Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

NuMotion Error Lists

Download as pdf or txt
Download as pdf or txt
You are on page 1of 53

100*

 Header
 GMP_GET_RTC_ADDRESS_FAILED
 Definition
 The Inter-Process Communication Shared Memory cannot be
created
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 Check for any memory crash in application program
 Remark
 Can locator verify the inter-process communication shared memory?
101*

 Header
 GMP_GET_RPT_POINTER_ERROR
 Definition
 The report buffer cannot be created in Host-to-RTC memory
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 Check for any error message about Motion Controller card in the
event log
 Remark
 Locator should able to get the reason from GMP
102
 Header
 GMP_SYSTEM_NOT_INITIALIZED
 Definition
 The Nu Motion system doesn't initialize when calling any GMP
command
 Action
 Service
 Check whether the system is successfully initialized or not
 Try to initialize the system again. If it doesn't successfully initialize,
please find error code listed in error log file and then refer to the
corresponding solution
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator should able to check the initialization status and point out the
fail reason of previous initialization
200*
 Header
 GMP_OPEN_SCF_FILE_ERROR
 Definition
 Fail to open the SCF file
 Action
 Service
 Check whether the machine configuration selection is correct
 Inform product team
 Production
 Check whether the machine configuration selection is correct
 Inform product team
 Product
 Please check for file path specified in gmp_init_system
 Remark
 Locator should able to catch the problem
223*

 Header
 GMP_READ_SCF_FILE_PATH_ERROR
 Definition
 Fail to open the SCF file
 Action
 Please check for file path specified in gmp_init_system
SCF, Activation and RTC
Error log file
 Remark
 Locator can get the specified path in gmp_init_system
 Able to check which file is missing?
300

 Header
 GMP_CREATE_SHARE_MEM_FAILED
 Definition
 Fail to create the Inter-Process Shared Memory
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 Check for the size of RAM in your PC
 Remark
 Locator is able to check the size and report the recommendation?
304

 Header
 GMP_MALLOC_MEM_EMPTY
 Definition
 Memory allocation is failed during initialization
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 Check for the size of RAM in your PC. 1G Ram is recommended
 Remark
 Locator is able to check the size of RAM and recommend the RAM need?
306

 Header
 GMP_SYSTEM_RESOURCE_NOT_CREATED
 Definition
 Inter-Process Shared memory cannot be successfully created
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 Check for the size of RAM in your PC. 1G Ram is recommended
 Remark
 Locator is able to check the size of RAM and recommend the RAM need?
400
 Header
 GMP_FIBRE_READ_DATA_ERR
 Definition
 Fail to read fiber data when getting the hardware node ID
 Action
 Service
 Check the fiber is connected in good condition by swapping a
known good cable
 Production
 Check the fiber is connected in good condition by swapping a
known good cable
 Product
 Check the fiber is connected in good condition by swapping a
known good cable
 Remark
 NA
401
 Header
 GMP_FIBRE_LOAD_SEEP_ERR
 Definition
 Fail to load fiber data when getting the hardware node ID
 Action
 Service
 Check the fiber is connected in good condition by swapping a
known good cable
 Production
 Check the fiber is connected in good condition by swapping a
known good cable
 Product
 Check the fiber is connected in good condition by swapping a
known good cable
 Remark
 NA
600

 Definition
 A packet cannot be transmitted within a period of time (5s). The
GMP command cannot be sent to motion controller board
 Action
 Service
Replace the motion controller board
 Production
Run FCT of motion controller board to verify the functionality
Replace the motion controller board
 Product
Replace the motion controller board
 Remark
 Can the locator run FCT to verify the control board before
suggest replacement?
603
 Header
 GMP_INVALID_PORT_NAME
 Definition
 An invalid port name is going to use
 Action
 Service
Inform product team
 Production
Inform product team
 Product
Identify the invalid port name from the error log file
Check whether the port(s) is deactivated in activation file.
Activate the port
Remove access to this port in application program
 Remark
 Locator check activation file to judge the reason whether the
invalid port is deactivate or doesn’t exist
605
 Header
 GMP_OPEN_FILE_FAILED
 Definition
 Fail to open files such as SCF, Activation file or RTC
 Action
 Service
Inform product team
 Production
Check whether the SCF, Activation file and RTC were put
into the folder or not
 Product
Check whether the SCF, Activation file and RTC were put
into the folder or not
 Remark
 Locator is able to show the incorrect SCF activation and RTC file
path
608
 Header
 GMP_CHANNEL_ERROR
 Definition
 Channel Error
 Action
 Service
 Same as production
 Production
 Error status = 0x1: search timeout
Error status = 0x2: Exceed software position limit
Error status = 0x4: hit limit sensor
Error status = 0x8: encoder fault
Error status = 0x10: driver fault
Error status = 0x20: motion abort
Error status = 0x40: fault in dependent channel
Error status = 0x80: Control error
Error status = 0x100: Brake locked
Error status = 0x200: Open wire fault
Error status = 0x400: Velocity jump
Error status = 0x40000000: Emergency stop
Error status = 0x80000000: Hit project software position limit
 Product
 Same as production
 Remark
 Locator is able to check the channel error type
609

