(For Research Use Only) 2. Catalyst (R2): 1 x 1.2 ml amber bottle. Intended Use 3. Standard: 1 x 1 ml clear bottle. 4. Diluent: 1 x 10 ml clear bottle. TM The SideroTec Assay is a colorimetric test for use 5. Assay tray: 1 x 96 well plate. in the detection of siderophores secreted by Note: Do not mix reagents from different batches bacteria [1] or fungi [2], or for the assessment of of kits synthetic iron chelators. The test can be used with liquid culture either directly or following filtration Additional Material Required to remove microorganisms. The test may also be 1. Sample. used with other aqueous based liquids. 2. Plastic or Eppendorf® tubes for Background preparation of reference curve. 3. Accurate pipettes and disposable pipette Siderophores are generally low molecular weight to dispense 100 µl of sample or reaction peptides which are specific for chelating iron in material. low iron environments. Siderophores are 4. A repeat 8-channel pipette (50-200 µl) for synthesised by a wide variety of both fungi and addition of reagent to a 96-well plate. bacteria. While there are many different 5. Microtitre plate reader capable of siderophores, in general they fall into three main measuring at 620-630 nm. groups depending on the chemical nature of the 6. Timer. reaction with iron: hydroxymates, catecholates and “mixed-type” i.e. they have both reactive Storage: If not used within 1 month of receipt, hydroxymate and catecholate groups. In general the standard should be stored frozen until bacteria produce either form of siderophore, and needed. in some cases both, where as fungi generate The kit should be stored at room temperature hydroxymate siderophores only. The SideroTec TM when not in use. Once opened, the kit should be Assay is a universal assay that will react with all used within 1 month. The kit has a shelf-life of 12 classes of siderophore regardless of chemistry or months from the date the kit was made. The origin and so can be used for detection of a wider expiry date is located on the outer label of the box. range of iron-binding compounds. TM Method SideroTec Assay Principle Sample Preparation The assay is based on the colour change that occurs as result of ferric iron transfer from the Samples may be aqueous based or be from more reagent complex to siderophore present in a complex media such as cell culture. With cell sample. The key reagent is a complex of a dye, culture, it is not necessary to remove cells from iron, and a detergent. The dye complex is initially cell culture fluid to detect the presence or absence blue but on removal of iron the dye changes colour of siderophore, as the reagent will change colour to purple or pink colour depending on the amount even in the presence of cellular material. However, of siderophore present in a sample. The test can the extent of colour change will depend on the be used as a qualitative indicator for siderophore amount of siderophore secreted, so very low levels detection or can be used as quantitative test to (less than 10 µg/ml) may not be detected where estimate the level of siderophore present in a the cell suspension is turbid. In this case, cells sample. In the absence of any siderophore, the should be removed either by filtration or by reaction colour remains blue. centrifugation of 1-2 ml of culture fluid in a minifuge prior to assay.
For optimal use or for quantitative analysis,
samples should be clear of particulate material.
