Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Service Man CH

Download as pdf or txt
Download as pdf or txt
You are on page 1of 287

TM

CARDIOCAP II
CH-SERIES

SERVICE MANUAL

All specifications subject to change without notice

Document No. 878910- 1

March 22nd, 1996

Datex-Engstrom Division, Instrumentarium Corp.


P.O.Box 446 SF-00101 Helsinki Finland
Until Oct. 11th, 1996 Tel. +358 0 39411 Fax +358 0 146 3310
After Oct. 12th, 1996 Tel. +358 9 39411 Fax +358 9 146 3310
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-1

1 CONTENTS .......................................................................................................... PAGE


Table of Contents ...................................................................................................... 1-1
List of Figures.. ........................................................................................................ 1-4
List of Tables ............................................................................................................ 1-7

2 WARNINGS AND CAUTIONS.. .............................................................................. 2- 1


Warnings .................................................................................................................... .2-1
Cautions ...................................................................................................................... 2-3

3 INTRODUCTION AND APPLICABILITY OF THIS MANUAL ......................... 3-1


3.1 Introduction and applicability of this manual ............................................... 3-1
3.2 Summary of revision changes ......................................................................... 3-3
3.3 Manual updates ................................................................................................ 3-4
3 . 3 . 1 CARDIOCAP™ II CH-series service manual changes.. .................. 3-4
3.3.2 Record of manual updates carried out .............................................. 3-5
3.4 Software changes ............................................................................................. 3-6

4 GENERAL DESCRIPTION.. .................................................................................... 4- 1


4.1 Typical performance of CARDIOCAP™ II CH-SERIES
monitor ............................................................................................................ 4-1
4.2 Technical specifications of CARDIOCAP™ II CH-SERIES
monitor ............................................................................................................ 4-7
4.3 Principle of operation .................................................................................... 4-11
4.3.1 Principle of ECG measurement ....................................................... 4-11
4.3.2 Principle of Respiratory rate measurement.. ................................... 4-11
4.3.3 Principle of NIBP measurement.. ..................................................... 4-12
4.3.4 Principle of SpO 2 measurement ....................................................... 4-13
4.4 General block diagram .................................................................................. 4-17
4.5 Wiring diagram ............................................................................................. 4-19
4.6 External connector configurations.. .............................................................. 4-20
4.6.1 Input/Output specifications.. ............................................................ 4-20
4.6.2 Connectors ......................................................................................... 4-20

5 DETAILED DESCRIPTION O F MODULES............................................................. 5-1


5.1 ECG Board ...................................................................................................... 5-1
5.2 RESP Board ...................................................................................................... 5-6
5.3 I P T (Invasive-Pleth-Temperature) Board....................................................... 5- 12

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-2

5.4 NIBP Measurement ....................................................................................... 5-17


5.4.1 NIBP board ........................................................................................ 5-17
5.4.2 Pneumatic unit .................................................................................. 5-27
5.4.3 NIBP air pump .................................................................................. 5-27
5.4.4 Safety valve ...................................................................................... 5-28
5.4.5 NIBP tubing ...................................................................................... 5-28
5.4.6 Twin hose ......................................................................................... 5-28
5.4.7 Blood pressure cuff ........................................................................... 5-29
5.5 SpO2 Measurement.. ...................................................................................... 5-30
5.5.1 SpO2 Measuring Board.. .................................................................... 5-30
5.5.2 SpO2 Processor Board.. ...................................................................... 5-35
5.6 CPU Board .................................................................................................... 5-38
5.7 Video ASIC Board.. ....................................................................................... 5-43
5.8 Power Supply Board ...................................................................................... 5-46
5.9 Mother Board ................................................................................................ 5-51
5.10 Keyboard ....................................................................................................... 5-51
5.11 Loudspeaker Unit ......................................................................................... 5-51
5.12 Internal connector configurations.. ............................................................... 5-55

6 SERVICE AND TROUBLESHOOTING .................................................................. 6-1


6.1 General service information ........................................................................... 6-1
6.2 Disassembly and reassembly ........................................................................... 6-2
6.3 Troubleshooting .............................................................................................. 6-4
6.3.1 Monitor start-up sequence .................................................................. 6-4
6.3.2 Troubleshooting in general ................................................................. 6-5
6.3.3 ECG board troubleshooting ............................................................... .6-7
6.3.4 RESP board troubleshooting .............................................................. .6-8
6.3.5 IPT board troubleshooting .................................................................. 6-9
6.3.6 NIBP board troubleshooting ............................................................. 6-10
6.3.7 SpO2 measuring electronics troubleshooting.. .................................. 6-12
6.3.8 SpO2 processor board troubleshooting ............................................. 6-14
6.3.9 CPU board troubleshooting .............................................................. 6-15
6.3.9.1 Instructions after replacing the software or CPU
board ........................................................................................... 6-15
6.3.10 Video ASIC board troubleshooting .................................................. 6-16
6.3.11 Power supply board troubleshooting.. .............................................. 6-17
6.3.12 Mother board/Keyboard troubleshooting.. ...................................... 6-18
6.3.13 Loudspeaker troubleshooting ............................................................ 6-18
6.4 Service mode ................................................................................................. 6-19

ADJUSTMENTS.. ..................................................................................................... 7-1


7.1 ECG board adjustment.. ................................................................................. 7-1
7.2 RESP board adjustment .................................................................................. 7-2
7.3 IPT board adjustment.. .................................................................................... 7-3

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-3

7.4 NIBP board adjustment.. ................................................................................. 7-4


7.5 SpO2 processor board adjustment ................................................................... 7-5
7.6 CPU board adjustment .................................................................................... 7-5
7.7 Video display unit (Philips) adjustment ......................................................... 7-6

8 FUNCTIONAL FIELD CHECK PROCEDURE ..................................................... 8-1


8.1 Preoperative check list .................................................................................... 8-1
8.2 Checks after component replacements ........................................................... 8-3
8.3 Preventive maintenance check list.. ................................................................ 8-5

9 SPARE PARTS ......................................................................................................... 9-1


9.1 Spare parts ....................................................................................................... 9-1
9.2 Service accessories ......................................................................................... 9-4a

12 APPENDICES ......................................................................................................... 12-1

A COMPUTER OUTPUT ........................................................................................... 12-1

B CCP-104 GRAPHICS PRINTER ............................................................................ 12-5

C TRILINGUAL LIST OF ERROR MESSAGES.. ................................................... 12-6

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-4

List o f Figures................................................................................................................ PAGE

Absorption coefficients of oxy- and deoxy-hemoglobin in the red and


near-infrared regions ............................................................................................ 4-14

Absorption of infrared light in the finger ........................................................... 4-15

Finger sensor parts layout and schematic diagram .............................................. 4-16

General block diagram .......................................................................................... 4-18

Wiring diagram ...................................................................................................... 4-19

ECG board block diagram ...................................................................................... 5-3

ECG board parts layout (CH, CH-2, CH-S, CH-2S) ............................................ 5-4

ECG board schematic diagram (CH, CH-2, CH-S, CH-2S) .............................. .5-4a

ECG board parts layout (CH-R, CH-1R, CH-RS, CH-1RS)


(board modification level 5 and higher) ................................................................. 5-5

ECG board schematic diagram (CH-R, CH-1R, CH-RS, CH-1RS)


(board modification level 5 and higher) ................................................................ 5-5a

RESP board block diagram ..................................................................................... 5-7

RESP board parts layout and schematic diagram


(board modification level 5 and higher) ............................................................... 5-11

IPT board block diagram ...................................................................................... 5-14

IPT board (board modification level 4 and higher) and


IPT connector board parts layouts and IPT connector board
schematic diagram ................................................................................................. 5-15

IPT board schematic diagrams ............................................................................ 5-16a

Block Diagram of NIBP System (USA Version) ................................................... 5-24

NIBP board parts layout and schematic diagram


(part 1 ) (board modification level 4 and lower). .................................................. 5-25

5.10a NIBP board schematic diagram


(part 2) (board modification level 4 and lower) ................................................... 5-25a

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-5

5.11 NIBP board parts layout schematic diagram


(part 1) (board modification level 5 and higher)....................................................5-26

5.11a NIBP board schematic diagram


(part 2) (board modification level 5 and higher) .................................................5-26a

5.12 SpO2 measuring board block diagram, parts layout


and timing diagram ............................................................................................. 5-32

5.13 SpO2 measuring board schematic diagram ......................................................... 5-33

5.14 SpO 2 measuring board signal waveforms...............................................................5-34a

5.15 SpO2 processor board block diagram and parts layout


(board modification level 6 and higher)................................................................ 5-36

5.16 SpO2 processor board schematic diagram


(board modification level 6 and higher) .............................................................. 5-37

5.17 CPU board block diagram and parts layout...........................................................5-40

5.18 CPU board schematic diagram (part 1) .............................................................. 5-41

5.19 CPU board jumper configuration and schematic diagram


(part 2) .................................................................................................................. 5-42

5.20 Video ASIC board parts layout............................................................................ 5-44

5.21 Video ASIC board schematic diagram ................................................................ 5-45

5.22 Power supply board block diagram....................................................................... 5-47

5.23 Transformer diagram and power supply board signal waveforms..............................5-48

5.24 Power supply board parts layout and schematic diagram


(part 1) ................................................................................................................. 5-49

5.25 Power supply board schematic diagram (part 2) ................................................. 5-50

5.26 Mother board parts layout and schematic diagram


(board modification level 3 and higher)................................................................ 5-52

5.27 Keyboard parts layout and schematic diagram.......................................................5-53

5.28 Loudspeaker unit parts layout and schematic diagram........................................5-54

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-6

7.1 Video display unit adjustment .............................................................................. 7-6

9.1 Exploded pictures of the monitor ........................................................................ 9-5

9.2 ECG, IPT, and RESP board assembly .................................................................. 9-6

9.3 Pneumatic unit parts layout.. ................................................................................ 9-7

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-7

List of Tables.. ............................................................................................................... PAGE

4.1 Pin order of the ECG connector ........................................................................... 4-20

4.2 Pin order of the invasive pressure connector (PRESSURE 1)............................. 4-21

4.3 Pin order of the invasive pressure connector (PRESSURE 2). ............................ 4-21

4.4 Pin order of the pulse oximeter sensor connector (SpO2) .................................... 4-21

4.5 Pin order of the SERIAL & ANALOG connector.. .............................................. 4-22

4.6 Pin order of the AUX I/O connector.. .................................................................. 4-23

5.1 Video ASIC board (X1) - Mother board (X1). ..................................................... 5-55

5.2 Main CPU board (X1) - Mother board (X2). ....................................................... 5-56

5.3 SpO 2 processor board (X1) - Mother board (X3). ................................................ 5-57

5.4 NIBP board (X1) - Mother board (X4). ................................................................ 5-58

5.5 Power supply board (X10) - Mother board (X5). ................................................. 5-59

5.6 SpO2 measuring board (X2) - SpO2 processor board (X2). .................................. 5-60

5.7 Keyboard (X1) - Mother board (X8) .................................................................... 5-60

5.8 ECG board - Mother board (X12). ........................................................................ 5-61

5.9 IPT board - Mother board (X10). .......................................................................... 5-61

5.10 Power supply board - Mother board (X6). ............................................................ 5-61

5.1 1 ECG board - RESP board (X7) ............................................................................ 5-62

5.12 Mother board test connector (X13). ....................................................................... 5-62

5.13 Video ASIC board (X2) - Video unit main pc board (X13). ............................... 5-62

5.14 Front panel SpO 2 connector - SpO 2 measuring board (X1). ................................ 5-63

5.15 Power supply board (X1) - Line transformer.. ..................................................... 5-63

5.16 Power supply board (X3) - Fan ............................................................................. 5-63

5.17 Power supply board (X6) - Video unit main pc board.. ...................................... 5-64

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 1-8

Power supply board (X8) - Loudspeaker .............................................................. 5-64

NIBP board (X2) - NIBP Pump.. ........................................................................... 5-64

General troubleshooting chart.. ................................................................................. 6-5

ECG board troubleshooting chart.. ........................................................................... 6-7

RESP board troubleshooting chart ........................................................................... 6-8

IPT board troubleshooting chart.. ............................................................................ 6 - 9

NIBP board troubleshooting chart.. ........................................................................ 6-10

SpO 2 measuring electronics troubleshooting chart ................................................ 6-13

Power supply board troubleshooting chart ............................................................ 6-17

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 2-1

2 WARNINGS AND CAUTIONS


WARNINGS

A WARNING indicates that there is a possibility of injury to


yourself or others.

PROPER GROUNDING: For protection against shock hazards:

Connect this monitor to a three-wire, grounded, hospital


grade receptacle.

The power cord and plug must be intact and undamaged.

Do not remove the grounding prong from the power plug. Do not
use extension cords or adapters of any type.

Replace the power cord if it becomes cracked, frayed, broken or


otherwise damaged.

Confirm that AC-operated equipment used with the


monitor is properly grounded.

Do not perform any testing or maintenance on medical


instruments while they are being used to monitor a patient.

Do not break or bypass the patient isolation barrier when


testing ECG, Invasive-Pleth-Temp or SpO2 measuring
boards.

Use only defibrillation protected invasive pressure sensors on


DATEX-ENGSTROM Monitors.

EXPLOSION HAZARD: Do not use this monitor in the presence


of flammable anesthetics.

FUSE REPLACEMENT: Replace the fuse with a fuse of the same


type and with the same rating.

PATIENT SAFETY: Do not modify patient cables.

Use only patient cables and accessories approved by DATEX-


ENGSTROM. Other cables and accessories may damage the
monitor or interfere with measurement.

Do not use ECG during MRI.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 2-2

ELECTRIC SHOCK HAZARDS:

High Voltage is present within the CRT display unit.

