Fabius Gs Service 342
Fabius Gs Service 342
Fabius Gs Service 342
Technical
Service
Manual
Part Number: 4117104
Rev: L
Date: 1 June 2004
© 2004 Draeger Medical, Inc.
Fabius GS®
Anesthesia System
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DESCRIPTION PAGE
SECTION 1: INTRODUCTION
1.0 Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
SECTION 4: DIAGNOSTICS
4.0 DIAGNOSTICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 System Service Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2 Main Service Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.3 Service Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.4 Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
4.5 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
4.6 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.7 Configure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
4.8 Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
i
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
CONTENTS (continued) FABIUS GS
DESCRIPTION PAGE
ii
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS CONTENTS (continued)
DESCRIPTION PAGE
iii
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ANESTHESIA SYSTEM
AUXILIARY DISPLAY
OXYGEN
FLOWMETER
CONTROL
PANEL
VENTILATOR
VAPORIZER
BREATHING
SYSTEM
CYLINDER
PRESSURE
GAUGES
PIPELINE
O2 FLUSH
PRESSURE
GAUGES
SCAVENGER FLOW
CONTROLS
TROLLEY
1-0
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS INTRODUCTION
1.0 Recommendations
Because of the sophisticated nature of Draeger Medical, Inc. anesthesia equipment and its
critical importance in the operating room setting, it is highly recommended that only
appropriately trained and experienced professionals be permitted to service and maintain
this equipment. Please contact DrägerService® at (800) 543-5047 for service of this
equipment in North America. For service in Europe call 49 (451) 882-4222. For service in
other countries call (215) 721-5402.
Draeger Medical, Inc. recommends that the Fabius GS be serviced at six month intervals.
Periodic Manufacturer’s Certification agreements are available for equipment manufactured
by Draeger Medical, Inc. Please contact us for further information concerning these
agreements.
Draeger Medical, Inc. products /material in need of factory repair shall be sent to:
1-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
INTRODUCTION (continued) FABIUS GS
WARNINGS
Warnings are used in this manual before procedures which if not performed correctly could
result in personal injury.
CAUTIONS
Cautions are used in this manual to alert service personnel to the possibility of damage to
the equipment if a procedure is not performed correctly.
Copyright
Copyright© 2004 by Draeger Medical, Inc. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, or stored in a retrieval system in any form or
by any means, electronic or mechanical, including photocopying and recording, without
written permission of Draeger Medical, Inc.
Trademark Notices
Datagrip, DrägerService, ORM, Quality Service For Life, Respitone, Narkomed, Vigilance
Audit, Vitalert, Vitalink, Narkomed GS, Fabius, Fabius Tiro, and Fabius GS are registered
trademarks of Draeger Medical, Inc. All other products or name brands are trademarks of
their respective owners.
Disclaimer
The content of this manual is furnished for informational use only and is subject to change
without notice. Draeger Medical, Inc. assumes no responsibility or liability for any errors or
inaccuracies that may appear in this manual.
1-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION
AUXILIARY DISPLAY
OXYGEN
FLOWMETER
CONTROL PANEL
VENTILATOR
VAPORIZER
BREATHING
SYSTEM CYLINDER
PRESSURE
GAUGES
PIPELINE
PRESSURE
O2 FLUSH GAUGES
FLOW
TROLLEY CONTROLS
2-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
SV00457
SERIAL COMMUNICATION
PORT
O2 SENSOR
CONNECTION
BREATHING PRESSURE
PILOT LINE CONNECTION
AC POWER CONNECTION
ON-OFF SWITCH
BATTERY FUSE
Figure 2-2. Back View Showing Interface Panel and Power Entry
2-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
SV00505
PIPELINE
HOSE
CONNECTIONS
O2 CYLINDER YOKE
N2O OR AIR
CYLINDER YOKE O2 CYLINDER YOKE
Figure 2-3. Back View Showing Gas Pipeline and Cylinder Yokes
2-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
O2 CYLINDER CONNECTION
PIPELINE
HOSE
CONNECTIONS
SV00551
Figure 2-4. Back View Showing Gas Pipeline and Cylinder Hose Connections
(machines without cylinder yokes)
2-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
AUXILIARY O2
AIR
VALVE LUNG
SAFETY
VALVE
VENTILATOR P VENT
BREATHING
BAG (AIR
ENTRAINMENT)
ABSORBER
PEEP
EXPIRATORY CONTROL
VACUUM
VALVE VALVE
*V
APL VALVE
APL
BYPASS
VALVE
PISTON
PUMP
SV88038
SCAVENGER
OUTLET
Figure 2-5. Function diagram of Fabius GS - Typical for Breathing System P/N 4116398 or
4117529
2-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
SV77028
Figure 2-6. Function diagram of Fabius GS - Typical for Breathing System P/N 4118378 or
4118379
2-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
The batteries are located within the controller housing and are accessible by
opening the ventilator compartment. The 3.15A battery fuse is located on the
power entry panel at the back of the machine.
POWER
FUSE
3.15A J6
28 V POWER SUPPLY
2-7
2.4
SV00454
YOKE OPTION
O2 YOKE
AIR GAUGE
FUNCTION DESCRIPTION (continued)
N2O GAUGE
2-8
SORC BY-PASS ASM
TOTAL
FLOWMETER
VAPOR
MOUNT
FLOW
SENSORS
FLOW CONTROL
VALVE MANIFOLD
O2 FLUSH
FABIUS GS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
Should the O2 supply fail or its pressure decrease below a certain limit, the O2
supply pressure alarm switch signals an alarm.
O2
FLUSH O2 SUPPLY
TO
PRESSURE
FRESH GAS
ALARM SWITCH
OUTLET
P8 P7
O2
P6 TO
AIR FLOW
METERS
P4 P5
N2O
PRESSURE
REGULATORS
O2 N2O
PRESSURE
GAUGES
P1 P2 P3 GAS INLET
ASSEMBLY
P11
O2
P10
AIR
P9
N2O
PIPELINE NON-RETURN
SUPPLIES FILTERS
VALVES SV00489
2-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
If the O2 flush button is pressed, oxygen is delivered to the fresh gas outlet. The
fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic
vaporizer. This avoids an increase in the anesthetic gas concentration.
O2 FLUSH
FRESH GAS
FLOWMETER TOTAL
ASSEMBLY FLOWMETER
(O2)
ANESTHETIC
FROM THE VAPORIZER
GAS BOX (O2)
FROM THE
SV00490
GAS BOX (N2O)
2-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
If the O2 flow control valve is closed or if the O2 flow is lower than or equal to
200 mL/min, the SORC interrupts the N2O flow.
N2O can be added when the O2 flow is approx. 300 mL/min. In this case, the
SORC also prevents O2 concentrations below 23 vol.%.
The SORC bypass allows O2 to bypass the restrictor in the SORC when O2 flows
above 10 L/min. are needed.
FRESH GAS
FROM
O2 FLUSH
FLOWMETER TOTAL
FLOW
ASSEMBLY FLOWMETER
SENSOR
FLOW
CONTROL
VALVE
ANESTHETIC
FROM THE VAPORIZER
GAS BOX (O2)
S-ORC
FROM THE
SV88062
GAS BOX (N2O)
2-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
The flow control valves are used to adjust the O2 and N2O flows.
The restrictors and the spring tension are dimensioned such that a minimum
concentration of 23 vol.% O2 is always ensured. The maximum O2 flow is
approx. 12 L/min.
O2 N2O
restrictors
control
diaphragms
N2O check valve
operating-point
adjusting screw
O2 N2O
flow control
valves
2-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
In the "SPONT" position the APL valve is open to atmosphere. This position is
used for spontaneous breathing.
The pressure limit (Pmax) can also be adjusted (through front panel interface)
during volume control from 15 cmH2O to 70 cmH2O using the control box and
the PEEP Pmax valve.
2-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
10 2 1
3
9
17
7
11
6
4
5
8
12
16
15
13
14
SV00493
Figure 2-13. Compact breathing system, CosyII - Typical for P/N 4116398 or 4117529
Key
2-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
17
10 2 1
3
11
7
4
5
8 16 12
13
14
SV77029
15
Figure 2-14. Compact breathing system, CosyII - Typical for P/N 4118378 or 4118379
Key
2-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
During manual ventilation, the APL valve is set to the "MAN" position.
The patient system safety valve is activated. The piston of the ventilator
is in the upper end position in order to reduce the volume of the
ventilator.
SV88035
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-15. Manual ventilation (inspiration) - Typical for Breathing System P/N 4116398
or 4117529
2-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
SV77030
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-16. Manual ventilation (inspiration) - Typical for Breathing System P/N 4118378
or 4118379
2-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
During expiration, the inspiratory valve remains closed and thus prevents the
expiratory gas from flowing back into the inspiratory branch.
After releasing the breathing bag, the expiratory gas from the lung flows
through the expiratory hose, the flow sensor, the PEEP Pmax valve, the
expiratory valve, and through the absorber into the breathing bag. At the same
time, new fresh gas flows into the breathing bag.
SV88036
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-17. Manual ventilation (expiration) - Typical for Breathing System P/N 4116398 or
4117529
2-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
During expiration, the inspiratory valve remains closed and thus prevents the
expiratory gas from flowing back into the inspiratory branch.
After releasing the breathing bag, the expiratory gas from the lung flows
through the expiratory hose, the flow sensor, the PEEP Pmax valve, the
expiratory valve, onto the breathing bag and through the absorber. At the same
time, new fresh gas flows into the breathing bag.
SV77031
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-18. Manual ventilation (expiration) - Typical for Breathing System P/N 4118378 or
4118379
2-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
The patient inhales the gas mixture (expiratory gas and fresh gas) from
the breathing bag. The gas mixture flows through the fresh-gas
decoupling valve, the inspiratory valve, the O2 sensor, the inspiratory
hose, and through the Y-piece into the lung. The pressure sensor
measures the airway pressure.
SV88042
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-19. Spontaneous breathing (inspiration) - Typical for Breathing System P/N
4116398 or 4117529
2-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
SV77032
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-20. Spontaneous breathing (inspiration) - Typical for Breathing System P/N
4118378 or 4118379
2-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
During expiration, the inspiratory valve remains closed thus preventing the
expiratory gas from flowing back into the inspiratory branch.
The expiratory gas flows from the lung through the expiratory hose, the flow
sensor, the PEEP control valve, the expiratory valve, and through the absorber
into the breathing bag. At the same time, new fresh gas flows into the breathing
bag.
When the breathing bag is full, any excess gas mixture flows through the non-
return valve into the anesthetic gas scavenging system.
SV88043
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-21. Spontaneous breathing (expiration) - Typical for Breathing System P/N
4116398 or 4117529
The CO2 is scrubbed from the expiratory gas by the soda lime contained in the
absorber. The fresh gas replaces the anesthetic and oxygen taken up by the
patient.
2-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
During expiration, the inspiratory valve remains closed thus preventing the
expiratory gas from flowing back into the inspiratory branch.
The expiratory gas flows from the lung through the expiratory hose, the flow
sensor, the PEEP control valve, the expiratory valve, the breathing bag, and
through the absorber. At the same time, new fresh gas flows into the breathing
bag.
When the breathing bag is full, any excess gas mixture flows through the non-
return valve into the anesthetic gas scavenging system.
SV77033
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
INACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE CLOSED
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-22. Spontaneous breathing (expiration) - Typical for Breathing System P/N
4118378 or 4118379
2-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
The APL bypass valve opens in volume mode, allowing excess gas to be
vented to the scavenging system regardless of the MAN-SPONT valve
setting.
The safety valve of the patient system makes sure that no pressures
greater than 75 cmH2O build up in the system.
During ventilation, the pressure limit (Pmax) can be adjusted at the user
interface.
2-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
During inspiration, the PEEP/Pmax valve remains closed. The control pressure
present at the PEEP/Pmax valve varies with the set pressure limit (Pmax).
Any excess fresh gas flows through the open APL bypass valve, and through the
non-return valve into the anesthetic gas scavenging system.
SV88044
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE VARIABLE
VENTILATOR FRESH GAS
ACTIVE DECOUPLER
CLOSED
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE OPEN
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-23. Volume control ventilation (inspiration) - Typical for Breathing System P/N
4116398 or 4117529
2-25
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
SV77034
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE VARIABLE
VENTILATOR FRESH GAS
ACTIVE DECOUPLER
CLOSED
EXPIRATORY
FRESH GAS INLET VALVE CLOSED
APL
APL BYPASS VALVE
VALVE OPEN
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-24. Volume control ventilation (inspiration) - Typical for Breathing System P/N
4118378 or 4118379
2-26
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
The expiratory gas from the lung flows through the expiratory hose, the flow
sensor, the PEEP/Pmax valve, the expiratory valve, and through the absorber
back into the breathing bag mixing with fresh gas also flowing into the
breathing bag.
The ventilator’s piston moves back drawing the gas mixture needed for the next
inspiration into the piston space.
Any excess fresh-gas flows through the APL bypass valve, and through the non-
return valve into the anesthetic gas scavenging system.
SV88045
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
ACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE OPEN
ABSORBER
BREATHING BAG
EXHAUST VALVE
OPENS AS
REQUIRED
Figure 2-25. Volume control ventilation (expiration) - Typical for Breathing System P/N
4116398 or 4117529
2-27
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
SV77035
OXYGEN SENSOR
PRESSURE SENSOR
PEEP/PMAX
VALVE OPEN
VENTILATOR FRESH GAS
ACTIVE DECOUPLER
OPEN
EXPIRATORY
FRESH GAS INLET VALVE OPEN
APL
APL BYPASS VALVE
VALVE OPEN
ABSORBER
EXHAUST VALVE
OPENS AS
REQUIRED
BREATHING BAG
Figure 2-26. Volume control ventilation (expiration) - Typical for Breathing System P/N
4118378 or 4118379
2-28
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
Expired soda lime changes its color. The soda lime must be replaced
when two thirds of the soda lime in the absorber canister is discolored.
2-29
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
2.8 Ventilator
The ventilator is located in a swing-out compartment at the left side of the
Fabius GS. A hose terminal is provided on the left side of the compartment for
connection to the breathing system. Fresh gas is delivered to the patient by a
piston that is driven by a motor and ball-screw arrangement. A sight window on
the compartment allows the operator to verify movement of the piston.
Two diaphragms (upper and lower) comprise a bag-type rolling seal that
surrounds the piston. Vacuum from the pneumatic assembly (described in a
later paragraph) is provided between the outside of the seal and the cylinder, to
ensure proper operation of the seal during piston movement.
During inspiration the ventilator delivers fresh gas at a given volume, pressure
and frequency. These parameters are set at the control panel. Refer to the
Operator’s Manual for details on ventilator settings, displays and controls.
During expiration, the bag-type rolling seal fills with expired gas from the
patient and with fresh gas stored in the breathing bag.
Power for the ventilator motor is distributed from the control PCB. A position
sensor on the ventilator signals the control PCB when the piston reaches its
lower limit. An incremental encoder on the motor shaft determines the number
of revolutions and provides piston travel information to the control PCB.
The ventilator pressure transducer is the same type as the one used for
measuring airway pressure. A hose connects the transducer’s positive pressure
port to a hose barb located on the top cover of the ventilator. The purpose of this
transducer is to allow the software to sense when a condition exists that would
cause the ventilator negative pressure relief valve to open. The threshold used
by the software for this condition is indicated in the table below. In normal use
the primary cause for this condition is an insufficient amount of reserve gas in
the breathing bag. The operator is alerted when this condition exists, with a
medium priority “FRESH GAS LOW” alarm. This alarm may be disabled via
Service Mode.
2-30
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
NEGATIVE PRESSURE
RELIEF VALVE
VENT CHAMBER
HIGH PRESSURE PRESSURE
VALVE SENSOR LINE
VACUUM LINE
TO PNEUMATIC
ASSEMBLY
UPPER
DIAPHRAGM
CYLINDER
PISTON
LOWER
DIAPHRAGM
MOTOR AND
BALLSCREW
ASSEMBLY
2-31
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
The top of the ventilator assembly (patient system) contains two valves:
screw
spring
valve disc
washer
threaded ring
valve seat
Software Version Threshold
2-32
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
TO UPPER PORT OF
LS1 ON PCB ASM
TO LOWER PORT OF
LS3 ON PCB ASM
PNEUMATIC ASSEMBLY
VENTILATOR CHAMBER
VARIABLE PRESSURE
FILTER
RESTRICTOR
VENTILATOR
V2
ELECTRONIC
V1 PEEP
VALVE
PUMP
RESERVOIR
APL BYPASS
VALVE
PEEP VALVE
INTERFACE BREATHING
CONTROL SIGNAL PANEL SYSTEM
FROM J7 ON PCB ASM
2-33
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
When the machine is operating in the Manual mode, the pump on the
pneumatic assembly (and the ventilator) is stopped, and the spring-
loaded APL bypass valve in the breathing system closes, directing
exhaled gas through the APL valve.
2-34
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
FRESH LOW O2
J3
SV88046 N2O AIR O2 GAS PRESS
DISPLAY SW
P2 P3 P4 J1
PNEUMATIC ASM J1
J16
J15
SERIAL PORT
J7
PEEP INTERFACE
VALVE
J17
O2 SENSOR
J8
PUMP J18
SPIROLOG
VOL SENSOR
MAIN CONTROLLER
PCB ASSEMBLY
J10
FRONT J2 MOTOR
PANEL
DISPLAY
J11 TACH
SENSOR
KEYPAD J1 J9
INTERFACE
J12
HOME
J2 J3 SENSOR
J4 J5 J6 J21
TABLE VENTILATOR
LAMP
ON/OFF BOOST
SW CAPS
ROTARY
ENCODER
BATT F4
J3
J1 POWER
MAINS F1
F2 SUPPLY
PWR ENTRY
2-35
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
2.11.1 Flowmeter
Receives flow rate data from the N2O, Air, and O2 flow sensors and
provides fresh gas display information for these gases.
Receives flow information from the Spirolog sensor and processes it for
display.
2.11.3 Ventilator
Provides ventilator motor drive, receives ventilator piston position and
movement information. Provides ventilator pressure information.
Receives information from the keypad and rotary encoder for making
display and operating selections, setting alarm limits, and service
functions.
2.11.7 Battery
Provides charging current for battery and monitors state of battery
charge.
2-36
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
SV88047
FLASH
LATCH BUFFER XCVR
MEM
FRESH GAS
FLOW SENSORS
VIDEO DISPLAY
RAM
CONTROL PANEL
O2 SENSOR
MICROCONTROLLER
SPKR
VOL
SENSOR
DRV LEDS
VENT
PRESS LS3 VENT
MOTOR VENTILATOR MOTOR
DRIVE
PWR IN
PWR
VENTILATOR HOMES SENS
BATT MGMT
+28V
OPTO DRIVER SERIAL
ISO RCVR PORT
2-37
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
Data and power for the display comes from the main controller PCB via a 20-
conductor ribbon cable. The keypad interface is connected to the main controller
PCB by a 30-conductor ribbon cable. A block diagram of the control panel
assembly is shown in the following illustration.
J1
SPEAKER
J2 J3
ROTARY
ENCODER
2-38
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
Descriptions of the numbered items in the following illustration are given in the
table below.
12
Service 10:50 AM
1 8
Service Log Preventive
Maintenance
General
13
2 Entries
Real Time Values
Activate Serial Number
Reset Hours Run 9
Reset Last Service Date
Select Market Kit
11 14
4
5 6
15
7
SV
Item Function
7 Soft keys: activate the corresponding function that appears on screen above the key
9 Setup key: activates sub-screens for monitoring functions. Refer to Operator’s Manual
11 Rotary control: moves the cursor on the screen; confirms selection when pressed
13 Alarm Silence key: silences all active alarms for two minutes
2-39
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
The O2 sensor contains an alkaline electrolyte, a lead anode, two gold cathodes,
and a Teflon membrane. The spatial separation of the two gold cathodes allows
a voltage comparison to be made as explained below.
The O2 sensor is an electrochemical cell that generates a voltage from the ion
current.
TEMPERATURE
COMPENSATING
RESISTORS
LEAD ANODE
ALKALAI
ELECTROLYTE
The internal resistance of the cell is determined by the surface area of the gold
cathodes, the O2 diffusion velocity, the distance between the gold cathodes and
the lead anode. This resistance is approximately 700 ohms.
