UPDATE
PAGE
STEP
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 8 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
Contents
1. Introduction.............................................................................................. 12
1.1. Conventions used in this Bulletin .................................................................................................12
2. Description of R10.6.2 ............................................................................. 13
2.1. Features .......................................................................................................................................13
2.2. Solved FPR’s in R10.6.2..............................................................................................................13
2.3. Known Problems and Workarounds ............................................................................................15
3. SW Installation ......................................................................................... 17
3.1. Check SW levels z......................................................................................................................17
3.2. Firmware-installation + PMS BIOS settings.................................................................................19
3.3. Reconstructor SW-installation......................................................................................................20
3.3.1. SW-installation on an XW6000/W6000 Reconstructor ................................................................20
3.4. Host SW-Installation (z)..............................................................................................................21
3.4.1. SW-installation on Host (z) .........................................................................................................22
3.4.2. Keyboard Layout and Input Language Configuration <MR User> (z) ........................................23
3.4.3. Host date & time settings z.........................................................................................................27
3.4.4. Virus scanner set-up ....................................................................................................................29
3.4.5. Defrag C-partition on Host z .......................................................................................................30
4. ASW Configuration.................................................................................. 32
4.1. Start Configuration Application z ................................................................................................32
4.2. Set System Reference Number z...............................................................................................35
4.3. Set System Type..........................................................................................................................35
4.4. Set System Parameters ¡...........................................................................................................36
4.5. Set Patient Name Format (Japan only) .......................................................................................37
4.6. Software Options .........................................................................................................................38
4.6.1. Re-activate Key(s) .......................................................................................................................38
4.6.2. Register Permanent key ..............................................................................................................39
4.6.3. Edit Temporary key(s)..................................................................................................................40
4.7. Coil Administration .......................................................................................................................41
4.8. Scanner Hardware Configuration ................................................................................................42
4.9. Logoff Service Application – Reboot Host z ...............................................................................43
4.10. Add R10.6.2 Scan Protocol(s) .....................................................................................................45
5. Network configuration............................................................................. 47
5.1. Host Network Configuration .........................................................................................................47
5.1.1. Host “Hospital Connection” settings z ........................................................................................47
5.1.1.1. Host “Hospital Connection” dynamic IP-address .........................................................................48
5.1.1.2. Host “Hospital Connection” static IP address ..............................................................................49
5.1.1.3. Host “Hospital Connection” speed/duplex settings (z) ...............................................................51
5.1.1.4. Host “Hospital Connection” Advanced TCP/IP settings (zc¡) ..............................................51
5.1.2. Host Computer and Workgroup name z.....................................................................................53
5.1.3. Host “Loopback Connection” settings..........................................................................................53
5.1.4. Host “Reconstructor Connection” settings ...................................................................................54
5.1.4.1. Host “Reconstructor Connection” static IP-address.....................................................................54
5.1.4.2. Host “Reconstructor Connection” speed/duplex settings ............................................................55
5.1.5. Host Port filter settings.................................................................................................................55
5.1.6. Host Certificates...........................................................................................................................55
5.1.6.1. Create Self-signed Root, MR-Host and PACS Certificates .........................................................56
5.1.6.2. Export Self-signed Root, MR-Host and PACS Certificates..........................................................57
5.1.6.3. Install Keys on Secure R10.6.2 Host ...........................................................................................60
5.1.6.4. Install Keys on Secure Remote DICOM Node.............................................................................70
5.1.7. Host Network configuration checks z .........................................................................................70
5.1.7.1. Host “Reconstructor Connection” check ......................................................................................72
5.2. Reconstructor Network Configuration..........................................................................................72
5.2.1. Reconstructor “Host Connection” settings ...................................................................................73
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 9 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
List of Figures
Figure 3-1 Start About ..................................................................................................................................... 18
Figure 3-2 About info R10.6.2 Host (example)................................................................................................ 18
Figure 3-3 Date/Time Properties ..................................................................................................................... 28
Figure 3-4 Time zone selection ....................................................................................................................... 28
Figure 3-5 Date/Time Properties ..................................................................................................................... 28
Figure 4-1 Configuration Main Menu ............................................................................................................... 34
Figure 4-2 System Type configuration............................................................................................................. 36
Figure 4-3 System Parameters........................................................................................................................ 37
Figure 4-4 Coil Administration ......................................................................................................................... 42
Figure 4-5 Edit Hardware Configuration .......................................................................................................... 43
Figure 4-6 Edit Action in Service Log Book ..................................................................................................... 44
Figure 5-1 Internet Protocol (TCP/IP) Properties ............................................................................................ 48
Figure 5-2 Static IP-address Hospital Connection .......................................................................................... 50
Figure 5-3 Advanced TCP/IP Settings............................................................................................................. 52
Figure 5-4 Computer name ............................................................................................................................. 53
Figure 5-5 Static IP-address “Loopback Connection” ..................................................................................... 54
Figure 5-6 Certificates to export ...................................................................................................................... 57
Figure 6-1 Local System is a secure node ...................................................................................................... 79
Figure 6-2 Basic Settings DICOM Node.......................................................................................................... 89
Figure 6-3 Enabled Software Options ............................................................................................................. 91
Figure 6-4 Dicom Network node advanced settings........................................................................................ 96
Figure 7-1 LAN Settings ................................................................................................................................ 110
Figure 7-2 Proxy server settings for NetForum ............................................................................................. 110
Figure 7-3 Control Panel ............................................................................................................................... 114
Figure 7-4 Mouse properties ......................................................................................................................... 115
Figure 7-5 Start Security Policies .................................................................................................................. 117
Figure 7-6 Local Security Settng ................................................................................................................... 118
Figure 7-7 Minimum password length ........................................................................................................... 118
Figure 10-1 Cancel Database server startup ................................................................................................ 146
Figure 11-1 Remote user settings ................................................................................................................. 151
Figure 11-2 Control Panel ............................................................................................................................. 154
Figure 11-3 Device manager ......................................................................................................................... 154
Figure 11-4 Location Information................................................................................................................... 155
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 11 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
1. Introduction
This Software Installation Manual contains information about the Intera Software package R10.6.2 .It is
mainly intended for the customer service engineer, who has to install/configure the new software release.
