ENDA-5000 Instruction Manual
ENDA-5000 Instruction Manual
ENDA-5000 Instruction Manual
ENDA-5000
Instruction Manual
CODE:GZ9100497217J
Preface
This manual describes the operation of the Stack Gas Analysis System, ENDA-5000.
Be sure to read this manual before using the product to ensure proper and safe operation of
the instrument. Also safely store the manual so it is readily available whenever necessary.
Product specifications and appearance, as well as the contents of this manual are subject to
change without notice.
HORIBA, Ltd. warrants that the Product shall be free from defects in material and
workmanship and agrees to repair or replace free of charge, at option of HORIBA, Ltd., any
malfunctioned or damaged Product attributable to responsibility of HORIBA, Ltd. for a period
of one (1) year from the delivery unless otherwise agreed with a written agreement. In any one
of the following cases, none of the warranties set forth herein shall be extended;
Any malfunction or damage attributable to improper operation
Any malfunction attributable to repair or modification by any person not authorized by
HORIBA, Ltd.
Any malfunction or damage attributable to the use in an environment not specified in this
manual
Any malfunction or damage attributable to violation of the instructions in this manual or
operations in the manner not specified in this manual
Any malfunction or damage attributable to any cause or causes beyond the reasonable
control of HORIBA, Ltd. such as natural disasters
Any deterioration in appearance attributable to corrosion, rust, and so on
Replacement of consumables
HORIBA, LTD. SHALL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM ANY
MALFUNCTIONS OF THE PRODUCT, ANY ERASURE OF DATA, OR ANY OTHER USES
OF THE PRODUCT.
Trademarks
Company names and brand names are either registered trademarks or trademarks of the
respective companies. (R), (TM) symbols may be omitted in this manual.
FCC Rules
Any changes or modifications not expressly approved by the party responsible for compliance
shall void the user's authority to operate the equipment.
WARNING
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in
which case the user will be required to correct the interference at his own expense.
Korea Certification
Safety Policy
We arrange warning labels on our products, and describe notes and cautions in this manual.
Make sure to follow these instructions for your safety.
WARNING 警 告
感 電 注 意
ELECTRICAL
通電中にカバーを開けると感電することがあります。
カバーを開けるときは、電源を元から切ってから
行ってください。
OPENING THE COVER WHILE POWERED ON COULD
RESULT IN ELECTRIC SHOCK BE SURE TO TURN
OFF POWER PRIOR TO OPENING THE COVER.
Pictograph Message
- Type of warning
- Why it is dangerous and
how to avoid injury.
- What would happen
if the warning is not
followed.
• CAUTION:
This indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury. It may also be used to alert against unsafe practices.
■ Labels and Location
● Label location
WARNING 警 告
感 電 注 意
ELECTRICAL
通電中にカバーを開けると感電することがあります。
カバーを開けるときは、電源を元から切ってから
行ってください。
OPENING THE COVER WHILE POWERED ON COULD
RESULT IN ELECTRIC SHOCK BE SURE TO TURN
OFF POWER PRIOR TO OPENING THE COVER.
WARNING 警 告
感 電 注 意
ELECTRICAL
感電防止のため、 装置の接地を
してください。
GROUND INSTRUMENT TO
AVOID ELECTRICAL SHOCK.
CAUTION 注 意
高温注意
HOT COMPONENT
ユニットに触れるとやけどをします。
冷めるまで分解しないこと。
YOU MAY BE BURNED IF YOU
TOUCH THE HEATING ELEMENT.
DO NOT DISASSEMBLE THE UNIT,
UNTIL IT HAS COOLED OFF.
■ Description in this manual
This manual contains the cautionary messages in the following style:
This product contains a part of high voltage and high temperature that may cause danger in human life
or burn injuries.
Fully take care for operation, inspection and adjustment of the product for the safety.
■ Operating precautions
Whenever handling (operating, maintaining, repairing, etc.) this product, you have to follow the general
safety precautions mentioned below:
• Ground the product:
In order to minimize the risk of electric shock
• Shut off the power supply in the operations of product assembling/disassembling:
In order to prevent electric shock and product damage
Do not connect the power cable and discharge the internal circuits before operations.
• Use protective equipment:
Wear protective glasses, gloves, etc., if necessary.
■ Original language
This is the English translation of an original Japanese document.
■ Reference Documents
ENDA-5000 Installation manual I1002205000
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3 Basic Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Preparation for Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Operation Stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Power Interruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Momentary power interruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.2 Power interruption 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.3 Power interruption 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.1 Calibration Screen (CAL. screen) . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.2 Preparation of Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.2.1 Gas connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.2.2 Setting calibration range and span gas concentration . . . . . . . . . . . . 19
5.3 AIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3.1 AIC setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.3.2 AIC start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6 Other Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.1 Operation Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.1.1 Setting measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
7 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.1 Contact for Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
7.2 Daily Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.3 Periodical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.3.1 Inspection every 1 to 6 months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
7.3.2 Annual inspection and replacement . . . . . . . . . . . . . . . . . . . . . . . . . 58
7.3.3 Inspection with Interval of 3 years or more . . . . . . . . . . . . . . . . . . . . 59
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8.1 Alarm Type and Countermeasure . . . . . . . . . . . . . . . . . . . . . . . . . 76
8.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
9 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
9.1 Measurement Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
9.1.1 Measurement principle using cross-modulation type NDIR . . . . . . . . 83
9.1.2 Measurement principle for O2 using magnetic pressure method . . . 84
9.2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
9.3 Analog Signal Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
9.4 List of Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
9.5 Accessories and Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
9.5.1 List of accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
9.5.2 List of spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
10 Optional Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1 Purge Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1.1 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.1.2 Purge setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.1.3 Purge start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
1 Introduction
The Stack Gas Analyzer ENDA-5000 series is a measuring instrument that continuously mon-
itors the concentration of NOx, SO2, CO, CO2 and O2 contained in the stack gas of fixed dis-
charge source. It can measure simultaneously up to 5 components.
For the analyzers of NOx, SO2, CO and CO2, non-dispersive Infrared ray absorptiometry with
cross modulation system (NDIR) is applied to enable excellent stability and accuracy.
For O2 analyzers, magneto-pneumatic system (MPA) is applied, which is free from cylinder
gas as a carrier gas.
The operation unit incorporates touch panel LCD.
As a pollution-monitoring instrument, the approval type of products is available for analyzers
NOx, SO2, CO and O2.
1
2 Name of Each Part
1
3
10
2
2 Name of Each Part
7 8 9 10 11
Fig. 2 Panel switch in cabinet
Note
The short circuit breaker, circuit protector, and switch cannot always be attached depending on the
specification.
3
2 Name of Each Part
Ventilation fan
Fluorescent lamp
Primary filter power output Primary filter CP (3 A)
Fluorescent lamp
100 V AC,
Approx. 100 VA Power outlet
(Purge specification: Approx. 200 VA) Panel heater CP (5 A)
Panel heater
Note
The actual power system may be different from the above figure according to the specification.
4
2 Name of Each Part
19 28
9
27 15 18
26
21 23 24 Silica 20
gel
Air inlet
29
30
SV-7
1 5 6 7 8 9 10 25
2
32 31 SV-8
3
Sample outlet
4
11 12 14 13 16 17
22 SV-2
SV-3
SV-4
SV-6
SV-1
5
2 Name of Each Part
6
2 Name of Each Part
2, 10, 25
1
15, 16, 17, 18, 19, 20
(Contained in the analyzer case)
3
12
32
8 14
28
9
27
26
30 23
(Scrubber)
22 21
29 13 5 11, 31
7
3 Basic Operation
3 Basic Operation
3.1 Preparation for Operation
Wiring check
Confirm that the wiring is correctly given according to the separate installation manual.
