BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 1 of 40
Contents:
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 2 of 40
1. KEY FEATURES
BX10 MB
BX11 MB
BX11 CO
BX11 AN
BX11 PN
BX11 EN
BX11 PB
BX10
BX11
1 000 to 999 999 display resolution Yes Yes Yes Yes Yes Yes Yes Yes Yes
High internal resolution up to 16 000 000 counts Yes Yes Yes Yes Yes Yes Yes Yes Yes
Up to 1600 conversion per second Yes Yes Yes Yes Yes Yes Yes Yes Yes
Serial interface RS 232C Yes Yes Yes Yes Yes Yes Yes Yes Yes
Serial interface RS 485 - Yes Yes Yes Yes Yes Yes Yes Yes
Analogue output - - - Yes - - - - -
Profibus DPV1 interface - - - - - Yes - - -
Profinet interface - - - - - - Yes - -
Ethernet interface - - - - - - - Yes -
CANopen interface - - - - - - - - Yes
Continuous data output Yes Yes Yes Yes Yes Yes Yes Yes Yes
Fast Continuous data output Yes Yes Yes Yes Yes Yes Yes Yes Yes
BSI data interface Yes Yes Yes Yes Yes Yes Yes Yes Yes
Modbus RTU - Yes - Yes Yes Yes Yes Yes Yes
Modbus TCP - - - - - - - Yes -
2 programmable digital input/output
Yes Yes - - - - - - -
(non-isolated)
4 digital input and 5 relay contact output - - Yes Yes Yes Yes Yes Yes Yes
Error and at zero outputs (non-isolated) - - Yes Yes Yes Yes Yes Yes Yes
Bidirectional signal input for force measurement Yes Yes Yes Yes Yes Yes Yes Yes Yes
Unit selection (g, kg, t, lb, klb, N, kN ) Yes Yes Yes Yes Yes Yes Yes Yes Yes
Peak function - - Yes Yes Yes Yes Yes Yes Yes
Hold function - - Yes Yes Yes Yes Yes Yes Yes
Auto-zero tracking and auto-zero at
Yes Yes Yes Yes Yes Yes Yes Yes Yes
power-up
Motion detection Yes Yes Yes Yes Yes Yes Yes Yes Yes
Zeroing and Taring via interface Yes Yes Yes Yes Yes Yes Yes Yes Yes
Adaptive digital filter for faster measuring Yes Yes Yes Yes Yes Yes Yes Yes Yes
Electronic calibration (eCal) without test weights Yes Yes Yes Yes Yes Yes Yes Yes Yes
Electronic calibration (eCal) over field bus - Yes - Yes Yes Yes Yes Yes Yes
Zero and Span calibrations over field bus - Yes - Yes Yes Yes Yes Yes Yes
Zero adjustment Yes Yes Yes Yes Yes Yes Yes Yes Yes
Span adjustment with test weights Yes Yes Yes Yes Yes Yes Yes Yes Yes
Span adjustment for filled tanks Yes Yes Yes Yes Yes Yes Yes Yes Yes
3 point calibration ( linearity correction ) Yes Yes Yes Yes Yes Yes Yes Yes Yes
Programming by BAYKON IndFace1X PC software Yes Yes Yes Yes Yes Yes Yes Yes Yes
8 load cells 350 Ω or 18 load cells 1100 Ω Yes Yes Yes Yes Yes Yes Yes Yes Yes
12 to 28 VDC power supply range Yes Yes Yes Yes Yes Yes Yes Yes Yes
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 3 of 40
2. THE FRONT VIEW AND KEY FUNCTIONS
2.1 Display
The weight display of BX11 is seven segments LED. At the right side of the display there are three
LED’s for indicating the net, gross and the unit ( standard kg ). The meanings of the announcement
LED’s on the display are:
Units g, kg, t, lb, klb, N, kN units are located on the right of the
display.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 4 of 40
2.2 Key Pad
The keys and the key functions of BX11 are:
GN / Set Point: Pressing this key indicates the Gross weight temporarily. Press long
time to enter the setpoint menu.
Tare / Clear: Pressing this key tares the scale and get into the Net mode. Press long
time to clear the tare.
Zeroing: In Gross mode, if the scale doesn’t show zero while there is no load on the
pan, you can zero the scale by pressing this key.
Print: By pressing this key weight data and other information depending on the setup
parameters sent to a printer or a PC via serial port.
2.4 Housing
BX11 housings are panel type with stainless steel front and back parts, and aluminum body.