Furthermore this manual contains additional information about the software procedures.
NOTE
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 12 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
2. Description of R10.6.2
2.1. Features
See Newsletters on InCenter for features of Intera R10.6.2 release.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 13 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
MR00028280, Re: Fw: SRN 05724 , RIS problem after upgrade to Rel 11 <Case ID:
MR00033690, MR20050829EM08> <Case ID: MR20050923FK01>
MR00028128 (Dakjes probleem)
MR00033489 Sybase Callback Error messages missing from Log files
MR00033499 VIPER: Dynamic Scan, incorrect sorting after DICOM
MR00033500
MR00033501
MR00034037 Study Comment (0032,4000) missing in DICOM
MR00034151 R11.1.3: GAD Tool Buffer overrun whem nr of repetitions is increased.
MR00035047 body coil frequency shifts and aborts using PV coil
MR00035086 SPIR fat suppression issues
MR00035267 recon PR den bosch, MR00035267
MR00035297 Zooming plan scan during HC
MR00035630 IAPServer should not use network connection but Reconstructor connection
MR00037784 Partly solve FSF robustness by automatically restarting the FSF server (copy R2)
MR00037791 System Locked up during Plan Scan
MR00037916 CCT: Configuration UI and Logging UI open in same window
MR00037922 CCT: hcserverprocess takes 1 cpu
MR00038173 Configuration should not support P10
MR00038198 Scanner Crashed
MR00038338 Shim tool calculates negative amount of iron for empty rails
MR00038338 Shim tool calculates negative amount of iron for empty rails
MR00038342 Shim tool shows iron distribution with wrong plate or shows non existing plates
MR00038342 Shim tool shows iron distribution with wrong plate or shows non existing plates
MR00038427 System crash
MR00038452 Shim tool does not show help information
MR00038467 Wrong allowed F0 freq. offset in prep phase parameters
MR00038596 PFEI MRX board causes bad image quality
MR00038785 Data transfer from R8.5.2 to R2: images lost, series added together
MR00038794 Software crash after "Copy from Temp".
MR00038836 FSF: FSF Crash occurred and problem with PMSSec
MR00038871 Nthost file corrupt after upgrade from VMS
MR00039013 AfterFCO78100133 Pat. pos has changed Plan Scan
PMSta12882 Slow DICOM transfers.
Customer: Long waiting times for DICOM network export jobs.
PMSta13426 Shift in spectral display in spectroscopic imaging.
Customer: In Spectroview, the displayed spectra are shifted with respect to the position of
the grid over a distance of half a voxel. The effect is mostly visible in spectroscans with a
low number of quite large voxels (like prostate spectroscopy), but it happens in all scans.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 14 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
PMSta13595 PPinstall loosing (Spectro) user protocols. Customer: After upgrading a VMS system to
R1/R11 sometimes customer protocols are lost (especially Spectro protocols). See
Appendix for pr PMSta13595 for the new upgrade procedure.
PMSta13685 PPinstall loosing (Spectro) user protocols.
Customer: After upgrading a VMS system to R1/R11 sometimes customer protocols are
lost (especially Spectro protocols).
PMSta13685
PMSta13795 SPT testing CTL spine coil on 3T systems is very critical.
FSE: The positioning of the CTL spine coil is too critical and as a consequence a lot of
CTL spine coils do not pass the SPT test and are rejected.
PMSta13797 Physiotrack full functionality to be supported in R1.2.x.
Customer: The added functionality contains:
- Non Invasive Blood Pressure measurements (NIBP).
- PMS VCG samples are returned to the Monitoring Equipment Box (MEB) and are
used for displaying on the remote monitors.
- Performance gain, which could cause scan aborts in the past (immediately after start
scan).
PMSta14364
PMSta14366 Hardcopy problems (GVHCserver crash). Customer: The solution is a temporary
workaround. The GVHCserver will be automatically restarted when crashed.
PMSta14393 Gyroview crash after subtraction. Customer: After a subtraction job is finished and
operator selects Exit, the message “Instruction at 0x00020026 referenced memory at
0x02c8950c. Memory could not be written to." and the system crashes.