Tube 1
Disconnect the tube from the upper of cabinet,
connect the watering pot to start water supply.
Tube 2
Connect the watering pot
to start the water supply
8
3 Basic Operation
Note
When the panel heater is attached, connect the attached cover to the cabinet ventilation filter in
winter.
5. After confirming the followings on the electric cooler, turn ON the pump 1 switch
and 2 switch.
Item Confirmation Reference
Power lamp Lit
" Fig. 6 Drain trap watering " (page 8)
Fan Rotated
0C to 8C (below 15C when the
" Fig. 31 ANALOG INPUT 2 screen "
Temperature temperature inside the cabinet is
(page 36)
over 40C)
Note
The operation lamp lights up in red for a moment at power ON, and turns green at once.
This happens by hardware check. It is not a sign of abnormality.
8. Regulate the sample flow into 0.6 L/min ±0.1 L/min by needle valve.
9
3 Basic Operation
1. Introduce zero calibration gas, or allow air suction from the sample inlet to purge
for 5 minutes or longer.
2. Turn OFF the pump 1 switch and 2 switch.
3. Turn OFF the analyzer switch.
4. Turn OFF the primary filter CP, heated tube CP, and panel heater CP.
5. Turn OFF the main power ELB.
6. In winter, discharge the water remained in the drain trap.
Otherwise, the drain trap may be cracked.
Note
Be careful in handling the water in the drain trap, because it may be strongly acidic depending on the
sample condition. If your hand or skin contacts the water in the drain trap, flush away with water
completely.
10
3 Basic Operation
Energized status
1 2 3 4
Note
When power failure occurs during “Battery” alarm, the internal clock, AIC start time, and the purge
start time is reset to 2090/1/1, 0:00 (initial value) after the power restoration.
11
3 Basic Operation
Note
Be careful in handling the water in the drain trap, because it may be strongly acidic depending on the
sample condition. If your hand or skin contacts the water in the drain trap, flush away with water
completely.
3. When the calibration gas is connected, ensure the main valve is closed.
Note
To store the instrument in a place other than the installation site, avoid direct sunlight, wind, or rain.
To start the operation after the storage, replace the following parts.
Primary filter element
Secondary filter element
Scrubber (reference side)
Mist catcher
Silica gel (when discolored to pink)
When a vinyl is covered over the instrument, dew condensation may occur inside and cause
corrosion.
Do not use a vinyl cover.
12
4 Measurement Screen (Basic Panel)
↓
System setting LCD adjustment → Refer to page 46
↓
Setting of calibration range
Calibration preparation → Refer to page 19
and span gas concentration
↓
Calibration → Refer to page 22 to page 30
↓
Setting of measurement range → Refer to page 33
↓
Measurement
13
4 Measurement Screen (Basic Panel)
Note
The screen applies a touch panel.
Directly press on the screen by the fingers for operation.
Do not use anything having hard or sharp tips like a ball-point-pen nor press the screen too strong.
The screen may be malfunctioned.
① [MEAS.] displayed ② Icon displayed ③ Current time displayed ④ [KEY LOCK] button
⑤ Measurement component
displayed
⑦ Measurement range
displayed
⑧ Function keys
Pressing the [COR.] key (optional) switches the MEAS. screen to the COR. screen.
14
4 Measurement Screen (Basic Panel)
① [MEAS.] display
Displayed during the sample gas is measured.
It is not displayed in performing automatic calibration (hereinafter called as “In AIC,” in
introducing calibration gas (including at replacement), or in maintenance.
② Icon display
The instrument status is displayed in the following icons.
In maintenance Displayed when the maintenance switch is ON.
(The concentration signal output is hold according to the
specification.)
Line Displayed or blinks in the calibration gas introduction and in sample
replacement. For calibration, refer to " 5 Calibration " (page 17).
In AIC Displayed or blinks in AIC.
For automatic calibration, refer to " 5.3 AIC " (page 22).
When the key lock function is set to ON, only the screen confirmation is enabled.
No button operation other than pressing the [KEY UNLOCK] button is available.
This function allows preventing irresponsible setting change due to incorrect operation.
15
4 Measurement Screen (Basic Panel)
Measurement range
⑧ Function keys
Pressing each key enables the corresponding functions as follows.
Menu Displays the MENU 1/5 (OPERATION menu) screen (page 31 Fig. 25).
COR. (optional) Displays the COR. display in the MEAS. screen 2 (Conversion value
display) (page 14 Fig. 8).
For O2 conversion function, refer to " 10.2 O2 Correction " (page 102).
CAL. Displays the CAL. screen (page 17 Fig. 11).
ALARM Displayed when any error occurs.
Pressing this key displays the ALARM screen (page 54 Fig. 52), enabling
to check the alarms currently occurred.
16
5 Calibration
5 Calibration
For a stable and accurate data acquisition, periodical calibration is necessary.
There are two calibration methods; automatic calibration (AIC) and manual calibration.
Reference
Refer to " 5.3 AIC " (page 22).
Manual calibration
A calibration method given by manual operation with flowing the calibration gas
voluntarily.
Reference
Refer to " 5.4 Manual Calibration " (page 28).
Note
In calibration, the concentration output is held according to the specification.
① Calibration mode
② Range
③ Alarm/Caution
④ Function key
17
5 Calibration
① Calibration mode
Displays the selected calibration mode.
② Range
In measurement and in sample replacement, the measuring range is displayed.
In the calibration gas introduction, the set calibration range is displayed.
③ Alarm/Caution
Displays the operation result of calibration coefficient.
ALARM The calibration coefficient is out of the normal range.
The calibration coefficient is not revised but the previous calibration coefficient
is used for the concentration operation.
CAUT. The calibration coefficient is within the normal range, but is close to the range
outside.
The calibration coefficient is revised to perform the concentration operation.
Reference
When ALARM or CAUT. is displayed, refer to " 8 Troubleshooting " (page 76).
Note
When zero calibration alarm occurs during AIC, the span gas flows though span calibration will not be
performed for the components.
In manual calibration, the span calibration will be performed also for the alarmed components of zero
calibration.
④ Function keys
Pressing each key enables the corresponding function as follows.
EXIT Returns to the MEAS. screen 1 (Measurement value display) (page 14 Fig. 7).
RELEASE When the calibration alarm (ALARM) or calibration caution (CAUT.) occurs, the
[RELEASE] key is displayed on the screen.
To release the alarm or caution, press this key after checking the cause of
problem. Even with ALARM or CAUT. released, the occurrence of ALARM or
CAUT. is recorded in the alarm history.
CAL. The manual calibration is performed.
For details, refer to " 5.4 Manual Calibration " (page 28).
AIC Starts automatic calibration.
18
5 Calibration
3. Press the [CALIBRATION SET] button. The CALIBRATION SET screen appears.
In this screen, the calibration range as well as the span concentration can be set.
19
5 Calibration
4. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 20 to page 21.
5. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the CALIBRATION SET screen.
Tip
To cancel the changes, press the [CANCEL] key.
The changes are canceled and the screen returns to the CALIBRATION SET screen.
20
5 Calibration
Note
If changing the calibration range or the span concentration during AIC, “AIC” appears on the screen
and rejects the operation.
21
5 Calibration
5.3 AIC
In the automatic calibration (AIC), the calibration can be automatically started by the
internal clock.
The calibration can be started also by the [AIC] key in the CAL. screen or the external
contact input. Adjust the calibration gas flow by referring to " Calibration gas flow
control" (page 30).