In case of mechanical damage, inspect the integrity of the


patient isolation circuits, CRT unit, the power supply
transformer and power entry module.

Do not immerse the monitor in any liquid.

Do not immerse the SpO2 sensor in any liquid.

Do not use a sensor that is suspected of being immersed in liquid.


It may cause burns during electrosurgery.

Switch the power off and unplug the power cord before
cleaning or service.

Do not touch any exposed wiring or conductive surface while the


cover is off and the monitor is energized. The voltages present
when the electric power is connected to the monitor can cause
injury or death.

Perform a final electrical safety check and current leakage


test after doing any repair or calibration procedure to the
monitor.

The manufacturer accepts no responsibility for any


modifications made to the monitor outside the factory.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 2-3

CAUTIONS

A CAUTION indicates a condition that may lead to equipment


damage or malfunction.

The tests and repairs described in this manual should only


be done by trained personnel with proper tools and test
equipment. Unauthorized service may void the monitor
warranty.

Check the rear panel voltage setting before connecting the


monitor to AC mains power outlet.

Leave space behind the monitor to allow for proper


ventilation.

Clean or replace the fan filter regularly.

Before use, allow five minutes for warm-up and note any
error messages or deviations from expected operation. See
the Operator’s Manual.

Switch the monitor off before making any connections with


external equipment.

Do not use ammonia-, phenol-, or acetone-based cleaners.


These cleaners may damage the monitor surface.

Electrostatic discharge through the PC boards may damage


the components.

Before replacing and repairing PC boards, wear a static control


wrist strap.

Handle all PC boards by their non-conductive edges and use anti-


static containers when transporting them.

Do not apply tension to the power cord.

Do not autoclave the monitor nor sensors.

Do not gas sterilize the monitor.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-1

3 INTRODUCTION AND APPLICABILITY OF THIS MANUAL

3.1 Introduction and applicability of this manual


This service manual (Doc. No. 878910) provides information
required to maintain and repair the DATEX-ENGSTROM
CARDIOCAP™ II CH-series monitors. This manual is applicable
for the current production revision of the monitors. Differences
between monitor revisions are summarized in Section 3.2 and the
technical details of earlier revisions given in Chapter 11. Section
3.3 lists the hardware changes made to the monitor and Section 3.4
the software changes.

The revision of a monitor is changed when technical changes are


made to the monitor resulting in new spare parts that are
incompatible with earlier units. The last two digits of the monitor
type designation denote the revision of the monitor (e.g. CH-RS-
23-00 is a revision -00 unit).

Major parts, assemblies, and PC boards will have ID code stickers


indicating the modification level of the production documentation.
The code is shown as xxxxxx-y, where the “xx...” represents the
part number and “y” the revision level, which is referred to when
hardware changes are indicated in this manual.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-2

The following list shows the models and their monitoring


parameters.

MODEL MONITORING PARAMETERS

CH ECG, NIBP, Temp


CH-2 ECG, NIBP, Temp, 2 x invBP
CH-R ECG, NIBP, Temp, Resp*)
CH-1R ECG, NIBP, Temp, invBP, Resp*)
CH-S ECG, NIBP, Temp, SpO2
CH-2S ECG, NIBP, Temp, SpO2, 2 x invBP
CH-RS ECG, NIBP, Temp, SpO2, Resp
CH-1RS ECG, NIBP, Temp, SpO2, invBP, Resp

*) Manufacturing discontinued from revision -04 on.

This manual describes all the functions offered by the


CARDIOCAP™ II CH-series monitors. Some of the functions
may not be available in the monitor you are using.

Please review the Operator’s Manual to obtain a clear


understanding of the monitor.

The manufacturer reserves the right to make changes in product


specifications at any time and without prior notice. The
information in this document is believed to be accurate and
reliable; however the manufacturer assumes no responsibility for its
use.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-3

3.2 Summary of revision changes


Revision -00

Initial production revision of the monitor.

Revision -01

The main differences to the revision -00 are:

- Main software

- NIBP software

- SpO2 software

- NIBP tubing redesigned

- One of the NIBP measurement patient groups, NEONATAL is


changed to INFANT. The specification of INFANT is 5 kg or
more, one to twelve months old.

Revision -02

Current production revision of the monitor. The main differences


to the revision -01 are:

- Main software

- NIBP software (except for Germany)

- SpO2 software

- High speed main CPU board (p/n 880523) installed to replace the
existing board (p/n 878822)

- Enlarged fan airway inlet and larger dust filter

- Power supply board modified

- ECG board modified

- NIBP pneumatic unit modified

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-3a

Revision -03

The main differences to the revision -02 are:

- Main software

- SpO2 software

- SpO2 measuring board

Revision -04

Current production revision of the monitor. The main differences


to the revision -03 are:

- CE approval and marking accordingly

- Improved EMC protection including new ECG board, IPT board


and SpO 2 board. The main mechanical modifications have been
made on the rear panel, which now has a transformer case and a
new mains switch.

- Video control board has been replaced by video ASIC board.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-4

3.3 Manual updates

3.3.1 CARDIOCAP™ II CH-series service manual changes


This is update number 3. We have made the following changes to
this manual:

- started printing on both sides of pages to decrease manual


thickness.

- gave up costly extensive pages.

Carry out the update by making the changes shown under this
update number in the list below:

- replace pages which already exist in the manual.

- insert pages which do not exist yet in the manual after the
previous entry in numerical or an alphabetical order.

- of all the pages removed from the manual, the following pages
should be filed in chapter 11. All other pages are thrown away.

5-5, 5a, 11, 11a, 15, 16, 33, 33a, 33b, 34, 34a, 34b, 37, 37a, 42,
45, 52

After having made the changes required, replace Section 3.3.1 of


the manual with this one and sign the record of updates in Section
3.3.2.

No Page Change Date

1 3-3 Revision 01 introduced. September 2nd, 1991

3-7,8 Revision 01 software added.

4-1...6 Typical performance revised.

4-12 Note on neonatal changed.

4-16 Figures changed.

5-16 A7 pin 15 corrected to pin 16.

5-24 Neonatal changed to infant.

5-25 Program Eprom text location corrected.

5-27...29 Neonatal changed to infant.

5-42 CPU board jumper table revised.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-4a

5-50 Power supply board schematic diagram replaced.

6-6 Cuff occlusion message added.

9-1...4 Spare parts list revised.

9-7 Pneumatic unit illustration changed.

12-3 Status string corrected.

12-4 Capnomac text deleted.

3-3 Revision 02 introduced. 20th March 1992

3-7, 9...11 Text revised. Revision 02 software described.

4-6 Trends table revised.

4-12...14 Principles of measurement revised.

4-22 Pin order table revised.

5-4 ECG board modified.

5-29 Pediatric cuff renamed to Child.

5-38...42 High speed CPU board introduced.

5-49, 50 Power supply board modified.

6-15 CPU board troubleshooting text revised.

7-4 R31 corrected to R6 in NIBP board adjustment

7-5 CPU board adjustment text revised.

9-1...4 Spare parts list revised.

9-7 Pneumatic unit parts layout changed.

12-1...6 Appendices pages revised. Page 12-5 removed.

1-1...8 Contents revised. 10th December 1993

3-3a Revision 03 added.

3-7, 9, 12 Text revised. Revision 03 software described.

4-3...6 Text revised.

4-5 Max. cuff pressure limited to 330 mmHg.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-4b

4-16 Text and figure revised.

4-21 Table 4.4 revised.

4-22, 23 Tables revised.

4-24 Page removed.

5-4...4a Page layout changed.

5-5...5a ECG board parts layout and schematic diagram updated. Page
layout changed

5-11 RESP board schematic diagram updated.

5-15...16 IPT board parts layout and schematic diagram updated. Page
layout changed.

5-24 Text and figure revised.

5-25...26a Page layout changed.

5-28 Safety valve check added.

5-30...31 Text revised.

5-32...34a SpO2 measuring board changed. Page layout changed.

5-36...37 SpO2 processor board parts layout and schematic diagram


updated. Page layout changed.

5-38...39 Text revised.

5-40...42 Jumper position changed. Page layout changed.


5-44...45 Video control board parts layout, timing diagram and schematic
diagram updated. Page layout changed.
5-48...50 Page layout changed.
5-52 Mother board parts layout schematic diagram updated.
5-55, 56 Tables revised.

5-63 Table 5.15 revised.


6-1 Text revised.

6-5...6 Error messages added.


6-11...12 Text revised.
6-15 Text added.
6-19 Service mode introduced.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-4c

9- 1...4 Spare parts added.

9-5 Figure revised. Page layout changed.

12-1...4 Text revised. The status characters of the standard string and
status code interpretation added.

12-16...17 Page layout changed.

12-21 Trilingual list of error messages added.

Company name & phone info changed. 22nd March 1996

1-1 Table of Contents updated

1-4 List of Figures updated

1-7 List of Tables updated

2-1 ECG/MRI warning added to Patient Safety

3-1 Panasonic Service Manual deleted

3-2 Note “Manufacturing discontinued...” added.

3-3a Revision 04 introduced

3-4c Revision 04 manual changes introduced

3-5 Revision 04 introduced

3-12 Word “current” deleted

3-13 New page added for 04 softwares

4-5 Word “Germany” replaced by

“in adaptation 70 monitors”.

4-10 Power req. changed from 100/115/220/240

to 100/115/220-240

4-16,17 Probe changed to sensor

4-21 Table 4.2: Pins 6-8 to not connected

Pin 9 to P Ground

5-4,4a ECG board parts layout and schematic changed

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-4d

5-5,5a ECG board parts layout and schematic changed

5-12,13 IPT Board text revised

5-14...16 IPT board artworks changed

5-32...34 SpO2 meas. board artworks changed.

5-39 Caution text revised.

5-43 Video control board changed to Video ASIC

board, text revised

5-44,45 New artworks for Video ASIC board

5-46 Supply voltage changed from +15 V to +5 V.

5-47 Power supply board block diagram changed

5-48 Transformer diagram changed

5-62 Video control board changed to Video ASIC

board

5-63 Probe changed to sensor

6-6 New message “NIBP INACTIVE ? SELF CHECKING...

and explanation added

6-16 Video control board changed to Video

ASIC board, text revised

7-6 New section 7.7 Video Display Unit (Philips)

Adjustments and one new artwork

9-1...5 Spare parts list and exploded monitor parts

pictures updated

12-6 CCK-104 Keyboard chapter deleted, old

appendix D (Trilingual list of error

messages) becomes appendix C.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-5

3.3.2 Record of manual updates carried out

Update Carried out by


number Name Date

1 DATEX September 2nd, 1991

2 DATEX March 20th, 1992

3 DATEX December 10th, 1993

4 DATEX-ENGSTROM March 22nd, 1996

10

11

12

13

14

15

16

17

18

19

20

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-6

3.4 Software changes


The software code (six numbers) and revision number (if other
than initial revision) are displayed on the screen during the startup
sequence.

The initial production software versions are as follows.

ITEM DESCRIPTION ORDER No.

CPU board
Software EPROM (-, 2, R, 1R) (English) 878107
(-, 2, R, 1R) (German) 878108
(-, 2, R, 1R) (French) 878109
(-, 2, R, 1R) (Germany) 878110
(S, 2S, RS, 1RS) (English) 878111
(S, 2S, RS, 1RS) (German) 878112
(S, 2S, RS, 1RS) (French) 878113
(S, 2S, RS, 1RS) (Germany) 878114

NIBP board
NIBP software EPROM 877039
EPROM (Germany) 877040

SpO2 processor board


SpO2 software EPROM 875360

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-7

The production revision 01 softwares are as follows.

ITEM DESCRIPTION ORDER No.

CPU board
Software EPROM (-, 2, R, 1R) (English) 879425
(-, 2, R, 1R) (German) 879426
(-, 2, R, 1R) (French) 879427
(-, 2, R, 1R) (Germany) 879428
(S, 2S, RS, 1RS) (English) 879529
(S, 2S, RS, 1RS) (German) 879430
(S, 2S, RS, 1RS) (French) 879431
(S, 2S, RS, 1RS) (Germany) 879432

NIBP board
NIBP software EPROM 879808
EPROM (Germany) 879809

SpO2 processor board


SpO2 software EPROM 878105

The following changes have been made to the initial software


version.

Main software

- Graphic trends are available in two pages:


First page: HR, SpO2, NIBP, invBP1, invBP2
Second page: SpO2, RR

- 4 and 8 hour trends added. Trends also in numeric form (HR,


NIBP, invBP1, SpO2).

- Numeric trends printed in CCP printer.

- Real time ECG waveform displayed while viewing the trend.

- PROBE OFF/NO PROBE alarm sequences revised.

- Audible ECG leads off alarms added.

- Return from deeper menu levels to the first menu level can be
short cut by hitting the respective menu hardkey.

- Straight access from a parameter menu to another is possible


without pressing the RETURN TO MONITOR key.

- ECG displayed in cascade form if pleth waveform is selected to


the lowest display field.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-8

- CO 2 waveform from DATEX-ENGSTROM Ultima monitor can


be displayed on the third display field.

- Audible indication of NIBP measurement readiness is available.

NIBP software

- NIBP autocycling mode is interrupted if a cuff loose state is


detected.

- Cuff occlusion alarm is added.

SpO2 software

- Two level measuring light intensity adjustment is done


automatically.

- Averaging time is selectable between 5 and 10 seconds.

NOTE: To upgrade CARDIOCAP II monitors from 00 to 01


revision, all three new software (CPU, NIBP, and SpO2) should
be installed simultaneously because none of the revision 00
software is compatible with revision 01 software.