If the O2 sensor fails, the control box will indicate an error on the graphics
display.
2-40
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
A second platinum wire (B) in the measuring tube is used to compensate for
interferences from different gases present in the respiratory gas. The heat
removed from the second platinum wire is measured during inspiration when
the gas flow is zero.
The different gases present in the respiratory gas have a different thermal
conductivity. The amount of heat removed from the second platinum wire is
thus an indicator of respiratory gas composition.
Internal calibration tables for O2/N2O mixtures, Air and 100% O2 are used to
linearize the measured flow.
SV88037
2-41
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FUNCTION DESCRIPTION (continued) FABIUS GS
ELECTRICAL
TUBING CONNECTION
CONNECTOR ELECTRONIC
COMPONENTS
GAS OUTLET
PORT
GAS INLET
(TO MANIFOLD)
ASSEMBLY
HEATED
CHAMBER
MOUNTING POSTS
FROM OXYGEN
FLOW CONTROL
VALVE
2-42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS FUNCTION DESCRIPTION (continued)
2.16 Vaporizer
Refer to separate technical documentation of the anesthetic vaporizer.
2-43
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE
3.0 TROUBLESHOOTING
This section contains information to assist the Draeger qualified Service
Representative in locating electrical faults affecting operation of the Fabius GS. A
simplified electrical block diagram is given in Figure 3-1. Since most troubleshooting
efforts begin with verifying power supply voltages, the following paragraph outlines
the voltage distribution scheme within the machine along with test points for each of
the voltages.
A piping diagram (Figure 3-4) and a pneumatic control diagram (Figure 3-5) are also
included in this section.
3.1 Power Supply and Voltage Distribution
In the Fabius GS, the power supply delivers 28 VDC to the main controller unit
PCB assembly. Some of the other voltages derived on this assembly can be
measured at the connectors listed in Table 3-1. Fuse data is listed in Table 3-2.
Controller PCB connector locations are shown in Figure 3-2.
J6-1, + 28 VDC
Power In 26.6 VDC to 29.4 VDC
J6-4, Common
J1-1, + 5 VDC Fresh Gas
Display 4.75 VDC to 5.25 VDC
J1-14, Common
J2-1, +12 VDC Front Panel
Display 11.5 VDC to 12.5 VDC
J2-20, Common
J16-1, +10 VDC Fresh Gas
Flow Sensors 9.99 VDC to 10.01 VDC
J16-9, Common
3-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
FRESH LOW O2
J3
SV88046 N2O AIR O2 GAS PRESS
DISPLAY SW
P2 P3 P4 J1
PNEUMATIC ASM J1
J16
J15
SERIAL PORT
J7
PEEP INTERFACE
VALVE
J17
O2 SENSOR
J8
PUMP J18
SPIROLOG
VOL SENSOR
MAIN CONTROLLER
PCB ASSEMBLY
J10
FRONT J2 MOTOR
PANEL
DISPLAY
J11 TACH
SENSOR
KEYPAD J1 J9
INTERFACE
J12
HOME
J2 J3 SENSOR
J4 J5 J6 J21
TABLE VENTILATOR
LAMP
ON/OFF BOOST
SW CAPS
ROTARY
ENCODER
BATT F4
J3
J1 POWER
MAINS F1
F2 SUPPLY
PWR ENTRY
3-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
J9 KEYPAD INTERFACE
J15 SERIAL
PORT
BT1 INTERFACE
JP3
VENT
CHAMBER
PRESSURE
SENSOR
BREATHING
LS1 LS3 PRESSURE
SV00503 INTERFACE
PNEUMATIC
ASM LS2
F3 F1 F2
J6 POWER SUPPLY
3-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
J9
KEYPAD
J2 INTERFACE
JP3 J15
FRONT PANEL
JP2 SERIAL PORT
DISPLAY
INTERFACE
BT1
J18
SPIROLOG
VENT
LS3 CHAMBER SENSOR
PRESSURE
SENSOR
LS1 LS2 J17
SV00612
O2 SENSOR
PNEUMATIC
BREATHING
ASM PRESSURE
INTERFACE
J12
F1 HOME
SENSOR
J16 J10
J8 J21
FLOW MOTOR
PUMP J7 BOOST
SENSORS
PEEP CAPS
J6
J1 VALVE
POWER
FRESH GAS
SUPPLY
DISPLAY
J4
J11 ON/OFF
J5 TACH SWITCH
BATTERY SENSOR
3-4
SV00454
FABIUS GS
YOKE OPTION
O2 YOKE
AIR GAUGE
N2O GAUGE
3-5
SORC BY-PASS ASM
TOTAL
FLOWMETER
VAPOR
MOUNT
FLOW
SENSORS
FLOW CONTROL
VALVE MANIFOLD
O2 FLUSH
TROUBLESHOOTING GUIDE (continued)
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
TO UPPER PORT OF
LS1 ON PCB ASM
SV88048
VENT CHAMBER
PNEUMATIC ASSEMBLY PRESSURE
VENTILATOR
ELECTRONIC
PEEP
VALVE
PUMP
RESERVOIR
APL BYPASS
VALVE
PEEP VALVE
INTERFACE BREATHING
CONTROL SIGNAL PANEL SYSTEM
FROM J7 ON PCB ASM
3-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
3.2 Battery
The Fabius GS backup battery comprises two series-connected 12 V
rechargeable batteries. While the machine is operating from AC, the total
battery voltage at full charge should be within the range of 27.0 to 29.6 V.
Battery voltage can be measured at J5 on the main controller PCB assembly.
During battery operation, the low battery cutoff voltage should be within the
range of 21.0 to 20.0 V.
3-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
START
SV77001
IS O2 Y
MONITOR WORKING
PROPERLY
CONNECT N
SENSOR
CORD TO
INTERFACE AND
RECALIBRATE
IS SENSOR
CORD CONNECTED TO
N INTERFACE PANEL?
INSTALL SENSOR
ELEMENT CORRECTLY OR Y
REPLACE IF EXPIRED
AND RECALIBRATE O2
SENSOR
IS SENSOR
N ELEMENT INSTALLED
CORRECTLY AND
NOT EXPIRED?
REPLACE
SENSOR HOUSING
AND
CHECK INTERNAL RECALIBRATE
CABLE FROM
N IS O2
INTERFACE PANEL
MONITOR WORKING
TO J17 ON
PROPERLY?
CONTROLLER PCB
Y
IS O2 Y
MONITOR WORKING
PROPERLY?
REPLACE
CONTROLLER PERFORM A DOES Y UNIT IS
PCB ASM COMPLETE PMS UNIT PASS FUNCTIONAL
AS OUTLINED IN ON THE UNIT PMS?
SECTION 5
IS O2 Y
MONITOR WORKING
PROPERLY?
N CONTACT LOCAL
REINSTALL ORIGINAL AUTHORIZED SERVICE
CONTROLLER PCB ORGANIZATION OR DMI
ASM DRAGERSERVICE
3-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
START
SV77002
IS
BREATHING
Y
PRESSURE
MONITOR WORKING
PROPERLY?
N
CHECK PILOT LINE
CONNECTIONS AT
BOTH ENDS AND IF
APPLICABLE,
BREATHING
GAUGE ASM AND
ASSOCIATED TUBING
IS PILOT LINE
N AND OPTIONAL
BREATHING
GAUGE ASM
CONNECTED
PROPERLY?
CHECK CALIBRATION/
PERFORM CALIBRATION
UNIT IS
IS
FUNCTIONAL
BREATHING
PRESSURE Y
MONITOR WORKING
PROPERLY?
Y
CHECK INTERNAL IS
TUBING FROM BREATHING DOES
Y PERFORM A
EXTERIOR OF PRESSURE UNIT PASS
COMPLETE PMS
INTERFACE PANEL TO MONITOR WORKING PMS?
ON THE UNIT
SENSOR ON PROPERLY?
CONTROLLER PCB ASM
N
N
IS
BREATHING CONTACT LOCAL
Y REINSTALL ORIGINAL AUTHORIZED SERVICE
PRESSURE
MONITOR WORKING CONTROLLER PCB ASM ORGANIZATION OR DMI
PROPERLY? DRAGERSERVICE
REPLACE CONTROLLER
PCB ASM AS OUTLINED IN
SECTION 5
3-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
START
IS
RESPIRATORY SV77003
VOLUME MONITOR Y
WORKING
PROPERLY?
CONNECT
SENSOR N
CORD TO
FLOW SENSOR
AND CALIBRATE
N IS SENSOR
CORD CONNECTED TO
FLOW SENSOR?
Y
CHECK COMPLETE
SPIROLOG VOLUME
SENSOR INSTALLATION
AND CONNECTIONS
IS SENSOR
N INSTALLED PROPERLY
AND ARE PNEUMATIC
CONNECTIONS
SECURE?
REPLACE SPIROLOG
VOLUME SENSOR DOES Y
AND RECALIBRATE UNIT IS
UNIT PASS FUNCTIONAL
PMS?
N
IS CONTACT LOCAL
RESPIRATORY AUTHORIZED SERVICE
Y PERFORM A Y ORGANIZATION OR DMI
VOLUME MONITOR
COMPLETE PMS DRAGERSERVICE
WORKING
ON THE UNIT
PROPERLY?
IS N
RESPIRATORY
N VOLUME MONITOR REINSTALL ORIGINAL
WORKING CONTROLLER PCB
PROPERLY? ASM
Y
REINSTALL ORIGINAL IS
N
VOLUME SENSOR RESPIRATORY REPLACE
----------------------------- VOLUME MONITOR CONTROLLER PCB
CHECK INTERNAL CABLE WORKING ASM
FROM INTERFACE PANEL PROPERLY? AS OUTLINED
TO J18 ON IN SECTION 5
CONTROLLER ASM
3-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
START
SV77004
ARE Y
AUDIO ALARMS
WORKING?
ARE Y
AUDIO ALARMS
WORKING?
SUBSTITUTE
ANOTHER FRONT
PANEL (MONITOR
BEZEL) AS
OUTLINED IN
SECTION 5
ARE Y REPLACE
AUDIO ALARMS MONITOR BEZEL
WORKING? ASM AS OUTLINED
IN SECTION 5
ARE Y
AUDIO ALARMS
WORKING?
N CONTACT LOCAL
REINSTALL ORIGINAL AUTHORIZED SERVICE
CONTROLLER PCB ORGANIZATION OR DMI
DRAGERSERVICE
3-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
START
SV77005
DOES
HOST UNIT Y
COMMUNICATE WITH
PERIPHERAL
DEVICE?
N
CONNECT
CABLE
AT INTERFACE
IS CONTACT LOCAL
N PERIPHERAL CABLE AUTHORIZED SERVICE
CONNECTED TO COMM ORGANIZATION OR DMI
PORT? DRAGERSERVICE
CHECK PERIPHERAL
DEVICE AND HOST REINSTALL DOES UNIT Y
UNIT IS
SERIAL PORT ORIGINAL PASS PMS? FUNCTIONAL
CONFIG SETTINGS CONTROLLER ASM
DOES DOES
Y HOST UNIT HOST UNIT PERFORM A
Y
COMMUNICATE COMMUNICATE COMPLETE PMS
WITH PERIPHERAL WITH PERIPHERAL ON UNIT
DEVICE? DEVICE?
DOES
HOST UNIT N
COMMUNICATE
WITH PERIPHERAL
DEVICE?
3-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
START
SV77006
IS
LOW O2 SUPPLY Y
PRESSURE ALARM
FUNCTIONAL?
VERIFY NO O2
PRESSURE IS IN THE
SYSTEM
IS
LOW O2 SUPPLY Y PERFORM A DOES
PRESSURE ALARM COMPLETE PMS UNIT PASS
FUNCTIONAL? ON THE UNIT PMS?
N
N
REPLACE FRESHGAS
DISPLAY PCB AS
OUTLINES IN SECTION 5
IS IS
LOW O2 SUPPLY Y LOW O2 SUPPLY
PRESSURE ALARM PRESSURE ALARM
FUNCTIONAL? FUNCTIONAL?
N N
CONTACT LOCAL
REPLACE CONTROLLER AUTHORIZED SERVICE
REINSTALL ORIGINAL
PCB AS OUTLINED IN ORGANIZATION OR DMI
CONTROLLER PCB
SECTION 5 DRAGERSERVICE
3-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
START
IS DISPLAY Y
FUNCTIONAL UPON
SYSTEM
POWER-UP?
SV77007
WAS A
SOFTWARE DOWNLOAD Y PERFORM BOOT STRAP
PROCEDURE PERFORMED DOWNLOAD PROCEDURE
IMMEDIATELY PRIOR TO AS OUTLINED
DISPLAY FAILURE? IN SECTION 8
IS DISPLAY
N FUNCTIONAL UPON Y
SYSTEM
POWER-UP?
CHECK POWER
SUPPLY VOLTAGE,
REPLACE POWER
SUPPLY IF OUT
OF TOLERANCE
IS DISPLAY Y
FUNCTIONAL UPON
SYSTEM
POWER-UP?
IS DISPLAY
Y
FUNCTIONAL UPON
SYSTEM
POWER-UP?
N
UNIT IS
POWER OFF, FUNCTIONAL
SUBSTITUTE ANOTHER
DISPLAY PANEL
N
N
REINSTALL ORIGINAL
DISPLAY PANEL IS DISPLAY CONTACT LOCAL
N REINSTALL AUTHORIZED SERVICE
____________________ FUNCTIONAL UPON
ORIGINAL ORGANIZATION OR DMI
REPLACE CONTROLLER SYSTEM
CONTROLLER PCB DRAGERSERVICE
PCB AS OUTLINED IN POWER-UP?
SECTION 5
3-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
START
SV77008
IS Y
KEYPAD WORKING
PROPERLY?
IS Y
KEYPAD WORKING
PROPERLY?
N
UNIT IS
FUNCTIONAL
SUBSTITUTE ANOTHER
FRONT BEZEL
(MONITOR BEZEL)
N
N
Y
RECONNECT ORIGINAL
MONITOR BEZEL IS N CONTACT LOCAL
REINSTALL ORIGINAL AUTORIZED SERVICE
-------------------------------- KEYPAD WORKING
CONTROLLER PCB ORGANIZATION OR DMI
REPLACE CONTROLLER PROPERLY?
PCB AS OUTLINED IN DRAGERSERVICE
SECTION 5
3-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS TROUBLESHOOTING GUIDE (continued)
START
SV77009
Y
IS VENTILATOR
WORKING
PROPERLY?
Y
IS VENTILATOR
WORKING
PROPERLY? IS VENTILATOR Y
WORKING
PROPERLY?
N
Y
IS VENTILATOR
WORKING Y
PROPERLY? IS VENTILATOR PERFORM A
WORKING COMPLETE PMS
PROPERLY? ON THE UNIT
N
N
CHECK: PNEUMATIC
CONNECTIONS, CORRECT
VACUUM FROM PNEUMATIC ASM, REINSTALL ORIGINAL DOES Y UNIT IS
CABLES TO J10, J11, J12 ON CONTROLLER PCB UNIT PASS FUNCTIONAL
CONTROLLER PCB, ERROR LOG PMS?
FOR SIGNS OF CAUSE
Y
IS VENTILATOR
WORKING
PROPERLY? CONTACT LOCAL
AUTHORIZED SERVICE
ORGANIZATION OR DMI
DRAGERSERVICE
N
CHECK VENTILATOR
DIAPHRAGMS, MOTOR; REPLACE
IF NEEDED AS OUTLINED IN
SECTION 5
3-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued) FABIUS GS
Y
ARE ALL
FRESH GAS DISPLAYS
WORKING
SV77010 PROPERLY?
N IS CONFIGURATION
SET PROPERLY?
(I.E. 2-GAS/3-GAS)
Y
RECONFIGURE
GAS SETTINGS
ARE ALL Y
FRESH GAS DISPLAYS
WORKING
PROPERLY?
RECALIBRATE
FRESH GAS
FLOW SENSORS
CHECK CABLE
CONNECTIONS AT ARE ALL Y
FLOW SENSORS: FRESH GAS DISPLAYS
CHECK CABLE FROM WORKING
FLOW SENSORS TO J16 PROPERLY?
ARE ALL N ON CONTROLLER PCB UNIT IS
FRESH GAS DISPLAYS FUNCTIONAL
WORKING N
PROPERLY?
Y REPLACE FRESH
GAS DISPLAY PCB
Y ARE ALL AS OUTLINED IN
FRESH GAS DISPLAYS SECTION 5 Y
WORKING
PROPERLY?
PERFORM A
DOES UNIT
N COMPLETE PMS ON
PASS PMS?
THE UNIT
N
ARE ALL Y
FRESH GAS DISPLAYS
REPLACE AFFECTED
WORKING
FLOW SENSOR AND
PROPERLY?
RECALIBRATE
CONTACT LOCAL
N AUTHORIZED SERVICE
ORGANIZATION OR DMI
DRAGERSERVICE
ARE ALL N
FRESH GAS DISPLAYS
WORKING
PROPERLY?
3-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.0 DIAGNOSTICS
NOTE: The screen illustrations contained in this section are for reference only
and therefore may or may not reflect the software version currently
installed.
The Fabius Tiro diagnostic system monitors and records the status of its internal
hardware when the machine is turned on. The status of each test is displayed on the
Power-up screen as shown in Figure 4-1. This screen is displayed for several seconds
before proceeding to the Standby screen. The power-up screen also displays one of
three messages at completion of the diagnostics:
FUNCTIONAL: This message indicates that the Fabius GS has passed all
power-up tests and is fully functional. The machine will
proceed to the Standby screen (Figure 4-2.) after a short
delay.
CONDITIONALLY This message indicates that a minor problem has been
FUNCTIONAL: detected. The Fabius GS may be used, but your local
authorized service organization or DrägerService should be
notified to correct the problem. Press the rotary control to
proceed to the Standby screen.
NON- This message indicates that a serious problem has been
FUNCTIONAL: detected, and the machine will not proceed to the Monitor
screen. Do not use the machine. Immediately notify your
local Authorized Service Organization or DrägerService to
correct the problem.
4-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
The following quick reference tree is for use in quickly locating the Service Screens
necessary to perform specific functions within the software. The indentions shown
below indicates the relationship of the Service Screen/function relative to the location
of each Service Screen within the software.
4-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
The Preventive Maintenance Due message will appear on the screen if the current
date exceeds the Periodic Manufacturer's Service due date stored in the machine.
4-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Further diagnostic functions are available through service screens that can be called
up at the display panel. The following paragraphs provide a description of each
service screen that can be accessed at the display. If no display is present upon
system power-up, refer to Section 3.0 of this manual for troubleshooting assistance.
NOTE: During display of the Standby screen, a five minute count-down appears
on the screen, after which the display changes to a Sleep Mode. Press any
key on the panel to return to the Standby screen.
To access the System Service screen, press and hold the and keys, and
press the rotary knob.
4-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Turn the rotary knob to move the scroll box to the desired category, then press
the knob to confirm. The Service Log category is illustrated.
4-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Turn the rotary knob to scroll down to select the desired function, then press the
knob to confirm. The service log screens are described in the following pages.
Service 14:05
Service Log
Entries
Real Time Values
4-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Use the rotary knob to scroll through the list of service log entries. The
service log holds up to 256 entries. When new entries are made, the
oldest entries are replaced with the new entries. The most recent entries
appear upon screen call-up.
CAUTION: Selecting Clear Service Log and pressing the knob will bring
up another screen that allows you to delete all entries.
Selecting YES will delete all entries.
Service 11:30
Service Log
Tables of service log entries, error codes and their descriptions begin on
the next page.
4-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
M004 BATTERY <10% Battery level below 10% of capacity while operating
without mains power
M005 BATTERY <20% Battery level below 20% of capacity while operating
without mains power
4-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
F002 Diag System Ram System RAM fails on pwr up, bad RAM chip, bad CPU,
bad trace, etc.