92 mm ( 3,62" )
47 mm
( 1,85" )
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 5 of 40
100 mm ( 3.94" ) 99 mm ( 3.90" )
55 mm 46 mm
( 2.16" ) ( 1.81" )
BX11 AN Panel type rear view BX11 & BX11 MB Panel type rear view
BX11 PB Panel type rear view BX11 PN Panel type rear view
BX11 EN Panel type rear view BX11 CO Panel type rear view
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 6 of 40
2.5 Accessories supplied with the instrument
The following accessories are supplied together with the BX11 instruments. If any part is missed,
please contact to your supplier.
BX11 MB
BX11 CO
BX11 AN
BX11 PN
BX11 EN
BX11 PB
BX11
3-pos and 3,81 mm pitch green plug
1 2 1 1 1 1 1
for power supply and analogue output
7-pos and 3,81 mm pitch black plug for
1 1 1 1 1 1 1
load cell cable
6-pos and 3,81 mm pitch black plug for
3 3 3 3 3 3 3
RS 232C & RS 485 and isolated I/O
2-pos and 3,81 mm pitch green plug for
1 1 1 1 1 1 1
non-isolated outputs
User manual 1 1 1 1 1 1 1
Installation CD (IndFace1X setup,
1 1 1 1 1 1 1
technical manual and technical documents)
BX11 CO
BX11 AN
BX11 PN
BX11 EN
BX11 PB
BX11
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 7 of 40
3. INSTALLATION
PRECAUTION: Please read this section carefully before installation of the instrument. Applying the
recommendations in this section will increase your system reliability and long term performance.
3.1 Recommendations
Control Cabinet Design
Warning: Please care the following warnings for designing the control cabinet which will increase
your system reliability.
The control cabinet should be designed so that Analog Digitizer can operate safely. The panel
should be placed clean area, not getting direct sun light if possible, with a temperature between
-15 ºC and +55 ºC, humidity not exceeding 85% non-condensing. All external cables should be
installed safely to avoid mechanical damages.
BX11 instruments are very low level signal measuring instruments. To avoid electrical noise, BX11
should be separated from the equipments that produce electrical noise. Preferable use metal
cabinet against radio frequency interference and the cabinet shall be connected to ground against
the electromagnetic disturbances. Load cell cable trays must be separated from others, if possible.
If there are noise-generating equipments such as heavy load switches, motor control equipments,
inductive loads etc., please be careful against the EMC interference in the cabinet. If possible
protect BX11 instruments with the faraday cage or install them in separate section or install them
far a way from this kind of equipments. Connect parallel reverse diodes to the DC inductive loads
like relays, solenoids etc. to minimize voltage peaks on the DC power lines.
Cabling
All cables coming to the control cabinet shall be shielded. Please use separate cable tray for these
low signal level cables. Distance from load cell cables, interface cables and DC power supply cables
to power line cables shall be minimum 50 cm.
Warning: Please always remember that BX11 instruments are very low voltage measuring
instruments. Your control cabinet design and proper installation increases reliability and
performance of the instrument. Please do not forget that the instrument must be powered off before
inserting or removing any peripheral connector.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 8 of 40
Step 1. Mechanical Installation
Take care to the housing dimensions and the suggested panel hole dimensions given in the Page 5.
Install the indicator on the panel with its panel fixing accessories. Be sure, all mechanical
installation of your mechanical system and panel are finalizing before starting the next step.
Load cell connection detail is shown in Figure 3.1. In 4-wire installations the sense and excitation
pins with the same polarity should be short circuited at the connector side. If you have junction
box, use 6 wire cable between BX11 and the junction box, and short circuit these pins at junction
box for better performance.
Warning: Always connect Sense pins to Excitation pins for 4 wire connection. Non-connected
sense pins may cause the wrong Excitation voltage measurement and create an accuracy problem.
Warning: Connect the load cell cable shield to the reference ground or shield pin of the load cell
connector.
Power supply voltage of the instrument shall be between 12 VDC and 28 VDC. The pin configuration
of the 24 VDC power supply connector located right - bottom of the instrument is shown in Figure
3.2 below. Be sure, the power supply is switched of before connection.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 9 of 40
The pin layout of the 24 VDC
connector of BX11 Series
( rear view )
Warning: Before interfering the instrument, turn off the power and wait at least for 30 seconds.
Warning: Connect the grounding pin to the reference ground.
If there is a DIP switch on BX11’s rear side and its position should be “ON” ( downward ) to change
the metrological related parameters including calibration. No need to open the housing to change
the position of this DIP switch.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 10 of 40
Step 6. Programming
You will program the instrument according to your application in this step. Adjust;
• Interface parameters in block [ 00- ] RS232 and [ 01- ] RS485 serial ports,
• Configuration parameters in block [ 1-- ] for scale functions related with application like
saving tare before power off and restart in Net mode and digital I/O related entries.