22
5 Calibration
Note
If changing the AIC setting during AIC, “AIC” appears on the screen and rejects the operation.
23
5 Calibration
Mode
Sets the mode to start AIC.
1. Pressing the right button of [MODE] in the AIC SET screen will display the AIC
MODE screen.
2. Press the button of the item to be set, and press the [SET] key.
The setting is changed and the screen returns to the AIC SET screen.
Note
AIC based on the internal clock and AIC from external contact input cannot be set at the same time.
24
5 Calibration
Start time
Note
In selecting the AIC based on the internal clock, it is required to set the start time.
Year Minute
Hour
Month
Day
2. Pressing the button showing the value to be changed will reverse the display
mode. Change the value by the [] and [] button, and press the [SET] key.
The instrument changes the setting and the screen returns to the AIC SET screen.
Note
The start time setting is based on the internal clock.
It is not available to change into an impossible date.
When this happens, the existing date closest to that before the set date is automatically selected.
When the start time previous to the current time is set, the setting value of start time is changed
into the minimum value which comes in future by adding multiple number of the interval set
values.
When the start time set value becomes a previous time by changing the internal clock setting
(refer to " 6.5.2 Time adjust " (page 48)), the start time is changed into the minimum value which
comes in future by adding multiple number to the current time set value.
When the AIC start time is set later than 2090/1/1/0:00, the AIC does not start.
When the “Battery” alarm occurs, confirm the setting time.
25
5 Calibration
Interval
Note
When the AIC based on the internal clock is selected, it is required to set an interval.
The following is a setting example of time span interval when AIC is started with a specific
time span.
1. Pressing the right button of [INTERVAL] in the AIC SET screen will display the AIC
INTERVAL screen.
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by the [] and [] button and press the [SET] key.
Change the setting and the screen returns to the AIC SET screen.
Note
The available interval to set is from “ 0 day 1 hour” to “99 day 0 hour.”
The recommended interval is “7 day 0 hour.”
26
5 Calibration
Manual start
1. In the MEAS. screen 1 (Measurement value display), press the [CAL.] key.
The CAL. screen appears.
Calibration mode
Note
Pressing the [AIC] key during the AIC can stop the AIC.
Note
The AIC by manual start using the [AIC] key is available except during the purge.
When the AIC command is input by either the internal clock or the external contact input during the
manual start AIC, the signal is not received but the AIC by manual start continues.
27
5 Calibration
Calibration mode
2. Pressing the right button of [MODE] will display the CAL. MODE screen.
28
5 Calibration
4. The solenoid valve of the selected calibration gas automatically opens, and the
calibration gas is introduced into the analyzer.
On the screen, the LINE icon starts blinking.
LINE icon
5. Confirm that the mode on the screen shows the set component and type of
calibration.
6. When the indication stabilizes, press the [CAL.] key.
The calibration of set component and type starts.
Note
When any abnormality occurs in calibration to cause a state where the further calibration is not
possible, either “ALARM” or “CAUT.” is indicated at the “component” and “type” of calibration in the
CAL. screen. When either of these signs appears on the screen, the [RELEASE] key also appears on
the screen.
For “CAUT.” The calibration result is revised, but for “ALARM,” the result is not revised.
To release “ALARM” or “CAUT,” press the [RELEASE] key.
To finish the calibration and retry the measurement, press the [MEAS.] button in the CAL.
MODE screen. The LINE icon is extinguished.
The solenoid valve is actuated, and the sample gas is introduced into the analyzer.
Note
Be sure to confirm that the LINE icon is extinguished on the MEAS. screen.
When the LINE icon is displayed, it means that the calibration gas is being introduced.
When the automatic calibration command either by internal clock, external contact input, or by
key input during the calibration through manual gas introduction, the system preferably selects
the automatic calibration.
29
5 Calibration
Note
When the air is used as zero gas (it is span gas for the O2 analyzer), flow control of zero gas (it is span
gas for the O2 analyzer) is not required.
At air zero calibration (it is span calibration for the O2 analyzer), the flowmeter indication is 0 L/min.
30
6 Other Functions
6 Other Functions
Pressing the [MENU] key in the MEAS. screen allows other various functions as follows.
31
6 Other Functions
2. Press the button for the item to be set. The setting screen appears or the operation
is performed.
Pressing each button allows the corresponding operations as follows.
MEASURE RANGE Displays the MEAS. RANG screen (page 33 Fig. 27).
PURGE START Optional function; not displayed in the standard specification.
(optional) For the operation procedure, refer to " 10.1.3 Purge start " (page 101).
ROLLING AVERAGE Optional function; not displayed in the standard specification.
RESET For the operation procedure, refer to " 10.3.1 Rolling average time set "
(optional) (page 108).
3. When the setting screen is displayed, change the setting and press the [SET] key.
Change the setting in each setting screen, and press the [EXIT] key.
The screen returns to the MENU 1/5 (OPERATION menu) screen.
4. Press the [EXIT] key in the MENU 1/5 (OPERATION menu) screen.
The screen returns to the MEAS. screen.
32
6 Other Functions
Range 1
Range 2
Press the button of the component to be set as a measurement component, and execute
the setting.
33
6 Other Functions
Note
For automatic range switch, the range automatically switches in the following conditions.
RANGE 1 to RANGE 2: At more than 90% of the concentration of RANGE 1
RANGE 2 to RANGE 1: At less than 80% of the concentration of RANGE 1
The concentration output to the external devices is switched at the same time.
34
6 Other Functions
3. Confirm the situation and press the [EXIT] key in each screen.
The screen returns to the MENU 2/5 (MAINTENANCE/DATA menu) screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and the screen returns to the
MENU 2/5 (MAINTENANCE/DATA menu) screen.
4. Press the [EXIT] key in the MENU 2/5 (MAINTENANCE/DATA menu) screen.
The screen returns to the MEAS. screen.
35
6 Other Functions
36
6 Other Functions
Note
When any problem occurs, each value may be required to be reported to solve the problem.
37
6 Other Functions
Note
After the setting, press the [EXIT] key twice to return to the MEAS. screen.
Pressing the [EXIT] key returns the output from all the channels to the [MEAS.] state.
For the configuration of output channel and output terminal, refer to " 9.3 Analog Signal Output "
(page 87).
When the analog output needs adjustment, contact HORIBA.
38
6 Other Functions
Maximum 15 records of calibration coefficient history can be memorized for each component.
Pressing the [] key will display the next screen.
Press the [] key and [] key to switch the screen.
Pressing the [COMPO.] key will display the COMPONENT screen.
Press the component button which calibration coefficient history is to be checked, and press
the [SET] key.
The displayed component is changed, and the screen returns to the CALIBRATION HISTORY
screen for the set component.
39
6 Other Functions
Reference
For detailed information of each alarm, refer to " 8.1 Alarm Type and Countermeasure " (page 76).
40
6 Other Functions
Reference
For the peak count setting, refer to " 10.5.1 Peak count set " (page 117).
41
6 Other Functions
2. Press the button for the item to be set. The SETTING screen appears.
Pressing each button allows the corresponding operations as follows.
AIC SET Displays the AIC SET screen.
For details, refer to " 5.3 AIC " (page 22).
CALIBRATION SET Displays the CALIBRATION SET screen.
For details, refer to " 5 Calibration " (page 17).
3. Change the setting in each setting screen, and press the [EXIT] key.
The setting is changed and the screen returns to the MENU 3/5 (SETTING menu)
screen.
4. Press the [EXIT] key in the MENU 3/5 (SETTING menu) screen.
The screen returns to the MEAS. screen.