PMSta14469 New exam crash
PMSta14103
PMSta13876
PMSta14799;
PMSta14469;
PMSta14858;
PMSta14568
PMSta14125
PMSta13877;
PMSta13445
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 15 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
This is an overview of the problems in the area of installation and test-software, which have been discovered
during or after the beta-test. These problems have not been resolved due to resource restrictions and priority
settings.
• Service PC needs Windows 2000 or higher, to support Remote Field Service Framework.
• Service PC fails to start LOTS session. Possible workaround on Service PC is:
• Start PMSSec by inserting smart card/dongle on Service PC and entering correct HW Key.
• Right click over PMSSec icon in task bar and select ‘Update Tasklist’ as follows for
FSFTasklist.pms:
• Verify that the new updated tasklist should not have IEXPLORE.exe in it.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 16 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
3. SW Installation
PLATFORM 12NC DESCRIPTION
XW8000 Host 4522 132 34892 Dual Pentium IV (Xeon) 2.8GHz
W8000 Host 4522 132 51141 Dual Pentium IV (Xeon) 2.4GHz
XW6000 Reconstructor 4522 132 51441 Dual Pentium IV (Xeon) 2.8GHz
W6000 Reconstructor 4522 139 51431 Dual Pentium IV (Xeon) 2.2GHz
Table 3 Required SW
Status (2006-12-19):
XW8000 Firmware 2.0
W8000 Firmware 1.29
XW6000 BOS 02.02.02.01
Firmware 1.22
W6000 BOSWXP_R 02.02.03.01
Firmware 1.29
Prerequisites Host:
Note • Reconstructor is switched on.
• Ethernet Host Reconstructor cable is connected.
• Login MRService + <Password> on Host.
Password = Manager (DVD installed Host).
Password = MRService (Factory installed Host).
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 17 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Number 1: Release
Number 2: Version
Number 3: Level
Number 4: Service Pack level
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 18 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
1. Insert the newly created floppy disk containing the Firmware + PMS BIOS-settings and
restart the computer.
4. The computer will now reboot and it will boot from floppy.
7. Remove the floppy from the system and restart the computer.
1. Insert the newly created floppy disk containing the Firmware + PMS BIOS-settings and
restart the computer.
2. Press <Return> to continue after:
The 'Welcome to ROMPAQ' window.
3. Select HP System BIOS, press <Return> to continue.
4. Select the latest version and press <Return> to continue.
5. Press <Return>, if necessary, after the message appears:
Do you still want to reprogram your firmware with this image?
6. Press <Return> after the message:
A backup image of your current system ROM has been created.
7. Press <Return> after the message:
If you continue, your firmware will be reprogrammed.
1
This section is optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 19 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
21. Remove the floppy from the system and restart the computer.
The sections below describe the various steps to install the initial R10.6.2 software on an XW6000 or W6000
Reconstructor.
The Reconstructor will boot from CD, the installation menu window will appear:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 20 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter “Y”:
XW6000 Version 02.02.02.01 Reconstructor software
Install Reconstructor Operating System
0 - Return to Menu
Y - Start installing Reconstructor Operating System
• Software upgrade.
1
Depends on result section 3.1Check SW levels z p. 17
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 21 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
A DVD SW installation takes about 45 minutes and requires user input only at the
beginning.
Prerequisites:
Host:
• Is connected via Ethernet with Hospital LAN
• USB-slots have no Memory sticks/Card readers/Hard disks connected.
Reconstructor:
• Is powered on
Notes • Is connected via Ethernet with Host.
Paper printer:
• Is disconnected (to avoid acknowledgement new hardware detected).
(Ex/In)ternal MO-drive:
• Is powered on.
• Has no MO-Disk inside!
• Is connected via PCI SCSI-adapter on Host.
• External MO-drive has SCSI-termination enabled (dipswitch F only)
• Power on or Host.
Reboot
• Place in the meantime the Preinstallation DVD 4522 210 27971 into the internal DVD Rom-drive.
When R10.6.2 Host crashes with Blue screen please verify firmware level
DVD-ROM (see Table 1-1 Checklist to Update or Install an R10.6.2
Warning System step 2)
The Software Installation Tool menu pops up.
• Please remove Preinstallation DVD form DVD-ROM drive and insert Application Software DVD.
• Click “Yes”:
1
Depends on result section 3.1Check SW levels z p. 17
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 22 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
Installation is started…
• The R10.6.2 Host will reboot a couple of times during the installation!
Notes • All error messages are related to the non-configured status of the Host!
3.4.2. Keyboard Layout and Input Language Configuration <MR User> (z) 1
1
Optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 23 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
1
MR-Host will ask for a new password.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 24 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Add:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 25 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Warning
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 26 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Login MRService
Password = Manager 1 (DVD installed Host).
Password = MRService (Factory installed Host)
1
MR-Host will ask for a new password.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 27 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click (If applicable) checkbox “Automatically adjust clock for daylight saving
changes:
• Click “Ok”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 28 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
1
Otherwise RSN support via SSH and SFTP is not possible.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 29 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Virus scanning slows down (re) store R1.1, R10.3 or R10.4 Spectro DICOM data
significantly!
Warning
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 30 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
When this message about lack of free space appears, please click “Yes”:
Note
• Wait until the Disk Defragmenter has finished (takes about 10 minutes).
