Compact Immunity Test System CIT-10, 10kHz - 400MHz

acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

IN
ONE UNIT:
erator
Signal Gen r (opt.)
l Couple
Directiona amplifier
RF-p er- eter
ow
m
RF-power-

General:
The CIT-10 is a complete test system for conducted RF- im-
munity tests according to IEC/EN 61000-4-6, ISO 11452-4, Special Features:
MIL STD 461E/F CS114, SAE-J1113-2, DC 10614 and similar

standards. Its internal RF generator and RF-power-amplifier • Conducted RF immunity tests acc. to IEC/EN 61000-4-6
produce output signals with max. up to 150W within a fre- and BCI tests acc. to ISO 11452-4 and MIL –STD 461E
quency range from 100 (10)kHz up to 400MHz. Generated • Signalgenerator, RF-power-amplifier, RF-power-meter
signals are measured via one of the max. 3 internal RF-Volt- and directional coupler (optional) in one 19“ case
meters. Furthermore via an optional, internal directional • Stand-alone operation possible with optional available netbook
coupler forward and reflected power can be measured. The • Control-software included
whole test system allows full automatic tests for the speci- • Most important parameters are shown on a integrated display
fied frequency range. As a “stand-alone” test system the • Automatic EUT-monitoring
CIT-10 is convincing by its easy and comfortable handling • Operation via USB port of a PC or Notebook
and the excellent cost-performance ratio. Add-ons like • Complete range of CDN´s available
coupling/decoupling devices are available as well.

Applications:

Immunity Testing:
Testing according to IEC/EN 61000-4-6, ISO 11452-4, MIL STD 461E/F CS114, DC10614 can be performed automatically.

Generation, amplification and verification of RF-Signals:


PAGE // 6

The internal amplifier amplifies any signal from 100 (10)kHz up to 400MHz. Using the internal generator can also generate a desired nar-
rowband signal. Signals up to 30dBm can be measured at the same time. If a directional coupler is installed, forward and reflected power
are measured as well.
Compact Immunity Test System CIT-10, 10kHz - 400MHz
acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

Features:

Internal RF-Power-Amplifier Internal RF-Voltmeter


Several amplifier modules are available. Highest output Accurate measurements of RF signals from -40 dBm up to +30
power can be 150 W over the specified frequency range. dBm are done by the internal RF-voltmeter which can be accessed
The amplifier input can be accessed via the back panel of (for separate use) via a BNC connector. Two internal voltmeters
the CIT-10, so that the amplifier can also be used with measure the forward and reverse power on an optional available
any external generator. 20W, 75W and 150W amplifiers directional coupler or if no directional coupler is installed, the
are available as standard. output voltage of the amplifier is measured.

Amplitude Modulation Internal RF-Signal Generator


Frequencies generated by the generator can also be modulated As the internal generator generates its output signal without any
with a LF signal. Modulation frequencies may vary from 1 Hz up internal mixing components, low harmonics and spurious fre-
to 100kHz, modulation levels are available from 0% to 100%. quencies are assured.

BCI-Tests with additional Power-Meter User defined signals


For BCI-Tests the CIT-10 can be equipped with up to 3 piec- External signals (e.g. EUT Fail or external instruments) can be con-
es internal power meters. nected and monitored using the application software.

Set-up:
The CIT-10 is a PC-controlled test equipment. It can be operated by any commercial IBM compatible PC (WIN/NT/2000/XP/VISTA) via USB port. All
settings of the equipment, e.g. start frequency, stop frequency, step width, test voltage etc. are made by means of the control software which
is also included in the delivery. The three functional units Signal generator, RF power amplifier and RF voltmeter are set automatically by the
software, depending on the pre-set test parameters. Each component, however, may also be called and operated as separate measuring and
testing equipment. This means: using the CIT-10 as testing system, you have three full, additional “single units” at your disposal, for which
separate inputs and outputs are available as BNC connections. Due to the computer-aided control of the CIT-10, any modifications which may
become necessary, for example, due to the revision of standards, may be performed without problems and without having to manipulate the
hardware of the equipment.