42
6 Other Functions
2. Press the button for the item to be set. The setting screen appears.
Pressing each button allows the corresponding operations as follows.
O2 CORRECTION Displays the O2 CORRECTION SET screen.
SET This is the optional function, and not displayed in the standard specifi-
cation.
For the operation procedure, refer to " 10.2 O2 Correction " (page 102).
ROLLING AVERAGE Displays the ROLLING AVERAGE TIME SET screen.
TIME SET This is the optional function, and not displayed in the standard specifi-
cation.
For the operation procedure, refer to " 10.3 Rolling Average/Integration
" (page 108).
PURGE SET Displays the PURGE SET screen.
This is the optional function, and not displayed in the standard
specification.
For the operation procedure, refer to " 10.1 Purge Functions " (page
93).
MEASURE VALUE Displays the MEASURE VALUE ALARM screen.
ALARM This is the optional function, and not displayed in the standard specifi-
cation.
For the operation procedure, refer to " 10.4 Measurement Value Alarm "
(page 113).
CO PEAK COUNT Displays the CO PEAK COUNT ALARM screen.
ALARM This is the optional function, and not displayed in the standard specifi-
cation.
For the operation procedure, refer to " 10.5 Peak Count (with CO Ana-
lyzer) " (page 117).
43
6 Other Functions
3. Change the setting in each setting screen, and press the [EXIT] key.
The setting is changed and the screen returns to the MENU 4/5 (OPTION menu)
screen.
4. Press the [EXIT] key in the MENU 4/5 (OPTION menu) screen.
The screen returns to the MEAS. screen.
44
6 Other Functions
3. Press the button for the item to be set. The setting screen appears.
Pressing each button allows the corresponding operations as follows.
LCD ADJUST Displays the LCD ADJUST screen (page 46 Fig. 40).
The setting of LCD backlight time period and brightness adjustment are
already set at the factory.
Change the setting in this screen as necessary.
TIME ADJUST Displays the TIME ADJUST screen (page 48 Fig. 42).
The time is already set at the factory.
Change the setting in this screen as necessary.
PASSWORD Displays the PASSWORD SETTING screen (page 49 Fig. 43).
SETTING
TOUCH PANEL Displays the TOUCH PANEL ADJUSTMENT screen (page 51 Fig. 46).
ADJUSTMENT
4. Change the setting in each setting screen, and press the [EXIT] key.
The setting is changed and the screen returns to the MENU 5/5 (SYSTEM menu)
screen.
5. Press the [EXIT] key in the MENU 5/5 (SYSTEM menu) screen.
The screen returns to the MEAS. screen.
45
6 Other Functions
The current BACK LIGHT TIME PERIOD setting is indicated in button configuration.
Pressing the button indicating the BACK LIGHT TIME PERIOD enables to change the time
setting (refer to page 47). The current BRIGHTNESS of LCD is indicated in [] position.
Pressing each key allows the corresponding operation as follows.
EXIT Returns to the MENU 5/5 (SYSTEM menu) screen.
46
6 Other Functions
Note
To protect the back light, normally set either to 10 min, 20 min, or 30 min (avoid to set to ON).
Even with time setting, the backlight will be always lit when the In-maintenance switch is ON.
Brightness adjustment
Pressing the following buttons enables to adjust BRIGHTNESS.
[] The screen is darkened.
[] The screen is brightened.
47
6 Other Functions
Year Minute
Hour
Month
Day
Soon after the screen appears, the current time is indicated in buttons as year, month, day,
hour, and minute. Pressing each item button allows to change the corresponding item value.
Press the following buttons for changing the value.
[] The value increases.
[] The value decreases.
Note
Pressing the [CANCEL] key during the setting change does not change the time but applies the
time before change.
The setting per second is not possible; pressing the [SET] key automatically sets to “00 second.”
It is not available to change into an impossible date.
When this happens, the existing date closest to that before the set date is automatically selected.
During AIC or purge function, it is not possible to change the current time setting.
Trying to change the setting is invalid, as the screen only shows “In AIC” or “In Purge.”
48
6 Other Functions
1. According to the message on the screen, input the password currently set.
The screen to input a new password will appear.
Note
The password at shipment is set to “0000.”
49
6 Other Functions
4. Input the new password again and press the [SET] key.
The input password is checked with the new password. When both of them accord each
other, the password is changed into a new one and the screen returns to the MENU 5/5
(SYSTEM menu) screen.
Note
When the input password does not accord with the new password, the password change is aborted
and the screen returns to the MENU 5/5 (SYSTEM menu) screen.
The password is not changed.
50
6 Other Functions
Note
The adjustment is given by positioning the touch panel response part closer to the center of .
It is not automatically correcting the position gap.
1. Pressing the [TOUCH PANEL ADJUST] button in the MENU 5/5 (SYSTEM menu)
screen will display the TOUCH PANEL ADJUSTMENT screen.
2. Touch the part on the upper left of screen, and search the position where
turns to .
3. Keep you finger on the position where turns to , and slide the finger to the
center of .
4. Leave the finger at the center of .
If turns to on the way to the center, leave the finger at the displayed
position nearest to the center.
5. Perform the adjustment in the same matter for the at the lower right of screen.
6. If it is hard to slide the finger to the center of , repeat the step 1. to 5. above.
51
6 Other Functions
Note
To use the key lock function, the password is required.
Set the password beforehand according to the " 6.5.3 Password setting " (page 49).
Key lock
2. Press the [Key Lock] button from the icon area in the upper of screen.
The KEY LOCK screen appears.
52
6 Other Functions
4. Change the setting in each setting screen, and press [SET] key.
The setting is changed and the screen returns to the KEY LOCK screen.
5. Press the [EXIT] key in the KEY LOCK screen.
The screen returns to the MEAS. screen.
KEY LOCK
Press the [KEY LOCK] button. The system is allowed to be KEY LOCK state immediately.
KEY UNLOCK
Press the [KEY UNLOCK] button. The PASSWORD screen appears.
Note
The password at shipment is set to “0000.”
53
6 Other Functions
Reference
For checking the alarm history, refer to " 6.2.4 Alarm history check " (page 40).
Reference
For detailed information of each alarm, refer to " 8.1 Alarm Type and Countermeasure " (page 76).
54
7 Maintenance
7 Maintenance
To use the exhaust gas analysis instrument for a long period of time in a good condition,
maintenance and inspection are always necessary.
The maintenance interval depends on the using condition as well as on the sample gas
property. In the beginning of the operation, follow the recommended interval shown below so
that it works also for the pre-maintenance.
After that, economical maintenance/inspection works will be discussed according to the actual
operating condition as necessary.
55
7 Maintenance
Analyzer
Check item Contents of inspection Reference
Sample gas flow Confirm a flow rate of 0.6 L/min ±0.1 L/min
Existence of alarm generation Confirm that the alarm is not generated. page 54
Sampling system
Check item Contents of inspection Reference
Secondary filter 1, 2 Check for dirt on surface. If dirty, replace it. page 63
Confirm that the temperature of electric cooler is 0C to 8C
(under 15C if the temperature inside the cabinet is over 40C) in
Electronic cooler page 36
the ANALOG INPUT 2 screen (Fig. 31 page 36).
Also confirm the fan rotation.
Replace the primary filter element if the water causes bubbling. page 62
If the water level is low, perform watering until the water starts
Drain trap 1 being discharged from the drain trap 2.
page 8
(The water level might be low by drying the atmosphere in
winter.)
Mist catcher Check for dirt. If dirty, replace it. page 65
If the residual pressure is below approximately 1 MPa, replace
Calibration gas page 19
the gas.