Ref: SideroTec 2301/6
Reagent Preparation culture medium or other aqueous matrix). Add 100 µl of each to wells. Proceed to Catalyst (R2) should be diluted 1:10 in the Dye step 3. reagent (R1). The volume of R1 and R2 solution 2. For quantitative assessment, add 100 µl used will depend on the number of samples to be of sample or standards to the appropriate tested and whether standard or controls are used. wells. For accuracy we recommend that Table 1 gives an indication of the amount of each all samples and standards are run in to be mixed depending on the number of test duplicate. However, single wells can be strips used in the plate provided. run where a qualitative result is required i.e. where it is only necessary to indicate Table 1 the presence or absence of siderophore. Number of Volume of R1 Volume of R2 3. Add 100 µl of pre-mixed R1 reagent/ R2 wells used (ml) ( µl) using an 8 channel pipette. 16 2 200 4. Incubate for 10 minutes at room 32 4 400 temperature. 48 6 600 5. Record results. Wells can be read visually, 64 8 800 photographically, or on a microplate 80 10 1000 reader at 630 nm. 96 12 1200 Interpretation of results Standards For qualitative interpretation, either visual recording or plate reading can be used. Reaction of Standards are prepared by serial dilution of the the dye complex with any iron-binding compound stock material (100 µg/ml) provided in the kit. It is present will result in both a colour change and a recommended that standards are run in duplicate. reduction in optical density relative to the The standard should be prepared following zero/blank control. addition and subsequent serial dilution of the 100 µg/ml standard (Table 2). For quantitative interpretation, results should be obtained by reading the microplate on a Table 2 microplate reader. A standard curve should be Tube Concentration Volume Volume Serial prepared by plotting optical density of standards Number (µg/ml) of of Dilution versus siderophore concentration (µg/ml). Calibrator Diluent (µl) (µl) (µl) C1 100 500 - Alternatively results can be expressed as a C2 50 - 250 250 C1 percentage of the zero standard/blank. The C3 25 - 250 250 C2 C4 12.5 - 250 250 C3 standard curve is calculated by dividing standard C5 6.25 - 250 250 C4 OD by the blank/zero standards multiplied by 100 C6 3.1 - 250 250 C5 C7 1.5 - 250 250 C6 i.e. (OD Standard / OD zero Standard) x 100 C8 0 - 250 -
Similarly, divide the sample OD by the OD reading
of the zero standard and multiply by 100. Note: there is sufficient standard to run two Extrapolation from the reference curve will give an complete standard curves. indication of the relative amount of siderophore in TM SideroTec Assay Procedure the sample. However, quantitative results should be interpreted with a degree of caution; different (Note: the assay procedure described is for a siderophores may react at a different rate with dye manual assay format. However, it is possible to reagent. run the assay as single reagent format on an autoanalyzer but the settings should be set Samples with an OD lower than the 100 µg/ml locally for each type of analyzer). standard should be diluted 1 in 5 in the diluent provided and retested. 1. For qualitative assay it is sufficient to run samples and both positive (undiluted standard) and negative controls (e.g. cell Ref: SideroTec 2301/6 Figure 1 is an example of a typical standard curve but it is only provided as an example of what is expected and should not be used for actual References determination of siderophore concentration. 1. Balado, M., Osorio, C.R. and Lemos, M.L. (2006) A gene cluster involved in the biosynthesis of vanchrobactin, a chromosome-encoded siderophore produced by Vibrio anguillarum. Microbiology. 152(12): 3517-3528.
2. Reiber, K., Reeves, E., Neville, C., Winkler, R.,
Gebhardt, P., Kavanagh, K. and Doyle, S. (2005) The expression of selected non-ribosomal peptide synthetases in Aspergillus fumigatus is controlled by the availability of free iron. FEMS Microbiol. Lett. 248(1): 83-91.
interfere with reading of a test. However, for visual reading, colour change should be visible even in the presence of microbes, but may be dependent on the amount of siderophore secreted.
The test works best within pH range 6-8. Acidic
(less than pH 5) and basic (greater than pH 9) can interfere with colour development. While most culture fluids are within this range, the pH of those outside may need to be adjusted before use in the Disclaimer test. Products available from the Company are intended for research Some salts at high concentrations (e.g. phosphate use only. It is the responsibility of the user to ensure that they ions) may interfere with the test depending on the possess the necessary technical skills to determine the concentration used. appropriateness of these products for the proposed application. The Company cannot accept responsibility for Sensitivity results obtained from these products, as their efficacy depends on conditions of use and the variability of other materials used which are beyond the control of the Company. In no event will The sensitivity of the test using Desferoxamine as the Company be held responsible for the loss of profits, or for reference material is approx 2 µg/ml indirect or consequential loss including, but not limited to, personal injury arising from the use of these products. Contact Details Customers are advised that the Company accepts no responsibility for the non-delivery of goods or for damages in For technical support, contact: Emergen Bio transit unless it can be proved to have resulted from its negligence. Claims for goods damaged in transit should be • e-mail: kieran.walshe@nuim.ie made within 3 working days of receipt. Any replacements are • Tel: 00 353 87 2823555 made strictly as an act of goodwill, and should in no way • Webpage: www.emergenbio.com construed as an acceptance of liability.
Emergen Bio operates at NUI Maynooth. The term "The