 Header
 GMP_SYS_INIT_OC_FAILED
 Definition
 Fail to download or setup the OCDB
 Action
 Service
Inform product team
 Production
Inform product team
 Product
Check the validity of OCDB files with Control Team members
or the revision of OCDB
 Remark
 Locator locates OCDB issue
611
 Header
 GMP_NODE_IS_NOT_EXIST
 Definition
 The Hardware node of the activated channel doesn't physically exist
 Action
 Service
 Check the error message from the error log to identify which
hardware causes this problem
 Correct the Node ID
 Swap by a healthy fiber cable. Replace the cable
 Swap by a healthy one. Replace the hardware
 Production
 Same as Service
 Product
 Same as Service
 Remark
 Locator locates which node has issue. Show the correct order
612

 Header
 GMP_DOWNLOAD_FIRMWARE_ERROR
 Definition
 Fail to download controller's firmware
 Action
 Service
Replace the motion controller card
 Production
Same as Service
 Product
Same as Service
 Remark
 NA
613

 Header
 GMP_GET_PCIE_BASE_ADDR_ERROR
 Definition
 Fail to download controller's firmware
 Action
 Service
Replace the motion controller card
 Production
Same as Service
 Product
Same as Service
 Remark
 NA
617

 Header
 GMP_CREATE_DEVICE_FAIL
 Definition
 Fail to create a device
 Action
 Service
Confirm the functionality of the device driver
 Production
Same as service
 Product
Same as service
 Remark
 Locator suggests the version of device driver?
618

 Header
 GMP_FREE_COMMON_BUFFER_FAIL
 Definition
 Fail to free up the datalog buffer in PC
 Action
 Service
Confirm the functionality of the device driver
 Production
Same as service
 Product
Same as service
 Remark
 Locator suggests the version of device driver?
623

 Header
 GMP_MAP_COMMON_BUFFER_FAIL
 Definition
 Fail to allocate continuous memory for datalog
 Action
 Service
Inform product team
 Production
Inform product team
 Product
Add more RAM or reduce the size of datalog buffer
 Remark
 Locator is able to display the datalog information
625
 Header
 GMP_DRIVER_FAULT
 Definition
 Encounter driver fault during stationary commutation
 Action
 Service
 Clear channel error and run again
 Inform product team if problem still exists
 Production
 Clear channel error and run again
 Inform product team if problem still exists
 Product
 Clear channel error and run again
 Inform motion team if problem still exists
 Remark
 Locator is able to know whether the error is triggered before or
during commutation
626

 Header
 GMP_WAIT_NUDRIVE_READY_TIMEOUT
 Definition
 NuDrive initialization fail
 Action
 Service
Check the functionality of NuDrive
 Production
Same as service
 Product
Same as service
 Remark
 Locator runs simple NuDrive checking?
628
 Header
 GMP_SWCOM_MTR_NOT_MOV
 Definition
 The motor doesn't move during Stationary Commutation
 Action
 Service
 Check whether the encoder value can be updated when the "End
effector" moves in user application. Launch NuDiagnostic
 Inform product team if problem still exists
 Production
 Clear channel error and run again
 Inform product team if problem still exists
 Product
 Clear channel error and run again
 Inform motion team if problem still exists
 Remark
 Locator launches encoder reading part
631