• Set up parameters to set up the scale in block [ 2-- ] like zeroing range. Please remember
these parameter values are restricted for scale usage in trade. You have to enter the scale
capacity and devision values in this block.
Warning: The programming and parameter descriptions are referred in Page 19. We recommend
you to save these parameters before next step against to lost the adjustments in this step if
calibration can not performed somehow.
Step 7. Calibration
You will calibrate the scale after programming the parameters. You may follow one the calibration
methods below.
• eCal Electronic calibration without test weight need via IndFace1x or fieldbus ( Page 22 ).
• Full calibration via keys on the instrument by using test weight. Refer to Page 22.
• Zero adjustment and span adjustment in sequence via keys or fieldbus. Span calibration is
needed test weight. You may save the zero adjustment before starting span adjustment.
Refer to Page 24.
• Zero adjustment and span adjustment under load in sequence via keys if the scale cannot be
emptied when the span adjustment is performed. Need test weight for span calibration. Refer
to Page 25.
You may access to the calibration menu from operation menu by following the description in fast
access to the calibration section ( Page 24 ). After the calibration go back to the operation menu
after saving adjustment ( Refer to Page 19 )
Remember: You may use parameter 905 to follow the load cell output voltage on loading, if you can
not perform the calibration.
Step 9. Bring the DIP switch to up position to unlock the scale adjustment
If the DIP switch is bring to the downward position for programming, push it to the upward position
to lock the scale adjustment against interfaring the un-autorized persons in to the scale. If the scale
is used in non-trade industrial weighing, this step is not must.
Warning: If the scale is used in trade, the scale should be sealed after bringing the DIP switch to
the upward position sealed before. Call the authorized person for sealing the scale, if need be.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 11 of 40
Step 10. Peripheral related parameters programming, if any
If you will connect any peripheral to the instrument, you have to set the related parameters up.
Refer to Page 12 for digital input/output, Page 25 for analog output programming and technical
manual for fieldbus interfacing.
Important : Powered off the instrument before any connector installing to the instrument.
Non-isolated outputs are power supply voltage – 1.5V, 100mA. Connection diagram is shown in
Figure 3.6.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 13 of 40
Standard Output:
Only one set point value is entered.
The output state is forced active high
when the weight is higher than SP1,
else the output is passive.
Threshold Output:
2 set point values are entered. SP1 is
the point that the output goes active
when the weight increased from
SP1_H. SP1_L is the point that the
output drops to passive state when
the weight decreased to SP1_L.
Inverse function is available.
Window Output:
2 set point values are entered. The
output is active when the weight is
between SP1_L and SP1_H.
Inverse function is available.
RS 232C Connection
RS 232C port usage and specifications are shown in the table below.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 14 of 40
RS 232C serial connection is done with three wire as indicated below in Figure 3.7.
Warning: Connecting the shield to the reference ground will protect your weighing system against
EMC disturbances.
RS 485 serial connection is done with three wire as indicated below in Figure 3.8. Line termination
resistor ( 110 ohm ) is needed both end of the RS 485 line.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 15 of 40
Warning: Connect the shield to the reference ground.
Warning: Disconnect IndFace1X PC software before starting Modbus-RTU interfacing.
Analogue connections are done as indicated below in Figure 3.9 and Figure 3.10.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 16 of 40
PROFIBUS Connector pin configuration (DB9F)
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 17 of 40
Step 12. Peripheral connections testing
Test your peripheral connections. You may change the related parameters in your testing for better
performance, if need be.
You may use also main parameter group [ 9-- ] for RS232, RS485, and digital In / Out testing.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 18 of 40
4. PROGRAMMING AND CALIBRATION
The signs those take place on the lower right corner of the keys indicate the function of the keys in
programming menu. The basic meanings of these keys are given the table below.
• Return to main Advancing next Select the digit will Changing parameter • Go to the next
block parameter be changed value or increasing parameter
• Exit without the blanking digit • Enter
saving • Save and Exit
Display Operation
[123.456 kg] Press key until [ PASSWr ] prompts seen.
Programming and Calibration menu consist of main blocks which are shown as [X-- ] and sub-
blocks. By using < ↑ > key you can reach next main blocks. After reaching the desired main
block, you can get in by pressing < Enter > key. As you enter the block you will reach the first sub-
block in that main block. The sub-block address will be seen on the display as [X0- ]. You can
also search between the sub-blocks by using < ↑ > key and reach the first parameter of the sub-
block seen on the display by pressing < Enter > key. The number of the parameter comes on
display as [XY0 ]. Again you can search between parameters by < ↑ > key. For entering
numerical value in the parameters, press the < Tare > key to select the digit and press the
< Zero > key the change the value.