F003 Diag Program Memory Bad flash download, bad flash chip
F004 Diag Video Test Video memory failed, bad video chip
F008 Diag Serial Port Failed pwr up loopback test, bad UART chip
4-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
V001 VENT PEEP NOT CAL Auto ventilation was attempted without valid PEEP
calibration
V002 ZERO POS INVALID Crossing of light barrier at bottom vent not detected
V037 EX VALVE LEAK Greater than 15 mL of Tidal Volume during inspiration,
and this Vt is greater than 10% of Vt flow during
expiration
V042 Motor Voltage Error Motor voltage has dropped below minimum required
for auto ventilation while in Volume or PCV Ventilation
mode
V043 MOTOR LIM STOP Incident of excessive motor current during auto
ventilation
V044 MEMBRANE PRES LOW Membrane vacuum below normal range: possible leak
V045 MEMBRANE PRES HIGH Membrane vacuum above normal operating range
V049 MOTOR OVER CURRENT Persistent excessive motor current during auto
ventilation: caused ventilator motor shut off
V050 LT Pres Failed>350 Leak in path from vent to hoses to peep valve to bag
V051 LT Pres Rose>250 Leak in path from vent to hoses to peep valve
V052 LT SYS FAIL Leak in path from vent to hoses to peep valve to bag
V053 LT PC FAIL Leak in path from vent to hoses to peep valve
V060 Comp Unexp State Compliance test entered an unexpected test state
V061 Comp Bag Pres Too High Compliance test start pressure too high
V062 Comp Cant Fully Pres Compliance test could not attain complete
pressurization
V063 Comp Cant Hold Pres Compliance test could not maintain plat. pressure
V064 COMP Cant DePres Pressure not dropping during compliance test
V065 COMP Inv Pres Pressure changing in the wrong direction during
compliance test
V066 COMP Inv Pres Pressure changing in the wrong direction during
compliance test
V067 COMP Sys < VentHose Irregular compliance test results
V069 Comp Unexpected Error Compliance test entered an unexpected error state
V070 COMP Sys Fail in Range Unusable compliance value
V071 COMP Sys Fail out of Range Unusable compliance value
V072 COMP VH Fail in Range Unusable vent hose compliance value
V073 COMP VH Fail out of Range Unusable vent hose compliance value
4-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Refer to Table 4-8: for Real Time Value item descriptions and acceptable
range data.
Figure 4-7. Real Time Values Screen Appearance with PCB 4116632
4-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Refer to Table 4-9: for Battery/Charger Status error message list and Table 4-10:
for PSC (Power Supply Controller) error message list as shown on the following
screen.
Figure 4-8. Real Time Values Screen Appearance with PCB 4118079
Press the knob to return to the Service Log category screen.
4-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Table 4-8: Real Time Values Screen Key and Acceptable Range Data
CAL Calibration State of O2 Sensor and Calibration Average stored valve between OFFSET
value Min and OFFSET Max
Pump Pump Motor Speed in RPM No MIN/MAX, typically 2400 to 2800 RPM
Speed when running
Pump Percentage of full power in use (only on 0 to 100%, typically 6 to 15% running
Power 4118079 PCB)
4-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Charger OFF The PSC has detected a +28V supply Refer to Table 4-10: for additional
failure or a Charge error. conditions.
Slow Charging The charger is trickle charging the Normal after the machine has had no
battery until voltage rises to 24V. AC power for an extended period of
time.
Bulk Charging The charger delivers the maximum Normal after the machine has run on
charging current to the battery. battery.
Full Charge The charger keeps the battery voltage Normal after the machine has run on
at a consistent charging voltage, battery and been recharged.
which is above the rated voltage, for 3
hours.
Holding Full Charge The charger is maintaining full charge Normal when battery is in good
on the battery. condition and has been charged
correctly.
No PSC Data The PSC is not communicating. If condition persists, check service
log, possible replace PSC.
Charger Error Charging process has timed out Check battery, connectors
+28V supply failure <25.2 or >30.8 Volts Check AC power, fuse or 28V supply
4-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Turn the rotary knob to select Activate, then press the knob to confirm. The next
preventive maintenance due date can then be set using the screen described on
the next page.
Service 14:43
Preventive
Maintenance
Activate
4-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
With Activate selected, the screen will appear as shown in Figure 4-10.
Turn the rotary knob to position the box under Mth and press the knob to
highlight the box. Turn the knob to select the next month preventive
maintenance is due, and press the knob to enter the month.
With the box under Day, press the knob to highlight the box. Turn the
knob to select the desired day, then press the knob to enter the day.
With the box under Year, press the knob to highlight the box. Turn the
knob to select the year, and press the knob to enter the year.
Service 14:48
Preventive
Maintenance
4-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.5 General
When the General category is selected, the screen shown in Figure 4-11. will
appear.
Turn the rotary knob to scroll down to select the desired function, then press the
knob to confirm. The general screens are described in the following pages.
Service 15:16
General
Serial Number
Reset Hours Run
Reset Last Service Date
Select Market Kit
4-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 15:30
General
Serial Number
1 2 3 4 5 6 7 8
4-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
When Reset Hours Run is selected, the screen will appear as shown in
Figure 4-13. Turn the rotary knob to select Reset Hours Run, and press
the knob to highlight the selection. Then turn the knob to select Yes or
No. Selecting Yes will reset the hours run to zero. Press the knob to enter
the selection.
Service 8:59
General
4-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
When Reset Last Service Date is selected, the screen will appear as
shown in Figure 4-14. Turn the rotary knob to select Reset Last Service
Date, and press the knob to highlight the selection. Then turn the knob
to select Yes or No. Selecting Yes will reset the last service date to the
current date. Press the knob to enter the selection.
Service 9:06
General
4-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
When Select Market Kit is selected, the screen will appear as shown in
Figure 4-15. or Figure 4-16. Turn the rotary knob to scroll to Select
Market Kit, and press the knob to highlight the selection. Then turn the
knob to select U.S. or Non-U.S.
When U.S. is selected, the only language supported is English and the
pressure unit of measure is set to cmH2O. The options to choose a
different language or pressure unit are eliminated while in the “U.S.”
mode.
Service 9:06
General
4-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
4-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.6 Calibration
When the Calibration category is selected, the screen shown in Figure 4-17. will
appear.
Turn the rotary knob to scroll down to select the desired function, then press the
knob to confirm. The calibration screens are described in the following pages.
Service 9:08
Calibration
4-25
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
When Fresh Gas Flow is selected, the screen shown in Figure 4-18.
appears.
Service 9:12
Freshgas Calibration
Stored Zero: PASS
Zero calibration procedure:
- Close all flow control valves.
- Verify no fresh gas flow is present.
- Touch “Store Zero” key.
- At completion of calibration
“Stored Zero” shall indicate
Pass/Fail.
Store Exit
Zero
Press the Exit key to return to the Calibration category screen, or the key
to return to the Standby screen.
4-26
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Service 9:24
Pressure Calibration
Current Airway Pressure: 709Current offset value: 716
Current Ventilator Pres.: 760Current offset value: 732
Zero calibration procedure:
- Remove pressure sample line and ventilator hose
from breathing system. Expose to air.
- Let current pressure value stabilize.
- Touch “Store Zero” key to enter the current
value as the calibration zero.
Store Exit
Zero
Press the Exit key to return to the Calibration category screen, or the key
to return to the Standby screen.
4-27
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 9:34
O2 Calibration
Current cell A: 1
Current cell B: 1
Press the Exit key to return to the Calibration category screen, or the key
to return to the Standby screen.
4-28
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Service 10:06
PEEP Calibration
Calibration: PASS
PEEP calibration procedure:
- Attach breathing hose from expiratory
valve to inspiratory valve and remove breathing bag.
- Touch “Calibrate” key.
- At completion of calibration
“Calibration” shall indicate
Pass/Fail. PWM, Pres: 170 44
Target Pres: 14
Cali- Exit
brate
Press the Exit key to return to the Calibration category screen, or the key
to return to the Standby screen.
4-29
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
4.7 Configure
When the Configure category is selected, the screen shown in Figure 4-22. will
appear.
Turn the rotary knob to scroll down to select the desired function, then press the
knob to confirm. The flowmeter screens are described in the following pages.
Service 10:10
Configure
System Settings
Standard Options
Secure Options
4-30
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
When the System Settings category is selected, the screen shown in Figure 4-23.
or 4.7.1.1 will appear.
Turn the rotary knob to scroll down to select the desired market kit, then press
the knob to confirm.
4-31
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
NOTE: Machines from the factory are configured with the appropriate model
type. Other than performing a software download or PCB replacement,
the model type should not be changed. Changing the model type will
disable some ventilator options previously configured and will require
reconfiguration.
4-32
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Turn the rotary knob to scroll down to select the desired function, then
press the knob to confirm.
4-33
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
4.7.2.1 Flowmeter
Service 11:02
Flowmeter
O2 Position
Gas Selection
Flowtube Res.
Pressure
4-34
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Service 11:11
Flowmeter
O2 Position
RIGHT
4-35
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 11:19
Flowmeter
Gas Selection
3
4-36
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Service 11:19
Change State
High Res
4-37
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
4.7.2.2 O2 Whistle
Service 11:27
O2 Whistle
Change State
ENABLE
4-38
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.7.2.3 Alarms
Service 11:27
Alarms
No Fresh Gas
Fresh Gas Low
Threshold Low
4-39
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 11:27
No Fresh Gas
Change State
ENABLE
4-40
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Service 11:34
Change State
ENABLE
4-41
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 11:34
Threshold Low
Change State
ENABLE
4-42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.7.2.4 Pressure
Service 11:34
Pressure
Ambient Pressure
Plateau/Mean Dis.
4-43
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
4-44
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4-45
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
When the Secure Options category is selected, the screen shown in Figure
4-38. will appear.
Turn the rotary knob to scroll down to select the desired function, then
press the knob to confirm.
4-46
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
Turn the rotary knob to select Parameters, then press the knob to confirm. The
serial port communication parameters can then be set using the screen
described on the next page.
Service 11:37
Serial Port
Parameters
4-47
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued) FABIUS GS
Service 11:42
STOP BITS 1 or 2 1
DATA BITS 7, 8 8
To return to the main service screen, scroll to the and press the knob.
4-48
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS DIAGNOSTICS (continued)
4.9.1 Access the ‘Real Time Values’ screen. Refer to Section 4.0.
4.9.2 Press the ‘PUMP ON/OFF’ softkey to toggle the pump between the On
and Off state.
4-49
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES
This section outlines removal and replacement procedures for the field-replaceable
assemblies of the Fabius GS Anesthesia System.
NOTE: The PMS PROCEDURE given in Section 7.0 must be performed after
any replacement, removal, calibration or adjustment procedure.
5-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
NOTE: The following procedure will require making adjustments and calibrations with
the possibility of removal of accessories and/or additional monitoring equipment.
Therefore, take note of these devices for reconfiguration and/or reinstallation
after the PMS procedure is complete.
Replacement of a yoke, check valve, or cylinder regulator requires that the yoke
be removed from the anesthesia machine. Figure 5-1. shows the cylinder yokes
mounting arrangement. Access to the yoke mounting screws and tubing
connections is at the underside of the table top at the back of the machine.
5.1.1 Disconnect all pipeline hoses and set the System Power switch to ON.
5.1.4 Open the other gas flow control valves to drain pressure from the
system.
5.1.5 Close the O2 cylinder valve, and close the flow control valves. Press the
O2 Flush valve to drain oxygen pressure from the system.
5.1.6 Set the System Power switch to OFF, disconnect AC power from the
machine, and remove any third party monitoring devices, if applicable.
WARNING: Ensure the cylinder is stored in a safe place and is laid on its side.
5.1.9 Hold the check valve assembly with a wrench, and carefully turn the
yoke counter-clockwise to remove it.
NOTE: If a yoke is being replaced, verify that the pin indexing arrangement
and the label are in agreement with the gas designation stamped on
the mounting surface of the yoke. Refer to the parts list in Section 9.0.
NOTE: Where a regulator is replaced, assemble the fittings and check valve to
the regulator with Loctite #271 (red).
5.1.11 Install the regulator assembly in the machine. Hold the check valve
assembly with a wrench and carefully tighten the yoke to the correct
horizontal position.
5-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
BOTTOM VIEW
SV00472
O2
O2
* *
** **
*
*
O2
N2O OR AIR
REGULATOR
O2 YOKE
O2 YOKE * HP COPPER TUBE CONNECTION N2O OR
** LP TUBE CONNECTIONS AIR YOKE
5-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.1.12 Reattach the yoke to the machine with the hardware previously
removed.
5.1.13 If a new cylinder is being installed, remove the old sealing washer from
the gas inlet of the yoke and install a new washer.
5.1.14 Install the correct cylinder on the yoke, making sure that the index
pins are properly engaged before tightening the handle bolt. The
cylinder should hang vertically after the handle is tight.
5.1.15.1 Open the cylinder valve and check for a pressure indication on
the corresponding cylinder pressure gauge.
NOTE: The cylinder used for this test must contain the following minimum
pressure:
O2 : 1000 Psi
N2O : 700 Psi
5.1.15.2 Close the cylinder valve and remove the cylinder from the
yoke.
5.1.15.3 For any gas, the pressure should not drop more than 50 psi in
two minutes.
5.1.18 Restore AC power to the machine and reinstall any third party
monitoring devices previously removed.
5-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Each pressure gauge is secured by two 6-32 kep nuts As shown in Figure 5-2.
Access to gauge mounting nuts and tubing connections is at the underside of the
table top at the back of the machine.
5.2.1 Disconnect all pipeline hoses and set the System Power switch to ON.
5.2.4 Open the other gas flow control valves to drain pressure from the
system.
5.2.5 Close the O2 cylinder valve, and close the flow control valves. Press the
O2 Flush valve to drain oxygen pressure from the system.
5.2.6 Set the System Power switch to OFF, disconnect AC power from the
machine, and remove any third party monitoring devices, if applicable.
WARNING: Ensure the cylinder is stored in a safe place and is laid on its side.
5.2.9 Remove the two kep nuts securing the gauge to its mounting plate.
5.2.10 Install the replacement gauge and secure it with the two kep nuts
previously removed. Ensure that the replacement gauge is mounted
straight onto the mounting plate.
5-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
NOTE: The cylinder used for this test must contain the following minimum
pressure:
5.2.13.1 Open the cylinder valve and check for a pressure indication on
the corresponding gauge.
O2 : 1000 Psi
N2O : 700 Psi
5.2.13.2 Close the cylinder valve and remove the cylinder from the
yoke.
5.2.13.3 For any gas, the pressure should not drop more than 50 psi in
two minutes.
5.2.16 Restore AC power to the machine and reinstall any third party
monitoring devices previously removed.
SV00471
TUBING CONNECTION
GAUGE MOUNTING
NUTS
5-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
The auxiliary oxygen flowmeter is attached to the outside of the ventilator door
on the left side of the machine, by a stud and nut arrangement - accessible by
opening the door. A flexible O2 supply tube from the flowmeter connects to a T-
fitting below the floor of the ventilator housing. Figure 5-3. shows the
mounting and tubing arrangement.
5.3.1 Disconnect all pipeline hoses and close all cylinder valves.
5.3.2 Press the O2 Flush button to drain oxygen pressure from the system.
5.3.3 Set the System Power switch to OFF and disconnect AC power from
the machine.
5.3.5 Cut the tie strap on the flexible tube at the T-fitting, and disconnect the
tube.
5.3.6 Cut the tie strap on the flexible tube just below the floor of the
ventilator housing.
5.3.7 Remove the screw nuts securing the auxiliary O2 flowmeter to the door,
and remove the flowmeter.
5.3.8 Position the replacement flowmeter on the ventilator door (feed the flex
tubing through the clearance hole) and secure the auxiliary O2
flowmeter with the two nuts that were previously removed.
5.3.9 Connect the flex tubing to the T-fitting and secure it with a tie strap.
5.3.10 Install a tie strap on the flex tube just below the floor of the ventilator
housing.
5.3.12 Connect the pipeline hoses and restore AC power to the machine.
5-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.4 Vaporizers
WARNING: For Vapor 19.x series, do not tilt a vaporizer that contains
anesthetic agent more than 45 degrees. Failure to observe this
precaution will render the handwheel calibration invalid.
5.4.3 Remove the filler and drain plugs, and drain the vaporizer into a
suitable container. Dispose of the residual agent in an approved
manner.
5.4.4 Make the necessary machine adjustments to direct gas flow through
the breathing system to the scavenger.
5.4.8 Turn the vaporizer handwheel to 0 (zero), and replace the filler and
drain plugs.
5.4.9 Turn the oxygen flow off, and set the Main Power switch to OFF.
5.4.10 Un-lock the vaporizer and carefully lift it from the manifold.
5-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV00518
LOCKING
LEVER
SELECTOR
LEVER
O-RINGS
VAPORIZER MOUNTING
AND EXCLUSION SYSTEM
5-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.4.12 Set the handwheel on the replacement vaporizer to its Zero position.
5.4.13 Inspect the O-rings on the manifold and replace them if needed.
5.4.14 Install the replacement vaporizer on the manifold and turn the locking
lever to secure the vaporizer to the manifold.
5.4.15.1 Move the selector lever to the left. The left vaporizer
handwheel should be locked, and the right vaporizer
handwheel should be free to turn. Return the right vaporizer
handwheel to zero.
5.4.15.2 Move the selector lever to the right. The right vaporizer
handwheel should be locked, and the left vaporizer handwheel
should be free to turn. Return the left vaporizer handwheel to
zero.
5-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
The O2 flush valve is accessible from the underside of the table top. Valve
removal requires disconnecting the plastic tubing and removal of an L-fitting.
Figure 5-5. shows mounting and tubing arrangement for the valve.
5.5.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.5.4 Press the O2 flush valve to drain oxygen pressure from the system.
5.5.5 Disconnect the tubing from the straight fitting and from the L-fitting
on the O2 flush valve.
5.5.6 Remove the L-fitting from the outlet of the O2 flush valve.
5.5.7 Un-screw the plastic retaining nut from the valve body, and withdraw
the valve from the front of the table.
5.5.8 Install the replacement valve in the table top, and secure it with the
plastic retaining nut.
5.5.10 Reconnect the tubing to the straight fitting and to the L-fitting.
5.5.11 Install the appropriate label on the replacement valve. Refer to Section
9.0 for the correct part number.
5-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
STRAIGHT FITTING
L-FITTING
O2 FLUSH VALVE
RETAINING NUT
5-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.6 Caster
The casters are threaded into the frame rails on the trolley. A typical
arrangement is shown in Figure 5-6. Caster replacement requires that the
machine be tilted to provide enough clearance for the caster stem to be
withdrawn from the frame rail.
WARNING: Do not tilt the machine more than 10 degrees or raise the casters
more than 3½ inches from the floor. Failure to observe this
precaution may result in a tip-over, causing personal injury.
Vaporizers containing anesthetic agent may also be damaged.
5.6.1 Obtain a brace capable of supporting one side of the machine with its
casters two to three inches from the floor.
5.6.2 Remove all unsecured equipment and accessories from the machine.
5.6.4 Using at least two people, tilt the machine until the casters on one side
are raised two to three inches from the floor, and position the support
brace under the trolley rail between the front and back casters.
5.6.5 Un-screw the caster from the frame using a caster wrench P/N
S010055.
5.6.6 Apply an ample amount of Loctite (#271 - Red) to the stem of each
replacement caster, then thread the caster into the frame and tighten
the caster.
5.6.7 Using at least two people, tilt the machine, remove the support brace
and carefully lower the machine to the floor.
5.6.8 Check for proper operation of the caster and ensure that the front
casters lock properly.
5.6.9 Perform the PMS procedure given in Section 7.0, including a vaporizer
calibration verification.
5-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
SV00507
THREADED STEM
*R
5-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.7 Ventilator
The ventilator and door assembly are retained in the monitor housing by a
hinge rod. Figure 5-7. and Figure 5-8. show the hinge arrangement and the
disconnects needed to remove the ventilator from the machine. Figure 5-9.
shows the lower cylinder and motor arrangement. While it is possible to service
some components without removing the ventilator assembly from the machine,
the complete disassembly sequence is given below.
Ventilator Removal:
5.7.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.7.3 Remove the back panel to expose the controller PCB assembly. Provide
support for the panel to prevent damage to the power supply cable
connections.
NOTE: Make note of the positioning and routing of all cables and pneumatic
lines for reassembly.