Reference
For replacement and cleaning of each part, refer to " 7.4 Maintenance of Sampling Devices " (page 62).
56
7 Maintenance
Sampling point
Check item Inspection cycle Content of inspection and action Reference
Filter element 1 month Replace
Primary filter Holder cap 3 months Replace page 62
O-ring 6 months Replace after applying Daifloil grease
If clogging is found, clean it by instrumental air. (Be
Heated tube (piping) 3 months
sure to clean it with the inlet tube disconnected.)
Probe 3 months Clean the clogging and dirt
Sampling system
Check item Inspection cycle Content of inspection and action Reference
Secondary filter 1, 2 1 month Replace page 63
Halogen scrubber Discoloration check
1 month page 64
(waste incinerator specification) Replace if discolored
Mist catcher 2 months Replace page 64
Pump1, 2 6 months Replace the diaphragm page 66
Electronic cooler 6 months Clean the radiating fin and fan
Cabinet
Check item Inspection cycle Content of inspection and action Reference
Clean if dirty
Ventilation filter 3 months page 67
Replace if deteriorated
Reference
For replacement and cleaning of each part, refer to " 7.4 Maintenance of Sampling Devices " (page 62).
57
7 Maintenance
CAUTION
The zero gas purifier is heated up to 190C. Be careful during replacement work.
CAUTION
The arrestor and the noise filter are applied with the voltage even when the main
power ELB is not charged. At inspection, do not touch the terminal part.
CAUTION
Analyzer
Check item Contents of inspection Reference
Perform the calibration and confirm that the zero and
Calibration span calibration are correctly performed (no alarm or
caution occurs).
Note
For the analyzer inspection, contact HORIBA.
Sampling system
Check item Contents of inspection Reference
NOx converter (for NOx analyzer) Replace (catalyst element) page 68
Scrubber Replace page 64
Zero gas purifier (for CO analyzer) Replace (catalyst element) page 69
Air filter Replace page 70
Silica gel (for O2 analyzer) Replace
page 71
Protective filter (for O2 analyzer) Replace
58
7 Maintenance
Cabinet
Check item Contents of inspection Reference
Measure the insulation resistance with the arrestor
Arrestor individually, and replace if the resistance is below
5 M at 500 V megger.
Checking Arrestor
In checking the arrestor, the breaker at the user side (or the leak breaker) that
supplies power to the analyzer needs to be OFF.
CAUTION This is a hazardous operation; contact HORIBA.
Reference
For replacement and cleaning of each part, refer to " 7.4 Maintenance of Sampling Devices " (page 62).
Note
The replacing interval is a typical example and not the warranty period of each part.
The interval may be shortened depending on the installation environment or measurement gas
property.
Recommended
Category Part Name Replacing Interval Remarks
3 years 5 years
Probe tube
Sample gas
Primary filter
collection point
Heater
Ventilation fan
Cabinet Door packing
peripherals Panel heater
Panel heater fan
Replace when broken
Leak breaker 30 A
Circuit protector 15 A
Circuit protector 7.5 A
Circuit protector 5 A
Circuit protector 3 A
Relay G2R-1-T
Panel switch
Solid relay G3NE-205T
With Electrolytic
Power supply ZWS50AF-24/J
condenser
Noise filter Replace when broken
Printed board EN-DAC-03A With Electrolytic
Printed board EN-PIO-02A condenser
59
7 Maintenance
Recommended
Category Part Name Replacing Interval Remarks
3 years 5 years
Electric cooler
Electric cooler power unit
Drain trap 1
Drain trap 2
Catalyst tube heating case
Secondary filter
Needle valve
Pump
Buffer tank
Sampling part Replace when broken
3-way solenoid valve
2-way solenoid valve
Air filter cap
Air filter pot
Silica gel cap
Regulator
Regulator O-ring
Regulator spectacle O-ring
Flowmeter
Power ZWS75AF-5/J With Electrolytic
Power ZWS150AF-24/J condenser
Solid relay G3R-ODX02SN Replace when broken
Analyzer casing With LCD touch panel With Backlight
Printed board AP-TMC-02
With Electrolytic
Printed board AP-CPU-02
condenser
Printed board EN-ADC-02A
60
7 Maintenance
Recommended
Category Part Name Replacing Interval Remarks
3 years 5 years
Light source
Cell (NDIR)
Heater
Cell (for CO2)
Taper block
Beam splitter
Replace when broken
Infrared analyzer Gas cell
Detector (CO2)
Detector (NO)
Detector (SO2)
Detector (CO)
Printed board AP-PRE-10
Solenoid valve Replace
Capillary SA-152-QJ
Detector (O2)
Printed board EN-PRE-02
Cell (O2) Replace when broken
O2 analyzer
Exciting coil
Capillary SA-131-QJ
Capillary SA-090-KJ
61
7 Maintenance
High Temperature
The primary filter is in high temperature of approximately 120C.
At filter replacement, be sure to wear the gloves to avoid heated injuries.
CAUTION
Element cap
(3200044271)
Filter element
(3200044236)
Holder cap
(3200044240)
O-ring
(3200043716)
Holder
Clamp
62
7 Maintenance
Pull
Note
For element, use PA-5L filter when the SO2 analyzer is included, or use GC-90 filter when the SO2
analyzer is not included.
Attach the F-packing with the correct orientation facing the surface with groove to the element.
F-packing
(3200044273)
O-ring
(3200043785)
Groove
Element
(Without SO2 analyzer (GC-90):3200043784
With SO2 analyzer (PA-5L):3200043874)
63
7 Maintenance
Note
Attach the scrubber with the correct orientation so that the arrow mark faces downward.
Joint
Scrubber
(3200092415)
Pull off
Joint
64
7 Maintenance
Note
Attach the mist catcher with the correct orientation so that the arrow mark faces downward.
Joint
Mist catcher
(3014006594)
Pull off
Joint
Fig. 56 Replacing the mist catcher
65
7 Maintenance
High Temperature
The pump surface is in a high temperature. Replace the diaphragm at earliest after
one hour from power OFF.
CAUTION
Note
When the temperature is low, it may take time to obtain a stable pump flow.
Diaphragm
(3200043977)
Screw 1 (4)
Oscillation piece
Cover
Transformer
Screw 3 (4)
Screw 2 (4)
66
7 Maintenance
Metal fitting
Wing nut
Filter
(3200082233)
Stop frame
Fig. 58 Cabinet ventilation filter
Note
For the heater inside the panel, attach the cover (included in the accessory) at the ventilation inlet in
winter.
67
7 Maintenance
High Temperature
The catalyst tubes as well as joints are in high temperature. Replace the catalyst
tubes at earliest after one hour from power OFF.
CAUTION
Note
Attach the catalyst tube with the correct orientation so that the arrow mark faces downward.
Joint 2
Catalyst tube
(3014038044)
Joint 1
(3012123572)
68
7 Maintenance
High Temperature
The catalyst tubes as well as joints are in high temperature. Replace the catalyst
tubes at earliest after one hour from power OFF.
CAUTION
Note
Attach the catalyst tube with the correct orientation so that the arrow mark faces downward.
Joint 2
Catalyst tube
(3014038045)
Joint 1
(3012176183)
69
7 Maintenance
Cap
Air filter
(3200044378)
Fixing plate
Pot
70
7 Maintenance
Note
In inserting the tube into the silica gel, be careful not to break the tube.
Relay joint 1
Joint
Protective filter
(3014038043)
Relay joint 2
Cap
Tube
Silica gel
(3200044017)
71
7 Maintenance
Note
Attach the scrubber with the correct orientation so that the arrow mark faces downward.