You may adjust your parameters and calibrate the scale by using the diagram below:
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 19 of 40
Press long time
PASSWr + + sequentially
---
INTERFACE 0_ _ 00_ 000 3 001 3 003 00 004 0 005 0 006 1 007 1 008 0
RS232 Data Format Baudrate Address Length&Parity Checksum CarrigeReturn Line Feed Response S.
0 : No 0 : 1200 0 : 8 None 1 0 : Disable 0 : Disable 0 : Disable 0 : Immediately
1 : Continuous 1 : 2400 1 : 7 Even 1 1 : Enable 1 : Enable 1 : Enable 1 : 20ms delay
2 : Print 2 : 4800 2 : 7 Odd 1
3 : BSI 3 : 9600 3 : 8 Odd 1
4 : Modbus HL 4 : 19200 4 : 8 Even 1
5 : Modbus LH 5 : 36400
6 : Fast Cont. 6 : 57000
7 : 115200
*
*
*
*
Address 0 : Immediately
1 : 20ms delay
2 : 50ms delay
06_ 060 0
Profinet Data Format
0 : Integer
1 : Floating
rate
SETUP 20_ 200 0 201 0 202 0 203 3 204 0 205 2 206 2 207 0.3
2_ _
PARAMETERS Setup Approved x10 indication Power on zero Zeroing range Auto zero Tare Motion detector Stability period
0 : No 0 : by x10 key 0 : Disable 0 : Disable tracking 0 : Disable 0 : ± 0,3e 0.1 sn
1 : OIML 1 : Always x10 1 : ± 2% 1 : ± 2% 0 : Disable 1 : Multitare 1 : ± 0,5e :
2 : Hopper&Tank 2 : ± 10% 2 : ± 20% 1 : ± 0,5e 2 : Tare/Clear 2 : ± 1e :
3 : ± 50% 2 : ± 1e 3 : ± 2e 9.9 sn
3 : ± 3e 4 : Disable
2:t 6 : kN
3 : lb
Next : Advancing next parameter.
** Changeable
look the Techical Manual for details.
2 : 0-5 VDC
3 : 0-10 VDC
DIGITAL
OUTPUT 7_ _ 70_ 700 0 701 0 702 0 703 0 704 0
FUNCTIONS Outputs Output-1 Output-2 Output-3 Output-4 Output-5
0 : No 3 : Window (Active High)
1 : Threshold (Active High) 4 : Window (Active Low)
2 : Threshold (Active Low) 5 : Fieldbus output Page 27
: YES
SAvE
: NO
Calibration
1. Press < Enter > at the [ 301 ] prompt to start the calibration.
2. At the [ ZEro.CA ] prompt, remove any weight on the platform, then press < Enter >.
3. The terminal automatically starts to capture zero and the [ WAıt ] message indicating
the operation is in progress.
4. After the [ Load ] prompt, the test weight value will be used for the calibration seen
on the display as [ XXXXXX ] . If the value of the test weights that will be used is
different from the value shown on the display, type the new value via < Tare > and
< Zero > keys. A minimum of 20% of scale capacity is necessary for calibration;
BAYKON recommends 50 to 100%. A calibration error will result if insufficient weight
is used.
5. Place the test weights or another practical weight on the scale.
6. Press < Enter > to start span calibration. [ WAıt ] message will shown on the display
10 seconds while span calibration is being performed.
7. At the [ SAvE ] prompt press < Enter > key to continue to the next parameter or
press < F > key to exit without saving the calibration.
Note: Refer to Technical manual for multipoint calibration to correct the linearity.
This parameter lets you to perform calibration without using any test weights. BX11 A/D
coefficients are adjusted in production for increasing eCal accuracy. The calibration
coefficients are calculated by scale capacity, total load cell capacity, load cell full scale
output, and estimated dead load.
Access to the parameter 313 and enter the values below as;
[LC.CAP ]
[XXXXXX]
Enter total load cell capacity via < Tare > and < Zero > keys and press < Enter > key to go
to the next step.
Example : If the weighing system has 4 pcs 1000 kg load cell, enter 4000.