• Click Close →
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 31 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
4. ASW Configuration
• Enter Username
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 32 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
• Click “Ok”
• Click “Ok”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 33 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
4.2 p. 35
4.4 p. 36
4.5 p. 37
4.6.2 p. 39
4.6.3 p. 40
4.7 p. 41
6.3.2 p. 80
6.3.3.1 p. 81
6.3.3.3 p. 87
6.3.3.2 p. 85
4.8 p. 42
6 p. 76
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 34 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
Fill in System
reference number.
• Click “Check”→“Apply”→“Ok”→“Yes”:
1
Warning the System Reference number cannot be changed afterwards.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 35 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 36 of 154 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Check”→“Apply”→“Ok”→“Yes”:
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 37 of 154
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 38 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 39 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Enter The permanent key obtained from PMS InCenter Server into the Value field (remove the
# symbols first):
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 40 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter The temporary key(s) into the Value field (remove the # symbols first), enable the key(s)
by selecting the checkboxes in the column to Activated:
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 41 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 42 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Check”→“Apply”→“Ok”→“Yes”:
• Click Logoff:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 43 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Ok”
• Click “Next” when you want to edit the current Site Log Entry or
“Cancel” when you are ready → “Yes” on Do you want to cancel?
• Click “Ok”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 44 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
This procedure will add R10.6.2 Philips Scan Protocol(s) to the original R10.3/R10.4/R10.6.1 Scan Protocol
Database (see step 26 Table 1-1 Checklist to Update or Install an R10.6.2 System)
• Login MRService + <Password> on Host
Prerequisite for this procedure is that the Knee-Foot coil is enabled, see section 4.7 Coil
Administration p. 41.
• Click “Ok”:
• When the enabled Knee-Foot coil is not used, it must be disabled again, see section 4.7 Coil
Administration p. 41.
• Click Exit
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 45 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Logoff
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 46 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
5. Network configuration
In the next table a default network overview is given for the R10.6.2 Host and
Reconstructor computer. This is the situation after SW loading or Factory Pre-installed
SW. This section describes the procedures to change the default settings to site-specific
settings.
The default TCP/IP network settings are:
In the table below, a default network overview is given for the Host and Reconstructor . This is the
situation after SW loading or Factory Pre-installed SW. This section describes the procedures to change
the default settings to site-specific settings.
R10.6.2 Host
Workgroup: PHILIPSMR
“Hospital Connection”:
Speed/duplex: AUTO/AUTO
Workgroup: WORKGROUP
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 47 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
When Hospital has a DHCP-server, that will assign an IP- address to the R10.6.2 Host,
see section 5.1.1.1 p. 48.
The alternative is to assign the “Hospital Connection” a fixed IP-address, see section
5.1.1.2 p. 49
• Click Properties
2. Click “Properties”
4. Click “Ok” or
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 48 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
• Click “Close” 2*
• Click Properties
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 49 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
3 4
5
1
7?
2
2. Click “Properties”
4. Enter <IP-Address>,
this address has the form of nnn.nnn.nnn.nnn, where 1<= nnn <= 254.
In this example, 172.16.0.201 has been assigned as the IP-address.
The R10.6.2 Host might incorrectly prompt the network mask of the previous Host
network configuration.
In principle, the subnet mask is configured according the network class (A, B or
Warning C) unless the network is divided into additional subnets.
6. Enter <IP-Address> Default Gateway Hospital.
This may left blank, when Hospital has no Gateway.
8. Click “Ok”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 50 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Properties
• Click “Ok”
• Click “Close”
At this point, all basic units of the Intera workgroup have been configured with the
correct network settings.
• Click “Properties”
1
Only required AutoDetect is not functioning
2
Only required when Hospital uses DNS Domain suffix
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 51 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
2
3
7 8
2. Click “Properties”
3. Click “Advanced”
6. Click “Add”
8. Click “Add”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 52 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Properties
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 53 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
3 4
1
5
2
2. Click “Properties”
4. Enter <IP-Address>,
this address has the form of nnn.nnn.nnn.nnn, where 1<= nnn <= 254.
In this example, 192.168.70.10 has been assigned as the IP-address.
The R10.6.2 Host might incorrectly prompt the network mask of the previous Host
network configuration.
In principle, the subnet mask is configured according the network class (A, B or
Warning C) unless the network is divided into additional subnets.
• Click “Ok” (2*)
• Click “Close”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 54 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
• Click “Close”
Note With the introduction of Windows XPE SP2 this procedure has become obsolete:
• All network traffic, initiated on Host, is allowed,
• Firewall is pre-configured, to support all required functionality.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 55 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
When R10.6.2 Host is set to secure DICOM node, it can transmit data to both
insecure and secure Remote DICOM nodes.
When R10.6.2 Host is set to insecure DICOM node, it cannot receive data from a
Warning secure Remote DICOM node.
• Enter CMD
1
When Hospital already has a self-signed Root certificate, this can be used to create the certificates.
2
PACS is used as an example.
3
Certificate is valid for 10 years
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 56 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 57 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 58 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Finish:
• Repeat These “Certificate Export steps” for YourRootCA and PACS1 certificates.