Joint
Halogen scrubber
(3014038046)
Pull off
Joint
Fig. 63 Replacing the halogen scrubber
72
7 Maintenance
Note
Since the gas remaining in the pressure regulator is discharged when removing the cylinder, provide
adequate ventilation. When using poisonous gas, use protective gear such as a gas mask.
Note
Take care so that the cylinder cannot fall over.
Note
In parts replacement, it is recommended that the gasket with ring to be replaced into the packing
without ring.
The packing can be used only once. In replacing the cylinder, use the new packing.
The old packing, if continuously used, causes irregularity in sealing surface which easily attracts
metal powder to deteriorate the sealing performance.
The typical replacing interval of gasket with ring is after 4 to 5 times of tightening.
In replacing it, remove a foreign material adhered to the packing or the gasket with ring if any.
73
7 Maintenance
7. Tighten the mounting nut with hands and give it 1/4 or 1/3 turn using the wrench
provided.
74
7 Maintenance
Reference
For introducing gas to the analyzer, refer to " 5.4 Manual Calibration " (page 28).
75
8 Troubleshooting
8 Troubleshooting
8.1 Alarm Type and Countermeasure
Item Description Check item Countermeasure Reference Contact output
Indicates that the power Analyzer alarm
Power supply
is charged. (at power OFF)
Indicates that the in-
In
maintenance switch is In maintenance
Maintenance
ON.
Indicates that the
battery for time saving
Battery Contact HORIBA. Analyzer caution
at power OFF is being
consumed.
Indicates that the
temperature
Temperature compensation sensor
sensor alarm for NDIR detector has Contact HORIBA. Analyzer alarm
for NDIR * abnormalities such as
disconnection or short
circuit.
Indicates that the
NDIR
temperature
Temperature
compensation sensor Contact HORIBA. Analyzer caution
Sensor
for NDIR detector is
Caution *
being abnormal.
Indicates that the
temperature
MPA
compensation sensor
Temperature
for MPA detector has Contact HORIBA. Analyzer alarm
Sensor Alarm
abnormalities such as
*
sensor breakage or
short circuit.
Indicates that the
MPA
temperature
Temperature
compensation sensor Contact HORIBA. Analyzer caution
Sensor
for MPA detector is
Caution *
being abnormal.
Indicates that the
temperature adjustment
function for NDIR
NDIR Cell sample cell has
Temperature abnormalities such as Contact HORIBA. Analyzer alarm
Alarm * heater disconnection,
temperature sensor
disconnection, or short
circuit.
Indicates that the
NDIR Cell temperature adjustment
Temperature function for NDIR Contact HORIBA. Analyzer caution
Caution * sample cell is being
abnormal.
76
8 Troubleshooting
77
8 Troubleshooting
78
8 Troubleshooting
Note
The alarm/caution messages marked by “*” are not output as analyzer alarm/analyzer caution
contact output when they occur within 1 hour from power ON.
They are not recorded on the alarm history. Only the presence history of alarms/cautions can be
confirmed from the screen. Refer to " 6.7 Alarm Check " (page 54).
Analyzer Alarm (contact output):
Indicates that the analyzer is in the state disabling the measurement or the state hard to maintain
the measurement accuracy. Handle it as soon as possible.
Analyzer Caution (contact output):
Indicates that the measurement accuracy is maintained but starting to be hard to be maintained.
Confirm the situation and prepare for the countermeasures.
79
8 Troubleshooting
8.2 Troubleshooting
This chapter describes the troubleshooting mainly for the parts replacement advices and
check items for customers.
If these counteractions do not solve the problems, be sure to contact HORIBA.
Note
Before trying the counteractions, check the followings again:
Power is connected
Power voltage/capacity meet the specification
Parts replacement is correctly performed
No indication
Cause Check item Counteraction Reference
page 16
Back light is extinguished Touch the screen.
page 47
Analyzer power is OFF Confirm the power switch. Turn the analyzer power ON. page 3
Others Contact HORIBA.
80
8 Troubleshooting
81
8 Troubleshooting
82
9 Technical Data
9 Technical Data
9.1 Measurement Principle
9.1.1 Measurement principle using cross-modulation type NDIR
The molecular consisting of different atoms absorbs infrared ray of specific wavelength.
The cross-modulation type non-dispersive infrared analyzer detects the gap of infrared
absorption caused by alternatively introducing the sample gas and the reference gas by the
certain period to the measurement cell using the condenser microphone detector.
This method is free from zero drift. As for maintenance, the optical adjustments are not
required because it does not apply a rotation sector, which is quite effective for the stack gas
analyzers requiring continuous operation and high measurement accuracy.
Dividing the infrared ray from light source into two categories: reflected light and transmitted
light using a beam splitter allows the optical layout of as much as 4 components, realizing the
compact, downsizing module.
Subtraction
SO2 detector Measured value of SO2
*1 MAIN
Light source Subtraction
Subtraction
NO detector Measured value of NOx
Measurement
cell
*1 COMP
CO2 cell
CO2
sensor
CO detector MAIN
COMP Subtraction
Measured value of CO
Sample gas
Solenoid valve
Reference gas
83
9 Technical Data
P = H2 • X • C
H: Magnetic field strength
X: Magnetization factor of paramagnetic component (O2)
C: Concentration difference between paramagnetic component (O2)
and carrier gas (O2)
The pressure increase at this time is taken out from the magnetic field using a non-paramag-
netic gas such as nitrogen. It is detected by a capacitor microphone detector and then con-
verted into electric signals.
Carrier gas uses air, which does not require any cylinder gas.
The magnetic field uses electromagnet of AC drive, and the signal is processed as AC
signals, contributing to a stable measurement value.
Sample gas
Magnet
Measured value of O2
84
9 Technical Data
9.2 Specifications
Model ENDA-5000
Measurement target NOx SO2 CO *1 CO2 O2 *2
Measuring principle NDIR NDIR NDIR NDIR MPA
200 ppm 200 ppm 200 ppm 5 vol% 10 vol%
Standard
Range to 5000 ppm to 5000 ppm to 5000 ppm to 25 vol% to 25 vol%
Option 100 ppm or more 50 ppm or more 100 ppm or more
Range ratio Within 10 times Within 10 times Within 10 times Within 5 times Within 2.5 times
Repeating accuracy ±0.5% of full-scale value
(Repeatability) (±1.0% of full-scale value when any optional range is included or for the O2 measurement)
Linearity
±1.0% of full-scale value
(Indication error)
±1.0% of full-scale value per week
Zero drift (within 5C in ambient temperature changes, except for air pressure interference)
(±2.0% of full-scale value when any optional range is included or for the O2 measurement)
±2.0% of full-scale value per week
Span drift
(within 5C in ambient temperature changes, except for air pressure interference)
Td+T90= 60 s maximum at system inlet (at Sample flow of 0.6 L/min)
Response speed
(240 s maximum for the SO2 measurement only)
±2.0% of full-scale value
Interference
(within the composition of standard gas in the standard range)
Display Touch panel LCD screen (with back light)
Ambient
5C to 40C without direct sunlight and radiant heat *3
temperature
Installation Ambient 90% maximum
conditions humidity
Vibration 100 Hz, 0.3 m/s2 maximum
Dust Less than environmental standard
Temperature 250C maximum
Dust 0.1 g/Nm3 maximum
NO: 500 ppm maximum
Sample NO2: 15 ppm maximum
Standard
conditions SO2: 1000 ppm maximum
gas
SO3: 50 ppm maximum
composition
*4 CO: 200 ppm maximum
CO2: 15 vol% maximum
H2O: 20 vol% maximum
Sampling method Dehumidified sampling at 5C using the electronic cooler
Flow rate of sample gas 2.5 L/min to 3.0 L/min
Sample inlet tube Teflon tube (8 mm O.D./6 mm I.D.)