NOTE: All three new software (CPU, NIBP, and SpO2) can be
installed in revision 00 monitors simply by replacing the
software EPROMs and shorting the jumper X2 pins 1 and 2 on
the CPU board.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-9

The production revision 02 softwares are as follows.

ITEM DESCRIPTION ORDER No.

CPU board
Software EPROM (-, 2, R, 1R) (English) 880788
(-, 2, R, 1R) (German) 880789
(-, 2, R, 1R) (French) 880790
(-, 2, R, 1R) (Germany) 880791
(S, 2S, RS, 1RS) (English) 880784
(S, 2S, RS, 1RS) (German) 880785
(S, 2S, RS, 1RS) (French) 880786
(S, 2S, RS, 1RS) (Germany) 880787

NIBP board
NIBP software EPROM 879808
EPROM (Germany) 879809

SpO2 processor board


S p O 2 software EPROM 878105

The following changes have been made to the revision 01 software


version.

Main software

- More complete numeric trends presentation

- More complete real time waveform presentation

- Capable of recognizing the hardware and software levels of the


main CPU board, NIBP and SpO2 modules of the monitor:

16 MHz CPU board or 11.7 MHz CPU board

NIBP hardware of CARDIOCAP II or NIBP of


CARDIOCAP I

SpO2 hardware of CARDIOCAP II or SpO2 of


CARDIOCAP I

- Automatically, when setting the system up, adjusts the


intermodular communication protocols and menu structures on the
screen accordingly. That is why this software can be installed also
into CARDIOCAP II revisions 00 and 01 units as well as into
CARDIOCAP I units (with the main CPU board p/n 878822).

- The patient group PEDIATRIC is renamed to CHILD.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-10

NOTE: This new software can be installed into all CARDIOCAP


II revision 01 units without any modifications. Note that some
new features need the new high speed main CPU board to be
activated.

NOTE: This new software can be installed into all CARDIOCAP


II revision 00 units with main CPU board jumper modifications
(pins 1 and 2 shorted in jumpers X2 and X4).

NIBP software

- Improved performance in cases like small children, moving and


shivering patients, high blood pressures, low blood pressures,
strong and weak pulses.

NOTE: This new software can be installed into all CARDIOCAP


II revision 00 and 01 units without any modifications.

- Improved performance in NO PROBE conditions.

NOTE: This new software can be installed into all CARDIOCAP


II revision 01 units without any modifications.

NOTE: This new software can be installed into all CARDIOCAP


II revision 00 units together with rev 01 or 02 main software
after the CPU board jumper modifications (pins 1 and 2 shorted
in jumpers X2 and X4).

Upgrading

To upgrade CARDIOCAP II units from 00/01 to 02 revision, all


three new software (CPU, NIBP, and SpO2) should be installed
simultaneously as well as the high speed CPU board.

If the new high speed CPU board is not desired, the three new
softwares can simply be installed into CARDIOCAP II units 01
revision monitors. All three new softwares (CPU, NIBP, and SpO2)
should be installed simultaneously.

If the new high speed CPU board is not desired, the three new
softwares can be installed into CARDIOCAP II units 00 revision
monitors with the CPU board jumper modification. All three new

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-11

software (CPU, NIBP, and SpO2) should be installed


simultaneously.

NOTE: All three new softwares have to be installed at the same


time. None of the previous revision softwares are compatible
with the new softwares.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-12

The production revision 03 softwares are as follows.

ITEM DESCRIPTION ORDER No.

CPU board
Software EPROM (-, 2, R, 1R) (English) 882597
(-, 2, R, 1R) (German) 882598
(-, 2, R, 1R) (French) 882599
(-, 2, R, 1R) (Germany) 882600
(S, 2S, RS, 1RS) (English) 882593
(S, 2S, RS, 1RS) (German) 882594
(S, 2S, RS, 1RS) (French) 882595
(S, 2S, RS, 1RS) (Germany) 882596

NIBP board
NIBP software EPROM 879808
EPROM (Germany) 879809

SpO2 processor board


SpO 2 software EPROM 882881

The following changes have been made to the revision 02 software


version.

Main software

- Automatically, when starting up the monitor the real-time clock


and the EEPROM are checked. A warning message “MEMORY
CIRCUIT FAILURE / CALL SERVICE” will be displayed if:
- The real-time clock is in disorder or has stopped.
- A fault has been detected in the memory circuit D4.

- If the temperatures or the invasive pressures have not been


calibrated after a factory reset the following messages will be
displayed:
- TEMPERATURE NOT CALIBRATED, RECALIBRATE
- INVASIVE BP NOT CALIBRATED, RECALIBRATE

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 3-13

The production revision 04 softwares are as follows.

ITEM DESCRIPTION ORDER No.

CPU board
Software EPROM (-, 2) (English) 882597
(-, 2) (German) 882598
(-, 2) (French) 882599
(S, 2S, RS, 1RS) (English) 882593
(S, 2S, RS, 1RS) (German) 882594
(S, 2S, RS, 1RS) (French) 882595

NIBP board
NIBP software EPROM 879808
EPROM (Germany) 879809

SpO2 processor board


SpO 2 software EPROM 882881

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-1

4 GENERAL DESCRIPTION

4.1 Typical performance of CARDIOCAP™ II CH-SERIES monitor

Measurement method

3-lead ECG
5-lead ECG (not with Respiratory rate measurement)

Lead selection

I, II, III with standard 3-lead cable


V connection with optional 5-lead cable (not with Respiratory rate
measurement)

Measuring range

QRS detection 0.5 - 5 mV, 40 - 200 ms


Allowable offset ±300 mV

Waveform display

Frequency response 0.5 - 30 Hz


Sweep speeds 12.5 and 25 mm/s
Gain range 0.2 - 4.0 mV/cm

Numerical display (heart rate)

Range 30 - 230 beats/min


Accuracy ±1 %
Resolution 1 beat/min
Averaging 10 s
Update interval 5s

Alarms

Asystole no QRS detected in 7 s

Heart rate
high limit adjustable 30 - 250 beats/min
low limit adjustable 30 - 250 beats/min
alarm delay 10 s after limit violation

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-2

RESPIRATION

Measurement method

Transthoracic impedance

Measuring range

0.3 - 3 ohm (dynamic), max. transthoracic impedance 3 kohm

Waveform display

Frequency response 0.1 - 3 Hz (-3 dB)


Sweep speed 6.25 mm/s
Length of trace approx. 16 s

Numerical display (respiratory rate)

Range 0, 2 - 180 breaths/min


Accuracy ±3 breaths/min
Resolution 1 breath/min
Averaging 30 s
Update interval 2s

Alarms

Apnea no breath detected during time speci-


fied by apnea alarm limit (3 - 30 s)

Respiratory rate
high limit adjustable 2 - 180 breaths/min, OFF
low limit adjustable 2 - 180 breaths/min, OFF
alarm delay 10 s after limit violation

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-3

INVASIVE BLOOD PRESSURE

Measurement method Transducer 50 uV/cmHg/V,


20 mA max. current

Measuring range -40 - 260* mmHg

Zero adjustment range ±100 mmHg

Calibration range ±50 %

Waveform display

Frequency response 0 - 20 Hz
Sweep speeds 12.5 or 25 mm/s
Scales -5 - 32 mmHg
-10 - 65 mmHg
-20 - 130 mmHg
-40 - 260* mmHg

Numerical display

Range -40 - 260* mmHg


Resolution 1 mmHg
Averaging 10 s
Update interval 5s

Alarms

Systolic pressure
high limit adjustable -40 - 260* mmHg
low limit adjustable -40 - 260* mmHg
alarm delay 10 s after limit violation

* 230 mmHg in Germanv

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-4

TEMPERATURE

The temperature channel is YSI 400 series sensors compatible

Numerical display in °C or °F

Update interval 24 s

Amplifier accuracy ±0.1°C (25.0 - 45.0°C) /


±0.2°F (77.0 - 113.0°F)
±0.2°C (15.0 - 24.9°C) /
±0.4°F (59.0 - 76.8°F)

Sensor accuracy ±0.1°C (15 - 45°C)/


(YSI 400 series) ±0.2°F (59.0 - 113.0°F)

NON-INVASIVE BLOOD PRESSURE

Measurement method Oscillometric

Measuring range Adult 25 to 260* mmHg


Pediatric 25 to 195 mmHg
Infant 15 to 145 mmHg

Accuracy Compared to auscultatory method in adults (slow deflation


speed):
systolic ±5 mmHg, STD < 8 mmHg, diastolic ±5 mmHg,
STD < 8 mmHg

Compared to auscultatory method in neonates (normal


deflation speed):
systolic ±5 mmHg, STD < 8 mmHg, diastolic ±5 mmHg,
STD < 8 mmHg

Accepted heart rate 30 to 250 beats/min

Measurement intervals manual, continuous for 5 min,


1, 2, 3, 5, 10, 15, 30, 60 min

Measurement time

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-5

Safety Limits Adult Pediatric Infant

Max. measurement time 2 min 2 min 1 min


Max. inflation pressure 280 mmHg 200 mmHg 150 mmHg
Overpressure limit, stops
measurement 320 mmHg 220 mmHg 165 mmHg
Initial inflation pressure 185 mmHg 185 mmHg 120 mmHg
Successive infl. pressures syst + 40 syst + 40 syst + 40
Next infl. pressure after
‘low infl press’ message +50 mmHg +50 mmHg +40 mmHg
Stasis time 2 min 2 min 1 min
Stasis pressure 80 mmHg 60 mmHg 40 mmHg

Mechanical safety valve limits max. cuff pressure to 330 mmHg.


Independent timing circuit limits pressurization to max. 5 min.

Alarms
Systolic pressure
high limit OFF, adjustable 30 - 260* mmHg
low limit OFF, adjustable 30 - 260* mmHg

* 230 mmHg in adaptation 70 monitors

OXYGEN SATURATION

Measurement method Red and infrared light absorption

Numerical display

Range 40 - 100 %
Accuracy 100 - 80 %, ±2 % SpO 2 (±l STD)
Resolution 1%
Averaging 6 s (rev 00), 5 s/10 s (01)
Update interval 2s

Alarms

SpO 2
high limit adjustable 40 - 99 %, OFF
low limit adjustable 40 - 99 %
alarm delay 10 s after limit violation

Pulse beep pitch corresponds to SpO2 level

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-6

PLETH

Automatic scale setting when sensor is attached.


User-adjustable scale during operation.

Heart rate

Range 30 - 250 bpm


Accuracy ±1 %, ±1 bpm
Resolution 1 bpm
Averaging 5 s or 10 s (selectable)
Update interval 5s

Alarms

Heart rate
high limit adjustable 30 - 250 beats/min
low limit adjustable 30 - 250 beats/min
alarm delay < 3 s after limit violation

TRENDS

Revision 00

2 hours graphic trends of Heart Rate, Respiration Rate, Systolic


and Diastolic Pressures, SpO2, and Pleth Amplitude. Sampled every
10 s except NIBP.

Revision 01 and up

0.5, 2, 4, 8 hours of graphic and numerical trends of Heart Rate


(0.5 hour trend Revision 02 and up), Respiration Rate, Systolic and
Diastolic Pressures, SpO2, and Pleth Amplitude. Sampled as follows
except NIBP.

Trend Time Sampling Rate Display Rate

0.5 h HR (10 s), RR (20 s) every 10 s


2 h SpO2 (30 s) Pleth (30 s) every 10 s value
4 h P1 (40 s, 20 s if no P2) every 20 s average
8 h P2 (40 s) every 40 s average

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-7

4.2 Technical specifications of CARDIOCAP™ II CH-SERIES


monitor
Subject to change without notice

The ECG channel features an integral electrosurgery filter, defibrillator


protection and pacemaker pulse detection and rejection.

Patient Safety Type CF isolation

RESPIRATION

Input impedance 6.4 Mohm at 10 Hz


1.3 Mohm at 50 Hz

Sensitivity 1 V/ohm

Excitation signal frequency 48.3 kHz (nominal)

Auxiliary patient current

The EGG/ Respiration channel features an integral electrosurgery filter,


defibrillator protection and pacemaker pulse detection and rejection.

Patient Safety Type CF isolation

Novemher 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-8

INVASIVE BLOOD PRESSURE

Amplifier

Patient Safety Type CF isolation

TEMPERATURE

The temperature channel is YSI 400 series sensors compatible

NON-INVASIVE BLOOD PRESSURE

Pressure transducer accuracy better than ±3 mmHg


or ±2 % (whichever greater)

Patient Safety Type BF isolation

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-9

OXYGEN SATURATION

The pulse oximeter accuracy measurements are statistically derived and


correlated to simultaneous SpO2 measured on an Instrumentation Laboratory
IL/282 CO-oximeter.

Accuracy 100 - 80 %, ±2 % SpO2 (±1 STD)


80 - 50 %, ±3 % SpO2 (±1 STD)
50 - 40 %, unspecified

Patient Safety Type BF isolation

PLETH

Automatic scale setting when sensor is attached.


User-adjustable scale during operation.

Amplifier Bandwidth 0.8 to 15 Hz (-3 dB)

Patient Safety Type BF isolation

ALARMS

Adjustable audio alarm sound, rear panel loudspeaker


(approximately 900 Hz)
Adjustable high/low alarm limits for heart rate, systolic pressures,
SpO2, respiratory rate
Adjustable alarm for apnea
Alarm for asystole

DISPLAY

9” green monochrome picture tube


Resolution 1024 x 256 pixels

Clock

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-10

GENERAL DATA

Safety standard fulfilled


IEC 601-1, Safety class I
CSA C22.2 No.125-M1584

Operating temperature 10 to 35°C / 50 to 95°F


Storage temperature -5 to 50°C / 23 to 122°F
Power requirements 100/115/220-240 V
50/60 Hz
120 W, 600 mA
DxWxH 340 x 330 x 210 mm
13.4 x 13.0 x 8.3 inches
Weight 12 kg / 26 lbs

EXTERNAL CONNECTIONS

Serial data output for computer interface (SERIAL & ANALOG


I/O connector) and graphics printer (AUX I/O connector)
Composite video output
ECG, PLETHYSMOGRAPH or PRESS1, PRESS2 analog waveform
output (SERIAL & ANALOG I/O connector)
ECG test signal
AC, DC power supply connection
Alarm signal (Nurse call)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-11

4.3 Principle of operation

4.3.1 Principle of ECG measurement


Electrocardiography is the process of analysing the electrical
activity of the heart by measuring the electrical potential produced
with electrodes placed on the surface of the body.