[LC.oUt ]
[XXXXXX]
Enter load cell output in mV/V via < Tare > and < Zero > keys. If the weighing system has
more than one load cell, calculate the mean value of load cells outputs mV/V indicated on the
certificates of the load cells. Press < Enter > key to go to the next step.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 22 of 40
Example: If load cell outputs are LC1: 2.0010, LC2: 1.9998, LC3:1.9986 and LC4:2.0002, the
mean value will be
Mean of LC outputs = ( 2.0010 + 1.9998 + 1.9986 + 2.0002 ) ÷ 4 = 1.9999 mV/V.
[ZEr.AdJ]
[XXXXXX]
If the scale is empty and you want to make automatic zero adjustment instead of entering
estimated dead load ( look next step ), press < Enter > key appears [ Zero.CA ] and press
< Enter > key for starting zero calibration. The display will show [ WAıt ] message during
zero adjustment. In this while the scale must be unloaded and stable. Approximately 10
seconds later display will prompt you to save the calibration by [ SAvE ] message below.
If the scale is not empty or you prefer to enter estimated preload value, press the up
< G/N > key.
[PrE-Ld]
[XXXXXX]
Enter the dead load value of the weighing system in current unit by using < Tare > and
< Zero > keys. Press the < Enter > key to go to the next step.
[ SAvE ]
Save your eCal calibration by pressing < Enter > key or press < F > key to go out without
saving your eCal calibration.
Note: If you want to make zero adjustment after entering estimated preload value, empty the
scale, change the preload value as (estimated value + display value at empty scale ) or enter
parameter [ 310 ] for zero adjustment.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 23 of 40
4.3 Fast Access to the Calibration
The instrument has fast access calibration feature to earn time to the service technician. If only the
calibration adjustment is needed, follow the steps below to access the calibration parameters fast.
Display Operation
[31-] Adjustment
In this sub-block you can only perform zero adjustment or span adjustment without full calibration
operation.
1. Press < Enter > at the [ 310 ] prompt to start the zero adjustment.
2. At the [ ZEro.CA ] prompt, remove any weight on the platform, then press < Enter >.
3. The terminal automatically starts to capture zero and the [ WAıt ] message indicating
the operation is in progress.
4. Press < Enter > to start zero adjustment. [ WAıt ] message will shown on the display
nearly 10 seconds.
5. At the [ SAvE ] prompt press < Enter > key to continue to the next parameter or press
< F > key to exit without saving the calibration.
1. Press < Enter > at the [ 311 ] prompt to start the span adjustment.
2. At the [ XXXXXX ] prompt, the test weight value will be used for the calibration seen
on the display. If the value of the test weights that will be used is different from the
value shown on the display, type the new value via tare and zero keys. A minimum of
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 24 of 40
20% of scale capacity is necessary for calibration; BAYKON recommends 50 to 100%.
A calibration error will result if insufficient weight is used.
3. Place the test weights or another practical weight on the scale.
4. Press < Enter > to start span calibration. [ WAıt ] message will shown on the display
10 seconds while span calibration is being performed.
5. At the [ SAvE ] prompt press < Enter > key to continue to the next parameter or press
< F > key to exit without saving the calibration.
1. Press < Enter > at the [ 312 ] prompt to start the span adjustment under load.
2. [ P.ZEro ] prompt appears on the display to indicate the scale load will be determined
as temporary zero.
3. Press < Enter > key and the display will show [ WAıt ] message during temporary
zero adjustment.
4. At the [ LoAd ] a little while and then [ XXXXXX ] prompt, the test weight value will
be used for the calibration seen on the display. If the value of the test weights that
will be used is different from the value shown on the display, type the new value via <
Tare > and < Zero > keys.
5. Place the test weights or another practical weight on the scale.
6. Press < Enter > to start span calibration. [ WAıt ] message will shown on the display
10 seconds while span calibration is being performed.
7. At the [ SAvE ] prompt press < Enter > key to continue to the next parameter or
press < F > key to exit without saving the calibration.
The calibration of the analogue output is performed automatically after calibration the scale. If you
want to adjust the analog output manually access to the parameters below as seen at the in the
programming parameters flow diagram ( Page 20 ) and follow the descriptions below.
Press < Zero > key one after another to increase the analogue signal level or press
< Tare > key to decrease the analogue signal level.
Press < Zero > key continuously one after another to increase the analogue signal level or
press < Tare > key to decrease the analogue signal level.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 25 of 40
[420] Coarse Span Adjustment
Press < Zero > key continuously one after another to increase the analogue signal level or
press < Tare > key to decrease the analogue signal level by taking the full capacity value
as reference without placing any weight.
Press < Zero > key continuously one after another to increase the analogue signal level or
press < Tare > key to decrease the analogue signal level by taking the full capacity value as
reference without placing any weight.
