Mbs 3802708
Mbs 3802708
Mbs 3802708
Cat.No: MBS3802708
Storage:2-8°°C.
om
validity:six months.
c
Standard concentration was followed by: 20,10,5,2.5,1.25,0ng/mL
e.
Both intra-assay CV and inter-assay CV is less than 15%.
rc
ou
Intended use
oS
This Human oxyhemoglobin kit is intended Laboratory for Research use only and is not for use in
diagnostic or therapeutic procedures. The Stop Solution changes the color from blue to yellow and the
i
intensity of the color is measured at 450 nm using a spectrophotometer. In order to measure the
yB
concentration of Human oxyhemoglobin in the sample, this Human oxyhemoglobin Kit includes a set of
M
calibration standards. The calibration standards are assayed at the same time as the samples and allow
the operator to produce a standard curve of Optical Density versus Human oxyhemoglobin
om
collection. Store samples at -20°C or -80°C.
freeze-thaw cycles.
Note: The samples should be centrifuged adequately and no hemolysis or granule was
c
allowed.
3. 37 °C incubator.
i
yB
Precautions
M
Materials supplied
om
Closure plate membrane 2 2
User manual 1 1
c
Sealed bags 1
e. 1
rc
ou
Reagent preparation
oS
Assay procedure
2. Add standard: Set Standard wells, testing sample wells. Add standard 50µl to standard well.
3. Add Sample: Add testing sample 10µl Then add sample diluent 40µl to testing sample well; Blank
minutes at37°C.
5. Aspirate each well and wash, repeating the process four times for a total of five washes.Wash by
filling each well with Wash Solution (400µl) using a squirt bottle, manifolddispenseror auto washer.
Complete removal of liquid at each step is essential to good performance. After the last wash, remove
any remaining Wash Solution by aspirating ordecanting. Invert the plate and blot it against clean paper
towels.
6. Add chromogen solution A 50µl and chromogen solution B 50µl to each well. Gently mix and
om
incubate for 15 minutes at 37°C. Protect from light.
7. Add 50µl Stop Solution to each well. The color in the wells should change from blue toyellow. If the
c
e.
color in the wells is green or the color change does not appear uniform,gently tap the plate to ensure
thorough mixing.
rc
ou
8. ReadtheOpticalDensity(O.D.)at450nmusinga microtiterplatereaderwithin15minutes.
Calculation of results
oS
1. This standard curve is used to determine the amount in an unknown sample. The standard curve is
i
yB
generated by plotting the average O.D. (450 nm) obtained for each of the six standard
concentrations on the vertical (Y) axis versus the corresponding concentration on the horizontal (X)
M
axis.
2. First, calculate the mean O.D. value for each standard and sample. All O.D. values, are subtracted
by the mean value of the zero standard before result interpretation. Construct the standard curve
3. To determine the amount in each sample, first locate the O.D. value on the Y-axis and extend a
horizontal line to the standard curve. At the point of intersection, draw a vertical line to the X-axis
age can cause variation in result. Each user should obtain their own standard curve.
5. Standard curve
c om
e.
rc
ou
i oS
yB
BEGINNING!