5.7.4 Swing open the ventilator door. Disconnect the ventilator cables from
J11 and J12 on the controller PCB, and separate the flying connector
on the cable attached to J10 on the controller PCB. Feed these three
cables forward into the ventilator compartment.
5.7.6 Disconnect the ventilator tubing from the T-fitting (see Figure 5-7. )
and feed this tube into the ventilator compartment.
5.7.7 If applicable, disconnect the O2 supply tube from the auxiliary O2 flow
meter.
5.7.8 Disconnect the ventilator ground wires from the stud on the ventilator
door.
5.7.9 Remove the E-ring from the hinge rod. While supporting the ventilator
assembly by the top handle, slide the hinge rod down and out through
the underside of the table.
5.7.10 Remove the ventilator from the door by removing Screw A (see Figure
5-8. ) and sliding out the upper and lower hinge bushings.
5-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
E-RING
UPPER
HINGE
BUSHING
VENTILATOR
ASSEMBLY
HINGE ROD
LOWER HINGE
BUSHING
SPACERS
J11 TACH
SENSOR
J12 HOME
SENSOR
DETAIL OF CONTROLLER
PCB WITH VENTILATOR JI0
CONNECTIONS MOTOR
T-FITTING
TUBE FROM
SV00523 VENTILATOR
5-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
TOP VIEW OF
VENTILATOR ASSEMBLY HINGE DETAILS
HINGE UPPER
ROD HINGE
BUSHING
VENT
PRESSURE
HOSE DOOR
HIGH PRESSURE CONNECTION
VALVE
NEGATIVE PRESSURE
RELIEF VALVE LOWER
HINGE
BUSHING
BAND CLAMP
TO LS3 ON
PATIENT CONTROLLER PCB
ASSEMBLY
UPPER DIAPHRAGM
UPPER
CYLINDER
SCREW A
GROUND WIRE
SV88075
*P CONNECTION
5-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.7.11 Release the three clamps holding the patient assembly to the upper
cylinder; lift the patient assembly up and out.
5.7.12 Lift the diaphragm out of the upper cylinder as shown in Figure 5-9. in
the direction of the arrow.
.
UPPER DIAPHRAGM
SV00546
5.7.14 Carefully reinstall the patient assembly, making sure the diaphragm
seats properly. Secure the three clamps holding the patient assembly to
the upper cylinder.
5-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.7.15 Note the position of the clamp for reassembly, then remove the bolt
securing the band clamp holding the upper and lower cylinders
together. Open the band clamp and move it down onto the lower
cylinder. Separate the upper cylinder from the lower cylinder.
5.7.16 Using the proper “pozi” driver, remove the three screws from the top of
the piston. See Figure 5-12.
5.7.17 Lift the piston and diaphragm up and out of the lower cylinder.
5.7.19 Place the piston upside down on a table. Remove the retaining ring
(circlip) securing the diaphragm at the inside of the piston, and pull the
diaphragm up off the piston. See Figure 5-10.
RETAINING RING
O-RING
LOWER DIAPHRAGM
SV00545
NOTE: The gap in the retaining ring (circlip) shall be aligned opposite the
mark on top of the piston as shown in Figure 5-12. This note does not
apply on later units, where the mark on the piston does not exist.
5-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
O-RING
LOWER DIAPHRAGM
SV00544
5.7.22 Slightly roll the diaphragm down the piston until the diaphragm
(overlaps) covers the O-ring evenly around the bottom.
5.7.23 Install the piston over the motor and reinstall the three screws that
were previously removed.
NOTE: The lip on the lower diaphragm is captured between the upper and
lower cylinders, and must be correctly installed.
5.7.24 Place the upper cylinder on the lip of the diaphragm, and carefully fit
the lip into the groove in the upper cylinder.
5.7.25 Carefully press the upper cylinder down until it is resting on the lower
cylinder. Note that the cylinders are keyed to fit together in only one
position. Look down through the upper cylinder and verify that the
lower diaphragm shows a uniform roll around its circumference.
5-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV88059
PISTON
SCREW (3X)
RETAINING RING
GAP
TOP VIEW
OF PISTON
PISTON
LOWER
DIAPHRAGM
O-RING
LIP
LOWER
CYLINDER
MOTOR/BALLSCREW
ASSEMBLY
RETAINING
RING
*P
GROUND PLATE
MOTOR ASSEMBLY CENTER CAP
SCREW (4X)
5-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.7.26 While holding the cylinders together, move the band clamp into
position and reinstall its bolt to join the cylinders. Make sure the band
clamp is positioned as shown in Figure 5-8. or the ventilator door may
not close properly.
Motor/Ballscrew Assembly:
NOTE: Upper and lower cylinders must be separated, and piston must be
removed.
5.7.27 At the bottom of the lower cylinder, remove the ground plate, and
remove the center cap.
5.7.28 Disconnect and remove the tach cable and positioning sensor cables.
Note the arrangement of the cables so they can be reinstalled in the
same manner. (The motor cable remains with the motor/ballscrew
assembly.)
5.7.29 Remove the four screws securing the motor/ballscrew assembly to the
lower cylinder. Note the orientation of the unit so the replacement
assembly can be installed in the same manner.
5.7.32 Reinstall the tach and sensor cables. Route all cables in the original
manner.
NOTE: The lip on the lower diaphragm is captured between the upper and
lower cylinders, and must be correctly installed.
5.7.35 Place the upper cylinder on the lip of the diaphragm, and carefully fit
the lip into the groove in the upper cylinder.
5.7.36 Carefully press the upper cylinder down until it is resting on the lower
cylinder. Note that the cylinders are keyed to fit together in only one
position. Look down through the upper cylinder and verify that the
lower diaphragm shows a uniform roll around its circumference.
5.7.37 While holding the cylinders together, move the band clamp into
position and reinstall its bolt to join the cylinders. Make sure the band
clamp is positioned as shown in the illustration, or the ventilator door
may not close properly.
5-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
Ventilator Reassembly:
5.7.38 Position the ventilator against the door; slide the upper and lower
hinge bushings in place, and reinstall Screw A (Figure 5-8. ) to attach
the ventilator to the door.
5.7.39 Locate the hinge rod that was previously removed. Orient it with the
E-ring groove toward its upper end. Hold the ventilator assembly by its
top handle and position its lower hinge in the monitor housing (make
sure the spacers are in place). Sight down through the hinge bushings
to align the lower hinge bushing with the hole in the monitor housing.
Slide the hinge rod up through the table and up through the lower
hinge bushing. Tilt the ventilator assembly into position and continue
sliding the rod up and into the upper hole in the monitor housing until
the E-ring groove on the rod is visible just above the upper hinge
bushing. Reinstall the E-ring on the hinge rod.
5.7.40 Feed the three ventilator cables through the opening in the back wall
of the ventilator compartment. Reconnect these to J11 and J12 on the
controller PCB, and reconnect the remaining one to the flying
connector on the cable attached to J10 on the controller PCB.
5.7.41 Feed the tubing from the ventilator patient assembly through the
opening in the back wall of the ventilator compartment and reconnect
it to the T-fitting.
5.7.46 Close the ventilator door and restore all external hose connections.
5.7.47 Restore AC power to the machine, and test all ventilator functions.
5-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Patient Assembly:
In case of repair, replace the high pressure valve, the negative pressure relief
valve, or the complete patient system. See Figure 5-8.
NOTE: Ensure that the correct Patient Assembly is being installed. Refer to
the Spare and Replacement Parts section of this manual to verify the
correct assembly per software version of the machine.
5.7.49 Disconnect the ventilator hose from the side of the machine.
5.7.52 Release the three clamps holding the patient assembly to the upper
cylinder, and lift the patient assembly up and out.
5.7.54 Remove all residual thread sealant from the high pressure valve mount
on the patient system.
NOTE: Ensure that the correct Negative Pressure Relief Valve is being
installed. Refer to Section 9.0 of this manual to verify the correct valve
per software version of the machine.
5.7.57 Unscrew the negative pressure relief valve using plastic-jaw pliers.
5.7.58 Remove all residual thread sealant from the negative pressure relief
valve mount on the patient system.
5-25
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.7.61 Carefully reinstall the patient system, making sure the diaphragm
seats properly.
5.7.62 Secure the three clamps holding the patient assembly to the upper
cylinder.
5-26
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
The gas inlet assembly is attached to the inside of the back panel on the gas box
assembly as shown in Figure 5-13. This assembly includes regulators, check
valves, filters, and the O2 supply pressure alarm switch. Refer to the parts list
in Section 9.0 for details on internal parts locations for replacement.
5.8.1 Set the main power switch to OFF, and close all cylinder valves.
5.8.3 Disconnect the pipeline hoses from the machine. Deplete all cylinder
and pipeline pressures.
5.8.4 Remove the back panel from the gas box. Provide support for the panel
while disconnecting the internal tubing from the gas inlet assembly.
5.8.5 Disconnect the wire harness from the O2 supply pressure alarm
switch.
NOTE: Make a note of the connections and their destinations in the following
step - for later reassembly.
5.8.7 Remove the screws securing the gas inlet assembly to the back panel.
5.8.8 Install the replacement assembly on the back panel in the same
manner as the original.
5.8.10 Reconnect the wire harness to the O2 supply pressure alarm switch.
5-27
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
GAS INLET
ASSEMBLY
P4
WIRE HARNESS
P1
P5
P6
P2
P7
O2 SUPPLY P8
PRESSURE
ALARM SWITCH P3
BACK PANEL
SV00473
5-28
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.9 Battery
The two 12 V rechargeable batteries are located in the vent & monitor assembly
behind the front bezel. The batteries and their connections are accessible
through the ventilator compartment as shown in Figure 5-14.
5.9.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.9.2 Remove the battery fuse from its holder (located on the back of the
machine next to the main power switch).
5.9.4 Loosen the retainer screw securing the battery retainer bracket.
Remove the bracket, and slide the batteries down to the lower level of
the ventilator compartment.
5.9.5 Note the battery wire harness colors to the battery terminal markings
for reconnection. Disconnect the battery wire harness from the battery
terminals and remove the batteries.
5.9.6 Install the replacement batteries into the lower level of the ventilator
compartment with their terminals oriented as shown in the
illustration.
5.9.7 Reconnect the battery wire harness to the battery terminals. Make
sure the connections are as shown in the illustration.
5.9.9 Reinstall the battery retainer bracket and secure it with the retainer
screw.
5.9.11 Reinstall the battery fuse and restore power to the machine.
5-29
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
RETAINER SCREW
BLACK WIRE
CABLE CLIP
YELLOW WIRES
RED WIRE
SV0052
VENTILATOR COMPARTMENT
5-30
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
The power supply is attached to the inside of the controller assembly back
panel. Figure 5-15. shows the mounting and connection arrangement for the
power supply.
5.10.1 Set the main power switch to OFF and remove AC power from the
machine.
5.10.2 Remove the screws securing the back panel on the controller assembly;
while supporting the panel, carefully open it as shown in the
illustration.
CAUTION: The power supply contains static sensitive devices. Use ESD
protection while handling this assembly.
5.10.3 Disconnect the input cable from J1 on the power supply, and disconnect
the output cable from J3 on the power supply.
5.10.4 Disconnect the ground wire from the back panel. Remove the wires
from the cable clips, and separate the panel from the machine.
5.10.5 Remove the two power supply mounting screws and lock washers
holding the power supply to the panel.
5.10.6 Install the replacement power supply on the panel - oriented as shown
in the illustration.
5.10.7 While supporting the panel, reconnect the ground wire, and reconnect
the input and output cables to the power supply at J1 and J3.
5.10.11 Set the main power switch to ON and verify that the power-up
sequence is successfully completed.
5-31
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV00464
WIRE CLIPS
GROUND WIRE
CONNECTION
INPUT
CABLE (J1)
POWER SUPPLY
BACK PANEL OF
CONTROLLER
ASSEMBLY
POWER SUPPLY
MOUNTING SCREWS
5-32
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
The control PCB assembly is accessible by removing the back cover of the
controller housing. Figure 5-16. and Figure 5-17. shows the mounting
arrangement and electrical connections to this assembly.
NOTE: Verify the original software level of the PCB prior to removal.
Replacement control PCBs may contain a different software level, a
software update may be necessary. If applicable, perform a software
download in accordance with the applicable Software Upgrade Service
Procedure after the replacement PCB is installed.
5.11.1 If possible, record the market settings on the unit for re-configuration
of the replacement PCB. Refer to Section 4.0 to access the appropriate
service screen. If these settings can not be displayed, refer to the tables
below for the settings that correspond to your region.
NOTE: The following step is applicable on units with Software Version 2.X.
5.11.3 Set the main power switch to OFF and disconnect AC power from the
machine.
5.11.4 Remove the back panel from the controller housing. Provide support
for the panel to prevent damage to the power supply cable connections.
CAUTION: The control PCB contains static sensitive devices. Use ESD
protection when handling this assembly.
5.11.5 Disconnect all of the cables from the board connectors indicated in the
illustration. Also disconnect the tubing from the pressure sensors on
the PCB.
5.11.6 Remove the screws holding the PCB assembly to the controller
housing, and carefully remove the PCB assembly.
5.11.7 Install the replacement PCB assembly with the hardware that was
previously removed.
NOTE: FOR CONTROL PCB P/N 4116632: Verify jumper at JP3 is installed
on both pins. This activates the 3V lithium battery (BT1) for the non-
volatile RAM.
NOTE: FOR CONTROL PCB P/N 4118079: Verify jumper at JP2 is installed
on both pins. This activates the 3V lithium battery (BT1) for the non-
volatile RAM.
5.11.8 Reconnect all of the cables and hoses that were previously
disconnected.
5-33
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
NOTE: Replacement PCB’s come with a new hose attached to the LS2 pressure
sensor. Remove and discard the original 2-hose assembly from the
breathing pressure interface panel connector inside the unit and attach
the new hose to the breathing pressure interface connector.
5.11.10 Restore power to the machine and observe the power -up diagnostic
display (see Section 4.0) to verify that the replacement PCB is working
properly.
5.11.11 Access the appropriate service screen and update the machine serial
number, clear service log and reset hours run. The machine serial
number is located on the rear of the machine.
5.11.12 Perform the market settings per the original machine configuration or
customer requirements.
NOTE: The default “Market Selection” for the replacement PCB is “U.S.” To
change the language and pressure measurement units, the “Market
Selection” must be changed to “Non-U.S. (see Section 4.0). The market
kit setting will vary depending upon the current software installed on
the machine. Table 1: is to be used for SW Versions 1.2n and 1.3n and
Table 2: is to be used for SW Version 2.X. Refer to the appropriate table
for the market settings that correspond to your region.
5.11.15 Perform the PMS procedure, including all calibrations and site
configurations, given in Section 7.0.
5-34
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Table 1: Default Market Kit Settings for SW Versions 1.2n and 1.3n
5-35
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5-36
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
VENT
CHAMBER
PRESSURE
SENSOR
BREATHING
LS1 LS3 PRESSURE
SV00503 INTERFACE
PNEUMATIC
ASM LS2
F3 F1 F2
J6 POWER SUPPLY
5-37
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
J9
KEYPAD
J2 INTERFACE
JP3 J15
FRONT PANEL
JP2 SERIAL PORT
DISPLAY
INTERFACE
BT1
J18
SPIROLOG
VENT
LS3 CHAMBER SENSOR
PRESSURE
SENSOR
LS1 LS2 J17
SV00612
O2 SENSOR
PNEUMATIC
BREATHING
ASM PRESSURE
INTERFACE
J12
F1 HOME
SENSOR
J16 J10
J8 J21
FLOW MOTOR
PUMP J7 BOOST
SENSORS
PEEP CAPS
J6
J1 VALVE
POWER
FRESH GAS
SUPPLY
DISPLAY
J4
J11 ON/OFF
J5 TACH SWITCH
BATTERY SENSOR
5-38
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.12.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.12.2 Remove the back panel from the controller housing. Provide support
for the panel to prevent damage to the power supply cable connections.
5.12.5 Remove the rear mounting screw and carefully slide the assembly out
of the controller housing. (Make sure the back panel with the power
supply attached is supported to prevent damage to its connections.)
5.12.6 If installing a pump retrofit kit, proceed with the following steps. If a
pump retrofit kit is not required, skip to step 5.12.7.
5-39
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV279001
CONTROLLER PCB
J8
PUMP CABLE LS1 SENSOR
CONNECTION CONNECTION
MUFFLER
ASSEMBLY
EXHAUST
PORT
PUMP
ASSEMBLY
CAPTIVE
MOUNTING
SCREW
DAMPER
DEVICE (X3)
REAR MOUNTING J7
SCREW AND MUFFLER PEEP VALVE
CABLE CLAMP ASSEMBLY CONNECTION
SUCTION
PORT PEEP HOSE
PNEUMATIC
ASSEMBLY VACUUM HOSE
5-40
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
PUMP
ASSEMBLY
FOAM
CUSHION
SP279002
5-41
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
CAP
SOCKET HEAD 14 IN.
SCREW AND HOSE
LOCK WASHER (INLET PORT TO
(3X) MUFFLER ASSEMBLY
SUCTION PORT)
J8 PUMP CABLE
CONNECTION 5.25 IN.
HOSE
(OUTLET PORT TO
MUFFLER ASSEMBLY
EXHAUST PORT)
TIE
STRAP FIXING
PLATE
PUMP ADAPTER
MOUNTING
BRACKET
PAN HEAD
ACORN SCREWS
NUTS
SP279003
REAR MOUTNING
SCREW AND CABLE
CLAMP
5-42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.12.6.3.7 Using the two pan head screws, attach the Motor Pump
Assembly to the Pump Adapter Mounting Bracket.
5.12.6.3.8 Secure ferrule from Motor Pump Assembly to Pump
Adapter Mounting Bracket using tie strap (P/N
4106068).
5.12.6.3.9 Install 14 in. hose (P/N 1190520) from inlet port of Motor
Pump Assembly to suction port of Muffler Assembly.
(Refer to Figure 1 for Muffler Assembly connection loca-
tions)
5.12.6.3.10Install 5.25 in. hose (P/N 1190520) connecting to outlet
port of Motor Pump Assembly to exhaust port of Muffler
Assembly.
5.12.7 Slide the replacement pneumatic assembly into the controller housing
and secure it with the captive mounting screw. reinstall the rear
mounting screw and cable clamp.
5-43
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
Following is the procedure for removing and replacing the flow meter bezel
assembly. See Figure 5-21. This assembly comprises the fresh gas flow meter,
flow control valves, flow sensors and filter, pipeline pressure gauges, SORC, and
the fresh gas display PCB. Replacement procedures for these individual sub-
assemblies are given in subsequent paragraphs.
5.13.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.13.2 Disconnect the pipeline and cylinder supplies, and deplete all
pressures.
5.13.3 Remove the two screws at the back of the machine securing the flow
meter bezel assembly.
5.13.4 At the front of the machine, carefully pull the assembly forward about
2.5 inches. Disconnect the ribbon cable and the O2 alarm switch wire
harness from the fresh gas display PCB, and disconnect each flow
sensor cable. Continue to pull the assembly forward.
---------------------------------------------------------------------------------------------------
5.13.5 When reinstalling the flow meter bezel assembly, push the assembly
into its compartment; reattach the ribbon cable and the O2 alarm
switch wire harness to the fresh gas display PCB, and reconnect each
flow sensor cable. Continue to push the assembly into its compartment
until it is fully seated.
5.13.6 Reinstall the two screws at the back of the machine to secure the flow
meter bezel assembly.
5-44
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
SV00506
FLOWMETER BEZEL
ASSEMBLY
RIBBON CABLE
FLOWMETER BEZEL
ASSEMBLY RETAINER
SCREWS
5-45
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV77041
FLOWMETER BEZEL
ASSEMBLY
RIBBON CABLE
N2O
AIR
O2
O2 ALARM
SWITCH
FLOWMETER BEZEL
ASSEMBLY RETAINER
SCREWS
5-46
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Access to the fresh gas flow meter requires removal of the flow meter bezel
assembly as described in a previous paragraph. Figure 5-23. shows the tubing
and mounting arrangement for the fresh gas flow tube. For clarity, other sub-
assemblies are not shown in the illustration.