1
No password is needed, when certificate files are kept in a safe place.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 59 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click Delete
• Click Yes:
• Store *.pfx files in C:\nmr\gyroscan\certificates in a safe place and store public key’s (.cer) on
an USB-key or floppy, to export them to (another) R10.6.2 Host or secure DICOM-node
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 60 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 61 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Next>”:
• Click “Next>”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 62 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Double-Click YourRootCA:
• Click “Ok”
• Right-click Personal
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 63 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 64 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Next>”:
• Click “Next>”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 65 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 66 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Double-Click MR1:
• Click “Ok”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 67 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 68 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Double-Click PACS1:
• Click “Ok”
• Delete the Private Key Certificate (*.pfx) files stored in C:\nmr\gyroscan\certificates folder
during creation process:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 69 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 70 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 71 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click Close
to ping continuously:
to stop pinging:
• Press <Ctrl> C
• Enter Exit
See section 5.1.4 Host “Reconstructor Connection” settings p. 54 and 5.2.1 Reconstructor “Host
Connection” settings p. 73 when result is nok.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 72 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter Reconstructor:
• Click Connect
• Click Ok
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 73 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Properties”
• Click Configure…:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 74 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
• After this modification on the Reconstructor, you have to perform 5.1.4.1 Host “Reconstructor
Connection” static IP-address p. 54.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 75 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• A DICOM File Export function to write to the local hard disk drive (E:\DICOM).
These files can subsequently be copied from R10.6.2 Host via SFTP or via other
file transfer/copy.
• A DICOM RIS (Radiology Information System) interface to retrieve the work list
of patients that are scheduled to be examined and to send an MPPS (Modality
Performed Procedure Step) request/status report to the RIS or PACS (Picture
Archiving and Communication System).
• Intera Network AE – application on R10.6.2 Host, that offers all DICOM SCU
(Service Class User) and SCP (Service Class Provider) services mentioned in
Table 2 DICOM Interoperability p. 77 except the Print SCU/SCP services.
The DICOM AETitle uniquely refers to an application on a DICOM network node that
offers DICOM service(s). The name cannot be used on another DICOM network node.
The characters a-z, A-Z, 0-9, composite characters Ä and ‘underscores’ _ are allowed;
spaces and special characters `-/ are not allowed.
The DICOM AETitles and IP-Hostnames are CaSe SeNsItIvE!. We advice to use
UPPERCASE characters only.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 76 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
On an R10.6.2 Host, the configuration of Remote DICOM nodes has been simplified,
through the use of so-called templates.
Each of the available templates is pre-configured with most of the DICOM settings.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 77 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
When applicable:
• Login MRService + <Password> on Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 78 of 155 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
1
See Help text for explanation of the other local configuration parameters.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 79 of 155
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
• Click “Up”
DICOM Image Number direction is used to sort the images in a DICOM viewer or PACS.
Some hospitals want the revert ordering of image slices in one or more of the main
anatomic directions. Sagital, Transversal or Coronal. This configuration setting applies
only for the slice order! For additional information refer to the Gyroweb link:
http://pww.mr-mkt.philips.com/cmt_mr/attachments/pk2_101/image_numbering_in_R10.htm
When applicable:
• Login MRService + <Password> on Host
• Click Transversal:T
Sagital:T
Coronal:T
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 80 of 156 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Check”→“Apply”→“Ok”→“Yes”:
When applicable:
• Login MRService + <Password> on Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 81 of 157
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
DICOM print defines the R10.6.2 Host as Service Class User (SCU) and the hardcopy
printer as Service Class Provider (SCP).
If the DICOM hardcopy printer provides print services to more than one MR-Host (e.g.
MRI 1, MRI 2, etc), an unique SCU AE Title must be defined for each MR-Host e.g.
MR1, MR2.
Note
For the AETitles the characters a-z, A-Z, 0-9 and underscore are allowed. Spaces and
other characters are not! To prevent problems, we advice to use Upper Case characters
only.
The AETitle must be unique for every DICOM node at the Hospital.
________________
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 82 of 157 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
All the other fields of the selected DICOM printer can also be modified if necessary.
http://incenter.medical.philips.com/ contains a list of tested printers including all the
parameters and their optimum values. In case of problems with the printer, check these
parameters first.
• Click “Check”→“Apply”→“Ok”→“Yes”:
You can also configure the same printer 2 times when the printer has more than 1 film
magazine (film size). It is important to realize that, in this example, to change from
AGFA_14x17 to AGFA_8x10, it is not enough to just select the printer with the other film
size in the Print Control UI. It is still necessary to manually change the selected film size
on the printer too.
• Click T → AGFA_14X17:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 83 of 157
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 84 of 157 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
When applicable:
Do NOT modify the standard DICOM printer templates starting with ISGDICOM
Warning
• Click T to select KODAK_14 (Example)
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 85 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Enter Modifications:
• Click “Check”→“Apply”→“Ok”→“Yes”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 86 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
DICOM printers
that can be deleted.
The ViewForum DICOM node has to be connected to the network and must be
switched on.
When applicable:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 87 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Not used!
• Click “Add”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 88 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter • Name
• AE Title
• IP HostName
• IP Address
When Hospital has a DNS-server (see also section 5.1 Host Network Configuration p.