 Header
 GMP_MOTOR_COMMUTATION_NOT_READY
 Definition
 No commutation is done on the specific motor during system
initialization or before enable driver
 Action
 Service
Report to product team
 Production
Report to product team
 Product
Commutate the motor before enabling the motor
 Remark
 NA
635
 Header
 GMP_SIO_LINK_TIMEOUT
 Definition
 Sending the data to remote SIO timeout
 Action
 Service
 Report to product team
 Production
 Check whether the SIO/ESIO, which cause timeout, physically exist or not
– Install the SIO/ESIO
 Check the power supply input to the SIO/ESIO
– Fix the power supply issue
 Verify whether the SIO/ESIO is malfunction by swapping another well-
function SIO/ESIO
– Order a new SIO/ESIO for replacement
 Verify the SIOLink cable by swapping a healthy cable
– Order a new cable for replacement
 Product
 Same as production
 Identify the problematic hardware node from the error log file. Check any
excessive active component stated in activation file
– Correct the activation setting
 Remark
 Locator is able to detect the existence of SIO/ESIO
636
 Header
 GMP_INVALID_ACTIVATION_FILE_ACTIVATION_STATE
 Definition
 The activation state in activiation.ini is incorrect. Other than 0 or
1 is invalid. 0: inactive, 1: active
 Action
 Service
Check whether using the correct activation file
Report to product team
 Production
Same as service
 Product
 Check activation.ini file and correct it by modify the file directly or
generate it by SCFMaker
 Remark
 Locator is able to check the missing information in activation file
637
 Definition
 Error in parsing the activation file
 Action
 Service
 Check the machine configuration selection
 Inform product team
 Production
 Check the machine configuration selection
 Inform product team
 Product
 Open the Activation file and then check whether the names of
each module’s component (i.e. channels and ports) between SCF
and SCFMaker
 Generate the activation again by SCFMaker
 Modify the SCF according to the activation file
 Remark
 Locator shows the mismatch part
638
 Header
 GMP_ACTIVE_RESOURCES_NOT_LINKED_TO_HW
 Definition
 The active component doesn't connected to the hardware
 Action
 Service
Inform product team
 Production
Inform product team
 Product
Check the error log file and find out the problematic active
component (i.e. port or channel)
Open SCF by SCFMaker and locate the problematic
component
Deactivate the component highlighted in red or link it to
corresponding hardware
 Remark
 Locator displays the deactivated component
639
 Header
 GMP_ENCODER_NOT_LINKED_TO_HW
 Definition
 The active encoder port doesn't connect to the hardware
 Action
 Service
 Check whether using correct activation file
 Connect the encoder component to the hardware
 Production
 Same as service
 Product
 Check whether the component should be activated
 Connect the component to the hardware
 Remark
 Locator is able to check the hardware connectivity status
644
 Header
 GMP_INVALID_SIM_CARD
 Definition
 The sim card in Motion Controller doesn't plugged
 Action
 Service
 Check the existence of sim card in motion controller card
 Inform product team
 Production
 Same as service
 Product
 Check for the revision of both GMPLibrary and RTC
 Remark
 Locator shows the library version and RTC version
645
 Header
 GMP_INCOMPATIBLE_DEVICE_DRIVER_VERSION
 Definition
 The device driver is imcompatible with the feature used (defined in SCF)
 Action
 Service
 Upgrade the device driver
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator shows device driver version currently using
646
 Header
 GMP_REMOVE_COM_PORTS_FAIL
 Definition
 Fail to remove the COM port
 Action
 Service
 Make sure the com port are well functioning
 Check any fault on installing the Device Driver
 Kill all com port related application

 Production
 Same as service
 Product
 Same as service
 Remark
 NA
647
 Header
 GMP_SET_READ_BUF_ADDR_FAIL
 Definition
 Fail to map the memory for the COM port
 Action
 Service
 Make sure the com port are well functioning
 Check any fault on installing the Device Driver
 Production
 Same as service
 Product
 Same as service
 Remark
 NA
648
 Header
 GMP_CREATE_COM_PORT_FAIL
 Definition
 Fail to create COM Port
 Action
 Service
 Make sure the com port are well functioning
 Check any fault on installing the Device Driver
 Production
 Same as service
 Product
 Same as service
 Remark
 NA
650
 Definition
 CRC Error
 Action
 Service
 Identify which SIOlink’s hardware got this error from error log. Then
check the existence of the SIOlink’s hardware from the machine
 If it exists, check the power cable functionality (electrical drawing).
Replace the power module if failure
 If the power module works properly, swap by healthy fiber cable.
 If problem still exists, replace by a new hardware
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator should locate which hardware not exist
 Electrical drawing. Able to check the power automatically?
660
 Header
 GMP_WAIT_DRV_CH_RDY_TIMEOUT
 Definition
 The Driver channel doesn't enable within the timeout period
 Action
 Service
 Check any driver fault exists in the driver channel
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator is able to display any fault exists in the driver channel
671
 Header
 GMP_EMERGENCY_STOP_TRIGGERED
 Definition
 Emergency stop was triggered
 Action
 Service
 Check whether the EMO button is pressed
– Contact the operator for the reason of pressing EMO button
 Check whether the signal of detecting the power down is triggered
– Report to product team
 Production
 Same as service
 Product
 Same as service
 Find the root cause of false alarm
 Remark
 NA
673
 Header
 GMP_HALL_SNR_ERROR
 Definition
 Hall sensor state is invalid or hardware timeout when performing index reference
commutation during system initialization
 Action
 Service
 Ensure the encoder and each hall sensors are functional
 Production
 Same as service
 Product
 Ensure the encoder and each hall sensors are functional