ECG reflects:
Electrical activity of the heart
Normal/abnormal function of the heart
Effects of anesthesia on heart function
Effects of surgery on heart function

See the Operator’s Manual for electrodes positions and other


information.

4.3.2 Principle of Respiratory rate measurement


Impedance Respiration is a technique of monitoring respiratory
rates of a patient based on fractional changes in thoracic
impedance between two electrodes. These electrodes are the same
electrodes used for ECG measurement.

When monitoring ECG and respiratory rates simultaneously the


position of the electrodes is crucial. See the Operator’s Manual for
more information.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-12

4.3.3 Principle of NIBP measurement


NIBP (Non-Invasive-Blood-Pressure) is an indirect method for
measuring blood pressure.

The NIBP measurement uses the oscillometric measuring principle.


The cuff is inflated with a pressure slightly higher than the
presumed systolic pressure, then slowly deflated at a speed based
on the patient’s heart rate, collecting data from the oscillations
caused by the pulsating artery. Based on these oscillations, the
monitor calculates values for systolic, mean, and diastolic pressures.

The NIBP measuring unit is a fully automatic, self-contained non-


invasive blood pressure measuring system which communicates
with the main CPU via an asynchronous serial channel. All NIBP
functions are controlled by the NIBP’s CPU in the NIBP board.
The NIBP system contains the following main parts:

- NIBP board
- Pneumatic
- NIBP air pump
- Safety valve
- NIBP tubing
- Twin hose
- Blood pressure cuff

The NIBP board contains the pressure transducer, bleed and


exhaust valves, and electronic parts of the system.

The pneumatic contains two damping chambers, two magnetic


valves, and tubes (see Figure 9.3).

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-13

4.3.4 Principle of SpO2 measurement


SpO2

Oxygen is the most acutely necessary substrate for survival.


A major concern during anesthesia is the prevention of tissue
hypoxia. Thus immediate and direct information about tissue
oxygenation is needed. When oxygen diffuses from the alveolus
into the blood it dissolves into the plasma and binds to hemoglobin.
Most of the oxygen needed by the body is transported bound to
hemoglobin. The total hemoglobin in the blood is composed of
oxygenated oxyhemoglobin (HbO2), reduced or deoxygenated
hemoglobin (Hb) and other forms of hemoglobin such as
Dyshemoglobins: carboxyhemoglobin (HbCO) and methemoglobin
(MetHb).

Pulse Oxymetry determines noninvasively the oxygen saturation of


hemoglobin (SpO2) on the basis of light absorption at only two
wavelengths. The limitation is that correspondingly only two
hemoglobin species can be discriminated by the measurement. In
principle, Pulse Oxymetry can be calibrated either against
fractional saturation SaO 2frac ,

S a O 2 f r a c = HbO2 / (HbO2 + Hb + Dyshemoglobin)


or against functional saturation SaO2func,

SaO2func = HbO2 / (HbO2 + Hb),

which is less sensitive to changes of Dyshemoglobin concentrations


in blood.

The oxygen saturation percentage measured by the DATEX-


ENGSTROM CARDIOCAP II monitor is calibrated against the
functional saturation SaO 2func . The advantage of this method is
that the accuracy of SpO2 measurement can be maintained even at
rather high concentrations of carboxyhemoglobin in blood.
Independent of the calibration method pulse oxymetry is not able
to correctly measure oxygen content of the arterial blood at
elevated Dyshemoglobin levels, which clinically may be harmful
for patient.

The absorption of light of normal human blood at different


wavelengths is mainly determined by oxygenated oxyhemoglobin
and by deoxygenated deoxyhemoglobin (see Figure 4.2). The
monitor measures the relative absorption of light at two
wavelengths, one in the near infrared (about 900 nm) and the other
in the red region (about 660 nm) of light spectrum. These
wavelengths are emitted by LEDs and detected by a PIN-diode in
the sensor. The total absorption can be divided into components of

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-14

arterial blood. Only the last component gives variations


synchronous with heart beat into the transmitted light intensity.
This fact is most essential for Pulse Oxymetry and eventually
makes feasible the measurement of oxygen saturation
noninvasively.

Plethysmographic pulse wave

The plethysmographic waveform is derived from the IR signal and


reflects the blood pulsation at the measuring site. Thus the
amplitude of the waveform represents the perfusion.

Pulse rate

The pulse rate calculation is done by peak detection of the


plethysmographic pulse wave. The signals are filtered to reduce
noise and checked to separate artifacts.

Figure 4.1 Absorption coefficients of oxy- and deoxy-


hemoglobin in the red and near-infrared regions

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-15

Figure 4.2 Absorption of infrared light in the finger

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-16

Sensor

The standard sensor is a finger clamp sensor which measures


through the finger (see Figure 4.3) and contains the light source
LEDs in one half and the photodiode detector in the other half.
Different kinds of sensors are also available from DATEX-
ENGSTROM.

Figure 4.3 Finger sensor parts layout and schematic diagram

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-17

4.4 General block diagram


The monitor consists of the following modular parts (see page 3-2
for the parts included in the monitor you are using):

ECG and respiration measuring electronics

NIBP measuring electronics, pump, and tubing

Invasive BP and temperature measuring electronics

Sensor and SpO2/pulse oximeter measuring board

Main processor board including analog signal multiplexer,


A/D converter, and real time clock

Video ASIC board to convert the CPU commands into


video signal

Video display module

Transformer and power supply board to generate necessary


voltages and I/O functions

Mother board including signal buses and analog input signal


buffers

Tactile membrane keyboard

Loudspeaker unit

See Figure 4.4 for the monitor block diagram.

For monitor parts locations see the exploded view (Figure 9.1) in
Chapter 9.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-19

4.5 Wiring diagram


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-20

4.6 External connector configurations

4.6.1 Input/Output specifications


Serial data output for computer interface (SERIAL & ANALOG
I/O connector)
Serial Output for Graphics Printer (AUX I/O connector)
Composite video output

ECG, PLETHYSMOGRAPH or PRESS1, PRESS2 analog waveform


output (SERIAL & ANALOG I/O connector)
ECG test signal
AC, DC power supply connection
Alarm signal (Nurse call)

4.6.2 Connectors
Table 4.1 Pin order of the ECG connector

5-lead cable ECG measurement not with Respiratory rate


measurement.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-21

Table 4.2 Pin order of the invasive pressure connector


(PRESSURE 1)

Table 4.3 Pin order of the invasive pressure connector


(PRESSURE 2)

Table 4.4 Pin order of the pulse oximeter sensor connector


(SpO2)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-22

Table 4.5 Pin order of the SERIAL & ANALOG connector

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 4-23

Table 4.6 Pin order of the AUX I/O connector

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-1

5 DETAILED DESCRIPTION OF MODULES

5.1 ECG Board


The ECG board is an isolated amplifier that accepts the low level
(approx. 1 mV) signal from the patient cable via the protection
circuitry and the lead selector. The signal is filtered and amplified
to A/D-conversion levels.

ECG input

The input circuit has a four position switch for selecting the
desired patient lead configuration and directing the signal into the
preamplifier input. In models which include RESP measurement,
the switch is limited to three positions.

Defibrillation and ESU protection

Input protection against defibrillation pulses (typically 5 kV/400 J)


is established by 4.7 kOhm/1 W resistors and 6800 uH coils in each
patient cable lead and by the spark gap F2.

NOTE: In CARDIOCAP™ II models which include the RESP


board, the defibrillation pulse protection circuits are located on
the ECG board.

At the input there is also the electrosurgery interference (ESU)


filter (R22, R23, R28, R29 and C6-C8) and a diode circuit that
serves as input limiter and provides input amplifier saturation bias
if the patient connection is broken, (LEADS OFF).

The patient isolation section is shielded by a metal enclosure to


prevent high frequency ESU interference. A second spark gap is
connected across the isolation barrier to allow a controlled passage
for static voltages which otherwise would destroy the isolation
components.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-2

Amplification

Each input is amplified by standard non-inverting amplifiers


which drive a differential amplifier, whose output is sent to a
pulse width modulator operating at approximately 5 kHz. The pulse
width modulated signal drives an optocoupler that transmits the
ECG signal across the isolation barrier.

ECG output

The output stage consists of a pulse width demodulator, low pass


filter and an adjustable 50/60 Hz notch filter. The signal is
demodulated, filtered and sent to an analog multiplexer on the
CPU board.

Leads off/Pacemaker pulses

Input amplifier saturation caused by a lead off or by a high slew


rate pulse is detected by the IN-OP comparators. In case of
LEADS OFF, the INOP output is a steady low voltage state. When
a pacemaker pulse is detected, a fast (0.5 - 2 ms) low pulse is
produced.

These signals are transmitted across the isolation barrier using an


optocoupler. The signals are sent to port C of the 8255 PPI on the
CPU board. The constant low signal will produce the LEADS OFF
message on the display and output data string and pacemaker pulse
will be clipped from the ECG waveform.

ECG power supply

Isolated power is generated by a small 75 kHz switching converter


(D1, V24, V25 and T1) the output of which is a regulated
+/- 12 V, using zener diodes.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-3
Figure 5.1 ECG board block diagram
November 15th, 1990
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-4

Figure 5.2 ECG board parts layout


(CH, CH-2, CH-S, CH-2S)

Figure 5.2a (on the next page)


ECG board schematic diagram
(CH, CH-2, CH-S, CH-2S)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-4a

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-5

Figure 5.3 ECG board parts layout


(CH-R, CH-1R, CH-RS, CH-1RS)
(board modification level 5 and higher)

Figure 5.3a (on the next page)


ECG board schematic diagram
(CH-R, CH-1R, CH-RS, CH-1RS)
(board modification level 5 and higher)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-6

5.2 RESP Board


The RESP board measures the respiratory rate of a patient based
on fractional changes in thoracic impedance between two
electrodes.

An oscillator produces a 48 kHz signal at a constant voltage that is


applied to the input of a preamplifier whose input impedance
varies with respiration. The output of the amplifier is rectified and
filtered producing a signal where a 1 Vpp change represents a 1.0
ohm change in the thoracic impedance of the patient.

The dynamic measuring range of the amplifier is 0.1 to 3.0 ohm,


with a frequency response of 0.2 to 3.0 Hz corresponding to
respiratory rate of 12 to 180 per minute.

The AC amplifier produces a signal from which respiratory rate


data is displayed. it operates within an impedance window
established by the static impedance level (approx. 3 kohms) and the
voltage peak to peak limit of the AC amplifier. If a limit is
exceeded, an IN-OP switch is activated which effectively turns off
the AC amplifier. An error message will be displayed.

The RESP board (connector Xl) is connected to the ECG board


(connector X8) with flexible cable.

The RESP board is included in models 1R, 1RS, R, and RS.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-7

Figure 5.4 RESP board block diagram

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-8

Patient hook up

The electrodes are placed on the patient in such a manner to


provide a static impedance of up to 3 kohms. As the patient
breathes, this impedance will change fractionally.

This variable patient or base impedance becomes part of the


preamplifier input circuit.

Resonant circuit

The resonant circuit consisting of the base impedance, transformer


T2, high voltage capacitors C35 and C36, choke L1 and a spark
gap F3. Transformer T2, provides the necessary patient isolation as
well as stepping up the impedance of the resonant circuit. The
effect is that small changes in the base impedance caused by
respiration, produce larger changes to the preamplifier input
impedance.

Choke L1, matches the resonant circuit so that its resonant


frequency is not dependant on the base impedance. The input
impedance of the preamplifier is approximately 1.3 Mohm at 50 Hz
and 6.4 Mohm at 10 Hz.

This circuit is located on the ECG board.

Oscillator

A1B is an oscillator whose frequency is adjusted by trimmer R54


to 48.3 kHz. Output waveform is sinusoidal with some limit applied
to the peaks.

Preamplifier

The output of the oscillator is applied to the input of the


preamplifier Al A. The amplifier has a gain of 4, but the actual
signal strength produced will vary as the input impedance varies.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-9

Detector

The output of the preamplifier is split so that some of the signal is


processed by the amplifier A1C and diodes V3 and V4 (which
form a full wave rectifier). This rectified signal is summed with
the non-rectified signal through R5, then low-pass filtered by
A1D. The result of which is to produce a low-frequency AC signal
with a DC component.
* The DC-component is produced from the static base
impedance.
* An AC-component is the result of the impedance variation
of the base impedance caused by respiration.

RESP amplifier

The AC-component of the detected signal is amplified by an AC


amplifier A2A and A2B (RESP-amplifier), then transmitted to the
amplitude limiter.

Amplitude limiter

Operational amplifiers A2C and A2D monitor the output signal of


the RESP-amplifier. If the output signal exceeds a pre-established
point, the INOP-switch is activated when the signal is rectified by
the limiter circuit and applied A3B.

The voltage applied through D1 is established by a pull-up resistor


at +15 V. When the inverted voltage from A3B is applied to D1,
the output of the RESP-amplifier becomes 0 volts.

