Ametek Model 3050OLVRev2
Ametek Model 3050OLVRev2
MOISTURE
Model 3050-OLV Process Moisture Analyzer
Fast, Reliable PPMV Moisture Measurement
www.ametekpi.com
Model 3050-OLV Process Moisture Analyzer
Performance Specifications
MOISTURE
Technology: Quartz-crystal Sensitivity: 0.1 ppmv or 1% of Ambient Temperature Limits:
microbalance reading, whichever is greater Analyzer: 5° to 50°C (41° to 122°F)
Range: Calibrated from 0.1 to 2500 Allowable Inlet Pressure: 1.3 Enclosed analyzer with sample
parts-per-million by volume (ppmv). to 3.3 Bar (20 to 50 psi) gauge; system: -20° to 45°C (-4° to 113°F)
Readout capability in ppmw, lb/ up to 200 Bar (3000 psi) with Voltage/Power Requirements:
mmscf, mg/Nm3, and dew point in optional pressure reducer; analyzer Analyzer: 24 VDC, 50 watts
°C or °F (requires process pressure performance is independent of
as an input) Analyzer with sample system:
process pressure
Reference Gas: Continuously 120 ±10% VAC, 50/60 Hz, 150 W
Exhaust Pressure: 0 to 1 Bar
produced using actual sample gas maximum
(0 to 15 psi) gauge
On-Line Verification: Internal 230 ±10% VAC, 50/60 Hz, 150 W
Sample Gas Temperature:
moisture source with NIST- maximum
0 to 100°C (32° to 212°F); analyzer
traceable calibration enables performance is immune to changes Approvals and Certifications:
on-demand verification of analyzer in sample gas UL/CSA General Safety
accuracy and responsiveness Gas Flow Requirements: Requirements (general purpose)
without uninstalling the analyzer Approximately 150 sccm. UL/CSA Class I, Division 2, Groups
Accuracy: ±10% of reading from Approximately 1.0 slpm bypass flow A, B, C, D, T4
1 to 2500 ppmv with standard for increased speed of response UL/CSA Class I, Division 1, Groups
calibration; special calibration Outputs: Isolated 4-to-20 mA B, C, D, T6
ranges not required analog signal, 12 bit (0.025%) ATEX II 2 G EEx d IIC T6 or EEx de
Reproducibility: ±5.0% of reading resolution, RS232 and RS485 serial IIC T6
from 1 to 2500 ppmv communication ports (supports
Modbus RTU) Russian GOST 1ExdIICT6X
Limits of Detection: 0.1 ppmv
Alarms: Three contact closures; Russian GOST Pattern Approval
A/D Resolution: 16 bit (0.0015%)
system, data valid, and Complies with all relevant European
QCM Response Time: Near directives
concentration alarms
real time. Computer-enhanced
response, which may lead to errors,
is not required to obtain quick wet-
up or dry-down response