 Remark
 Locator checks the encoder functioning?
675
 Header
 GMP_SRH_IDX_FAIL
 Definition
 Search index failed when performing index reference commutation
 Action
 Service
 Inform product team
 Production
 Inform product team
 Product
 ensure the channel is enabled, closed loop and not moving, and index
sensor is functional
 Remark
 Can locator searches index?
676
 Header
 GMP_UNKNOWN_IDX_ELECTRICAL_ANGLE
 Definition
 The channel cannot perform index reference commutation
 Action
 Service
 Check whether using correct SCF and activation file
 Inform product team
 Production
 Same as service
 Product
 Make sure the motor and driver supports index reference commutation type
and the channel has selected correct encoder as feedback source
 Remark
 Locator check whether existing system supports index reference commutation
677
 Header
 GMP_MOTOR_COMMUTATION_INACCURATE
 Definition
 The commutation hasn't complete
 Action
 Service
 Check whether using correct SCF and activation file
 Inform product team
 Production
 Same as service
 Product
 Use any of the commutation function to ensure more consistent
commutation angle
 Remark
 Locator is able to run different type of commutation
679
 Header
 GMP_UNKNOWN_ABS_ENC_OFFSET_ELECTRICAL_ANGLE
 Definition
 Absolute encoder position and electrical angle relationship is unknown
 Action
 Service
 Check whether using correct SCF and activation file
 Inform product team
 Production
 Same as service
 Product
 Make sure the motor has built-in encoder that supports absolute encoder
commutation type and the channel has selected this encoder as feedback
source.
 Check whether hardware support to compensate encoder offset count
 Remark
 Check whether hardware support the feature
680
 Header
 GMP_DRV_CURRENT_SNR_ERROR
 Definition
 Some current sensors in driver have wrong response
 Action
 Service
 Replace the driver
 Production
 Make sure motion driver team has entered correct parameters to SCF
 Driver power supply is ON and functional
 Driver loading is connected properly
 Motor does not move to cause back EMF
 Check connection and direction of current sensors are correct and
functionality of current sensors are ok
 Product
 Same as production
 Remark
 Locator is able to check the driver sensor error?
683
 Header
 GMP_HEALTH_MONITOR_NOT_EXIST
 Definition
 Health Monitor should exist during initialization
 Action
 Service
 Retry
 Execute it inside NuSpace
 Report to product team
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator should able to run health monitor automatically
684
 Header
 GMP_SIO_LINK_ERR
 Definition
 SIO link error
 Action
 Service
 Check the hardware connectivity
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator should able to check the connectivity of linkage
694
 Header
 GMP_BOARD_I2C_ERROR
 Definition
 I2C error
 Action
 Service
 Check the board and board connector
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator is able to locate the I2C information?
695
 Header
 GMP_INVALID_FIBER_NODE_TYPE
 Definition
 Invalid fiber node type
 Action
 Service
 Ensure the node ID of each node is correct
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator is show the node ID information
697
 Header
 GMP_SIO_READ_CONFIG_ERROR
 Definition
 Failed to read SIO config data
 Action
 Service
 Check the communication linkage
 Production
 Same as service
 Product
 Same as service
 Remark
 Locator is able to check the communication linkage
698
 Header
 GMP_HALL_SNR_COMM_REQUIRED
 Definition
 Hall sensor commutation is required
 Action
 Service
 Check whether using correct SCF and activation file
 Production
 Same as service
 Product
 Ensure hall sensor commutation was performed successfully
 Remark
 Locator is able to show the commutation information
2016
 Definition
 Driver NOT on when running motion command
 Action
 Service
Ensure the channel is ON before any motion command
Turn it ON and retry the motion
 Production
Ensure the channel is ON before any motion command
Turn it ON and retry the motion
 Product
Ensure the channel is ON before any motion command
Turn it ON and retry the motion
 Remark
 Locator is able to tell any protection is triggered that may cause
missing turn ON
9012

 Definition
 Invalid port is in use when setting the datalog port entries
 Action
 Service
Report to product team
 Production
Report to product team
 Product
Modify the port name of SCF
Modify the port name in the program
Modify the activation file
 Remark
 Locator is able to tell the invalid port name user enterd
9016

 Definition
 The datalog possession Key length is too long
 Action
 Service
Report to product team
 Production
Report to product team
 Product
Correct the key such that the length is less than
129 characters. The length doesn't longer than 128
characters
 Remark
 Locator is able to tell the datalog key users entered