Free Chlorine Analyser (Swan)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

SWAN Analytische Instrumente AG

CH-8340 Hinwil/Switzerland Monitor AMI Codes-II


Tel. +41 44 943 63 00
swan@swan.ch . www.swan.ch Data sheet no. DenA254411000

Complete monitoring system for the automatic, continuous measurement of chlorine


and other disinfectants (e.g. monochloramine) in potable water, swimming pools,
cooling water, hot water loops and effluent.

Monitor AMI Codes-II


 For the continuous online determination of disin-
fectants based on the DPD colorimetric method
(EN ISO 7393-2; APHA 4500 Cl-G).
 Measurement values : free chlorine / chlorine di-
oxide / iodine or bromine using DPD+Buffer or
monochloramine / ozone using DPD+Buffer&KJ,
flow and if installed pH and temperature.
 Also applicable for water containing additives like
cyanuric acid.
 Complete system including measurement and
control electronics, photometer, flow indicator, re-
action chamber, reagent dosing system and rea-
gent containers.
 Integrated pH measurement with temperature
compensation (available as option).
 All usual dosing devices for disinfectants and pH
control can be connected either through relays or
analog output signals. Two independent control-
lers can operate simultaneously.
 Dosing of disinfectant can be interrupted automat-
ically with an external signal, e.g. during sample
flow interruption or filter backwashing.
 Two (optionally three) selectable measurement
values are available as analog output signals.
 Alarm display and activation of alarm relay when
user defined, critical limits for a measurement
value are reached.
 Continuous, automatic monitoring of main instru-
ment functions (dirty photometer, sample flow,
reagents).
 Large back-lit LCD display showing all measured Options:
 Communication interface
values and status information simultaneously.
 pH option containing pH sensor, temperature
 Factory tested, ready for installation and opera- sensor, cables and electronics board.
tion. Accessory:
 Chemical cleaning module. For details please
see separate data sheet no. DenA82312000.

Order Nr. Monitor AMI Codes-II A–25.441.100.0


rd
Option: [ ] 3 current signal output (0/4 – 20mA) A-81.410.020
[ ] Profibus DP interface A-81.420.020
[ ] HyperTerminal interface (RS-232) A-81.420.010
[ ] Modbus interface A-81.420.022
[ ] USB interface A-81.420.040
Option: [ ] pH and temperature measurement A-87.127.020

07/2014 Subject to change without notice DenA254411000_AMI_Codes-II.docx


SWAN Analytische Instrumente AG
CH-8340 Hinwil/Switzerland Monitor AMI Codes-II
Tel. +41 44 943 63 00
swan@swan.ch . www.swan.ch Data sheet no. DenA254411000

Disinfectant measurement
Electrical Connection Scheme
Measuring range: Accuracy:
Ozone
0.000 – 1.000 ppm  0.005 ppm
HOCl, free chlorine, monochloramine
0.00 – 1.00 ppm  0.01 ppm
1.00 – 3.00 ppm  0.06 ppm
0.00 – 6.00 ppm  0.12 ppm
Chlorine dioxide, iodine, bromine
0.00 – 2.00 ppm  0.02 ppm
2.00 – 6.00 ppm  0.12 ppm

Response time:
90% of change of excessive Cl in 60 sec-
onds after sample entered flow cell.

Cycle time free chlorine: 1 – 12 min.

pH (option):
Measuring range: pH 2 - 12
Resolution: 0.01 pH

Temperature measurement (Option)


with Nt5k sensor
Measuring range : -30 to +100 °C
Resolution : 0.1 °C

Transmitter Specifications and


Functionality

Electronics case: Aluminum


Protection degree: IP 66 / NEMA 4X
Display: backlit LCD, 75 x 45 mm Real-time clock with calendar 1 Communication interface (option)
Electrical connectors: screw clamps For action time stamp and prepro- - RS232 interface for logger download to
Ambient temperature: -10 to +50 °C grammed actions. PC with SWANTerminal
Limit range of operation: -25 to +65 °C - RS485 interface (galvanically separated)
Storage and transport: -30 to +85 °C 1 Alarm relay with Fieldbus protocol Modbus or Profi-
Humidity: 10 to 90 % relative, One potential free contact for summary bus DP
non condensing alarm indication for programmable alarm - 3rd Signal output
Power supply values and instrument faults. - USB interface
Voltage: 100 - 240 VAC ( 10 %), Maximum load: 1A / 250 VAC
50/60 Hz ( 5 %)
or 24 VDC ( 10 %) 1 Input Monitor Data
Power consumption: max. 30 VA One input for potential-free contact. Pro-
grammable hold or remote off function. Sample conditions
Operation Water consumption: min. 10 l/h
Easy operation based on separate menus 2 Relay outputs Water inlet pressure 0.15 to 2 bar
for “Messages”, “Diagnostics”, “Mainte- Two potential-free contacts programmable Sample temperature: 5 to 50 °C
nance”, “Operation” and “Installation”. as limit switches for measuring values,
User menus in English, German, French controllers or timer with automatic hold Flow cell and connections
and Spanish function. Made of acrylic glass with water inlet filter
Separate, menu specific password protec- Rated load: 1A / 250 VAC and needle valve.
tion. Openings for pH and temperature sen-
Display of process value, alarm status and 2 Signal outputs (3rd optional) sors.
time during operation. Two programmable signal outputs for Inlet tubing: 6 x 8 mm
Storage of event log, and alarm log and measured values (freely scaleable, linear Outlet pressure: atmospheric drain
calibration history. or bilinear) or as continuous control out- Outlet tubing: 15 x 20 mm (1/2")
Storage of the last 1’500 data records in puts (control parameters programmable).
logger with selectable time interval. Current loop: 0/4 - 20 mA Panel
Maximum burden: 510  Panel dimensions: 400 x 850 x 200 mm
Safety features Panel material: PVC
No data loss after power failure, all data is Control functions Weight: 12.0 kg
saved in non-volatile memory. Over- Relays or current outputs programmable
voltage protection of in- and outputs. for 1 or 2 pulse dosing pumps, solenoid
Galvanic separation of measuring inputs valves or for one motor valve.
and signal outputs. Programmable P, PI, PID or PD control
parameters.
Transmitter temperature monitoring
With programmable high/low alarm limits.

07/2014 Subject to change without notice DenA254411000_AMI_Codes-II.docx