The output of the amplitude limiter is also transmitted to the


analog multiplexer on the CPU board via the ECG board.
Respiratory rate is then calculated and displayed.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-10

Comparator

The DC signal is applied to a comparator circuit (A3A). The


voltage is directly proportional to the impedance in the patient
circuit. The comparator has a threshold level of about 4.2 V
(corresponding to impedance of 3 kohm). When the voltage at the
input exceeds this level (that is, the base impedance has exceeded 3
kohms), the comparator output changes state (+15 V). This voltage
is applied through diode V13, amplified and inverted by A3B.

In addition, the comparator output is transmitted to port B of the


8255 PPI on the CPU via the ECG board. The change in state
causes an interrupt to be produced resulting in an error message to
be displayed.

Defibrillation protection

The defibrillation protection is realized by five resistors (R58


through R62) in the ECG board.

Over voltage protection

Transient suppressor diodes VI, V2, V19, and V20 protect the
measuring amplifier against an Over voltage pulse which might
pass through the transformer.

Voltage regulator

The operating voltage of oscillator, preamplifier, and detector is


regulated and stabilized with A3C and A3D. This reduces noise in
the first stages of the amplifier and keeps the amplitude of the
oscillator constant.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-11

Figure 5.5 RESP board parts layout and schematic diagram


(board modification level 5 and higher)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-12

5.3 IPT (Invasive-Pleth-Temperature) Board


The IPT board is designed to support two pressure channels and
one temperature channel. However, in models CH-1R and CH-
1RS, only one pressure channel is available. In models CH, CH-R,
CH-S, and CH-RS, no invasive pressure channels are provided.

The boards have two common features with the ECG board; the
isolating power switcher and the pulse width modulator/
optoisolator/ demodulator chain.

The two-pressure channel IPT board generates two real time


waveforms:

1. Invasive pressure P1 waveform.

2. Invasive pressure P2 waveform.

Temperature

The temperature signal is produced by a voltage divider, part of


which is the patient sensor (YSI 400-series thermistor). The output
is amplified by the calibrated amplifiers. Offset adjustment and
linearization is done by software.

Invasive pressure

The monitor provides +5 V excitation voltage and a ground


reference. The transducer output is sent to a single differential
amplifier IC. The excitation voltage is sourced from the isolated
power supply on the IPT board and all variations in this voltage
are compensated for by software.

The Cardiocap senses when a pressure sensor has been inserted into
a connector by providing a line on each connector that is held high
(pressure flag) by a pull-up resistor. The line is shorted to ground
when a pressure transducer is plugged into the monitor. The
pressure flag lines are polled by the multiplexer and when the CPU
senses the line is grounded, the pressure display is activated.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-13

Signal output

The following signals are multiplexed to a pulse width modulator:

- the temperature channel


- two pressure flags
- zero and +5 V reference
- pressure differential signal

The multiplexer control is provided by the CPU.

If a second pressure is provided, its differential output is directed


to its own pulse width modulator.

After all signals are transmitted across the isolation barrier, they
are demodulated, low-pass filtered and transmitted to the analog
multiplexer resident on the CPU board.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-15

Figure 5.7 IPT board (board modification level 4 and higher)


and IPT connector board parts layouts and IPT
connector board schematic diagram

Figure 5.8 (on the next page)


IPT board schematic diagram
(board modification level 4 and higher)m

(Connector X4 or both X3 and X4 not included in some models)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-16a

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-17

5.4 NIBP Measurement

5.4.1 NIBP board

PNEUMATIC SYSTEM

The NIBP board consists of the following pneumatic parts:

- Pressure transducer
- Bleed system
- Exhaust valve

See Figures 5.10 and 5.11 for NIBP board parts layout and
schematic diagram.

Pressure transducer

The piezoresistive pressure transducer (B1) is connected to the


cuff. It measures the absolute pressure of the blood pressure cuff
and the pressure fluctuations caused by arterial wall movement.

Bleed system

Bleed valve (Y2) releases the cuff pressure. The opening of the
valve is pulse-width controlled between 100% and 0%. When the
pulse-width ratio is 100% the signal is constantly high and the
valve is open all the time. When the ratio is 0% the signal is
constantly low and the valve is closed all the time. The driving
signal frequency is 40 Hz.

Exhaust valve

Valve Y1 is used as an exhaust valve to quickly deflate the cuff.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-18

ANALOG CIRCUITS

Pressure transducer excitation

The pressure transducer B1 is excited by a 4 mA constant current


source. This is produced by floating constant voltage of 1.200 V
that is produced by the voltage reference diode V7. Operational
amplifiers A5A and A5B are used as voltage followers. The voltage
difference through resistor R5 (TP X8/7, X8/8) is 1.200 V,
producing a current of 4 mA through the resistor for the pressure
transducer. The excitation voltage of B1 is about 12 V (TP X8/8,
X6/5).

Differential amplifier

The output of the pressure transducer B1 is a differential signal


between the midpoints of the pressure sensing resistor bridge legs
(TP X8/2, X8/3). This signal is fed to a differential amplifier
(operational amplifiers A4A and A4B, and resistors R12, 13, 14,
and 15) that produces a gain of 31.1 (TP X8/4).

Pressure zero and gain control

At zero pressure, the output voltage difference of the pressure


transducer B1 may be up to 100 mV. This zero pressure voltage is
compensated for by R6 (TP X8/5) in the amplifier stage A4C. This
stage is an inverting amplifier with unity gain and low-pass
filtering.

The next stage, A4D, is an inverting amplifier whose gain can be


adjusted with trimmer R11. The gain is adjusted to compensate for
pressure transducer gain differences, so that a linear signal is
produced throughout the transducers range (TP X8/6).

Pressure DC channel range and bias circuits

The DC channel is used to measure the static or non-oscillating


pressure of the blood pressure cuff. The static signal from A4D has
a 5 V range (0 to -5 V). This voltage is sent to A2A which limits
the voltage range to 1 V (+0.755 V to -0.755 volts) for the A/D
converter.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-19

Two input ports of the analog multiplexer (A3) are dedicated to


the DC channel. The microprocessor selects the desired port
depending on the cuff pressure required for best resolution. If the
cuff pressure is between 0 and 255 mmHg the signal produced by
A2A is sent directly to the multiplexer port 6, because the pressure
corresponds directly to the 8 bit range of the A/D converter.

If the cuff pressure exceeds 255 mmHg, the signal is directed


through a voltage divider network (R30/1-2 and R30/3-4) to port
5 of the multiplexer. This doubles the pressure range represented
by the 1 volt signal and since the voltage measured at the divider
is referenced to ground the pressures represented are between -128
and +382 mmHg.

To allow for a certain amount of zero pressure drift, the pressure


zero points are actually set at 0 to 235 mmHg and -108 to 362
mmHg respectively.

Pressure AC channel filters

The AC component of the cuff pressure data is amplified to allow


the processor to analyze the small cuff pressure fluctuations which
are used as a basis for blood pressure determination. This
additional amplifier is AC coupled. The first stage, A2B, is a
second order high-pass filter to effectively block out the DC
component of the pressure data (TP X9/2). The second stage, A2C,
is the amplifier stage (G = 73) and low-pass filter (TP X9/3). The
third stage A2D blocks the offset voltages of the previous two
stages (TP X9/4).

AC reset

During cuff inflation or if an abnormal situation arises, the AC


signal is sent to ground through D6A and D6B with AC reset signal
P16 from the processor.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-20

AC channel range selection

The output from A2D is connected to a voltage divider chain (R24


through 28). The 5 outputs of this resistor chain make up a binary
division chain. The division ratio is selected by the analog
multiplexer (A3). If voltage at the beginning of the chain is 1 volt
the following outputs have voltages 1/2, 1/4, 1/8 and 1/16 volts.
The processor selects the proper division ratio according to the
incoming signal level. This effectively extends the dynamic range
of the A/D converter from 8 bits to 12 bits.

Multiplexer and A/D converter

The analog multiplexer circuit A3 is used to select either the DC or


AC channel to the A/D converter. The same multiplexer is also
used to select suitable ranges for both channels. As described
previously, the DC channel has two different ranges and the AC
channel has 5 different ranges. The multiplexer inputs are allocated
according to the following table.

INPUT DESCRIPTION DIVISION RATIO

The operational amplifier A1C is a non-inverting gain stage


(G = 6.62), which amplifies the incoming signals to -5 V to +5 V
level (TP X9/5). Resistor network R30/5-6 and 7-8 shifts the -5 to
+5 V range to 0 - 5 V for the A/D converter U1. U1 is an 8 bit
successive approximation A/D converter. The converter is timed by
its own internal clock (TP X6/2), the conversion rate is determined
by the external RC circuit (R29, C34), conversion time is about
200 us. The converted signal is led to buffer D13 output when
ADCRD-signal is activated. Decoder D5 activates the ADCRD-
signal when Al 5 and RD-signal are active.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-21

Reference voltage source

The reference voltages are generated with the reference diode V6


and operational amplifiers A5C and A5D. The reference voltages
are typically +5.30 V (TP X6/4) and -10.6 V (TP X6/5).

JUMPERS

The NIBP board contains two jumpers.

Jumper X13 selects the watchdog time. Normally the time is


adjusted to 5 minutes (no jumper). In test purpose the time can be
changed to 5 seconds (X13 short-circuited).

Jumper X10 selects the RAM memory capacity used.

PIN RAM CAPACITY SELECTED

1-2 32 kbytes
2-3 8 kbytes

PUMP AND VALVE DRIVERS

The magnetic valves and the air pump are controlled by the open
collector darlington driver circuit D11. A separate power source
(+15 V) is used to drive these circuits. An additional power switch
(V1, V3), controlled by the processor, connects power to valve
circuits. The watchdog timer prevents a prolonged inflation.

SAFETY CIRCUITS

Pressure control switch

A comparator circuit A1A compares cuff pressure to a fixed


reference level. If cuff is inflated the circuit prevents CPU from
zeroing the watchdog circuit (TP X9/6).

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-22

Watchdog timer

The watchdog timer circuit (D8, D9) controls the power of the
magnetic valves and the air pump. Normally, power to the valves is
not connected. The CPU is able to reset the watchdog circuit and
connect power to the pump only when the cuff is not inflated. If
the cuff is inflated, the CPU cannot control the watchdog circuit,
which will deflate the cuff after 5 minutes.

PROCESSOR CIRCUITS

The NIBP board contains its own 8051FA CPU and memory
circuits to allow independent operation. The 64 kbyte code memory
resides in D2 (EPROM) and the 8 kbyte data memory in D3
(RAM). Ports P3 and P1 are used to control directly various
operations of the NIBP.

Port P3 also provides RS-232 output which is used to transmit


NIBP data to the main CPU.

Memory organization

The code memory and the external data memory both start from
zero. The decoded address space is 64 kbytes.

CPU Control ports

CPU ports P3 and P1 are used to directly control several NIBP


functions.

PORT P3 FUNCTION

P 3.0 RXD/
P 3.1 TXD/
P 3.2 Mux address 0
P 3.3 Mux address 1
P 3.4 Mux address 2
P 3.5 15 V power sense
P 3.6 WR
P 3.7 RD/

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-23

PORT P1 FUNCTION

Reset circuit

The reset line is connected to the common reset line RESET to


guarantee simultaneous resetting with the main processor at power
down conditions.

TEST POINT SIGNALS

PIN SIGNAL

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-24
Figure 5.9 Block Diagram of NIBP System
Two different kinds of twin hose are used: one for standard adult
cuff and another which contains a built-in choke for infant cuff.
December 10th, 1993/3
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-25

Figure 5.10 NIBP board parts layout and schematic diagram


(part 1) (board modification level 4 and lower)

Figure 5.10a (on the next page)


NIBP board schematic diagram (part 2)
(board modification level 4 and lower)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-25a

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-26

Figure 5.11 NIBP board parts layout and schematic diagram


(part 1) (board modification level 5 and higher)

Figure 5.11a (on the next page)


NIBP board schematic diagram (part 2)
(board modification level 5 and higher)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-27

5.4.2 Pneumatic unit


The pneumatic unit includes three damping chambers and two
magnetic valves.

Damping chamber I

The damping chamber I prevents a rapid increase of pressure


caused by the pump.

Choke

A choke is used to slow down the pressure change in infant


measurement.

Damping chamber II

The damping chamber II is used with small cuff sizes to increase


volume so that the pressure can be precisely adjusted.

Damping chamber III

The damping chamber III smooths down rapid pressure pulses


caused by the bleed valve.

Shunt valves

If the cuff volume is too small, the two shunt valves open and
connect damping chamber II to the blood pressure circuit.

5.4.3 NIBP air pump


The NIBP air pump module contains a membrane type pump and a
DC motor. The module is enclosed in a separate foam rubber-filled
case to attenuate noise.

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-28

5.4.4 Safety valve


The NIBP is equipped with a separate mechanical safety valve to
prevent accidental cuff over-pressurization. The valve operates
nominally at 330 mmHg.

The safety valve is checked by connecting a hand pump with a


manometer to it and pumping slowly until the valve opens. It
should open at 300-330 mmHg.

5.4.5 NIBP tubing


The pneumatic components are assembled together as shown in
Figure 9.3.

WARNING: PATIENT SAFETY: It is important to attach the


tube from pressure transducer to the inner connector of the
NIBP-connector and the tube from the pneumatic unit is
attached to the outer connector of the NIBP-connector.
Reversing the tubes will inhibit deflation of the infant cuff.

5.4.6 Twin hose


The twin hose is used to connect cuffs to the monitor. The choke
which is designed for infant measurement is located in the infant
adapter (USA) or inside the input port of the infant twin hose (all
others). The monitor automatically detects the hose type in use.