5.14.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.14.3 Remove the screws securing the bezel assembly (refer to the Flow
Meter Bezel Assembly instructions, Section 5.13) and bring the
assembly forward for access to the fresh gas flow meter.
5.14.4 Disconnect the tubing from the flow tube ports and remove the Luer
fittings.
5.14.5 Remove the flow tube mounting nuts to release the flow tube.
5.14.6 Install the replacement flow tube with the mounting brackets and nuts
supplied. Ensure that the float moves freely without sticking, and that
the flow tube is oriented properly.
5.14.8 Reinstall the flow meter bezel assembly in the machine (refer to the
Flow Meter Bezel Assembly instructions, Section 5.13).
5.14.9 Reconnect the pipeline supplies and restore power to the machine.
5-47
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
TUBE TO
VAPOR FLOW
MOUNT TUBE
TUBE FROM
FLOW SENSOR
MANIFOLD
5-48
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Flow control valves can be replaced from the front panel while the flow meter
bezel assembly is in place. Figure 5-24. shows the hardware arrangement for a
typical flow control valve. Should the valve manifold need to be replaced, its
mounting arrangement is shown in the illustration. For clarity, other sub-
assemblies are not shown in the illustration.
5.15.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.15.3 Remove the cover from the knob of the valve to be replaced.
5.15.4 Loosen the nut securing the knob to the valve, and pull the knob from
the valve.
5.15.5 Un-screw the flow control valve from the valve manifold.
NOTE: Use plastic jaw pliers (P/N 7910296) to hold the knob while tightening
the nut in the next step.
5.15.7 Reinstall the knob and tighten the nut to secure the knob to the valve.
5.15.9 Reconnect the pipeline supplies and restore power to the machine.
5.15.10 Set the main power switch to ON and ensure that the gas flow can be
controlled over its entire range.
5-49
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
COVER
NUT
SPRING
WASHER
FLAT
WASHER
KNOB
CHUCK
COLLET
SPACER
FLOW CONTROL VALVE
5-50
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Access to the flow sensors and filter assembly requires removal of the flow
meter bezel assembly as described in Section 5.13. Figure 5-25. and Figure 5-
26. shows the tubing and mounting arrangement for this sub-assembly. For
clarity, other sub-assemblies are not shown in the illustration.
NOTE: The safety valve, filter, and flow sensors can be serviced without
complete removal of this sub-assembly from the flow meter bezel
assembly. Figure 5-25. and Figure 5-26. show exploded views to
indicate connections that would not otherwise be visible.
5.16.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.16.3 Remove the screws securing the bezel assembly (refer to the Flow
Meter Bezel Assembly instructions, Section 5.13) and bring the
assembly forward for access to the flow sensor sub-assembly.
5.16.5 The filter is accessible by un-screwing the safety valve from the
manifold.
The flow sensors and manifold are held in place by a spring clip
arrangement.
The SORC bypass valve is secured to the flow sensor bracket by two
screws and kep nuts as shown in the illustration.
5.16.6 Apply Loctite (#425 - Blue) to the adapter fitting on the end of the
flowsensor, then install the replacement angle connector on the
flowsensor hand tight.
CAUTION: While securing the plastic hex located on the flowsensor, tighten
the angle connector to the flowsensor and align per original
location.
5.16.7 Install the O-ring on the white manifold squarely by twisting the O-
ring side to side until it is fully seated against the shoulder. Do not
allow the O-ring to roll, it will not allow proper sealing once the flow
valve is installed.
5-51
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5.16.8 Place the flowmeter opening squarely over the O-ring and twist the
flowmeter side to side until it is seated onto the manifold. Do not force
the flowmeter over the O-ring or it will not seat properly.
---------------------------------------------------------------------------------------------------
5.16.10 Reinstall the flow meter bezel assembly in the machine (refer to the
Flow Meter Bezel Assembly instructions, Section 5.13).
5.16.11 Reconnect the pipeline supplies and restore power to the machine.
5-52
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
MANIFOLD
AIR
SORC BYPASS
VALVE
O2
FLOWMETER
BEZEL
ASSEMBLY
SV00484
5-53
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
FLOWMETER
BEZEL ASSEMBLY
SORC BYPASS
VALVE
SV77042
SAFETY VALVE
FILTER
MANIFOLD
FLOW SENSOR
HOUSING COVER
5-54
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Access to the pipeline pressure gauges requires removal of the flow meter bezel
assembly as described in Section 5.13. Figure 5-27. shows the tubing and
mounting arrangement for the gauges. For clarity, other sub-assemblies are not
shown in the illustration.
5.17.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.17.3 Remove the screws securing the bezel assembly (refer to the Flow
Meter Bezel Assembly instructions, Section 5.13 given previously in
this section) and bring the assembly forward for access to the pipeline
pressure gauges.
5.17.4 Disconnect the tubing from the fitting on the gauge to be replaced.
5.17.5 Remove the retainer plate screws, and separate the retainer plate with
the gauges from the bezel.
5.17.6 Un-thread the straight fitting from the back of the gauge to be
replaced.
5.17.7 Remove the gauge mounting nut, and remove the gauge from the
retainer plate.
5.17.8 Install the replacement gauge in the retainer plate - make sure it is
oriented properly when tightening the gauge mounting nut.
5.17.12 Reinstall the flow meter bezel assembly in the machine (refer to the
Flow Meter Bezel Assembly instructions, Section 5.13).
5.17.13 Reconnect the pipeline supplies and restore power to the machine.
5-55
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV88012
FLOW METER
BEZEL ASSEMBLY
TUBES FROM
GAS INLET
ASSEMBLY
N2O
AIR
O2
STRAIGHT
FITTINGS (3X)
RETAINER PLATE
SCREW (4X) PIPELINE PRESSURE GAUGE (3X)
5-56
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
Access to the SORC requires removal of the flow meter bezel assembly as
described in Section 5.13. Figure 5-28. shows the tubing and mounting
arrangement for the SORC. For clarity, other sub-assemblies are not shown in
the illustration.
5.18.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.18.3 Remove the screws securing the bezel assembly (refer to the Flow
Meter Bezel Assembly instructions, Section 5.13) and bring the
assembly forward for access to the SORC.
5.18.4 Disconnect the tubing from the SORC where shown in Figure 5-28.
Mark these tubes so they can be reattached to the replacement SORC
in the same manner.
5.18.5 Remove the four SORC mounting nuts and remove the SORC from the
flow meter bezel assembly.
5.18.6 Install the replacement SORC with the four mounting nuts that were
previously removed.
5.18.7 Re-connect the O2 and N2O tubing to the SORC in the same manner as
the original.
5.18.8 Reinstall the flow meter bezel assembly in the machine (refer to the
Flow Meter Bezel Assembly instructions, Section 5.13).
5.18.9 Reconnect the pipeline supplies and restore power to the machine.
5-57
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
5-58
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
Access to the fresh gas display PCB requires removal of the flow meter bezel
assembly as described in Section 5.13. Figure 5-29. shows the cable
connections and PCB mounting details. For clarity, other sub-assemblies are not
shown in the illustration.
5.19.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.19.3 Remove the screws securing the bezel assembly (refer to the Flow
Meter Bezel Assembly instructions, Section 5.13) and bring the
assembly forward.
5.19.4 Remove the two screws securing the flow sensor sub-assembly to the
bezel assembly, and move the sub-assembly aside for access to the fresh
gas display PCB. (Refer to the Fresh Gas Flow Sensors and Filter
Assembly instructions given in Section 5.16.)
5.19.5 Disconnect the cables from the fresh gas display PCB where shown in
Figure 5-29. .
5.19.6 Remove the PCB mounting screws and lock washers, and remove the
PCB.
5.19.7 Install the replacement fresh gas PCB, with the insulator in place
using the hardware that was previously removed.
5.19.10 Reinstall the flow meter bezel assembly in the machine (refer to the
Flow Meter Bezel Assembly instructions, Section 5.13).
5.19.11 Reconnect the pipeline supplies and restore power to the machine.
5-59
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV00483 FLOWMETER
BEZEL
ASSEMBLY
PCB MOUNTING
SCREWS AND
LOCKWASHERS (4X)
RIBBON CABLE
FROM
CONTROLLER
(J1)
INSULATOR
5-60
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
The monitor bezel assembly comprises the keypad, flat panel display, rotary
switch, and LED lamp PCB assembly. Access to the bezel assembly mounting
screws is through the ventilator compartment as shown in Figure 5-30.
5.20.1 Set the main power switch to OFF and disconnect AC power from the
machine.
5.20.3 While supporting the bezel assembly, remove the two screws securing
the left side of the bezel.
CAUTION: Do Not pull the bezel outward, as this will break the locking tabs at
the right side of the bezel.
5.20.4 Slide the bezel to the right to release its locking tabs from the machine.
See Figure 5-30.
NOTE: Verify that bezel locking tabs are clear of the machine before pulling
the bezel away from the machine.
5.20.5 Disconnect the large ribbon cable from the keypad connector, and the
smaller ribbon cable from the display panel.
Refer to the parts list in Section 9.0 for identification of individual components.
5.20.7 Reconnect the ground wire and the two ribbon cables to the
replacement assembly.
5.20.8 Carefully fold the cables behind the bezel assembly as you bring it into
position on the machine. Hold the bezel against the machine and slide
it to the left to engage the locking tabs in the slots in the machine
housing.
5.20.9 While supporting the bezel assembly, reinstall the two screws at the
back of the bezel on the left side.
5.20.11 Restore power to the machine and verify all of the keypad and display
functions.
5-61
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued) FABIUS GS
SV00463
SEE TEXT
KEY PAD
INSERT
BEZEL SCREWS
GROUND
WIRE
KEYPAD
CONNECTOR
L OCKING
TABS
5-62
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS REPLACEMENT PROCEDURES (continued)
5.21.1 Set machine power switch to OFF and disconnect AC power from the
machine.
5.21.2 Remove the expiratory hose terminal from the breathing system, and
slide the Spirolog sensor out of the breathing system.
5.21.3 Disconnect the cable from the sensor by squeezing the connector at the
points opposite the latch while pulling the connector away from the
sensor.
5.21.4 Remove the back panel from the controller housing, and disconnect the
Spirolog cable from J18 on the controller PCB.
5.21.5 Loosen the nut on the strain relief bushing at the interface panel.
Remove the strain relief bushing from the panel while supporting the
rear nut.
5.21.6 Install the replacement Spirolog strain relief bushing into the interface
panel, and tighten the nut.
5.21.7 Connect the replacement Spirolog cable to J18 on the controller PCB.
5.21.8 Route the cable from the machine to the breathing system in the same
manner as the original.
5.21.9 Connect the outboard end of the Spirolog cable to the flow sensor. Note
that the sensor and connector are keyed.
5.21.10 Reinstall the back panel on the controller housing, and restore AC
power to the machine.
5-63
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES
NOTE: Figure 6-1. shows test connections for the O2 regulator adjustment. If
you are adjusting the N2O regulator, connect the test gauge in series
with the N2O pipeline hose.
6.1.2 Connect test pressure gauge (P/N 4114807) between machine’s pipeline
inlet connector and the pipeline supply hose.
6.1.3 Open the cylinder valves and set the System Power switch to ON.
6.1.4 Set the oxygen flow to 4 liters per min. (If you are adjusting the N2O
regulator, also set the nitrous oxide flow to 4 liters per minute.)
6.1.6 Release the push button. After the pressure decay stabilizes, the gauge
should indicate 32-40 psi.
6.1.7 Remove the acorn nut from the regulator to expose the adjusting screw.
For N2O, turn the screw until the test gauge indicates 35 psi. (50 psi
for CSA machines). For O2 and Air, use the compensated regulator
output setting based on the cylinder pressure given in the following
table.
6-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
SV00465
O2 REGULATOR
N2O
REGULATOR
ACORN NUT
O2 PIPELINE
HOSE
TEST GAUGE
6-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
NOTE: Cylinder pressure compensation for the N2O regulator is not required.
* Canada settings
6.1.9 If the O2 cylinder regulator output was adjusted, perform the following
test:
6.1.12 Press the O2 Flush and verify the Lo O2 Supply alarm is not active. If
the alarm is active refer to Section 6.3: Oxygen Supply Pressure Alarm
Switch Adjustment.
6.1.14 Close the cylinder valves and allow pressure to drain from the system.
6.1.15 Close all of the flow control valves and set the System Power switch to
OFF.
6-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
O2 Inlet Regulator:
6.2.1 Bleed all cylinder and pipeline pressures. Disconnect all cylinders and
pipeline hoses.
6.2.2 Remove the rear cover supporting the gas inlet assembly.
6.2.3 Remove the pneumatic hose from the O2 outlet of the gas inlet block
and interconnect a digital manometer as shown in Figure 6-2.
ADJUSTMENT
SCREW
O2 INLET
REGULATOR
TO O2
FLOW CONTROL
VALVE
TEST 'T' O2 OUTLET (TO O2
CONNECTOR FLOW CONTROL VALVE)
M32803
6-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
6.2.6 After the digital manometer stabilizes, adjust the O2 regulator output
pressure to 30 psi.
6.2.8 Close the flow control valve and disconnect the O2 pipeline hose from
the inlet.
6.2.9 Remove test equipment and reconnect the O2 pneumatic hose leading
from the O2 flow control valve to the O2 connector on the inlet block.
6.2.10 Reinstall rear cover and perform the PMS Procedure given in Section
7.0.
6.2.11 Disconnect the N2O pneumatic hose from the outlet connector on the
gas inlet block that connects to the N2O flow control valve.
6.2.13 Reconnect the N2O pipeline hose to the inlet connector and activate
the N2O supply.
6.2.14 Set the O2 and N2O flows to 4 L/min. After the digital manometer
stabilizes, adjust the N2O regulator output pressure to 30 psi.
6.2.16 Close the O2 and N2O flow control valves, and disconnect both pipeline
hoses from the inlets.
6.2.17 Remove the test equipment and reconnect the N2O hose from the flow
control valve to the N2O outlet on the inlet block.
6.2.18 Reinstall the rear cover and perform the PMS procedure given in
Section 7.0.
6-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
6.3.1 Disconnect all pipeline hoses and set the System Power switch to ON.
6.3.4 Open the other gas flow control valves to drain pressure from the
system.
6.3.5 Close the O2 cylinder valve, and close the flow control valves. Press the
O2 Flush valve to drain oxygen pressure from the system.
6.3.7 Remove the rear cover supporting the gas inlet assembly.
6.3.8 Connect test pressure gauge (P/N 4114807) between machine’s oxygen
pipeline inlet connector and the oxygen pipeline supply hose.
6.3.9 Open the O2 cylinder valve and set the System Power switch to ON.
6.3.12 As the pressure drops, the O2 SUPPLY alarm should activate when the
pressure is between 16 and 24 psi as shown on the test gauge.
6.3.13 If the alarm activates when the pressure is below 16 psi or above 24
psi, turn the adjustment set screw (see Figure 6-3. ); repeat the test
and adjust as necessary to bring the set point into the correct range.
6-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
O2
SUPPLY
SWITCH
ADJUSTMENT
SCREW
O2 PIPELINE
HOSE
TEST GAUGE
6-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
6.4.2 Remove the two screws at the back of the machine securing the flow
meter bezel assembly. Carefully pull the assembly out of the machine
for access to the SORC.
6.4.3 Connect a test pressure gauge to the O2 port on the SORC shown in
Figure 6-4.
6.4.8 The test pressure gauge should read 25 ±1 cmH2O. If needed, turn the
O2 control pressure adjustment screw until the test gauge reads 25 ±1
mbar. Apply a small amount of thread sealant to the screw head.
6.4.12 Tighten the lock nut and verify the N2O flow is 50 ± 30 mL/min. Apply
a small amount of thread sealant where the lock nut meets the threads
of the adjustment point.
6.4.13 With the N2O flow control valve fully open, set the oxygen flow to 1 L/
min.
6-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
O2 CONTROL
PRESSURE
ADJUSTMENT
SORC
FLOWMETER BEZEL
ASSEMBLY
SV77011
6-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
If needed, turn the N2O flow adjustment screw until the flow is 3 ±0.3
mL/min, resulting in an O2 concentration of 25 ± 3 vol. % O2 at the
fresh gas outlet. Apply a small amount of thread sealant on the
adjustment screw.
O2 Concentration Test:
6.4.15 Verify O2 concentration at the fresh gas outlet, with the O2 flow
settings in the following table:
N2O Flow
O2
O2 Flow Control
Concentration
Valve
Flow Test:
6.4.18 As the O2 flow increases, the N2O flow should also increase, to over 10
L/min. at the maximum.
6.4.19 Slowly close the O2 flow control valve. The N2O flow should decrease
proportionally.
6.4.20 Close the flow control valves and remove all test equipment.
6.4.21 Set the System Power switch to OFF, and reinstall the flow meter bezel
assembly in the machine.
6-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
Zero Calibration:
6.5.3.1 Remove the oxygen sensor capsule from its housing and allow
several minutes for the displayed offset readings to stabilize.
NOTE: The difference between the Cell A and Cell B readings should be no
greater than 8.
6.5.3.2 Press the Store Zero soft key to store the current values as the
new zero calibration.
21% Calibration:
6.5.3.4 Expose the sensor to ambient air only (away from any open
part of the breathing system) and allow it to stabilize for
several minutes.
6.5.3.7 Press the rotary knob to start the 21% O2 calibration. During
calibration the ‘Calibrate O2 Sensor’ softkey shall illuminate.
Following calibration the screen returns to normal.
6-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
SENSOR CAPSULE
SV00492
Service 9:34
O2 Calibration
Current cell A: 66
Current cell B: 70
6-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
Zero Calibration:
6.6.3.1 Remove the pressure sample line and ventilator hose from the
breathing system, and expose the sample line to air.
6.6.3.3 Touch the “Store Zero” softkey to enter the current value as
the calibration zero.
6.6.4 Re-connect the pressure sample line and the ventilator hose to the
breathing system.
Service 9:24
Pressure Calibration
Current Airway Pressure: 709 Current offset value: 716
Current Ventilator Pres.: 760 Current offset value: 732
Zero calibration procedure:
- Remove pressure sample line and ventilator hose from
breathing system - expose to air.
- Let current pressure value stabilize.
- Touch “Store Zero” key to enter the current
value as the calibration zero.
Store Exit
Zero
6-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
6.7.3 Scroll to Calibration and select Fresh Gas Flow to bring up the flow
calibration screen. See Figure 6-7.
Service 9:12
Freshgas Calibration
Stored Zero: PASS
Zero calibration procedure:
- Close all flow control valves.
- Verify no fresh gas flow is present.
- Touch “Store Zero” key.
- At completion of calibration
“Stored Zero” shall indicate
Pass/Fail.
Store Exit
Zero
6-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
6.8.3 Scroll to Calibration and select PEEP to bring up the PEEP calibration
screen.
6.8.4 Attach test hose (P/N 9995112, 22 mm x 12 in long) from the expiratory
valve terminal to the inspiratory valve terminal and remove breathing
bag.
6.8.5 Touch the Calibrate softkey. (The calibration process can take several
minutes.)
6.8.8 Remove the short hose from the inspiratory and expiratory valves.
6-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
INSPIRATORY
VALVE TERMINAL
TUBE, CORRUGATED
22mm x 12 in. LONG
P/N 9995112
EXPIRATORY
SV00494
VALVE TERMINAL
Service 10:06
PEEP Calibration
Calibration: PASS
PEEP calibration procedure:
- Attach breathing hose from expiratory
valve to inspiratory valve and remove breathing bag.
- Touch “Calibrate” key.
- At completion of calibration
“Calibration” shall indicate
Pass/Fail. PWM, Pres: 170 44
Target Pres: 14
Calibrate Exit
6-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
6.9.1 Remove the back panel from the controller housing. Provide support
for the back panel to prevent damage to the power supply cable
connections.
6.9.2 Set the Main Power switch to ON and set vent control to Volume mode.
6.9.3 Remove the APL bypass hose from the breathing system and attach it
to a digital manometer.
6.9.5 Disconnect the APL bypass hose from the digital manometer and
reconnect it to the APL port on the breathing system.
6-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued) FABIUS GS
NOTE: The ability to calibrate the pump is dependant upon the current PCB
installed in the Core Unit.