[130 0 ] Outputs
The outputs operate with respect to the weight value seen on the display.
0 : No.
1 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Sp5 )
2 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Stable )
3 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Error )
4 : Not used
5 : Functional outputs ( Refer to [ 7-- ] parameter group )
[131 0] Input 1
[132 0] Input 2
[133 0] Input 3
[134 0] Input 4
In this block, the digital outputs are programmed for the functions indicated in the table below.
Threshold Output :
2 set point values are entered. SP1 is
the point that the output goes active
when the weight increased from
SP1_H. SP1_L is the point that the
output drops to passive state when
the weight decreased to SP1_L.
Inverse function is available.
Window Output:
2 set point values are entered. The
output is active when the weight is
between SP1_L and SP1_H. Inverse
function is available.
[700 0] Output 1
0 : No function on output 1
1 : Threshold (Active High)
2 : Threshold (Active Low)
3 : Window (Active High)
4 : Window (Active Low)
5 : Fieldbus output
[701 0] Output 2
0 : No function on output 2
1 : Threshold (Active High)
2 : Threshold (Active Low)
3 : Window (Active High)
4 : Window (Active Low)
5 : Fieldbus output
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 27 of 40
[702 0] Output 3
0 : No function on output 3
1 : Threshold (Active High)
2 : Threshold (Active Low)
3 : Window (Active High)
4 : Window (Active Low)
5 : Fieldbus output
[703 0] Output 4
0 : No function on output 4
1 : Threshold (Active High)
2 : Threshold (Active Low)
3 : Window (Active High)
4 : Window (Active Low)
5 : Fieldbus output
[704 0] Output 5
0 : No function on output 3
1 : Threshold (Active High)
2 : Threshold (Active Low)
3 : Window (Active High)
4 : Window (Active Low)
5 : Fieldbus output
Diagnostics
The tests and measurings below are available to find the error source.
Short circuit the RXD and TXD pins of the RS232 serial port. Access to the parameter [901].
The characters in the alphabet will be seen double on the display by pressing < Zero > key
sequentily.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 28 of 40
RS 485 Serial Interface testing
Access to the parameter [902].The characters in the alphabet will sequentially be transferred
from RS 485 serial interface port by pressing < Zero > key one after another. Received
numerical data is also seen on display.
Parallel Inputs
Parallel Outputs
Measuring load cell signal value is very difficult after installation the system. You can
measure the load cell signal value at parameter [905] without disconnecting the load cell
connector.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 29 of 40
5. ANALOGUE ( ONLY BX11 AN )
BX11 AN is programmable to 4 – 20 mA, 0 – 20 mA, 0 – 5 V or 0 – 10 V analogue output types.
Analog output is automatically adjusted to the weighing range after the calibration. The mid value of
the analog output is set to zero load at bipolar usage. The manual analog output adjustment is
available in parameter group [ 4-- ] ( Refer to Page 25 and technical manual ).
The analogue output is related with the gross load of the scale. The analog output signal operates
as described next.
When the gross indication drops below zero, the analog output
Under Zero reduces the analog output to 0mA or - 4 V to indicate error on the
analog output.
The analog output will reflect the gross value to the programmed
Normal Range analog output 4 – 20 mA, 0 – 20 mA, 0 – 5 V or 0 – 10 V.
When the gross value exceeds the high limit, the analog signal
increase to approximately 24 mA or 11 V and remains there until the
Over High Limit
weight display is no longer blanked or the analog signal returns to
within range.
The following table indicates the analog output value when the gross indication is out of the range
and if there is any error indication on the display.
The error data indicated above can be used to follow the errors at PLC.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 30 of 40
6. RS232C AND RS485 DATA OUTPUTS
In this section, you will find the data structure of different type of the data outputs via these serial
ports.
The including of the status bytes STA, STB and STC are ;
Definition Table for Status A ( STA )
Bits 0, 1 and 2 Bits 3 and 4 Bit 5 Bit 6 Bit 7
0 1 2 Decimal point 3 4 Increment size
0 0 0 XXXXOO 1 0 X1
1 0 0 XXXXXO 0 1 X2
0 1 0 XXXXXX 1 1 X5
Always 1
Always 1
1 1 0 XXXXX.X X
0 0 1 XXXX.XX
1 0 1 XXX.XXX
0 1 1 XX.XXXX
1 1 1 X.XXXXX
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 31 of 40
CHK (Checksum) = 0 – ( STX + STATUS A + ..... + LF )
Error Messages: UNDER, OVER, A.OUT, L-VOLT, H-VOLT, are represented in Indicated data fields.
Note: The weight data is represented with right aligned, error messages are represented with left
aligned.