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-29

5.4.7 Blood pressure cuff


The NIBP measuring unit is designed to operate with seven
different standard blood pressure cuffs. The cuff sizes are as
follows:

CUFF SIZE CUFF

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-30

5.5 SpO2 Measurement


SpO2 measurement is included in models 2S, RS, 1RS, and S.

5.5.1 SpO2 Measuring Board


The board is intended to perform the following tasks:

Control the LED light sources of the sensor.

Amplify the signal coming from the detector and separate


the red and infrared signal components to respective
channels.

Multiplex in both channels the alternating component of the


signal (plethysmographic pulse) with the signal proportional
to the total intensity measured with the respective
wavelength.

Provide isolated output from the multiplexer channels (red


channel and IR channel) to the SpO2 Processor board.

Power supply

The isolated power supply consists of the following functional


blocks:

- 32,768 Hz oscillator.
- Half-bridge converter with isolation transformer.
- Stabilization and filtering of the output voltages with linear
regulators.
- Protection of the overloading with PTC-type thermistor.

Timing/LED control

The timing pulses are produced by a PAL (Programmable Array


Logic) D3. The input signal for D3 (SYNC.) is taken from the
switching power supply as a 32,768 Hz square wave. All timing
signals are synchronized at this switching frequency. The timing
circuit controls the LED driver circuitry (signals LEDR and
LEDIR), the RC time constants in amplifier chain (MEASURE)
and sampling (SAMPLER, SAMPLEIR) (Figure 5.14).

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-31

Referring to Figures 5.13 and 5.14, the LEDs in the sensor are
driven with constant current pulses, (90 or 300 mA). The pulse
duration and duty cycle can be seen in timing diagram in Figure
5.13. A positive voltage pulse at 1/X1 corresponds to the red LED
current and a negative one to the IR-LED, respectively.

Detector signal processing

The signal produced by the detector is a current. The first


amplifier stage is a current-to-voltage converter. A signal current
passes through the resistors between pins 13 and 14 of A3 and
produces a negative voltage pulse at 14/A3. Notice that the part of
the feedback resistance is located in the sensor connector.

The bias voltage of the detector (4.2V) is the voltage difference


between the connector pins 3/X1 (5V) and 5/X1 (0.8V).

At 8/A3 the detected voltage pulses are inverted to positive value.

The digitally controlled amplifier is a Digital to Analog Converter


(DAC), D5. The signal is fed to the reference input of D5. The 8-
bit digital control word is transferred over the patient isolation
barrier in serial mode (PA2) and is converted into parallel mode by
a shift register D4. The signal level at the output, 7/A3, is adjusted
to 3 to 8 V by the CPU.

The amplified signal pulses are separated to red and infrared


channels by sample-and-hold circuitry (S/H). Voltages VR and VIR
are proportional to the total intensity of the light detected at the
respective wavelength. VRac and VIRac are the amplified alternating
components (plethysmographic pulses).

The signals are multiplexed into two channels by a 2 x 4 MUX,


A5. Also +5 V and GND are connected to MUX input. The value
of the resistor RC in the sensor connector can be read through the
red channel, if needed.

The two output channels of MUX A5 are transferred across the


patient isolation by two identical pulse width
modulator/optoisolator/demodulator-chains. The frequency of the
pulse width modulator is about 20 kHz. The demodulated signal is
inverted.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL
Figure 5.12 SpO2 measuring board block diagram, parts layout
and timing diagram
Figure 5.13 (on the next pages)
SpO2 measuring board schematic diagram
March 22nd, 1996/4
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-34

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-34a

Figure 5.14 SpO2 measuring board signal waveforms

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-35

5.5.2 SpO2 Processor Board


The SpO2 processor board accepts the multiplexed signal from the
SpO2 measuring board, converts the signal to digital data, processes
the information and transmits the data to the main CPU.

The SpO2 processor board (see Figure 5.15) contains, the 8031
CPU, EPROM, RAM, the parallel interface circuit, analog
multiplexer/demultiplexer, and the analog/digital converter.

The CPU uses the internal bus to access most of the peripheral
circuits. CPU port 1 is used to control the analog multiplexer (A3).
The serial channel 0 (ASCII computer output) is not used on this
board.

The two memory chips are jumper selected for 27256 program
EPROM, 8 x 8 kbit low current CMOS RAM 5564 powered by the
data retention voltage.

SpO2 input

The analog signals from the SpO2 measuring board are sent to an
analog MUX, A3 that is used as a demultiplexer. The output goes
to a sample/hold circuit and then transmitted to the A/D
converter, the output of which is put on to the data bus.

SpO2 processor board output

The SpO2 data is transmitted to the main CPU through the 8255
PPI. In addition, the measuring board MUX and DAC control
signals are also sent through the PPI.

November l5th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-36

Figure 5.15 SpO2 processor board block diagram and parts layout
(board modification level 6 and higher)

Figure 5.16 (on the next page)


SpO2 processor board schematic diagram
(board modification level 6 and higher)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-37

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-38

5.6 CPU Board


The high speed CPU board contains 16 MHz oscillator to replace
that of 11,059 MHz. Serial I/O signals that used the CPU’s own
channel in the previous CPU board are changed and they occupy
one channel in QUART.

The board contains, in addition to the 80C32 CPU, the standard


EPROM, SRAM, and several analog and digital I/O functions. See
the CPU board block diagram.

The CPU (D5, refer to Figure 5.18) uses the CPU board internal
bus to access most of the peripheral circuits. Processor port 1 is
used to control the analog multiplexers (MUX).

Communication with the NIBP board is established in serial mode


through QUART (D15) pins 4 (in) and 3 (out).

There are three memory chips on the CPU board. The EPROM D1,
1M bit or 4M bit, is jumper selected, see table 5.19, 32 x 8 kbit
low current CMOS SRAM (D6) powered by the data retention
voltage, and battery back-up 8 kbit SRAM (D4) for permanent
calibration value memory. This IC has also a built-in Real Time
Clock. The lifetime of the battery inside the IC is minimum 10
years. If the SRAM is defective or the battery is dead the monitor
gives warnings every 30 seconds. Please refer to section 6,
Troubleshooting in general. See the jumper configuration.

The 16 MHz oscillator clock is first passed through control logic


circuit (D2, D16, and D19), which is for wait state control, and fed
to the CPU.

When a key is pressed (short-circuit) keyboard scanner (D9)


interrupts the microprocessor and this reads from the scanner
which key was pressed.

Input signal processing

The analog input signals from the IPT, ECG, and RESP (if used)
boards are read through the multiplexer (A3) to a sample and hold
circuit and to the A/D-converter A2.

Output signal processing

Up to eight analog signals are sent to the output connectors. Digital


patient data is sent to a D/A converter A1 and after conversion,
sent to the demultiplexer A7. The eight analog signals produced are
sent to sample and hold circuits and then transmitted to the power
supply board and the connectors in the back of the monitor.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-39

Control signals of MUX are in port 1 on the microprocessor as follows:

pins 3-5 MUX A0-A2 (both)


pin 6 MUX enable (both)
pin 7 MUX 0 Write (ADC)
pin 8 MUX 1 Write (DAC)
ADC 0 IMP. RESP
ADC 2 ECG
ADC 4 IPT signal
ADC 6 PB2 signal
ADC 7 EXT INPUT
DAC 2 IPT signal
DAC 4 ECG TEST
DAC 5 Loudspeaker volume
DAC 6 PB2 signal
DAC 7 Loudspeaker pitch

Ports on the PPI is used for as follows:

When a key is pressed (short-circuit) keyboard scanner (D9) interrupts the


microprocessor and this reads from the scanner which key was pressed.

The 16 MHz oscillator clock is first passed through control logic circuit (D2,
D16, and D19), which is for wait state control, and fed to the CPU.

Because of slowness of the some internal peripheral circuits, their bus


(IODATA) is separated from the high speed bus (DATA) by D18. Function of
real time clock is included in battery back-up SRAM (D4).

Software features are described in the Operator’s Manual and main differences
between software revisions in Section 3.4.

CAUTION: The SRAM IC (D4) contains a lithium battery. Danger of


explosion if IC is incorrectly replaced. Replace only with same or equivalent
type recommended by DATEX-ENGSTROM. Do not dispose of faulty IC:s in
fire. IC should be disposed of according to local regulations.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-41

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-42

Figure 5.19 CPU board jumper configuration and schematic


diagram (part 2)

CONNECTOR JUMPER MEMORY TYPE

The value of resistor network R5 is 4 x 47k.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-43

5.7 Video ASIC Board


The video ASIC board replaces video control board from revision
04 on. ASIC board includes ASIC IC and some other components.
Due to the number of components we recommend changing the
complete board in case of failure.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 5.20 Video ASIC board parts layout

Figure 5.21 (on the next page)


Video ASIC board schematic diagram

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-46

5.8 Power Supply Board


The primary of the power supply is designed to double insulation
requirements for added safety. There are two fuses. The primary
operating voltage is factory selected by insulating and folding the
unused primary leads inside the additional insulation tube.

The mains transformer is magnetically shielded to minimize screen


disturbance.

The power supply board contains basically four DC sources:

+5 V switched, for the digital circuitry.

+15 V switched, for valves, pump and other components.

+/-15 V regulated for the analog amplifiers.

Data retention voltage generation circuit supplies +5 V DRV


voltage for memory from switched +5 V supply.

Also, +12 V/1 A for the CRT unit and serial drivers/receivers is
derived from the +15 V switched voltage. The -12 V for the serial
I/O is derived from -15 V.

In addition to the power supply functions the board contains


drivers for two serial channels (including the modem control
signals CTS and RTS), a RESET control, which generates a 200 ms
reset pulse to the CPU if the +5 V line goes below 4.75 V, and
miscellaneous I/O functions like a loudspeaker driver. Some signals
from the mother board are passed directly to the rear panel
connectors.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-47
Figure 5.22 Power supply board block diagram
March 22nd, 1996/4
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-48

Figure 5.23 Transformer diagram and power supply board signal


waveforms

March 22nd, 1996/4


DATEX CARDIOCAP TM II CH-SERIES SERVICE MANUAL 5-49

Figure 5.24 Power supply board parts layout and schematic


diagram (part 1)

Figure 5.25 (on the next page)


Power supply board schematic diagram (part 2)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-50

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-51

5.9 Mother Board


The mother board contains mainly the system bus interconnections
and connectors. Also on the board are power bypass capacitors and
driver transistors for the sampling system magnetic valves (gas zero
and pressure valves; not used in Cardiocap II CH-series monitors).

For signals in each bus, see Table 4.7.

5.10 Keyboard
The keypad pcb is a simple 4x4 matrix which is scanned by the
keyboard controller on the CPU board.

5.11 Loudspeaker Unit


Audible alarms and beeps are generated by a separate loudspeaker
unit. It contains an 8 ohm/0.4 W speaker and the associated driving
circuitry.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-52

Figure 5.26 Mother board parts layout and schematic diagram


(board modification level 3 and higher)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-53

Figure 5.27 Keyboard parts layout and schematic diagram

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 5.28 Loudspeaker unit parts layout and schematic diagram

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-55

5.12 Internal connector configurations

Table 5.1 Video ASIC board (X1) - Mother board (X1)

NC = not connected
AIN is an AD-converter in and AOUT is a DA-converter out
ADC is an AD-converter and DAC is a DA-converter

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-56

Table 5.2 Main CPU board (X1) - Mother board (X2)

NC = not connected
AIN is an AD-converter in and AOUT is a DA-converter out
ADC is an AD-converter and DAC is a DA-converter

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-57

Table 5.3 SpO2 processor board (X1) - Mother board (X3)

NC = not connected
AIN is an AD-converter in and AOUT is a DA-converter out
ADC is an AD-converter and DAC is a DA-converter

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-58

Table 5.4 NIBP board (Xl) - Mother board (X4)

NC = not connected

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-59

Table 5.5 Power supply board (X10) - Mother board (X5)

NC = not connected
AIN is an AD-converter in and AOUT is a DA-converter out
ADC is an AD-converter and DAC is a DA-converter

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-60

Table 5.6 SpO2 measuring board (X2) - SpO2 processor board

Table 5.7 Keyboard (X1) - Mother board (X8)

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-61

Table 5.8 ECG board - Mother board (X12)

Table 5.9 IPT board - Mother board (X10)

Table 5.10 Power supply board - Mother board (X6)

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Table 5.11 ECG board - RESP board (X7)

Table 5.12 Mother board test connector (X13)

Table 5.13 Video ASIC board (X2) - Video unit main


pc board (X13)

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-63

Table 5.14 Front panel SpO2 connector - SpO2


measuring board (X1)

Table 5.15 Power supply board (X1) - Line transformer

Table 5.16 Power supply board (X3) - Fan

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-64

Table 5.17 Power supply board (X6) - Video unit main pc board

Table 5.18 Power supply board (X8) - Loudspeaker

Table 5.19 NIBP board (X2) - NIBP Pump

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-1

6 SERVICE AND TROUBLESHOOTING

6.1 General service information


Usually field service is limited to replacing the faulty circuit
boards or mechanical parts. The boards are then returned to
DATEX-ENGSTROM for repair.

DATEX-ENGSTROM is always available for service advice. Please


provide the unit serial number, full type designation, program
revision (displayed at monitor startup) and a detailed fault
description.

NOTE: After any component replacements see Section 7.1


(Adjustments) and after any service perform the functional field
check procedure in Chapter 8.

CAUTION: The tests and repairs outlined in this section should


only be attempted by trained personnel with the appropriate
equipment. Unauthorized service may void warranty of the unit.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-2

6.2 Disassembly and reassembly


The monitor is disassembled in the following way (see page 3-2 for
the parts included in the monitor you are using). See Figure 9.1 for
the exploded view of the monitor:

Disconnect the power cord.

