MG Instructions For Use Magnesium: Au/Dxc Au
MG Instructions For Use Magnesium: Au/Dxc Au
MG Instructions For Use Magnesium: Au/Dxc Au
AU/DxC AU MG
Instructions For Use Magnesium
© 2022 Beckman Coulter, Inc. All rights reserved.
OSR6189 4 x 40 mL R1
PRINCIPLE
INTENDED PURPOSE
Magnesium reagent is an in vitro diagnostic medical device intended to be used by healthcare professionals for the
quantitative photometric colour measurement of magnesium in human serum, plasma and urine using the automated
Beckman Coulter AU/DxC AU analyzers.
Measurement of magnesium is intended to be used as an aid in the diagnosis and treatment of hypomagnesemia
(abnormally low) and hypermagnesemia (abnormally high).
Reference1,2
Magnesium is an essential factor in many important enzymatic reactions, either as an integral part of a metalloenzyme
or as an activator, and plays an important role in glycolysis, cellular respiration and transmembranous calcium transport.
Magnesium is mainly regulated by the rate of renal magnesium excretion, which along with calcium is subject to the
effects of parathyroid hormone. Increasing calcium reabsorption thus leads to competitive inhibition of magnesium
absorption.
Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low) and
hypermagnesemia (abnormally high). The best-defined manifestation of magnesium deficiency is impairment of
neuromuscular function e.g. hyperirritability, tetany, convulsions, and electrocardiographic changes. Hypomagnesemia
is observed in cases of diabetes, chronic alcoholism, forced diuresis, hyperthyroidism, hypoparathyroidism,
hypocalcemia, malabsorption and acute pancreatitis. Increased serum magnesium levels have been found in cases of
renal failure, dehydration, severe diabetic acidosis and Addison’s Disease.
METHODOLOGY
Reference3,4,5
The Magnesium reagent utilises a direct method in which magnesium ions form a coloured complex with xylidyl blue
in a strongly basic solution. The colour produced is measured bichromatically at 520/800 nm and is proportional to the
magnesium concentration in the sample. Calcium interference is eliminated by glycoletherdiamine-N,N,N`,N`-tetraacetic
acid (GEDTA).
pH 11.4
Mg2+ + Xylidyl blue Purple complex
Serum, Na/Li-heparinized plasma or Urine 6 Do not use EDTA, oxalate or citrate plasma.
Separate serum from red blood cells immediately.
Haemolysed samples should be avoided due to the higher magnesium concentration in erythrocytes.
REAGENTS
WARNING AND PRECAUTIONS
Exercise the normal precautions required for handling all laboratory reagents.
Dispose of all waste material in accordance with local guidelines.
This product contains material of animal origin. The product should be considered as potentially capable of transmitting
infectious diseases.
ACTIVE INGREDIENTS
CAUTION
Sodium azide preservative may form explosive compounds in metal drain lines.
See NIOSH Bulletin: Explosive Azide Hazard (8/16/76).
To avoid the possible build-up of azide compounds, flush wastepipes with
water after the disposal of undiluted reagent. Sodium azide disposal must be in
accordance with appropriate local regulations.
Magnesium DANGER
REAGENT PREPARATION
The reagent is ready for use and can be placed directly on board the instrument.
INDICATIONS OF DETERIORATION
Visible signs of microbial growth, gross turbidity, precipitate, or change in color may indicate degradation and warrant
discontinuance of use.
CALIBRATION
CALIBRATOR REQUIRED
Use System Calibrator Cat. No. 66300 for serum and plasma application and Urine Calibrator Cat. No. B64606 for
urine application.
The reagent blank and calibration curves should be visually reviewed for acceptability. Instructions on how to perform
the review are provided in the relevant AU/DxC AU analyzer IFU and Reference Manual. Quality control procedures
should be undertaken immediately following calibration in accordance with good laboratory practice.
TRACEABILITY
The magnesium values of System Calibrator Cat. No. 66300 are traceable to the National Institute of Standards
and Technology (NIST) Standard Reference Material (SRM) 909b Level 2. The magnesium values of the calibrator
Urine Calibrator Cat. No. B64606 are traceable to the National Institute of Standards and Technology (NIST) Standard
Reference Material (SRM) 929a.
QUALITY CONTROL
Controls Cat. No. ODC0003 and ODC0004 or other control materials with values determined by this Beckman Coulter
system may be used for the serum/plasma application.
Control materials with values determined by this Beckman Coulter system may be used for the urine application.
Each laboratory should establish its own control frequency however good laboratory practice suggests that controls be
tested each day patient samples are tested and each time calibration/blanking is performed.
The results obtained by any individual laboratory may vary from the given mean value. It is therefore recommended that
each laboratory generates analyte specific control target values and intervals based on multiple runs according to their
requirements. These target values should fall within the corresponding acceptable ranges given in the relevant product
literature.
