Silica Analyzer
Silica Analyzer
Silica Analyzer
Sodium
monitoring
for high
purity water
POLYMETRON 9245
Operation
The HACH ULTRA POLYMETRON 9245 mono-channel, sodium
analyzer is designed for ease of installation and use, requiring
The POLYMETRON 9245 only power connection and sample hydraulics. Set-up is
provides low-level managed through the readily accessible controller with clear
sodium measurement messages guiding the user through menus and sub-menus
in high purity water
applications. The POLYMETRON 9245 is fitted with a quick-loop valve at
the bottom of the analyzer, a constant head vessel and minimum
With a detection sample flow detection to allow for variations in sample pressure,
limit of 0.01ppb and a flow, and distance from the analyzer to the “fresh-sample” point.
range of 0 – 10,000ppb
this analyzer is The principle of operation is that of ion selective electrode
ideally suited for measurement after pH conditioning. Sample pH conditioning is
monitoring sodium essential for limiting the interference of temperature or other ions.
in demineralized Constant buffering is assured using regulated reagent addition
water, boiler feed, across ambient and sample temperature changes.
condensate To maintain optimum response time even in systems of continuous
and all parts of low sodium concentration the POLYMETRON 9245 incorporates
the steam/water cycle. automatic reactivation. Reactivation uses non-hazardous chemicals
and eliminates the need for manual reactivation and the use of
This on-line process hazardous solutions.
monitoring analyzer is
designed
for use in the power Calibration
generation industry
but it may also be Due to the difficulty in preparing low ppb sodium solutions for
applied in a number calibration purposes the POLYMETRON 9245 is supplied with
of other industrial an automatic calibration system. Using easy to prepare 10ppm
applications requiring samples the fully automated system follows a multiple step
effective sodium calibration cycle to accurately calibrate the analyzer, eliminating
monitoring, such as user variability and possible sample contamination.
process water
in the semiconductor
industry. Benefits
Q Measures sodium levels from 0-10,000ppb
with a detection limit of 0.01ppb.
Comprehensive information is available Q Warning messages (reagent level low, small calibration
at a glance from the large display, deviation),
including: 24 hour sample trend curve, Q System alarms (no reagents, no sample, no calibration,
bar graph tracker of the current analysis, no power supply).
sample name and alarms. Clear mes-
sages guide the user through menus
and sub-menus. A built-in data logger Maintenance
allows measurement values, calibration
results and alarm information to Normal maintenance of the POLYMETRON 9245
be recorded (capacity = 3200 data). has been reduced to:
The four sets of isolated analog outputs
can be configured in 0/4-20 mA with Q Replenishment of the calibration solution as required,
three outputs assigned to sample typically every 100 days,
concentration or temperature readings Q Replenishment of reagents and electrolyte of
and the fourth reporting electrode the reference electrode every 100 days,
live signal, including calibration and grab Q Annual replacement of the reagent tube.
sample. Additional digital communication
is available with JBUS/MODBUS Clear step-by-step instructions are given by the analyzer
or Profibus DP. to simplify maintenance operations such as instrument
start-up, long-term stand-by, and reagent replenishment.
Sample No. of sample streams 1, for all-integrated 1 to 4 channels see POLYMETRON 9240
Insoluble < 10ppm , no oil , no grease. For boiler sample type , install approx. 100um filter.
Temperature 5–45°C (41–113°F)
Pressure / Flow rate 0.2–6 bar (3–87 psi), 5 L/h during sampling phase
Connections Sample line Simple fittings for 6mm O.D. tubing or 1/4” O.D. in PE-low density
For 1/4” O.D. in PEHD-PTFE-SS, request Imperial kit
Drain Barbed stem for 12 mm (1/2” I.D.) hose
Ambient temperature 5–50°C (41–122°F)
Power supply 90-240 VAC, ±10%, 50/60Hz, 80 VA
Analysis Measuring ranges 0 to 10,000 ppb freely programmable / 0 to 200ppm with K Kit option
Repeatability < 0.02ppb or 1.5% reading, whichever is greater, within 10°C variation
Detection limit 0.01 ppb
Response time (t = 90%) 180s
Calibration As standard: Manual, 1 or 2 points
Option (sample < 0.1ppb) : Automatic known addition 2 points
Sample conditioner Di-isopropylamine (1L / 100 days),
Concentrated ammonia (sample >1ppb - 2.5 L / 90 days)
Total acidity As standard less than 50ppm (equivalent CaCO3)
Option K kit (cationic) less than 250ppm (equivalent CaCO3)
Maintenance Every 100 days Refill electrolyte, reagents and calibration solution
© 2006 HACH ULTRA ANALYTICS. Trademarks are property of their respective owners. Specifications are subject to change without notice.