Remove the upper two galvanized screws on the side panels


and lift off the top cover.

Remove the lower two screws and detach the side panels.

Now the interior of the monitor is exposed. The pneumatic unit for
the NIBP measurement is located on the left side and printed
circuit boards are located on the right side of the monitor.

Video screen frame can be removed. Keyboard pc board


with plastic key panel is attached to the front plate. It can
be removed by detaching the ECG lead selector knob and
disconnecting X8 connector on the mother board.

Behind the Keyboard pc board an assembly of pc boards is


located. They are (front to rear) the ECG board, IPT board,
a supporting board, and the Respiration board. The
connector board (no.2) with pressure and temperature
connectors is attached to the IPT board, beneath the ECG
board. This whole assembly can be detached by removing 6
supporting screws on the front plate (two on top and one on
each side of two connector openings) and disconnecting
X10 and X12 connectors after the Keyboard pc board is
removed. Each board can now be detached from the
assembly by removing the screws and spacers.

Four boards are attached to the mother board on the right


half of the chassis: (moving front to rear) the NIBP
processor board, the SpO2 processor board, the CPU board,
and the Video ASIC board. These boards are removed by
pulling upwards.

Beneath the mother board the intermediary bottom plate


houses SpO2 measuring board.

Power supply board is attached to the rear plate with four


screws.

Transformer, power cord receptacle block, loudspeaker, fan


and dust filter, fuse housing(s), NIBP cuff connector, and
ECG test signal connector are all attached to the rear panel.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-3

The rear panel can be tilted back by first detaching the power cord
receptacle block and removing the two screws beside the fan which
are keeping the rear panel in the upright position and sliding the
panel in sideways.

The main PC board for the picture tube, NIBP pump and
vertically positioned supporting plate, and NIBP safety
valve are attached to the bottom plate with screws.

Two magnetic valves, damping chambers II and III, and a choke


are screwed to the horizontally positioned plate, which in turn, is
attached to the vertically positioned supporting plate.

CAUTION: When the pneumatic unit is disassembled, make sure


that the tubes are reconnected properly.

The picture tube unit is screwed to the front plate.

Most parts can be removed by loosening the appropriate screws.

Reassembling is essentially reversing what was described


above.

CAUTION: When attaching the top cover, make sure that the
tubes and cables are not pinched between the boards and the
cover.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-4

6.3 Troubleshooting

6.3.1 Monitor start-up sequence


After the monitor is switched on

Fan on the rear panel starts to run.

The self test text appears on the screen. All the memory in
the monitor is checked. Program code and the date of its
completion are also displayed.

Within a half minute fields for waveforms and numeric


values will appear as horizontally running straight lines and
dashes on the screen. The real time clock also appears. If
SpO2 sensor is not connected to the monitor, the message
“no probe” appears. If the sensor is connected to the
monitor but not to the patient, the message “probe off
appears. When the sensor is connected to the patient, the
message “pulse search” appears and shortly pulse waveform
appears on the screen.

The NIBP software is performing its own selfcheck. If the


selfcheck is completed with no error detected, the program
code for NIBP appears in the lower left corner of the
screen.

The monitor is ready for use.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-5

6.3.2 Troubleshooting in general

NOTE: Please read “Troubleshooting and Displayed Messages” in


the Operator’s Manual first. The part of this chart is duplicated
from it.

Table 6.1 General troubleshooting chart

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-6

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-7

6.3.3 ECG board troubleshooting


In troubleshooting the ECG board it is important to note that the
isolated ground has no connection to the system electronics ground.
Also, if the problem is excessive high frequency noise, it may be
that altering the normal ground by connecting an oscilloscope
ground to the isolated ground may cancel the noise.

When checking ECG board problems, confirm that the isolated


power (±12 V) is available and stable. The waveform at the
secondary of T1 should be 100 kHz square wave, 50 % duty cycle
and 25 - 30 Vpp.

Table 6.2 ECG board troubleshooting chart

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-8

6.3.4 RESP board troubleshooting

Table 6.3 RESP board troubleshooting chart

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

6.3.5 IPT board troubleshooting

Table 6.4 IPT board troubleshooting chart

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-10

6.3.6 NIBP board troubleshooting


The NIBP board has its own diagnostics software that tests the
NIBP hardware every time the monitor is switched on. Should
there be any trouble with the NIBP, the monitor should be
switched off and on again and check for a display of a NIBP error
message. If the NIBP unit is ok it will display the NIBP VER X.X.
message.

Table 6.5 NIBP board troubleshooting chart

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-11

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-12

6.3.7 SpO2 measuring electronics troubleshooting


In troubleshooting the measuring board it is important to note that
the isolated ground has no connection to the system electronics
ground. Also, if the problem is excessive high frequency noise, it
may be that altering the normal ground by connecting an
oscilloscope ground to the isolated ground may cancel the noise.

When checking measuring board problems first confirm that the


isolated DC power (+12 V, -12 V, and +5 V) are present and are
stable.

The troubleshooting chart of the measuring electronics is shown in


Table 6.6.

For the parts layout and schematic diagram of the SpO2 measuring
board see Section 5.

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-13

Table 6.6 SpO2 measuring electronics troubleshooting chart

Finger in probe and connected to CARDIOCAP™

*) see Figure 5.14 for signal waveforms.

NOTE: When doing any work on the measuring board care has
to be taken that the patient isolation is not violated.

NOTE: When reassembling the measuring board see Figure 9.1


for the proper placement of the shields.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-14

6.3.8 SpO2 processor board troubleshooting


Due to the complexity of the LSI circuitry there are few faults in
the CPU digital electronics that can be located without special
equipment. The following checks may, however, be performed:

All integrated circuits are properly inserted into the sockets.

The 11.059 MHz clock signal at the CPU pins 18 and 19


(use a high impedance sensor to check).

PSEN (CPU pin 29) shows that instructions are being


fetched. If this line is static, the processor is not running.

RESET (CPU pin 9) is normally low but pulled up to +5 V


for a moment after power up. If RESET is constantly high,
check the +5 V supply line for spikes or too low level.

The analog part is easier to debug. The input analog voltages are
easily tracked to the MUX input. After the MUX, the voltages are
multiplexed so that the resulting waveform frequency is at 200 -
600 HS.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-15

6.3.9 CPU board troubleshooting


Due to the complexity of the LSI circuitry there are few faults in
the CPU digital electronics that can be located without special
equipment. The following checks may be performed:

The RAM, EPROM, CPU, and other socketed I.C.‘s are


properly installed and the memory configuration jumpers
are correct (see Figure 5.19).

The 16 MHz (revision 02 and up) or 11.059 MHz clock


signal at the CPU pins 18 and 19 (use a high impedance
sensor to check).

PSEN (CPU pin 29) shows that instructions are being


fetched. If this line is static, the processor is not running.

RESET (CPU pin 9) is normally low, but pulled to +5 V


for a moment after power up. If RESET is constantly high,
check the +5 V supply line for spikes or low voltage.

The analog sections are easier to troubleshoot. The input analog


voltages are easily tracked to the MUX input. After the MUX, the
voltages are multiplexed so that the resulting waveform frequency
is at 200 - 600 HS.

All analog output channels are updated every 40 ms. Thus the D/A
conversion interval is 5 ms.

6.3.9.1 Instructions after replacing the software or CPU board


After replacing the software, the CPU board or the memory IC D4
on the CPU board the monitor displays “MEMORY CIRCUIT
FAILURE / CALL SERVICE”. The content of D4 is incorrect.
Performing factory reset will set the content correct.

Perform factory reset in the engineering menu (see the


Operator’s Manual Section 2).

Re-establish previously used settings or inform the monitor


user that all other settings are default values.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-16

6.3.10 Video ASIC board troubleshooting


Due to the number of components in video ASIC board, we
recommend changing the complete board in case of failure.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-17

6.3.11 Power supply board troubleshooting


The following troubleshooting chart will help in pinpointing
a malfunctioning component. The only part that requires good
understanding of operating principles are the switching supplies.
Consult the IC data sheet for more information on the LM 3524
switcher controller.

Table 6.7 Power supply board troubleshooting chart

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-18

6.3.12 Mother board/Keyboard troubleshooting


Fault finding on the mother board/Keyboard is reduced to:

visual inspection of board surface and connectors

continuity and short circuit testing with an ohmmeter

measuring of power supply voltages.

6.3.13 Loudspeaker /troubleshooting


Loudspeaker can be evaluated by checking its supply voltage (red
wire), volume (black), and pitch (green) signals. The measurement
can be performed easily by detaching the loudspeaker from the
rear panel.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 6-19

6.4 Service mode


The service mode is designed originally for factory use. However
some simple checks can be done in the service mode to ensure the
proper operation of the monitor without having to disassembly it.

Enter the service mode as follows:

Turn on the monitor and allow it to go into normal


monitoring mode.

Press the softkeys 1, 2, 3, and 4 in order and then


immediately press and hold down the NORMAL SCREEN
key

The character ‘S’ blinks and the text ‘SERVICE MODE ON’
appears to indicate that you have entered the service mode.

Press soft key 4. The service mode menu will appear.

The menu is:


STAND BY
INFO
SERVICE OFF

Press the soft key 1 to move the box from selection to


another. Press the NORMAL SCREEN key to enter the
menu.

In STAND BY, normal monitoring can be continued even


though you are in service mode.

In info, the revision and the date of the program, the date
of the last factory reset performed and a list of observed
faults can be seen.

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-1

ADJUSTMENTS

7.1 ECG board adjustment


The only adjustment on the ECG board is the 50/60 Hz notch.
This is also the only difference between 50/60 Hz monitors.

To adjust the notch, feed a 1 - 10 mV 50/60 Hz signal into the


ECG input, between LA and RA in LEAD 1 configuration, either
through a resistive divider from system bus pin 31c (22 Vac) or
from a signal generator. Adjust R48 (accessible from top of the
ECG board) to minimize the displayed amplitude (fast speed).

WARNING: When doing any work on the ECG and IPT boards,
care has to be taken that the patient isolation is not violated.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-2

7.2 RESP board adjustment

Test equipment

Oscilloscope 10 mV/div.

Frequency counter

Oscillator adjustment

Connect oscilloscope and frequency counter to test point TP1 (pin


1 in connector X2). A nearly sinusoidal signal of amplitude almost
+10 V should be seen on the oscilloscope.

Adjust the frequency to 48.3 kHz ±100 Hz with trimmer R54.

LEADS OFF threshold adjustment

Connect 3 kohm resistor between yellow and black patient cables.

Turn trimmer R53 counter-clockwise until the message LEADS


OFF appears on the screen of the host monitor. Turn the trimmer
carefully and slowly clockwise until the message disappears.

Remove the 3 kohm resistor and make sure that the message
returns to the screen.

Connect the cables together and see that the message disappears.

Confirm that the message does not appear when a resistor whose
value is less than 3 kohm is connected between the cables.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-3

7.3 IPT board adjustment


The only adjustments on the IPT board are the temperature gain
set trimmers, which are adjusted as follows:

Set the 38°C TEMP CAL via the engineering menu. To


enter the engineering menu keep MARK RESET softkey
depressed during power up until the menu appears. With
temperature measuring jack disconnected choose
CALIBRATE and TEMP CAL, depress the TEMP TEST
button and adjust the reading to 38.0°C.

Insert a plug with a precision resistor listed below into the


jack receptacle and adjust R11 to show the correct
temperature. The trimmers are accessible from above
without removing the IPT-ECG board assembly.

YSI 400 series temperature-resistance table:

°C Ohms

15 3539
20 2814
25 2253
30 1815
35 1471
38 1301
40 1200
44 1024

NOTE: For temperature gain calibration use a precision resistor


between 3539 and 2000 ohms.

WARNING: When doing any work on the ECG and IPT boards,
care has to be taken that the patient isolation is not violated.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-4

7.4 NIBP board adjustment


The NIBP board automatically performs all adjustments except
pressure calibration, which must be done manually. Pressure zero is
automatically maintained by the processor. If the zero point drifts
too much NIBP transmits an error message. Normally zero point
trimmer R6 should not be touched.

CALIBRATION

NIBP calibration can be performed as follows:

Enter engineering menu by pressing MARK RESET key


during power up. Keep the key down until the
“STARTING” sequence has been completed.

Press soft key 2 to enter calibration mode.

Press soft key 2 again to enter NIBP calibration.

In the calibration mode, the automatic zero adjustment is


not operating to allow the adjustment of the pressure
channel zero trimmer R6 (upper trimmer on the NIBP
board). Adjust R6 until the mean pressure value M is zero.
Make sure that the upper pneumatic connector is not
connected to a pressure source to allow proper zeroing.

Remove the female connector from an adult cuff and


connect it to the male connector in the twin hose. Connect
an external pump and mercury manometer to it and pump
the pressure up to 200 mmHg and adjust the calibration
trimmer R11 (lower trimmer on the NIBP board) until the
mean value M shows 200.

Repeat steps d) and e) until zero and 200 mmHg readings


are both correct.

Check the linearity of the pressure transducer at the


pressure of 100 mmHg.

Press “RETURN TO MONITOR” to enter normal operating


mode.

NOTE: In the normal operating mode the zero pressure drift is


compensated by software and this drift does not affect the
accuracy of the device. If zero drift is more than ±20 mmHg
NIBP issues an error message, see troubleshooting chart.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-5

7.5 SpO2 processor board adjustment


There are only two adjustments on the SpO2 processor board:

A/D-converter zero (R9).

A/D-converter 10 V reference (R8).

The A/D-converter adjustments are performed only at the factory.

7.6 CPU board adjustment


There is one adjustable capacitor on the CPU board for real time
clock oscillator frequency. It is factory adjusted, but may need to
be readjusted if some components are replaced.

