Savina

Technical Service Bulletin # 8

Re: New software version 2.10 and new Operating Instructions

Date: May 12, 2003

Reference Doc: N/A

Reason: Draeger Medical AG & Co. KGaA has released software version
2.10 which includes the following improvements:

♦ Improved battery management


- Battery capacity display. The battery symbol in the display shows the
remaining capacity of the internal battery.
- Energy saving mode. The speed of the blower is reduced when running on
internal battery.
- Improved Alarm concept of internal batteries
“Caution” when remaining capacity is below 30%
“Warning” when remaining capacity is below 10%
♦ Lung mechanics: Measurement of static Compliance and Resistance

♦ Improved synchronization between patient and ventilator:


Automatically leak compensated trigger criteria for in- and expiration

♦ Extended flow sensor lifetime


“Hot calibration” will be carried out only if required.

♦ Non invasive ventilation via face mask as an option

Devices affected: The software update with software version 2.10 is mandatory for all
Savina units in the USA and Canada.

When: By September 30, 2003

Cost: Free of charge

Ordering Info: SW 2.10 Kit Savina P/N 8415594

5664.900 Savina TSB No. 8 05.2003 1/3


Reporting: Feedback for DrägerService (direct) will be captured using the
DrägerService Dispatch System. All other distributors will be
required to complete the attached Mandatory Action Completion
form and either fax or mail to the following:

Mail: Drager Medical Inc.


3122 Commerce Drive
Telford, PA 18969
Attn: Regulatory Affairs Dept.

FAX: (215) 721-5784 Attn: Tony Achey

Additional Info: Replace all US Operating Instructions P/N 9037222 with new
Operating Instructions 2.1n P/N 9037626, included in the SW 2.10
kit.

Distribution: DrägerService Personnel and authorized Service Organizations for


CCS products. ( Lifetronics, Brathwaites, ABS Medical, Freedom
Medical ).

Addition Info 1: After downloading SW 2.10 you will see an error 65.0020 on the initial
power up. This is to be expected after download of SW 2.10. Enter
DS mode once and cycle power and the error will be cleared.

Additional Info 2: After downloading SW 2.10 the following steps must be performed:

1. Verify that valve classes have remained accurate (test 15)


2. Reset/verify ambient pressure (test 27)
3. Perform insp/exp airway pressure sensors (test 23)
4. Perform O2 calibration (test 22)
5. Perform PMC of unit.
6. Clear error log.
7. Leave updated manual in place of old version manual. Enter s/n of Savina
in block on back page of manual.

NOTE: Steps 1 through 4 in Additional Info 2 are executed via the


Internal Service Mode. See “Service Mode” section of Savina service
documentation on CD-ROM.

If you have any questions, please contact Technical Support by


phone at 1-800-543-5047 or by fax at 1-215-721-5789.

Dräger Medical, Inc.


Technical Product Manager

5664.900 Savina TSB No. 8 05.2003 2/3


Savina
Mandatory Field Correction Form
TSB # 8

Device: Savina
Part Number: 8413600

Location Of Device(s):______________________________________________________
(include address) ______________________________________________________
______________________________________________________
______________________________________________________

Device serial numbers: (list all devices at location that have been completed)

1._______________ 6._______________ 11.______________


2._______________ 7._______________ 12.______________
3._______________ 8._______________ 13.______________
4._______________ 9._______________ 14.______________
5._______________ 10_______________ 15.______________

The above listed devices have been modified according to the instructions as listed in
Savina Technical Service Bulletin # 8.

Mail or Fax to:

Mail: Drager Medical Inc.


3122 Commerce Drive
Telford, PA 18969
Attn: Regulatory Affairs Dept.

FAX: (215) 721-5784 Attn: Tony Achey

Name:_____________________________ENG#__________Date:__________________

5664.900 Savina TSB No. 8 05.2003 3/3