(1) 1.96 kPa to 4.9 kPa
±4.9 kPa (3-point selection method)
Pressure of sample gas (2) ±3.43 kPa
(Without sample gas backpressure)
(3) 4.9 kPa to 1.96 kPa
Normal pressure control by regulator and pump, depressurized sampling
Pressure control method
Control pressure: 4.9 kPa
4 mA to 20 mA DC (Insulation output)
Concentration output (0 mA to 16 mA DC/0 V to 1 V DC/1 V to 5 V DC is optional)
Max. 12 systems output available
Analyzer alarm, analyzer caution, range display, in-calibration, in-maintenance, in-purge
External contact output (optional)
Contact capacity: 30 V DC, 1 A, 250 V AC, 1 A in resistance load
85
9 Technical Data
DRY calibration
Automatic calibration
Calibration method
(Calibration interval: 7 days (standard), setting available from 1 to 99 days)
Manual calibration
Zero gas N2 or ambient air
Carrier gas for
Calibration gas O2 Ambient air
measurement
Span gas Gas cylinder for measured component
Flange: JIS 10 K, 40 AFF
Probe tube length: 1000 mm
Probe
Material: SUS-316
Filter at sampling point
Element: SUS-304 + quartz wool bellows of 2 µm in thickness
Electric heating: 100 VA with drip-proof case
Power voltage 100 V ±15 V AC
Power frequency 50/60 Hz switchable
Approximately 800 VA (Including primary filter and service plug outlet)
Power consumption
Heated tube 30 m (+1100 VA) and panel heater (+300 VA)
600 (W) mm x 1770 (H) mm x 300 (D) mm
(High-pressure gas cylinder, maximum 3 gas cylinders of 3.4 L are contained.)
Dimensions Approximately 180 kg (Without cylinders)
Mass 600 (W) mm x 1770 (H) mm x 500 (D) mm
(High-pressure gas cylinder, maximum 6 gas cylinders of 3.4 L are contained.)
Approximately 200 kg (Without cylinders)
SUS-316, SUS-304, Teflon, polypropylene, polyethylene, fluoro-rubber, PVC, PVDF, and
Materials exposed to gas
glass
Standalone type for outdoor installation
Plate thickness: 2.3 mm for steel plates of main unit, door. and top plate,
Cabinet 3.2 mm for channel base
Door: Front
Connections: Right side
Color Munsell 5Y7/1 semi-gloss for inner and outer surfaces
*1: The analyzer against N2O interference for CO analyzer applies the standard range of 200 ppm or more
(no optional range).
*2: No carrier gas cylinder is necessary.
*3: For the temperature range of 15C to 40C (cold district) and for the specification of 5C to 50C, we will
separately discuss the design.
*4: When the coexisting gas contains NH3, NH3 scrubber is prepared for an optional part.
When CH4 coexists in the sample gas for SO2 analyzer, the SO2 analyzer of CH4 interference compensation type is
prepared.
When N2O coexists in the sample gas for CO analyzer, the CO analyzer of N2O interference compensation type is
prepared.
86
9 Technical Data
87
9 Technical Data
88
9 Technical Data
28
Daifloil grease
1 1g
(DG-203) 2
5
50
89
9 Technical Data
Note
At test operation and maintenance, the accessories can be used by HORIBA service person.
50
3200044236
Primary filter element 1 month 12
98 (12 pcs)
60
3200043716
Primary filter O-ring 6 months 2
(1 pc)
for SE3
21
3200044240
Primary filter holder cap 3 months 4
10
(10 pcs)
48
3200044271
Primary filter element cap 3 months 4
(10 pcs)
14
90
9 Technical Data
70
3200044260
Primary filter element 1 month 12
170 (1 pc)
75
3200043880
Primary filter O-ring 6 months 2
(1 pc)
for LE3
39
3200044258
Primary filter holder cap 3 months 4
15
(1 pc)
55
3200044259
Primary filter element cap 3 months 4
12
(1 pc)
62
55
55
248
Converter catalyst tube 3014038044
1 year 1 With NOx analyzer
(COM-50) (1 pc)
14
91
9 Technical Data
Scrubber 193
3200092415 For other cases than
1 year 1
(ESU-050A) (1 pc) above
23
70 45
90
With O2 analyzer or
3200044017
Silica gel 1 year 1 for the specification
180 (1 pc)
using air as zero gas
193
3014038046 Depending on the
Halogen scrubber 6 months 2
(1 pc) sample gas condition
23
92
10 Optional Functions
10 Optional Functions
10.1 Purge Functions
When there is a large amount of dusts in the stack gas, the purge function is effective:
the separate blowback instrument enables the backflush of the dusts collected by the primary
filter using the purge air.
According to the purge time setting, the periodical purge is available.
Solenoid valve
Solenoid valve
F P
To analyzer
Sample line
Primary filter
93
10 Optional Functions
94
10 Optional Functions
3. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 96 to page 100.
4. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the PURGE SET screen.
Tip
To cancel the change, press the [CANCEL] key.
The change is canceled and the screen returns to the PURGE SET screen.
Note
Trying to change the purge setting during the purge is invalid. The screen only shows “IN PURGE.”
95
10 Optional Functions
Purge mode
Sets the purge start mode.
1. Pressing the button at the right of [MODE] in the PURGE SET screen displays the
PURGE MODE screen.
2. Press the button for the item to be set and press the [SET] key.
The setting is changed and the screen returns to the PURGE SET screen.
96
10 Optional Functions
Start time
Sets the next purge start time.
1. Pressing the button at the right of [START] in the PURGE SET screen displays the
PURGE START screen.
Year Minute
Hour
Month
Day
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by the [] and [] button and press the [SET] key.
The setting is changed and the screen returns to the PURGE SET screen.
Note
The start time setting is based on the internal clock.
It is not available to change into an impossible date.
When this happens, the existing date closest to that before the set date is automatically selected.
When the start time previous to the current time is set, the setting value of start time is changed
into the minimum value which comes in future by adding multiple number of the interval set
values.
When the start time set value becomes a previous time by changing the internal clock setting
(refer to " 6.5.2 Time adjust " (page 48)), the start time is changed into the minimum value which
comes in future by adding multiple number to the current time set value.
When the AIC start time is set later than 2090/1/1/0:00, the AIC does not start.
When the “Battery” alarm occurs, confirm the setting time.
97
10 Optional Functions
Interval
Sets the time span (interval) for the purge start.
1. Pressing the button at the right of [INTERVAL] in the PURGE SET screen displays
the PURGE INTERVAL screen.
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by the [] and [] button and press the [SET] key.
The setting is changed, and the screen returns to the PURGE SET screen.
Note
The recommended purge interval is 12 hours. When the above recommended time does not meet the
actual condition, change the setting after confirmation.
98
10 Optional Functions
In blow time
Set the IN BLOW TIME in purge.
1. Pressing the button at the right of [IN BLOW TIME] in the PURGE SET screen
displays the PURGE IN BLOW TIME screen.
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by the [] and [] button and press the [SET] key.
The setting is changed, and the screen returns to the PURGE SET screen.
Note
The recommended purge interval is 1 minute. When the above recommended time does not meet the
actual condition, change the setting after confirmation.
99
10 Optional Functions
Delay time
Sets the purge delay time.
1. Pressing the button at the right of [DELAY TIME] in the PURGE SET screen displays
the PURGE DELAY TIME screen.
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by [] and [] button and press the [SET] key.