Real time clock oscillator frequency

The real time clock oscillator frequency is set by connecting a


counter to pin 15 of D16 and adjusting C49 to read 256.00 Hz
corresponding to a cycle time of 3906.25 us.

NOTE: The high speed CPU board does not require any
adjustments.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 7-6

7.7 Video Display Unit (Philips) Adjustments


The following adjustments can be done in the Philips CRT display
board:

Basic brightness adjustment Trimmer potentiometer P4


Horizontal width Trimming indutance L1
Vertical height Trimmer potentiometer P2
Horizontal positioning Trimmer potentiometer P1
Horizontal linearity Trimmer magnet L2
Focus Trimmer potentiometer P3

Figure 7.1 Video Display Unit Adjustment

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 8-1

8 FUNCTIONAL FIELD CHECK PROCEDURE

8.1 Preoperative check list


Perform the following tests daily to assure proper operation of the
monitor.

WARNING: If the monitor fails to respond as described, stop


using the monitor, troubleshoot, and correct the situation.

Power up

Turn the power on. The monitor performs a start-up and self test
procedure.

Confirm that no error messages appear.

ECG Check

Plug an ECG cable into the ECG connector receptacle on


the front panel.

Connect the ECG lead wires to the clip-on connectors on


the monitor’s rear panel according to the letter codes.

Turn the lead selector to position I.

The ECG waveform should display a train of 1 mV pulses


with a frequency of 60 ±1 bpm.

Confirm that the QRS indicator ‘*’ is flashing to the left


and above the ECG channel.

Remove the ECG leads from the test pins. Ensure that the
message ‘LEADS OFF’ appears in the upper right corner of
the screen.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 8-2

NIBP Check

Assemble the blood pressure cuff with the cuff hose and
connect it to the NIBP connector on the rear panel.

Attach the cuff to your arm.

Take a blood pressure measurement by pressing the START


key. Confirm that the pump inflates the cuff. Confirm that
the cuff pressure is displayed during measurement and that
final systolic, diastolic, and mean values are displayed.

SpO2 Check

Connect the pulse oximeter sensor to the monitor. The


message ‘NO PROBE’ is replaced by ‘PROBE OFF’.

Confirm that the red light is seen from the sensor, and that
the sensor surface is undamaged.

Attach the pulse oximeter sensor to your finger.

Wait for the pulse search to be completed and readings to


stabilize. Verify that the SpO2 readings are in the expected
range and that the plethysmographic pulse waveform is
regular.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 8-3

8.2 Checks after component replacements


The following test should be performed after any service to ensure
proper operation of the monitor.

Check visually that tubes, valves, and pump are properly


connected. Tubes are not pinched, clogged, or jammed.
There should be no sharp bends either.

Connect power cord and switch the monitor on.

Check that no error messages are shown.

Check the operation of the real time clock:


- Verify the date (use engineering menu).
- Verify time.
- Check that the clock is running.

Connect the ECG cable to the rear panel test connectors.


Verify that the test signal is shown on the display. Heart
rate display should read 60 ±1.

Connect the invasive blood pressure transducer(s) to the


front panel connector(s) PRESSURE 1 (and 2). Zero the
sensor(s) by pressing PRESSURE key and soft key 1 (and
2). Verify proper operation by connecting a manometer and
hand pump in series with the transducer(s). Check
calibration at 200 mmHg pressure.

Connect the NIBP cuff. Perform one measurement with


yourself. Verify operation. Perform venous stasis (see the
Operator’s Manual Section 7) and check for leaks. Check
NIBP calibration (see Section 7.4).

Check temperature channel by pressing the TEMP TEST


switch. Reading should be 38.0 ±0.1°C (100.4 ±0.2°F).

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 8-4

Connect the SpO2 sensor and observe the message PULSE


SEARCH followed by PROBE OFF. Put the sensor on your
finger and keep the hand still.

The message PULSE SEARCH should appear first.

A numeric value should appear next in the SpO2 field


(normally between 94 and 98 in non-anesthetized
people).

The plethysmographic pulse waveform from SpO2


sensor should be shown on the screen,

Keep the SpO2 sensor on your finger for several minutes.


Press the TRENDS key and check that the measurement
results are enough to draw trend waveform.

Check the operation of the loudspeaker by creating an


alarm state.

Connect serial printer to serial & analog I/O connector on


the rear panel of the monitor. Turn the power on and check
that the printer prints data correctly.

Reset the monitor by pressing the MARK RESET key for


about five seconds.

Check the leakage current of the unit. Check the condition


of all cables and connectors.

Disconnect power cord for 30 minutes. Then connect it


back and turn the monitor on. When the initialization is
over, make trend displayed on the screen. The trend which
had been measured before the power cut-off should be
displayed.

Check that there are no dirt stains in the monitor. Top


cover, side panels, and bottom plate are properly attached
with screws.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 8-5

8.3 Preventive maintenance check list


We recommend that you perform these checks at least once every
six months to keep the monitor in good working condition.

1. VISUAL INSPECTIONS

Remove the top cover and check internal tubing


system for sharp bends and dirt.

Rear panel dust filter (clean at least once a month).

Screen for distortion (adjust if necessary).

Software version numbers (update if necessary).

2. MAIN FUNCTIONAL CHECKS

Check ECG functions (see Section 8.2 step e)).

Check invasive blood pressure measurement


(see Section 8.2 step f)).

Check NIBP operation (see Section 8.2 step g)).

Check temperature (see Section 8.2 step h)).

Check Pulse oximeter functions


(see Section 8.2 step i)).

Check date in the engineering menu.

Check real time in SETUP menu.

Check the function of the loudspeaker in SETUP menu.

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-1

9 SPARE PARTS

9.1 Spare parts


For the locations of the main parts see Figure 9.1.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-2

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-3

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-4

In your rear panel or front panel order, please specify the serial number and the
full monitor code as found in the type shield on the rear panel.
The part is recommended for stock.
The part is sold as an accessory.

March 22nd, 1996/4


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-4a

9.2 Service accessories

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-5

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-6

Figure 9.2 ECG, IPT, and RESP board assembly

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 9.3 Pneumatic unit parts layout

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-4a

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-4a/2

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 5.3 ECG board parts layout and schematic diagram


(CH-R, CH-1R, CH-RS, CH-1RS)
(board modification level 4 and lower)

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-5/2

Figure 5.3 ECG board schematic diagram (part 2)

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-5a

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-5a/2

Figure 5.3 ECG board schematic diagram (part 2)

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 5.5 RESP board parts layout and schematic diagram


(board modification level 2 and lower)

November 15th, 1990


September 2nd, 1991/1
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-15

Figure 5.7 IPT board (board modification level 4 and higher)


and IPT connector board parts layouts and IPT
connector board schematic diagram

Figure 5.8 (on the next page)


IPT board schematic diagram
(board modification level 4 and higher)

(Connector X4 or both X3 and X4 not included in some models)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-32
Figure 5.12 SpO2 measuring board block diagram, parts layout
and timing diagram
Figure 5.13 (on the next pages)
SpO2 measuring board schematic diagram
December 10th 1993/3
DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33/2

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33/3

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33a/2

September 2nd, 1991/1


DATEX CARDIOCAPTM II CH-SERIES SERVICE MANUAL 5-33a/3

September 2nd, 1991/l


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33b/2

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-33b/3

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-34

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-37

Figure 5.16 SpO2 processor board parts layout and schematic


diagram (modification level 4 and lower)

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL

Figure 5.16 SpO2 Processor board schematic diagram (part 2)

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-37a

Figure 5.16a SpO2 processor board parts layout and schematic


diagram (modification level 5 and higher)

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-37a/2

Figure 5.16 SpO2 Processor board schematic diagram (part 2)

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-44

Figure 5.20 Video control board block diagram, parts layout and
timing diagram
(board modification level 9 and higher)

Figure 5.21 (on the next page)


Video control board schematic diagram
(board modification level 9 and higher)

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 5-50/2

September 2nd, 1991/1


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-5

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 9-5

November 15th, 1990


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-15

Figure C.5 Keyboard parts layout and schematic diagram

March 20th, 1991


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-16

Figure C.6 Control board parts layout and troubleshooting chart

Figure C.7 (on the next page)


Control board schematic diagram

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-17

December 10th 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-20

Figure C.9 Exploded view of CCK-104 Keyboard

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-1

12 APPENDICES

COMPUTER OUTPUT (Serial & Analog I/O Connector)


Transmission rate 1200 Baud. One series of output strings in 10 seconds except
NIBP string which comes out immediately after NIBP measurement. 8 data bits,
no parity. 1 start bit and 1 stop bit. Each parameter has 3 digits. Parameters are
separated by commas.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-2

If DATEX-ENGSTROM gas monitor CAPNOMAC™ II is connected to


CARDIOCAP™ II monitor (see the preceding page for NIBP string):

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-3

If DATEX-ENGSTROM gas monitor CAPNOMAC ULTIMA™ is connected to


CARDIOCAP™ II monitor (depending on the Ultima in use, some strings may
not be available:

(See the preceding pages for NIBP, SpO2, and basic strings).

U01 string U03 string U04 string

String type String type String type

Marker no. MV exp (0.1 1/min) ETCO 2

ETO2 (% x 10) MV insp (0.1 1/min) FiCO 2

FiO2 (% x 10) TV exp high (1) CO 2 unit

ET balance gas TV exp low (ml)

Fi balance gas TV insp high (1)

ASX status TV insp low (ml)

Automatically V1.0 (%) for adults or


identified gas V0.5 (%) for paediatrics
Reserved Compliance

ET (primary gas, 10=1%) Ppeak

ET (secondary gas) Pplat

Fi (primary gas) PEEP

Fi (secondary gas) I:E

MAC (x 10) Reserved

Pulse rate Reserved

SpO2 Reserved

Ired ampl. Reserved

Red ampl. Recorder output

Gain Units status

Status Status

U03-string comes out when CAPNOMAC ULTIMA™ monitor includes V-option.


See operator’s manual for strings’ interpretation.
The status data are followed by: CR (Carriage return, hex D) LF (Line feed, hex A)

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-3a

The Status Characters of the Standard String

*) Mean alarm when p2 scale = 12.5 mmHg

December lOth, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-3b

Status Code Interpretation

For example:

740

7 (first character) indicates that the status code is about Pleth/SpO2


status and alarm (see the table on the preceding page).

4 (second character) indicates bit b2 is active (see the table above).


It is “probe off” according to the table on the preceding page.

0 (third character) indicates that there is no error among the items


listed in the third character box in the same table.

So the status code 740 indicates “probe off”.

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-4

Alarm Activation/Deactivation String

Alarm activation string appears immediately when an alarm is


given.

CARDIOCAP™ monitor alarm


activation string identification
Alarm parameter

Alarm activation value

< or > depending on whether low


or high limit
Limit

CARDIOCAP™ alarm deactivation


string identification
Alarm parameter

*) When Capnomac Ultima™ is connected

Command/Info String

CARDIOCAP™ command/
info string identification
RES for reset, POW for power-up

Program code

December 10th, 1993/3


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-5

B CCP-104 GRAPHICS PRINTER


The CCP-104 is an HP ThinkJet graphics printer. Earlier printers
(CCP-104-xx-00, see the type plate on the bottom of the unit)
require an isolated connecting cable (P/N 873152) to satisfy
medical electronics safety requirements. Printers currently in
production (CCP-104-xx-01) are IEC 601-1 class II devices and do
not require an isolated cable. The non-isolated connecting cable’s
part number is 875370. Due to mechanical changes the isolated
connecting cable can not be connected to the current printers.

For troubleshooting the printer please refer to the HP ThinkJet


service manual.

See figure below for the correct DIP switch settings in the printer.

March 20th, 1992/2


DATEX CARDIOCAP™ II CH-SERIES SERVICE MANUAL 12-6

C TRILINGUAL LIST OF ERROR MESSAGES


Listing of error messages in English, German and French.

English version German version French version

ARTIFACT ARTEFAKT ARTEFACT

CHECK PROBE CHECK SENSOR VERIF CAPTEUR

CUFF LOOSE MANSCH. LOCKER BRASSARD LACHE

CUFF OCCLUSION MAN. OCCLUSION OCCL BRASSARD

EXTERNAL RAM EXTERN. RAM ERREUR DE RAM


ERROR FEHLER

INVASIVE BP INVASIVER DRUCK PRESS. INV.


NOT CALIBRATED, NICHT KALIBRIERT, NON ETALONNEE,
RECALIBRATE NEU-KAL RE-ETALONNER

LEADS OFF EKG AUS ECG INOP

LOW PLETH SIGNAL GERINGES PULS- SIGNAL PLETH PEU


SIGNAL FIABLE

MEMORY CIRCUIT SPEICHER-FEHLER/ ERREUR CIRCUIT


FAILURE/CALL SERVICE MEMOIRE/APPELER
SERVICE ERFORDERLICH SERVICE

NIBP AIR LEAKAGE NIBD UNDICHT FUITE AIR BRAS

NIBP ARTIFACT NIBD ARTEFACT ARTEFACT PNI

NIBP HW ERR X NIBD HW FEHL X PNI HW ERR X

NTBP ZERO ERROR NIBD NULL FEHLER ERREUR 0 PNI

NO NIBP KEIN NIBD PAS DE PNI

NO PROBE KEIN SENSOR PAS DE CAPTEUR

PROBE OFF SENSOR LOSE CAPTEUR DECONN.

ROM CHECKSUM ROM CHECK ERREUR DE ROM


ERROR FEHLER

TEMPERATURE NOT TEMPERATUR NICHT TEMPERATURE NON


CALIBRATED, KALIBRIERT, ETALONNEE,
RECALIBRATE NEU-KALIBR. RE-ETALONNER

December 10th, 1993/3

You might also like