Functioning:
The equipment is ready for operation immediately after connec-
tion of the USB port, installation of the drivers and the control
software. After starting of the control software, the main menu
offers the manual control of <RF-Generator> and <RF-Power
Meter>. Further options in the menu are <Calibration> (<CDN-
Calibration>, <Self-Calibration>) and <Test> (<Complete Test>,
<Selective Test>).
7 // PAGE
Compact Immunity Test System CIT-10, 10kHz - 400MHz
acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

CDN-Calibration:
The CDNs (Coupling/Decoupling Networks) serve to inject the complete selected frequency range; in this case the test frequency is in-
test voltage into the lines to be tested and/or to decouple any creased by the control software according to the selected step width and
connected peripheral equipment from the EUT. The charac- the entered dwell time. If there is a malfunction of the EUT, the test may
teristics of the CDNs as well as of the power amplifier are not be stopped at any time. It is then possible to either increase or reduce
absolutely linear over the whole frequency range, i.e. the the frequency by any number of steps, as well as to switch on and off
amount of power required to generate a constant test voltage the modulation and test voltage. Besides, a description of the malfunc-
over the whole frequency range varies slightly, depending on tion occurred may be entered in a comment line which is included in
the frequency. In the calibration run, the frequency-depen- the test record.
dent output level of the signal generator, which is necessary
for a constant test voltage, will be determined and stored in
the software, together with the defined frequency range and
the desired test voltage. The data records thus created may
then be stored and recalled for tests.

<Selective Test> offers the possibility of testing the EUT at discrete frequen-
Self-Calibration:
cies. This can be done either with a fixed test voltage or, optionally, with
When selecting this menu option, the test equipment will per- a ramp function. In case of the ramp function, the start and stop voltage,
form a self-calibration. In this case, the output of the signal the step width by which the test voltage is to be increased, as well as the
generator must be connected to the input of the voltmeter. dwell time between the individual steps may be preset by the tester.

The standard <Protocol> consists of the head of the protocol and a dia-
gram which shows the test results. In the head of the protocol the date
Test:
and time are taken over from the computer; in addition, details like
The menu option <Test> offers the selection possibilities <Com- temperature, air humidity, tester, as well as testing set-up and EUT,
plete Test>, <Selective Test> and <Protocol>. The settings for a may be registered. The protocol may be printed directly. It is also pos-
test, e.g. start and stop frequency, step width and test voltage sible to edit the protocol individually.
are made automatically via the calibration file of the selected
coupling unit. It is now possible to decide whether the test is
to be performed exactly according to these pre-settings, i.e.
exactly as in the calibration, or whether modifications of the
pre-settings shall be admissible. If the calibration run was per-
formed, for example, for a test voltage of 10V, and the test is to
be performed now with 3V without having to perform a new
calibration run for this purpose, this can be done by selecting
menu item <Extrapolation>.

Is a suitable measuring instrument connected to the specified se-


rial port of the CIT-10, EUT can be monitored automatically. Data
PAGE // 8

are shown graphically. During all test routines the amplifier out-
put is monitored in a bar display. This guarantees correct tests.
In the case of <Complete Test>, a test is performed over the
Compact Immunity Test System CIT-10, 10kHz - 400MHz
acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

Technical data: Technical data:


RF Voltmeter (external in-/output) Power amplifier

Frequency range 10 kHz to 400MHz Frequency range 100kHz (10kHz) to 400MHz (75W, 150W)

Measuring range +30 dBm to - 40 dBm 100kHz to 230MHz (20W)

Accuracy ±0.5 dB Gain 51 dB ± 1.5 dB

VSWR < 1.1 : 1 Output power 75W/150W (optional)


20W (100kHz to 230MHz)
Input BNC, 50 Ohm
Distortion <20 dBc at 75W
RF Generator
Input impedance 50 Ohm, VSWR < 1.5:1
Output BNC, 50 Ohm
Output impedance 50 Ohm nom.
Frequency range 10 kHz to 400 MHz
EUT-fail input
Frequency resolution 1 Hz
Input resistance 2.2 kOhm
Output level range 0 to - 60 dBm
Level TTL/CMOS compatible, optical decoupled
Output level resolution 0.1 dB

Output level accuracy ±0.5 dB (± 1 dB max)


EUT-Monitor input
Accuracy (frequency) ±5 ppm (TCXO)
Input voltage 0-10 V
Harmonics < -30 dBc
Input impedance 100 kOhm
Non harmonics < -45 dBc
Amplifier monitor
Amplitude modulation 0 to 100%; resolution 0.5%
(internal) (internal AF Generator) Output BNC, 50 Ohm

