Calibration Certificate
P-17 F03-00
Certificate/ULR No.: CC205522000006330F Page 1 of 1
MIS MARUTI ISPAT & ENERGY Instrument Receipt Date 22-Aug-2023
PVT.LTD.
Customer Details: SRF No. TNC1/JC/23-24/635
167, Madhavaram Village, Taluk Date of Calibration 22-Aug-2023
Mantralayam, Karnool Dist., A.P Pin Next Calibration Date
Code: 518345 22-Aug-2024
Certificate Issue Date 27-Aug-2023
A) ldentification On UUC (Unit Under Calibration)
Instrument Name Universal Testing Machine Make /.Model No. -UTE-FC 100
Range 1000 KN Instrument Condition OK
Least Count As per lnstrument Location of Calibration At Site
Serial No. 1012 Location LAB
Instrument ID. -

Parameter Force
ENVIRONMENTAL Temperature:- 24+2 °C,Relative Calibration Reference Standard IS 1828(Part1) :2015
CONDITION Humidity: - 50+10 Calibration Procedure TNC -01-F 2.0
Details of Test Equipments Used
Name L.D. NO.J SR NO Certificate/ULR No. Valid upto
Proving Ring VMW.17265.1000 kN
FRI/O6/20/12455 28-Aug-23
Calibration Results
Force(kN Division Division Observed ObservedObserved Average
) at Std. at roomReading Reading Reading of
AverageRelative
of
Relative Relative Relative Expende
Temp. Temp. (Division (Division (Division Reading Reading accuracy
error
Repeatab resolutio Zero
ility error n Error
24°C 24.4°C s)0° s) 120° s) 240° (Division) (KN) Error Uncertain
(q)% (b)% (a)% (fo)% ty of
Measure
ment t %
100 107.5 107.5 107.5 107.6 107.7 107.60 100.09 0.09 0.19 0.10
300 316.4 0.00 1.00
316.4 316.4 316.5 316.7 316.53 300.12 0.04 0.09 0.03 0.00 1.09
500 527.3 527.4 527.4 527.5 527.6 527.50 500.19 0.02 0.04
800 842.7
0.02 0.00 1.00
842.8 842.8 642.9 842.9 842.87 800.16 0.01 0.01 0.01 0.00 1.00
1000 1056.3 1056.4 1056.4 1056.5 1056.6 1056.50 1000.19 0.01 0.02 0.01 0.00 1.00
0 0
Note: 0
1. The Calibration results reporied in this certificate are valid at the time of and
the started
2. This report should not be reproduced except
3. Calibration certificate without in fullwithout our prior permission in writing.condition of measurement.
signature
4 All our certificates are Traceabie to are not valid and this certificate is inteded of only for guidance and not for legel purpose or for
5. This Certificate relates to the item national/International standard. advertisement.
Calibrated.Totalliabilityof Testncal Laboratory. willbe limited to the invoiced amount only.