47), the IP Address that belongs to “IP HostName”, is filled in automatically when you
click “Check” or “Apply”.
• Check in the Intera application if it is possible to export images to the just configured Remote
DICOM node.
• Check with the Queue Manager if all the queues are enabled and if the transfer passes with
status “success”.
In case of status “failed” one can continue with the advanced configuration as described
in section 6.3.4.8.1 DICOM Network Node Advanced settings p. 95 or do a DICOM
Verify test as described in section 6.3.5 DICOM Verify z p. 100.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 89 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Be aware that only 1 Worklist Management and only 1 MPPS node should be
configured! These may be the RIS Node, but the MPPS node can also be a PACS. In
this latter case the applicable SOP classes should be (un) configured in the applicable
nodes with Advanced Settings.
Warning • Worklist Management is only available when Software Option “RIS” is enabled.
• MPPS is only available when Software option “IHE 2/3: MPPS and SC” is enabled:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 90 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
Example:
Warning
For the configuration of an Archive with the Storage Commitment service you can select
one of the Philips templates EasyAccess (Store), EasyWeb, Xcelera or use the non-
Philips templates “Other - without private SOPs” or “Other - with private SOPs”.
For example in case of an EasyAccess 8.2 with different AETitles for Store and for
Query/Retrieve, configure the Store AETitle with the template EasyAccess (Store), and
the Query/Retrieve AETitle with the template EasyAccess (Query/Retrieve). In cases
where a different StorageCommit AETitle has to be configured, this AETitle can be set
while configuring the EasyAccess (Store).
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 91 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
To configure such a remote node, use the template “Other - without private objects”. This
template almost has all SOP classes configured. Manually, unwanted SOP classes can
be moved to the un-configured column.
• Click T to select Other - without Private SOPs in the list of available templates.
• Click Add:
• Enter • Name
• AE Title
• IP HostName
• IP Address
In case the Remote DICOM Node doesn’t support the Grayscale Softcopy
Presentation state SOP class un-configure this as follows:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 92 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
In the SCU SOP classes’ window, in the configured column, select the Grayscale
Softcopy Presentation State storage by clicking on this field and move it to
the un-configured column with the mouse pointer.
In case the Remote DICOM Node doesn’t support Secondary Capture Image Storage
un-configure this as follows:
In the SCU SOP classes window, in the configured column, select the Secondary
Capture Image Storage by clicking on this field and move it to the un-configured column
with the mouse pointer:
In case the Remote DICOM Node doesn’t support MPPS un-configure this as follows:
In the SCP SOP classes window, in the configured column, select the Modality
Performed Procedure Step by clicking on this field and move it to the un-configured
column with the mouse pointer.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 93 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Yes” for the parameter “Combine MR Rescale” (to not export rescaling parameters):
• Click “Apply”
• Check in the Intera application if it is possible to export images to the just configured Remote
DICOM node.
• Check with the Queue Manager if all the queues are enabled and if the transfer passes with
status “success”.
In case of status “failed” one can continue with the advanced configuration as described
in section 6.3.4.8.1 DICOM Network Node Advanced settings p. 95 or do a DICOM
Verify test as described in section 6.3.5 DICOM Verify z p. 100.
To configure such a remote node, use the template “Other - with private objects”:
This template has almost all SOP classes configured. Manually SOP classes can be
moved to the unconfigured column. These template are meant for Remote DICOM
Nodes that support the receiving / processing of "private attributes" (Spectro data and
Series (scan) Data Blob sets)
Example procedure for PACS with private objects see section 6.3.4.6 Non-Philips
Remote DICOM Node without private objects support p. 92.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 94 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Modify
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 95 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
1. The ArtimTimeout value (default 60 sec.) determines the timeout between the
establishment of a tcp/ip connection and the association request. Decline it
specifies also the maximum timeout between an association reject or association
release and the actual disconnect of the tcp/ip connection.
2. The value of AssociationTimeoutSCU (default 60 sec.) is the allowed period of
inactivity (in seconds) for an open association. Even if the association is released
after the timeout period due to inactivity, a ’virtual’ association still exists on the
SCU side; i.e., DICOMNetworkRepositoryConnection at the repository level stays
open while the DICOM network association at tcp/ip level is temporarily closed and
reopened when needed:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 96 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
3. the Transfer Syntaxes can be adjusted by shifting the Syntaxes from “Configured”
to “Unconfigured” or vice versa:
4. The SCU and SCP SOP classes can be (un)configured. Choose “Custom” to
(un)configure individual SOP classes. The selection “Full” contains the same SOP
classes as “Custom”, only in Custom the (un)configured SOP classes can be made
visual. The selection “Standard DICOM” is the same as the “Full” setting except
the private objects Spectro data and Series Data (Blob) sets are removed. The
following parameters can be adjusted:
5. Finally
The DICOM node can be set as a SecureNode. That enables other entries like
• Encryption (do not use!)
• Default Certificate
• Click “Apply”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 97 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application – Reboot Host
z p. 43
In case you still have problems with DICOM networking please refer to the document
“WinMRI and DICOM”, which is available via the http://InCenter.Medical.Philips.com
Here the DVD settings can be configured. For a new installation, these settings can be
kept to the default values. The default settings for Name is “DICOM-DVD” and all other
settings are set to “No”.