The setting is changed, and the screen returns to the PURGE SET screen.
Note
The purge delay time depends on the tube length. Set the time by actually measure the tube length.
100
10 Optional Functions
Note
During the purge, the concentration output is held.
Manual start
1. Press the [MENU] key in the MEAS. screen.
Note
When the purge starts, the [PURGE START] button changes into the [PURGE STOP] button.
Pressing the [PURGE STOP] button interrupts the purge.
During the purge interruption, the [PURGE STOP] button changes into the [PURGE START] button.
Note
The purge by manual start via the [PURGE START] button is available except for during the AIC.
When the purge command is input by the internal clock or via the external contact input during the
purge of manual start mode, the signal is not received but the purge by the manual purge continues.
101
10 Optional Functions
10.2 O2 Correction
The stack gas from the fixed discharge source, when measured, gives a diluted, low
concentration value due to the air mixture amid the stack path.
This is why when measuring the nitrogen oxides from the fixed discharge source, the suitable
correction according to the oxygen concentration in the discharged gas is included in the
enforcement regulations of Clean Air Act.
For the measurement of carbon monoxide correction for industrial waste incinerator, too, the
correction according to the oxygen concentration is required.
O2 set value (vol%) is a concentration set depending on the fuel facility as well as fuel type.
Two types of O2 values can be set.
102
10 Optional Functions
3. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 104 to page 107.
4. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the O2 CORRECTION SET screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and the screen returns to the
O2 CORRECTION SET screen.
103
10 Optional Functions
Mode
Sets the O2 correction mode.
Pressing the button at the right of [MODE] in the O2 CORRECTION SET screen displays
the O2 CORRECTION MODE screen.
Pressing the button of the item to be set, and press the [SET] key.
The setting is changed, and the screen returns to the O2 CORRECTION SET screen.
104
10 Optional Functions
O2 SET VALUE 1
Pressing the button at the right of [O2 SET VALUE 1] in the O2 CORRECTION SET
screen displays the O2 SET VALUE 1 screen.
105
10 Optional Functions
O2 SET VALUE 2
Pressing the button at the right of [O2 SET VALUE 2] in the O2 CORRECTION SET
screen displays the O2 SET VALUE 2 screen.
106
10 Optional Functions
O2 Limit value
Pressing the button at the right of [O2 LIMIT VALUE] in the O2 CORRECTION SET screen
displays the O2 LIMIT VALUE screen.
107
10 Optional Functions
108
10 Optional Functions
3. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 110 to page 111.
4. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the ROLLING AVERAGE TIME SET
screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and screen returns to the
ROLLING AVERAGE TIME SET screen.
5. Press the [EXIT] key in the ROLLING AVERAGE TIME SET screen.
The screen returns to the MENU 4/5 (OPTION menu) screen.
6. Press the [EXIT] key in the MENU 4/5 (OPTION menu) screen.
The screen returns to the MEAS. screen.
109
10 Optional Functions
Time 1
Pressing the button at the right of [TIME 1] in the ROLLING AVERAGE TIME SET screen
displays the ROLLING AVERAGE TIME SET 1 screen.
Note
Setting the ROLLING AVERAGE TIME 1 more than 60 minutes automatically sets the time to 60
minutes.
110
10 Optional Functions
Time 2
Pressing the button at the right of [TIME 2] in the ROLLING AVERAGE TIME SET screen
displays the ROLLING AVERAGE TIME SET 2 screen.
Note
Setting the ROLLING AVERAGE TIME 2 more than 240 minutes automatically sets the time to 240
minutes.
Note
In purge, in calibration, and in maintenance (with the output hold specification), the rolling
average time as well as the integration time are accumulated and calculated just before they are
held.
In changing the rolling average time, the data before the change is temporarily reset, and the time
newly changed is used for the calculation.
111
10 Optional Functions
NO The screen returns to the MENU 1/5 (OPERATION menu) screen without the data
reset.
YES Resets the data and the screen returns to the MENU 1/5 (OPERATION menu) screen.
112
10 Optional Functions
113
10 Optional Functions
3. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 114 to page 116.
4. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the MEASURE VALUE ALARM screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and screen returns to the
MEASURE VALUE ALARM screen.
Note
The following state stops the alarm judgment and releases an alarm in occurrence.
・5 minutes from power ON ・In maintenance
The following state maintains the condition just before the alarm occurrence.
・In AIC (In introducing the calibration gas) ・In purge
Component
Pressing the button showing the component to be set in the MEASURE VALUE ALARM
screen displays the COMPONENT screen.
Press the button of item to be set and press the [SET] key.
114
10 Optional Functions
The setting is changed and the screen returns to the MEASURE VALUE ALARM screen.
To set the integration value or rolling time average, press the [INT.] key.
Press the button of item to be set as MEASURE VALUE ALARM and press the [SET] key.
The setting is changed and the screen returns to the MEASURE VALUE ALARM screen.
Alarm type
Pressing the button showing the alarm type in the MEASURE VALUE ALARM screen
displays the ALARM TYPE screen.
Press the button of item to be set and press the [SET] key.
The setting is changed and the screen returns to the MEASURE VALUE ALARM screen.
115
10 Optional Functions
Alarm value
Pressing the [VALUE] button in the MEASURE VALUE ALARM screen displays the
ALARM VALUE screen.
Note
In the state without selecting COMPONENT cannot set TYPE or ALARM VALUE.
Even without setting the H.HIGH alarm, the H.HIGH alarm can be used.
The H.HIGH alarm and HIGH alarm are independent each other.
At input these alarm values, set so that H.HIGH value is more than HIGH value.
116
10 Optional Functions
117
10 Optional Functions
3. Press the button corresponding to the item to be set. The setting screen appears.
For the detailed information on each setting screen, refer to page 118 to page 121.
4. Change the setting in each setting screen, and press the [SET] key.
The setting is changed and the screen returns to the CO PEAK COUNT ALARM screen.
Tip
To cancel the change, press the [CANCEL] key. The change is canceled and screen returns to the CO
PEAK COUNT ALARM screen.
Note
The following state stops the peak count function.
・Five (5) minutes from the power ON
・In AIC (In introducing the calibration gas)
・In purge
・In maintenance
Mode
Pressing the [MODE] button in the CO PEAK COUNT screen displays the CO COUNT
MODE screen.
118
10 Optional Functions
Note
Operation from [ON] to [OFF] resets the current count value, set count value and the peak value.
When CO PEAK COUNT ALARM is output, this operation releases the alarm.
Operation from [OFF] to [ON] starts counting time.
The peak count value counts time by the interval of 1 minute.
When 60 minutes are elapsed, the peak count for the first minute (60 minutes ago) is deleted, and
the counting continues.
Total
Peak count 2 times 1 time 0 time 0 time 1 time 1 time 1 time 6 times Alarm output
After 1 minute
Total
Peak count No count 1 time 0 time 0 time 1 time 1 time 1 time 0 time 4 times Alarm release
Tip
In changing the peak count value, to start the count by resetting the current count value, switch MODE
to the following way: ON to OFF to ON.
The current count value can be reset by this operation.
119
10 Optional Functions
Count value
Pressing the [COUNT] in the CO PEAK COUNT screen displays the COUNTER screen.
Note
When the CO PEAK COUNT mode is OFF, the count value setting is not available.
120
10 Optional Functions
Peak value
Pressing the [PEAK VALUE] in the CO PEAK COUNT screen displays the PEAK COUNT
VALUE screen.
121
2 Miyanohigashi, Kisshoin Minami-ku, Kyoto 601-8510 Japan
http://www.horiba.com