Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

CLX-Brochure 영문

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

ES-HF-CLX

Online Chlorine Monitoring


Job Name –––––––––––––––––––––––––––––––––––––––––––– Contractor –––––––––––––––––––––––––––––––––––––––––––––

Job Location –––––––––––––––––––––––––––––––––––––––––– Approval ––––––––––––––––––––––––––––––––––––––––––––––

Engineer –––––––––––––––––––––––––––––––––––––––––––––– Contractor’s P.O. No. –––––––––––––––––––––––––––––––––––

Approval –––––––––––––––––––––––––––––––––––––––––––––– Representative –––––––––––––––––––––––––––––––––––––––––

Reliable Chlorine Monitoring


Specifications
HF scientific developed the CLX OnLine Residual Chlorine
Monitor specifically for drinking water, wastewater, industrial and
marine residual chlorine monitoring applications.

The CLX OnLine Residual Chlorine Monitor is manufactured


using microprocessor based technology making it the most
reliable and cost efficient instrument for monitoring
residual chlorine.

Colorimetric DPD chemistry technology provides a proven,


stable and reliable reading along with the most economical
method available.

Features
• Colorimetric DPD Chemistry
Proven reliable and economical, colorimetric DPD
(N,N-diethyl-p-phenylenediamine) chemistry is the most accurate
method for measuring free or total residual chlorine.
• Rugged Enclosure
The CLX's small footprint and corrosion resistant NEMA
4X (IP66) enclosure allows for a simple installation that will
fit in almost any space.
• Optical Design
New optical design uses similar optics to the hand held portable
Chlorine Photometer. The new design increases accuracy and
allows for comparable and precise readings between the OnLine
Monitor and hand held portable instruments.
• Removeable Sample Cuvette
A removeable sample cuvette allows for simple cleaning and
maintenance. Viewable sample chamber allows a clear view of Changing the
the sample cuvette while the instrument is sampling. Sample Cuvette
• Cycle Times
User selectable cycle times allow the user to set the cycle time
from 110 seconds up to 10 minutes (optional 60 seconds to
10 minutes). Low volume reagent and sample chamber allow for
affordable reagent costs and decreases water consumption.
• Certification
Listed or Certified to CE, UL, CSA (ETL,ETLc)
Sample Specifications
The Residual Chlorine Monitor shall be able to measure free or Range: 0 - 10 mg/l.
total chlorine using the DPD colorimetric method. The Chlorine Cycle Time: User Selectable: 110 seconds to
Monitor shall have a measurement range of 0-10 mg/l. The 10 minutes. (optional 60 seconds to
Chlorine Monitor shall include the power supply, sensor and 10 minutes)
display in one NEMA 4X (IP66) corrosion resistant enclosure. The
Chlorine Monitor shall have readings consistent with laboratory Resolution: 0.01 mg/l
and portable Chlorine Photometers. The Chlorine Monitor shall Accuracy: 0 - 6 mg/l - ±5% or 0.03 mg/l of Cl2
have standard 4-20 mA outputs and RS-485 with Modbus whichever is greater.
Protocol. The 4-20 mA output shall be standard or reversible. 6 - 10 mg/l - ±10%
The Chlorine Analyzer shall have a removable sample cuvette.
The sensor shall have a clear view to the optical chamber Method: USEPA accepted DPD method of
allowing viewing of the sample and reaction of the reagents, analysis for measuring Free or
before, during and after a reading. Accuracy shall be ±5% Total Residual Chlorine
with a resolution of 0.01 mg/l for the range of 0-6 mg/l and Standard Outputs: 4-20 mA and RS-485 with Modbus.
±10% with a resolution of 0.01 mg/l for the range of 6-10 mg/l. User Alarms: 2 User Selectable Alarms for Sample
The Chlorine Monitor shall have a user selectable cycle time Concentration (Form C 240VAC 2A)
between 110 seconds and 10 minutes (optional 60 seconds to
10 minutes). The Chlorine Monitor shall be capable of 30 days Operating Temperature: 0°C to 40°C (32°F to 104°F)
of unattended operation. The Chlorine Monitor shall take a zero Input Pressure: 5 - 150psi
reading before each reading to compensate for color in the Enclosure: ABS Plastic, NEMA 4X, IP66
sample. The Chlorine Monitor shall have microprocessor based
technology. The Monitor shall have two user selectable alarm Power: 100 - 240 Volt Auto Switchable
relays. 47-63hz
Certifications: CE, UL, CSA, (ETL, ETLc)
Shipping Dimensions: 16" x 16" x 10"
(41cm x 41cm x 26cm)
Shipping Weight 8lbs (3.6kg)

Changing the Reagents

Ordering Information Catalog No.


Catalog No. Description
J.A.W.* Liquid
20040  LX OnLine Residual Chlorine
C
Monitor for Free or Total Chlorine Free Chlorine 30 day supply 09951 09947
Complete with operating manual. Re-
agents must be purchased separately. Total Chlorine 30 day supply 09952 09948
If the application combines high humidity

with cold sample water, purchase a *Just Add Water
#09944 desiccant cartridge separately. **User must add Sulfuric Acid, Soduim Hydroxide and supplied DPD
Accessories reagent before use
09950 Replacement tubing/cuvette kit
25018S Flow through cuvette CLX
28316 Low pressure adapter
Dimensions
All Dimensions are in millimeters (inches).

325.74
(12.82)

311.11
(12.25)

Front View

219.05 These 4 holes will


128.50
accept up to 8 MM
(8.625) (5/16 inch) hardware (5.06)

298.45
(11.75)

Rear View Side View


CLX with cover removed

LCD
Display

Latch
Sample Cuvette
Power Backlit - Reaction
fully visible even with
cover in place
Alarm Relays

RS-485 & 4-20 mA


Pressure
Regulator

Inlet Tube
Buffer Reagent
Indicator 1/4" NPT
Bottle
Solution
Drain Bottle
1/2 inch I.D.
drain tubing

USA: T: (239) 337-2116 • Toll-Free (888) 203-7248 • F: (239) 332-7643 • HFscientific.com


Latin America: T: (52) 81-1001-8600 • F: (52) 81-8000-7091 • HFscientific.com

ES-HF-CLX 1726 © 2017 HF scientific

You might also like