Memmert CO2-incubator ICO105 EN
Memmert CO2-incubator ICO105 EN
Memmert CO2-incubator ICO105 EN
CO2 incubator
ICO105
On this page, you can find all the essential technical data on the Memmert CO2 incubator ICO. Our customer
relations team will be pleased to help if you want further information. If you should require a customised special
solution, please contact our technical specialists at sales@memmert.com.
Page 1/4 | 02.01.2023 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail info@memmert.com | 0
DS_Memmert CO2 incubator ICO105 - en
Temperature
Working-temperature range 5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an
error
Humidity
Humidity control (standard) Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element
limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Adjustment range 1 to 20 % O2
O<sub>2</sub>
Control technology
ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition
TFT-colour displays.
Function SetpointWAIT the process time does not start until the set temperature is reached
Adjustable parameters temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime
Page 2/4 | 02.01.2023 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail info@memmert.com | 0
DS_Memmert CO2 incubator ICO105 - en
Communication
Interface Ethernet LAN, USB
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes
via Ethernet interface or USB port
Safety
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the
setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating
function is switched off in case of overtemperature, cooling function in case of undertemperature
Heating concept
6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid
condensation
Standard equipment
Works calibation certificate Works calibration certificate at +37 °C, 5 % CO2 and 90 % rh (requires option K7); standard equipment
for units with active humidity control
Works calibation certificate Works calibration certificate at +37 °C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option
T6); standard equipment for units with O2 control
Works calibation certificate incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard
units
Door fully insulated stainless steel door with 2-point locking (compression door lock)
Standard equipment CO2 connection set: hose with coupling and clamp
Standard accessories Membrane filter (in order to remove impurities and pollutants, all incoming gases pass through a
membrane filter before they reach the chamber)
Volume 107 l
Page 3/4 | 02.01.2023 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail info@memmert.com | 0
DS_Memmert CO2 incubator ICO105 - en
Electrical data
Voltage 230 V, 50/60 Hz
Electrical load approx. 1300 W
Ambient conditions
Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance
from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances
must not be less than 5 cm.
Ambient temperature 10 °C to 35 °C
Overvoltage category II
Pollution degree 2
Packing/shipping data
Transport information The appliances must be transported upright
WEEE-Reg.-No. DE 66812464
Page 4/4 | 02.01.2023 Memmert GmbH + Co. KG | Tel. +49 (0) 9122/925-0 | E-Mail info@memmert.com | 0