Important : The CR and LF shall be enabled for Baykon remote display interfacing.
Examples :
S+000123.4 (weight is stable and 123.4)
D+000123.4 (weight is dynamic and 123.4)
+ (Over load)
- (Under load)
O (ADC out error)
Single Line
You can send the data in single line like below by pressing < Enter > key as;
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 32 of 40
Multi Line
You can send the data in multiple lines as seen in the label given below by pressing < Enter > key.
The data output structure can be programmed with printer parameters.
Command Table:
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 33 of 40
7. TECHNICAL SPECIFICATIONS
Common Specifications
A/D Converter
A/D converter type 24 bit Delta-Sigma ratiometric with integral analog and digital filters
Conversion rate Up to 1600 measurement values per second
Input sensitivity 0.4 μV/d ( Approved ); 0.1 μV/d (Non approved )
Analog input range 0 mV to +18 mV ( unipolar ) ; - 18 mV to +18 mV ( bipolar )
Internal resolution up to 16 000 000
External Resolution
Display resolution up to 10 000 increment (Approved ); up to 999 999 increment (Non approved )
Scale Calibration and Functions
Calibration Calibration is performed with or without test weights ( eCal )
Digital filter 10 steps programmable adaptive filter
Taring, zeroing, auto zero tracking, motion detection, auto zero at power up, net
Weighing functions
indication at power on, increased resolution
Linearity
Within 0.0015% FS, ≤ 2 ppm/°C
Load cells
Excitation 5 VDC max. 300 mA
Number of load cells Up to 8 load cells 350 Ω or 18 load cells 1100 Ω in parallel
4- or 6-wire technique.
Connection
Cable length: maximum 1000 m/mm² for 6-wire connection
Communication
RS-232 1200 to 115200 baudrate, 8N1 / 7O1 / 7E1 / 8O1 / 8E1
Response speed Up to 4 ms. response delay after read/write commands
Digital Inputs and Outputs
Error & at Zero outputs U0= Power supply voltage – 1.5V , 100mA. Non-isolated transistor outputs.
Power consumption
12 to 28 VDC max. 300 mA
Environment and Enclosure
Operation temperature -15 °C to +55 °C; 85% RH max, non-condensing
Panel type, front panel and rear panel are stainless steel;
Enclosure
Aluminum body.
Protection Front panel IP65
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 34 of 40
BX10 MB Modbus-RTU
Communication
RS-485 1200 to 115200 baudrate, 8N1 / 7O1 / 7E1 / 8O1 / 8E1
Response speed Up to 4 ms. response delay after read/write commands
Max Stations Up to 31 stations per segment
Digital I/O Ports
2 ports can be programmed as digital input or digital output, non-isolated.
BX11 AN Analogue
Communication
Voltage output 0-5 VDC, 0-10 VDC
Current output 4-20mA, 0-20mA
D/A Converter 16 bit
Max. cable length 300 meter
Max. load resistance
500 Ω
(current output )
BX11 MB Modbus-RTU
Communication
RS-485 1200 to 115200 baudrate, 8N1 / 7O1 / 7E1 / 8O1 / 8E1
Response speed Up to 4 ms. response delay after read/write commands
Max Stations Up to 31 stations per segment
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 35 of 40
BX11 PN Profinet
Communication
Data rate 100 Mbit/s, full duplex
GSDML file Generic GSDML-file provided
DHCP or manual IP assign over EtherX PC Software.
TCP/IP settings
Device identity customization
Topology Line, Bus, Star or Tree topology depending on physical media
Switched Ethernet transmission with shielded twisted pair cables and RJ-45
Installation
connectors.
Isolation Galvanically isolated bus electronics
Response speed Up to 4 ms. response delay after read/write commands.
BX11 EN Ethernet
Communication
Transmission rate 10 Mbit/s, half duplex
TCP/IP settings Manual IP assign over EtherX PC Software or by keys in programming mode.
Switched Ethernet transmission with shielded twisted pair cables and RJ-45
Installation
connectors.
Web client Available
Response speed Up to 4 ms. response delay after read/write commands
BX11 CO CANopen
Communication
Data rate 10 kbit/s – 1 Mbit/s (selectable) kbit/s
ESD file Generic EDS-file provided
Line with Trunkline, Dropline structure and Termination at both Ends
Topology
Line length depending on baudrate 25 – 500 meter.
2 wires shielded twisted pair cable.
Installation
Alternatively 4 wire with 24 Volt power over the bus
Max. Stations Up to 127 stations per network
Isolation Galvanically isolated bus electronics
Response speed Up to 4 ms. response delay after read/write commands
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 36 of 40
8. TROUBLE SHOOTING
BX11 weighing indicator had been designed as a very reliable and virtually error free instrument.