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application – Reboot Host
z p. 43
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 98 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 99 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Here the DICOM File settings can be configured. For a new installation, these settings
can be kept to the default values.
In the “Suppressed Patient Info” list, the attributes can be specified which will be
suppressed in case of anonymous export to DICOM Files. By default, the patient name,
Patient ID (in the <MR User> Interface this is called Registration ID) and Patient Birth
date will be suppressed.
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application – Reboot Host
z p. 43
With an Intera MR it is possible to verify several items via one test button.
The following tests are performed one after the other in the listed order:
1. Perform a TCP/IP ping to the remote DICOM node (Network Connectivity);
2. Check what DICOM services are supported by both the Intera Host and the
Remote DICOM Node (Associate Configured DICOM Node);
3. Make a DICOM association with the Remote DICOM Node via the DICOM
Verification service class (Verify Configured DICOM Node);
4. Check if the Remote DICOM Node accepts a DICOM secondary capture image and
if so, check the storage of that image via the query SOP class (Network Store
Query).
The ViewForum DICOM node has to be connected to the network and must be
switched on.
When applicable:
• Login MRService + <Password> on Host
Modified automatically
after logoff/logon
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 100 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
Here the Name, ImageBitsStored (16), ImageColumns (128) and ImageRows (128)
can be specified. The defaults can be left as they are.
Select which of the above tests have to be done. Normally the NetworkStoreQuery test
will fail so it is better to de-select this test.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 101 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Ok”:
Here the results of the tests are shown in detail. When the Remote DICOM Node is
switched on and configured correctly, at least the first three tests should pass correctly.
The result of the 4th test is dependent of the store/query capabilities of the Remote
DICOM Node.
In case you still have problems with DICOM networking please refer to the document
“WinMRI and DICOM”, which is available on http://InCenter.Medical.Philips.com.
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application p. 43
When applicable:
• Login MRService + <Password> on Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 102 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Next”
Every medium (DICOM File or MOD) or DICOM Node (LOCAL or Remote) can be
expanded to see the basic and advanced settings.
If you want to send the DICOM configuration e.g. by e-mail to the Helpdesk for
reviewing than you can save the DICOM survey frame as HTML file as follows:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 103 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application – Reboot Host
z p. 43
In order to comply to HIPAA , the Security Eventlog can be shadowed to an External Syslog
When applicable:
• Login MRService + <Password> on Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 104 of 158 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
When the Syslog can not be logged on an external system it needs to be disabled!
Note UDP is preferred protocol: no local buffering/hangup when syslog server is not available.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 105 of 158
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Check”→“Apply”→“Ok”→“Yes”:
• When modification(s) on configuration are ready, perform 4.9 Logoff Service Application – Reboot Host
z p. 43
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 106 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
1
Passwords are case sensitive and expire default after 42 days.
2
Additional settings: M: User must change Password at first login; C: User cannot change password; N: Password never
expires; A Account Disabled; D: Dial in permission granted
3
Password is MRService when system is Pre-installed in factory.
4
Optional
5
Password is Hospital, when system is via DVD or factory pre-installed.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 107 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Add”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 108 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
1
Optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 109 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 110 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter Subsequently the following fields, to identify the Host in the Hospital:
• Click “Ok”
• Click “No”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 111 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Close”
• Close NetMeeting
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 112 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
The Host Login screen reflects the language used by the previous <MR User>!
Warning
2. Click “Apply”
3. Click “Ok”
1
Optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 113 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
4. Click “Ok”:
• Repeat this procedure for every <MR User> that needs UI language to be changed.
Optionally Keyboard layout and Input language can be adjusted (see 3.4.2 Keyboard
Layout and Input Language Configuration <MR User> (z) p. 23)
• Double-Click Mouse:
1
Optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 114 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
HospitalAdmin:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 115 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click Ok
<MR User>:
• Press Ctlr+Alt+Delete
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 116 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 117 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click T to 0 characters:
When user changes password now (see section 7.6 Change Password <MR User>
p. 115), blank passwords are allowed.
• Enter compmgmt.msc:
• Click Users
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 118 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Properties:
• Click “Ok”
To remove the lock out from an account (after 5 times entering with a wrong password:
the account is locked) proceed with this procedure or wait 30 minutes (the lockout will
disappear).
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 119 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Ok”
Remark 1: If the MRService account password is not known anymore, you can use the
Remote account if this account was enabled before. Then you can set a new password
on the MRService account. Another possibility is remotely login into the Remote account
and give the command via SSH: > net user MRService <new password>.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 120 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Enter compmgmt.msc:
• Click Users
• Click “Proceed”
1
See section 7 <MR User> Management on Host and Reconstructor p. 107 for value
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 121 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• You may repeat this procedure for all <MR Users> with unknown password.
• Restart Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 122 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
8. HW Configuration
• Right-click Desktop:
• Click “Properties”:
2. Click “Monitor 2”
1
Optional
2
This disables “Look over the shoulder/Take over” until you logon again!
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 123 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
3. Click “Advanced”:
2
3
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 124 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Apply”:
• Click “Ok”:
• Click “Yes”:
• Click “Ok”:
• Click “Ok”:
Likely, the Host will automatically install the appropriate printer driver when the printer is
connected.