If any trends or sudden shifts in values are detected, review all operating parameters.
Each laboratory should establish guidelines for corrective action to be taken if controls do not recover within the specified
limits.
TESTING PROCEDURE(S)
Refer to the appropriate Beckman Coulter AU/DxC AU analyzer Instructions For Use (IFU) for analyzer-specific assay
instructions for the sample type as listed in the Intended Purpose statement.
CALCULATIONS
The Beckman Coulter analysers automatically compute the magnesium concentration of each sample.
REPORTING RESULTS
REFERENCE INTERVALS
Reference1
Serum Men 0.73 - 1.06 mmol/L 1.8 - 2.6 mg/dL
Serum Women 0.77 - 1.03 mmol/L 1.9 - 2.5 mg/dL
Urine Adults 3 - 5 mmol/24 h 73 - 122 mg/24h
Expected values may vary with age, sex, sample type, diet and geographical location. Each laboratory should verify
the transferability of the expected values to its own population, and if necessary determine its own reference interval
according to good laboratory practice. For diagnostic purposes, results should always be assessed in conjunction with
the patient's medical history, clinical examinations and other findings.
Results of serum studies conducted to evaluate the susceptibility of the method to interference were as follows:
Icterus: Interference less than 10% or 0.12 mmol/L up to 28 mg/dL or 479 μmol/L bilirubin
Calcium: Interference less than 3% or 0.12 mmol/L up to 30 mg/dL or 7.5 mmol/L calcium
Haemolysis: Interference less than 10% or 0.12 mmol/L up to 1.5 g/L haemoglobin
Lipemia: Interference less than 10% or 0.12 mmol/L up to 500 mg/dL Intralipid
Results of urine studies conducted to evaluate the susceptibility of the method to interference were as follows:
PERFORMANCE CHARACTERISTICS
PERFORMANCE CHARACTERISTICS
Data contained within this section is representative of performance on Beckman Coulter systems. Data obtained in your
laboratory may differ from these values.
mmol/L mg/dL
Serum and Plasma 0.2 - 3.3 0.5 - 8.0
Urine (AU480/AU680) 0.2 - 9.25 0.5 - 22.5
Urine (AU5800/DxC 700 AU) 0.2 - 7.8 0.5 - 18.9
SENSITIVITY
Correctly operating AU/DxC AU Systems should exhibit the following sensitivity values:
METHODS COMPARISON
Regression
Equation r n Range
y=1.018x-0.0125 0.988 114 0.215 - 2.971 mmol/L
Patient Urine samples were evaluated in method comparison studies.
Results of Deming regression analysis were as follows:
Regression
Equation r n Range
Y=0.983x+0.031 0.999 103 0.4 - 7.8 mmol/L
PRECISION
Correctly operating AU/DxC AU systems should exhibit the following precision values:
Serum/Plasma Urine
TYPE OF
IMPRECISION % CV SD % CV
Repeatability ≤3 ≤0.06 mmol/L @ ≤3
(Within-run) ≤0.62 mmol/L Mg
Within Laboratory ≤5 N/A ≤5
(Total)
Estimates of precision, based on CLSI recommendations9 are consistent with typical performance.
AU5800
AU5800
Notice to User
Any serious incident that has occurred in relation to this device should be reported to Beckman Coulter and the competent
authority of the Member State in which the user and/or patient is established.
Refer to the Beckman Coulter Chemistry Systems Reagent Guide (BLGUIDE) for specific chemistry information for the
AU/DxC AU clinical chemistry systems and guidance on symbols used on all AU/DxC AU product labelling.
REVISION HISTORY
2. Jacob RA. Trace Elements. In: Tietz NW, ed. Fundamentals of clinical chemistry. Philadelphia:WB Saunders
Company, 1987:521-524.
5. Bohuon C. Microdosage du magnésium dans divers milieux biologiques. Clin Chim Acta 1962;7:811-817.
6. Kazmierczak SC. Magnesium. In: Kaplan LA, Pesce AJ, eds. Clinical chemistry theory, analysis, and correlation.
St Louis: Mosby, 1996:551pp.
7. NCCLS. Urinalysis and collection, transportation, and preservation of urine specimens; approved guideline.
NCCLS Document GP16-A2, 2nd ed. Pennsylvania: NCCLS, 2001.
8. AACC Effects on Clinical Laboratory Tests: Drugs, Disease, Herbs and Natural Products
https://clinfx.wiley.com/aaccweb/aacc/
9. Clinical and Laboratory Standards Institute. Evaluation of Precision of Quantitative Measurement Procedures;
Approved Guideline - Third Edition. CLSI document EP05-A3, Wayne PA 2014.
Beckman Coulter Ireland Inc., Lismeehan, O’Callaghan’s Mills, Co. Clare, Ireland +(353) (0) 65 683 1100