However if there is an error occurs, do not attempt to repair the equipment before understanding
what caused the error. Note the problems you have with your instrument and the error messages
shown on the display. Then try to solve the problem according to the error table given below.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 37 of 40
9. FREQUENTLY ASKED QUESTIONS
Question : My PC could not interface with BX11. How can I check the com port?
Answer : − Connect the instrument to the PC and run Hyper Terminal.
− Check com ports as described in The Diagnostic Test section on Page 28.
Question : IndFace1X installation needs restart every time. How can I install it?
Answer : − Read and follow the installation notes in the installation directory.
− Update your computer (visit http://update.microsoft.com).
Question : IndFace1X could not connect to instrument. What can I do?
Answer : − Check the power, data cabling and PC port setting.
− Remove other connections. Re-energize the BX11 instrument and then make connection.
Question : My PC doesn’t have any COM port. How can I connect instrument to my PC?
Answer : You can use RS-232 / USB converter for serial interfacing via USB port. And select com port
with Connection Settings menu.
Question : My PC have a COM port but I couldn’t see COM port in Connection Setting menu. How can I
solve that problem?
Answer : Another software may be connected to that COM port. Close all applications before running
IndFace1X.
Question : My PC could not interface with BX11. How can I check the com ports?
Answer : Short circuit your com port RXD and TXD pins. Check if the sending data is received or not
by using any terminal software. You may test also BX11 com ports as described in The
Diagnostic Tests section on Page 28 by short circuiting RXD and TXD terminals.
Question : I need very fast interfacing. What is the response delay time of BX11?
Answer : BX11 response delay is max. 4 milliseconds for weight data. Extremely fast interfacing.
Question : What is the external conversion rate of BX11?
Answer : Only continuous data output rate might be called as an external conversion rate which is
depend on the baud rate and data length and up to 65 conversion/second.
Question : How can I check Ethernet connection?
Answer : − BX11EN has a dummy web page. You can easily open web page with any browser
installed on any PC in network.
Question : EtherX is searching but it could not find any instrument over Local Network Area.
Answer : − Check the Exceptions tap in Windows firewall settings. EtherX should be marked.
− Check the firewall of Anti-virus program, If you have.
Question : I done crossover connection with BX11 EN but EtherX could not find it.
Answer : − Check the Internet Protocol (TCP/IP) Properties in Local Area Connection of Windows.
IP address blocks and gateway address of BX11 and PC should be the same in cross
connection.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 38 of 40
Declaration of Conformity
DECLARATION OF CONFORMITY
Uygunluk Beyanı
AT Tip Onayı
DK 0199.452
[code]
M
Baykon, March 2014 Baykon, March 2014
Mart 2014 Mart 2014
Weighing instruments verified at the place of manufacture bear the preceeding mark on the packing label and
a green "M" sticker on the descriptive plate. They may be set to work immediately.
M Üretim yerinde doğrulanan tartı aletleri, ambalajında yandaki işareti ve tanıtım plakasında yeşil "M" etiketi taşırlar. Hemen
kullanıma alınabilirler.
Weighing instruments which are verified in two steps have no green "M" on the descriptive plate, bear the
aforementioned identification on the packing label. The second step of the verification must be carried out by
the Baykon or by another authoritiesed firm. Please contact to Baykon.
M Đki aşamalı olarak doğrulanan tartı aletlerinin tanıtım plakasında yeşil "M" etiketi yoktur, sadece ambalajında yandaki işareti
taşırlar. Doğrulamanın ikinci aşaması Baykon veya diğer bir yetkili kuruluş tarafından yerine getirilmelidir. Lütfen Baykon'la
bağlantıya geçiniz.
The first step of the verification has been carried out in the manufacturing company. It comprises all tests according EN
45501-8.2.2. If national regulations in individual countries limit the period of validity of the certification, the operator of such
a scale is himself responsible for its timely re-certification.
Doğrulamanın birinci aşaması üretici firmada gerçekleştirilmiştir. Bu, TS EN45501-8.2.2 'de yer alan tüm testleri içerir. Eğer tartı aletinin
kullanıldığı ülkelerde ulusal hükümler sertifikanın geçerlilik periyodunu sınırlıyorsa, sertifikanın yenilenmesi tartı aletininin kullanıcısının
sorumluluğundadır.
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 39 of 40
NOTES:
BX11 Smart Process Indicator, User Manual, Rev. 1.4, June 2015 Page 40 of 40