Amplitude modulation BNC jack Level - 40 dB (amplifier output), ±3 dB


(external) 1 Hz to 100 kHz, 0 to 100%
Input impedance > 100 kOhm

Pulse modulation variable duty cycle 10 - 90%; Interfaces


resolution 1 % (internal AF Generator)
USB-A Multimeter (for EUT control)
VSWR < 1.5:1
USB-A Relays switching unit
AF Generator
USB-B Connection to computer
Output jack BNC

Frequency range 1 Hz to 100 kHz


General data
Frequency resolution 0.1 Hz
Temperature range 0 to 40°C
Output voltage 0 to 1 V amplitude; resolution 5mV
Warm-up time 15 min.
Accuracy (frequency) ±50 ppm
Housing 19” Subrack or desktop case
Signal Sine wave / square wave / triangular
Width / height / depth 449 mm x 133 mm x 435.5 mm
RF Voltmeter (internal, 2 pcs.)
AC input 100 - 240 VAC; 50/60 Hz
Frequency range 10 kHz to 400 MHz
Volume of delivery CIT-10 ( basic equipment),
Measuring range +53 dBm to - 0 dBm cabling, system software
Accuracy ±0.5 dB Part Number CIT-10/20 with integrated 20W
Directional coupler (optional) power-amplifier;

Frequency range 10 kHz to 400 MHz CIT-10/75 with integrated 75W


power-amplifier;
Power 200 W CW
CIT-10/150 with integrated 150W
9 // PAGE

Insertion loss 0.5 dB max


power-amplifier;
VSWR 1.25 : 1 max
CIT-10/W without internal
Directivity 20 dB min power-amplifier;
Compact Immunity Test System CIT-10, 10kHz - 400MHz
acc. to IEC/EN 61000-4-6 / ISO 11452-4 / MIL-STD 461E

Circuit diagram of a test system according to IEC/EN 61000-4-6:

max. Power Meter


100W Channel 1
Generator Relay 1 max. Amplifier 1 Relay 2
107 dBµV

CIT-10 RS
Directional
Coupler DC
CIT-10 RS CIT-10

Amplifier 2
C5019 Power Meter
GPIB Channel 2

No amplifier
CIT-10 RS
PC
Reference RS232
Device
RS232

CIT-10 RS
~
Device 1 Device 2 Device 3 Device 1 ~ -
AE EUT
Keithley 2000 CAN-Bus Test HP34401A R&S CMD 55

Coupling Network CDN801 M3

Circuit diagram of a BCI test system:

max. Power Meter


100W Channel 1
Generator Relay 1 max. Amplifier 1 Relay 2
107 dBµV

CIT-10 RS
Directional
Coupler DC
CIT-10 RS CIT-10/75

Amplifier 2
C5019 Power Meter
Channel 2
GPIB

No amplifier
CIT-10 RS
Relay 3

PC RS232
Reference
Device
RS232

CIT-10 RS

Device 1 Device 2 Device 3 Device 1

EUT AE
PAGE // 10

Keithley 2000 CAN-Bus Test HP34401A R&S CMD 55


Current Probe Injection Clamp
FCC F-52 FCC F-120-8
Compact Immunity Test System CIT-10
Coupling / Decoupling Networks (CDNs)

Description
Immunity tests according to IEC/EN require coupling of RF disturbance voltages into any
conducting cable of an EUT. Furthermore these disturbances should not be coupled into
any further equipment so that a decoupling path to any auxiliary equipment is provided.
We offer a wide range of CDN´s for different types of interconnected lines which are fully
calibrated for the frequency range from 150kHz to 230MHz. The following CDNs are avail-
able: M-, AF-, S-, T-, RJ, USB-types. Almost any network can be assembled on customer´s
requests. Guidance for selecting the appropriate CDN is given in the following table:

Type Interconnected lines

M1, M2, M3, M4, M5, M2+M3 Unscreened supply (mains)


AF2, AF4, AF5, AF6, AF8 Unscreened nonbalanced lines
S1, S2,S4, S8, S9, S15, S25, S36 Screened lines
T2, T4, T8 Unscreened balanced lines
RJ11, RJ45 Unscreened data lines
RJ11/S, RJ45/S, USB Screened data lines

Test procedure with Coupling/Decoupling (CDNs) Networks acc. to IEC/EN 61000-4-6:

• The EUT shall be placed on an isolating support, 0.1 m above the ground reference plane.
For table-top equipment, the ground reference plane may be placed on a table.
• On all cables to be tested, coupling and decoupling devices shall be inserted.
• The coupling and decoupling devices shall be placed on the ground reference plane, making direct contact with it at about 0.1 – 0.3 m from the EUT.
• The cables between the coupling and decoupling devices and the EUT shall be as short as possible and shall never be bundled or wrapped.
• The height above the ground reference plane shall be between 30 and 50mm (where possible).
• The 6dB attenuator shall be placed to the coupling and decoupling device as near as possible.
• The test shall be performed with the test generator connected to each of the CDN´s in turn while the other non-exited RF-input ports of the
CDN´s are terminated by a 50Ω load resistor.

Calibration of Coupling Devices:

The coupling and decoupling devices shall be placed on a ground reference plane:

*) The 150Ω loading (e.g. a 150Ω to


50Ω adaptor terminated with a 50Ω
11 // PAGE

load) at the AE-port shall only be ap-


plied to unscreened cables (screened
cables will have their screen con-
nected to the ground reference plane
at the AE-side).
Compact Immunity Test System CIT-10
Coupling / Decoupling Networks (CDNs)

Set-up for level setting at the EUT-port of coupling and decoupling devices:

1. The test generator (RF-out) shall be connected to the RF-input port of the coupling device via the 6dB-attenuator.
2. The EUT port of the coupling device shall be connected in common-mode through the 150Ω to 50Ω adaptor to the RF-Voltmeters (calibration).
3. The AE-port shall be loaded in common-mode with a 150Ω to 50Ω adaptor, terminated with 50Ω.

With direct injection to screened cable (CDN S types), the 150Ω load at the AE-port is not required as the screen will be connected to the
ground reference plane at the AE-port side.

Although the 150Ω load at the AE-port is mandatory with CDN T, AF and M types calibration data are identical with the AE-port open or
short. This is because a capacitor is connected to ground at the AE-port side, which leads to a RF-short-circuit similar to the CDN S types.
This means that even with CDN M, AF and T types the 150Ω load at the AE-port is not required.

To calibrate a CDN the following items are required:

• adaptor • fastening angle • 50Ω/150Ω adaptor


Nut
Fastening angle and 50Ω/150Ω adaptor should be ordered
for the first CDN. For each following CDN only the specific adap-
Adapter with BNC jack, specific for each CDN, Fastening 50 / 150 Ohm adpter with
tor has to be ordered. e.g. with 3 banana plugs for CDN M3 angle BNC connectors

Direct Injection: Any shielded connection to an EUT can also be connected to the RF disturbance voltage via a direct injection adapter.

Electromagnetic Coupling Clamp acc. to IEC/EN 61000-4-6

Type F-203I-23MM
Frequency range 10KHz to 1000MHz
Description Max. allowed power:
The EM coupling clamp is used in all cases where CDNs are 10KHz to 100 MHz: 100 watts CW for 15 minutes
not advisable or not available. The lines to be tested are in 100 MHz to 230 MHz: 100 watts CW for 10 minutes
this case inserted in the clamp and the test voltage is inject-
230 MHZ to 1GHz: 50 watts CW for 10 minutes
ed inductively. RF electromagnetic fields frequently degrade
electronic equipment by generating common mode currents The test level of the immunity test with the specified values
on cables. The effect of these E and H fields on the equip- corresponds to a field strength of 100V/m
ment can be simulated by injecting common mode currents Directivity >10dB above 20 MHz
into the cables of the equipment being tested for RF im-
Dimensions (L x W x H) 23 x 610 x 135 mm
munity. IEC/EN 61000-4-6 defines the methods for testing
the immunity of electronic equipment to conducted mode RF-disturbance connector N type
currents between 150kHz and 230MHz. The electromag- Max. cable diameter 23mm
netic (EM) clamp is a high efficiency broadband clamp-on Weight approx. 7.5 kg
injection device developed to test the immunity of electronic
Options: Calibration set consisting of:
equipment when the standard IEC/EN 61000-4-6 CDN using
PAGE // 12

- 2 pcs. calibration corners


the direct capacitive coupling technique is not possible nor
incl. 150/50 W adapter
appropriate. The EM Clamp is often used to test unshielded
- 1 pcs. calibration cable
multiple conductor cables.