6.10.1.1 Access the ‘System Service Screen’ screen and note the
installed ‘Hardware Part Number’. Refer to Section 4.0.
6.10.1.3 Access the ‘Real Time Values’ screen, Refer to Section 4.0.
6.10.1.4 Press the ‘PUMP CAL’ softkey to activate the pump calibration
operation. The ‘PUMP CAL’ label will be highlighted in reverse
upon activation and will remain in this condition until the
calibration operation is either complete or aborted.
6-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE
NOTE: Verify the dates on test equipment calibration labels. DO NOT USE any test
equipment having an expired calibration date. Notify your supervisor
immediately if any equipment is found to be out of calibration.
In the space provided at the bottom of the PMS checklist form, record the Model
and ID number of all calibrated test equipment used. Also record the calibration
due dates. Examples are: multimeter, digital pressure meter, Riken gas analyzer,
safety analyzer, volumeter, trace gas analyzer, simulators.
Description Page
Test Equipment Required: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Fabius GS Parts Replacement Schedule: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Periodic Manufacturer’s Certification General Instructions . . . . . . . . . . . . . . . . . . 7-11
7.1 Electrical Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
7.2 System Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
7.3 Battery Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
7.4 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
7.5 Service Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
7.6 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
7.7 Site Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
7.8 Scavenger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
7.9 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-26
7.10 Vapor Interlock System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-40
7.11 Yokes & Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-41
7.12 Gas Inlet Regulator Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-42
7.13 Cylinder Regulator & Pipeline Gauges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-46
7.14 High Pressure Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-51
7.15 Oxygen Supply Failure Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-53
7.16 Flowmeters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-55
7.17 Oxygen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-57
7.18 Oxygen Concentrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-59
7.19 SORC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-60
7.20 Pressure Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-61
7.21 Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-63
7.22 Volume Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-70
7.23 Audio Silence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-72
7.24 Oxygen Flush Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-72
7.25 Final Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-74
7-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
Alternate
Quantity Description Part Number
Part Number
As Required
1 Flow Sensor (5 pack) 8403735
1 O2 Sensor 6850645
2 O-ring, Vent Hose 2M08777
2 Valve Disc 2123249 or
M23225
1 Hose asm, PEEP/Pmax - APL Bypass 4117027
1 Hose, Pressure 1190520
1 Filter, AGS M33294
Annually: Fabius GS/Tiro 1 year kit 4117360-001 4199912
2 Filter, Pressure & Pneumatic asm 8402868
1 Diaphragm, Patient 2600650
4 O-ring, Vapor 4115864
1 O-ring, for Diaphragm (piston) 8604831
Every 3 Years: Fabius GS/Tiro 3 year kit 4117360-002 4199911
1 Diaphragm, Piston 2600651
2 Battery, 12V Rechargeable 4114229
1 Canister Assembly M29320
1 Lip Seal M30455
1 Packing Ring M30456
1 PM Kit; 1 Year 4117360-001 4199912
7-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
4113119
FRESH GAS LEAK TEST DEVICE
CONNECT TO
TEST PRESSURE GAUGE
TEST
CONNECTOR
4117361
SV00517
7-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
8401892
SIEMENS TEST LUNG
SV00025
7-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
TEST TERMINAL
S010159
BREATHING SYSTEM LEAK TEST DEVICE
TEST TERMINAL
4104389
TEST TERMINAL
ADAPTER
SV00515
7-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
SV00559
S000081
FLOW METER
TEST STAND
0-250cc/MIN
7-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
SV00552
15
Liter
900 100
10 5
800 200
vol/min 2212300
ml MINUTE
VOLUMETER
700 300
600 400
500
TEST
VOLUME
METER
7-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
M33972
TEST CAP
SV00516
7-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
SV00376
4114830-002 4114830-001
4114830-004 4114830-003
4114830-006 4114830-005
7-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
The purpose of these procedures is to provide detailed instructions for performing a Periodic
Manufacturer’s Certification (PMC) inspection on the Fabius GS anesthesia machines.
Several additional documents have been created to assist the technician through the
process. Following is a brief description of the purpose of each document.
The first list in the Recommendation Guidelines is a reference chart for machine
certification based on equipment status. The second is an abbreviated summary of all NAD
Recommendations and Failure Codes including the Condition Number, Equipment
Condition, Recommended Corrections, Certification Code, and Tests Affected when
applicable.
There is also a matrix classified as “Failure Codes” which identifies the correct manner in
which to document equipment tests that fail, or were unable to be performed due to
circumstances beyond the control of the service technician performing the inspection. (Ex:
Air cylinder supply is unavailable to perform Air High Pressure Leak test.) The Failure
Codes section also indicates suggested resolution of the situation. Failure Code numbers
begin at 34 and use the same certification levels strategy, and carry the same weight as
NAD Recommendation equipment condition codes.
The next section of the guideline lists all NAD Recommendations identified at a machine’s
major assembly level. This section is divided into subsections titled: “Anesthesia System”,
“Vaporizers”, “Absorber System”, “Ventilator” and “Scavenger System”. The final matrix is
the most comprehensive index sorted by machine model and includes Equipment Condition,
Certification Code, and NAD Recommendations. It also specifies any suggested upgrade
path including ordering information that should be taken such as installing a Bellows with
Pressure Limit Control 4109664-S01 Kit, after market modification kit to a machine not
equipped with pressure limit control.
7-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
The letters A, B, C, D and the Roman Numerals I, II are used as codes in the individual
matrix for each model of anesthesia machine. The letters A, B, C, and D are used in
descending order to indicate the certification level of the equipment. They are as follows:
A = Certified
B = Certified with Recommendations
C = Conditionally Certified
D = No Certification
Roman Numerals I and II do not affect the certification level but rather are provided to give
further instructions to the end user as follows:
I= The system in its present configuration shall only be used with a CO2 monitor
incorporating an apnea warning. The operator of the system is advised to
frequently scan the CO2 readings and alarm thresholds.
II = The present configuration of equipment requires that the unit operate at all times
with an oxygen analyzer that includes a low oxygen warning. The operator of the
system is advised to frequently scan the oxygen readings and alarm limits.
When multiple recommendations apply, use the above list in descending order from bottom
to top. For example, “No Certification would take precedence over “Conditionally Certified”
and “Certified with Recommendations”. “Conditionally Certified” would take precedence
over “Certified with Recommendations”.
For example:
7-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
A Narkomed 3 could have recommendation number 21 and failure code 61.1 apply.
21 - No ventilator pressure limit control. Code is B.
61.1 - Enflurane agent is unavailable to test. Code is BC.
Correct certification for this machine is BC, which means CONDITIONALLY CERTIFIED
WITH RECOMMENDATIONS.
A Narkomed 6000 could have no NAD recommendations or failure codes apply. The correct
certification level for this machine is Code A, “CERTIFIED”.
Code D, which means “NO CERTIFICATION”, also means the machine shall not
receive a Periodic Manufacturer’s Certification label. The machine shall also
receive a “WARNING - This System is Not Certified” label, P/N 4114857. This label
shall be placed at a prominent location on the right side of the machine after all
other previous PM and “Vigilance Audit® Validation” labels have been removed.
7-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
1. Use the PM Certification form P/N 4117715 for the Fabius GS Anesthesia
System.
3. Perform the vapor concentration test on all Dräger vapor vaporizers every six
months in accordance with SP00073 at a six month maximum interval. Perform
the vaporizer concentration test on all Desflurane vaporizers in accordance with
SP00091 for fixed mount vaporizers and SP00189 for user removable D-tec
vaporizers at a six month maximum interval. For every vaporizer tested, fill out
a “VAPOR VAPORIZER CALIBRATION CHECK” label (part # S010016).
Information on this label shall include your signature, type of agent, date
tested, test results @ 1%, 2.5%, 4% for H, E, I, or S vaporizers, or @ 4%, 6%, 10%,
12%, 16% for Desflurane vaporizers, and a PASS or FAIL indication. This label
shall be attached to the upper right side of the vaporizer. If vaporizer fails the
concentration verification, internal leak, or exclusion system tests, check “NO”
in the “RECOMMENDED FOR USE” section on the PM Certification form.
Place a “CAUTION DO NOT USE” label (part # 4114327) on the vaporizer, and
issue a departmental alert. The TSR shall also seek permission from the
equipment operator to remove the failed vaporizer from the machine and install
a replacement vaporizer or an adapter block onto the mount.All nonfunctional
Dräger vaporizers must be removed from service for machine to receive
certification.
7-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
5. Based on the “EQUIPMENT CONDITION” inspect the machine for any “NAD
RECOMMENDATIONS” that would apply. Use the Fabius GS section of the
“NAD RECOMMENDATION GUIDELINES INDEX”, P/N S010250. Note all
applicable NAD recommendations on the Executive Summary. NOTE: If using a
carbon form, indicate the Equipment Condition number and to see reverse side
under “RECOMMENDATIONS / GENERAL COMMENTS” section of the form.
6. Determine the correct certification level of the machine based on the combined
lowest common denominator of “Equipment Conditions” and “Failure Codes”. If
the machine is at least conditionally certified, fill out the “PM
CERTIFICATION” label. Check the box(s) on the validation label where
appropriate. Write the month and year, (six months from date of PM
Certification) next to “NEXT VISIT DUE:” If certification level is “D”, machine
shall not receive a “PM CERTIFICATION” label. Any machine not receiving a
PM Certification label shall receive a “WARNING NOT CERTIFIED” label. This
label shall be placed at a prominent location on the left side of the machine after
all other previous PMS and Vigilance Audit Validation labels have been
removed.
10. Have the customer sign each PM Certification form or the Executive Summary,
and review the equipment conditions and recommendations with the customer.
11. Return the top copy to Draeger Medical, Inc. Service Department, keep middle
copy for service organization records, give bottom copy to customer.
7-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
NOTE: The following procedure will require making adjustments and calibrations
with the possibility of removal of accessories and/or additional monitoring
equipment. Therefore, take note of these devices for reconfiguration and/or
reinstallation after the PMS procedure is complete.
Plug the unit into the safety analyzer, and plug the power cord of
the safety analyzer into an AC receptacle. Turn analyzer to ON
position.
NOTE: The BIOTECH 501 PRO will automatically test the source outlet
for open ground (or ground resistance of 31 Ω or higher), reverse
polarity, open neutral and open line. (The latter two conditions
will prevent the analyzer from powering up.)
7.1.1.2 Set the safety analyzer function switch to the GROUND WIRE
RESISTANCE position. Attach the test lead to the SINGLE
LEAD connector of the analyzer. Connect the other end of the red
test lead to the ground socket of the front panel outlet on the
safety analyzer. Verify a displayed resistance of 0.000 Ω or, if
necessary, press the CALIBRATE key on the front panel of the
analyzer to zero the device.
7.1.1.3 Set the safety analyzer GROUND switch to NORMAL. Set the
POLARITY switch to OFF.
7.1.1.4 The safety analyzer shall indicate 0.1 Ω or less with its test lead
applied to the following points:
7-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
NOTE: This test is not applicable on machines with serial numbers 12191 to
12334.
7.1.2.1 With a multimeter set to its highest resistance range, check for
continuity between the serial port cable retaining nut on
processor and the O2 sensor connector pin shown in the
illustration. There shall be no continuity between these points.
O2 SENSOR
NOTE: This test will check the auxiliary strip outlets for fault conditions
such as open ground (>31 Ω), reverse polarity, open line and open
neutral. This is done each time the BIOTECH 501 PRO is
powered up and allowed to cycle through its self test.
7.1.3.1 Shut off and unplug the safety analyzer. Remove the anesthesia
machine plug from the analyzer, and plug it into the same outlet
that was being used by the analyzer.
7.1.3.2 Plug the safety analyzer into the first outlet to be tested, and
turn the analyzer power switch ON. Allow the analyzer to cycle
through its Auto Test sequence. If no wiring fault is indicated,
shut off the analyzer and move its plug to the next outlet. Test
this outlet in the same manner, and continue until all
convenience outlets on the auxiliary strip outlets are tested.
7.1.4.2 Attach the safety analyzer test lead to a rear GCX rail.
7-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.1.4.3 Record the total leakage current with the Polarity and Ground
switches set as follows:
Ground Polarity
Open Normal
Normal Normal
Open Reversed
Normal Reversed
Fabius GS SW 1.03
XX CRC 6DBA
(9) 7.2.4 Record the machine software version on the header of the checklist
form.
7-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.3.1 Press the Man Spont key, and press the rotary dial to confirm.
7.3.2 Unplug the AC power cord and verify the “Power Fail” message
appears on the display within one minute.
7.3.3 Restore AC power to the machine and verify the “Power Fail” message
disappears within one minute.
7.4.1 Press the Standby key, then press the rotary key to confirm.
7.4.2 Press the Setup key, and press the rotary dial to confirm.
Default Settings:
7.4.3 Verify all parameters listed in the ‘Default Settings’ screen are
adjustable.
7.4.4 Adjust the Alarm Volume to its highest setting and press the rotary
dial to confirm.
7.4.5 Press the rotary dial once to return to the Setup screen.
Configuration:
7.4.6 Using the rotary dial, scroll to Configuration and press the rotary dial
to confirm.
7.4.7 Verify all parameters listed in the ‘Configurations Settings’ screen are
adjustable and correct per local configurations. Press to confirm two
times to exit Setup mode.
7.5.1 Press and hold the Home and Standby keys, and press the rotary dial.
Verify the System Service screen appears.
(9) 7.5.2 Verify on-screen serial number with machine serial number located on
rear of unit, and record on header of PMS form.
(9) 7.5.3 Record the Last Service Date, Hours Run Since Last Service, Total
Hours Run, and Total Ventilator Hours Run on the PMS form.
7-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.5.7 Inspect the condition of the ventilator (piston diaphragm) O-ring. Is
the O-ring in good condition? (Y)
(9) 7.5.8 If applicable, install the appropriate PM kit(s) and record in Parts
Replacement section of PM form. See Parts Replacement Schedule on
page 3.
7.6 Calibrations
7.6.3 Follow on-screen instructions and perform Fresh Gas Flow calibration.
Pressure:
O2 Offset:
(9) 7.6.9 Follow the on-screen instructions and perform the O2 zero calibration.
Reassemble O2 housing and install the inspiratory dome plug in
inspiratory dome.
PEEP:
7-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
(9) 7.7.2 Scroll to Flowmeter and confirm. Scroll to O2 Position and confirm.
Select position of virtual flowmeter (left or right) in accordance with
local requirements or customer demand. Press to confirm. record
position on PMS form.
Gas Selection:
(9) 7.7.3 Press to confirm, scroll to Gas Selection and confirm. Verify Gas
Selection type (2 or 3) is in accordance with machine configuration.
Press to confirm (x2). Record selection type on PMS form.
Flowtube Resolution:
7.7.4 Scroll to Flowtube Res. and press to confirm. Scroll to change state and
confirm. Scroll to change state in accordance with customer demand.
O2 Whistle:
Alarms:
(9) 7.7.6 Scroll to No Fresh Gas and confirm. Verify state of No Fresh Gas alarm
is enabled. Press to confirm (x2). Record position on PMS form.
(9) 7.7.7 Scroll to Fresh Gas Low Alarm. Verify Fresh Gas Low Alarm position
(enabled or disabled) is set to enabled position. Press to confirm (x2).
Record position on PMS form.
(9) 7.7.8 Scroll to Threshold Low and confirm. Verify state of Threshold Low
Alarm is enabled. Press to confirm (x2). Record position on PMS form.
Pressure:
(9) 7.7.9 Scroll to Ambient Pressure and confirm. Turn rotary dial to adjust.
Press to confirm (x2). Record pressure setting on PMS form.
NOTE: Using the table found on the following page, set the ambient pressure
(mbar) in accordance with the local elevation. To verify the local
elevation, contact the nearest regional airport, or contact:
DMI - Technical Support - 1-800-543-5047
DMT - Tech Line - 49-451-882-4222
7.7.10 Scroll to Plateau/Mean Dis and verify setting is in accordance with
customer demand. Press to confirm (x3).
7-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
NOTE: Machines from the factory are configured with the appropriate model
type. Other than performing a software download or PCB replacement,
the model type should not be changed. Changing the model type will
disable some ventilator options and will require reconfiguration.
Serial Ports:
(9) 7.7.12 Scroll to Serial Ports and confirm. Verify parameters are adjustable to
the following protocols. If necessary, set the protocols for any third
party monitoring device connected to the machine. Refer to the Fabius
GS Operator’s Manual and third party Operator’s Manual.
7-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
Scavenger Cleaning
7.8.1.1 Remove all scavenger hoses one at a time and drain all
accumulated moisture. Inspect all scavenger hoses for
deterioration and replace any worn hoses.
7.8.1.3 Remove the reservoir canister from the scavenger by partial turn
counter-clockwise of the canister.
7.8.1.7 Verify vacuum hose is connected to one of the two gas supply
connections.
7.8.1.9 Verify all hose connections to the scavenger are complete at all
destinations.
7.8.1.11 Reinstall canister and adjust scavenger flow control valve; verify
float moves freely between the upper and lower limits.
7.8.2.1 Inspect all scavenger hoses for signs of wear and deterioration.
Replace any worn hoses.
7-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.8.2.5 Place the filter back, making sure that it lays flat on the valve
port orifice.
7.8.2.11 Open the oxygen flow control valve to a flow of 10 L/min and
occlude the passive scavenger exhaust port connection.
7.8.2.12 After the breathing bag inflates, the absorber systems’ breathing
pressure gauge must indicate a pressure of less than 5 cm H2O.
7.8.2.13 Systems that are not equipped with a pressure gauge, observe
the generated pressure waveform, set the threshold to 5 cm H2O.
The pressure waveform shall not rise and remain above the
dotted threshold trace during this performance test.
7-25
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.9.1.1 Record the serial number (located on the side of the COSY) on
the PMS form.
7.9.1.7 Are the valve dome washers in good condition? ___ (Y)
7.9.1.10 Remove the flow sensor from the breathing system housing. Is
the O-ring inside housing in good condition? ___(Y)
7.9.1.12 Remove the canister and inspect the canister, gaskets, and
condition of soda lime. Are these components in good condition?
___(Y)
7.9.1.13 Remove the breathing pressure gauge from its mount and
inspect the O-rings in the mount and on the 90° hose connection -
if applicable.
7.9.1.14 Is the breathing pressure gauge in good condition and set at zero
(0)? ___ (Y) (if applicable)
7.9.1.15 Inspect the O-rings on the inspiratory dome plug assembly and
O2 sensor housing. Are the O-rings in good condition? ___(Y)
7-26
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.9.1.17 Inspect the breathing circuit and manual breathing bag. Is the
breathing circuit and manual breathing bag in good condition?
___(Y)
7.9.1.19 Inspect the fresh gas hose connected to the breathing system. Is
the fresh gas connector and washer in good condition? ___(Y)
7.9.1.20 Inspect APL valve labeling for legibility. Are all markings on
APL valve easy to see and legible? ___(Y)
7.9.2.2 Remove the fresh gas connector from the breathing system.
7.9.2.3 Connect the fresh gas hose from the breathing system to the
fresh gas leak test fixture (4113119 modified) or equivalent, via
the fresh gas test connector 4117361.
7.9.2.4 Connect a digital pressure manometer to the fresh gas leak test
device.
(9) 7.9.2.6 After thirty (30) seconds, what is the pressure on the
manometer? ___ (>40 cm H2O)
(9) 7.9.2.9 After thirty (30) seconds, what is the pressure on the
manometer? ___ (>40 cm H2O)
7-27
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.9.2.11 If applicable, turn on the right mounted vaporizer to the first
graduated marking; repeat Steps 7.9.2.5 thru 7.9.2.10 ___ (>40
cm H2O)
7.9.2.12 Remove the test equipment from the fresh gas hose.
7.9.2.14 Open the O2 flow control valve to 5 L/min., purge the system for
5 seconds, then close the O2 flow control valve.
y
7.9.2.15 Push the System Power switch OFF ( { ).
7.9.2.16 Reconnect the fresh gas connector from the machine to the
freshgas outlet connector on the breathing system.
7-28
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
NOTE: Use the appropriate column in the following table - according to the
software version of the machine you are testing:
7-29
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
SV00549a
SET APL VALVE TO MAN, 70 cmH2O
TEST TERMINAL
4104389
7-30
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7-31
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.9.4.3 Verify the fresh gas hose is connected to the breathing system.