Exceptions like HP Deskjet 3940 need to be installed from HP-CD, while language
Note setting for MRService user is English! (see section 7.4 Language set-up <MR User>
(zc
¡)22F p. 113)
Printers verified by PMS-MR are: Hp DeskJet 697/840/940/5150/3940 and HP LaserJet
1200/1300
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 125 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Some printers may get a name longer than 27-characters (HP 840 DeskJet for
example). They need to be renamed! (see 8.2.1.1 Rename Printer p. 126)
1
E.g. when Printer Name ≥ 27 characters
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 126 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Rename”:
Every user on the Host needs to set default printer settings (Letter, A4 etc)!
Warning
• Login For every interactive user on Host.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 127 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 128 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
9. Update/Save/Restore Configuration
• Press <Enter>
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 129 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Browse to g:\log
logfile:
• Double-click
• Right-Click G:\Site
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 130 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 131 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Press <Enter>
• Enter 2 <Enter>
• Place MO-Disk with saved R10.6.2 configuration into MO-drive and wait until busy led is off.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 132 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Browse to g:\log
logfile:
• Double-click
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 133 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 134 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
2) Free Space on D:
1) Calculate
SUM = (Total Size – Free Space) drive C:
+ (Total Size – Free Space) drive E:
+ (Total Size – Free Space) drive G:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 135 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Place in the meantime the System Backup/Restore DVD 4522 210 27672 into the internal DVD Rom-
drive.
When R10.6.2 Host crashes with Blue screen please verify firmware level
DVD-ROM (see Table 1-1 Checklist to Update or Install an R10.6.2 System step 2).
Warning
The System Backup/Restore Tool menu pops up.
• Remove the System Backup/Restore DVD 4522 210 27672 from internal DVD Rom-drive.
• Insert MO-disk into MO-drive and wait until busy LED is green.
*
Optional
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 136 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “OK”:
When asked for, insert next MO-Volume into MO-drive and click “Ok”:
• Click “Ok”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 137 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Ok”:
• Insert next backup MO-disk in MO-drive, when system asks for and
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 138 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”:
• Click “Ok”:
• Remove MO-Disk from MO-Drive and DVD from internal DVD ROM-drive.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 139 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “Ok”:
• Click “Ok”:
• Click “Ok”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 140 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”:
• Click “OK”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 141 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
60 GB USB Disk
18 GB System Disk
18 GB Data Disk
• Click T → USB-drive
1
Selection based upon 1) Size and 2) System disk always above Data disk.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 142 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
1
Filename ≤ 8 characters.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 143 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click “OK”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 144 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click
• Click “Ok”:
• Click “Yes”
1
Selection based upon 1) Color black; 2) Size and 3) System disk always above Data disk.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 145 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Remove System Backup/Restore Tool 4522 210 27672 from internal DVD Rom-drive.
Database Server and other processes are unable to start, because of mismatch
between Patient Database on Data Disk and transaction logging on System Disk.
This requires a “Reinstall Databases”. See following steps:
Warning
• Click “Cancel”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 146 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “OK”:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 147 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Check http://incenter.medical.philips.com/ for all Service Packs and Security fixes applicable for this
release R10.6.2, please see Table 1-1 Checklist to Update or Install an R10.6.2 System step 3 for how to
obtain.
When saved R10.6.2 configuration on DVD is newer than Backup system disk:
• Logoff MRService
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 148 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 149 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Enter compmgmt.msc:
• Click Ok
• Click Users
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 150 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click “Ok”
• Close Window
Look Over The Shoulder/Take Over makes it possible to view or take over the control
of the operator’s desktop.
These menu items are only accessible when being remotely connected to an
Note
Intera MR Host.
• Login MRService + <Password> on Host
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 151 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
Connection
status
These menu items are only accessible when being remotely connected to an
Intera MR Host.
Note The other party will automatically accept the request and a connection has been
made.
When the operator presses “Accept” a window with the operators desktop will
popup
• Click “Control”
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 152 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
When the operator presses “Accept” the window can be remotely controlled. The
Note:
operator can always interrupt the take over session by pressing Any Key.
• Click “End Session” to end the Look Over The Shoulder / Take Over session.
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 153 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Double-click System:
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 154 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
• Click Country/Region
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 155 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 156 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Intera ALL RIGHTS RESERVED Page 157 of 159
2006-12-19 SW Update Manual R10.6.2 4522-210-27552 / 1.1
• Click Ok
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.
Page 158 of 159 ALL RIGHTS RESERVED Intera
4522-210-27552 / 1.1 SW Update Manual R10.6.2 2006-12-19
document by unauthorized persons is strictly prohibited. This document must be returned to Philips when the user is no longer licensed and in any event upon Philips’ first written
request.
Liability / Warranty Disclaimer:
Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for errors or omissions and reserves the right to make changes
without further notice to any products herein to improve reliability, function, or design. Philips may make improvements or changes in the product(s) or program(s) described in this
document at any time.
ENDOFDOCUMENT
(CSIP Level 1)
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and may not be reproduced,
Copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written permission of the Philips Legal Department.
© 2005 Koninklijke Philips Electronics N.V.