Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

2018 02 - Hoentsch U276 Hfa Dop e

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Hand-held unit HFA

A handy and easy-to-use All associated Höntzsch vane wheel


anemometer for measuring flow flow sensors with appropriate plug
velocity, volume flow and can be connected to the HFA display
temperature of air/gases and unit:
water/liquids. • sensors designed as probes or
measuring tubes, also for
±directional sensing of flow
Measurable variables are the flow • sensors with integrated tempera-
velocity v in m/s, the volume flow ture probe PT100
V/t in m³/h or l/min and the
temperature t in °C. • temperature probes PT100

Examples of Application

• measurements in outgoing air, • using velocity area method to


exhaust gases, flue gases, determine the flow rate or the
process gases … also in average flow velocity by approval
aggressive mediums up to 550 °C tests and inspections for proof of
guarantee according
• regulating and adjusting air condi- to VDI/VDE 2640
tioning and ventilating systems Corresponding Documents

U117 Data Sheet Vane Wheel


• measuring on filtering devices and • laminar flow measurements in
pharmaceutical application Flow Sensors FA
ventilating fans
U116 Overview FA
• measurements in clarified waste
U205 User’s Information
water, streams, rivers and lakes • flow measurement in all indus-
FA Probes
monitoring/measuring backflow trial spheres, in communal com-
when using ±directional sensing merce, in ministries, research U278 Operating Instructions
sensors and development HFA
U276_HFA_DoP_e_110803 http://www.hoentzsch.com 1/3
Hand-held unit HFA

Hardware Software

Input (8-pin flange socket) Analog output Measured value display


for connection of designated FA: Instantaneous values every
sensor type: 0 … 4 V = 0 … x m/s second. Display of actual velocity
FAR: v, actual volume flow V/t or
v/FA 0 … 2 … 4 V = -x … 0 … +x m/s, temperature (resolution 0.1 °C),
for non-directional sensing vane x = variable, impedance 2 kOhm selectable. Display of
wheel flow sensors (6-pin flange socket) instantaneous value in the START
FA (v-transmitters). and also START/STOP mode also
Power supply with long-term measurement.
by 9 V battery (operating time Display units:
v/FAR m/s, m³/h, l/min, °C.
30 hours) or by using the
for ±directional sensing vane
appropriate plug-in power supply
wheel flow sensors Operator information
in the 6-pin flange socket.
FA (v-transmitters). Status messages. Display of time
lapse during long-term measure-
Housing
vt/FA ment.
ABS plastic
for non-directional sensing vane W/H/L = 66/30 … 40/175 mm
wheel flow sensors Inputs, parameters and
FA (v-transmitters) with integrated measurement data
Protective system non-volatile storage, retrieval
temperature probe PT100.
IP64 for the housing. Splash-proof possible even after switching
protection can only be guaranteed OFF/ON or after power failure.
t/PT100 if both the flange sockets in the
temperature probe PT100. housing are connected to a water- Instrument settings
Connection cable with overall tight connection plug or a screw Terminal value analog output,
shielding: twin, 3 or 4-wire. cap according to regulations. measuring tube diameter or
rectangle area, profile factor,
Keypad Working temperature range display unit, types of sensor and
8 multifunctional keys 0 °C … +50 °C medium, instantaneous
value/long-term measurement
Weight SM/LM, long-term duration of
Interface RS232 approx. 265 g measurement, long-term
for direct transfer of the stored
commencement of measurement
measured values to a PC
automatic or after START, end of
(6-pin flange socket)
measurement follows
automatically or after STOP.
Display panel Short-term time constant
high-contrast LCD liquid crystal between 1 … 60 s variable.
display, measured value display 13 LM duration of measurement
mm high four-digit with decimal variable up to 9999 s.
point as well as affiliated display
unit, display of duration Time constant
of measurement in seconds The time constant set for the
6 mm high, adjustment and measured value display as well as
configuration instructions, the adjustment of the profile
operator’s information, status factor effect the instantaneous
displays 3 mm high. values of the analog output.

Datalogger

Data and measured value storage The data can be read out on the
for 1000 data records. When being display panel or on a PC using
stored the data records are the HFA RS232 interface and the
automatically marked with a group Höntzsch software HLOG for
number (up to 99 maximum) and WIN9X.
serial.

2/3 http://www.hoentzsch.com U276_HFA_DoP_e_110803


Hand-held unit HFA

Display unit, sensors and accessories


Article-no.
HFA display unit for measuring flow and temperature, a000/550
display unit incl. 9 V
Connection cable for analog output v approx. 1.5 m long, one end a000/503
fitted with connector 680-6, the other with exposed
Plug-in power supply HFA a000/552
for 230 VAC, 50/60 Hz, with connector 680-6
Replacement batterie 9 V, 450 mAh, IEC/DIN6LR61 a000/005
Software Set HLOG II V1.1 / HFA, HFA-Ex, HTA-Ex a000/520
for transfer and evaluation of the measurement values stored
in the data logger of the hand-held unit HFA, HFA-Ex, HTA-Ex
Connection cable fitted with connector 680-6 / D-Sub 9-pin, approx.
1.5 m long. CD ROM and connection cable in carrying case type E.
Interface converter USB / RS232 a010/100
connects PC with USB interface and Höntzsch programming adapter
with RS232 interface
PC connection : USB plug type A
programming adapter : sub-D 9-pin
Vane wheel flow sensors FA also for combined measuring of flow
and temperature, probes, probe extensions, probe guides, see leaflet
and Price List FA Vane Wheel Flow Sensors
Temperature probe TF500, immersion probe for measuring range a000/022
-100 °C ... +500 °C, probe length w/o handle 300 mm, Ø 4,2 mm
Temperature probe TF350, surface probe for -60 °C ... +350 °C a000/012
Extension cables
suitable for above mentioned sensors, 8-wire, resistant up to +60 °C,
connector plug 680-8 on both sides 3 m long a000/013
5 m long a000/014
connector plug 423-8 on both sides, waterproof 3 m long a000/015
5 m long a000/016
extension cable longer than 5 m standard price a000/025
extension cable longer than 5 m per meter a000/026
Y-Cable a000/511
for simultaneous connection of a flow sensor and a temperature probe
PT100 at the input of the hand-held units HFA and HFA-Ex, resistant up
to +60 °C, connector plug 680-8
Carrying case type G a000/515
Dimensions 497x411x140 mm
Capacity suggestion:
1 HFA
1 plug-in power supply
1 extension cable
1 temperature probe
1 flow sensor
1 extension rod
Light metal carrying case type A a000/019

Höntzsch GmbH
Gottlieb-Daimler-Straße 37
D-71334 Waiblingen (Hegnach)
Telefon +49 7151 / 17 16-0
Telefax +49 7151 / 5 84 02
E-Mail info@hoentzsch.com
Internet www.hoentzsch.com Subject to alteration

U276_HFA_DoP_e_110803 http://www.hoentzsch.com 3/3