7.9.4.4 Verify the APL valve is in the MAN position and set to 10 cmH2O.
7.9.4.5 Activate the flush to inflate the reservoir bag, then release.
(✔) 7.9.4.6 Adjust the O2 and Air flows to 10 L/min. (20 L/min. total). Verify pressure
remains between 8 and 12 cmH2O.
(✔) 7.9.4.7 Repeat Steps 7.9.4.4 thru 7.9.4.8 Repeat Steps 7.9.4.4 thru 7.9.4.8 with the
with the following settings, and following settings and verify APL Valve
verify APL valve accuracy: accuracy:
30 27 33 40 34 46
50 45 55
7-32
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
Inhalation:
SV00553
TEST TERMINAL
4104389
TEST CONNECTOR
4117361
FRESH
GAS HOSE
TEST
TERMINAL
4104389
(INSTALL ON FROM MACHINE
VENTILATOR FRESH GAS LEAK
PORT) TEST DEVICE
(MODIFIED)
0-250 cc
TEST FLOWMETER
7-33
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.9.5.3 Remove the freshgas hose from the breathing system and
connect it to the freshgas leak connector (4117163).
7.9.5.4 Connect the fresh gas leak connector to the fresh gas leak test
device (4113119 modified) as shown.
7-34
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
Exhalation:
7.9.5.8 Occlude ventilator and inspiratory port, each with test cap
M33972.
7.9.5.10 Connect Fresh gas hose from machine to fresh gas port of
breathing system.
7.9.5.13 Remove all test equipment except for test cap on inspiratory
port.
SV00563
TEST CAP
M33972
OCCLUDE
VENTILATOR PORT TEST TERMINAL
WITH PLUG 4104389
0-250 cc
TEST FLOWMETER
7-35
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.9.6.1 Verify the breathing bag port, fresh gas connection and
expiration port connections are open and not occluded.
7.9.6.3 Verify occlusion of the inspiration port with test cap M33972.
7.9.6.4 Connect the fresh gas hose to the ventilator port on the COSY
with the test connector as shown.
SV00554 TO BREATHING
PRESSURE
CONNECTOR
TO BREATHING
ON COSY
PRESSURE
INTERFACE
CONNECTOR
ON MACHINE
TEST CAP
M33972
0-250 cc
TEST FLOWMETER
TEST CONNECTOR
4117361
TEST TERMINAL
4104389 FRESH GAS
(INSTALL ON HOSE
VENTILATOR PORT)
7-36
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.9.6.5 Verify the APL valve is in the MAN position and set to 70
cmH2O.
(9) 7.9.6.7 Verify the flow on the 0-250 cc test flowmeter indicates less than
10 mL/min. (0.01 L/min.).
7.9.6.9 Remove the occlusion plugs and test equipment from the
breathing system.
7-37
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.9.7.1 Disconnect the APL and PEEP control hoses from the interface
panel at rear of machine.
7.9.7.2 Connect the PEEP control hose to the input of the 0-250 cc test
flowmeter.
SV00562
0-250 cc
TEST FLOWMETER
APL BYPASS
VALVE HOSE APL BYPASS VALVE
(OPEN)
TEST
TERMINAL
4104389
PEEP 12 IN.
VALVE TEST HOSE
HOSE 9995112
PEEP/PMAX VALVE
7-38
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.9.7.3 Short circuit the inspiratory and expiratory ports, attach a test
terminal to the breathing bag port and attach a digital
manometer to the test terminal.
7.9.7.6 Remove the 0-250 cc test flowmeter, digital manometer and short
circuit hose from machine.
7-39
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.10.2 Adjust the handwheel on the left vapor to any concentration above zero
(0).
7.10.5 Adjust the handwheel on the right vapor to any concentration above
zero (0).
7-40
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.11.1.2 Disconnect all pipeline hoses and close all cylinder valves.
7.11.1.5 Are the two (2) yoke pins installed securely in each yoke? ___ (Y)
7.11.1.6 Is there only one (1) cylinder washer on each yoke assembly? ___
(Y)
7.11.1.7 Is there a yoke plug attached to each yoke assembly? ___ (Y)
7.11.1.8 Is the proper gas I.D. label affixed to each yoke assembly? ___ (Y)
7.11.1.9 Attach a cylinder to each yoke assembly, open the cylinder valve,
let the pressure stabilize, close the cylinder valve, and remove
the cylinder from the yoke assembly.
7.11.1.10 Does the yoke check valve assembly prevent the escape of
excessive pressure? ___ (Y)
7.11.2.1 Are the pressure gauges correct for the gases indicated by the
flowmeters? ___ (Y)
7.11.2.2 Bleed all pressure from the cylinder circuits using the flow
control valves.
7.11.2.3 Are the cylinder gauges at zero (0) PSI? ___ (Y)
7-41
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.12.1.2 Remove the pneumatic hose from the O2 outlet of the gas inlet
block and interconnect a digital manometer as shown.
ADJUSTMENT
SCREW
O2 INLET
REGULATOR
TO O2
FLOW CONTROL
VALVE
TEST 'T' O2 OUTLET (TO O2
CONNECTOR FLOW CONTROL VALVE)
M32803
7-42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
(9) 7.12.1.5 After the digital manometer stabilizes, what is the regulator
output pressure? ___psi (28 - 32 psi)
7.12.1.7 Close the flow control valve and disconnect the O2 pipeline hose
from the inlet.
7.12.2.1 Disconnect the N2O pneumatic hose from the outlet connector on
the gas inlet block that connects to the N2O flow control valve.
7.12.2.3 Reconnect the O2 and N2O pipeline hoses to the inlet block and
activate the pipeline supplies.
(9) 7.12.2.4 Set the O2 and N2O flows to 4 L/min. After the digital
manometer stabilizes, what is the N2O regulator output
pressure? ___psi (28 - 32 psi)
7.12.2.6 Close the O2 and N2O flow control valves and disconnect both
pipeline hoses from the inlet block.
7.12.2.7 Remove test equiupment and reconnect the N2O hose from the
flow control valve to the N2O outlet on the inlet block.
7-43
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.12.3.2 Detach the hose from the Pressure Test adapter (P/N 4114807)
gauge assembly and attach it to the inlet of the test flowmeter as
shown.
TEST FITTING
4109318
S000081
TEST
FLOWMETER
QUICK DISCONNECT
O2 ADAPTER
SV00584
7-44
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.12.3.5 Attach the appropriate N2O adapter to the N2O connector on the
inlet block, and transfer the end of the hose with the quick
disconnect fitting to the N2O adapter.
(9) 7.12.3.6 Open the N2O (reserve) cylinder. What is the flow as indicated
on the test flowmeter? ___cc ≤5 cc/min.
7.12.3.7 Attach the appropriate Air adapter to the Air connector on the
inlet block, and transfer the end of the hose with the quick
disconnect fitting to the Air adapter.
(9) 7.12.3.8 Open the Air (reserve) cylinder. What is the flow as indicated on
the test flowmeter? ___cc ≤5 cc/min.
7.12.3.9 Remove all test equipment and reinstall the gas inlet block and
back panel on the machine.
7-45
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
Minimum cylinder pressure requirements for this test are: N2O: 600 psi;
O2, Air: 1000 psi.
SV00465
O2
PIPELINE
HOSE
O2 REGULATOR
N2O
REGULATOR ACORN NUT
7-46
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.13.1.2 Connect the test fixture hose to the machine’s nitrous oxide
pipeline inlet.
7.13.1.3 Does the back panel correctly identify the nitrous oxide inlet?
___(Y)
7.13.1.4 Connect the nitrous oxide pipeline supply hose to the test fixture.
7.13.1.5 Open the nitrous oxide and the oxygen cylinder valves.
(9) 7.13.1.8 Release the push button. After the pressure decay stabilizes,
what is the regulator output pressure? ___psi (32 - 40)
NOTE: f a pressure decrease does not occur, either the hospital’s supply
pressure is too low or the regulator pressure is set too high.
7.13.2.1 Close the nitrous oxide cylinder valve and drain all nitrous oxide
pressure.
7.13.2.2 Depress and hold the push button on the test device.
(9) 7.13.2.4 After the pressure stabilizes, are the indicated pressures on the
test gauge and the nitrous oxide pipeline pressure gauge within
the normal operating range? 41 - 87 psi (non-USA), 50 - 55 psi
(USA).
7-47
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.13.3.3 After 30 seconds, what is the pressure loss? ___psi (<5)
7.13.3.4 Remove the test equipment and reconnect the nitrous oxide
pipeline hose.
7.13.4.1 Configure the test gauge using the appropriate adapters between
the pipeline inlet of machine and central supply hose.
7.13.4.2 Connect the test fixture hose to the machine’s air pipeline inlet.
7.13.4.3 Does the back panel correctly identify the air inlet? ___(Y)
7.13.4.4 Connect the air pipeline supply hose to the test fixture.
(9) 7.13.4.7 Release the push button. After the pressure decay stabilizes,
what is the regulator output pressure? ___psi (based upon
cylinder pressure as given in the following table)
Compensated Regulator
Cylinder Pressure (psi)
output tolerances (-4/+2)
* Canada settings
NOTE: If a pressure decrease does not occur, either the hospital’s supply
pressure is too low or the regulator pressure is set too high.
7-48
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.13.5.1 Close the air cylinder valve and drain all air pressure.
7.13.5.4 After the pressure stabilizes, are the indicated pressures on the
test gauge and the air pressure gauge within the normal
operating range? 41 - 87 psi (non-USA), 50 - 55 psi (USA).
(9) 7.13.6.3 After thirty (30) seconds, what is the pressure loss? ___psi (<5)
7.13.6.4 Remove the test equipment and reconnect the air pipeline hose.
7.13.7.2 Connect the test fixture hose to the machine’s oxygen pipeline
inlet.
7.13.7.3 Connect the oxygen pipeline supply hose to the test fixture.
7.13.7.4 Does the back panel correctly identify the oxygen inlet? ___(Y)
7-49
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.13.7.7 Release the push button. After the pressure decay stabilizes,
what is the regulator output pressure? ___psi (based upon
cylinder pressure as given in the following table)
Compensated Regulator
Cylinder Pressure (psi)
output tolerances (-4/+2)
* Canada settings
NOTE: If a pressure decrease does not occur, either the hospital’s supply
pressure is too low or the regulator pressure is set too high.
7.13.8.1 Close the oxygen cylinder valve and drain all oxygen pressure.
7.13.8.4 After the pressure stabilizes, are the indicated pressures in the
test gauge and the oxygen pressure gauge within the normal
operating range? 41 - 87 psi (non-USA), 50 - 55 psi (USA).
(9) 7.13.9.2 After thirty (30) seconds, what is the pressure loss? ___psi (<5)
7-50
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.14.1.4 Close the oxygen cylinder valve and remove the cylinder.
(9) 7.14.1.6 After two (2) minutes, what is the pressure loss? ___ PSI (<50)
7.14.2.2 Open one (1) oxygen cylinder valve and one (1) nitrous oxide
cylinder valve.
7.14.2.5 Close the nitrous oxide cylinder valve and remove the cylinder.
(9) 7.14.2.7 After two (2) minutes, what is the pressure loss? ___ PSI (<50)
7-51
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.14.3.2 Open one (1) oxygen cylinder valve and the air cylinder valve.
7.14.3.5 Close the air cylinder valve and remove the cylinder.
(9) 7.14.3.7 After two (2) minutes, what is the pressure loss? ___ PSI (<50)
7-52
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.15.1.4 Does the flow of nitrous oxide cease when the oxygen flow is
depleted? ___ (Y)
7.15.1.6 Close the nitrous oxide cylinder valve and bleed the pressure
from the circuit.
7.15.1.9 Does the flow of nitrous oxide cease when the oxygen pressure is
depleted? ___ (Y)
7-53
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.15.2.5 What is the pressure on the test gauge when the "O2 SUPPLY
LOW !!!” message and associated red indicator appear? ___ PSI
(16 - 24)
(9) 7.15.2.6 Does the “O2 SUPPLY LOW !!!” message appear on the display?
___ (Y)
7.15.2.7 Bleed the remaining O2 pressure from the system, then close the
flow control valve.
7.15.2.8 Remove the test gauge from the machine and reconnect the
pipeline supply hose.
SV00465
O2
PIPELINE
HOSE
7-54
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.16 Flowmeters
7.16.1.4 Connect the O2 pipeline supply, and verify the operation of the
oxygen flowmeter.
7.16.1.5 Is the correct flow control knob and label attached to the oxygen
flow control valve? ___ (Y)
7.16.2.4 Close the nitrous oxide cylinder valve and bleed the pressure.
7.16.2.5 Connect the N2O pipeline supply, and verify the proper operation
of the N2O flow.
7.16.2.6 Is the correct flow control knob attached to the N2O flow control
valve? ___ (Y)
7.16.2.7 Close the oxygen and nitrous oxide flow control valves.
7-55
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.16.3.3 Close the air flow control valve and disconnect the Air pipeline
supply.
7.16.3.4 Is the correct flow control knob attached to the air flow control
valve? ___ (Y)
7.16.4.4 Slowly open the Aux O2 flow control valve to apply a pressure of
50 cm H2O to the manometer, then close control valve and start
timer.
(9) 7.16.4.5 After 30 seconds, what is the pressure on the manometer? ___(40
- 50 cm H2O)
7.16.4.7 Is it possible to adjust the flow over the full range of the auxiliary
oxygen flowmeter? ___ (Y)
(9) 7.16.4.10 After the value stabilizes, what is the oxygen concentration? ___
% (97-100)
7.16.4.11 Remove the oxygen sensor from the auxiliary oxygen flowmeter,
and insert it into the inspiratory valve dome adapter.
7-56
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.17.1 Press the Man Spont key and press the rotary knob to confirm.
7.17.3 Disconnect the oxygen sensor cable from the Oxygen Sensor interface.
7.17.4 The following message shall appear on the display: O2 Sensor Fail!.
NOTE: Make sure that the sensor has stabilized in ambient air for several
minutes.
7.17.6 Press the Setup key, then perform O2 calibration per on-screen
instructions.
7.17.7 After calibration is completed, verify the “O2 Sensor Fail” message
disappears.
7.17.10 Press the key and select the OXYGEN LOW alarm limit. Does
the low alarm limit illuminate? ___ (Y)
7.17.11 Verify that the low alarm limit has a range from 18 to 99%. Adjust the
alarm limit above current O2 monitor value.
(9) 7.17.12 INSP O2 LOW !!! shall appear on the display and the heading shall be
flashing with a corresponding audible alarm.
7.17.13 Place the oxygen sensor into the inspiratory valve dome adapter and
set the APL valve to 70 and MAN position. Attach a 12-inch hose to the
inspiratory port and occlude the bag mount.
7.17.15 Verify that the INSP O2 LOW !!! message has cleared.
7.17.16 Select the OXYGEN HIGH alarm limit. Does the high alarm limit
illuminate? ___ (Y)
7-57
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.17.18 Verify that the high alarm limit has a range from 100 to 19%.
7.17.19 Set the high alarm limit to below the correct O2 reading.
(9) 7.17.20 Does the message INSP O2 HIGH !! appear on the display? ___ (Y)
(9) 7.17.23 Within 3 minutes, what is the oxygen concentration? ___ % (97-100)
7-58
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.18.1.3 Does the oxygen monitor read 97-100% after the value
stabilizes?__(Y)
(9) 7.18.1.5 After the value stabilizes on the oxygen sensor, what is the O2
concentration? ___ % (62-72)
7.18.1.6 If measured value is within the range, close the N2O flow control
valve and proceed to Section 7.18.2.
(9) 7.18.1.7 If measured value is not within range, close the nitrous oxide
flow control valve and adjust the oxygen flow until total flow
meter reads 4 L/min. What is the corresponding display flow
rate? ___ L/min (3.9 - 4.7)
(9) 7.18.1.8 Set nitrous oxide flow to 2.5 L/min and oxygen flow to 2.5 L/min
on the displays. What is the corresponding flow visually
approximated on the total flow meter? ___L/min (4.5 - 5.5)
7.18.2.2 Does the oxygen monitor read 97-100% after the value
stabilizes?__(Y)
7-59
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.19 SORC
(9) 7.19.4 What is the oxygen concentration after the value stabilizes? ___ %
(≥23%)
(9) 7.19.6 What is the oxygen concentration after the value stabilizes? ___ %
(≥23%)
(9) 7.19.8 What is the oxygen concentration after the value stabilizes? ___ %
(≥23%)
(9) 7.19.10 What is the oxygen concentration after the value stabilizes? ___ %
(≥23%)
(9) 7.19.11 Reduce the O2 flow to 0.5 L/min. Verify that the N2O flow is greater
than or equal to 0.6 L/min. ___ L/min.
7.19.12 Reduce O2 flow to zero, then slowly increase O2 flow until N2O flow is
present.
(9) 7.19.13 What is O2 flow when N2O starts to flow (0.1 - 0.2 L/min.)
7-60
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.20.2 Disconnect the breathing pressure sensor line from the breathing
system.
7.20.3 Connect a test pressure gauge and syringe to the breathing pressure
sensor line.
7.20.5 What is the pressure apnea threshold default? ___ cm H2O (8)
7.20.6 Verify that the pressure apnea threshold limit has a range from 5 to 30
cm H2O.
7.20.9 What is the high alarm limit default? ___ cm H2O (40)
7.20.10 Verify that the high alarm limit has a range from 10 to 70 cm H2O.
7.20.11 Set the high alarm limit to 40 cm H2O, and confirm. Set the ventilator
to Volume Control mode and confirm.
(9) 7.20.13 What is the time when APNEA PRESSURE !! appears as a medium
alarm tone with associated flashing yellow LED? ___ sec (26-34)
(9) 7.20.14 What is the time when the APNEA PRESSURE !!! appears as a high
alarm with flashing red LED and corresponding audible alarms? ___
sec (56-64)
NOTE: For software version >1.20, with threshold low alarm enabled a
corresponding threshold low alarm shall appear.
7.20.15 After the APNEA PRESSURE !!! alarm is displayed as a high, slowly
increase the test pressure.
7-61
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.20.16 At what pressure does the APNEA PRESSURE alarm deactivate? ___
cm H2O (7-13)
7.20.18 Increase the pressure to 20 cm H2O, maintain the pressure, and start a
stopwatch.
(9) 7.20.19 What is the time when CONTINUOUS PRESSURE appears as a high
alarm? ___ sec (12-18)
(9) 7.20.20 Decreasing the pressure slowly, what is the pressure at which the
CONTINUOUS PRESSURE alarm deactivates? ___ cm H2O (15-21)
(9) 7.20.22 At what pressure does the Airway Pressure High !!! alarm activate? ___
cm H2O (38-42)
7.20.24 Using a syringe in place of the squeeze bulb, slowly create a sub-
atmospheric pressure.
(9) 7.20.25 At what pressure does the Pressure Negative !!! alarm activate? ___ cm
H2O (-2 to -8)
7.20.28 Using the syringe, increase the pressure to 20 cmH2O; then decrease
the pressure to zero.
7.20.30 Does the threshold limit adjust to within four (4) cmH2O of peak
pressure? ___ (Y)
7.20.32 Reconnect the breathing pressure sensor line to the breathing system.
7-62
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.21 Ventilator
7.21.2.2 Verify that spontaneous breathing is possible with the test lung.
___ (Y)
7.21.3.3 Press the Calibrate Flow Sensor key and perform the on-screen
instructions. At completion of calibration verify “Flow
Calibration Completed” message appears.
7-63
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.21.4.2 Switch to Volume Control mode and set the fresh gas flow to 3 L/
min.
(9) 7.21.4.12 Verify Minute Volume on test volumeter and MV on display are
within 20% of each other.
(9) 7.21.5.2 Verify displayed values for the following PEEP settings:
0 0 + 2 cmH2O
10 10 ± 2 cmH2O or 20% of setting, which ever is greater
15 15 ± 2 cmH2O or 20% of setting, which ever is greater
7-64
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.21.6.1 Connect a test pressure gauge in line with the pressure port
connector of breathing system.
(9) 7.21.6.3 Verify test gauge display and breathing gauge values for the
following Pmax settings:
25 25 ± 5 cmH2O
40 40 ± 5 cmH2O
60 60 ± 5 cmH2O
70 70 ± 5 cmH2O
7.21.7.2 Set O2 and Air Fresh Gas flows to 10L/min ea. (20 L/min. total)
7.21.7.3 Set APL valve to MAN, Set APL Valve to 40 cm H2O
30 cmH2O.
7.21.7.8 Reconnect test lung to breathing circuit and remove the test
pressure gauge from the breathing system.
7-65
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7-66
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
(9) 7.21.10 Verify relief valve opens between 70 and 80 mbar (cmH2O) when
ventilator is in inspiratory phase.
SV00561
TEST TERMINAL
7-67
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
(9) 7.21.11.1 Close all flow control valves. Verify vacuum valve opens at the
pressure indicated in the following table, when ventilator is in
expiratory phase.
-3 mbar -8 mbar
Software Version
neg relief valve neg relief valve
NOTE: For software version ≤ 1.20, proceed to the next step. For software
version >1.3x, skip to Part (b) of Step 7.21.12.5
7.21.12.2 Connect test setup as shown in the illustration on the next page.
NOTE: Do not run the leak test per on-screen instructions. Continue with
piston chamber leak test as follows.
30 20
(9) 7.21.12.5 (a) Verify leakage rate is <20 mL/min. as shown on display.
(b) For software version 1.3x, perform the Leak/Compl Test and
verify ventilator leak test indicates a Pass condition.
7-68
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
AIRWAY
PRESSURE
CONNECTION
TEST
TERMINAL
FRESH GAS
HOSE
FRESH GAS
TEST
CONNECTOR
4117361
SV00548
FRESH GAS LEAK TEST
DEVICE 4113119 (MODIFIED)
7.21.13.2 Remove APL bypass hose from breathing system and attach to
digital manometer.
(9) 7.21.13.3 Verify vacuum is between -150 and -240 cmH2O (mbar) (-110 and
-175 mmHg) as shown on digital manometer. ___ cmH2O
7-69
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.22.3 Press the Alarms key, scroll to MV High Alarm and confirm. Does a box
appear around the Minute Volume High Alarm Limit? ___(Y)
7.22.4 What is the High Minute Volume Alarm default? ___ (12) Press to
confirm.
7.22.5 Scroll to Minute Volume Low Alarm and confirm. What is the minute
Volume Low default? (3.0)
(9) 7.22.6 Adjust the low minute volume alarm to 5.0 liters and confirm. Does the
MINUTE VOLUME LOW !! alarm appear as a mid alarm with a
corresponding yellow LED? ___ (Y).
7.22.7 Adjust minute volume alarm to 2.0 and confirm. Does the Minute
Volume Low alarm disappear? ___ (Y).
7.22.8 Disconnect breathing hose from expiratory connection and start a stop
watch.
(9) 7.22.9 What is the time when APNEA FLOW !! appears as a mid alarm? ___
sec (13 - 17)
(9) 7.22.10 What is the time when APNEA FLOW !!! appears as a high alarm? ___
sec (26 - 34)
7.22.11 Disconnect the respiratory volume sensor cord from the Flow sensor.
(9) 7.22.12 Does the “FLOW SENSOR FAIL !” message appear as an Advisory
with a single audible tone? ___ (Y)
7.22.14 Do the numerical values associated with volume disappear? ___ (Y)
7.22.15 Reconnect the respiratory volume sensor cord to the flow sensor, and
expiratory limb of breathing circuit to breathing system.
7-70
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
(9) 7.22.17.3 After 30 seconds verify the APNEA PRESSURE !!!, APNEA
FLOW !!!, MINUTE VOLUME LOW !!! and FRESH GAS LOW !!
alarm messages appear on the display with a corresponding
audible alarm.
7.22.17.5 Verify the FRESH GAS LOW !! alarm disappears after two (2)
seconds following the next ventilator cycle. ___ (Y)
7.22.17.6 For software version 1.3n, with No Fresh Gas enabled, verify a
No Fresh Gas !!! alarm appears.
7-71
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.23.2 Does the LED on the Silence Alarms key light? Is the audio alarm
silenced for 120 sec.?
7.23.4 Cycle the system power and wait for the Standby screen to appear.
7.24.5 Insert the sensor from a calibrated O2 Med into the valve dome adapter
on the inspiratory valve.
(9) 7.24.8 What is the O2 concentration after the value stabilizes?___%O2 (97-
100)
7.24.9 Remove the O2 sensor from the inspiratory dome and install the plug.
7.24.10 Disconnect the fresh gas hose from the breathing system. Using the
fresh gas test connector (4117361), test terminal (4104389), and
appropriate adapter (4115087), connect to fresh gas hose and attach
test terminal to in port of test volumeter.
7.24.11 Press and hold the O2 FLUSH button for 15 seconds; multiply the
value by 4.
7-72
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS PMS PROCEDURE (continued)
7.24.13 Remove the test minute volumeter and test fixture, and reconnect the
fresh gas hose.
7-73
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
PMS PROCEDURE (continued) FABIUS GS
7.25.3.1 Verify that all cylinder pressure gauges indicate zero (if
applicable).
7.25.3.2 Verify that the pipeline hoses are connected to the hospital
pipeline.
7.25.3.4 Verify that the O2 sensor is removed from the valve dome and the
plug is inserted in the inspiratory valve dome.
7.25.3.7 Return all machine controls and settings to their original state.
7-74
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SOFTWARE UPDATE PROCEDURE
This section outlines the basic software installation procedure, including the
equipment needed and its connections. For specific software versions, refer to
applicable Service Procedures.
NOTE: The screen illustrations contained in this document are for reference only
and therefore may or may not reflect the current equipment or software
version being installed.
8.1 REQUIREMENTS
NOTE:
U.S. units: Verify original machine software level prior to performing software
download procedure. Software downloads from version 1.20 to 1.3x shall only be
performed using Upgrade Kit P/N 4117917. Refer to Product Bulletin 143 for details.
Non-U.S. units: Verify original machine software level prior to performing software
download procedure. Software download from version > 1.22 to 1.3x shall only be
performed using the appropriate upgrade kit. Refer to Product Bulletin 143 for
details.
Equipment required:
The following Service Procedures are available via Lotus Notes and provides detailed
instructions to install/upgrade specific software versions. Refer to the following
Service Procedures or contact DrägerService - Technical Support: Phone 1-800-4-
Drager, Phone: 215-721-5402, or e-mail to techsupport@draegermed.com.
-- SP00255 Fabius GS Kit, Software Version 1.3N Upgrade
-- SP00269 Fabius GS Software Version 1.3N Upgrade
-- SP00312 Software Version 1.3N to 1.3N Upgrade
-- SP00297 Software Version 2.0N to 2.N Upgrade
8-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SOFTWARE UPDATE PROCEDURE (continued) FABIUS GS
This procedure uses Version 2.x of the SEND.IMG program. V2 is a new Win32
program that adds the Fabius GS boot Strap load function to the original
SENDIMG.EXE program that is currently used in existing products.
When a Fabius GS does not have a valid bootable image installed, it can not
perform an Automatic Image Reload Operation. The same PC and files are used
(including SENDIMG.EXE V2).
NOTE:
U.S. units: Verify original machine software level prior to performing software
download procedure. Software downloads from version 1.20 to 1.3x shall only be
performed using Upgrade Kit P/N 4117917. Refer to Product Bulletin 143 for details.
Non-U.S. units: Verify original machine software level prior to performing software
download procedure. Software download from version > 1.22 to 1.3x shall only be
performed using the appropriate upgrade kit. Refer to Product Bulletin 143 for
details.
8.2.1 Record the market setting on the unit for reconfiguration of the
controller PCB after download. Refer to Section 4 to access the
appropriate service screen.
8.2.3 Remove the back panel and change the JP2 (on control PCB P/N
4116632) or JP3 (on control PCB P/N 4118079) connector on the control
PCB from the Boot Flash position to the Boot Strap position .
Refer to
Figure 8.1 or Figure 8.2 for proper connector locations.
8.2.4 Connect the 3-pin plug end of the boot strap download cable (P/N
4117459) to J13 on the control PCB. See Figure 8.1 or Figure 8.2 for
proper connector locations.
8.2.5 Connect the DB9 end of the cable to the Com1 port on the PC.
8-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SOFTWARE UPDATE PROCEDURE (continued)
8.2.6 Power up the PC and double-click the “send_img” icon on desk top. The
following screen will appear:
8.2.7 Choose “Boot Strap and Image transfer”, then click OK or press Enter.
8.2.8 Set the machine power switch to ON. The bootstrap load of the FLASH
programmer load program then proceeds according to the status
messages on the PC. The following screen will be displayed:
8.2.9 Follow the same sequence as a normal download procedure to access the
software. See Step 8.2.5. Once the load program is executing on the
machine, the image load will proceed according to messages on the
machine and the PC until a message on the machine indicates
completion.
8.2.10 Close the PC Loader screen and power down the PC.
8-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SOFTWARE UPDATE PROCEDURE (continued) FABIUS GS
8.2.12 Return the boot selector jumper at JP2 (on control PCB P/N 4116632) or
JP3 (on control PCB P/N 4118079) to the Boot Flash position.
8.2.13 Cycle machine power and verify all self-diagnostics indicate a PASS
condition. NOTE: NVRAM settings will be set when Step 8.2.16 is
performed.
8.2.15 Access the appropriate service screens (see Section 7.5 of the Fabius GS
Service Manual) and update the machine serial number, clear service log
and reset hours run. The machine serial number is located on the rear of
the machine.
8.2.16 Perform PMS test Steps 7.6, 7.7, 7.17, 7.18, 7.20 and 7.21 in Section 7 of
the Fabius GS Service Manual. Verify the correct market kit settings are
entered per original machine configuration or customer requirements.
Refer to Fabius GS Market Kit setting located within the Service
Procedure.
LS1 LS3
LS2
F3 F1 F2
FIGURE 8-1. Boot Strap Jumper and Download Connections on Control PCB P/N
4116632
8-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SOFTWARE UPDATE PROCEDURE (continued)
JP3 J13
LS3
LS1 LS2
SV00611
F1
FIGURE 8-2. Boot Strap Jumper and Download Connections on Control PCB P/N
4118079
8-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS
Part numbers for field-replaceable items on the Fabius GS anesthesia system are listed on
the following pages, along with part numbers for related hardware and cables.
The item numbers are keyed to the accompanying illustrations to aid in identifying the item
and its location.
NOTE: Certain items in the following parts listings include an “alternate part number”
corresponding to those original part numbers that exceed seven (7) digits. Use the original
part number whenever possible when ordering spare or replacement parts.
ASSEMBLY/PART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PAGE
9-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
27
26
WRITING TRAY
ASSEMBLY
19
24
20 23
21
28
18 22
17
4
5
6
7 13 12 11 15
8
2
14
13
SV77015
9-2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
11. Valve assembly, Fresh Gas (without Label - See item #15 for Label P/Ns). . . . . . .4117226
12. Straight connector, M5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30952
13. Washer, M4 white . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M31602
14. Angle connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30953
15. Label, O2 flush:
Germany, Austria, France, Spain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2600494
USA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M28934
16. Deleted
17. Caster, locking (2x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8411945
18. Caster (2x). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8411946
19. Drawer lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117035
20. Drawer (top) includes clips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117036
21. Drawer (middle & bottom) includes clips (2x). . . . . . . . . . . . . . . . . . . . . . . . . . AF00007
22. Drawer Slide (6x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117008
9-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
7 5
6
SV00558
9-4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
3. Deleted
4. Sensor capsule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6850645
5. Housing assembly, less capsule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117454
6. Adapter (cover) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4106385
7. O-ring, #025 (Viton) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4105791
8. O-ring, #016 (Viton) (2x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4106388
9-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
2
6a
6b
8 6
1 6c
9
FRESH GAS
NOTE 15 DECOUPLING VALVE
10
A
NOTE 11a
A 15
11
12a
12 3
13a
APL BYPASS
VALVE
13
PEEP/PMAX
VALVE
5
14
SV77012
9-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
1 2
BAG SV77037
9-8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
1. Pole Assembly, Breathing Bag Extension (for use with item #3). . . . . . . . . . . . . . .4117753
2. Pole Assembly, Breathing Bag Extension (for use with item #4). . . . . . . . . . . . . . .4118380
3. Breathing System Configuration (for reference only) . . . . . . . . . . . . . . . . . . . . . . . . . . N/A
4. Breathing System Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SE4118378
9-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
3
5
9 2
SV00418
6
9-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV00419
5
6
4
1
3
9-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV77026
3
2
12
0.19 M
7
[7 1/2 IN]
13
14
0.33 M
[13 IN] 7
10
11
9
0.94 M
7 8
[37 IN]
9-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-15
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV00477
1
1
2 3 1
2 3
9-16
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-17
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
8
9
5
7
3 4
13 14
13 15
16 14
6
14 17
16 15
18
15 10
17
SV77016
9-18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
O2 Air N2O
Country
Color P/N Color P/N Color P/N
9-19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
15 14
3
14 15 4
2
5
6
19
17 13 7
14 15
8
16 13
9 10
13 14
20
SV77017 12 11
9-20
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
Fresh gas flow sensors and filter assembly (Fabius GS Machines with Serial Numbers of
12191 and lower)
[part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)]
1. Deleted
2. Flow sensor assembly (3x) (w/o angle connector (12) and O-ring (8)) . . . . . . . . . . .4116938
3. Cable assembly, gas flow sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117066
4. Safety valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4115950
5. O-ring, #113 (silicone). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4107961
6. Filter assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M16201
7. Flow sensor manifold, 3-gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4116978
Flow sensor manifold, 2-gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4116978-001
Alternate part number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4199937
8. O-ring, #106 (neoprene) (3x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4109322
9. Collar, 6 mm white . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M31603
10. Angle connector, M5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30961
11. Kep nut, M5 (2x). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HW55013
12. Angle adapter, 1/4 MPT to M5 (3x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4116986
13. Straight connector, M5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30952
14. Washer, M4 white . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M31602
15. Y-plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30962
16. Ferrule, Pilot to Divan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30963
17. Washer, 4 mm blue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30937
18. Deleted
19. Input valve, complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117293
20. Screw, M4 x 8 (2x). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1340778
21. Deleted
9-21
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
27
14 18
15 14
15
19
4
1
5
6
8
2
7
N2O
24
AIR 9
10
O2
14 13 13 13
15 17 16 14
12
9-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
Fresh gas flow sensors and filter assembly (Fabius GS Machines with Serial Numbers of
12191 and higher)
[part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)]
9-23
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV77018
4 5
6 5
7 5
9-24
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
1. Deleted
2. Pipeline pressure gauge (3x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117144
3. Screw, M4 x 12 mm self tap (4x) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4117021
4. Washer, 4 mm blue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30937
5. Straight connector, M5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30952
6. Ferrule, pilot to Divan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30963
7. Washer, M4 white . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M31602
9-25
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SORC
3
7
4
2
5
SV77019
9-26
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
SORC
[part of flow meter bezel assembly 4116543 (3-gas) or 4116542 (2-gas)]
1. Deleted
2. SORC (proportional valve) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M32940
3. Resistor, O2 red . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M27543
4. O-ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E20274
5. O-ring, DIN 3771 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M30308
6. Resistor, N2O blue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M27539
9-27
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV77020
1
3
2
4
9-28
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-29
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
20
22
17 18
O2
14
14
O2
19
23 12
O2
15
9
13
23
11 13
10
16
9
6 3
4 SV77021
9-30
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-31
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV77022
10
11
12
9
13
17
15
20 19 18
16 14
8
5
21
22
9-32
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
1. Deleted
2. Deleted
3. Deleted
4. Deleted
5. Connector, NIST N2O male . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4111382
Connector, DISS N2O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4111384
9-33
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
1 NOTE A SV77023
17
4 5
13 12
15
15
11
16
9-34
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-35
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
18
17 19
15
11
16
14
SV77024
12 13
8
6
7
9 4
5
3
1
2
10
9-36
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-37
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
11
10
13
4
6
15
1
7
9
8
7 REF
12
SV77036a
7 2
3 14
REF
5
9-38
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-39
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
8 8A 18
9 11
12
10 10A
6
7
17
13 16
15
4
3
14
19
11
1
SV77025
9-40
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
* Machine software level dictates the threshold of the negative pressure relief valve used:
-3mbar -8mbar
Software Negative Negative Patient
Market
Version Press relief Press relief Asm
valve (orange) valve (black)
9-41
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
9-42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-43
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV77027
4 5 6
1A
7
13 3 8
***G 11
10
***G
12 ***G - MOLYKOTE 55M GREASE
**P = LOCTITE #222 (PURPLE)
*R *R - LOCTITE #271 (RED)
9
9-44
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
NOTE: The adapter fitting is needed when ordering a two-vapor mounting system as a
replacement part for Fabius GS units with serial numbers < 10685.
Related parts:
Miscellaneous parts:
9-45
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
AGS 4
3
1
SV00560
9-46
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-47
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
SV00598
3
5
9-48
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-49
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
SPARE AND REPLACEMENT PARTS (continued) FABIUS GS
7 SV77040
4
SCAVENGER
9-50
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FABIUS GS SPARE AND REPLACEMENT PARTS (continued)
9-51
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DrägerService is a division of
Draeger Medical, Inc.
3122 Commerce Drive
Telford, PA 18969
Tel: (215) 721-5402
(800) 543-5047
Fax: (215) 721-5784
Web: www.draegermedical.com
Printed in the U.S.A.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
Page Description
1 continued
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
2
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
7-21 . . . . Switched steps 7.7.6 and 7.7.7, added new step 7.7.11 and associated
NOTE, changed (3x) to (5x) in step 7.7.10
7-22 . . . . Changed ‘Stop Bits’ to replace dash with comma in table within step
7.7.12
7-24 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Added checkmark by para. 7.8
7-28 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Added checkmark by para. 7.9.3
7-31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Deleted steps 7.9.4.6 and 7.9.6.7
7-42 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Expanded in step 7.12.1.1
7-57 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changed step 7.7.11
7-58 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Deleted checkmark from step 7.17.8
7-61 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changed para. 7.20.2
7-62 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changed paras. 7.20.26 and 7.20.32
7-64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Added (cmH2O) to step 7.21.4.6
7-66 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Added (cmH2O) to step 7.21.7.6
7-67 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Added (cmH2O) to step 7.21.8.2
7-70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . added (cmH2O) to steps 7.22.2
8-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removed second paragraph on page
8-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Added NOTE
8-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Added NOTE
8-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changed step 8.4.2
8-11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changed step 8.4.13, deleted 8.4.15
9-1 . . . . Updated TOC, removed Aux. Fresh Gas Outlet figure and lister page
9-2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-17 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-21 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
9-23 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated parts list
9-24 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Updated illustration
7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
Page Description
8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Page Description
4-1, 4-2, 4-4 . . . . . . . . . . . . . . . . . . . Revised Figs 4-1, 4-2, 4-4 to reflect sw 1.20
4-19 . . . . . . . . .Revised Fig 4-16 to show current offset value for vent pressure
4-21 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Revised Fig 4-18 to reflect sw 1.30
4-22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Revised Fig 4-19 caption
4-23 . . . . . . . . . . . . . . . . . . . . . . . . . . . Added Fig 4-20A to show sw 1.30 screen
4-27A. . . . . . . . . . . . . . . . . . . . . . . . . . Added Fig 4-24B to show sw 1.30 screen
5-27 . . Added Notes to ref P/L for correct patient asm & neg press relief valve
5-28 . . . . . . . . . . . . . . . . Text spillover to accommodate notes on previous page
5-35 . . . . . . . . .Added note to verify software level of replacement control PCB
5-36 . . . . . . . . Revised table of market kit settings to include 6 more countries
10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
7-67, 7-68 . Revised aux air valve & piston leak test for two software versions
7-70 . . . . . . . . . . . . Added step No Fresh Gas alarm check on SW version 1.30
8-1, 8-2 . . . . . . . . . . . . . . Revised page layout for added software version notes
8-8 thru 8-11 . . . . . Revised page layout to expand table of market kit settings
11